WIKA PCA-HP User Manual

NI-281E
c
c
A
A
INSTRUCTION MANUAL
Rev. 5 11/02
COMPACT PRESS. SWITCHES SERIES PCS & PCA
WEATHERPROOF EXPLOSIONPROOF
SERIES PCS SERIES PCA
B
B = Cable entry A = Pressure connection WEIGHT 1kg dimensions in mm
NOTE: dimensions and weights are not binding unless released on certified drawings.
CAUTION
Before installing, using or carrying out maintenance on the instrument it is necessary to read and understand the indications given in the attached Instruction Manual.
The instrument must only be installed and maintained by qualified personnel.
INSTALLATION IS TO BE CARRIED OUT ONLY AFTER CHECKING THAT INSTRUMENT CHARACTERISTICS ARE
CONSISTENT WITH PROCESS AND PLANT REQUIREMENTS
The functional features of the instrument and its degree of protection are shown on the identification plate fixed to the case.
CONTENTS:
GENERAL NOTES
1 -
2 - OPERATING PRINCIPLE
3 - IDENTIFICATION PLATE AND MARKINGS
4 - SET POINT REGULATION 5 - SET POINT CALIBRATION 6 - MOUNTING AND CONNECTIONS 7 - INSTRUMENT PLUMBING 8 - PUTTING INTO OPERATION 9 - FUNCTIONAL VERIFICATION 10 - FUNCTIONAL VERIFICATION 11 - STOPPING AND DISMOUNTING 12 - DEMOLITION
SAFETY INSTRUCTIONS FOR USE IN HAZARDOUS ATMOSPHERES. RECOMMENDATIONS FOR PRESSURE SWITCH SAFE USE.
B = Cable entry A = Pressure connection WEIGHT 1kg dimensions in mm
B
RELATED DOCUMENT
To authentified document with certificate
N° CESI 02 ATEX 118
ll data, statements and recommendations supplied with this manual are based on information believed by us to be reliable. As the onditions of effective use are beyond our control, our products are sold under the condition that the user himself evaluates such onditions before following our recommendations for the purpose or use foreseen by him.
The present document is the property of ETTORE CELLA SPA and may not be reproduced in any form, nor used for any purpose other than for which it is supplied.
INSTRUCTION MANUAL
1 - GENERAL NOTES
1.1 FOREWORD
The wrong choice of a series or a model, as well as the incorrect installation, lead to malfunction and reduce instrument life. Failure to abide by the indications given in this manual can cause damage to the instrument, the environment and persons.
1.2 ALLOWED OVERRANGE
Pressures exceeding the working range can be occasionally tolerated provided they remain within the limits stated in the instrument features (proof pressure). Continuous pressures exceeding the working range can be applied to the instrument, provided they are clearly stated in the instrument features. The current and voltage values stated in the technical specifications and ratings must not be exceeded. Transitory over-ranges can have a destructive effect on the switch.
1.3 TEMPERATURE
Due to the temperature of both the environment and the process fluid, the temperature of the instrument could exceed the allowed limits (normally from -20°C to +70°C). Therefore, in case it does, suitable measures (protection against heat radiation, fluid separators, cooling coils, heated lockers), aimed at limiting the value, must be taken.
2 - OPERATING PRINCIPLE
A pressure-tight diaphragm (or an O-ring sealed piston) applies a force to a stiff disc. This force is directly proportional to the pressure value and is contrasted by an helicoidal spring charged by a suitable bush. When the force balance point is exceeded, the stiff disc shifts and, by means of a rigid rod, actuates one or two simultaneous release electric microswitches. The microswitches are of the snap acting type with automatic reset. When the pressure moves away from the set values, returning towards the normal values, the switch is reset.
3 - IDENTIFICATION PLATE AND MARKINGS
The instrument is fitted with a metal plate bearing all its functional characteristics and – in case of explosionproof execution (Series PCA) – also the markings prescribed by standard CEI EN 50014. Fig. 1 shows the plate mounted on explosionproof instruments.
Fig. 1 - Explosionproof instruments plate
NI-281E
Rev. 5 11/02
Prior to installation the instrument must be calibrated and the definitive calibration values written on the label using a suitable indelible ink pen.
If the instrument has been ordered with a specific calibration, it is a good rule to check the calibration values marked on the relevant adhesive label, prior to installation.
The position of the adjustment bush given in Fig. 2.
Fig. 2 - Electrical connections and adjustment screws
(one contact/two contact instruments)
1 - External earth screw 2 - Hole for test plug 3 - Adjustment rod for set point calibration 4 - Adjustment bush access plate 5 - Adjustment bush 6 - Screw for fixing the adjustment bush access plate 7 - Internal earth screw 8 - Terminal block for the second microswitch
9 - Terminal block for the first microswitch 10 - Pre-insulated thimbles - Ventilation path Microswitch electrical circuit. State of the contacts at atmospheric pressure
1 CE marking and identification number of the notified body
responsible for production surveillance.
2 Apparatus classification according to ATEX 94/9 CE
directive..
3 Type of protection and ambient temperature limits of
operation.
4 Notified body that issued the type certificate and number of
said certificate.
4 - SET POINT REGULATION
The charge of the helicoidal spring can be regulated by means of the bush (for adjustment) in such a way that the switch is released when the pressure reaches (increasing or decreasing) the desired values (set point). The instrument is usually supplied with the switches set at the setting range value nearest to zero (factory calibration). The instrument is supplied with an adhesive label showing the set point calibration value. With factory calibration the values are not indicated on the label as these are temporary and will be modified with the definitive values.
Designation of the contacts:
C - Common NA - Normally open NC - Normally closed
The effect of the direction of rotation of the adjustment bush has is described on the adhesive plate. The bush must be turned by means of the special adjustment rod which comes with the instrument (Fig.2).
5 - SET POINT CALIBRATION
In order to proceed with the calibration and the periodical functional verification of the instrument a suitable calibration circuit (Fig. 4) and an adequate pressure source is required..
INSTRUCTION MANUAL
5.1 PRELIMINARY OPERATIONS
CAUTION: do not open the cover of explosionproof pressure
switches (Series PCA) when energized, in explosive atmospheres.
With reference to Fig. 3 unscrew the screw (b) until the blocking device (a) can be turned 180°; then unscrew the cover.
Fig. 3 - Weatherproof pressure switch cover blocking
CALIBRATION CIRCUIT AND The warning lamps should be connected to the contacts in the NA or NO position according to the required contact action.
If the instrument is equipped with two microswitches, take into account that they actuate simultaneously within rated tolerances.
The warning lamps can either be connected by means of a thimble with a maximum diameter of 2.5 mm or by means of a test plug with a diameter of 2 mm to be inserted in the appropriate holes situated frontally beside the terminal screw (see Fig. 2).
Connection of C and NA terminals
If the circuit is open at the working pressure, the switch closes
the circuit as the pressure increases when the desired values is reached (MAX. closing).
If the circuit is closed at the working pressure, the switch opens the circuit as the pressure decreases when the desired value is reached (MIN. opening).
Connection of C and NC terminals
If the circuit is closed at the working pressure, the switch opens
the circuit as the pressure increases when the desired value is reached (MAX. opening).
If the circuit is open at the working pressure, the switch closes the circuit as the pressure decreases when the desired value is reached (MIN. closing). The test instrument should have a measurement range approximately equal to or slightly wider than the pressure switch range and should have an accuracy consistent with the precision required to calibrate the set point. The pressure switch must be mounted in the normal installation position, i.e. with the pressure connection downwards. Avoid forcing the microswitch by hand or with tools. This could affect the instrument functioning. With reference to Fig.3, free the access to the adjustment bush by loosening the screw (c) which holds the closure plate (f). Increase the pressure in the circuit up to the desired microswitch set point value. Turn the adjustment bush using the adjustment rod with which the instrument is equipped (Fig. 2) until the relative lamp turns on (or turns off); then turn it in the opposite direction until the lamp turns off (or on). Slowly turn the bush again until the lamp turns on (or off).
device and plumbing
a. Cover blocking
device b. Blocking screw c. Screw blocking
the adjustment
bush access
plate d. Flexible steel
wire e. Plumbing f. Adjustment
bush access
plate
NI-281E
Rev. 5 11/02
Fig. 4 - Calibration circuit
PS - Pressure switch CA - Test gauge V1 - Inlet valve V2 - Discharge vale P - Pressure source
Check the calibration value (varying the pressure in the circuit accordingly) and register it, using a pen with indelible ink, on the ratings.
5.2 CALIBRATION USING REGULATION SCALE
See the specific instructions attached with instruments with this option.
5.3 FINAL OPERATIONS
Disconnect the instrument from the calibration circuit. With reference to Fig. 2, insert the adjustment rod into the appropriate seat; close the access to the adjustment bush by rotating the closure plate (4) and tighten the relative screw (6). Take the cover, ensure that the sealing gasket is correctly fitted into its seat, insert the cover onto the case and turn it clockwise until the cover is closed. With reference to Fig. 3 turn the blocking device (a) 180° sliding the tongue into the appropriate seat in the cover; tighten the blocking screws (b).
Mount on pressure connectionn and cable entry the protection caps supplied with the instrument. CAUTION: the protection caps should only be definitively removed during the connection steps (see § 7).
6 - INSTRUMENT PLUMBING
The plumbing, aimed as a guarantee against possible tampering of the calibrations, can be carried our using a flexible steel wire (d) inserted into the holes in the screw (c) and the adjustment bush closure plate provided for this purpose (see Fig. 3).
7 - MOUNTING AND CONNECTIONS
7.1 MOUNTING
Surface mount the instrument by mean of the holes, or pipe mount using the appropriate bracket (see Fig. 8) or mount directly on the plant in a vertical position (with the pressure connection downwards). The chosen position must be such that the possibility of shocks or temperature changes are within tolerable limits. The above also applies to direct mounting. With gas or vapour process fluid, the instrument must be positioned higher than the pipe inlet (see Fig. 7). With a liquid process fluid, the instrument can be positioned higher or lower, indifferently (see Fig. 6 and 7). In this case, during set point calibration the
negative or positive head must be taken into account.
CAUTION: positions other than vertical are allowed provided
environmental conditions do not cause condensation to form or water to enter the instrument through the ventilation path.
7.2 PRESSURE CONNECTIONS
For a correct installation it is necessary to: Mount a shut-off valve with drain (root valve) on the process tube to allow the instrument to be excluded and the connection tubing to be drained. It is recommended that said valve has a capstan­blocking device aimed at preventing it being activated casually and without authorisation. Mount a service valve near the instrument to permit possible functional verification on site. It is recommended that the service
Test fluid: air for P < water for P > 10 bar
10 bar
valve is closed with a plug to prevent the outlet of the process fluid caused by the incorrect use of said valve. Mount a three piece joint onto the threaded attachment of the instrument to permit the easy mounting or removal of the instrument itself. Carry out the connection using a flexible tube in such a way that the tube itself does not force the instrument attachment due to variations in temperature. Ensure that all the pressure connections are airtight. It is important that there are no leakages in the circuit.
Close the root valve and the relative drain device. Close the service valve fitted with a safety plug.
INSTRUCTION MANUAL
7.3 ELECTRICAL CONNECTIONS
It is recommended to carry out the electrical connections according to the applicable standards. In case of explosionproof instruments (Series PCA) see also the Standard EN-60079-14. If the electrical connection is carried out in a protected tube, it shall be made so that condensate is prevented from entering instrument enclosure.
The arrangement shown in Fig. 5 or 6 is therefore recommended.
CAUTION: fittings used for the electrical connection of the
pressure switch Series PCA (explosionproof) shall be certified to Standards EN 50014 and 50018, and shall guarantee instrument degree of protection (IP65).
Check that there is no power in the lines. Remove the cover and carry out the cabling and connections to the terminal block (see Fig. 2). Flexible cables with a maximum section of 1,2 mm recommended using the pre-insulated thimbles with a maximum diameter of 2,5 mm supplied with the instrument.
When inserting cables into the enclosure pay attention not to force the microawitch with cable or tools, otherwise instrument calibration or even its operation could be compromised. The microswitch has been factory mounted and positioned in order to obtain the best performances. Any tampering made on site without following instructions authorised by the E. CELLA SPA may result in instrument malfunction.
Ensure that no deposits or wire ends remain inside the case. Once the connection operations have been completed, replace the cover and ensure that it is properly sealed and blocked. See Fig. 3.
7.4 SPECIAL NOTE FOR INSTALLATION OF CATEGORY 1 / 2 G PRESSURE SWITCHES
Explosionproof pressure switches (Series PCA) can be installed on processes requiring apparatus of group II category 1 in an ambient requiring apparatus of group II category 2 (see Fig. 5).
2
(16AWG) are
NI-281E
Rev. 5 11/02
Fig. 5 - Installation of Group II Cat. 1 / 2 G instruments
8 - PUTTING INTO OPERATION
As the signal transmitted by the instrument is used in a complex system, it is necessary that the means of putting it into operation are established by personnel in charge of the plant.
The instrument comes into operations as soon as the root valve is opened. Any possible drainage of the connection tubing can be carried out by removing the safety plug and opening the service valve with the necessary caution.
In case of explosionproof instruments (Series PCA), initial inspections are to be carried out according to customer procedures and at least in accordance with Standard EN-60079-
17.
9 - FUNCTIONAL VERIFICATION
This will be carried out according to the Client’s control procedures. Series PCS instruments can be verified on the plant if installed as illustrated in Fig. 6 and 7.
The instruments Series PCA may be checked on site only if apparatus suitable for explosive atmosphere are used and provided that the electric line is not energized.
If this is not the case it is necessary to stop operation, dismount by means of the three piece joints and carry out the verification in a test room.
CAUTION: do not open the cover of explosionproof pressure switches (Series PCA) when energized, in explosive atmospheres.
Verification consists in check the calibration value and possibly regulating the adjustment bush (see §5).
In case of explosionproof instruments (Series PCA), inspections of the electrical installation are to be carried out also according to customer procedures and at least in accordance with Standard EN-60079-17.
NI-281E
INSTRUCTION MANUAL
10 - TROUBLESHOOTING
IMPORTANT NOTE: operations involving replacement of essential components must be carried out at our workshop, especially for instruments with explosionproof certificate; this is to guarantee the user the total and correct restoration of the product original characteristics.
MALFUNCTION PROBABLE CAUSE REMEDY
Permanent deformation of the sensitive element
due to fatigue or excess over-ranges.
Set point shift
Poor repeatability
Slow response
No actuation or
undue actuation
Undue actuation
Variation of the elastic features of the sensitive
element due to its chemical corrosion.
O-ring wear (only PCS2P, PCS3P, PCA2P and
PCA3P).
O-ring wear (only PCS2P, PCS3P, PCA2P and
PCA3P).
Air bubbles or condensation (only for types with
pressure < 1 bar).
Clogged or obstructed connection line. Root valve partially closed. Too viscous fluid. Root valve closed. Microswitch contacts damaged. Loosened electrical joints. Interrupted or short-circuited. Accidental shocks.  Modify the mounting.
11 - STOPPING AND DISMOUNTING
Before proceeding with these operations ensure that the plant or machines have been put into the conditions foreseen to allow these operations.
With reference to figures 6 or 7
Remove the power supply (signal) from the electrical line. Close the root valve (6) and open the drain. Remove the plug (2), open the valve (3) and wait until the process fluid has drained from the tubing through the drain.
Do not dispose of the process fluid into the environment, if this can cause pollution or damage to people.
Unscrew the three piece joint (8).
CAUTION: do not open the cover of explosionproof pressure switches (Series PCA) when energized, in explosive atmospheres.
Unscrew the three piece joint (10) (electrical cable tubing).
Remove the instrument cover and disconnect the electrical cables from the terminal block and earth screws. Remove the screw fixing the case to the panel (or pipe) and remove the instrument, taking care to slide the electrical conductors out from the case.
Mount instrument cover. Insulate and protect cables around, if any. Temporarily plug pipes not connected to the instrument. In case of explosionproof instruments (Series DE) it is recommended to follow - at least – the standard EN-60079-17 for the withdrawal from service of electrical apparatus.
12 - DEMOLITION
The instruments are mainly made of stainless steel and aluminium and therefore, once the electrical parts have been dismounted and the parts coming into contact with fluids which could be harmful to people or the environment have been properly dealt with, they can be scrapped.
Recalibrate or replace the sensitive element. Recalibrate or replace the sensitive element with
another made of a suitable material. If necessary apply fluid separator.
Replace the piston subgroup and recalibrate. Replace the piston subgroup and recalibrate. Drain the process connection lines and if necessary
modify them.
Check and clean line. Open valve. Provide instrument with suitable fluid separator. Open the valve. Replace the microswitch. Check all electrical joints. Check the conditions of the electric line.
Rev. 5 11/02
NI-281E
INSTRUCTION MANUAL
WEATHERPROOF EXPLOSIONPROOF
Fig. 6 - Example of connections
Fig. 7 - Example of connections
Rev. 5 11/02
Fig. 8 - Surface or bracket mounting for 2" pipe
(weatherproof and explosionproof instruments)
LEGEND 1 - Fitting 9 - "T" fitting 2 - Drain plug 10 - Three piece fitting 3 - Service valve 11 - Blocking joint 4 - Piping 12 - Curve 5 - Three piece fitting 13 - M5 screws (No. 2) 6 - Root valve with drain 14 - Bracket for 2" pipe 7 - Process piping 15 - Vertical pipe 8 - Three piece fitting 16 - Horizontal pipe
NOTE: With gas or vapour process fluid, the instrument must be positioned higher than the pipe inlet (see Fig. 5). With a liquid process fluid, the instrument can be positioned higher or lower, indifferently (see Fig. 5 and 6). In this case, during set point calibration the negative or positive head must be taken into account (distance h in Fig.5 and 6).
ETTORE CELLA SPA Viale de Gasperi, 48 - Casella Postale (P.O. Box) 96 - I 20010 Bareggio (MILANO) ITALY
Telefoni +39 029036.1146/1237/1241 - FAX +39 029036.1331 e-mail: cella@ecellaspa.com
Loading...