The YSI Model 5239 probe is used for measuring dissolved
oxygen in the field. It is the first field probe that features cap
membranes for easy membrane replacement (YSI 5906
membrane caps). The probe is rugged, with the DO and
temperature sensors enclosed in a heavy duty probe guard. A
10, 25 or 50 foot cable is directly connected to the probe body
making it waterproof. A 5-pin MS connector at the end of the
cable makes the 5239 fully compatible with all YSI 50 series
DO meters retaining the accuracy of the system. The 5239 has
been optimized for YSI meters which have a micro-processor
for sampling and processing data, such as YSI Models 50B
and 52. The system specifications, when used with a YSI 50
Series DO meter, are as follows:
P
ROBE CONFIGURATION
• Cathode: Gold
• Anode: Silver
• Membrane: FEP Teflon
• Electrolyte: YSI standard probe solution (Half
saturated KCl)
• Polarizing voltage: -0.8 V vs. the potential of the Ag
anode/reference electrode
S
YSTEM SPECIFICATIONS
• Temperature range: -5° to 45° C
• Temperature accuracy: ±0.2° C
• Temperature response time: 30 sec for 95% of change.
1
Page 6
• DO range: 0 to 200 % air saturation or 0 to 20 mg/L.
• DO accuracy: ±2 % or ±0.2 mg/L whichever is greater.
• DO response time: 30 sec for 95% of change at 25°C.
T
EMPERATURE COMPENSATION
Temperature compensation of DO readings is provided by the
YSI DO meter. For best results, calibrate within ±5° C of the
sample temperature.
ROBE OPERATION
P
NOTE: The 5239 probe is shipped dry. A shipping membrane
was installed to protect the electrodes.
A new membrane cap
must be installed before first use.
M
EMBRANE INSTALLATION
Follow the steps below to install a membrane cap:
1. Unscrew and remove the probe guard. See Figure 1.
2. Unscrew the old membrane cap and remove it from the
probe.
3. Thoroughly rinse the entire area of the silver anode,
gold cathode and threads with clean water (distilled or
deionized) to remove any dry crystals of the old
electrolyte.
4. Hold the membrane cap and fill it with 6 to 7 drops of
YSI standard electrolyte (KCl probe solution)
provided.
2
Page 7
5. Tap the bottom of the membrane cap with your finger a
few times (don’t touch the membrane) to shake loose
any trapped air bubbles.
6. Screw the membrane cap onto the probe moderately
tight. A small amount of electrolyte should overflow.
7. Rinse off the excess electrolyte with clean water.
8. Screw the probe guard onto the probe body.
9. Place the probe in the calibration bottle provided.
Verify that sponge in the bottle is wet.
Fill membrane cap
with 6-7 drops of
KCl solution.
Unscrew guard
Unscrew cap
Screw cap on
moderately tight
Screw guard on
moderately tight
Tap cap with finger
to remove bubbles.
Figure 1
C
ALIBRATION
Dissolved oxygen calibration must be done in an environment
with a known oxygen content. Three such environments will
be discussed here: calibration in air (water-saturated),
calibration in air-saturated water and calibration by Winkler
3
Page 8
titration. Choose the
Calibration in air is the simplest and most accurate method of
calibration.
which best fits your application.
one
• For
• For
• To calibrate
YSI recommends that the following procedure to calibrate the
5239 probe:
air calibration
humidity. To achieve this, place the probe in the plastic
calibration bottle (provided) with a wet sponge or paper
towel in the bottom.
air-saturated water calibration
volume of water (300 to 500 ml) by aerating for at least 15
minutes at a relatively constant temperature. Place the
probe in the aerated water and provide adequate stirring (at
least 1 foot per second) while calibrating.
to a Winkler-titrated sample
dissolved oxygen value of a sample by Winkler titration.
Place the probe in the sample and provide adequate stirring
(at least 1 foot per second) while calibrating.
1. Connect the 5239 probe to one of the YSI 50 series DO
meters.
2. Turn on the meter and wait for the probe to polarize
and the reading to stabilize. Depending on the
condition of the probe, the time for stabilization could
vary between 15 and 30 minutes. For a new probe with
clean electrodes and a properly installed membrane
cap, it usually takes 15 to 20 minutes. If the electrodes
are fouled or the membrane or electrolyte are not in
good condition, it may take longer to stabilize than
anticipated.
, place the probe in air at 100% relative
, air-saturate a
, determine the
4
Page 9
3. Check the membrane to make sure there are no water
beads attached to it. Wipe to remove the water beads
with a soft tissue if necessary.
NOTE: Make sure that the probe has been in the
calibration bottle for more than 2 minutes before
calibrating so that both the thermistor and DO
sensor are at the same temperature.
4. Calibrate the system according to the instructions in
the operations manual for the particular YSI 50 series
meter you are using. Be sure to take into consideration
the local barometric pressure and adjust to the proper
calibration value. All YSI 50 series DO meters have a
table on the back which relates barometric pressures to
%-air saturation values.
Checking the Probe Zero
The probe zero (background current) is checked by immersing
the probe in a sodium sulfite solution (0.08M or 3g
Na2SO3/300 ml), or in an inert gas (e.g. nitrogen, argon). The
meter should read less than 1% dissolved oxygen in either of
these environments. If it does not, change the membrane or
clean the probe.
E
RRONEOUS READINGS
Erratic readings will result from loose, wrinkled or fouled
membranes, or from bubbles larger than 1/8” in the electrolyte
solution. Erroneous readings may occur if the membrane is
coated with oxygen consuming bacteria or oxygen evolving
algae. Heavy residue may coat the membrane causing
incorrect readings. Frequent membrane changes will eliminate
this problem.
5
Page 10
I
NTERFERENCES
Hydrogen sulfide, sulfur dioxide, halogens, carbon monoxide,
chlorine, nitric oxide, and nitrous oxide can cause the probe to
give erroneous readings. If you suspect erroneous readings, it
may be necessary to determine if these are the cause.
A
CIDS
Avoid any environment that contains substances such as
concentrated acids, caustics, and strong solvents that may
attack the probe. Probe materials that may be damaged by
these substances include FEP Teflon, EPR rubber, ABS
plastic, and stainless steel.
S
TORAGE
When the probe is not in use, store the probe in a humid
environment, such as the plastic calibration bottle (with a wet
sponge or paper towel inside) or a BOD bottle containing at
least 1 inch of water (remove probe guard first).
6
Page 11
RINCIPLES OF OPERATION
P
T
HE CLARK OXYGEN SENSOR
The 5239 probe is a Clark type polarographic (voltammetric)
dissolved oxygen sensor. The sensor is made of a silver anode
and a gold cathode (see Figure 2), and is separated from the
measured medium by a semi-permeable Teflon membrane.
The temperature sensing element (thermistor assembly) is
mounted next to the oxygen sensor vertically, providing
temperature readings for the DO system.
Temperature sensor
Anode (silver)
Cathode (gold)
Figure 2
The membrane selectively allows oxygen to permeate into the
sensor, but prevents most interfering molecules and fouling
materials from entering. Upon permeating through the
membrane, oxygen is reduced at the gold cathode. The current
resulting from this reduction is diffusion-limited and is
proportional to the partial pressure of oxygen in the sample.
The counter reaction is the oxidation of silver at the
7
Page 12
anode/reference electrode which completes the overall
electrolytic reaction in the chloride medium (KCl electrolyte)
behind the membrane. These reactions, at the cathode and the
anode, are as follows:
Cathode reaction:O
+ 2H2O + 4e- ==> 4OH
2
Anode reaction:Ag + Cl
T
DO R
HE
EADINGS FROM THE CATHODE REDUCTION
-
==> AgCl
-
The oxygen reduction current is sampled and processed, by the
meter, and displayed as either %-air saturation or mg/L. While
the parameter of %-air (partial pressure) is independent of
temperature and salinity, mg/L (solubility of oxygen) is a
function of temperature and salinity. For instance, the same %air reading (same partial pressure) would give a higher mg/L
reading at a lower temperature than at a higher temperature.
Also, the higher the salinity, the lower the solubility for the
same %-air reading at the same temperature.
T
HE FORMATION OF
AgCl
AT THE ANODE
While the oxygen reduction current passes through the internal
circuit to be reported as the DO reading, it also passes through
the anode oxidizing the silver and forming a thin layer of
silver chloride. This oxidation of silver at the chloride medium
provides a stable potential which the cathode potential is
referenced to (for instance, the polarization potential of the
cathode is -0.8 V versus the potential of the Ag/AgCl redox
couple at the silver anode). As more and more silver chloride
8
Page 13
(an insoluble solid and poor conductor of electricity)
accumulates at the anode, it begins to block the passage of
current in the electrolytic cell. The silver anode must be
cleaned periodically to prevent this thick layer of silver
chloride from reducing the sensitivity of the sensor.
T
HE FUNCTION OF THE ELECTROLYTE
There are two main functions for the electrolyte:
1. Supply the chloride (Cl-) to the anode/reference
electrode for the counter reaction of the oxygen
reduction at the cathode.
2. Provide the ionic conduction of electricity inside the
cell, especially in the thin layer between the gold
cathode and the membrane.
Under normal operating conditions, such as measuring oxygen
around 100 %-air (8.27 mg/L) at 25° C, the strength of the
chloride concentration should last up to 500 hours before the
chloride becomes a limiting factor in the operation of the
oxygen sensor. 500 hours translates into about 62.5 working
days at 8 hours per day operation.
9
Page 14
AINTENANCE AND TROUBLESHOOTING
M
M
EMBRANE CAP LIFE
If the probe is properly maintained, one membrane cap should
last two to four weeks depending on how often the probe is
used and the type of samples measured. Keep the probe in a
moist atmosphere as much as possible to avoid drying out the
electrolyte. Erratic readings will result from loose, wrinkled or
fouled membranes, or from bubbles larger than 1/8” in the
electrolyte solution. The following things are likely to cause
short membrane life:
1. The electrolyte has dried out because the membrane
cap was not tight or the probe was exposed to dry air
for a long time.
2. The membrane was not properly installed. For
instance, wrinkles have formed after membrane
installation.
3. The probe was not properly rinsed with water after a
field measurement, allowing mud or algae to attach to
the membrane.
4. The membrane was damaged by an object, such as
reeds or small sticks in the water.
5. The probe has struck a hard object and the sensor or
membrane has been damaged.
10
Page 15
E
LECTRODE CLEANING
It is always a good practice to keep the probe in good
operating condition.
Silver Anode
After extended use, a thick layer of AgCl builds up on the
silver anode reducing the sensitivity of the sensor. The anode
must be cleaned to remove this layer and restore proper
performance. The cleaning can be chemical or mechanical:
•
Chemical cleaning:
Remove the membrane cap and
soak the entire anode section in a 14% ammonium
hydroxide solution for 2 to 3 minutes, followed by a
thorough rinsing with distilled or deionized water.
The anode should then be thoroughly wiped with a
wet paper towel to remove the residual layer from the
anode.
•
Mechanical cleaning:
Sand off the dark layer from
the silver anode with 400 grit wet/dry sandpaper.
Wrap the wet sandpaper around the anode and twist
the probe. Rinse the anode with clean water after
sanding, followed by wiping thoroughly with a wet
paper towel.
Gold Cathode
For correct probe operation, the gold cathode must be textured
properly. It can become tarnished or plated with silver after
extended use. The gold cathode can be cleaned by using the
adhesive backed sanding disc and tool provided in the YSI
5238 probe reconditioning Kit.
11
Page 16
Stick the disc to the sanding tool, then wet sand the gold with
a twisting motion about 3 times or until all silver deposits are
removed and the gold appears to have a matte finish. If the
cathode remains tarnished, return the probe for service.
Wipe the gold cathode thoroughly with a wet paper towel
before putting on a new membrane cap.
T
ROUBLE SHOOTING GUIDE
Below is a quick reference of symptoms with steps to follow
to cure these symptoms.
SymptomsActions
Unable to calibrate• Extend the stabilization time from 15 min.
to 30 min. before calibrating
• Change membrane
• Check for loose or wrinkled membrane or
deposits between gold cathode and
membrane, especially after changing
membrane
• Clean probe (see instructions under
Electrode Cleaning)
12
Page 17
SymptomsActions
Does not hold
calibration
• Allow sufficient time for temperature and
oxygen stabilization before calibration
• Check for water droplets on membrane
• Check membrane for a pinhole or split
• Change membrane
• Clean probe (see instructions under
Electrode Cleaning)
Unstable readings• Check membrane for a pinhole or split
• Allow 60 sec. for reading to stabilize
• Change membrane
• Clean probe (see instructions under
Electrode Cleaning)
Membrane cap is
tight
S
ERVICING
• Use pliers to loosen the cap (turn counter
clockwise while facing sensor tip)
The 5239 probe has only two replaceable parts (other than the
membrane cap), the probe guard and the 5-pin MS connector
on the end of the cable. The cable is permanently attached to
the probe and cannot be replaced.
YSI part numberDescription
1102755-pin MS connector kit
059956Probe guard
5238Probe reconditioning kit
13
Page 18
ARRANTY AND REPAIR
W
YSI Model 5239 Dissolved Oxygen Probes are w arran t ed f or on e year from
date of purchase by the en d u s er again s t def ects in materials and
workmansh ip. Within th e warranty period, YSI will repair or replace, at its
sole discretion, free of charge, any product that YSI determines to be cov ered
by this warranty.
To exercise this warranty, write or call your local YSI representative, or
contact YSI Customer Service in Yellow Springs, Ohio. Send the produ ct an d
proof of purchase, tran sport at ion prepai d, to the Authorized Serv ice C en t er
selected by YSI. Repair or replacement will be made and the product
returned, transportation prepaid. Repaired or replaced products are warranted
for the balance of the ori g in al warranty period, or at l eas t 90 days from date
of repair or replacement.
Limitation of Warranty
This Warranty does not appl y to any YSI product damage or failure cau s ed
by (i) failure to install, operate or us e th e produ ct in accordan ce with YSI’s
written instructions, (ii) abuse or misuse of the product, (iii) failure to
maintain the product in accordance with YSI’ s written instructions or
standard industry procedure, (iv) any i mproper repairs to the product, (v) use
by you of defective or improper components or parts in se rv ic in g or repai ri n g
the product, or (vi) modi f i cati on of the product in any way not ex pres sl y
authorized by YSI.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
YSI’s LIABILITY UNDER THIS WARRANTY IS LIMITED TO
REPAIR OR REPLACEMENT OF THE PRODUCT, AND THIS SHALL
BE YOUR SOLE AND EXCLUSIVE REMEDY FOR ANY DEFECTIVE
PRODUCT COVERED BY THIS WARRANTY. IN NO EVENT
SHALL YSI BE LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING
FROM ANY DEFECTIVE PRODUCT COVERED BY THIS
WARRANTY.
14
Page 19
YSI Authorized Service Centers
For the nearest authroized service center contact:
YSI Technical Support • 1725 Brannum Lane • Yellow Springs, OH • 45387
Phone: +1 937 767-7241 • Fax: 937 767-9353 • environmental@ysi.com • www.ysi.com
15
Page 20
C
LEANING INSTRUCTIONS
NOTE: Before they can be serviced, equipment exposed to
biological, radioactive, or toxic materials must be cleaned
and disinfected.
Biological contamination is presumed for
any instrument, probe, or other device that has been used with
body fluids or tissues, or with waste water. Radioactive
contamination is presumed for any instrument, probe or other
device that has been used near any radioactive source.
If an instrument, probe, or other part is returned or presented
for service without a Cleaning Certificate, and if in our
opinion it represents a potential biological or radioactive
hazard, our service personnel reserve the right to withhold
service until appropriate cleaning, decontamination, and
certification has been completed. We will contact the sender
for instructions as to the disposition of the equipment.
Disposition costs will be the responsibility of the sender.
When service is required, either at the user's facility or at YSI,
the following steps must be taken to insure the safety of our
service personnel.
In a manner appropriate to each device, decontaminate all
1.
exposed surfaces, including any containers. 70% isopropyl
alcohol or a solution of 1/4 cup bleach to 1 gallon tap
water are suitable for most disinfecting. Instruments used
with waste water may be disinfected with .5% Lysol if this
is more convenient to the user.
The user shall take normal precautions to prevent
2.
radioactive contamination and must use appropriate
decontamination procedures should exposure occur.
16
Page 21
If exposure has occurred, the customer must certify that
3.
decontamination has been accomplished and that no
radioactivity is detectable by survey equipment.
Any product being returned to the YSI Repair Center,
4.
should be packed securely to prevent damage.
Cleaning must be completed and certified on any product
5.
before returning it to YSI.
P
ACKING INSTRUCTIONS
Clean and decontaminate items to insure the safety of the
1.
handler.
Complete and include the Cleaning Certificate.
2.
Place the product in a plastic bag to keep out dirt and
3.
packing material.
Use a large carton, preferably the original, and surround
4.
the product completely with packing material.
Insure for the replacement value of the product.
5.
17
Page 22
Cleaning Certificate
Organization ________________________________
Department _________________________________
Address ___________________________________
City _______________ State ______ Zip ________
Country __________________
Model No. of Device ______ Lot Number _________
Contaminant (if known) _____________________
Cleaning Agent(s) used _____________________
Radioactive Decontamination Certified?
(Answer only if there has been radioactive exposure)
___ Yes ___ No
Cleaning Certified By _________________________
NameDate
18
Page 23
1725 Brannum Lane
Yellow Springs, Ohio 45387 USA052395 - Web
937 767-7241 • 800 765-4974 A52395B
Fax 937 767-9353September 99
Environmental@YSI.com • www.YSI.com
1997 YSI Incorporated
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.