YSI EXO PAR User Manual

Page 1
EXO PAR
PHOTOSYNTHETICALLY ACTIVE RADIATION WITH THE EXO PLATFORM
USER MANUAL
XA00014
Page 2
EXO PAR
The information contained in this manual is subject to change without notice.
Effort has been made to make the information in this manual complete, accurate, and current.
The manufacturer shall not be held responsible for errors or omissions in this manual.
Thank you for purchasing a YSI EXO PAR System. This manual covers setup, operation, and functionality of the EXO PAR system.
EXO PAR features include:
• Simple setup and operation
• Depth rating of 250 meters
• Single or dual sensor option
• Non-corrosive materials
• Data output in µmol s-1 m-2
• Use with Kor or EXO handheld
• Easily removed for non-par applications
Safety Information
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all precautionary statements. Failure to do so could result in serious injury to the operator or damage to the equipment. Do not use or install this equipment in any manner other than that specified in this manual.
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
Precautionary Symbols
NOTE: Information that requires special emphasis
NOTICE: Indicates a situation which, if not avoided, may cause damage to the instrument
CAUTION: Indicates a potentially hazardous situation that may result in minor or moderate injury
WARNING: Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death
or serious injury
Product Components
Carefully unpack the instrument and accessories and inspect for damage. If any parts or materials are damaged, contact YSI ISS at 727-565-2201 or authorized YSI distributor.
Page 3
TABLE OF CONTENTS
1. Introduction
1.1 PAR Sensor
1.2 Overview
1.3 Components
2. Installation
2.1 Attach EXO Calibration Cup Extender
2.2 Attach the EXO PAR System Frame
3. Setup
3.1 Setup the Kor Software
3.2 Setup the EXO Handheld Display
4. Ordering
4.1 Ordering Replacement Components
4.2 Ordering
THIS IS AN INTERACTIVE DOCUMENT
When viewing this document as an AdobeTM PDF, hovering your cursor over certain phrases will bring up the finger-point icon. Clicking elements of the Table of Contents, website URLs, or references to certain sections will take you automatically to those locations.
Page 4
1. Introduction
1.1
Some users of the YSI EXO sonde may wish to incorporate a photosynthetically active radiation (PAR) sensor into their field monitoring equipment. This sensor can be added to the YSI EXO2 sonde in the form of a special adapter engineered by YSI’s Integrated System & Services division. This appendix is designed to give potential users of this PAR adapter information on how the system is configured and the steps necessary to acquire and log PAR data with the EXO PAR adapter.
PAR Sensor
TM
Figure 1 EXO PAR adapter with Sonde
1.2
Li-Cor PAR sensors are required for the EXO PAR adapter and these sensors can either be supplied to YSI by the customer (brass mounting screws provided with the sensor are required for installing on the par system) or can be purchased from YSI as part of the complete system. These sensors are then mounted to arms which extend from the sonde on either side. Output cables are then attached to the sensors which connect to the EXO PAR adapter, which in turn connects to the EXO2 Auxiliary port on the top of the sonde, as show in Figure 1.
Figure 1 shows the installation of two sensors (one pointing up, the other pointing down), but it is also possible to use a single PAR sensor in either an upward or downward looking configuration.
Your EXO PAR system is shipped separate from your EXO2 sonde and can be installed onto the EXO2 sonde following these easy steps.
4
Overview
Introduction
Page 5
1.3
TM
TM
The EXO PAR Systems consists of 3 main parts: The EXO PAR adapter cylinder and cables, the black plastic frame system and calibration cup extender, and the Li-Cor PAR sensors.
The EXO PAR system attaches to the EXO2 sonde via an upper and lower clamp system. The EXO PAR adapter attaches to the AUX port on the top bulkhead of the EXO2 sonde. For shipment and storage, the EXO PAR system can fold its sensor and support arms in, along the body of the EXO2 sonde (figure 2).
Components
TM
Figure 2 The EXO PAR system attaches to the EXO2 sonde via an upper and lower clamp system
Introduction
5
Page 6
2. Installation
2.1
Once you’ve unpacked the EXO PAR system, attach the EXO calibration cup extender to the bottom of your calibration cup. It simply clicks on by inserting the bottom of the calibration cup into the top of the extender and pushing down on a hard surface (figure 3).
The calibration cup extender provides you the extra height you will need to prevent your PAR sensors and cables from impacting the ground. It’s ideal for in lab use but we recommend using a lab stand as well for stability, to prevent your sonde and PAR system from tipping over and potentially causing damage.
CAUTION: Potential pinch hazard, be mindful of fingers.
Attach EXO Calibration Cup Extender
Figure 3 Push down to mate the extender onto
the calibration cup.
2.2
Now that your calibration cup extender is installed, stand the sonde in an upright position and slide the EXO PAR system frame over the EXO2 sonde. The two clamps; upper and lower are fastened to the EXO2 sonde body via two, 5/16-18 x 2.75” socket cap bolts. You will need to use a ¼” Allen wrench to secure these two bolts. Tighten the bolts until the clamp fits snug to the body of the sonde. Proper alignment of these clamps, as described below, should be followed as to not damage either system.
NOTE: Install the EXO PAR system with the EXO PAR adapter cylinder on the same side of the sonde where the AUX port connector is located. You will connect the EXO PAR adapter into this AUX port.
6
Attach the EXO PAR System Frame
Figure 4 The extender merged with the
calibration cup.
Installation
Page 7
1. The top of the upper clamp should align with the yellow bulkhead of the sonde (where the blue meets the yellow).
Proper alignment is necessary so that the PAR sensor arms are deployed at the correct angle
NOTICE: Proper alignment of clamps is necessary to prevent damage.
Figure 5 Align the upper clamps with the yellow bulkhead of the sonde
2. The bottom of the lower clamp should align with the upper edge of the top chamfer on the clear plastic section
covered with the black “EXO²” sonde label. Again proper alignment is necessary so the PAR sensor arms are deployed so that the PAR sensor faces axis is parallel to the sonde axis.
NOTICE: Proper alignment of clamps is necessary to prevent damage.
Figure 6 The lower clamp should align with the upper edge of the top chamfer on the clear plastic section
Installation
7
Page 8
3. Once you have the EXO PAR system secured to the EXO2 sonde body it is time to deploy the sensor arms. To do so simply lift the arms and install the pins where the two arms meet, as shown (figure 7).
Figure 7 Lift the arms and install the pins where the two arms meet
CAUTION: Potential pinch hazard, be mindful of fingers.
4. After you have installed the upper and lower clamps, it’s time to address cable management. There are two cables
that connect to your PAR sensors coming from the EXO PAR adapter cylindar. Supplied with your EXO PAR system will be several black, UV resistant Zip Ties. Fasten the sensor cables to the EXO PAR frame support arms as follows: (3 locations)
NOTE: Dummy plugs are supplied for these cable’s connectors, plug these connectors when sensors are removed or if only deploying 1 of your 2 PAR sensors. When not in use, the plugs can be stored in the calibration cup extension.
Figure 8 Fasten the sensor cables to the EXO PAR frame support arms
8
Installation
Page 9
3. Setup
Now that your EXO PAR system is installed, it’s time to connect your EXO PAR Sonde to either your PC running KOR or your EXO handheld system. For PC applications, you can use the Bluetooth functionality of the EXO to connect with KOR. For EXO handheld applications, you will need to connect an EXO field cable between the EXO communications bulkhead connector and the EXO handheld communications bulkhead connector. The EXO field cable comes in differing lengths depending on your needs, its part number is; 599040-xx (the –xx signifies the cable length; ex -2 equals 2 meters) Cable lengths start at 2 meters.
If the EXO PAR sonde is connecting to a data logger, you can use the EXO flying lead cable, part number 599008-xx (again the –xx signifies the cable length) flying lead cables start at 10 meters. You will also need the EXO DCP adapter to connect to a data logger, part number 599820.
Each Li-Cor PAR sensor is supplied with a Certificate of Calibration, shown below (figure 9). The calibration certificate contains the multipliers for the sensor. YSI uses the in-water multiplier for our EXO adapter, providing PAR engineering units out as data. In the following sections, we will review where and how to end this multiplier into either KOR EXO software or the EXO handheld display.
Setup
Figure 9 Certificate of Calibration for Li-Cor Sensor
9
Page 10
3.1
In the following steps we will cover the setup of the PAR sensor utilizing the YSI Kor software.
1. You will first need to enable the PAR(s) sensor in Kor. To do so you will need to select File, and then select Settings, and from the list of parameters select PAR and you will see figure 8 if done correctly. From this window
you can enable either the PAR sensor or sensors by clicking the boxes shown.
Setup the Kor Software
Figure 10 Enable the PAR sensor in Kor
2. Using the Calibration Certificate supplied with your Li-Cor PAR sensor, you will now enter the multipliers into YSI’s
Kor software. To do so first select Instruments and Sensors and from the list of sensors displayed select PA R sensor. If successful you will see the screen shown in figure 11. From the Calibration Certificate you will use the In-Water multiplier that is listed under the Handheld Meters and Loggers section. You will enter the multipliers exactly as they are on the certificates in the fields shown for Channel 1 and Channel 2. If you are only using one PAR sensor, the default channel will be Channel 1. Leave the default value in Channel 2, which will be (1).
10
Setup
Page 11
Figure 11 Enter the multipliers into YSI’s Kor software.
3. Now your PAR sensor(s) are ready to use. You can now see PAR data begin to show up in your Kor Dashboard
(example shown in figure 12).
Setup
Figure 12 PAR data will begin to show up in your Kor Dashboard
11
Page 12
3.2
In the following section we will cover setting up EXO PAR utilizing the EXO Handheld Display.
1. You will first need to enable the PAR(s) sensor in the EXO Handheld. To do so you will need to select Handheld, and then select Display, and from the list of Units select PAR and you will see figure 13. On the next screen you can enable either the PAR sensor or sensors by clicking the boxes shown, figure 14.
Setup the EXO Handheld Display
Figure 13 Select PAR from the list
of Units
2. Using the Calibration Certificate supplied with your Li-Cor PAR sensor, you will now enter the multipliers into EXO Handheld Display. To do so first select Calibration and from the list of sensors displayed select PAR sensor. If successful you will see the screen shown in figure 15. From the Calibration Certificate you will use the In-Water multiplier that is listed under the Handheld Meters and Loggers section. You will enter the multipliers exactly as they are on the certificates in the fields shown for Channel 1 and Channel 2 shown in figure 16. If you are only using one PAR sensor, the default channel will be Channel 1. Leave the default value in Channel 2, which will be (1).
12
Figure 14 Enable sensors
Setup
Page 13
Figure 15 Select calibration, then PAR from the list of sensors
3. Now your PAR sensor(s) are ready to use. You can now see PAR data begin to show up in your EXO Handheld Display (example shown in figure 17).
Figure 16 Enter the multipliers exactly as they are on the certificates
Setup
Figure 17 PAR data on EXO Handheld Display
13
Page 14
4. Ordering
4.1
4.2
Telephone: 727 565 2201 (USA)
Monday through Friday 8:00 AM to 5:00 ET
Ordering Replacement Components
YSI Item # Description
351070 EXO PAR Adapter Assembly
351073 EXO PAR, Upper Clamp
351074 EXO PAR, Lower Clamp
351075 EXO PAR, Wide Extension Arm (Lower)
351076 EXO PAR, PAR Sensor Bracket
351077 EXO PAR, Narrow Extension Arm (Upper)
351087 EXO PAR, Clip on Calibration Cup Extender
Ordering
Fax: 866 778 8431 (orders)
Email: YSISystemsOrders@Xyleminc.com
Mail: YSI Inc - Systems & Services Division
Attn: Order Entry 7100 Business Park Drive, Suite B Houston, TX 77041
When placing an order please have the following available:
1. YSI account number (if available)
2. Name and phone number
3. Purchase Order or Credit Card number
4. Copy of Quote (if applicable)
5. YSI Item Number or brief description
6. Billing and shipping addresses
7. Quantity
14
Ordering
Page 15
This page intentionally left blank
Page 16
1) The tissue in plants that brings water upward from the roots;
2) a leading global water technology company.
We’re a global team unified in a common purpose: creating advanced technology solutions to the world’s water challenges. Developing new technologies that will improve the way water is used, conserved, and re-used in the future is central to our work. Our products and services move, treat, analyze, monitor and return water to the environment, in public utility, industrial, residential and commercial building services settings. Xylem also provides a leading portfolio of smart metering, network technologies and advanced analytics solutions for water, electric and gas utilities. In more than 150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise with a strong focus on developing comprehensive, sustainable solutions.
For more information on how Xylem can help you, go to www.xylem.com
USER MANUAL XA00014
YSI, a Xylem brand
7100 Business Park Drive, Suite B Houston, TX 77041
+1.713.681.6101 info@ysi.com YSI.com
EXO is a trademark of Xylem or one of its subsidiaries. © 2019 Xylem, Inc. XA00014 0719
YSI.com/Systems