YSI 52 User Manual

YSI Model 52
Dissolved Oxygen Meter
Operations Manual

CONTENTS

SECTION 1 INTRODUCTION................................................................................................................. 1
1.1 PRINCIPLES
1.2 DISPLAY
1.3 FUNCTION
1.4 POWER
OF OPERATION...................................................................................................... 1
AND FUNCTION KEYS............................................................................................... 2
SWITCH ..................................................................................................................... 2
ON SELF-TEST................................................................................................................ 2
SECTION 2 PREPARING THE METER ............................................................................................... 3
SECTION 3 CALIBRATION ................................................................................................................... 5
PERCENT AND MG/L.................................................................................................................... 5
3.1
3.2 DETERMINING
3.3 CALIBRATION
3.4 CALIBRATION
3.5 CALIBRATION
3.6 ZERO
CALIBRATION ................................................................................................................... 8
3.7 RESETTING
CALIBRATION VALUE....................................................................................5
IN AIR.................................................................................................................. 6
IN AIR-SATURATED WATER.......................................................................... 7
BY WINKLER TITRATION............................................................................... 7
THE DEFAULT ZERO CALIBRATION ............................................................... 10
SECTION 4 SALINITY........................................................................................................................... 11
SECTION 5 OPERATION...................................................................................................................... 13
5.1 TEMPERATURE
5.2 DISSOLVED
5.3 MEASURING
5.4 RECORDING
5.5 REVIEWING
5.6 PRINTING
5.7 SENDING
5.8 ERASING
MEASUREMENTS ......................................................................................... 13
OXYGEN MEASUREMENTS .............................................................................. 13
OXYGEN IN FLUIDS OTHER THAN WATER.................................................. 13
TO INTERNAL MEMORY................................................................................... 14
RECORDED LIVE DATA..................................................................................... 15
DATA FROM MEMORY .......................................................................................... 16
RECORDED DATA TO A PERSONAL COMPUTER .............................................. 16
THE MODEL 52 MEMORY ....................................................................................... 17
SECTION 6 SYSTEM SETUP................................................................................................................18
6.1 SETTING
6.2 SETTING
6.3 SETTING
6.4 LAST
6.5 CONTRAST
THE TIME AND DATE............................................................................................... 18
UP AUTOSTABLE...................................................................................................... 18
COMMUNICATION PARAMETERS ........................................................................ 20
DIGIT SUPPRESSION....................................................................................................... 21
ADJUSTMENT........................................................................................................ 21
SECTION 7 PERSONAL COMPUTER INTERFACE........................................................................ 23
7.1 WRITING
YOUR OWN APPLICATIONS SOFTWARE ............................................................ 23
SECTION 8 MAINTENANCE AND TROUBLESHOOTING............................................................ 25
8.1 BATTERIES.................................................................................................................................. 25
8.2 POWER
ON SELF-TESTING AND ERROR DISPLAY MESSAGES ........................................ 25
SECTION 9 DISCUSSION OF MEASUREMENT ERRORS............................................................. 27
9.1 TYPE
9.2 TYPE
9.3 TYPE
9.4 ERROR
1 ERRORS INSTRUMENT COMPONENTS.................................................................. 27
2 ERRORS PROBE ACCURACY ................................................................................... 27
3 ERRORS - CALIBRATION............................................................................................. 28
CALCULATION EXAMPLE.......................................................................................... 28
SECTION 10 WARRANTY AND REPAIR ...................................................................................... 31
i
AUTHORIZED CLEANING PACKING
U.S. SERVICE CENTERS............................................................................................ 32
INSTRUCTIONS............................................................................................................... 34
INSTRUCTIONS.................................................................................................................. 34
APPENDIX A GENERAL SPECIFICATIONS................................................................................... 37
APPENDIX B REQUIRED NOTICE ................................................................................................... 39
APPENDIX C OXYGEN SOLUBILITY.............................................................................................. 41
APPENDIX D CALIBRATION TABLES............................................................................................ 43
ii

SECTION 1 INTRODUCTION

The YSI Model 52 is a microprocessor-based instrument designed for field and laboratory measurement of dissolved oxygen. Readings are automatically compensated for temperature. When calibrated to percent air saturation, the display simultaneously shows dissolved oxygen in mg/L and in percent air saturation as well as temperature in degrees Celsius.
An RS232 port permits results to be sent directly to your serial printer or personal computer. Salinity compensation can be adjusted manually. To assure the highest accuracy, an auto stable feature indicates when the readings have reached user-defined criteria for stability. To assure correct operation, the meter performs a self-testing routine each time it is turned on; any error is then signaled on the display.
Calibration is quick and easy. Procedures for calibrating with and without compensation for temperature, altitude, pressure, and salinity are described in this manual. The YSI Model 52 is also equipped with non-volatile memory, capable of storing up to 70 DO and temperature readings. Any data stored in internal memory can be reviewed on the LCD, or batch uploaded to a PC or an RS232 printer.
Power can be provided by an AC adapter (optional), or batteries that permit field use.

1.1 PRINCIPLES OF OPERATION

YSI dissolved oxygen probes use membrane covered, Clark-type polarographic sensors with built-in thermistors for temperature measurement and compensation. A thin semi permeable membrane, stretched over the sensor, isolates the sensor elements from the environment, but allows oxygen and certain other gases to enter. When a polarizing voltage is applied across the sensor, oxygen that has passed through the membrane reacts at the cathode, causing a current to flow.
Oxygen diffuses through the membrane at a rate proportional to the pressure difference across it. Since oxygen is rapidly consumed at the cathode, it can be assumed that the oxygen pressure inside the membrane is zero. Hence, the amount of oxygen diffusing through the membrane is proportional to the absolute pressure of oxygen outside the membrane. If the oxygen pressure increases, more oxygen diffuses through the membrane and more current flows through the sensor. A lower pressure results in less current.
YSI, Incorporated Model 52 1
Introduction Section 1

1.2 DISPLAY AND FUNCTION KEYS

A rotary function switch and four keys provide complete control of the Model 52. The four keys are SKIP,¿ ,À , and CONFIRM. ¿ andÀ are used to modify a setting either up or down (for example, calibration or salinity values can be modified for each calibration process if required). The CONFIRM key is used to accept your setting.
The display often asks questions such as “Calibrate in percent?” Use CONFIRM to answer yes. Use SKIP to go on to the next choice. The SKIP key can also be used to select among several default values.

1.3 FUNCTION SWITCH

The rotary function switch allows easy selection among the meter’s basic functions. In O
-TEMP, the instrument is measuring dissolved oxygen and temperature. To begin
2
calibrating, switch to CALIBRATE. To begin setting the clock, communication, and auto-stable parameters, switch to SYSTEM SETUP. EXTERNAL CONTROL activates the RS232 port. Salinity can be set after switching to SALINITY.

1.4 POWER ON SELF-TEST

A Power On Self-Test (POST) takes place whenever the meter is turned on. This ensures correct performance. If anything is wrong, the operator is signaled and the error identified. An error message will appear on the display. For more information see Section
8.2, Power On Self-Testing and Error Display Messages.
YSI, Incorporated Model 52 2

SECTION 2 PREPARING THE METER

To prepare the meter for use, connect a prepared probe to the meter. Set the function switch to O Autostable. Since complete equilibration can take as long as 15 minutes after the meter is turned on, it is customary to leave the instrument on throughout the working day.
When the instrument is turned on, a beep signals the beginning of a Power On Self Test (POST). A series of displays will appear. A final beep and a return to the normal display signal the conclusion of the test. The whole process takes about 10 seconds. If the instrument detects an error, it will display an error message. These messages, with appropriate corrective actions, are listed in the Maintenance and Troubleshooting section of this manual. If you have a problem you cannot resolve, call YSI Technical Support, (937) 767-7241 or (800) 765-4974.
-TEMP. Observe temperature and oxygen readings for stability, or set up
2
6.33 mg/L 19.9C
Stirring: Since the oxygen level in the layer of liquid at the membrane surface is continuously depleted, it is essential that water movement of 1 foot per second or more be maintained while making dissolved oxygen measurements. When using the Model 52 in field applications, a moving stream will usually provide adequate motion, as will moving the probe through the sample by hand. For lab applications, we recommend one of the YSI self-stirring BOD probes.
When the temperature and oxygen readings have stabilized, you are ready to begin calibration.
88.1 % 11: 52
YSI, Incorporated Model 52 3
Preparing the Meter Section 2
YSI, Incorporated Model 52 4

SECTION 3 CALIBRATION

Calibration is accomplished by exposing the probe to a known concentration of oxygen and then adjusting the calibration value to match. Since the oxygen content of our atmosphere varies predictably according to relative humidity and atmospheric pressure, we recommend that you calibrate in air, see Section 3.3, Calibration in Air.
Daily calibration is usually appropriate. Calibration can be disturbed by a physical shock, touching the membrane, fouling of the membrane or drying out of the electrolyte. Check calibration after each series of measurements, and in time you will develop a realistic schedule for recalibration.

3.1 PERCENT AND MG/L

100 percent air saturation corrected to standard pressure is the dissolved oxygen value of water saturated with air at 760 mm Hg (1013 millibars or 29.92 inches Hg). The mg/L values describe the concentration of oxygen. Since calibrations in mg/L do not necessarily yield correct percent readings, the percent readings are disabled after calibrations in mg/L. Calibrations in percent are correct for both percent and mg/L readings.

3.2 DETERMINING CALIBRATION VALUE

Typically, local barometric pressure is not equal to "standard pressure." Therefore, unless you are calibrating with a known sample determined by a Winkler titration, you should determine the proper calibration point for the local barometric pressure. This corresponds to a reading from a mercury barometer. Do not use the pressure reported by the Weather Bureau that has been adjusted to sea level. The calibration value for a given barometric pressure can be read from the chart on the back of the meter, or the one in Appendix D of this manual.
The chart also lists calibration values according to altitude. These values are only approximate because they are derived from the average pressure at the given altitudes, and do not take pressure variations into account. The typical error resulting from calibration to altitude alone will be ±4% of your reading. Use altitude values only when you cannot determine the true barometric pressure.
YSI, Incorporated Model 52 5
Calibration Section 3

3.3 CALIBRATION IN AIR

Three separate calibration techniques are discussed here: calibration in air, calibration in air-saturated water, and calibration by Winkler titration. Choose the one technique which best fits your application. Calibration in air is the simplest and most accurate method of calibration.
To calibrate in air:
1. Place a prepared probe in air at 100% relative humidity. To achieve this, BOD probes
can be placed in a BOD bottle with approximately 1” of water. Field probes can be placed in the plastic calibration bottle with a moistened sponge, or calibrated with the 5075A Calibration Chamber.
2. Switch to CALIBRATE. The display will read:
Select percent calibration by pressing CONFIRM. A display similar to the one below will be shown.
3. Using ¿ and À, adjust the number to the calibration value determined from the altitude or pressure chart on the back of the Model 52. SKIP sets the value to 100.0%; and shows the following display:
you may wish to do this to bring the setting quickly to a value near the one you want, then continue with ¿ and À. Pressing SKIP again sets the value to 0.0% (see Zeroing the Probe). Pressing SKIP a third time brings it back to the last calibration value.
E n t e r c a l v a l u e
L a s t = 9 6. 4 %
E n t e r c a l v a l u e
S a t . = 1 0 0 %
C a l i b r a t e i n p e r c e n t ?
YSI, Incorporated Model 52 6
Calibration Section 3
4. When you have adjusted the reading to the desired value (let’s say you chose 98.0%), press CONFIRM. The instrument will display the legend “Please wait” for a few seconds. Then it will display:
You are now ready to measure dissolved oxygen.
C a l i b r a t e d t o
9 8. 0 %

3.4 CALIBRATION IN AIR-SATURATED WATER

A second alternative for calibration is to calibrate in air-saturated water. This procedure is the same as the procedure for calibration in air except that the calibration medium is saturated water rather than saturated air.
To calibrate in air-saturated water:
1. Air-saturate a volume of water (300 to 500 mL) by aerating for at least 15 minutes at a relatively constant temperature.
2. Place the probe in the sample and provide adequate stirring (at least 1 foot per second). Switch to O take 5 minutes for the temperature to come to equilibrium. If you have just turned the instrument on, allow 15 minutes for the system to equilibrate.
3. Switch to CALIBRATE, select percent calibration, then adjust the number in the display to the correct value based on current barometric pressure or altitude and press
CONFIRM.
-TEMP and observe temperature and oxygen readings for stability. It may
2

3.5 CALIBRATION BY WINKLER TITRATION

A third alternative for calibration is to calibrate to a known oxygen value determined by a Winkler titration.
To calibrate using a Winkler-titrated sample:
1. Determine the dissolved oxygen value of a sample by Winkler titration.
2. Place the probe in the sample and provide adequate stirring (at least 1 foot per second).
Switch to O
-TEMP and observe temperature and oxygen readings for stability.
2
YSI, Incorporated Model 52 7
Calibration Section 3
3. Switch to CALIBRATE. The display will read:
Press SKIP. The display will read:
Press CONFIRM to select mg/L calibration. A display similar to the one below will be shown.
C a l i b r a t e
i n p e r c e n t ?
C a l i b r a t e
i n m g / L ?
4. Use ¿ and À to set the calibration value determined by Winkler titration. You may be able to set the value more quickly by using the SKIP key to bring the value near the one you want, then continuing with ¿ and À. Pressing SKIP once sets the value to the saturation value of oxygen in water at the probe temperature. A display similar to the one below will be shown.
Pressing SKIP again sets the value to 0.00 mg/L. (See Zero Calibration, below.) Pressing
SKIP a third time brings it back to the last calibration value.
5. When the correct value is set, press CONFIRM. The display will show the legend
“Please wait” for a few seconds, and then display “Calibrated to...” To take dissolved oxygen readings in mg/L, switch to O
Note: Since there is no way to determine percent air saturation after a calibration in mg/L, the % reading will be disabled.
E n t e r c a l v a l u e
L a s t = 8 . 1 5 m g / L
E n t e r c a l v a l u e
S a t . = 8 . 5 8 m g / L
-TEMP.
2

3.6 ZERO CALIBRATION

In rare applications, the accuracy of the calibration can be improved by performing a zero calibration along with one of the procedures discussed previously.
All oxygen probes have a small background current, even in the absence of oxygen. Model 52 compensation is based on the average background current of YSI 5700 and
YSI, Incorporated Model 52 8
Calibration Section 3
5900 Series probes. This is the default zero value. Although using this average will result in errors with probes whose background current differs from the average.
Note: Errors will be insignificant in most applications. See Discussion of Measurement Errors. For highest accuracy measurements, a zero calibration should be performed to compensate for the specific background current of the probe in use.
To calibrate to a true zero, place the probe in a zero oxygen environment and adjust the calibration value to zero.
A standard method for creating such an environment is to dissolve in water (preferably the water should come from the sample to be measured) excess sodium sulfite (Na and a trace of cobalt chloride (CoCl
). These chemicals will remove all oxygen from the
2
2SO3
),
sample (See Standard Methods for the Examination of Water & Wastewater, 17th edition, 1989, page 4-161.) Alternatively, you may place the probe in 100% nitrogen gas. Place the probe in the zero oxygen sample, switch to O
-TEMP and allow at least 20
2
minutes for the probe to come to equilibrium.
Switch to CALIBRATE. The display will show:
C a l i b r a t e
i n p e r c e n t ?
Press CONFIRM to select percent calibration. A display similar to the one below will be shown:
E n t e r c a l v a l u e
L a s t = 9 2 . 5 %
Press SKIP twice to set the calibration value to 0.0%. The following display will be shown.
E n t e r c a l v a l u e
Z e r o = 0. 0 %
Press CONFIRM to calibrate to zero. The display will ask you to wait, then show "Calibrated to 0.0 %." After zeroing, rinse the probe thoroughly to remove any residual trace of chemicals.
The Model 52 is calibrated at two points: the calibration value you select, and its default zero.
YSI, Incorporated Model 52 9
Calibration Section 3
Note: When you change the zero calibration point, you offset the other value, so that after zeroing the probe for zero oxygen, you must recalibrate in an oxygen environment. If you try to take readings without recalibrating, the display will provide this warning:
You may also select mg/L calibration and then zero-calibrate and the result will be the same.
8 . 7 4 m g / L 2 5 . 6 C
R e c a l i b r a t e

3.7 RESETTING THE DEFAULT ZERO CALIBRATION

To reset the zero calibration to its default value, turn the instrument off and then hold down SKIP and ¿ while turning the instrument back on again. The display below will show briefly, then the POST operation and display will resume.
This procedure resets the meter to its original state. You must reset the time, date, salinity, and the RS232 parameters after restoring the default zero. As with the zero calibration, you must also recalibrate after restoring the default zero.
S y s t e m r e s e t t o
d e f a u l t v a l u e s
YSI, Incorporated Model 52 10

SECTION 4 SALINITY

Salt reduces the ability of water to hold oxygen in solution. Enter the salinity of the sample you are measuring, and the meter will automatically compensate for the effect of salinity on dissolved oxygen. The default salinity setting is 0.0 ppt. You can enter any salinity setting between 0.0 and 40.0 ppt. To do this, turn the function switch to SALINITY. A display similar to the one below will show:
Use ¿ and À to adjust the salinity setting. You might find it quicker to press SKIP first, which will set the value to 35.0 ppt. The following display will be shown:
Pressing SKIP again will set the value to 0.0 ppt for fresh water measurements. The following display will be shown:
E n t e r s a l i n i t y
S e t t i n g = 0 . 0
E n t e r s a l i n i t y
S e a w a t e r = 3 5 . 0
E n t e r salinity
Fresh = 0. 0
Pressing SKIP once more will bring the setting to its original value. Once you have set the salinity correctly, press CONFIRM. The instrument will ask you to wait, then show you the new salinity setting.
If you have calibrated using the air saturation method, there is no need to recalibrate after changing the salinity setting. If you have calibrated in mg/L, then you must recalibrate after setting the salinity.
After calibrating in air or in air-saturated water, you may change the meter's salinity setting without recalibrating. You may, for instance, calibrate only once and then proceed to measure oxygen in a variety of samples at different salinity by simply setting the proper salinity before making each measurement.
YSI, Incorporated Model 52 11
Loading...
+ 33 hidden pages