YOKOGAWA US300PM User Manual

Page 1
Manual Change for IM 01G05B02-01E 1st Edition
: No.15-021-E Nov. 24 2015-00
Attention!
The instrument must not be used in a corrosive atmosphere.
Attention!
The current outputs may temporarily turn unstable during the switch-on sequence and parameter display / setting mode. Take care of your process not to be affected by this behavior.
Users Manual
US300PM Ultrasonic Flowmeter
Manual Change
No. 15-021-E
Addition to the chapter 2. Handling Precautions” (page 7) :
Changes to the section 2.1 Scope of Delivery” (page 7) and the section “3.4.2 Rear Panel” (page 13):
Change the figure of a nameplate to as below.
Addition to the section 2.3 General Precautions” (page 8) :
・The power supply adapter and the AC cable are dedicated to US300PM. Please do not use them with other
devices.
Do not disassemble or remodel.
Addition to the chapter “4. Getting Started” (page 15) :
Page 2
Manual Change for IM 01G05B02-01E 1st Edition
: No.15-021-E Nov. 24 2015-00
RESET
INIT
BRK
ENTER
[Operation during the measurement or menu display] RESET: Press these three keys simultaneously to recover from an error. This has
the same effect as restarting the unit. Data will not be affected.
INIT (cold start): Press these three keys simultaneously and release ENTER first.
After acknowledging the display of main menu, release (i) BRK key first and then (ii) C key. This will initialize the instrument. Most parameters and settings are reset to the factory default values. The memory will not be cleared. Note that when the data logging function is activated, the DELETE MEAS.VAL. display will appear instead of main menu. In this case, after releasing (i) BRK key first and then (ii) C key, select NO or YES and then press ENTER to finish the procedure.
ON
INI
T
BRK
[Operation when powering the instrument on] INIT (coldstart): Pressing these two keys simultaneously and after acknowledging the display
of main menu, release (i) BRK first and then (ii) C. This will initialize the instrument. Most parameters and settings are reset to the factory default values. The same procedure is required as above when the DELETE MEAS.VAL. display appears.
A:Reflection Mod no >YES<
Attention!
When using mounting fixtures, there may arise some air gap between the transducer surface and pipe wall because of any distortion of the pipe wall. Make sure to avoid having such air gap between them.
Addition to the section 4.2.1 Key Operations” (page 16) and related pages :
Refer to the following table for the function of corresponding key operations. The correct explanations for these key operations in the related pages are also as the table below.
Related pages: 16.2 US300PM doesnt react anymore” (page 113)
Addition to the section “5.6 Selection of the Sound Path Factor” (page 30) :
When you select “auto” for “Sound Path” described at the section “11.2 Settings for Dialogues and Menus”, the
setting for the sound path factor will be automatic. In this case, “Reflection Mod” display will appear and you
can select either “Diagonal Mode” by “no” or “Reflection Mode” by “yes”.
When the pipe outer diameter is more than 600mm, the sound path factor as “1” (one) is recommended. Otherwise, the measurement may become unstable when flow velocity or fluid temperature changed.
Addition to the section “5.7.2 Mounting of the Transducers” (page 31) :
The pipe wall thickness may slightly vary from part to part. Check it in advance by applying a wall thickness probe or some other ways and avoid mounting the transducers on such parts.
Addition to the section “5.7.2.4 Mounting with Fixtures” (page 32) :
Page 3
Manual Change for IM 01G05B02-01E 1st Edition
: No.15-021-E Nov. 24 2015-00
Attention!
In case of mass flow, select Other Medium as a measured fluid. When Other Medium has been selected, US300PM requests to enter the density which is used to calculate mass flow.
To reset the two flow totalizers to zero:
Press key
ON
O
three times when a totalizer is displayed.
To deactivate flow totalizing:
Press key
OFF
O
three times when a totalizer is displayed.
Sound Path auto >USER<
Attention!
It is necessary to activate flow totalizers by pressing key
ON
O
to get the actual pulse
output. Refer to the section 7.2 for flow totalizers.
Addition to the section “6.1 Selection of the Physical Quantity and of the Unit of Measurement” (page 37) :
Changes to the section 7.2 Flow Totalizers (page 41) :
Two of the key operations have been changed as below.
Addition to the section “11.2 Settings for Dialogues and Menus” (page 72) :
“Sound Path” setting appears next to “Meas.Point No.” and you can select “auto” or “user”. When you select
“auto”, the setting for the sound path factor will be automatic.
Changes to the section “11.6 Charging the Battery” (page 77) :
Change The chargeable NiCd-batteries guarantee an operating time of approximately 14 hours. to The chargeable NiCd-batteries guarantee an operating time of approximately 10 hours at room temperature (20°C / 68°F), with full charge.
Addition to the section “15.5 Activation of a Pulse Output” (page 101) :
Maximum pulse output rate is 2 pulse per second (+/-20%).
Set the Pulse Value and the Pulse Width so that the actual maximum flow in the pipe is less than half the displayed value (Max – Value). Example: The actual maximum flow 120m3/h Pulse Value = 0.02m3 Pulse Width = 150ms
INFO: Max – Value
240.0m3/h
Changes to the chapter A Standard Specification” (page 117 to 124) : Please use the following sheets for this chapter.
Page 4
Standard Specifications
A Standard Specifications
US300PM
General
Fluid:
Liquid (Turbidity < 10,000 mg/L, Sound velocity 800 to 3,500 m/s, Temperature –30 to +200°C / –22 to +392°F)
Measured Quantities:
Volume flow, mass flow (by setting density), flow
velocity, sound velocity in the fluid
Wall thickness of the pipe (available when optional
wall thickness probe is provided)
Measuring Principal:
Transit time method using ultrasonic signal
Pipe Sizes:
25 to 6,500 mm (1 to 255 inches) (covered by three types of transducers)
Pipe and Lining Materials:
Carbon steel, Stainless steel, Grey cast iron, Ductile iron, Copper, Glass, PVC, etc
Flow Velocity Range:
±0.01 to ±25 m/s (±0.033 to ±82 ft/s)
Resolution:
0.025 cm/s (0.01 in/s)
Repeatability:
0.15% of reading ±0.01 m/s (0.033 ft/s)
Accuracy:
(Note) Under fully developed rotationally symmetrical flow
profile
Volumetric flow:
±1 to 3% of reading ±0.01 m/s (0.033 ft/s) depending on pipe geometry and accuracy of entered pipe dimensions.
Flow velocity:
±0.5% of reading ±0.01 m/s (0.033 ft/s) over sonic path
Measuring Cycle:
100 to 1000 Hz (per one channel)
Straight Pipe Run in the Upstream:
10 to 50 pipe diameters, depending on the kind of flow disturbances
Ultrasonic Flowmeter, Main Unit
Housing Material:
Aluminum
Painting:
Powder coating
Dust and Water-proof:
IP54 (EN60529)
Measuring Input (flow / wall-thickness):
Two as standard (Channel A, Channel B). Both flow transducers and a wall thickness probe can be connected freely to whichever channel.
Current Output:
None (standard), one, or two (optional) Range 4 to 20 mA (Load resistance 0 to 500 )
Note: The current outputs may temporarily turn unstable
during the switch-on sequence and parameter display / setting mode. Take care of your process not to be affected by this behavior.
Frequency Output:
None (standard), or one (optional) Range 0 to 1 kHz Contact type: Open-collector, 24 V / 4 mA
Binary Output:
None (standard), one, or two (optional) Contact type: Open-collector, 24 V / 4 mA
Input / Output Terminal Configuration:
Round socket connectors for flow and wall-thickness measuring inputs on the front panel. Banana plug jacks (+,–) for current, frequency, and binary outputs in the rear panel.
LCD Display:
2 line x 16 character LCD display with backlight that can be switched on/off. Configurable to display two measured values (e.g., flow rate and total flow) simultaneously, or to display values from dual input channels alternately.
LED Lamp:
SIGNAL Lamp (green/red):
Indicates input signal status with green/red light, equipped for each channel.
BATTERY Lamp (red):
Indicates battery charging status with red light.
Keyboard:
15 keys (numerical, functional, or both) including four arrow-shaped keys for cursor operation, enabling easy access through its interactive menu structure.
Display Language:
Czech, Danish, Dutch, English (default), French, German, Norwegian, Polish, Turkey
Flow Measurement:
Flow velocity, Volume flow, Mass flow
Sound Velocity / Signal Amplitude Measurement (On­line):
Sound velocity and signal amplitude in the fluid available on-line simultaneously with flow measurement.
Totalization Function:
Totalizes the volume flow or mass flow. Ten-digit number for both forward/reverse directions of each channel.
Damping Function:
Time constant 0 to 100 seconds, moving average.
IM 01G05B02-01E 1st Edition 1
Page 5
Standard Specifications
Sound Velocity Measurement (Off-line):
Measures the sound velocity of unknown fluid starting from its estimated sound velocity. The result can be transferred to the current fluid parameter.
Wall Thickness Measurement (Off-line):
Measures the wall thickness of unknown pipe (material­known) using optional wall thickness probe. The result can be transferred to the current pipe parameter.
Arithmetic Operation on Dual-Channel / Dual-Path Inputs:
Arithmetic operations get the outcomes for calculation channel Y and Z by taking sum, average, or difference of two flow values from input channel A and B. Taking absolute values of each input independently in the calculation is also possible.
Note: Two sets of transducers are necessary.
Pulse Output (optional):
Available via optional binary outputs.
Pulse value: 0.01 to 1000 of totalization unit Pulse width: 100 to 1000 ms (±20%) Maximum output rate: 2pps (pulse/second) ±20%
Alarm Output (optional):
Available via optional binary outputs where each alarm item is assigned one-to-one. Alarm properties are also selectable for each alarm one-by-one.
Alarm items: High limit, Low limit, Flow direction
change, Quantity limit (for batch operation), and Error (measurement impossible)
Alarm properties: Normal open / Normal close, Non-
hold / Hold (at the alarm detection)
Output Signal Configuration:
Freely configurable including independent dual flow value outputs.
Current / frequency outputs: Flow velocity, Volume
flow, Mass flow, Sound velocity, or Signal amplitude
Binary outputs: Pulse or Alarm
Data Logging Function:
Data storage capacity of 27,000 values (standard) or 100,000 values (optional). Storage rate is selectable from 1 s, 10 s, 1 min, 10 min, 30 min, 1 h, or any other rate between 1 s to 43200 s (12 h) by the second. For each period of measurement, stored values are grouped by a user-defined measuring point name. Stored values can be transferred to a personal computer via RS232 serial communication port.
Communication Function:
Transfers the measured values to a personal computer or a serial printer. Both on-line/off-line transfer during/after the measurement available.
Output item: Flow value (flow velocity, volume flow,
or mass flow), Totalization (forward, reverse), Sound velocity, Signal amplitude
Output format setting: Spacing (for printer), Decimal
point character, Data delimiter
Communication interface: RS232 Communication port: D-sub 9-pin, male
Time-Programmable Measurement Function:
Automatic start / stop of the measurement without human operation using internal clock for specific systems. Can be used with current output, binary output, data logging function, and/or communication function for recording the measurement.
Site Parameters Storage Function:
Eliminates the necessity of re-entering parameters for additional measurement on a site. Totally 80 sets of site parameters for pipe/fluid settings are available with user-defined site names. Additionally, 14 sets of separate site parameters for not only pipe/fluid but also output-options setting in sets available.
Material / Fluid List Customization Function:
The lists of material / fluid in the parameter menu are editable. Unnecessary items can be cut off from the
menu for user’s convenience. Registration of new
material / fluid data is also possible totally up to 13 items with user-defined names.
Power Supply:
Power supply system:
Built-in rechargeable battery (6V/4Ah) Power supply adapter (Input: 100 to 240VAC,
50/60 Hz, works also as battery charger)
Battery operating time (with full charge):
Maximum around 10 hour continuous measurement at room temperature (20°C / 68°F)
Power consumption: Less than 15W
Safety and EMC Standard:
General safety: EN61010-1:2010 EN61010-2-030:2010
Altitude at installation site:
Max. 2000 m above sea level
Overvoltage category:
Overvoltage category II
Pollution degree:
Pollution degree 2
EMC regulation: EN61326-1:2006
Operating Conditions:
Ambient temperature: –10 to +60°C (14 to 140°F) Ambient humidity: 0 to 95% RH or less (Non-condensation)
(Note) Lengthy operation at 80% or more is not
recommended.
Transducers
Basic Construction:
A set of transducers are composed of a pair of sensor
elements (often called just as “transducers”), and
transducer cables with a connector at one end, which is to be inserted into the measuring input sockets in the front panel of US300PM main unit. The transducer cables are armed with stainless steel flexible tube.
(Note) Fixing hardware (mounting fixtures, chains) and
acoustic couplant are usually included in a set of transduces, specified in its model code by their kinds or with/without.
Material:
Case of sensor elements:
Stainless steel EN/DIN 1.4571 (JIS SUS 316Ti, AISI 316Ti SS equivalent)
Sensing surface of sensor elements:
General temperature type: PEEK (Poly Ether Ether Keton) High temperature type: Polyimid
2 IM 01G05B02-01E 1st Edition
Page 6
Standard Specifications
Dust and Water-proof:
General type: IP65 (EN60529) Immersible type: IP67 (EN60529)
Applicable Pipe Sizes (inner diameter):
Medium type: 25 to 400 mm (1 to 16 inches) Lager type: 100 to 2,500 mm (4 to 98 inches) Very large type: 2,000 to 6,500 mm (78 to 255 inches)
(Note) Only “6,500 mm” above is the size for outer diameter.
Fluid Temperature Range:
General temperature type:
–30 to +130°C (–22 to +266°F)
High temperature type:
–30 to +200°C (–22 to +392°F)
Note: Pay attention also to temperature specification of the
couplant you choose.
Mounting Fixture Variety and Usage:
Retaining clip:
Inserted into the groove on the upper side of transducers. Transducers are fixed onto the pipe by fixing chains hooked at the retaining clip.
Mounting fixture standard type:
Consisted of two metal blocks with a slide-in ruler to hold transducers in place and adjust their distance easily. Transducers are fixed onto the pipe by fixing chains hooked at the mounting fixture.
Mounting fixture magnetic type:
Pieces of magnet are attached onto both sides of blocks for the “standard type” above. When the pipe material is what magnet pulls, transducers are fixed onto the pipe by magnetic power eliminating the use of fixing chains. If necessary, magnet pieces can be easily added for stronger magnetic power.
Optional Extension Cable
Extends the length of transducer cable when the standard length is not enough for the site conditions. Select the extension length from 5 m (16 ft), 10 m (32 ft), or 20 m (65 ft).
Wall Thickness Probe
Specified in the option code of US300PM main unit. Model codes as accessories also available for separate/additional orders (listed later).
Fluid Temperature Range:
General temperature type: –20 to +60°C (–4 to +140 °F) High temperature type: 0 to +200°C (32 to 392°F)
Measuring Performance:
Measuring range: 1.0 to 200 mm (0.04 to 7.8 inches), depends on the material Resolution: 0.01 mm (0.0004 inch)
Accessories
Standard Accessories for US300PM Main Unit:
Transportation case, Measuring tape, and Built-in battery set always come with US300PM.
Fixing Hardware, Couplant, etc:
Specified in the suffix code of US300FT transducers. Model codes as accessories also available for separate/additional orders (listed later).
Power Supply Adapter and AC cable:
The power supply adapter and the AC cable are dedicated to US300PM. Please do not use them with other devices.
Data Transfer Software
General:
The software installed on a personal computer receives one or more records of logging data and parameter sets stored in US300PM main unit via RS232 communication port. Data can be viewed or graphed on a PC monitor, or exported as a text file. This software works while US300PM is off-line (not measuring flow) where current output goes down to zero.
Function:
Displaying Parameter Record:
Display parameter record of selected measuring data set.
Displaying Measured Data in Table:
Display measured data of selected measuring data set in table format.
Displaying Data in Graph:
Display measured data of selected measuring data set in graphic format. Marker type and color for each line of values selectable. Scales for time-axis and value-axis can be changed from default condition of automatic scaling. Graph printing function embedded.
Displaying Statistical Data:
Display statistical data of the measurement. Total data points, minimum, maximum, average and standard deviation of the measured data can be shown. Data range for statistical processing can be designated if necessary.
Exporting Text File:
Parameter record and measured data can be exported to a text file. Options for exporting items or their formats are available.
Entering Remarks:
User’s remarks for each measured data can be
entered and edited in the transferred data file on a personal computer. Remarks can be displayed in the main window of the software.
Display Language:
English, German (Note) Help is available only in English.
Operating Environment:
Personal Computer:
Microsoft® Windows® hardware compatible, one or more RS232 port
Operating System:
Microsoft® Windows® 98, ME, NT, 2000, XP
Standard Accessories:
RS232 cable, RS232 adapter 9/25
* Microsoft® and Windows® are registered trademarks
of Microsoft Corporation in the United States and/or other countries.
IM 01G05B02-01E 1st Edition 3
Page 7
Units of Measurement
Volume flow
Flow velocity
Mass flow
Totalizers
Sound velocity
Volume
Mass
m3/h
m/s
g/s
m3 g m/s
m3/min
inch/s
t/h
l
kg
m3/s
kg/h
gal t
l/h
kg/min
l/min
l/s
USgph
USgpm
USgps
bbl/d
bbl/h
bbl/m
1 gallon [US] = 3.78 l; 1 barrel = 42 gallons = 158.76 l
Standard Specifications
4 IM 01G05B02-01E 1st Edition
Page 8
Standard Specifications
Model
Suffix code
Specification
US300PM
·····························
Ultrasonic flowmeter, portable type
Output
-A0 ······················
No current output
-A1 ······················
One current output
-A2 ······················
Two current outputs
Power Supply Adapter and AC cable
1 ···················
Japan
2 ···················
USA
3 ···················
Europe
4 ···················
United Kingdom
5 ···················
Australia
6 ···················
South Africa
-2 ···············
Always 2
-N ··········
Always N
Option
/PU1 ····
One binary (pulse or alarm) output (open-collector) (*1)
/PU2 ····
Two binary (pulse or alarm) outputs (open-collector) (*1)
/FQ1 ····
Frequency output (open­collector, 0 to 1kHz)
/DLX ····
Data logging extension (100,000 values)
/BGT ····
Tag number on nameplate (in the nameplate label, maximum 16 characters)
/WTG ···
Wall thickness probe for general temperature (–20 to +60°C / –4 to +140°F)
/WTH ···
Wall thickness probe for high temperature (0 to +200°C / 32 to 392°F)
Model
Suffix code
Specification
US300PC
····················
Optional extension cable for portable type
Length
-A005 ··········
-A010 ··········
-A020 ··········
Cable length 5 m (16 ft) Cable length 10 m (32 ft) Cable length 20 m (65 ft)
Model
Suffix code
Specification
US300SA
····················
Data transfer software (Windows version) Including connecting kit (RS232 cable for connection, RS232 adapter 9/25)
Language
-1 ················
English / German version
00 ····
Always 00
Model
Suffix code
Specification
US300PT
······························
Transducers for portable type
Usage
-G ·························
General purpose (IP65)
-W ························
Immersible (IP67)
Pipe Size / Fluid Temperature (*3)
BG ·····················
Medium & General (with 3 m / 9.8 ft cable)
BH ·····················
Medium & High (with 3 m /
9.8 ft cable)
CG ·····················
Large & General (with 4.4 m / 14.4 ft cable)
CH ·····················
Large & High (with 4.4 m /
14.4 ft cable)
DG ·····················
Very large & General (with 12 m / 39.4 ft cable)
Mounting fixture (*4)
-C ···················
Retaining clip type (set of two clips)
-S ···················
Standard type (set of two blocks, including ruler marked length 320 mm, equivalent to 12.6 in.)
-M ···················
Magnetic type for general temp.
(–30 to +100°C / –22 to +212°F, set of two blocks, including ruler marked length 320 mm, equivalent to 12.6 in.)
-N ···················
None
Fixing chain (*4)
B ·················
For 25 to 1200 mm (1 to 47 in.)
Fixing chains of 2 m / 6.5 ft length (2 x 2)
C ·················
For 1200 to 3000 mm (47 to 118 in.)
Fixing chains of 2 m / 6.5 ft length (5 x 2)
D ·················
For 3000 to 6500 mm (118 to 255 in.)
Fixing chains of 2m / 6.5 ft length (11 x 2)
N ·················
None
Acoustic couplant
G ··············
General temperature type (– 30 to +130°C / –22 to +266°F)
H ··············
High temperature type (–30 to +200°C / -22 to +392°F)
N ··············
None
Option
/TTP ······
Transducer tag plate (maximum 16 characters)
Model and Suffix Code
Ultrasonic flowmeter, portable type
Transducers for portable type (*2)
*1: Option /PU1 and /PU2 are exclusive.
Optional extension cable for portable type
Data transfer software
*2: Two sets of transducers are necessary when applying dual
channel/path measurement.
*3: The alphabetic characters in the suffix code represent pipe
sizes and fluid temperature ranges are below. B: Medium type (25 to 400 mm / 1 to 16 in.) C: Large type (100 to 2,500 mm / 4 to 98 in.) D: Very large type (2,000 to 6,500 mm / 78 to 255 in.) G: General temperature (–30 to +130°C / –22 to +266°F) H: High temperature (–30 to +200°C / –22 to +392°F)
*4: When selecting code –C or –S for mounting fixture, always
select code B, C, or D for fixing chain. When selecting code –M for mounting fixture, fixing chains are optional. When selecting code –N for mounting fixture, select also code N for fixing chain.
IM 01G05B02-01E 1st Edition 5
Page 9
Standard Specifications
Model
Description
USPA301
Wall thickness probe for general temperature (–20 to +60°C / –4 to +140°F)
USPA302
Wall thickness probe for high temperature (0 to +200°C / 32 to 392°F)
USPA311
Power supply adapter (100 to 240 V AC input)
USPA321
AC cable (Japan)
USPA322
AC cable (USA)
USPA323
AC cable (Europe)
USPA324
AC cable (United Kingdom)
USPA325
AC cable (Australia)
USPA326
AC cable (South Africa)
USPA331
Battery set (6 V 4Ah) (*5)
USPA341
Transportation case (*5)
Model
Description
USPA401
RS232 cable (*6)
USPA402
RS232 adapter 9/25 (*6)
USPA411
Measuring tape (*7)
Model
Description
USPA001
Fixing strap (10 m / 32 ft length)
USPA002
Fixing strap (20 m / 65 ft length)
USPA011
Fixing clips (medium type, for pipe size 40 to 100 mm / 1.5
to 4 in., set of two clips)
USPA012
Fixing clips (large type, for pipe size 100 to 6500 mm / 4 to
255 in., set of two clips)
USPA021
Fixing bands (only for pipe size 25 to 50 mm / 1 to 2 in.)
USPA032
Fixing chains (set of two extensible chains) (2 m / 6.5 ft length, equal to 600 mm / 23 in.
diameter)
USPA033
Repair set for fixing chains
USPA034
Retaining clips (set of two clips, used with fixing chains)
USPA036
Fixing chains (set of two extensible chains) (0.5m / 1.6 ft length, equal to 150mm / 5.9 in.
diameter)
USPA037
Fixing chains (set of two extensible chains) (1 m / 3.2 ft length, equal to 300 mm / 11.8 in.
diameter)
USPA053
Mounting fixture short ruler type (for transducers medium pipe size type, temperature range –30 to +200°C / –22 to +392°F, set of two blocks with 120 mm / 4.7 in. ruler)
USPA054
Mounting fixture standard type (for transducers medium pipe size type, temperature range –30 to +200°C / –22 to +392°F, set of two blocks with 320 mm / 12.6 in. ruler)
USPA055
Mounting fixture magnetic general temperature type (for transducers medium pipe size type, temperature range –30 to +100°C / –22 to +212°F, set of two blocks with 320 mm / 12.6 in. ruler)
USPA057
Mounting fixture standard type (for transducers large or very large pipe size type, temperature range –30 to +200°C / –22 to +392°F, set of two blocks with 320 mm / 12.6 in. ruler)
USPA058
Mounting fixture magnetic general temperature type (for transducers large or very large pipe size type, temperature range –30 to +100°C / – 22 to +212°F, set of two blocks with 320 mm /
12.6 in. ruler)
USPA073
Additional magnets for mounting fixture magnetic general temperature type
(for transducers medium pipe size type, temperature range –30 to +100°C / –22 to +212°F, set of two magnets)
USPA075
Additional magnets for mounting fixture magnetic general temperature type
(for transducers large or very large pipe size type, temperature range –30 to +100°C / –22 to +212°F, set of two magnets)
USPA081
Ruler for mounting fixture (marked length 120 mm, equivalent to 4.7 in.)
USPA082
Ruler for mounting fixture (marked length 320 mm, equivalent to 12.6 in.)
USPA091
Acoustic couplant general temperature type (100 g, –30 to +130°C) (0.22 lb, –22 to +266°F)
USPA092
Acoustic couplant high temperature type (100 g, –30 to +200°C) (0.22 lb, –22 to +392°F)
Accessories (for ultrasonic flowmeter US300PM)
*5: Always comes with US300PM.
Accessories (others)
*6: Included in data transfer software US300SA as standard. *7: Included in US300PM as standard.
Accessories (for transducers US300PT)
6 IM 01G05B02-01E 1st Edition
Page 10
Dimensional Drawings
Length / Weight
Transducer
X m (inch)
Y m (inch)
X+Y m (inch)
Weight kg (lb)
US300PT-B
2.0 ( 78.7)
1.0 ( 39.4)
3.0 (118.1)
0.6 (1.32)
US300PT-C
2.0 ( 78.7)
2.4 ( 94.5)
4.4 (173.2)
1.2 (2.65)
US300PT-D
5.0 (196.9)
7.0 (275.6)
12.0 (472.4)
2.2 (4.85)
Length / Weight
Optional extension cable
L m (inch)
Weight kg (lb)
US300PC-A005
5.0 (196.9)
0.4 (0.88)
US300PC-A010
10.0 (393.7)
0.6 (1.32)
US300PC-A020
20.0 (787.4)
1.6 (3.53)
Ultrasonic flowmeter US300PM
Standard Specifications
Unit: mm (inch)
Transducers US300PT-B
Transducers US300PT-C, US300PT-D
Optional extension cable US300PC-A□□□
IM 01G05B02-01E 1st Edition 7
Page 11
Standard Specifications
Length / Weight
Wall thickness probe
L m (inch)
Weight kg (lb)
/WTG or USPA301
1.5 (59.1)
0.17 (0.37)
/WTH or USPA302
1.5 (59.1)
0.24 (0.53)
Wall thickness probe for general temperature (–20 to +60°C / –4 to +140°F) (option /WTG or model USPA301)
Wall thickness probe for high temperature (0 to +200°C / 32 to 392°F) (option /WTH or model USPA302)
Unit: mm (inch)