General
Specifications
VMA0055 Sensor Module
mV, V, and mA Measure & Simulate plus 24 Vdc Isolated
Loop Supply for use with YPC 4000 and YPC 4010
• Source precise voltage or mA signals to check calibration or
operation of receiving devices
- Key in exact output value desired
- Increase or Decrease output value by
pre-defined increments
Auto-step tests are user defined for start and
stop values, number of steps, dwell time, and
manual or automatic advance
• Read volts or mA from YPC meter while sourcing volts or mA from
module
VMA
- Test input isolation buffers with one YPC;
apply mA value to buffer input with VMA
and measure mA value output of buffer
- Select mA or V measurement on YPC meter
from the VMA output display without
interruption of output
• Simulate transmitter output in externally powered loop
- VMA modulates loop current to any selected
value
- Increase or Decrease simulated value by predefined increments
- Auto-step tests are user defined for start and
stop values, number of steps, dwell time, and
manual or automatic advance
The VMA0055 is a dc voltage and current measurement and
simulation module for use with the Yokogawa YPC 4000 Series
Multi-Function Tester
DVM and loop calibrator for terrific versatility. When used with other
YPC functions, the VMA / YPC combination provides valuable
functionality not available in other devices. Measurement ranges
include ±500 mV, ±55 V, and ±100 mA. Sourcing capabilities include
±1000 mV, 0 – 24 V and 0 – 24 mA. The VMA module will also
simulate transmitter outputs when connected to a loop or it can be
used to temporarily supply 24V dc loop power to unpowered devices
during commissioning or calibration activities.
VMA0055 goes far beyond the
capability of the YPC base unit and has its own isolated power supply
so common ground problems are note an issue. Consider the
following advantages for YPC users:
• Power a transmitter directly from VMA
- Use YPC 4000 and 4010 to calibrate transmitters
anywhere, anytime.
- Concurrently use YPC base unit V / mA meter to
simultaneously check transmitter output
- Use YPC 4010 HART® communications to
commission or reconfigure unpowered
transmitters
• Measure mV, V or mA as you would with a dc DVM
- Reduce the number of tools the technician carries
- Resolution to 1/100,000 counts
- Select desired V range for best possible accuracy
. The module features the functionality of a
50 volts / mA measurement
±
Available certifications include:
• NIST traceable – standard
• CE Mark
• Intrinsically Safe – MET Laboratories
per CSA C22.2 & UL 913 Class I
Division I, Groups A, B, C & D
VMA Module Accessories
P/N A900529-00015 VMA Test Lead Kit banana plugs on 9" breakouts (both ends),
assorted connectors (required for source and
simulate functions)
Yokogawa Corporation of
America
Dart Road, Newnan, GA 30265, USA ©Copyright November 2004
GS 61A
YPCVMA-01E-A
Vma Setup
Auto-Step 1:
Start
Mode: mA Source
Low pt: 4.000 mA
Upr pt: 20.000 mA
Steps: 5
Auto-Step 3
Dwell Time: Manual
Up Down Select Back
Measure
S1: 24.00 V
S2: 0.000 PSIG
S3: 72.05 deg F
IV: 0.000 mA
Pon Poff More
Sample
Transmitter Simulation M ode
Xmitter Sim
Output = 16.000 mA
IV = 16.042 mA
Up Down Select Back
YPC Displays and Wiring Diagrams for VMA0055 module with YPC 40xx
24V dc Loop P
ower
V or mA Sourcing
All Rights Reserved ©Copyright November 2004,
Subject to change without notice.
ransmitter Sim
T
Yokogawa Corporation of America GS 61A YPCVMA-01E-A
ulation Mode