2. Measurement
2.1 Preparation for Measurement
CAUTION
● The jaw se ction is a delicate, precision sensor. Do not subject the jaw to
unreasonably st rong sh ock, vibratio n, or forc e when usi ng it.
● If dust gets into the tops of the jaws, r emove it immed iately. Do not c lose t he
jaws whe n dust is trapped in its joint s as the sen sor may bre ak.
● Please c heck that the Funct ion Sel ector sw itch is se t to the des ired pos ition
2.2 AC Current Measurement
WARNING
Never use the instrument on a circuit above 600V AC. This may cause
elect rical s hock ha zard and damage to t he instr ument or the circ uit unde r test.
Keep your f inger s and hand s behind t he barr ier during measu rement .
(1) Set the Functio n Selec tor switch to the "
(2) Press the o pen/close lever to open the tran sformer jaws and clamp them
onto a s ingle conductor and take the readin g on th e disp lay. The most
accur ate r eading will be obtai ned by kee ping the cond uctor at the center of
the transform er jaws.
NOTE
● During current measurement, keep the trans forme r jaws fully closed.
Other wise, accurate measurement s cannot be ta ken. Maximum conductor
size is 30 mm in diameter.
● When measuri ng larg er cur rent, t he transform er jaws may buz z. Thi s in not a
fault and does not af fect t he accu racy eit her.
2.3 AC Voltage Measurements
WARNING
Never use the instrument on a circuit above 600V AC. This may cause
elect rical s hock ha zard and damage to t he instr ument or the circ uit unde r test.
Keep your f inger s and hand s behind t he barr ier during measu rement .
(1) Set the Functio n Selec tor switch to the "
(2) Plug the red test lead into Hi te rminal and the black test lead into th e Lo
terminal.
(3) Connec t the t ip of t he r ed and black test lead s to the cir cuit under test and
take the reading on t he displ ay.
2.4 Resistance Measurement and Continuity Check
WARNING
Never use t he instr ument on a n energi zed circ uit.
(1) Set the Functio n Selec tor switch to the "
(2) Plug the red test lead into Hi te rminal and the black test lead into th e Lo
termin al. Chec k that "1" (over in dication) is indic ated on the LCD display.
(3) Short t he tip of th e test lea ds and che ck whether the display reads "0".
(4) Connect the tip of the test lead s to the c ircuit unde r tes t. If the resis tance is
30.0Ω o r less, the buzze r beeps.
NOTE
When shortin g the tip of the
test leads, the display may
read a very small resistance
instead of "0 ". This is the
resistance of the test leads on
the disp lay.
3. Other Functions
3.1 Data Hold Function
This is a fu nctio n us ed to fr eeze the measur ed va lue on t he d isplay. Press the
Data Hold button to freeze the reading. The reading will be held regardless of
subsequent variation in input. " " is shown on the upper right corner of the
display w hile the i nstrument is in th e Data Ho ld mode.
To exit the Data H old mod e, press t he Data Hold button again .
User's
Manual
CL130/CL135
Clamp-on Tester
IM CL130
Contents
Precautions for Safety Use of the Instrument
■
1. Instrument Layout
2. Measurement
2.1 Preparation for Measurement
2.2 AC Current Measurement
2.3 AC Voltage Measurements
2.4 Resistance Measurement and Continuity Check
3. Other Functions
3.1 Data Hold Function
3.2 Optional Accessories
4. Battery Replacement
5. Specification
6. Calibration and After-sales Service
8th Edition: May 2018 (KYOU)
■ Precautions for Safe Use of the Instrument
When handlin g the instr ument, ALWAYS obser ve al l of t he ca utiona ry n otes on
safety given bel ow. Yokogawa M&C C orpor ation is not at all li able for damag e
resulting from misuse of this produ ct by th e user that is c ontrar y to these
cautionary n otes.
Various symbols are used on the in strument and in this manual to ensure the
produc t is used safet y and to protec t oper ators and property from possible
hazar ds or damag e. The foll owing safety symbols ar e use d where app ropri ate.
Read the explanations carefully and familiarize yourself with the symbols
before reading the text.
The inst rument a nd this ma nual use t he follo wing safety symb ols:
Danger! Handle with Care.
This sy mbol in dicate s that the ope rator m ust refe r to an explanat ion
in t he User’s Manua l in order to avoid the risk of perso nal injur y or
death an d/or dama ge to the in strument.
Double Insulation
This symbol indicates do uble insulation.
AC Voltage/Current
This symbol indicates AC voltage or c urrent.
Ground
This symbol indicates gr ound (ear th)
Indicates that this instrume nt can clamp on b are conductor s when
Measuring a voltage corresponding to the applicable Measurement
Catego ry, which i s marked n ext to thi s symbol.
WARNING
Indicates that there is a possi bilit y of serio us pers onal injury or loss of life if
the operating procedure is not followed correctly and describes the
precautions for avoiding such inj ury or l oss of lif e.
CAUTION
Indicates that there is a po ssibil ity of serious p ersonal injury o f dama ge to the
instrument if the operating procedure is no t followe d corre ctly and describes
the prec autio ns for avoi ding suc h injur y or damag e.
NOTE
Draws attention to information essential for understanding the operation and
features.
IM CL130
WARNING
● Never make m easurem ent on a circuit above 600V AC.
● Do not u se the instr ument in an atmosphe re where any f lamma ble or
explosive gas is pr esent.
● The t ransfo rmer jaws a re mad e of metal and t heir t ips a re not insul ated. Be
espec ially careful about the hazard of possible shor ting whe re the
equipm ent under t est has exp osed met al par ts.
● Avoid using the instrument if it has been exposed to rain o r moist ure or if
your hand s are wet.
● Do not exce ed the ma ximum allowabIe i nput of any measurem ent rang e.
● Never ope n the bat tery c ompar tment c over when m aking me asurem ent.
● Do n ot u se t he i nstrum ent if t here is any d amage to the casin g or when the
casing is removed.
● Do not turn the Functi on Selecto r switch with plugged in test leads
conne cted to th e circuit under te st.
● Do not install substitute parts or make any modificat ion to th e instr ument.
Return the instrument to Yokogawa M&C o r your distributor for repair or
re-calibration.
● Always switch off the instrument before opening the b atter y compartme nt
cover for batter y repla cement.
● Do not use t he test le ads that have deteri orated o r are defec tive.
● Check the test lea ds continuity.
WARNING
To avoid damage t o the inst rument o r elect ric sho ck!
The restrictions on t he maximum voltage level for which the CL130/CL135
tester s c an be used, de pend on the measurment c ategories specified by t he
safety standar ds. These cate gory specif icati ons are formulated to protect
operat ors agai nst tran sient impulse vol tage in po wer lines.
Model Function
CL1 3 0
CL1 3 5
O (None, O ther)
Other c ircuit s that are n ot direc tly con nected t o MEAI NS.
Measur ing categor y II (CAT.II)
Local level, applia nce, port able equipme nt etc., with small er tr ansient over-
voltag es than CAT.III.
Measur ing categor y III (CAT.III):
Distribution level, fixed installation, with smaller transient over-voltages
than CAT.IV.
A, V
A, V
Maxim um Allowable Inp ut
Measur ment ca tegor y Ⅲ
Measur ing cir cuit voltage : 60 0Vrms AC
Measur ing cir cuit voltage : 60 0Vrms AC
600A AC
600A AC
CAUTION
● Always make sure to insert each plug of the test leads fully into the
approp riate te rminal o n the inst rument.
● Make sure to remove the test leads from the instrument before making
current measurement.
● Be sure to set the Functi on Selector switch to the "OFF " p osition a fter use.
When the inst rument will no t b e i n u se for a long period of time, Place it in
storag e after removing t he battery.
● Use a damp cloth and detergent for c leaning the instrument. Do n ot use
abrasi ves or solvents.
NOTE
● Radiati on immunity affects the accur acy of CL130/CL135 testers under the
conditions specified in EN61000-4-3.
● If equip ment ge nerating strong elec tromag netic in terfe rence is loca ted near ly,
the testers may mal funct ion.
1. Instrument Layout
Barrier: It is a part providing protection against electric al shock and ensuring
the mini mum requi red air an d creep age distances.
200A " or " 600A " p ositio n.
200V " or " 600V " po sitio n.
" positi on.
3.2 Optional Accessories
Clamp Adapter Model 99 025 (For AC cu rrent me asurem ent only)
Clamp A dapter Model 99025 is de signed to inc rease
the m easuri ng capabili ty of a c lamp meter. With the
use of the Clamp Adapt er, you can not only extend
current range over 3000A, but also clamp on a large
bus-bar or con ductor.
(1) Set the Functi on S elect or sw itch to t he "
" or "
(2) As shown in the fig ure below, clamp Model
(3) Clamp Model 99025 onto the bus-bar or
(4) Take the reading on Model CL130/CL135 and
600A " p ositio n.
CL130/CL135 onto the pickup coil of Model
99025.
conductor und er test.
multip ly it by 10.
200A
NOTE
The clamp adapter (99025) has been d iscont inued.
4. Battery Replacement
WARNING
To avoid elec tric shoc k ha zard, make sure to set t he Fun ction Selector switch
to "OFF" and r emove the test leads from the instru ment before trying to
replac e battery.
CAUTIOIN
Make sure to install battery in correct polarit y as indic ated in battery
compartment.
If the battery voltage beco mes too low for t he instr ument to ope rate normally,
"
" is show n on the display. Then, replace the batter y. Note that w hen the
batter y is co mpletel y exhauste d, the dis play blan ks witho ut "
(1) Set the Functio n Selec tor switch to the "O FF" pos ition.
(2) Unscrew and remove the battery comp artm ent on the bott om of the
instrument.
(3) Replace the batter y obser ving correct p olari ty. Use a new 6LR61 ( Alkali ne
batter y) or 6F2 2 (Manganese bat tery).
(4) Re-place and screw the bat tery c ompar tment cover.
" shown.
5. Specifications
■ Instrument Specifications
● Measur ing Ranges and Ac curacy (at 2 3±5°C, re lative hu midit y up to 75%)
CL1 3 0
Range Frequency Accuracy Crest Fac tor
200A AC
600A AC
200V AC
600V AC
200Ω
CL1 3 5
Range Frequency Accuracy Crest Fac tor
200A AC
600A AC
200V AC
600V AC
200Ω
Conversion method : AC cou pled, tr ue rms re spondi ng, cali brated to the rms
■ General Specifications
● Operating Syst em : Dual integrati on
● Measure ment Function : AC cur rent, DC curr ent, Resistanc e, Continuity
check
● Display : Liquid crystal display w ith max imum co unts of 199 9
● Overrange Indi catio n : "1" is displayed on the highest digit.
● Respon se Time : Approx. 1 se cond (app rox.2 se conds o n resist ance ra nge)
● Temperature and Humidity for Guaranteed Accuracy : 23°C ±5°C, relat ive
humidity up to 75% without co ndensation
● Operating Temperature and Humidit y : -10 to 50°C (wit hout co ndensation)
● Storag e Temperature and H umidit y : -20 to 60° C, relative humidi ty up to 75%
without condensation
50/60Hz 1.5% rdg +6 dgt sin
40 to 1kHz 2% rdg +5dgt
50/60Hz 1% rdg +3dgt
40 to 1kHz 2% rdg +5dgt
50/60Hz 1% rdg +2dgt
40 to 1kHz 1.5% rdg +4d gt
50/60Hz 1% rdg +2dgt
40 to 1kHz 1.5% rdg +4d gt
-
50/60Hz 1.5% rdg +4d gt
40 to 1kHz 2% rdg +5dgt sin
50/60Hz 1.5% rdg +4d gt
40 to 1kHz 2% rdg +5dgt sin
50/60Hz 1% rdg +2dgt
40 to 1kHz 1.5% rdg +4d gt sin
50/60Hz 1% rdg +2dgt
40 to 1kHz 1.5% rdg +4d gt sin
-
1.2% rdg +4 dgt
Beeps be low about 30Ω
1.2% rdg +4 dgt
Beeps be low about 30Ω
up to 30°C , 90% rel ative humidity
up to 40°C, 75% relative humidity
up to 50°C, 45% relat ive humid ity
≦3
≦3
≦3
≦3
↑
↑
↑
↑
↑
↑
↑
● Effec t of condu ctor po sition :
CL130 Within 2% of indi cated value at the c enter to a 10 mm-dia conductor
carr ying 10 0A, at ever y par t inside t he jaws
CL135 Within 3% of indi cated value at t he c enter to a 10 mm -dia conductor
carr ying 10 0A, at ever y par t inside t he jaws
● Effec t of exter nal magn etic fi eld : 2 A or les s in AC magnetic field of 40 0 A/m
● Power Sou rce : 6F2 2(9V DC) or equ ivalent b atter y
● Batter y Life : A pprox. 2 00 hour s (contin uous)
● Curren t Consum ption : A pprox. 2m A
● Withst anding Voltage : 6300V AC for 5 sec. betwe en housing case and metal
part of j aws
● Insulat ion Re sistan ce : 10MΩ o r grea ter at 1000V betwe en hou sing c ase and
metal pa rt of jaw s
● Conduc tor Size : A pprox. 3 0mm diameter ma x.
● Dimens ions : Ap prox. 91(W ) x 208(H) x 40(D) mm
● Weig ht : Approx. 400g (batteries inc luded)
● Safe ty Stan dard: EN 61010-1, EN61010- 2-03 3
● EMC Sta ndard: EN 61326 , EN 5502 2
● Radiation immunity : EN 61000- 4- 3
Environ mental standar d: EN 50581
● Acces sories : Test leads Mo del 98071 ..................1s et
EN 61010- 031
EN 61010- 2-32
(600V AC C AT.III, Pollutio n degree2, indoo r use)
6F22 battery ....................................1
Carr ying ca se Mode l 93032 ............1
User’s Manual ..................................1
6. Calibration and After-sales Service
Should any fai lure occ ur while you are using t he tester, follow t he instruction s
given below. If t he test er stil l fails to ope rate co rrectly and needs repair, contact
the vendor from whom you purc hased the instr ument or t he nearest Yokogawa
Meters & I nstruments sales off ice.
● Turn off the P OWER swi tch onc e, then tur n it back o n again.
● If the tes ter does n ot turn on , replac e the bat tery wi th a new one.
Calibration
It is rec ommend ed that th e instrument be c alibrated once every year.
This manual is valid only in China.