Welch Allyn CareTemp Touch Free Directions For Use Manual

CareTemp™ Touch Free Thermometer
Directions for use
© 2018 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn. Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.
Caution Federal US law restricts sale of the device identified in this manual to, or on the order of, a licensed physician.
Software in this product is Copyright 2018Welch Allyn or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse­engineered, disassembled, or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title, and ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, call Welch Allyn Technical Support (http://www.welchallyn.com/
support) or visit www.welchallyn.com/en/about-us/locations.html.
DIR 80022132 Ver. B
Revised: 2018-07
AViTA Corporation 9F., No. 78, Sec. 1 Kwang-Fu Rd., San-Chung District, 24158 New Taipei City, Taiwan
This manual applies to 901094 Non Contact Thermometer
Wellkang Ltd 29 Harley St. W1G 9QR London, U.K.

Contents

Symbols ................................................................................................... 1
Introduction ............................................................................................. 3
Using the thermometer .......................................................................... 7
iii
Intended use ........................................................................................................ 3
General warnings and cautions ............................................................................ 3
Button functions .................................................................................................. 7
Take a measurement ........................................................................................... 8
Recall a measurement ......................................................................................... 9
Use the timer ....................................................................................................... 9
Change the temperature scale ............................................................................. 9
Troubleshooting and error messages ................................................. 11
Maintenance .......................................................................................... 13
Replace the batteries ......................................................................................... 13
Clean and disinfect the thermometer ................................................................ 13
Calibration testing .............................................................................................. 14
Disposal ............................................................................................................. 14
Specifications ........................................................................................ 17
Guidance and manufacturer's declaration .......................................... 19
EMC compliance ................................................................................................ 19
Emissions and immunity information ................................................................. 19
Warranty ................................................................................................ 25
Contents CareTemp™ Touch Free Thermometer
iv

Symbols

Documentation symbols

1
WARNING The warning statements in this manual identify conditions or practices that could lead to illness, injury, or death.
Consult operating instructions/directions for use (DFU).
A copy of the DFU is available on this website.
A printed copy of the DFU can be ordered from Welch Allyn for delivery within 7 calendar days.
Caution The caution statements in this manual identify conditions or practices that could result in damage to the equipment or other property, or loss of data. This definition applies to both yellow and black and white symbols.
Meets essential requirements of European Medical Device Directive 93/42/EEC.

Shipping, storing, and environment symbols

Temperature limits Keep dry
This device shall be disposed of in accordance with national laws after its useful life.
Fragile
Humidity limitation
IP22
This product meets the basic safety and essential performance requirements indicated in the IP22 conditioning test (IP22: Protection level against solid foreign objects and falling water)
ETL CLASSIFIED
Symbols CareTemp™ Touch Free Thermometer
2

Miscellaneous symbols

Manufacturer
Product identifier European authorized representative
Global trade identification number Serial number
Conforms to AAMI STD ES 60601-1, IEC STD 60601-1-6, ISO STD 80601-2-56; Certified to CSA STD C22.2 NO. 60601-1, NO. 60601-1-6, No. 80601-2-56
By prescription or order of licensed healthcare professional

Indicator symbols

One short beep indicating a button press Three beeps indicating the completion of
Reorder/Catalog number
Type BF applied part
a process
Two-second beep indicating an error

Battery level indicators

100 to 76 percent of useable battery capacity remaining
Approximately 3,300 measurements
50 to 26 percent of useable battery capacity remaining
Approximately 880 measurements
10 to 0 percent of useable battery capacity remaining
When the battery flashes, no measurements can be taken
75 to 51 percent of useable battery capacity remaining
Approximately 1,900 measurements
25 to 11 percent of useable battery capacity remaining
Approximately 200 measurements

Introduction

Intended use

The CareTemp™ Touch Free thermometer is a clinical-grade device intended for the intermittent measurement of human body temperature in patients of all ages in a professional-use environment.
3

General warnings and cautions

Warning and caution statements can appear on the thermometer, the packaging, the shipping container, or in this document.
The thermometer is safe for patients and clinicians when used in accordance with the instructions and with the warning and caution statements presented in this manual.
Before using the thermometer, familiarize all operating personnel with the general safety information in this summary. Specific warnings and cautions are also found throughout this manual.
Failure to understand and observe any warning statement in this manual could lead to patient illness, injury, or death.
Failure to understand and observe any caution statement in this manual could lead to damage to the equipment or other property, or loss of patient data.
WARNING Patient safety. The thermometer is designed for the intermittent measurement of the human body's temperature, and can be used upon people of all ages. The following recommendations must be carefully observed during the product's use. Any activities that are inconsistent with or do not take into account these recommendations could result in personal injury or could affect the accuracy of the thermometer itself.
WARNING Patient safety. If the accuracy of any measurement is in question, check the patient's temperature with an alternate method and then check to verify the device is functioning properly.
WARNING Safety risk. The thermometer battery must be kept strictly out of the reach of children, as ingestion of the battery could result in poisoning or other serious health risks.
WARNING Safety risk. Always dispose of batteries in accordance with applicable legal regulations.
Introduction CareTemp™ Touch Free Thermometer
4
CAUTION Always use new batteries of the type and specification indicated in this manual. Mixing old and new batteries will shorten the battery life.
CAUTION Do not use rechargeable batteries, as these may be of inferior quality and duration. The use of rechargeable batteries could compromise the performance of this device.
CAUTION Leaking batteries can damage the device. Remove the batteries whenever the device is not expected to be used for an extended period of time (e.g. multiple months).
CAUTION If a battery has leaked, put on protective gloves and clean the battery compartment with a dry cloth.
CAUTION Federal US law restricts sale of the device identified in this manual to, or on the order of, a licensed healthcare practitioner
CAUTION Proper measurement distance between 4 and 6 cm from the patient's forehead is essential to the accuracy of the temperature measurement determination.
CAUTION Unauthorized modifications to the device are not permitted. Do not modify the product in any way without the manufacturer's prior authorization.
CAUTION In order to avoid incorrect measurements, make sure that the lens on the thermometer's infrared sensor lens is clean and intact prior to use.
CAUTION Avoid touching the infrared sensor lens directly with your fingers.
CAUTION Do not expose the thermometer to extreme temperatures or humidity levels. Make sure that you follow the instructions provided in this manual. Do not expose to direct sunlight.
CAUTION The thermometer is NOT waterproof.
CAUTION Avoid dropping the device.
CAUTION Do not autoclave. Follow only the cleaning procedures
described in this manual.
Note Both the patient and the thermometer should be kept within the same
environmental conditions for at least 30 minutes prior to each measurement.
Note Avoid taking temperature measurements for at least 30 minutes after
physical activity, bathing, swimming, consuming food or beverages, or spending time outdoors.
Note Wait at least 1 minute between temperature measurements.
Measurements performed too frequently could result in incorrect temperature readings.
Note Contact Welch Allyn or your local Welch Allyn representative for any
assistance with configuration, use, and maintenance operations.
Directions for use Introduction 5
Note Avoid pointing the infrared sensor at any heat source other than a patient.
This might affect patient measurements.
Note Avoid exposing the device to external heat sources. This might affect
patient measurements.
Note If possible, measurements should always be taken by pointing the infrared
sensor at the same area of the forehead. Temperatures measured at different locations on the temples or on opposite sides of the head can vary considerably.
Note Holding the thermometer in your hands for too long or exposing the device
to external heat sources could result in distorted temperature readings. For this reason, the body temperature reading could result as being higher or lower than the actual value.
Introduction CareTemp™ Touch Free Thermometer
6

Using the thermometer

Button functions

7
1. Display 4. Memory
2. LED bar 5. C/F (Celsius/Fahrenheit)
3. Measure 6. Pulse timer
8
Using the thermometer CareTemp™ Touch Free Thermometer

Take a measurement

1. Press Measure to power on the thermometer.
When the device is ready, a single dash appears on the screen, and a beeps sounds.
2. Position the thermometer between 4 and 6 cm from the center of the patient's forehead, and aim between the eyebrows.
Note If the eyebrow area is covered with hair, sweat, or dirt, clean
the area and wait 10 minutes to before taking a measurement.
Note Hold the thermometer and the forehead steady during
measurement. Movement can impact the measurement.
3. Press Measure. Slowly move the device toward or farther from the forehead.
The arrow on the display flashes until the correct distance is achieved.
Note After pressing Measure, the thermometer allows 20
seconds to capture the measurement. If you press Measure again before the 20 second time limit is up, the process restarts. If the measurement is not completed in 20 seconds, the device displays Err . Press Measure to restart the process.
4. When the correct distance is reached the LED bar will illuminate, three beeps sound, and the measurement is displayed.
To repeat the process, return to step 2.
Note The complete measurement or Err ↑ message displays for 3
seconds. After 3 seconds, the thermometer returns to ready.
5. The thermometer automatically turns off after 10 seconds of inactivity.
Directions for use Using the thermometer 9

Recall a measurement

Pressing the Memory button (MEM) will display the temperature in memory with a MEM indicator. Memory is available directly from the off state, when no measurement is displayed, or immediately following a new measurement if a comparison to the prior reading is needed. Once the thermometer is powered off or a new measurement is initiated, the last temperature taken becomes the reading in memory.
1. Press and release MEM.
The LCD displays the temperature stored in the thermometer's memory for 10 seconds.
2. Press and hold MEM for 3 seconds to force the measurement on the display to memory and clear the display.
The thermometer beeps three times to indicate the measurement is moved to memory and the thermometer is ready to take a new measurement with a clear display.
Note It is not required to force a reading to memory. A new
reading will automatically move to memory when the thermometer turns off or a new measurement is initiated.

Use the timer

1. Press and release the Timer button.
The timer is activated. Audible beeps sound at 0, 15, 45, and 60 seconds.
2. Press the Timer button again to exit at any time.
After the full 60 second cycle, the thermometer turns off automatically after 2 seconds.

Change the temperature scale

1. Press C/F to temporarily view the measurement in the alternate temperature scale.
This scale conversion feature is available any time a temperature value is displayed, and can be used to continuously alternate between temperature scales.
2. Press and hold both the MEM and C/F buttons for 3 seconds to change the default scale.
Three beeps sound to confirm the setting is changed.
The alternate scale remains until one of the following occurs, returning the thermometer to the default temperature scale:
The C/F button is pressed again.
A new measurement is taken.
Any other action is taken.
The thermometer turns off.
Using the thermometer CareTemp™ Touch Free Thermometer
10

Troubleshooting and error messages

Error message displayed Possible cause Suggested action
11
The thermometer distance to the patient is out of range
The ambient temperature of the room is outside the operating range of 60.8 °F to 104 °F (16 °C to 40 °C)
The measurement is lower than 93.2 °F (34 °C)
The measurement is higher than 109.4 °F (43 °C)
Flashing battery symbol on the display The battery is low on power Replace the battery with two AAA
Blank display The thermometer does not have power Make sure the battery has been loaded
Take the measurement again, and make sure the thermometer is 4 to 6 cm away from the center of the patient's forehead, just between the eyebrows.
Move to a room with the proper ambient temperature and wait 30 minutes for the thermometer to stabilize.
Take the measurement again. Follow the steps in the "Take a measurement" section.
Take the measurement again. Follow the steps in the "Take a measurement" section.
(LR03) alkaline batteries.
properly. Contact your local Welch Allyn service
center or representative if the thermometer still does not function.
Troubleshooting and error messages CareTemp™ Touch Free Thermometer
12

Maintenance

Replace the batteries

The thermometer comes with two AAA (LR03) batteries. Replace with two new AAA (LR03) batteries when the flashing battery symbol appears on the display
1. Use a Phillips screwdriver to loosen the battery cover screw.
2. Slide open the battery cover and remove the batteries.
3. Replace the batteries. Make sure to align the batteries as indicated inside the battery compartment.
4. Replace the battery cover and tighten the battery cover screw.
13
Remove the batteries before storing the thermometer for an extended period of time.

Clean and disinfect the thermometer

The thermometer can be cleaned and an intermediate-level of disinfection can be achieved using the following method.
CAUTION Never submerge the thermometer in water or any other liquid
CAUTION Never use abrasive cleaning agents, thinners or benzene for
cleaning and never immerse the instrument in water or other cleaning liquids.
CAUTION Never insert a sharp object into the scanner area or any other open surface on the thermometer.
CAUTION Do not use unapproved cleaning or disinfection agents. Use of these agents may cause damage to components.
CAUTION Do not use chemicals other than isopropyl or ethyl alcohol on the lens.

Cleaning the sensor window

Slightly moisten a cotton swab or cloth with isopropyl or ethyl alcohol and gently wipe the surface of the lens using a side-to-side (not circular) motion. Avoid touching the lens except when cleaning is required.
14
Maintenance CareTemp™ Touch Free Thermometer

Cleaning the thermometer

Use a pre-moistened wipe or soft cloth slightly moistened with an approved cleaner to gently clean the thermometer.

Disinfecting the thermometer

Use a pre-moistened wipe or soft cloth slightly moistened with an approved disinfectant to disinfect the thermometer. Follow the manufacturers recommendations.
Note After cleaning or disinfecting, wait at least 10 minutes before taking
another measurement.

Approved cleaning agents

Use only the following approved cleaning agents on the thermometer:
CaviWipes Super HDQ® L10
Sani-Cloth® Plus Cleancide
Super Sani-Cloth® Clorox HealthCare Bleach Germicidal Cleaner
Sani-Cloth® Bleach Virex II (256)
Oxiver Tb Tuffie5 Cleaning Wipes
Clinell® Universal Wipes 70% IPA
Accel INTERVention 10% bleach (.5% - 1% Sodium Hypochlorite)
Accel TB Desko
Additional cleaning agents may be periodically evaluated for compatibility. If your cleaning agent is not listed, contact Welch Allyn to determine if additional cleaning agents are approved for use.
Welch Allyn has validated these instructions as being capable of preparing the thermometer for re-use. You must ensure that cleaning and disinfection as actually performed by your personnel, with your equipment and materials, achieves the desired result. This may require validation and routine monitoring of your actual process.

Calibration testing

The thermometer is calibrated at the time of manufacture. If the thermometer is operating according to these instructions, periodic readjustment is not required.

Disposal

These recommendations do not supersede any legal requirements. You must always comply with legal requirements for the control of the measurement, functionality, and accuracy of the device. These controls are required by the scope of relevant laws, directives or ordinances where the device is used.
Thermometer
Directions for use Maintenance 15
The thermometer contains no hazardous materials. Discard without environmental risk. Remove the batteries before disposal.
Battery
Dispose of empty batteries in accordance with national or local regulations.
Maintenance CareTemp™ Touch Free Thermometer
16

Specifications

Type Touch-free infrared thermometer
Model no. NT18, REF 901094, #105801
Displayed temperature range 93.2 to 109.4 °F (34 to 43 °C)
Display resolution 0.1 °F (0.1 °C)
17
Laboratory accuracy ±0.4 °F (±0.2 °C) for the range 93.2 to 109.4 °F (34 to 43 °C)
Clinical accuracy ±0.5 °F (±0.3 °C) for the range 93.2 to 109.4 °F (34 to 43 °C)
Display Liquid crystal display, four digits plus special icons
Operating temperature 60.8 to 104 °F (16 to 40 °C )
Operating humidity Up to 85% non condensing
Storage temperature –13 to 131 °F (–25 to 55 °C)
Storage humidity Up to 85% non condensing
Automatic power off Approximately 10 seconds after last measurement has been
taken
Battery 2 × 1.5V AAA (LR03)
Battery life 3,300 measurements
Weight 100.5g (with battery), 78.0g (without battery)
Dimensions 159.3 x 41.3 x 55.8 mm
Software version NT18 V001

Standards

ASTM 1965-98 EN 12470-5:2003
MDD (93/42/EEC) Annex II ISO STD 80601-2-56
Specifications CareTemp™ Touch Free Thermometer
18

Guidance and manufacturer's declaration

EMC compliance

Special precautions concerning electromagnetic compatibility (EMC) must be taken for all medical electrical equipment. This device complies with IEC EN 60601-1-2:2014.
All medical electrical equipment must be installed and put into service in accordance with the EMC information provided in this document, the Welch Allyn® OAE Hearing Screener Directions for use.
Portable and mobile RF communications equipment can affect the behavior of medical electrical equipment.
19
The device complies with all applicable and required standards for electromagnetic interference.
It does not normally affect nearby equipment and devices.
It is not normally affected by nearby equipment and devices.
It is not safe to operate the central station in the presence of high-frequency surgical equipment.
However, it is good practice to avoid using the device in extremely close proximity to other equipment.

Emissions and immunity information

Electromagnetic emissions
The Welch Allyn CareTemp™ Touch Free Thermometer is intended for use in the electromagnetic environment specified below. The customer or user of the Welch Allyn CareTemp™ Touch Free Thermometer should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1 The Welch Allyn CareTemp™ Touch Free Thermometer uses RF energy
only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The Welch Allyn CareTemp™ Touch Free Thermometer is suitable for use in all establishments including domestic and those directly
Guidance and manufacturer's declaration CareTemp™ Touch Free Thermometer
20
Electromagnetic emissions
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Not applicable
Not applicable
connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Electromagnetic immunity
The Welch Allyn CareTemp™ Touch Free Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Welch Allyn CareTemp™ Touch Free Thermometer should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
±8 kV contact ±15 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±8 kV contact ±15 kV air
Not applicable Mains power quality should be that of a typical
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
commercial or hospital environment.
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
±1 kV differential mode
±2 kV common mode
<5 % UT (>95 % dip in UT) for 0.5 cycle
40 % UT (60 % dip in UT) for 5 cycles
70 % UT (30 % dip in UT) for 25 cycles
<5 % UT (>95 % dip in UT) for 5 s
30 A/m
50Hz/60Hz
Not applicable Mains power quality should be that of a typical
commercial or hospital environment.
Not applicable Mains power quality should be that of a typical
commercial or hospital environment.
30 A/m
50Hz/60Hz
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Electromagnetic immunity
The Welch Allyn CareTemp™ Touch Free Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Welch Allyn CareTemp™ Touch Free Thermometer should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Directions for use Guidance and manufacturer's declaration 21
Electromagnetic immunity
Portable and mobile RF communications equipment should be used no closer to any part of the Connex Central Station, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distances:
d= 1.2
d= 2
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
150KHz to 80MHz
Not applicable
3Vrms
6Vrms (in ISM and amateur radio bands)
80% Am at 1kHz
10V/m 10V/m
d= 1.2 80 MHz to 800 MHz
d= 2.3 800 MHz to 2.7 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site
surveya, should be less than the compliance level in each frequency rangeb. Interference
may occur in the vicinity of equipment marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Welch Allyn CareTemp™ Touch Free Thermometer is used exceeds the applicable RF compliance level above, the Welch Allyn CareTemp™ Touch Free Thermometer should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Connex Central Station.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
22 Guidance and manufacturer's declaration CareTemp™ Touch Free Thermometer
Recommended separation distances between portable and mobile RF communications equipment and the Welch Allyn CareTemp™ Touch Free Thermometer
The Welch Allyn CareTemp™ Touch Free Thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the Welch Allyn CareTemp™ Touch Free Thermometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Welch Allyn CareTemp™ Touch Free Thermometer as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter (m)
Rated max. output power of transmitter (W)
150 kHz to 80 MHz
d= 1.2
150 kHz to 80 MHz
d= 2
80 MHz to 800 MHz
d= 1.2
800 MHz to 2.7 GHz
d= 2.3
0.01 0.12 0.2 0.12 0.23
0.1 0.38 0.632 0.38 0.73
1 1.20 2 1.20 2.30
10 3.8 6.32 3.8 7.3
100 12 20 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Test specifications for enclosure port immunity to RF wireless communications equipment
Test frequency (MHz)
Band MHz
a
Service
a
Modulation
b
Maximum power (W)
Distance (m) Immunity
test level (V/
m)
385 380 - 390 TETRA 400
450 430 - 470 GMRS 460,
FRS 460
710 704 - 787 LTE band 13, 17
745
780
810 800 - 960 GSM 800/900,
TETRA 800,
870
iDEN 820, CDMA 850,
Pulse modulation 18 Hz
FM c±5 kHz deviation
1 kHz sine
Pulse modulation 217 Hz
Pulse modulation 18 Hz
b
1.8 0.3 27
2 0.3 28
b
0.2 0.3 9
b
2 0.3 28
Directions for use Guidance and manufacturer's declaration 23
Test specifications for enclosure port immunity to RF wireless communications equipment
930
1720 1700 - 1990 GSM 1800;
1845
1970
2450 2400 - 2570 Bluetooth,
LTE Band 5
CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4, 25; UMTS
WLAN, 802.11 b/g/n, RFID
Pulse modulation 217 Hz
Pulse modulation 217 Hz
b
2 0.3 28
b
2 0.3 28
2450, LTE Band 7
5240 5100 - 5800 WLAN 802.11
a/n
Pulse modulation 217 Hz
b
0.2 0.3 9
5500
5785
a
For some services, only the uplink frequencies are included.
b
The carrier shall be modulated using a 50 percent duty cycle square wave signal.
c
As an alternative to FM modulation, 50 percent pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
The MANUFACTURER should consider reducing the minimum separation distance, based on RISK MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following equation:
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the IMMUNITY TEST LEVEL in V/m.
Guidance and manufacturer's declaration CareTemp™ Touch Free Thermometer
24

Warranty

Welch Allyn warrants the product to be free of defects in material and workmanship and to perform in accordance with manufacturer's specifications for the period of two years from the date of purchase from Welch Allyn or its authorized distributors or agents.
The warranty period shall start on the date of purchase. The date of purchase is: 1) the invoiced ship date if the device was purchased directly from Welch Allyn, 2) the date specified during product registration, 3) the date of purchase of the product from a Welch Allyn authorized distributor as documented from a receipt from said distributor.
25
This warranty does not cover damage caused by: 1) handling during shipping, 2) use or maintenance contrary to labeled instructions, 3) alteration or repair by anyone not authorized by Welch Allyn, and 4) accidents.
Shipping cost to return a device to a Welch Allyn Service center is not included.
A service notification number must be obtained from Welch Allyn prior to returning any products or accessories to Welch Allyn's designated service centers for repair. To obtain a service notification number, contact Welch Allyn Technical Support.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. WELCH ALLYN'S OBLIGATION UNDER THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF PRODUCTS CONTAINING A DEFECT. WELCH ALLYN IS NOT RESPONSIBLE FOR ANY INDIRECT OR CONSEQUENTIAL DAMAGES RESULTING FROM A PRODUCT DEFECT COVERED BY THE WARRANTY.
Loading...