1. 安全措施.......................................................................................................................... 1
1.1 安全措施概述.................................................................................................................... 1
2. 综述.................................................................................................................................. 4
2.1 摘要.................................................................................................................................... 4
2.2 打开仪器包装.................................................................................................................... 4
2.3 YSI 556型 多参数测量仪 的外观 .................................................................................... 5
2.4 电池.................................................................................................................................... 6
2.4.1 电池寿命....................................................................................................................... 6
2.4.2 安装4节2号电池........................................................................................................... 6
2.4.3 安装充电电池............................................................................................................... 7
2.4.4 给充电电池充电........................................................................................................... 7
2.4.5 储存电池....................................................................................................................... 8
2.4.6 可选点烟器充电器....................................................................................................... 8
2.5 开电源................................................................................................................................8
2.6 设置显示对比度................................................................................................................ 8
2.7 背景灯................................................................................................................................8
2.8 显示区说明........................................................................................................................ 9
2.9 键盘说明............................................................................................................................ 9
2.10 仪器复位.......................................................................................................................... 10
2.11 菜单流程表...................................................................................................................... 11
3. 探头模块........................................................................................................................ 12
3.1 概述.................................................................................................................................. 12
3.2 打开探头包装.................................................................................................................. 12
3.3 YSI 5563 探头特点......................................................................................................... 12
3.4 准备安装探头.................................................................................................................. 12
3.4.1 安装探头..................................................................................................................... 12
3.4.2 安装盖膜..................................................................................................................... 14
3.5 运输/校准杯 .................................................................................................................... 14
3.5.1 安装运输/校准杯........................................................................................................14
3.6 连接仪器/电缆 ................................................................................................................ 14
4. 传感器............................................................................................................................ 15
5. 显示................................................................................................................................16
6. 校准................................................................................................ 错误!未定义书签。
6.1 准备校准.......................................................................................................................... 18
6.1.1 用于校准的容器......................................................................................................... 18
6.1.2 校准技巧..................................................................................................................... 18
6.1.3 推荐的校准液用量..................................................................................................... 18
6.2 校准步骤.......................................................................................................................... 19
6.2.1 进入校准窗口............................................................................................................. 19
6.2.2 校准电导..................................................................................................................... 19
6.2.3 校准溶解氧................................................................................................................. 21
6.2.4 校准pH ................................................................................................................... 24
6.2.5 校准氧化还原电位..................................................................................................... 26
6.3 返回出厂设定值.............................................................................................................. 27
1
7. 运行................................................................................................................................29
7.1 实时数据采集.................................................................................................................. 29
8. 文件................................................................................................................................30
8.1 进入文件窗口.................................................................................................................. 30
8.2 目录.................................................................................................................................. 30
8.3 浏览文件.......................................................................................................................... 31
8.4 上传文件至计算机.......................................................................................................... 32
8.4.1 上传设置..................................................................................................................... 32
8.4.2 上传一个数据文件..................................................................................................... 33
8.4.3 上传一个校准记录文件............................................................................................. 34
8.5 文件内存.......................................................................................................................... 34
8.6 删除所有文件.................................................................................................................. 35
9. 记录................................................................................................................................37
9.1 进入记录设置窗口.......................................................................................................... 37
9.2 设置记录间隔.................................................................................................................. 37
9.3 储存大气压读数.............................................................................................................. 37
9.4 创建采样点列表.............................................................................................................. 38
9.5 编辑一个采样点列表...................................................................................................... 40
9.6 无采样点列表的数据记录.............................................................................................. 41
9.7 由采样点列表时记录数据.............................................................................................. 42
9.8 在已存在的文件中追加数据.......................................................................................... 43
10. 系统设置........................................................................................................................ 45
10.1 进入系统设置窗口.......................................................................................................... 45
10.2 日期和时间设置.............................................................................................................. 45
10.3 数据滤波.......................................................................................................................... 46
10.3.1 改变数据滤波设置................................................................................................. 46
10.4 自动关机时间.................................................................................................................. 47
10.5 逗点基数.......................................................................................................................... 47
10.6 ID .....................................................................................................................................48
10.7 GLP 文件名..................................................................................................................... 48
10.8 总溶解固体(TDS)常数.............................................................................................. 48
10.8.1 改变 TDS常数........................................................................................................48
10.9 气压计单位...................................................................................................................... 48
10.10 校准气压计...................................................................................................................... 49
11. 维护................................................................................................................................51
11.1 传感器维护与保养.......................................................................................................... 51
11.1.1 维护溶解氧传感器................................................................................................. 51
11.1.2 更换溶解氧传感器................................................................................................. 52
11.1.3 清洗5564 pH传感器或5565pH/ORP复合传感器................................................. 52
11.1.4 温度/电导传感器清洗 ........................................................................................... 53
11.2 YSI 556 MPS 软件升级.................................................................................................. 54
12. 存放................................................................................................................................56
12.1 短期存放的有关建议...................................................................................................... 56
12.2 关于长期存放的建议...................................................................................................... 56
12.2.1 存放探头模块......................................................................................................... 56
12.2.2 存放传感器 ............................................................................................................ 56
2
13. 故障诊断........................................................................................................................ 58
14. 附录 A YSI 556 MPS性能指标................................................................................60
14.1 传感器指标...................................................................................................................... 60
14.2 仪器技术指标.................................................................................................................. 61
15. 附录 B 仪器附件.........................................................................................................62
16. Appendix C Required Federal Communications Notice ..........................................63
17. 附录 D 健康与安全.....................................................................................................64
18. 附件 E 用户服务信息................................................................................................67
18.1 YSI 技术服务中心.......................................................................................................... 67
18.2 清理说明.......................................................................................................................... 67
18.3 包装步骤.......................................................................................................................... 68
18.4 清洗证明书...................................................................................................................... 68
18.5 保修.................................................................................................................................. 68
19. 附录 G EcoWatch软件..............................................................................................69
19.1 Windows操作系统中安装EcoWatch软件..................................................................69
19.2 EcoWatch向导.................................................................................................................69
19.2.1 工具栏的功能概述................................................................................................. 74
19.2.2 其它功能 ................................................................................................................ 75
20. 附录H 校准记录信息.................................................................................................. 76
20.1 查看校准记录(.glp)文件............................................................................................ 76
20.2 上载校准记录(.glp)文件............................................................................................ 76
20.3 校准记录(.glp)文件的内容........................................................................................ 76
3
1. 安全措施
1.1 安全措施概述
在使用 YSI 556MPS 型 多参数水质测量仪前,请仔细阅读本手册中所有的 安全需知。某些校准液和缓冲液对身体具有危害性,请浏览
安全
警告
仪器使用不当可能造成人员的重伤和死亡时在本手册中使用“警告”语 句。
当心
仪器使用不当可能造成人员的轻微伤害时在本手册中使用“当心”语句。
重要安全指示
使用 YSI 556MPS 最重要的安全原则是按照手册中的规定使用。对于含有镍
氢化合物的可充电电池盒尤其如此。用户在使用仪器之前请务必阅读预防 安全措施。.
附录D 健康与
儿童注意事项
YSI 6117 可充电电池盒安全知识
使用限制
1. 禁止暴露在火中。
2. 不要试图拆卸 YSI 6117 型 电池盒
3. 不要损坏电池盒中的电池或电子成分,否则会由于过热、过重等原因引 起安全危害。电子线路或电池的损坏不在保修范围内。
4. 充电时的环境温度应在0-40℃之间。
5. 不要在高温下使用或储存,如强光、高温下的汽车等。
6. 不要把电池暴露在水中或使终端变潮湿。
7. 避免击打或摔落电池盒。如果击打或摔落后电池盒出现损害或故障时, 请不要擅自拆卸,请联系 YSI 技术服务中心,参阅
8. 电池盒不可放在装有金属物(如钥匙等)的袋中,因为这样很容易导致
短路。
将电池盒置于儿童接触不到的地方。
附录E 客户服务。
危险提示——错误使用会导致重伤或死亡。
不遵循以下安全使用措施可能会导致电池泄露、爆炸及严重的人员伤亡。
1. 不可将电池盒暴露在火中。
2. 避免电极短路。
3. 不可擅自拆卸电池盒或损坏电池盒中的电子成分。
4. 电池盒中 NIME 电池盒含有强烈的碱性溶液,这种碱性溶液对皮肤和身
体的其它组织具有极强的腐蚀性。若电池盒中的溶液不慎溅入眼睛,请
1
第一章 安全措施
立即用水冲洗并寻求医生的帮助,碱性溶液会伤害眼睛并会导致永久失 明。
警告指示——使用不当会造成人员的伤亡。
1. 电池盒不可接触淡水、海水或其它氧化剂,否则将导致其生锈,并产生 热量。若电池生锈,热量无法散失,就会产生爆裂。
2. 若电池电解液接触了皮肤或衣服,立即用清水冲洗,电解液会腐蚀皮 肤。
警示信息——错用会导致轻微或严重伤害,也可能造成仪器损伤。
1. 不要击打或摔落电池盒,若怀疑电池盒出现硬损伤,请联系 YSI 技术服 务中心,参阅
2. 将电池盒存放于小孩拿不到的地方。
3. 20 -30°C 之间存放电池盒。
4. 在使用电池盒前,仔细阅读操作手册和注意事项,并在阅读后妥善保 存,以备使用时查阅。
YSI 616 点烟器充电器安全操作说明
附录E客户服务。
1. 这一部分包含 YSI 556MPS 点烟器充电器(YSI 616RadioShack Number 270-1533E)安全操作说明,一定要妥善保存。
2. 在使用 YSI 616 点烟器充电器前,需阅读所有说明和在电池充电器、电 池盒及 YSI 556 MPS 上的警告信息。
3. 在使用 YSI 616 点烟器充电器前,先确认 YSI 6117 电池盒已安装在 YSI 556 MPS 上。
4. YSI 616 点烟器充电器要防水、防潮、防雪。
5. 使用非厂家生产或非厂家推荐的辅助设备,有可能导致起火、电击、或 造成人身伤害。
6. 为避免造成充电器的损伤,在断开充电器时,拔充电器不要拔连线。
7. 确保连线所在处不会被人踩到,不会绊倒人,不会损伤和压迫连线。
8. 当连线或连接头损坏时,不要使用充电器,立即更换它们。
9. 当充电器被锐器碰伤、被摔坏或有其他硬损伤时,不要使用它,联系 YSI 技术服务中心,参阅
10. 除更换保险丝外,不要拆开充电器,如需维修送交 YSI 技术服务中心,
附录E客户服务
参阅
11. 为避免电击,在保养或维护前,拔下充电器。单纯关掉开关不会消除电 击危险。
附录E客户服务
。不正确的组装会导致起火或电击。
YSI 556 MPS 水密安全信息
The YSI 556 MPS 经过认证,符合 IP67 防水标准。例如,将仪器放在水下 一米处 30 分钟,主机和电池盒都不会进水。但若超过 30 分钟,电池盒或主 机将有可能进水造成损伤。
当使用 2 号电池时,出现电池盒进水,立刻取出电池,适当的换掉浸水的 电池,将电池仓完全弄干(比较理想的方法是使用压缩空气),若电池仓 中的电池电极金属片被腐蚀,送交 YSI 技术服务中心,参阅
附录E客户服
2
第一章 安全措施
当使用 YSI 6117 充电电池时,出现电池盒进水,立刻取出电池盒并烘干, 联系 YSI 技术服务中心,做出可能的损伤评估。最后,将电池仓完全弄干 (比较理想的方法是使用压缩空气),若电池仓中的电池电极金属片被腐 蚀,送交 YSI 技术服务中心,参阅
警告:若充电电池盒已接触水,在 YSI 技术服务中心做出评估前不要再使
用,参阅 若怀疑水由气压计的孔进入主机内,立即将电池取出,送交 YSI 技术服务
中心做损伤评判,参阅
警告:在任何情况下,用户不可试图打开主机。
附录E客户服务
附录E客户服务
。不履行这一说明会对使用者造成伤害。
附录E客户服务
3
2. 综述
2.1 摘要
高质量的 YSI 556MPS 型 多参数水质测量仪易于使用、便于读数, 可同时 测量溶解氧、温度、电导、盐度、pH、氧化还原电位和总溶解固体,所有 数据同时显示在屏幕上。手机式键盘设计和大显示屏都使仪器便于使用。
YSI 556 MPS 和适用于 Windows YSI 556 MPS 符合 GLP 标准,保证质量。附电量显示,用户可以选择标准
碱性电池和可充电电池。 512KB 内存可以储存 49,000 多组数据。可选件包括电缆长度、充电电池、
气压计和流动室。内置的气压计可由用户校准,并可与其它参数一起显 示,同时可存在内存中跟踪气压改变。
特征
IP67 防水等级设计
TM
YSI EcoWatchTM软件相匹配。
2.2 打开仪器包装
所有探头均可在野外由操作人员自行更换
可直接与计算机连接,通过 EcoWatch 计算机软件进行数据分析
符合 GLP 标准
使用新一代快速反应∕低搅拌依赖性溶解氧 PE 盖膜
通过 YSI 网站用户可免费升级软件
仪器保修三年,探头和电缆保修一年
可选 4 米、10 米 和 20 米电缆
可储存 49,000 组数据(含日期和时间)
自动温度补偿
可选配件:电缆长度、充电电池、气压计和流动室
1. 从运输箱中取出仪器。
注意:探头安放在另外的盒子里,将在 3.2 中说明怎样打开它。 注意:不要遗失任何部件。
2. 查看收到的设备是否与装箱单中的材料列表一致。
3. 查看部件是否受损。
注意:如发现有损坏或零件不全,请立即通知您所在区域的代理商或 YSI 公司。
4
2.3 YSI 556 型 多参数测量仪 的外观
第二章 综述
显示
背景灯键
取消键
气压计通气 孔
/关键
箭头键
2.1 YSI 556 MPS前视图
回车键
字母/数字键盘
电缆接口
附件安装孔
后盖
2.2 YSI 556 MPS后视图
5
电池盖螺丝
2.4 电池
2.4.1 电池寿命
标准碱性电池
带有 4 2 号碱性电池的 YSI 556 MPS 能连续使用 180 小时。假设一天采 样三小时,则可使用 60 天。
可选可充电电池包
充满电后约可使用 50 小时。
2.4.2 安装 4 2 号电池
第二章 综述
2.3 安装2号电池
警告:正确安装电池,避免损坏仪器。
1. 拧下仪器后盖上的 4 个螺丝。
2. 移走电池盖。
3. 按照电池槽中标注的极性装入四节电池。
4. 检查电池盖上的垫圈是否安装到位。
5. 装回电池盖,拧紧螺丝。
注意:不要拧得过紧。
6
2.4.3 安装充电电池
第二章 综述
图 2.4 安装充电电池
警告:使用前先阅读电池后面的警告。
1. 拧下仪器后盖上的 4 个螺丝。
2. 移走电池盖储存起来;如有电池,也移走。
3. 检查电池盖上的垫圈是否安装到位。
4. 安装充电电池,拧紧 4 个螺丝。
2.4.4 给充电电池充电
充电适配电缆 (YSI 6119)
警告:不要把电池放在太阳直射的地方,如高温下的汽车里、靠近加热器的
地方等。
1.把充电电池安装在仪器上,参见 2.4.3。 2.将充电适配电缆 (YSI 6119) 安装到仪器上。
注意:不要拧得过紧。
充电器 (YSI6114
图 2.5 给充电电池充电
注意:美国和加拿大之外的充电器的使用请参阅
附录B仪器附件
3.连接充电器插头与适配电缆。
警告:充电时间不可超过 48 小时。
警告:不要接触到水。
警告:不要在低于 0°C 或高于 40°C 的温度下充电。
4.将充电器插头插入交流电源插座。2 3 小时可充到 80% 90%的电 量,6 小时可以充满。
注意:无论仪器开或关充电电池都可以充电。
7
2.4.5 储存电池
当充电电池不用时,将它卸下放到厚塑料袋中,防止接线端意外短路,储 存在–20 30°C 的环境中。
2.4.6 可选点烟器充电器
警告:使用充电器前阅读所有的警告。
警告:只有使用充电电池时才能用点烟器充电器。
警告:不要对点烟器充电器进行错误操作,不要暴露在潮湿的环境中。
1. 把充电器的插头插到充电适配电缆 YSI 6119 的一端。
2. YSI 6119 适配电缆另一端的 MS-19 接头接到仪器上。
3. 对其它充电器做以下调整。
第二章 综述
用美国和日本的设备需要将环状适配器从插头上拿走。
2.5 开电源
2.6 设置显示对比度
图2.6 充电插头适配器的使用 对欧洲的设备来说,将适配器固定到插头上。把插头上的弹簧片与适配
器上的窄槽对齐。
2.7 欧洲充电插头适配器的使用
注意:如果充电器出现问题,参见 13 部分故障诊断。
On/Off 键打开或关闭仪器,见图 2.1
显示对比度可根据温度的改变进行自动补偿。然而在一些特殊的温度下, 需要手动调节以优化显示。
2.7 背景灯
1.按住背景灯键不放,按U键升高对比度。 2.按住背景灯键不放,按V键降低对比度。
按小键盘右上角的背景灯键可以开/关背景灯,见图 2.1。 注意:两分钟内仪器无操作,背景灯会自动关闭。
8
第二章 综述
2.8 显示区说明
2.9 键盘说明
主显示区 状态栏
大气压读数(可选项)– 可实时更新 且没有标定到海平面的读数
电池电量 – 带NiMH 标签表示使 用的是可充电电池, 以脉冲形式 出现表示正在充电, 以闪动形式 出现表示电池将要用完。
当前时间 当前日期
图 2.8 主菜单窗口
开/关键 退出键 – 在菜单中用
于返回前一位置
字母/数字键– 用于输 入字母和数字
减号/连字号 键
背景灯/对比度调整 箭头键
回车键
日期/十进制小数点 键
2.9 键盘说明
字母/ 数字
1 1 2 ABC2abc3 3 DEF3def3 4 GHI4ghi4 5 JKL5jkl5 6 MNO6mno6 7 PQRS7pqrs7 8 TUV8tuv8 9 WXYZ9wxyz9 0 0
2.10 键盘字母/数字键说明
1. 如图 2.10 所示,重复按键直到显示所需的数字或字母。
9
2.10 仪器复位
第二章 综述
注意:按键时要快速连续直到出现所需的字母。如果中断超过一秒钟, 光标会自动向右滚动,为下一个字符输入做准备。
例 1:按一次 6 键显示大写字母“M”; 例 2:按四次 6 键显示数字“6”; 例 3:按五次 6 键显示小写字母“m”。
2. Y键重新输入需要改变数字或字母。
3. 按 回车 键完成输入。
注意:在某些情况下只允许输入数字,如在设定时钟或输入校准参数 时。
YSI 556 MPS 的软件是非常完善的,具有一定的容错能力。但是和其它高性 能的软件一样,用户也会遇到仪器对输入不产生响应的时候。如果遇到这 种情况,用以下方法可快速恢复:移走电池,30 秒后再安上,参见 2.4
10
2.11 菜单流程表
第二章 综述
Run
Report
Sensor
Calibrate
File
Logging Setup
System Setup
Log One Sample Start Logging
Temperature Conductivity
Dissolved Oxygen pH ORP
DIrectory Upload to PC
View Fi le File Memory Delete All Files
Version Date & Time
Data Filter Shut Off Time Comma Radix ID GLP Filename TDS Constant Barometer Units Calibrate Barometer
Temperature Specific Conductance
Conductivity Resistivity TDS Salinity ppt DO % saturation DO mg/L pH pH mv ORP mv
Conductivity DO
pH ORP
Interval Use Site List
Store Barometer Store Site Number Edit Site List
11
3. 探头模块
3.1 概述
YSI 5563 探头可以测量溶解氧、温度、电导和可选的 pH 与氧化还原电位。 探头非常结实,并且用带有加重器的保护罩保护。4, 10 20 米的电缆直接 跟探头相连,符合防水标准。电缆末端的 MS-19 连接器使得电缆和主机非 常易于连接。
3.2 打开探头包装
1. 从运输箱中取出 YSI 5563 探头。
注意:不要遗失任何部件。
2. 查看收到的部件是否与装箱单中的材料列表一致。
3. 查看部件是否受损。
注意:如发现有损坏或零件不全,请立即通知您所在区域的代理商或 YSI 公司。
3.3 YSI 5563 探头特点
电缆
溶解氧探头
运输/校准杯
3.4 准备安装探头
应力舒缓器
金属探头连接螺帽
pH/ORP 探头
电导/温度探头
图3.1 探头模块
校准和使用仪器前,把所有探头安装到相应的插槽中。除探头外,还需要 安装一个新的溶解氧盖膜。
探头保护罩
3.4.1 安装探头
为防止水进入插槽,安装或移动探头时保持探头和插头干燥是非常重要 的。如果插槽中有水,一定要用压缩空气吹干。如果接口已经腐蚀,请把 探头寄回 YSI 客户服务部。
安装电导/温度和 pH 探头
1. 拧下保护罩。
2. YSI 5511 维护包中的工具拧下插口保护塞,如图 3.2 所示。
12
第三章 探头模块
3.2 移走端口保护堵头
3. 槽定位,如图 3.3 所示。
pH插槽
溶解氧插槽
电导/温度插槽
3.3 探头插槽识别
4. 在密封圈上涂上润滑油(由 YSI 5511 维护包提供),如图 3.4 所示:
O-Rings
Sensor nut
3.4 密封圈润滑油
警告:确保密封圈和探头之间无污染物。
5. 确保探头插槽中没有水。轻轻转动探头,对齐探头。
6. 使用专用工具拧紧螺帽。
3.5 安装探头
警告:螺帽具有防水性,故千万不要划破螺帽上的螺纹,也不要拧得过
3.6 防水外壳
7. 重复第 3-6 步骤安装其它探头 。
8. 拧上探头保护罩
13
安装溶解氧探头
密封圈
3.4.2 安装盖膜
3.5 运输/校准杯
第三章 探头模块
按照上述步骤安装 YSI 5563 型 溶解氧探头。关于如何替换 YSI 559 溶解氧 电极模块,请参见 11.12 溶解氧探头替换。
注意:YSI 5563 型 溶解氧传感器(已经安在探头模块上)在运输时是干燥 的,为了保护电极安装了一个运输盖膜。使用前必须更换新盖膜。
1. 移走探头保护罩。
2. 拧下旧盖膜。
3. 用蒸馏水清洗探头顶部。
4. 根据电解液瓶上的说明准备电解液。
5. 在新盖膜中加入电解液,至少占到整个容积的一半。
6. 将盖膜拧到探头上,适度拧紧,会有少量电解液溢出。
警告:不要碰到盖膜表面。
YSI 5563 探头模块提供一个使用方便的运输/校准杯。这个杯子是理想的校 准容器,可以最少量的使用校准液,参见第 6 章校准。
3.5.1 安装运输/校准杯
1. 移走探头保护罩。
2. 确保密封圈已安装在丝扣末端的凹槽中。
3. 拧上杯子运输/校准杯,适度拧紧,如图 3.7 所示。
3.6 连接仪器/电缆
注意:不要拧得过紧,以免损坏丝扣。
3.7 安装运输/校准杯
按以下步骤把电缆安装到仪器上:
1. 对齐电缆接头上的凸起与 YSI 556 末端接头上的凹槽。
2. 顺时针转动电缆接头螺帽,直到拧紧。 注意:反时针转动电缆接头螺帽,可使电缆和仪器分开。
14
4. 传感器
在下面的界面中,用户可激活或屏蔽探头,被屏蔽的探头不会显示在屏幕 上。
1. On/Off 键显示运行窗口。
2. 按退出 键显示主菜单窗口,如图 4.1 所示。
被激活探头
被屏蔽探头
4.1 主菜单窗口
3. 按△/▽键选中 Sensor 选项。
4. 按回车键显示探头激活窗口,如图 4.2 所示。
4.2 探头激活窗口
注意:左边的圆圈中有黑点表示探头被激活,无黑点表示探头被屏蔽。 注意:温度探头不能被屏蔽,因为其它参数需要温度作为参考。
5. 按△/▽键选中所需的探头,按回车键确认。
6. 重复第五步激活其它探头。
7. 按退出键返回主菜单。
15
5. 显示
在显示设置窗口中,用户可选择显示参数和单位。但此窗口无法决定哪个 参数会被存入内存,参阅第 四章。
1. On/Off 键显示运行窗口。
2. 按退出 键显示主菜单窗口,如图 5.1 所示。
选中要显示
的参数
没有选中
5.1 主菜单窗口
3. 按△/▽键选中 Report 选项
4. 按 回车键显示 Report setup 窗口,如图 5.2 所示。
5.2 显示设置窗口
注意:左边的圆圈中有黑点表示参数被选中,无黑点表示参数未被选 中。
注意:向下滚动屏幕可显示所有参数。
5. 按△/▽键选中所需参数,然后按回车键。如果找不到所需参数,说明 这个探头已被屏蔽。参见第 4 章。
6. 如果选择温度、比电导、电导、电阻和总溶解固体,则会显示单位选择 窗口,如图 5.3 所示。
16
第五章 显示
5.3 单位窗口
7. 按△/▽键选择所需单位,然后按 回车 键回到显示设置窗口。
8. 如果选择盐度、溶解氧%、溶解氧毫克/升、pHpH 毫伏值、氧化还原 电位,则不用选择单位。
9. 要设定其它参数,请重复第 5、6 步。
注意:所有参数可同时被选中,如图 5.4 所示。
5.4 所有参数同时显示
10. 按 退出 键回到主菜单窗口。
17
6. 校准
除了温度探头外,其它探头均需定期校准以保证较高的精度。本章将介绍 各探头的详细校准步骤。如果你购买的模块中不包括下面列出的某一探 头,跳过它,继续完成其它校准。
警告:校准液可能对你的健康有害。迅速浏览
6.1 准备校准
6.1.1 用于校准的容器
6.1.2 校准技巧
附录D 健康和安全
准液需要特别处理。
专用校准杯可用来完成所有的校准,并且可以减少校准液的用量。 除专用校准杯外,也可使用实验室里的玻璃杯校准。如果不使用专用校准
杯,一定要注意以下几点:
校准时戴上探头保护罩,以免碰坏探头。
用环形夹夹住探头以防止探头碰到玻璃杯底部。
保证所有探头均没入校准液中,因为探头的读数是相互联系的(例如:
温度)。电导顶部的透气孔也要没入在校准液中。
1. 如果用专用校准杯校准溶解氧,一定要放松杯口使内外压力达到平衡。 溶解氧的校准需在水蒸气饱和空气中。
2. 校准成功的关键是要保证探头完全浸在校准液中(校准溶解氧%除外)
。校准时,请使用推荐容量。
,许多校
3. 为获得最高精度,可使用以前用过的校准液预先冲洗探头模块,故需积
4. 在两次校准之间,用室温水冲洗探头模块。
5. 浸泡完后一定要用滤纸或棉布把探头擦干。特别是装有保护罩的时候,
6. 如用玻璃杯校准,浸入水中时不必拿开保护罩,只需把外面的水擦干就
7. 如用玻璃杯校准,把保护罩底部的加重头逆时针拧下。取下加重头可使
8. 确保在没有探头的端口上安上保护堵头,使端口保持干燥是非常重要
6.1.3 推荐的校准液用量
用专用校准杯校准时参考以下步骤:
确保校准杯底盖上附有密封圈,并且拧紧底盖。
移走探头保护罩。
取下探头上的密封圈,检查有没有明显的损坏,如有损坏请更换新的密
攒旧校准液。
一定要把它擦干,防止污染校准液降低精度。
行。
校准液更容易接触到探头,并减少粘带校准液。
的。
注意:不要拧得过紧以免损坏螺纹。
封圈。
许多校准需要探头竖放或反放,因此需要一个架子或支架(如环形 夹)。
18
第六章 标定
校准步骤参见 6.2,下表列出了校准所需校准液的体积。
校准溶解氧饱和百分比时,拧松杯口,使里外气压平衡,并在杯底放置
3 毫米深的水。
校准的探头 正放 反放
6.2 校准步骤
6.2.1 进入校准窗口
电导 pH/氧化还原电位
55ml 55ml 30ml 60ml
6.1 校准液体积
1. On/Off 键进入运行窗口 。
2. 按 退出键进入主菜单窗口,如图 6.1 所示。
6.2.2 校准电导
6.1 主菜单窗口
3. 按△/▽键选中 Calibrate 选项。
4. 按回车 键进入校准窗口,如图 6.2 所示。
6.2 校准窗口
这个步骤用来校准比电导、电导和盐度。校准一个选项会自动校准另外两 个。
1. 进入校准窗口,参见 6.2.1