Y S I Environmental
Water quality handhelds for field work
Oxygen, Conductivity, Salinity, pH, Temperature
• Simple push-button calibration
• Built-in probe storage chamber
• Backlit display for easy viewing in any light
• Convenient multiparameter probe designs
• Automatic temperature compensation
• Water-resistant to IP-65 standard
• Two-year instrument warranty
YSI 55
oxygen & temperature
• Dissolved oxygen in mg/L or % saturation
• Salinity compensation
• Industry standard
• Easy calibration
YSI Model 55 and
Model 85 (below)
YSI Models 30, 60,
and 63
Pure
Data
for a
Healthy
Planet.
Convenient
field measurement.
®
YSI 85
oxygen, conductivity, salinity, & temperature
• 50-reading memory with ID
• User-selected ranges or autoranging for conductivity
• Automatic salinity compensation
• Screw-on DO membranes
YSI 60
pH & temperature
• 50-reading memory with ID
• Replaceable pH electrode with one-year warranty
• Cable lengths up to 100 feet
YSI 63
pH, conductivity, salinity, & temperature
• 50-reading memory with ID
• Four-electrode conductivity cell
• User-selected ranges or autoranging for conductivity
• Replaceable pH electrode with one-year warranty
• Cable lengths up to 100 feet
YSI 30 & 30M
conductivity, salinity, & temperature
• Four-electrode conductivity cell
• User-selected ranges or autoranging for conductivity
• YSI 30M has 50-reading memory with ID (available with 100-foot
cables)
www.YSI.com
To order, or for more
information, contact
YSI Environmental.
+1 937 767 7241
800 897 4151 (US)
www.ysi.com
Y S I Environmental
55 oxygen & temperature (System Specifications)
Salinity compensation: 0 to 40 ppt (manual)
Altitude compensation: 0 to 10,000 feet
Cable lengths: 12, 25, 50 feet (3.7, 7.6, 15.2 meters), available with detachable cable
Measurement Range Resolution Accuracy (meter & probe)
Dissolved Oxygen 0-200% 0.1% ±2% saturation
0-20 mg/L 0.01 mg/L ±0.3 mg/L
Temperature -5 to +45°C 0.1°C ±0.2°C
YSI Environmental
+1 937 767 7241
Fax +1 937 767 9353
environmental@ysi.com
Endeco/YSI
+1 508 748 0366
Fax +1 508 748 2543
systems@ysi.com
SonTek/YSI
+1 858 546 8327
Fax +1 858 546 8150
inquiry@sontek.com
YSI Environmental Gulf Coast
+1 225 753 2650
Fax +1 225 753 8669
environmental@ysi.com
YSI Hydrodata (UK)
+44 1462 673 581
Fax +44 1462 673 582
europe@ysi.com
YSI Middle East (Bahrain)
+973 1753 6222
Fax +973 1753 6333
halsalem@ysi.com
YSI (Hong Kong) Limited
+852 2891 8154
Fax +852 2834 0034
hongkong@ysi.com
YSI (Qingdao) Limited
+86 532 575 3636
Fax +86 532 571 0101
ysiqd@ysiqd.com.cn
YSI Nanotech (Japan)
+81 44 222 0009
Fax +81 44 221 1102
nanotech@ysi.com
ISO 9001
ISO 14001
Pure Data for a Healthy Planet and
Who’s Minding the Planet? are registered
trademarks of YSI Incorporated.
©2006 YSI Incorporated
Printed in USA 0107 W42-03
Y S I i n c o r p o r a t e d
Whoʼs Minding
the Planet?
™
85 oxygen, conductivity, salinity & temperature (System Specifications)
Range: Autoranging for dissolved oxygen. User-selected or autoranging for conductivity
Adjustable conductivity reference temperature: 15 to 25°C
Adjustable conductivity temperature compensation factor: 0 to 4%
Salinity compensation: 0 to 80 ppt (automatic)
Altitude compensation: 0 to 10,000 feet
Cable lengths: 10, 25, 50, 100 feet (3, 7.6, 15.2, 30.5 meters), available with detachable cable
Measurement Range Resolution Accuracy (meter & probe)
Dissolved Oxygen 0 to 200% 0.1% ±2% saturation
0 to 20 mg/L 0.01 mg/L ±0.3 mg/L
Conductivity 0 to 499.9 µS/cm 0.1 µS/cm ±0.5% full scale
0 to 4999 µS/cm 1 µS/cm ±0.5% full scale
0 to 49.99 mS/cm 0.01 mS/cm ±0.5% full scale
0 to 200.0 mS/cm 0.1 mS/cm ±0.5% full scale
Salinity 0 to 80 ppt 0.1 ppt ±2% or ±0.1 ppt
Temperature -5 to +65°C 0.1°C ±0.1°C (±1 lsd)
-5 to +45°C for oxygen
60/63 pH & pH/conductivity & temperature (System Specifications)
pH electrode: Combination with gel reference
Range: Autoranging for pH, user-selected or autoranging for conductivity
Temperature compensation: Automatic
Adjustable conductivity reference temperature: 15 to 25°C
Adjustable conductivity temperature compensation factor: 0 to 4%
Cable lengths: 10, 25, 50, 100 feet (3, 7.6, 15.2, 30.5 meters)
Measurement Range Resolution Accuracy (meter & probe)*
pH 0 to 14 0.01 unit ±0.1 unit within 10°C of calibration
±0.2 unit within 20°C
Conductivity 0 to 499.9 µS/cm 0.1 µS/cm ±0.5% full scale
0 to 4999 µS/cm 1 µS/cm ±0.5% full scale
0 to 49.99 mS/cm 0.01 mS/cm ±0.5% full scale
0 to 200.0 mS/cm 0.1 mS/cm ±0.5% full scale
Salinity 0 to 80 ppt 0.1 ppt ±2% or ±0.1 ppt
Temperature -5 to +75°C 0.1°C ±0.1°C (±1 lsd)
*Meter accuracy is ±0.02 unit.
30 salinity, conductivity & temperature (System Specifications)
Range: User-selected or autoranging
Adjustable reference temperature: 15 to 25°C
Adjustable temperature compensation factor: 0 to 4%
Cable lengths: 10, 25, 50, 100 feet (3, 7.6, 15.2, 30.5 meters)
Measurement Range Resolution Accuracy (meter & probe)
Conductivity 0 to 499.9 µS/cm 0.1 µS/cm ±0.5% full scale
0 to 4999 µS/cm 1 µS/cm ±0.5% full scale
0 to 49.99 mS/cm 0.01 mS/cm ±0.5% full scale
0 to 200.0 mS/cm 0.1 mS/cm ±0.5% full scale
Salinity 0 to 80 ppt 0.1 ppt ±2% or ±0.1 ppt
Temperature -5 to +95°C 0.1°C ±0.1°C (±1 lsd)