1Acu-Rite Companies Inc.
Page 4
Introduction / Supplied ItemsSENC 50
The SENC 50 Precision Glass Scale Linear Encoder provides
the accuracy and reliability of an Acu-Rite Companies Inc.
measuring system in a very small cross-section with digital
output (analog output available). Features and options include:
• Resolutions of 0.5, 1.0 or 5µm.
• Accuracy grades of ± 3 or ± 5 µm / 1000mm.
• Position Trac or 50mm interval Reference Mark.
• Vinyl or Armor cables of 5, 13, & 19 ft. lengths.
• Mounting fasteners, center supports, and backup spar.
• Installation tools, brackets, and accessories.
The SENC 50 is especially useful for applications with limited
space such as X/Y stages and compact metrology equipment.
Contact your Authorized Distributor for assistance with selection
of product options and accessories.
Shipping carton
contents
A
C
For future ordering information or warranty service, record the
linear encoder catalog and serial numbers.
Mounting options can be adapted to machine mounting surfaces
using spacers, standoffs, and leveling screws.
• Measuring length and mechanical configuration of your
equipment determines your options.
• Length related mounting options shown.
• When a backup spar is used, mount the backup spar following
the same procedure as required for mounting without a backup
spar, using the same tolerances.
End mounting holes
Center support
assembly
• 8” and larger: End mounting holes with center support.
• 7” and smaller: End mounting holes with encoder mounted to
a continuous flat surface, or with center surface on same plane
as end mounting surfaces.
Acu-Rite Companies Inc.6
Page 9
SENC 50Typical mounting
A variety of mounting conditions can be accommodated.
• Mechanical configuration of the equipment determines
brackets and adapters required to install the linear encoder.
• Three typical conditions are shown for reference.
Flush mounting surfaces
// .005
Spar mounting
// .030
• Flush or offset mounting surfaces using a spar.
• Reading head mounting bracket required (bottom mounting
bracket example shown).
Offset mounting surfaces
// .030
• Mounting surfaces are flush within .005”.
• Installation without a backup spar.
• Offset mounting surfaces using a reading head bracket.
7Acu-Rite Companies Inc.
Page 10
Encoder Installation ProcedureSENC 50
These steps apply to all mounting conditions. Although this
may not pictorially represent your application, your installation
procedure should follow these steps.
• Acu-Rite Companies Inc. Bracket Kit instructions
supercede this section.
• General steps for small X & Y stage.
• Adjust drill depths and fastener lengths as required.
• Contact your Authorized Distributor if assistance is required.
Keep head and
Scale case
Shipping
brackets (2)
C
L
Reading head
scale case
surfaces flush
Strain relief
and
cable assembly
Center mounting axis
C
Mark center of axis
• Center the axis.
• Mark the axis for easy return to center.
Underside of end caps to be
flush with parting line of
upper slide.
End cap
Align top of scale case to
within .015” of axis travel
Mark end mounting
L
Scale case
hole location
• Center the reading head on the scale case by sliding the head
and shipping brackets together along the scale case.
Acu-Rite Companies Inc.8
• Locate the linear encoder along the axis parting line.
• Mark location of one end mounting hole in the scale case.
Page 11
SENC 50Installation Procedure
Axis
travel
Align to within
.005”
to axis travel
Drill / tap for 6-32
(M3).
Use shipping brackets
to maintain .059
gap between
head and
scale case.
Drill / tap
for 6-32 (M3).
6-32 x 3/4” SHCS & M3
flat washer (M3 x 20mm)
• Drill / tap the first end mounting hole / Attach the scale case.
• Align to within .005” TIR. and drill / tap second end hole.
• Attach scale case & align to within .005” TIR. of the axis travel.
6-32 x 3/4” SHCS & M3 flat washer
(M3 x 20mm)
Retaining clip
Center support
• Use center supports when provided.
• Place at uniform intervals along the scale case.
C
L
6-32 x 3/4” SHCS & M3
flat washer (M3 x 20mm)
• Center the axis and mark the reading head mounting holes.
• Move the axis and drill / tap holes for 6-32 (M3).
• Attach head and align to scale case to within .005” TIR.
Slide bracket from reading head
and twist 45°
Shipping bracket
(2)
• Slide shipping brackets from reading head and twist to remove
from the scale case.
• Save the shipping brackets with the Reference Manual.
9Acu-Rite Companies Inc.
Page 12
Checking the InstallationSENC 50
Connecting
Provide slack
loops
These steps will confirm proper operation of your installation.
Counting tests check channel A and B output function and
Reference Mark tests check Reference Mark operation.
• Configure the readout for proper counting resolution.
• Move axis and compare the display to the movement.
• Repeat test for mechanical integrity and repeatability.
• Configure the readout for sensing reference mark.
• Move the reading head approximately 20mm.
Secure excess
cable
• Locate an indicator on scale end and zero out axis and
indicator.
• Move axis out and back over full travel and return to dial
• Route the cables with slack loops to allow for axis motion.
• Secure excess cable by fastening with clips or ties.
zero.
• Electronics should read zero ± 1 count.
• Attach the linear encoder connectors to the readout.
Counting Test
Readout
• Check function of the linear encoder.
Acu-Rite Companies Inc.10
Place indicator at end of
moving component (scale case
shown, or reading head).
Repeatability Test
Readout
• Zero the display and indicator.
• Move axis to the end of it’s travel, then return to dial zero.
• Readout should read zero ± 1 count.
Page 13
SENC 50Electrical Shielding
Connect a ground wire from the terminal on the back of the
readout to the machine or earth ground. Attach a ground wire
from the machine to a solid earth ground.
With the encoder attached to the machine and the cable
connected to the readout, check shielding by measuring
resistance between connector housing and scale unit. Desired
value: 1
ΩΩ
Ω max.
ΩΩ
11Acu-Rite Companies Inc.
Page 14
Trouble ShootingSENC 50
If you experience difficulties with your installation, there are
methods to analyze the difficulty. Operating difficulties can be
caused by either the linear encoders or the readout.
Checking the readout
Follow the steps below to determine if your difficulties are
associated with the readout:
• Insure that the linear encoder connectors are correctly seated.
• Swap linear encoder cables at the readout to see if the problem
is still shown in the same display.
• If the problem remains in the same display, the readout is in
error.
• If the problem follows the connection change, the linear
encoder may be in error.
• Follow this procedure for any intermittent problem
experienced.
If the readout is at fault, refer to “What to do” to arrange for
the parts necessary to repair your system. If the linear encoders
appear to be the cause, perform the next step.
Checking the Linear Encoders
Linear encoder difficulties can be caused by improper
installation, loose or misaligned bracketry, or a damaged / failed
linear encoder.
Follow the steps below to determine the cause of your system
difficulties:
• Confirm that your bracketry and installation does not interfere
with other machine structures through the length of the linear
encoder travel.
• Check for loose fasteners. If you find loose fasteners, first
confirm that the linear encoder is installed to the tolerances
specified and then retighten the fasteners as required.
• Confirm that the linear encoder is installed to the required
tolerances by checking the alignment tolerances specified on
Page 3 “Key Points”. If the installation does not meet the
tolerances, reinstall the linear encoder according to the
procedures in “Installation”.
• Perform a Repeatability Test as described on Page 10,
“Checking Your Installation”. If the linear encoder is
installed to the required tolerances, the bracketry and linear
encoder have been checked for interferences and loose
fasteners, and the linear encoder fails the repeatability test,
the linear encoder is likely at fault.
Do not attempt to repair the linear encoder. The units are only
serviceable by assembly replacement. Attempted repair can
permanently damage the unit and void the warranty.
What to do
If an Acu-Rite Companies Inc. linear encoder or readout is found
to be at fault, prior to removing the linear encoders or readout
contact your Authorized Distributor or OEM/OEI for repair
instructions.
Acu-Rite Companies Inc.12
Page 15
SENC 50Mechanical Specifications
Mechanical SpecificationsDigitalAnalog
Resolution0.5µm1µm5µm
Grating pitch20µm
Scale mediumLight transmission reflective off of chrome coated glassAccuracy (@ 20° C) / 1000mm± 3µm, ± 5µm
Max. slew speed (M/s)1
Force required to move< 0.5 (lbs)
reading head
Operating Environment
Temperature0° to 50° C
Relative Humidity20% to 95% (non-condensing)
Storage Environment
Temperature- 20° to 70° C
Humidity20% to 95% (non-condensing)
Differential current or 1 volt peak to peak output
2-8µApp or 1.2 V
pp
w/100 K Ohm load
Power Supply
5.1 ± 0.1 VDC @ 220 mA max.
15Acu-Rite Companies Inc.
5.0 ± 0.1VDC @ 75 mA max.
Page 18
Acu-Rite Companies Inc. WarrantySENC 50
ACU-RITE Companies, Inc. is proud to offer the 3-Year Hassle-Free Warranty for all readout systems, and precision glass scales.
This warranty will cover all of the ACU-RITE repair and replacement costs for any readout or precision glass scale returned
during the three (3) year warranty period. ACU-RITE will repair or replace the damaged components regardless of the product’s
condition absolutely free, no questions asked. This warranty covers both materials, and factory labor. In addition, authorized
ACU-RITE service representatives will provide service labor (field service) for a period of one (1) year at no charge. Notice of the
claimed defect must be received by ACU-RITE within the warranty period.
This warranty applies only to products and accessories installed and operated in accordance with this reference manual. ACURITE shall have no obligation, with respect to any defect, or other condition caused in whole or part by the customer’s incorrect
use, improper maintenance modification of the equipment, or by the repair, or maintenance of the product by any person except
those deemed qualified by ACU-RITE.
Responsibility for loss of operation, or diminished performance due to conditions beyond ACU-RITE’s control cannot be accepted
by ACU-RITE.
The foregoing warranty obligations are in lieu of all expressed or implied warranties. ACU-RITE Companies, Inc. shall not be
liable under any circumstances for consequential damages.
Acu-Rite Companies Inc.16
Page 19
Page 20
Acu-Rite Companies Inc.Acu-Rite Companies Inc.
Acu-Rite Companies Inc.
Acu-Rite Companies Inc.Acu-Rite Companies Inc.
IS ANIS AN
IS AN
IS ANIS AN
ISO 9001ISO 9001
ISO 9001
ISO 9001ISO 9001
CERTIFIEDCERTIFIED
CERTIFIED
CERTIFIEDCERTIFIED
MANUFMANUF