Yokogawa Yokogawa CL155 Manual

User's
CL155
Manual
Clamp-on Tester
クランプテスタ
IM CL155
保証書付
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
IM CL155

Contents

Precautions for Safety Use of the Instrument ……………………………… i
1. Instrument Layout …………………………………………………………… 1
2. Measurement ………………………………………………………………… 3
2.1 Preparation for Measurement ……………………………………………… 3
2.2 Current Measurement ……………………………………………………… 3
2.2.1 AC Current Measurement (Normal Mode) ……………………………… 4
2.2.2 Peak Current Measurement (Peak Mode) ……………………………… 4
2.3 Voltage Measurement ……………………………………………………… 5
2.3.1 DC Voltage Measurement ………………………………………………… 5
2.3.2 AC Voltage Measurement ………………………………………………… 5
2.4 Resistance Measurement …………………………………………………… 6
2.4.1 Resistance Measurement (Normal Mode) ……………………………… 6
2.4.2 Continuity Check (400Ω range fixed) …………………………………… 6
3. Other Functions ……………………………………………………………… 8
3.1 Sleep Function ……………………………………………………………… 8
3.2 Data Hold Function ………………………………………………………… 8
3.3 OUTPUT Terminal (For current measurement only) …………………… 9
3.4 Optional Accessories ……………………………………………………… 10
4. Battery Replacement ……………………………………………………… 11
5. Specifications ……………………………………………………………… 12
6. Calibration and After-sales Service ……………………………………… 15
■PrecautionsforSafeUseoftheInstrument
When handling the instrument, ALWAYS observe all of the cautionary notes on safety given below. Yokogawa M&C Corporation is not at all liable for damage resulting from misuse of this product by the user that is contrary to these cautionary notes. Various symbols are used on the instrument and in this manual to ensure the product is used safety and to protect operators and property from possible hazards or damage. The following safety symbols are used where appropriate. Read the explanations carefully and familiarize yourself with the symbols before reading the text.
The instrument and this manual use the following safety symbols:
Danger! Handle with Care.   This symbol indicates that the operator must refer to an explanation
in the Users Manual in order to avoid the risk of personal injury or death and/or damage to the instrument.
Double Insulation  This symbol indicates double insulation.
AC Voltage/Current  This symbol indicates AC voltage or current.
DC Voltage/Current  This symbol indicates DC voltage or current.
Ground  This symbol indicates ground (earth)
This symbol Indicates that this instrument can clamp on bare conductors when measuring a voltage corresponding to the applicable Measurement category, which is marked next to this symbol.
WARNING
Indicates that there is a possibility of serious personal injury or loss of life if the operating procedure is not followed correctly and describes the precautions for avoiding such injury or loss of life.
CAUTION
Indicates that there is a possibility of serious personal injury of damage to the instrument if the operating procedure is not followed correctly and describes the precautions for avoiding such injury or damage.
NOTE
Draws attention to information essential for understanding the operation and features.
IM CL155
i
WARNING
Never make measurement on a circuit above 750V AC or 1000V DC.
Do not use the instrument in an atmosphere where any flammable or
explosive gas is present.
Do not attempt to make measurement in the presence of flammable gasses,
fumes, vapor or dust. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion.
Avoid using the instrument if it has been exposed to rain or moisture or if
your hands are wet.
Do not exceed the maximum allowable input of any measurement range.
Never open the battery compartment cover when making measurement.
Do not use the instrument if there is any damage to the casing or when the
casing is removed.
Do not turn the Function Selector switch with plugged in test leads
connected to the circuit under test.
Do not install substitute parts or make any modification to the instrument.
Return the instrument to Yokogawa M&C or your distributor for repair or re-calibration.
Always switch off the instrument before opening the battery compartment
cover for battery replacement.
Do not use the test leads that have deteriorated or are detective.
Check the test leads continuity
WARNING
To avoid damage to the instrument or electric shock! The restrictions on the maximum voltage level for which the CL155 testers can be used, depend on the measurement categories specified by the safety standards. These category specifications are formulated to protect operators against transient impulse voltage in power lines.
Maximum Allowable Input
Function
A
V, V
Input terminal-to­ground voltage
IM CL155
MEASUREMENT
CATEGORY II
AC 2000A rms Measuring circuit voltage: AC 750V rms
AC 750V rms /DC 1000V
AC 750V rms /DC 1000V
MEASUREMENT
CATEGORY III
AC 2000A rms Measuring circuit voltage : AC 600V rms
AC 600V rms /DC 600V
MEASUREMENT
CATEGORY
AC 2000A rms Measuring circuit voltage : AC 300V rms
AC 300V rms /DC 300V
ii
O (None, Other)   Other circuits that are not directly connected to MEAINS. Measurement category II (CAT.II)   Local level, appliance, portable equipment etc., with smaller transient over-
voltages than CAT.III.
Measurement category III (CAT.III):   Distribution level, fixed installation, with smaller transient over-voltages than
CAT.IV.
Measurement category IV (CAT.IV):  Outside and service entrance.
CAUTION
Always make sure to insert each plug of the test leads fully into the
appropriate terminal on the instrument.
Make sure to remove the test leads from the instrument before making
current measurement.
Be sure to set the Function Selector switch to the "OFF" position after use.
When the instrument will not be in use for a long period of time, Place it in storage after removing the battery.
Use a damp cloth and detergent for cleaning the instrument. Do not use
abrasives or solvents.
NOTE
Radiation immunity affects the accuracy of CL155 testers under the
conditions specified in EN 61000-4-3.
If equipment generating strong electromagnetic interference is located
nearly, the testers may malfunction.
IM CL155
iii
1.InstrumentLayout
(1)Transformer Jaws : Pick up current flowing through the conductor. (2) Open/Close Lever : Operates the transformer jaws. Press to open the
Transformer Jaws.
(3) Function Selector Switch : Selects ranges. Also used to power the
instrument on.
(4) Data Hold Button : Freezes the display reading. " " is shown on the
display when Data Hold is enabled.
(5) Mode Selector Button : On a AC current (" 400A" or " 2000A" or
the resistance ( ) range. Press the mode button to cycle through the measurement modes. The instrument is initially set to the normal mode and can be switched to peak or continuity check mode by means of the mode button.
IM CL155
1
A / V
(ACA/ACV)
Normal
 ↓
Peak
Display Display
(Resistance/Continuity)
  Resistance
Continuity check
Ω
(6) Button : Reset the measurement in peak mode. (7) LCD Display : Field effect type of liquid crystal display with maximum counts
of 3999. Function symbols and decimal point are controlled by the microprocessor based on the selected function and measuring mode.
(8) Terminal Cover : Slides over Hi and Lo Terminals to prevent access to them
when OUTPUT terminal is in use.
(9) OUTPUT Terminal (For " 400A" or " 2000A" range only) : Provides DC
voltage output in proportion to the AC or DC current reading. (See section
3.3, OUTPUT Terminal.) The output is connected to a recording device such as a chart recorder for long hour monitoring. No output is available on voltage and resistance ranges.
(10) Lo Terminal : Accepts the black test lead for voltage or resistance
measurement.
(11) Hi Terminal : Accepts the red test lead for voltage or resistance
measurement.
(12) Safety Hand Strap : Prevents the instrument from slipping off the hand
during use.
(13) Test Leads (Model 98072) : Connected to Lo and Hi terminals for voltage
or resistance measurement.
(14) Barrier: It is a part providing protection against electrical shock and
ensuring the minimum required air and creepage distances.
IM CL155
2
2.Measurement
2.1PreparationforMeasurement
CAUTION
The jaw section is a delicate, precision sensor. Do not subject the jaw to
unreasonably strong shock, vibration, or force when using it.
If dust gets into the tops of the jaws, remove it immediately. Do not close the
jaws when dust is trapped in its joints as the sensor may break.
Please check that the function and mode are set to the desired setting
before measurement.
2.2CurrentMeasurement
WARNING
Do not make measurement on a circuit above 750VDC. This may cause
shock hazard or damage to the instrument or equipment under test.
Do not make current measurement with the test leads connected to the Hi
and Lo terminals.
When measuring current is not less than 1000A, make sure to stop
measurement within the maximum measuring time shown below. Otherwise, transformer jaws may heat to cause a fire or deformation of molded parts, which will degrade insulation.
1000 to 1500A : 15min. 1500 to 2000A : 5min.
Keep your fingers and hands behind the barrier during measurement.
IM CL155
3
2.2.1ACCurrentMeasurement(NormalMode)
(1) Set the Function Selector switch to the " 400A" or " 2000A" position.
"AC" should be shown on the upper left corner of the display.
(2) Press the open/close lever to open the transformer jaws and clamp them
onto the conductor under test, then take the reading on the display. The most accurate reading will be obtained by keeping the conductor at the center of the transformer jaws.
NOTE
During current measurement, keep the transformer jaws fully closed.
Otherwise, accurate measurement cannot be made. The maximum measurable conductor size is approx. 54.5mm in diameter.
The transformer jaws may buzz when measuring large current. This has no effect on the instruments performance or safety.
2.2.2PeakCurrentMeasurement(PeakMode)
(1) Set the Function Selector switch to the "
400A" or " 2000A" position.
(2) Press the button once to enter from the normal mode to the PEAK
mode. " " should be shown on the display.
(3) Press the open/close lever to open the transformer jaws and clamp them
onto the conductor under test. Then, press the button.
(4) The display shows the currents crest value divided by the square root of
two. Therefore, when the current is sinusoidal, the reading equals RMS value.
(5) To reset the display, press the button.
(6) After the measurement is over, press the
button to return to the
normal mode.
NOTE
In the Peak measurement mode, the data hold feature is disabled.
In the Peak measurement mode, the response time is 10ms.
IM CL155
4
2.3VoltageMeasurement
WARNING
Do not make measurement on a circuit above 750V AC or 1000V DC. This may cause shock hazard or damage to the instrument or equipment under test.
2.3.1DCVoltageMeasurement
(1) Set the Function Selector switch to the " V " position. "DC" should be shown
on the upper left corner of the display.
(2) Slide the terminal cover to the left to disclose the Hi and Lo terminals. Plug
the red test lead into the Hi terminal and the black test lead into Lo terminal.
(3) Connect the tip of the red and black test leads to the positive (+) and
negative (-) sides of the circuit under test respectively. Take the reading on the display.
2.3.2ACVoltageMeasurement
(1) Set the Function Selector switch to the " V " position. "AC" should be shown
on the upper left corner of the display.
(2) Slide the terminal cover to the left to disclose the Hi and Lo terminals. Plug
the red test lead into the Hi terminal and the black test lead into Lo terminal.
(3) Connect the tip of the test leads to the circuit under test respectively. Take
the reading on the display.
NOTE For high sensitivity, there are parts which do not indicate "0".
IM CL155
5
2.4ResistanceMeasurement(NormalMode)
WARNING
Never use the instrument on an energized circuit.
2.4.1ResistanceMeasurement(NormalMode)
(1) Set the Function Selector switch to the "
" position. The "Ω" should be
shown on the upper right corner of the display.
(2) Slide the terminal cover to the left to disclose the Hi and Lo terminals. Plug
the red test lead into the Hi terminal and the black test lead into the Lo
terminal. Check that "OL" (over indication) is indicated on the LCD display. (3) Short the tip of the test leads and check whether the display reads "0". (4) Connect the tip of the test leads to the circuit under test respectively. Take
the reading on the display.
NOTE When shorting the tip of the test leads, the display may read a very small resistance instead of "0". This is the resistance of the test leads on the display.
2.4.2ContinuityCheck(400Ωrangefixed)
The continuity check mode is enabled by pressing the
button on
resistance range. " " andΩis indicated on the display to show the instrument in the continuity check mode. The buzzer beeps, if the resistance under test is 50Ω or less.
(1) Set the Function Selector switch to the " " position. (2) Slide the terminal cover to the left to disclose the Hi and Lo terminals. Plug
the red test lead into the Hi terminal and the black test lead into the Lo
terminal. Check that "OL" (over indication) is indicated on the LCD display.
IM CL155
6
(3) Press the
button once to enter from the normal mode to the
continuity check mode. " " should be indicated on the display. (4) Short the tip of the test leads and check whether the display reads "0". (5) Connect the tip of the test leads to the circuit under test. If the resistance is
50Ω or less, the buzzer beeps.
NOTE When shorting the tip of the test leads, the display may read a very small resistance instead of "0". This is the resistance of the test leads on the display.
IM CL155
7
3.OtherFunctions
3.1SleepFunction
This is a function to prevent the instrument from being left powered on in order to conserve battery life. This function causes the instrument to enter the Sleep (powered-down) mode about 10 minutes after the last switch or button operation.
To exit the Sleep mode, press the Data hold turn the Function Selector switch back to "OFF", then to any other position, or press any button. The current is consumed a little in the Sleep mode.
HowtoExittheSleepMode
Turn the Function Selector switch from "OFF" to another position with the data hold function button pressed. Then, "P.OFF" is shown on the display. This disables the sleep function and enables continuous use of the instrument.
NOTE
Connecting the plug to the OUTPUT terminal disables the Sleep function.
The function is enabled on removing the plug from the terminal.
The Sleep function is disabled in the PEAK measurement mode.
3.2DataHoldFunction
This is a function used to freeze the measured value on the display. Press the Data Hold button to freeze the reading. The reading will be held regardless of subsequent variation in input. " " is shown on the upper right corner of the display while the instrument is in the Data Hold mode. To exit the Data Hold mode, press the Data Hold button again.
button or button or
NOTE
The Data hold function is disabled in the peak measurement mode on the
AC current range.
When the Sleep function is activated, the Data hold mode turns to the
normal mode.
IM CL155
8
3.3OUTPUTTerminal(Forcurrentrangesonly)
WARNING
Never use the instrument on a circuit above 750VAC or 1000VDC. This
may cause electrical shock hazard and damage to the instrument or the circuit under test.
Never apply voltage to the OUTPUT terminal.
Only on the " 400A" or " 2000A" range, DC voltage proportional to the input current is output from the OUTPUT terminal.
(1) Attach the output plug cable (sold separately : 98076 or 98077) to a
connection lead so that the output voltage can be connected to a recording device such as a chart recorder.
(2) Slide the terminal cover to the right to disclose the OUTPUT terminal and
insert the output plug into the terminal. Make connection to the recording
device.
(3) Set the Function Selector switch to the " 400A" or " 2000A" position and
follow appropriate measurement instructions.
NOTE
Output voltage is 1mV/A on the " 400A" range and 0.1mV/A on
the " 2000A" range. Set an appropriate input sensitivity on the recorder.
The peak hold function does not apply to the recorder output even if the
instrument is in the peak hold mode.
For long term measurement, disable the sleep function.
IM CL155
9
3.4OptionalAccessories
Clamp Adapter Model 99025 (For AC current measurement only) The clamp adapter (99025) has been discontinued.
The clamp adapter (99025) is designed to increase the measuring capability of a clamp meter. With the use of the Clamp Adapter, you can not only extend current range over 3000A, but also clamp on a large bus-bar or conductor. (1) Set the Function Selector switch to the "
A " position.
(2) As shown in the figure below, clamp Model CL155 onto the pickup coil of
Model 99025. (3) Clamp Model 99025 onto the bus-bar or conductor under test. (4) Take the reading on Model CL155 and multiply it by 10.
IM CL155
10
4.BatteryReplacement
WARNING
To avoid electric shock hazard, make sure to set the Function Selector switch to "OFF" and remove the test leads from the instrument before trying to replace battery.
CAUTION
Do not mix new and old batteries.
Make sure to install battery in correct polarity as indicated in battery
compartment.
If the battery voltage becomes too low for the instrument to operate normally, "
" is shown on the display. Then, replace the battery. Note that when the battery is completely exhausted, the display blanks without " " shown. (1) Set the Function Selector switch to the "OFF" position. (2) Unscrew and remove the battery compartment on the bottom of the
instrument. (3) Replace the battery observing correct polarity. Use two new R6P batteries. (4) Re-place and screw the battery compartment cover.
IM CL155
11
5.Specifications

Instrument Specifications

Measuring Ranges and Accuracy (at 23 ±5℃, 45 to 75% relative humidity)
AC Current
Ranges
400A 0 to 400.0A 0.1A
0 to 1000A
2000A
Conversion method : AC coupled, true rms responding, calibrated to the rms *1 : 400A range : 5 to 100% of range , 2000A range : 200A to 2000A
1000 to 1500A 15min. 1500 to 2000A ±3.0% rdg (50/60Hz) 5min.
AC Voltage ( V) Auto-ranging
Range Measuring range 40V 0 to 40.00V 0.01V 400V 15.0 to 400.0V 0.1V 750V 150 to 750V 1V
Conversion method : AC coupled, true rms responding, calibrated to the rms Initially set to the 40V range. Input impedance is about 1MΩ.
DC Voltage ( V) Auto-ranging
Range Measuring range Resolution Accuracy 40V 0 to ±40.00V 0.01V
1000V 150 to ±1000V 1V
Initially set to the 40V range. Input impedance is about 1MΩ.
400A, 2000A
Measuring
range
Resolution
1A
Accuracy (frequency range)
±1.0% rdg ±3dgt (50/60Hz) *1 ±2.0% rdg ±3dgt (40 to 1kHz)
±1.0% rdg ±3dgt (50/60Hz) *1 ±3.0% rdg ±3dgt (40 to 1kHz)
Resolution
Accuracy
±1.0% rdg ±2dgt (50/60Hz) ±1.5% rdg ±3dgt (40 to 1kHz)
±1.0% rdg ±2dgt400V 15.0 to ±400.0V 0.1V
Maximum
Measurement
Time
Continuous
Resistance (Auto-ranging)
Range Measuring range Resolution Accuracy
Ω 0 to 400.0Ω 0.1Ω
400
Ω 0.150 to 4.000kΩ 1Ω
4k 40kΩ 1.50 to ±40.00kΩ 10Ω
±1.5% rdg ±2dgt
400kΩ 15.0 to ±400.0kΩ 100Ω
Initially set to the 400Ω range. In the continuity check mode, fixed to the 400Ω range and when the reading is not more than . 50 ±35Ω, the buzzer beeps.
IM CL155
12
Peak Current (10ms)
Range Measuring range Accuracy (Frequency Range)
400A 0 to 400.0A
2000A
0 to 1500A
±3.0% rdg ±6dgt (50/60Hz) ±3.0% rdg ±10dgt (40 to 1kHz)
±4.0% rdg ±6dgt (50/60Hz) ±4.0% rdg ±10dgt (40 to 1kHz)
1500 to 2000A ±4.5% rdg (50/60Hz)
OUTPUT (AC Current Ranges) DC Output : 100.0mV per 1000 counts (Output impedance : about 10k
Ω)
Range Measuring range Accuracy (Frequency Range) 400A 0 to 400.0mV/0 to 400A ±1.5% rdg ±0.5mV (50/60Hz) *1
±2.5% rdg ±0.5mV (40 to 1kHz)
2000A 0 to 150.0mV/0 to 1500A ±1.5% rdg ±0.5mV (50/60Hz) *1
±3.5% rdg ±0.5mV (40 to 1kHz)
150.0 to 200.0mV/1500 to 2000A
*1 : 400A range : 5 to 100% of range , 2000A range : 200A to 2000A
±3.5% rdg (50/60Hz)
■GeneralSpecifications
Operating System : Dual integration
Measurement Function : AC current, DC current, AC voltage, resistance,
continuity check
Display : Liquid crystal display with maximum counts of 3999
Overrange Indication : "OL" is shown on the display
Response Time : Approx. 2 seconds.
Temperature and Humidity for Guaranteed Accuracy : 23℃ ±5℃, relative
humidity 45 to 75% without condensation
Operating Temperature and Humidity : 0 to 40, relative humidity up to 85%
without condensation
Storage Temperature and Humidity : -20 to 60, relative humidity up to 85%
without condensation
Effect of conductor position : At every part inside the jaws
Range Measuring range Conductor size Effect of conductor position 400A 0 to 400A φ10mm or less ±2.0%rdg±3dgt
2000A
0 to 1500A φ25mm or less ±2.0%rdg±3dgt 1500 to 2000A φ25mm or less ±5.0%rdg
Effect of external magnetic field : 1A or less in AC or DC magnetic field of
400 A/m
IM CL155
13
Power Source : Two R6P 1.5V battery
Battery Life : Approx. 80 hours (continuity)
Current Consumption : Approx. 7mA max. (Sleep function : Approx. 20μA)
Sleep function : Automatically powered down in Approx. 10 minutes after the
last switch operation
Withstanding Voltage : 6300V AC, 50/60Hz for 5 sec. between electrical
circuit and housing case or metal part of the jaws
Insulation Resistance : 50MΩ or greater at 1000V between electrical circuit
and housing case or metal part of the jaws
Conductor Size : Approx. 54.5mm diameter max.
Dimensions : Approx.105(W) x 247(H) x 49(D) mm
Weight : Approx. 470g (with batteries)
Safety Standard: EN 61010-1, EN 61010-2-033          EN 61010-031, EN 61010-2-032          AC/DC300V CAT IV,AC/DC 600V CAT III,           AC/DC 1000V CAT II, Pollution degree2, indoor use
altitude 2000m or less
EMC Standard : EN 61326-1, EN 61326-2-1
Radiation immunity : EN 61000-4-3
Environmental standard: EN 50581
Accessories : Test leads Model 98072 1set         R6P batteries 2         Carrying case Model 93034 1         Users Manual 1
Optional Accessories : Output cable Model 98076
Output cable (for terminal screw) Model 98077
6.CalibrationandAfter-salesService
Should any failure occur while you are using the tester, follow the instructions given below. If the tester still fails to operate correctly and needs repair, contact the vendor from whom you purchased the instrument or the nearest Yokogawa Meters & Instruments sales office.
Turn off the POWER switch once, then turn it back on again.
If the tester does not turn on, replace the battery with a new one.
Calibration It is recommended that the instrument be calibrated once every year.