Welch Allyn Acuity User Manual [en, de, es, fr, it]

Acuity

Update Aktualisierung Actualización Mise à jour Aggiornamento

System
®
Central Monitoring

Update to Directions for Use

Acuity Central Monitoring System Acuity LT and Mobile Acuity LT Wireless System Acuity LT and Mobile Acuity LT Hardwired System
Deutsch Aktualisierung für Softwareversion 6.31.01
Español Actualización de la versión de software 6.31.01
Français Mise à jour pour la version logicielle 6.31.01
Italiano Aggiornamento per la versione software 6.31.01
ii Copyright Information Welch Allyn Acuity Central Monitoring System
Copyright 2007, Welch Allyn Inc. All rights are reserved. No one is permitted to reproduce or duplicate, in any form, this manual or any part thereof without permission from Welch Allyn.
Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.
®
Welch Allyn
, Propaq®, Acuity®, Micropaq®, Flexible Monitoring®, and FlexNet® are registered trademarks
of Welch Allyn.
Software in this product is copyright by Welch Allyn Inc. or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title and ownership of the software remains with Welch Allyn or its vendors.
For information about any Welch Allyn product, call Welch Allyn Technical Support:
USA 800 289 2501
503 530 7500
Canada 800 561 8797 China 86 21 6327 9631
European Call Center 353 46 9067790 France 33 1 6009 3366
Germany 49 7477 9271 86 Japan 81 3 3219 0071
Latin America 305 669 9003 Netherlands 31 157 505 000
Singapore 65 6419 8100 South Africa 27 11 777 7555
United Kingdom 44 1332 363812
Manual Part Number 810-2020-01 Rev. B 04/2007
Printed in USA
Australia 61 2 9638 3000
800 074 793
Contents
1 - English Update for Acuity System. . . . . . . . . . . . . . . . . . . . . . . . . . . 1
iii
Heart Rate Calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Set Acuity LT Localization Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Acuity LT Hardwired System - Sort Windows Button . . . . . . . . . . . . . . . . . . . . . . 4
Changes in Acuity Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2 - Deutsches Update für das Acuity-System . . . . . . . . . . . . . . . . . . . . 13
Herzfrequenzberechnung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Einstellen von länderspezifischen Acuity LT-Parametern . . . . . . . . . . . . . . . . . . 15
Acuity LT-Kabelsystem - Sort Windows Button. . . . . . . . . . . . . . . . . . . . . . . . . . 16
Änderungen bei Acuity-Meldungen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Technische Daten . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Garantie . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3 - Actualización en español del sistema Acuity . . . . . . . . . . . . . . . . . 25
Cálculo de frecuencia cardíaca . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Definición de parámetros de localización de Acuity LT. . . . . . . . . . . . . . . . . . . . 27
Sistema por cableado fijo Acuity LT - Botón Ordenar Ventanas . . . . . . . . . . . . . 28
Cambios en los mensajes del Acuity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Especificaciones . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Garantía. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4 - Mise à jour en Français pour le système Acuity. . . . . . . . . . . . . . . 37
Calcul de la fréquence cardiaque . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Définition des paramètres de localisation Acuity LT. . . . . . . . . . . . . . . . . . . . . . 39
Système câblé Acuity LT - Sort Windows Button . . . . . . . . . . . . . . . . . . . . . . . . 40
Modifications apportées aux messages de l'Acuity . . . . . . . . . . . . . . . . . . . . . . 41
Spécifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Garantie . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5 - Aggiornamento del sistema Acuity . . . . . . . . . . . . . . . . . . . . . . . . . 49
Calcolo della frequenza cardiaca . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Impostazione dei parametri di localizzazione di Acuity LT. . . . . . . . . . . . . . . . . . 51
Sistema via cavo Acuity LT - Pulsante per l'ordinamento delle finestre . . . . . . . 52
Modifiche nei messaggi Acuity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Specifiche . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
iv Contents Welch Allyn Acuity System
Garanzia . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
1
1

English Update for Acuity System

This document describes changes in operation for the Acuity Central Monitoring System (software version 6.31.01). This document replaces information in the following manuals:
810-1598-02 B Acuity System Directions For Use (software version 6.3X)
810-1600-02 B Acuity LT and Mobile Acuity LT Wireless System Directions For Use (software version 6.3X)
810-1608-02 B Acuity LT and Mobile Acuity LT Hardwired System Directions For Use (software version 6.3X)
For other operating information not contained in this document, refer to the preceding manuals.

Heart Rate Calculation

Understanding Differences Between Acuity System and Monitor
It is possible for the Acuity Central Station with Arrhythmia Analysis Option to display a different heart rate than the bedside patient monitor.
The Acuity System with the Arrhythmia Analysis Option calculates the patient heart rate by using the Mortara arrhythmia library analysis algorithms that are integrated into the Acuity System software. This approach analyzes beat morphology using multiple leads (when available). The result is a more sophisticated assessment of heart rate that is less affected by noise and other artifact. Noise and other artifact can affect the single-lead heart rate calculation performed by the bedside monitor. The Acuity System heart rate is also in greater harmony with the ECG waveform, arrhythmia detection, and arrhythmia event alarms.
The bedside monitor displays the heart rate that it calculates using the single-lead heart rate analysis. It does not use arrhythmia analysis. It is possible that this heart rate may be different from the heart rate at the Acuity Central Station. The differences are most likely to occur with rare ECG morphologies, such as “tall ‘T’ waves,” certain paced rhythms, and noise or artifact.
2 Chapter 1 English Update for Acuity System Welch Allyn Acuity System
Understanding Differences in Heart Rate Alarms
The possible differences between ECG-based heart rates at the Acuity Central Station and the bedside ECG monitor can affect the frequency of heart rate alarms. For example, when heart rate alarm limit settings are the same at the Acuity Central Station and the bedside monitor, the bedside monitor could display a higher heart rate (due, perhaps, to noise or artifact) and present a heart rate alarm event at bedside.
For patients monitored using a Propaq CS or Encore monitor, if a high-quality pulse oximetry waveform is available for the patient, you can try selecting pulse rate oximetry as the heart rate source. This may result in a more accurate heart rate, thereby reducing erroneous or nuisance heart rate alarm events at the bedside monitor.
To select pulse rate oximetry as the heart rate source:
From Main Menu, press SETUP, MORE to access Setup Menu 2.
Press NEXT as needed to highlight the HR/PR selection for SELECTED SOURCE (typically set to ECG).
Press CHANGE as needed to highlight SpO2.
Press Home or MAIN MENU key to exit.
Acuity Message for Monitor Heart Rate Alarm
If a heart rate alarm occurs at the bedside monitor but does not occur at the Acuity Central Station for a defined period of time (see table below), the Acuity Central Station displays an equipment alert to inform the operator of this condition:
When the heart rate alarm is no longer occurring at the bedside monitor, the Acuity system dismisses this message. The operator can also dismiss the message.
Acuity System (Software Version 6.31) Time Delay Between Monitor HR Alarm Occurrence and
Display of Acuity Message
Acuity Central Monitoring System Configurable by system administrator: 5 to 120 seconds, or off (no
Acuity LT/Mobile Acuity LT Wireless System 30 seconds
Acuity LT/Mobile Acuity LT Hardwired System 30 seconds
message)

Set Acuity LT Localization Parameters

Note
Note
This feature applies to the following systems:
Acuity LT and Mobile Acuity LT Wireless System
Acuity LT and Mobile Acuity LT Hardwired System
You can change the following Acuity LT system parameters according to the needs of your facility.
Time zone
Acuity language
Micropaq language
Printout paper size
EN1060 compliance
Temperature units
Enable or disable the optional parallel printer
Set Facility Name (appears on printouts)
Set Unit Name (appears in upper right corner of Acuity LT system display)
Chapter 1 English Update for Acuity System 3
These parameters have been added for Acuity LT system software version 6.31: Set Facility Name, and Set Unit Name.
To change any of these parameters:
1. Ac uit y LT:
Hold down the “diamond” key (next to the spacebar), move the cursor to the upper right corner of the screen, and click the right mouse button.
Mobile Acuity LT (Tadpole Computer laptop):
Hold down the “ * ” key (between “Fn” and “Alt” keys), move the cursor to the upper right corner of the screen, and click the right mouse button.
2. Click “Quit Application”, then click Confirm.
3. Wait for the login screen to appear (within 1 to 2 minutes), then type “iupdate” and press Enter.
4. Follow the prompts to make the desired changes, then press q to quit.
5. If the change requires a system reboot (Acuity will inform you), choose 2 to quit and reboot the system. The system will reboot and startup Acuity LT application.
If the change does not require a system reboot (Acuity will inform you), choose 1 to quit, then wait for the login prompt, type “Acuity” and press Enter to return to the Acuity LT application.
4 Chapter 1 English Update for Acuity System Welch Allyn Acuity System
Note

Acuity LT Hardwired System - Sort Windows Button

This feature is now available for the 28-monitor Acuity LT Hardwired System. It is not required for the 12-monitor Acuity LT Hardwired System.
Figure 1. Sort Windows Button
This sorts all Waveform Windows by room number in alpha-numeric order. Waveform Windows for monitors without confirmed patient IDs are located at the end.

Changes in Acuity Messages

In the following displayed buttons or messages, “Acuity” has been removed or replaced by “Central”.
Table 1. Summary of Changes in Acuity Messages and Buttons
New Button Previous Button
Presets Acuity Presets
New Message Previous Message
PRESETS: USE CENTRAL SETTINGS PRESETS: USE ACUITY SETTINGS
Central Print Requests Acuity Print Requests
Device Connecting to Central Device Connecting to Acuity
Device Connected to Central Device Connected to Acuity
Device Disconnecting from Central Device Disconnecting from Acuity
Chapter 1 English Update for Acuity System 5
Monitor Disconnected from Central at time XX Monitor Disconnected from Acuity at time XX
Room is not valid for this Central System Room is not valid for this Acuity System
Incompatible with Central Incompatible with Acuity
MONITOR CO2 UNITS CHANGED TO MATCH CENTRAL UNITS.
CAN’T CHANGE CO2 UNITS WHEN PROPAQ USED WITH CENTRAL.
MONITOR CO2 UNITS CHANGED TO MATCH ACUITY UNITS.
CAN’T CHANGE CO2 UNITS WHEN PROPAQ USED WITH ACUITY.
6 Chapter 1 English Update for Acuity System Welch Allyn Acuity System

Specifications

Heart Rate and Arrhythmia Analysis Option
Method for Calculating Heart Rate
Micropaq monitor
Propaq CS/ Propaq Encore monitor
Acuity System with Arrhythmia Option
ST Analysis ST Analysis can be performed for any or all of seven leads, depending on the operator selection.
Heart Rate Heart rate information is available in the trend data which can be viewed on the display or printed.
Determined by monitor (displayed at monitor)
Heart rate = 60 / latest average interval in seconds. For higher heart rates, latest average interval = 7/8 of previous average interval + 1/8 of latest interval For lower heart rates, latest average interval = 3/4 (previous average interval) + 1/4 latest interval. Transition rates for choice of formula include hysteresis and are 70 and 80 beats per minute.
Determined by Acuity Arrhythmia Option software (displayed at Acuity Central Station)
The beat-to-beat heart rate (HR) value is calculated as follows:
HR = 60000/actual RR (bpm). Actual RR = time between last detected QRS complex and previously detected QRS complex (ms). Average HR is calculated on the basis of the mean RR interval in the last 6 seconds or 8 RR intervals (whichever is shorter).
Arrhythmia Analysis Option
The operator can select a measurement offset.
ST segment shifts are recorded in continuous trend data every second. The operator can inspect trend data to see the duration and elevation or depression for each episode for any time period recorded. The operator can also inspect a summary of ST segment shift data within tabular trends.
The operator can inspect the trend data to see the lowest, highest, and median (averaged) heart rates. Trend data also includes the total beats per range of time.
Definition of Pause Arrhythmia Event
A pause is defined as the R-R interval which is greater than or equal to two times the average R-R.
EMC Compliance
Acuity System:
Acuity Central Monitoring System (Hardwired and Wireless)
Acuity LT Wireless System
Acuity LT Hardwired System
The Acuity System complies with all applicable and required standards for electromagnetic interference.
It does not normally affect nearby equipment and devices.
It is not normally affected by nearby equipment and devices.
It is safe to operate the Acuity System in the presence of high-frequency surgical
However, it is good practice to avoid using the Acuity System in extremely close
Chapter 1 English Update for Acuity System 7
equipment.
proximity to other equipment.
Guidance and manufacturer’s declaration—electromagnetic emissions
The Acuity System is intended for use in the electromagnetic environment specified below. The customer or the user of the Acuity System should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment—guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions IEC 61000-3-3
Group 1 The Acuity System uses RF energy only for its internal
Class A The Acuity System is suitable for use in all establishments,
Class A
Complies
function.Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
other than domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
8 Chapter 1 English Update for Acuity System Welch Allyn Acuity System
Guidance and manufacturer’s declaration—electromagnetic immunity
The Acuity System is intended for use in the electromagnetic environment specified below. The customer or the user of the Acuity System should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment—
guidance
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient/burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions, and voltage variations on power-supply input lines IEC 61000-4-11
Power frequency (50/ 60 Hz) magnetic field IEC 61000-4-8
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
±2 kV for power supply lines ±1 kV for input/output lines
±1 kV differential mode ±2 kV common mode
±2 kV for power supply lines ±1 kV for input/output lines
±1 kV differential mode
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
±2 kV common mode
<5% U
t
(>95% dip in Ut) for 0.5 cycle 40% U
t
(60% dip in Ut) for 5 cycles 70% U
t
(30% dip in Ut) for 25 cycles <5% U
t
(>95% dip in Ut) for 5 sec
<5% U
t
(>95% dip in Ut) for 0.5 cycle 40% U
t
(60% dip in Ut) for 5 cycles 70% U
t
(30% dip in Ut) for 25 cycles <5% U
t
(>95% dip in Ut) for 5 sec
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Acuity System requires continued operation during power mains interruption, it is recommended that the monitor be powered from an uninterruptible power supply or a battery.
3 A/m 3 A/m Power frequency magnetic fields should be
at levels characteristic of a typical location in a typical commercial or hospital environment.
Note U
is the AC mains voltage prior to application of the test level.
t
Chapter 1 English Update for Acuity System 9
P
P
P
Guidance and manufacturer’s declaration—electromagnetic immunity
The Acuity System is intended for use in the electromagnetic environment specified below. The customer or the user of the Acuity System should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance
level
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m 80 MHz to 2.5 GHz
3 V
rms
3 V/m d = 1.2 80 MHz to 800 MHz
Electromagnetic environment—guidance
Portable and mobile RF communications equipment should be used no closer to any part of the Acuity System, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2
d = 2.3 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recommended separation distance in meters. Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey be less than the compliance level in each frequency
b
range
.
a
, should
Interference may occur in the vicinity of equipment marked with the following symbol:
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
and reflection from structures, objects and people.
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Acuity System is used exceeds the applicable RF compliance level above, the Acuity System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Acuity System.
10 Chapter 1 English Update for Acuity System Welch Allyn Acuity System
PPP
Recommended separation distances between portable and mobile RF communications equipment and the
Acuity System
The Acuity System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Acuity System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Acuity System as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
m
power of transmitter
Separation distance according to frequency of transmitter
W
150 kHz to 80 MHz
d = 1.2
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5 GHz
d = 2.3
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.

Warranty

Note
Chapter 1 English Update for Acuity System 11
The warranty for Acuity LT Monitoring System software and hardware components is 12 months as described below.
LIMITED WARRANTY: ACUITY LT MONITORING SYSTEM AND ACCESSORIES
This product is sold by Welch Allyn under the warranties set forth in the following paragraphs. These warranties are extended only to the end user with respect to the purchase of this product directly from Welch Allyn or Welch Allyn’s authorized distributors as new merchandise.
For a period of 12 months from the date of original delivery to the buyer, the Acuity LT Monitoring System software and hardware components are warranted to be free from functional defects in materials and workmanship and to conform in all material respects to the description of the product contained in the Directions For Use and accompanying labels and/or inserts. For a period of 90 days this same warranty is made for accessories provided by Welch Allyn. Warranty of accessories purchased separately from listed suppliers will be the responsibility of the listed suppliers.
This warranty is valid only if (a) all network equipment and cabling are approved for use with the Acuity LT system by Welch Allyn and are installed according to instructions provided by Welch Allyn or its authorized distributors; (b) the product is properly operated under conditions of normal use in accordance with applicable safety and regulatory requirements; (c) replacements and repairs are made in accordance with the instructions provided by Welch Allyn; (d) only the Acuity LT system or other software authorized by Welch Allyn is used on the workstation; (e) the product has not been configured, modified, adjusted or repaired other than by Welch Allyn or by persons expressly authorized by Welch Allyn, or in accordance with written instructions provided by Welch Allyn; (f) the product has not been subject to misuse, negligence or accident.
Welch Allyn’s sole and exclusive obligation, and buyer’s sole and exclusive remedy under the above warranties, is limited to repairing or replacing, free of charge, a product which is reported to Welch Allyn at one of the telephone numbers listed below not later than seven days after the expiration of the warranty:
Welch Allyn 8500 S.W. Creekside Place Beaverton, Oregon 97008-7107 USA Telephone: (503) 530-7500 or (800) 289-2500 Facsimile: (503) 526-4200
WELCH ALLYN SHALL NOT BE OTHERWISE LIABLE FOR ANY DAMAGES INCLUDING, BUT NOT LIMITED TO, INCIDENTAL, CONSEQUENTIAL, OR SPECIAL DAMAGES.
THERE ARE NO EXPRESS OR IMPLIED WARRANTIES WHICH EXTEND BEYOND THE WARRANTIES IN THIS DOCUMENT. WELCH ALLYN MAKES NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
12 Chapter 1 English Update for Acuity System Welch Allyn Acuity System
Deutsches Update für das Acuity-
13
2
System
Dieses Dokument beschreibt Änderungen hinsichtlich des Betriebs des zentralen Acuity­Überwachungssystems (Software-Version 6.31.01). Die Angaben in diesem Dokument ersetzen die entsprechenden Informationen in den folgenden Handbüchern:
810-1413-02 B Acuity-System Gebrauchsanweisung (Software-Version 6.3X)
810-1563-01 B Acuity LT- und Mobile Acuity LT-Funksystem Gebrauchsanweisung (Software-Version 6.3X)
810-1621-01 B Acuity LT- und Mobile Acuity LT-Kabelsystem Gebrauchsanweisung (Software-Version 6.3X)
Alle sonstigen Informationen zum Betrieb, die nicht im vorliegenden Dokument enthalten sind, finden Sie in den genannten Handbüchern.

Herzfrequenzberechnung

Unterschiede zwischen Acuity-System und Monitor
An einer Acuity-Zentralstation mit Arrhythmieanalyseoption kann eine von der am bettseitigen Monitor des Patienten abweichende Herzfrequenz angezeigt werden.
Das mit der Arrhythmieanalyseoption ausgestattete Acuity-System berechnet die Herzfrequenz des Patienten mit Hilfe von Analyse-Algorithmen, die in die Software des Acuity-Systems integriert sind und die die Mortara-Arrhythmie-Bibliothek verwenden. Bei diesem Verfahren wird die Herzschlagmorphologie unter Einsatz mehrerer Ableitungen (sofern verfügbar) analysiert. Hieraus resultiert eine differenziertere Bewertung der Herzfrequenz, die weniger stark von Störungen und anderen Artefakten verfälscht ist. Störungen und andere Artefakte können die am bettseitigen Monitor mittels nur einer Ableitung durchgeführte Herzfrequenzberechnung nachteilig beeinflussen. Die vom Acuity-System bestimmte Herzfrequenz befindet sich außerdem in höherer Übereinstimmung mit der EKG-Kurve, der Arrhythmie-Erkennung und den Arrhythmie-Ereignisalarmen.
Am bettseitigen Monitor wird die Herzfrequenz angezeigt, die der Monitor anhand der Herzfrequenzanalyse mit nur einer Ableitung berechnet. Hierbei wird keine Arrhythmieanalyse eingesetzt. Es ist möglich, dass diese Herzfrequenz von der an der Acuity-Zentralstation angezeigten Frequenz abweicht. Ein Auftreten solcher Abweichungen ist bei seltenen EKG-Morphologien wie „hohen 'T'-Wellen“, bestimmten stimulierten Rhythmen oder Störungen bzw. Artefakten am wahrscheinlichsten.
14 Kapitel 2 Deutsches Update für das Acuity-System Welch Allyn Acuity-System
Unterschiede zwischen Herzfrequenz-Alarmen
Die möglichen Unterschiede zwischen EKG-basierten Herzfrequenzen an der Acuity­Zentralstation und Herzfrequenzen am bettseitigen EKG-Monitor können sich auf die Häufigkeit der Auslösung von Herzfrequenz-Alarmen auswirken. Wenn beispielsweise die Einstellungen der Herzfrequenz-Alarmgrenzen an der Acuity-Zentralstation und dem bettseitigen Monitor identisch sind, könnte am bettseitigen Monitor eine höhere Herzfrequenz angezeigt werden (möglicherweise durch Störungen oder Artefakte verursacht) und somit bettseitig ein Herzfrequenz-Alarmereignis darstellen.
Bei Patienten, die mit einem Propaq CS- oder Encore-Monitor überwacht werden, können Sie versuchen, die Pulsfrequenzoximetrie als Herzfrequenzquelle einzustellen, wenn für diesen Patienten eine Pulsoximetriekurve von hoher Qualität zur Verfügung steht. Dieses Vorgehen kann zu einer exakter erkannten Herzfrequenz führen, wodurch sich die Häufigkeit von Herzfrequenz-Fehlalarmen am bettseitigen Monitor verringern würde.
So stellen Sie Pulsfrequenzoximetrie als Herzfrequenzquelle ein:
Im Hauptmenü drücken Sie SETUP, WEITER, um ins Setup-Menü 2 zu gelangen:
Drücken Sie so oft WAHL wie nötig, um den Eintrag „HF/PF“ für GEWÄHLTE QUELLE (normalerweise auf EKG eingestellt) zu markieren.
Drücken Sie so oft ÄNDERN wie nötig, um „SpO2“ zu markieren.
Drücken Sie die Taste „Pos 1“ oder HAUPTMENÜ, um das Menü zu verlassen.
Acuity-Meldung für Monitor-Herzfrequenz-Alarm
Tritt über einen festgelegten Zeitraum (siehe folgende Tabelle) am bettseitigen Monitor ein Herzfrequenz-Alarm auf, an der Acuity-Zentralstation aber nicht, dann zeigt die Acuity­Zentralstation einen Gerätealarm an, um den Bediener über diesen Zustand zu informieren:
Wenn der Herzfrequenz-Alarm am bettseitigen Monitor nicht mehr auftritt, hebt das Acuity-System diese Meldung auf. Die Meldung kann ebenfalls vom Bediener aufgehoben werden.
Acuity-System (Software-Version 6.31) Zeitverzögerung zwischen dem Auftreten eines
Monitor-HF-Alarms und der Anzeige der Acuity-Meldung
Acuity-Zentralüberwachungssystem Vom Systemadministrator konfigurierbar: 5 bis 120 Sekunden, bzw.
ganz aus (keine Meldung)
Acuity LT-/Mobile Acuity LT-Funksystem 30 Sekunden
Acuity LT-/Mobile Acuity LT-Kabelsystem 30 Sekunden
Kapitel 2 Deutsches Update für das Acuity-System 15
Hinweis
Hinweis

Einstellen von länderspezifischen Acuity LT-Parametern

Diese Funktion ist nur für folgende Systeme verfügbar:
Acuity LT- und Mobile Acuity LT-Funksystem
Acuity LT- und Mobile Acuity LT-Kabelsystem
Die folgenden Parameter des Acuity LT-Systems können je nach Bedarf Ihrer Einrichtung festgelegt werden.
Zeitzone
•Acuity-Sprache
Micropaq-Sprache
Druckerpapierformat
EN1060-Konformität
Temperatureinheiten
Aktivieren oder Deaktivieren des optionalen Paralleldruckers
Einstellen des Namens der Einrichtung (Anzeige auf Ausdrucken)
Einstellen des Namens der Abteilung (Anzeige in der rechten oberen Ecke der Acuity LT-Systemanzeige)
Die folgenden Parameter wurden für das Acuity LT-System, Software­Version 6.31, hinzugefügt: „Einstellen des Namens der Einrichtung“ und „Einstellen des Namens der Abteilung“.
So ändern Sie diese Parameter:
1. Ac u i t y LT:
Halten Sie die „Diamanten“-Taste (neben der Leertaste) gedrückt, bewegen Sie den Cursor in die obere rechte Bildschirmecke und klicken Sie mit der rechten Maustaste.
Mobile Acuity LT (Tadpole Laptop-Computer):
Halten Sie die „*“-Taste (zwischen den Tasten „Fn“ und „Alt“) gedrückt, bewegen Sie den Cursor in die obere rechte Bildschirmecke und klicken Sie mit der rechten Maustaste.
2. Klicken Sie zuerst auf Anwendung verlassen und dann auf Bestätigen.
3. Warten Sie, bis der Anmeldebildschirm angezeigt wird (nach 1 bis 2 Minuten), geben Sie dann iupdate ein und drücken Sie die Eingabetaste.
4. Folgen Sie den Aufforderungen, um die gewünschten Änderungen vorzunehmen, und drücken Sie dann q, um den Einstellungsmodus zu beenden.
5. Wenn die erfolgten Änderungen einen Systemneustart erfordern (Acuity meldet, ob ein Neustart erforderlich ist), wählen Sie die Option 2 zum Beenden und zum Neustarten des Systems. Das System wird neu gestartet und startet im Anschluss die Acuity LT-Anwendung.
Wenn die erfolgte Änderung keinen Systemneustart erfordert (Acuity meldet, ob ein Neustart erforderlich ist), wählen Sie die Option 1 zum Beenden. Warten Sie dann auf die Anmeldeaufforderung, geben Sie Acuity ein und drücken Sie die Eingabetaste, um zur Acuity LT-Anwendung zurückzukehren.
16 Kapitel 2 Deutsches Update für das Acuity-System Welch Allyn Acuity-System
Hinweis

Acuity LT-Kabelsystem - Sort Windows Button

Diese Funktion steht jetzt für das 28-Monitor-Acuity LT-Kabelsystem zur Verfügung. Sie wird für das 12-Monitor-Acuity LT-Kabelsystem nicht benötigt.
Figure 2. Bildschaltfläche Fenster sortieren
Hiermit werden alle Kurvenfenster in alphanumerischer Reihenfolge nach Zimmernummer sortiert. Kurvenfenster für Monitore ohne bestätigte Patienten-IDs befinden sich am Ende.
+ 44 hidden pages