Welch Allyn Calibration Verification Braun Thermoscan Quick Reference Guide

Page 1 of 5Calibration/verification
Calibration/verification of the Braun ThermoScan device
A Welch Allyn authorized service center must perform individual module calibration and verification activities.
NOTE: The tests are product specific.
Welch Allyn service test equipment and training is available. Contact Welch Allyn Customer Service for information.
Display Definition/Description Required action
Display indicates normal operation after reading is taken.
Display indicates normal operation after device is turned off.
Display indicates reading stored in memory.
Setting used to change the default temperature scale.
Calibration mode.
None.
None.
Turn device on. None.
See manufacturer's instruction guide.
See manufacturer's instruction guide.
See manufacturer's instruction guide.
Setting displays hours remaining before anti-theft function of the device is enabled and device is locked.
Anti-theft function of the device is enabled; device is locked.
None.
See manufacturer's instruction guide.
Return device to station for unlocking.
See manufacturer's instruction
9/28/2010file://C:\Documents and Settings\grahamw1\Local Settings\Temp\~hhD427.htm
guide.
Page 2 of 5Calibration/verification
Battery is low, but thermometer will still operate correctly.
Battery is too low to take correct temperature measurements.
Anti-theft function of the device is enabled; device is locked. Batteries are low.
No probe cover is attached. Attach new, clean probe cover.
No probe cover is attached and batteries are low.
Insert new batteries.
Insert new batteries.
Return device to station for unlocking. Insert new batteries.
See manufacturer's instruction guide.
Attach new, clean probe cover.
Insert new batteries.
Position error.
Ambient temperature is not within the allowed operating range of 50 - 104 °F (10 - 40 °C) or is changing too rapidly.
Temperature taken is not within typical human temperature range.
HI will be displayed when temperature is higher than 108 °F (42.2 °C).
LO will be displayed when temperature is lower than 68 °F (20 °
Clear the display by pushing the Start button once.
Ensure that the position of the probe is correct and remains stable.
Allow the thermometer to remain in a room where the temperature is 50 - 104° F (10 ­40° C) for 30 minutes.
Ensure that the probe tips and lens are clean and a new, clean probe cover is attached. Ensure that the thermometer is properly inserted into the ear canal.