YSI 556 MPS Operating Manual

Y
Pure Data for a Healthy Planet.™
SI 556 MPS
Multi Probe System
Operations Manual
Contents
1. Safety .........................................................................................................................1
1.1 General Safety Information ...............................................................................1
2. General Information ................................................................................................7
2.1 Description.........................................................................................................7
2.2 Unpacking the Instrument..................................................................................8
2.3 Features of the YSI 556 Multi-Probe System....................................................9
2.4 Batteries ...........................................................................................................10
2.5 Power On .........................................................................................................15
2.6 Setting Display Contrast..................................................................................15
2.7 Backlight..........................................................................................................16
2.8 General Screen Features ..................................................................................16
2.9 Keypad Use......................................................................................................17
2.10 Instrument Reset..............................................................................................18
2.11 Menu Flowchart...............................................................................................19
3. Probe Module..........................................................................................................21
3.1 Introduction......................................................................................................21
3.2 Unpacking the Probe Module ..........................................................................21
3.3 Features of the YSI 5563 Probe Module .........................................................22
3.4 Preparing the Probe Module............................................................................22
3.5 Transport/Calibration Cup...............................................................................26
3.6 Instrument/Cable Connection ..........................................................................27
4. Sensors.....................................................................................................................29
5. Report ......................................................................................................................33
6. Calibrate..................................................................................................................37
6.1 Getting Ready to Calibrate ..............................................................................37
6.2 Calibration Procedures.....................................................................................40
6.3 Return to Factory Settings ...............................................................................56
7. Run...........................................................................................................................59
7.1 Real-Time Data................................................................................................59
8. File............................................................................................................................61
8.1 Accessing the File Screen................................................................................61
8.2 Directory..........................................................................................................62
8.3 View File .........................................................................................................63
8.4 Upload to PC....................................................................................................64
8.5 File Memory ....................................................................................................69
YSI Incorporated YSI 556 MPS i
Contents
8.6 Delete All Files................................................................................................ 70
9. Logging.................................................................................................................... 73
9.1 Accessing the Logging Setup Screen .............................................................. 73
9.2 Setting Logging Interval.................................................................................. 74
9.3 Storing Barometer Readings ........................................................................... 74
9.4 Creating a Site List.......................................................................................... 75
9.5 Editing a Site List............................................................................................ 79
9.6 Logging Data Without a Site List.................................................................... 81
9.7 Logging Data With a Site List......................................................................... 84
9.8 Adding Data to Existing Files ......................................................................... 85
10. System Setup........................................................................................................... 89
10.1 Accessing the System Setup Screen................................................................ 89
10.2 Date and Time Setup....................................................................................... 90
10.3 Data Filter........................................................................................................ 92
10.4 Shutoff Time.................................................................................................... 94
10.5 Comma Radix.................................................................................................. 95
10.6 ID..................................................................................................................... 95
10.7 GLP Filename.................................................................................................. 95
10.8 TDS Constant .................................................................................................. 96
10.9 Barometer Units .............................................................................................. 97
10.10 Calibrate Barometer .................................................................................... 98
11. Maintenance .........................................................................................................101
11.1 Sensor Care and Maintenance....................................................................... 101
11.2 Upgrading YSI 556 MPS Software............................................................... 107
12. Storage................................................................................................................... 111
12.1 General Recommendations for Short Term Storage...................................... 111
12.2 General Recommendations for Long Term Storage...................................... 111
13. Troubleshooting ...................................................................................................115
14. Appendix A YSI 556 MPS Specifications...........................................................119
14.1 Sensor Specifications .................................................................................... 119
14.2 Instrument Specifications .............................................................................. 120
15. Appendix B Instrument Accessories................................................................... 121
16. Appendix C Required Federal Communications Notice .................................. 123
17. Appendix D Health and Safety............................................................................125
18. Appendix E Customer Service............................................................................ 129
YSI 556 MPS YSI Incorporated
Page ii
Contents
18.1 YSI Environmental Authorized Service Centers...........................................129
18.2 Cleaning Instructions.....................................................................................131
18.3 Packing Procedure .........................................................................................132
18.4 Cleaning Certificate.......................................................................................132
18.5 Warranty ........................................................................................................133
19. Appendix F Ferrite Bead Installation.................................................................135
20. Appendix G EcoWatch.........................................................................................137
20.1 Installing EcoWatch for Windows.................................................................137
20.2 EcoWatch Tutorial.........................................................................................139
21. Appendix H Calibration Record Information ...................................................151
21.1 Viewing the Calibration Record (.glp) File ...................................................151
21.2 Uploading the Calibration Record (.glp) File ................................................151
21.3 Understanding the Calibration Record (.glp) File..........................................151
YSI Incorporated YSI 556 MPS Page iii
Contents
YSI 556 MPS YSI Incorporated
Page iv
1. Safety
1.1 General Safety Information
Read all safety information in this manual carefully before using the YSI 556 Multi-Probe System (MPS). Reagents that are used to calibrate and check this instrument may be hazardous to your health. Take a moment to review Appendix D Health and Safety.
WARNING
Warnings are used in this manual when misuse of the instrument could result in death or serious injury to a person.
CAUTION
Cautions are used in this manual when misuse of the instrument could result in mild or serious injury to a person and/or damage to equipment.
IMPORTANT SAFETY INSTRUCTIONS!
SAVE THESE INSTRUCTIONS!
In essence, the most important safety rule for use of the YSI 556
MPS is to utilize the instrument ONLY for purposes documented in this manual. This is particularly true of the YSI 6117 rechargeable battery pack that contains nickel metal hydride (NiMH) batteries. The user should be certain to read all of the safety precautions outlined below before using the instrument.
YSI 6117 Rechargeable Battery Pack Safety Information
Restrictions on Usage
1. Never dispose of the battery pack in a fire.
2. Do not attempt to disassemble the YSI 6117 battery pack.
3. Do not tamper with any of the electronic components or the batteries within the battery pack. Tampering with either the electronic circuitry or the batteries will result in the voiding of the warranty and the compromising of the system performance, but, more importantly, can cause safety
YSI Incorporated YSI 556 MPS Page 1
Safety Section 1
hazards which result from overcharging such as overheating, venting of gas, and loss of corrosive electrolyte.
4. Do not charge the battery pack outside the 0–40°C temperature range.
5. Do not use or store the battery at high temperature, such as in strong direct sunlight, in cars during hot weather, or directly in front of heaters.
6. Do not expose the battery pack to water or allow the terminals to become damp.
7. Avoid striking or dropping the battery pack. If the pack appears to have sustained damage from these actions or malfunctions after an impact or drop, the user should not attempt to repair the unit. Instead, contact YSI Customer Service. Refer to Appendix E Customer Service.
8. If the battery pack is removed from the YSI 556 MPS, do not store it in pockets or packaging where metallic objects such as keys can short between the positive and negative terminals.
Precautions for Users with Small Children
Keep the battery pack out of reach of babies and small children.
Danger Notifications – Misuse creates a STRONG possibility of death
or serious injury.
FAILURE TO CAREFULLY OBSERVE THE FOLLOWING PROCEDURES AND PRECAUTIONS CAN RESULT IN LEAKAGE OF BATTERY FLUID, HEAT GENERATION, BURSTING, AND SERIOUS PERSONAL INJURY.
1. Never dispose of the battery pack in a fire or heat it.
2. Never allow the positive and negative terminals of the battery pack to become shorted or connected with electrically conductive materials. When the battery pack has been removed from the YSI 556 MPS, store it in a heavy plastic bag to prevent accidental shorting of the terminals.
YSI 556 MPS YSI Incorporated
Page 2
Safety Section 1
3. Never disassemble the battery pack and do not tamper with any of the electronic components or the batteries within the battery pack. The battery pack is equipped with a variety of safety features. Accidental deactivation of any of these safety features can cause a serious hazard to the user.
4. The NiMH batteries in the battery pack contain a strong alkaline solution (electrolyte). The alkaline solution is extremely corrosive and will cause damage to skin or other tissues. If any fluid from the battery pack comes in contact with a user’s eyes, immediately flush with clean water and consult a physician immediately. The alkaline solution can damage eyes and lead to permanent loss of eyesight.
Warning Notifications – Misuse creates a possibility of death or serious
injury
1. Do not allow the battery pack to contact freshwater, seawater, or other oxidizing reagents that might cause rust and result in heat generation. If a battery becomes rusted, the gas release vent may no longer operate and this failure can result in bursting.
2. If electrolyte from the battery pack contacts the skin or clothing, thoroughly wash the area immediately with clean water. The battery fluid can irritate the skin.
Caution Notifications – Misuse creates a possibility of mild or serious
injury or damage to the equipment.
1. Do not strike or drop the battery pack. If any impact damage to the battery pack is suspected, contact YSI Customer Service. Refer to Appendix E Customer Service.
2. Store the battery pack out of reach of babies and small children.
3. Store the battery pack between the temperatures of -20 and 30°C.
4. Before using the battery pack, be sure to read the operation manual and all precautions carefully. Then store this information carefully to use as a reference when the need arises.
YSI Incorporated YSI 556 MPS Page 3
Safety Section 1
YSI 616 Cigarette Lighter Charger Safety Information
1. This section contains important safety and operating instructions for the YSI 556 MPS cigarette lighter battery charger (YSI 616; RadioShack Number 270-1533E). BE SURE TO SAVE THESE INSTRUCTIONS.
2. Before using the YSI 616 cigarette lighter charger, read all instructions and cautionary markings on battery charger, battery pack, and YSI 556 MPS.
3. Charge the YSI 6117 battery pack with the YSI 616 cigarette lighter charger ONLY when the YSI 6117 is installed in the YSI 556 MPS.
4. Do not expose charger to rain, moisture, or snow.
5. Use of an attachment not recommended or sold by the battery charger manufacturer may result in a risk of fire, electric shock, or injury to persons.
6. To reduce risk of damage to cigarette lighter and cord, pull by cigarette lighter rather than cord when disconnecting charger.
7. Make sure that the cord is located so that it will not be stepped on, tripped over, or otherwise subjected to damage or stress.
8. Do not operate charger with damaged cord or cigarette lighter connector – replace it immediately.
9. Do not operate charger if it has received a sharp blow, been dropped, or otherwise damaged in any way; contact YSI Customer Service. Refer to Appendix E Customer Service.
10. Do not disassemble charger other than to change the fuse as instructed. Replace the part or send it to YSI Product Service if repair is required (refer to Appendix E Customer Service). Incorrect reassembly may result in a risk of electric shock or fire.
11. To reduce risk of electric shock, unplug charger before attempting any maintenance or cleaning. Turning off controls will not reduce this risk.
YSI 556 MPS YSI Incorporated
Page 4
Safety Section 1
YSI 556 MPS Water Leakage Safety Information
The YSI 556 MPS has been tested and shown to comply with IP67 criterion, i.e. submersion in 1 meter of water for 30 minutes with no leakage into either the battery compartment or the main case. However, if the instrument is submersed for periods of time in excess of 30 minutes, leakage may occur with subsequent damage to the batteries, the rechargeable battery pack circuitry, and/or the electronics in the main case.
If leakage into the battery compartment is observed when using alkaline C cells, remove batteries, dispose of batteries properly, and dry the battery compartment completely, ideally using compressed air. If corrosion is present on the battery terminals, contact YSI Customer Service for instructions. Refer to Appendix E Customer Service.
If leakage into the battery compartment is observed when using the YSI 6117 rechargeable battery pack, remove the battery assembly and set aside to dry. Return the battery pack to YSI Product Service for evaluation of possible damage. Finally dry the battery compartment completely, ideally using compressed air. If corrosion is present on the battery terminals, contact YSI Customer Service for instructions. Refer to Appendix E Customer Service.
CAUTION: If water has contacted the rechargeable battery
pack, do not attempt to reuse it until it has been evaluated by YSI Product Service (refer to Appendix E Customer Service). Failure to follow this precaution can result in serious injury to the user.
If it is suspected that leakage into the main cavity of the case has occurred, remove the batteries immediately and return the instrument to YSI Product Service for damage assessment. Refer to Appendix E Customer Service.
CAUTION: Under no circumstances should the user attempt to
open the main case.
YSI Incorporated YSI 556 MPS Page 5
Safety Section 1
YSI 556 MPS YSI Incorporated
Page 6
2. General Information
2.1 Description
The rugged and reliable YSI 556 MPS (Multi-Probe System) combines the versatility of an easy-to-use, easy-to-read handheld unit with all the functionality of a multi-parameter system. Featuring a waterproof, impact-resistant case, the YSI 556 MPS simultaneously measures dissolved oxygen, conductivity, temperature, and optional pH and ORP. A simple cellular phone style keypad and large display make the instrument easy to use. The YSI 556 MPS is compatible with YSI EcoWatch Windows
The YSI 556 MPS assists the user in conforming to Good Laboratory Practice (GLP) standards which help ensure that quality control/quality assurance methods are followed. Battery life is displayed with a fuel gauge, and the user can choose standard alkaline batteries or an optional rechargeable battery pack.
The 1.5 MB memory can store more than 49,000 data sets. Other options include a flow cell and barometer. The internal barometer can be user-calibrated and displayed along with other data, used in dissolved oxygen calibrations, and logged to memory for tracking changes in barometric pressure.
TM
software.
TM
for
Features
Waterproof - meets IP67 specifications
Field-replaceable DO electrode module; pH and pH/ORP sensors
Compatible with Ecowatch
software
Assists with Good Laboratory Practice Standards (GLP)
Choice of DO membrane material for different applications
Easy-to-use, screw-on cap DO membranes
User-upgradable software from YSI website
Three-year warranty on the instrument; one-year on the probe
modules
Available with 4,10, and 20 m cable lengths
Stores over 49,000 data sets, time and date stamped
YSI Incorporated YSI 556 MPS Page 7
TM
for WindowsTM data analysis
General Information Section 2
Auto temperature compensating display contrast
Optional barometer
Optional rechargeable battery pack or standard alkaline batteries
2.2 Unpacking the Instrument
1. Remove the instrument from the shipping box. Note that the probe module and sensors are shipped in a separate box and will be unpacked later in Section 3.2 Unpacking the Probe Module.
NOTE: Do not discard any parts or supplies.
2. Use the packing list to ensure all items are present.
3. Visually inspect all components for damage.
NOTE: If any parts are missing or damaged, contact your YSI Service Center immediately. Refer to Appendix E Customer Service or www.ysi.com.
YSI 556 MPS YSI Incorporated
Page 8
General Information Section 2
2.3 Features of the YSI 556 Multi-Probe System
On/Off Key
Arrow Keys
Escape Key
Display
Backlight Key
Enter Key
Alpha/Numeric Keys – Used to enter letters and numbers
Cable Connector
Figure 2.1 Front View of YSI 556 MPS
YSI Incorporated YSI 556 MPS Page 9
General Information Section 2
Accessory Mounting Holes
Barometer Vent Patch
Lid
Battery Lid Screws
Figure 2.2 Back View of YSI 556 MPS
2.4 Batteries
2.4.1 Battery Life
Standard Alkaline Batteries
With the standard battery configuration of 4 alkaline C cells, the YSI 556 MPS will operate continuously for approximately 180 hours. Assuming a standard usage pattern when sampling of 3 hours of “on time” in a typical day, the alkaline cells will last approximately 60 days.
Optional Rechargeable Battery Pack
When fully charged, the optional rechargeable battery pack will provide approximately 50 hours of battery life.
YSI 556 MPS YSI Incorporated
Page 10
General Information Section 2
2.4.2 Inserting 4 C Batteries
Figure 2.3 Inserting C Cells
CAUTION: Install batteries properly to avoid damage to the
instrument.
1. Loosen the four screws in the battery lid on the back of the
instrument using any screwdriver.
2. Remove the battery lid.
3. Insert four C batteries between the clips following the
polarity (+ and -) labels on the bottom of the battery compartment.
4. Check gasket for proper placement on the battery lid.
5. Replace the battery lid and tighten the 4 screws securely and
evenly.
NOTE: Do not over-tighten the screws.
YSI Incorporated YSI 556 MPS Page 11
General Information Section 2
2.4.3 Inserting Optional Rechargeable Battery Pack
Figure 2.4 Inserting Battery Pack
CAUTION: Read all cautions and warnings that come with
the battery pack before using the battery pack.
1. Loosen the four screws in the battery lid on the back of the instrument using any screwdriver.
2. Remove the C battery lid and store for future use. Remove C batteries, if installed.
3. Check for proper placement of gasket on the rechargeable battery pack and lid.
4. Install the rechargeable battery pack and lid and tighten the 4 screws securely and evenly.
NOTE: Do not over tighten the screws.
YSI 556 MPS YSI Incorporated
Page 12
General Information Section 2
(
)
2.4.4 Charging the Optional Rechargeable Battery Pack
Charger adapter cable (YSI 6119)
Wall power supply
YSI 6114
Figure 2.5 Charging the Battery Pack
CAUTION: Do not use or store the battery pack at extreme
temperatures such as in strong direct sunlight, in cars during hot weather or close to heaters.
1. Install the rechargeable battery pack into the instrument as
described in Section 2.4.3 Inserting Optional Rechargeable Battery Pack.
2. Attach the charger adapter cable (YSI 6119) to the
instrument.
NOTE: Wall power supplies for use in countries outside the US and Canada can be found in Appendix B Instrument Accessories.
3. Insert the barrel connector of the wall power supply into the
barrel of the adapter cable.
CAUTION: Do not charge the battery pack continuously for
more than 48 hours.
CAUTION: Do not drop or expose to water.
CAUTION: Do not charge the battery pack at temperatures
below 0°°°°C or above 40°°°°C.
YSI Incorporated YSI 556 MPS Page 13
General Information Section 2
4. Plug the wall power supply into an AC power outlet for approximately 2 hours to obtain an 80% to 90% charge and for 6 hours to get a full charge.
NOTE: The battery pack can be recharged whether the instrument is on or off.
2.4.5 Storing the Battery Pack
Remove the battery pack from the instrument when the instrument will not be used for extended periods of time to prevent over discharge of the battery pack.
Store the battery pack in a heavy plastic bag to prevent accidental shorting of the terminals. Store between –20 and 30°C.
2.4.6 Optional Cigarette Lighter Charger
CAUTION: Read all warnings and cautions that come with
the charger before using the charger.
CAUTION: Only use cigarette lighter charger when
rechargeable battery pack is inserted into instrument.
CAUTION: Do not mishandle cigarette lighter charger. Do
not expose to moisture.
1. Plug the barrel connector of the cigarette lighter charger into the mating end of the YSI 6119 Charger Adapter Cable.
2. Attach the MS-19 end of the YSI 6119 Charger Adapter Cable to the instrument.
3. Make one of the following modifications to the other end of the charger:
Slide the adapter ring off the plug to use the device with an American or Japanese vehicle.
YSI 556 MPS YSI Incorporated
Page 14
General Information Section 2
Figure 2.6 Charger Plug Adapter Use
Leave the adapter ring on the plug and position it so that the slots on the adapter ring line up with the plug’s spring clips to use the device on a European vehicle.
Figure 2.7 European Charger Plug Adapter Use
NOTE: If the charger stops working properly, refer to
Section 13 Troubleshooting.
2.5 Power On
Press and release the on/off button in the upper left corner of the instrument keypad to turn the instrument on or off. See Figure
2.1 Front View of YSI 556 MPS.
2.6 Setting Display Contrast
The display contrast automatically compensates for temperature changes. However, under extreme temperature conditions you may wish to optimize the display by manual adjustment as follows:
YSI Incorporated YSI 556 MPS Page 15
General Information Section 2
1. Press and hold down the backlight key in the upper right corner of the keypad and press the “up” arrow to increase (darken) the contrast.
2. Press and hold down the backlight key in the upper right corner of the keypad and press the “down” arrow to decrease (lighten) the contrast.
2.7 Backlight
Press and release the backlight key in the upper right corner of the keypad to turn the backlight on or off. See Figure 2.1 Front View of YSI 556 MPS.
NOTE: The backlight turns off automatically after two minutes of non-use.
2.8 General Screen Features
Figure 2.8 Main Menu Screen
Main Display
Status Bar
Barometer Reading (optional)– Updated in real time, not corrected to sea level
Battery Charge – NiMH label indicates use of optional rechargeable battery pack, pulsing indicates that battery is charging, flashing indicates batteries almost exhausted
Current Time
Current Date
YSI 556 MPS YSI Incorporated
Page 16
General Information Section 2
2.9 Keypad Use
On/Off Key
Escape Key – Use to return to previous position in menu
Alpha/Numeric Keys – Used to enter letters and numbers
Minus/Hyphen (-) Key
Figure 2.9 Keypad Features
KEY LETTER/ NUMBER
1 1 2 ABC2abc3 3 DEF3def3 4 GHI4ghi4 5 JKL5jkl5 6 MNO6mno6 7 PQRS7pqrs7 8 TUV8tuv8 9 WXYZ9wxyz9 0 0
Backlight/Contrast
Arrows Keys
Enter Key
Period/Decimal Point Key
Figure 2.10 Keypad Letters & Numbers
1. See Figure 2.10 Keypad Letters & Numbers and press the
appropriate key repeatedly until letter or number desired appears in display.
NOTE: Press the key repeatedly in rapid succession to get to the desired letter or number. If you pause for more than a
YSI Incorporated YSI 556 MPS Page 17
General Information Section 2
second, the cursor automatically scrolls to the right to prepare for the next input.
EXAMPLE 1: Press the 6 key once and release to display an uppercase “M.”
EXAMPLE 2: Press the 6 key four times and release to display the number “6.”
EXAMPLE 3: Press the 6 key five times and stop to display a lowercase “m.”
2. Press the left arrow key to go back and reenter a number or letter that needs to be changed.
3. Press the Enter key when your entry is complete.
NOTE: The instrument software permits only numeric
entries in many instances, such as when setting the clock or entering calibration parameters.
2.10 Instrument Reset
The YSI 556 MPS is characterized by sophisticated software that should provided trouble-free operation. However as with all high-capability software packages, it is always possible that the user will encounter circumstances in which the instrument does not respond to keypad entry. If this occurs, the instrument function can easily be restored by removing and then reapplying battery power. Simply remove either your C-cells or rechargeable battery pack from the battery compartment, wait 30 seconds and then replace the batteries. See Section 2.4 Batteries for battery removal/reinstallation instructions.
YSI 556 MPS YSI Incorporated
Page 18
General Information Section 2
2.11 Menu Flowchart
Run
Report
Sensor
Calibrate
File
Logging Setup
System Setup
Log One Sample
Start Logging
Temperature
Conductivity
Dissolved Oxygen
pH
ORP
DIrectory
Upload to PC
View File
File Memory
Delete All Files
Ver sion
Date & Time
Data Filter
Shut Off Time
Comma Radix
ID
GLP Filename
TDS Constant
Barometer Units
Calibrate Barometer
Temperature
Specific Conductance
Conductivity
Resistivity
TDS
Salinity ppt
DO % saturation
DO mg/L
pH
pH mv
ORP mv
Conductivity
DO
pH
ORP
Interval
Use Site List
Store Barometer
Store Site Number
Edit Site List
YSI Incorporated YSI 556 MPS Page 19
General Information Section 2
YSI 556 MPS YSI Incorporated
Page 20
3. Probe Module
3.1 Introduction
The YSI 5563 Probe module is used for measuring dissolved oxygen, temperature, conductivity, and optional pH and ORP. The probe module is rugged, with the sensors enclosed in a heavy duty probe sensor guard with attached sinking weight. A 4, 10 or 20 meter cable is directly connected to the probe module body making it waterproof. An MS-19 connector at the end of the cable makes the YSI 5563 fully compatible with the YSI 556 Multi-Probe System.
3.2 Unpacking the Probe Module
1. Remove the YSI 5563 Probe module from the shipping
boxes.
NOTE: Do not discard any parts or supplies.
2. Use the packing list to ensure all items are present.
3. Visually inspect all components for damage.
NOTE: If any parts are missing or damaged, contact your YSI Service Center immediately. Refer to Appendix E Customer Service or www.ysi.com.
YSI Incorporated YSI 556 MPS Page 21
Probe Module Section 3
3.3 Features of the YSI 5563 Probe Module
Cable
Dissolved Oxygen (DO) Probe
Transport/ Calibration Cup
Figure 3.1 Probe Module
3.4 Preparing the Probe Module
To prepare the probe module for calibration and operation, you need to install the sensors into the connectors on the probe module bulkhead. In addition to sensor installation, you need to install a new DO membrane cap.
Strain Relief
Metal Probe Connector Nut
pH/ORP Probe
Conductivity/ Temperature Probe
Probe Sensor Guard
3.4.1 Sensor Installation
Whenever you install, remove or replace a sensor, it is extremely important that the entire probe module and all sensors be thoroughly dried prior to the removal of a sensor or a sensor port plug. This will prevent water from entering the port. Once you remove a sensor or plug, examine the connector inside the probe module sensor port. If any moisture is present, use compressed air to completely dry the connector. If the connector is corroded, return the probe module to your dealer or directly to YSI Customer Service. Refer to Appendix E Customer Service.
YSI 556 MPS YSI Incorporated
Page 22
Loading...
+ 132 hidden pages