3B Scientific Protective Adapter User Manual

3B SCIENTIFIC
Protective Adapter, 2-pole 1009961 Protective Adapter, 3-pole 1009960
06/12 NF/ALF
®
PHYSICS
1 Connector socket: cathode
1 / 2 Connector sockets: heater filament
3 / 5 Connector sockets: heater filament
4 Connector socket cathode
1. Safety instructions
The protective adapter conforms to safety regu­lations for electrical measuring, control and labo­ratory equipment as specified in DIN EN 61010 Part 1. It is to be operated in dry rooms as ap­propriate for the use of electrical equipment.
Safe operation of the adapter is guaranteed as long as it is used as stipulated. However, there is no guarantee of safety if the adapter is used incorrectly or carelessly.
If there is any suspicion that the adapter can no longer be operated without risk (e.g. if visible damage is detected), the adapter must immedi­ately be withdrawn from use and secured in such a way as to prevent its inadvertent opera­tion.
Do not subject the adapter to high humidity, extreme temperatures or heavy shocks.
Only use safety experiment leads for con­necting circuits.
2. Description
The 2-pole or 3-pole protective adapters are for connecting the heater voltage for the D-series electron tubes via safety experiments leads and also provide protective circuitry to prevent ex­cess voltage across the cathode heating ele­ment causing damage to the fragile filament.
A built-in relay cuts out when the heater voltage reaches approximately 10.5 V DC or 8.5 V AC and turns back on when the heater voltage has fallen back into range again.
The pin marked with a minus sign is connected to the brown socket. On the 3-pole, the red socket is connected to a 2-mm pin.
2-pole or 3-pole adapters should be chosen depending on the tube in use (for a list of which tubes match which adapter, see section 4).
1
3. Technical data
r
f
5. Care and maintenance
Terminals: 4 mm safety sockets Cut off voltage: DC: 10.5 V approx.
AC: 8.5 V approx. Dimensions: 75 mm x 45 mm dia. approx. Weight: 70 g approx.
4. Operation
The protective adapters match the following tubes:
2-pole (1009961): Diode D 1000646 Triode D 1000647 Luminescence tube D 1000648 Maltese cross tube D 1000649 Perrin tube D 1000650 Electron deflection tube D 1000651 Gas triode D 1000653
3-pole (1009960): Electron diffraction tube D 1000655
Plug the adapter into the sockets of the tubes in use in such a way that the adapter pin marked with a minus sign (brown socket) plugs into the socket marked with a minus on the base of the tube.
Insert the tube into the D-series tube holder and turn it on.
Before cleaning the adapter, disconnect it
from the power supply.
Use a soft, damp cloth to clean it.
6. Disposal
The packaging should
be disposed of at local recycling points.
Should you need to
dispose of the adapte itself, never throw it away in normal domes­tic waste. Local regula­tions for the disposal o electrical equipment will apply.
Elwe Didactic GmbH Steinfelsstr. 5 08248 Klingenthal Germany www.elwedidactic.com
3B Scientific GmbH Rudorffweg 8 21031 Hamburg Germany www.3bscientific.com
Subject to technical amendments
© Copyright 2012 3B Scientific GmbH