Yokogawa VMA0055 Data sheet

General Specifications
VMA0055 Sensor Module
mV, V, and mA Measure & Simulate plus 24 Vdc Isolated
Loop Supply for use with YPC 4000 and YPC 4010
• Source precise voltage or mA signals to check calibration or operation of receiving devices
- Key in exact output value desired
- Increase or Decrease output value by pre-defined increments Auto-step tests are user defined for start and
­stop values, number of steps, dwell time, and manual or automatic advance
• Read volts or mA from YPC meter while sourcing volts or mA from module
VMA
- Test input isolation buffers with one YPC; apply mA value to buffer input with VMA and measure mA value output of buffer
- Select mA or V measurement on YPC meter from the VMA output display without interruption of output
• Simulate transmitter output in externally powered loop
- VMA modulates loop current to any selected value
- Increase or Decrease simulated value by pre­defined increments
- Auto-step tests are user defined for start and stop values, number of steps, dwell time, and manual or automatic advance
The VMA0055 is a dc voltage and current measurement and simulation module for use with the Yokogawa YPC 4000 Series
Multi-Function Tester
DVM and loop calibrator for terrific versatility. When used with other YPC functions, the VMA / YPC combination provides valuable functionality not available in other devices. Measurement ranges include ±500 mV, ±55 V, and ±100 mA. Sourcing capabilities include ±1000 mV, 0 – 24 V and 0 – 24 mA. The VMA module will also simulate transmitter outputs when connected to a loop or it can be used to temporarily supply 24V dc loop power to unpowered devices during commissioning or calibration activities.
VMA0055 goes far beyond the capability of the YPC base unit and has its own isolated power supply so common ground problems are note an issue. Consider the following advantages for YPC users:
• Power a transmitter directly from VMA
- Use YPC 4000 and 4010 to calibrate transmitters anywhere, anytime.
- Concurrently use YPC base unit V / mA meter to
simultaneously check transmitter output
- Use YPC 4010 HART® communications to
commission or reconfigure unpowered transmitters
• Measure mV, V or mA as you would with a dc DVM
- Reduce the number of tools the technician carries
- Resolution to 1/100,000 counts
- Select desired V range for best possible accuracy
. The module features the functionality of a
50 volts / mA measurement
±
Available certifications include:
• NIST traceable – standard
• CE Mark
• Intrinsically Safe – MET Laboratories per CSA C22.2 & UL 913 Class I Division I, Groups A, B, C & D
VMA Module Accessories
P/N A900529-00015 VMA Test Lead Kit ­banana plugs on 9" breakouts (both ends), assorted connectors (required for source and
simulate functions)
Yokogawa Corporation of America Dart Road, Newnan, GA 30265, USA ©Copyright November 2004
GS 61A
YPCVMA-01E-A
Vma Setup
Auto-Step 1:
Start Mode: mA Source
Low pt: 4.000 mA Upr pt: 20.000 mA Steps: 5 Auto-Step 3
Dwell Time: Manual
Up Down Select Back
VOLTS
COM
mA
Milliamp
Receiver
Loop Load
VMA
Module
Measure
S1: 24.00 V
S2: 0.000 PSIG S3: 72.05 deg F
IV: 0.000 mA
Pon Poff More
Sample
Transmitter Simulation M ode
Xmitter Sim
Output = 16.000 mA
IV = 16.042 mA
Up Down Select Back
VOLTS
COM
mA
Power Supply
Loop Load
VMA
Module
+
YPC Displays and Wiring Diagrams for VMA0055 module with YPC 40xx
24V dc Loop P
ower
V or mA Sourcing
All Rights Reserved ©Copyright November 2004, Subject to change without notice.
ransmitter Sim
T
Yokogawa Corporation of America GS 61A YPCVMA-01E-A
ulation Mode