YOKOGAWA US300FM User Manual

User’s Manual
US300FM Ultrasonic Flowmeter
Manual Change
No. 15-010-E
This is the supplement for the original manual (No. IM 01G05B03-01E, 1st Edition) regarding the added and changed items. Please also refer to this supplement when you read the manual.
Addition to the section “3.5.4 Connection of the Transducers” (page 18) :
(1) Cable Finish
The cable finish should be done as the drawings (a) and (b) shown in the right.
(2) Cable Gland
When tightening the gland, not only pull back the cable shield (meshed part) but also roll it out into a ring-shaped form and let it be pushed by the compression part (grey-colored plastic) so that the cable shield will have electrical contact with the gland body. Also, be sure to align the convexity part outside the compression part and the concavity part inside the gland body. Refer to the drawings below.
Cable shield rolled out into a ring-shaped form
Compression part (convexity part outside)
Components of the cable gland
Gland body (concavity part inside)
(a) Main unit side
Perform soldering
7±2
13±3
White White
60±5
12±2
Cable shield
(meshed)
(*) Finish the brown cables as well as white cables. (**) You do not connect the thin lines to any terminals.
Gland body
Cap nut
Rubber ring
Compression
part
Cutaway view of the cable gland
Unit : mm
Shield line
Brown (*)
Less than
10mm
Thin lines
Outer skin of the cable
Cable shield (ringed-shape)
Signal lines
(b) Junction box side
52±5
Cable shield
(**)
(meshed)
Outer wall of the main unit or junction box
Perform soldering
7±2
Shield line
10±3
12±2
(3) Grounding
The connection of the cable shield will be realized as shown in the drawing below after going through the procedures (1) and (2) in the above. Connect the main unit to the ground through the terminal “PE” of the power supply terminals in the main unit. The grounding resistance should be less than 100 ohm.
Transducers
Sensors
Junction box
Connection
cable
Main
unit
Grounding resistance less than 100 ohm
GND
Connection of the cable shield
(4) Connecting the Sensor ROMs
Connect the sensor ROMs to the corresponding terminals in the main unit when connecting transducers. The sensor ROMs are packed and affixed to the transducers when shipped from the factory.
Brown (*)
Less than 5mm
Thin lines
(**)
Manual Change for IM 01G05B03-01E 1st Edition : No.15-010-E July 30 2015-00
Yokogawa Electric Corporation
Changes to the section “3.1 Scope of Delivery” (page 11) :
Change the figure of a nameplate to as below.
Addition to the section “3.1 Scope of Delivery” (page 11) :
When ordering a couplant weatherproof type (by couplant code “R” in the transducer specification or model code USPA097), a package of “Shin-Etsu Silicone, 1 COMPONENT RTV (KE45T)” is delivered.
Addition to the section “3.3 General Precautions” (page 12) :
Do not disassemble or remodel.
Addition to the section “3.5.1 Location” (page 13) :
Attention!
The instrument must not be installed in a corrosive atmosphere.
Attention!
Use Submersible(IP68) type transducers when they are installed where condensation occurs. A Tokuchu request is necessary for Submersible type. However the diameter must be over 100 mm.
Changes to the section “3.5.5 Connection to the Power Supply” (page 20) :
(1) Terminal Block (page 19)
The outer shapes of the terminal block for power supply connection, the printed terminal names, and the design on the board below the terminal block have been changed as the drawing in the right.
(2) Terminal Assignment (page 21)
Regarding the terminal assignment, the terminal names were revised as below from the original manual. The needed power supply is given in the nameplate on the housing, not in the plate below the terminal block.
N(-)PE L(+)
KL3
The terminal names are the same both for AC and DC power supply.
This can be used as the earth terminal instead of the terminal “PE”.
AC power supply DC power supply
Terminal Connection Terminal Connection PE Earth PE (Earth) N(-) Neutral N(-) - DC L(+) Phase 100...230 V AC±10% L(+) + DC 24V DC±10%
Manual Change for IM 01G05B03-01E 1st Edition : No.15-010-E July 30 2015-00
Addition to the section “3.5.5 Connection to the Power Supply” (page 20) :
Install an external switch or circuit breaker as a means to turn the power off (capacitance; 15A, conforming to IEC947-1 and IEC947-3). Locate this switch either near the instrument or in other places facilitating easy operation. Affix a “Power Off Equipment” label to this external switch or circuit breaker.
Changes to the section “3.6 Sensor ROM” (page 22) :
The sensor ROMs are packed and affixed to the transducers when shipped from the factory, not inserted to the terminals in the main unit. They shall be connected to the corresponding terminals in the main unit when installation.
Addition to the section “3.7 Replacement of the Fuse” (page 22) :
Be sure to use the fuse which satisfies the specification below.
Components may otherwise be damaged.
Changes to the section “3.7 Replacement of the Fuse” (page 22) :
Change the fuse specification “250V 1.25A, delayed action Type T” to as below.
“250V 1A delay type” for AC power supply “250V 1.6A delay type” for DC power supply
Addition to the chapter “4. Getting Started” (page 23) :
Attention!
The current outputs may temporarily turn unstable during the power-on sequence (including re-starting after power failure) and parameter display / setting mode. Take care of your process not to be affected by this behavior.
Addition and deletion to the section “4.1.1 Key Operations” (page 23) and related pages :
Refer to the following table for the function of corresponding key operations. The correct explanations for these key operations in the related pages are also as the table be low.
Related pages: “4.1.1 Interruption of Power Supply” (page 27) “15.2 US300FM doesn’t react anymore” (page 106)
LIGHT
Switches the background lighting ON/OFF.
BRK
INIT
RESET
ENTER
[Operation when powering the instrument on]
[Operation during the measurement or menu display] RESET: Press these three keys simultaneously to recover from an error. This
has the same effect as restarting the unit. Data will not be affected.
INIT (cold start): Press these three keys simultaneously and release ENTER
first. After acknowledging the display of main menu, release (i) BRK key first and then (ii) C key. This will initialize the instrument. Most parameters and settings are reset to the factory default values. The memory will not be cleared. Note that when the data logging function is activated, the DELETE MEAS.VAL. display will appear instead of main menu. In this case, after releasing (i) BRK key first and then (ii) C key, select NO or YES and then press ENTER to finish the procedure.
BRK
INIT
INIT (coldstart): Pressing these two keys simultaneously and after acknowledging the
display of main menu, release (i) BRK first and then (ii) C. This will initialize the instrument. Most parameters and settings are reset to the factory default values. The same procedure is required as above when the DELETE MEAS.VAL. display appears.
Manual Change for IM 01G05B03-01E 1st Edition : No.15-010-E July 30 2015-00
Addition to the section “4.4.1 Interruption of Power Supply” (page 27) :
Attention!
The current outputs may temporarily turn unstable during the power-on sequence after power failure. Take care of your process not to be affected by this behavior.
Addition to the chapter “5. Basic Measurement” (page 29) :
Note!
When changing any settings in "PARAMETER" program branch, be sure to also go through "MEASURING" program branch to the end where the meas urement will start. In this case, you do not always have to fix the transducers on to the pi pe, and thus you do not always need to get the actual flow measurement.
If you shut off the power supply without taking this procedure, the settings to be c hang ed would not be effective and keep the same settings as before.
Addition to the section “5.4 Selection of the Sound Path Factor” (page 34) :
When the pipe outer diameter is more than 600mm, the sound path factor as “1” (one) is recommended. Otherwise, the measurement may become unstable when flow velocity or fluid temperature changed.
Addition to the section “5.5.2 Mounting of the Transducers” (page 35) :
The pipe wall thickness may slightly vary from part to part. Check it in advance by applying a wall thickness probe or some other ways and avoid mounting the transducers on such parts.
Attention!
Instructions for using “Acoustic couplant weatherproof type” (Shin-Etsu Silicone, 1 COMPONENT RTV) are as below.
The operating condition of ambient temperature
Before and during curing : 0 to +50°C / 32 to 122°F After curing : -40 to +180°C / -40 to 356°F
When applying this couplant, cut the tip of attached nozzle to appropriate length and
set it to the tube container. Also, remove matters like dust or oil on the surface to be applied. Otherwise, its adhesion force may become lower.
When curing this couplant, keep the ambient and installation position temperature
between 0 to +50°C (32 to 122°F). If not, the flow measurement may fail because of air bubbles formed inside the rubber before curing, etc.
It usually takes one to three days for complete curing depending on the conditions.
Addition to the section “5.5.2.2 Mounting with Fixtures and Chains” (page 36) :
Attention!
When using mounting fixtures, there may arise some air gap between the transducer surface and pipe wall because of any distortion of the pipe wall. Make sure to avoid having such air gap between them.
Addition to the section “6.1 Selection of the Physical Quantity and of the Unit of Measurement” (page 39) :
Attention!
In case of mass flow, select “Other Medium” as a measured fluid. When “ Other Medium” has been selected, US300FM requests to enter the density which is us ed to calculate mass flow.
Manual Change for IM 01G05B03-01E 1st Edition : No.15-010-E July 30 2015-00
Loading...
+ 8 hidden pages