Update
Aktualisierung
Actualización
Mise à jour
Aggiornamento
System
®
Central Monitoring
Update to Directions for Use
Acuity Central Monitoring System
Acuity LT and Mobile Acuity LT Wireless System
Acuity LT and Mobile Acuity LT Hardwired System
EnglishUpdate for Software Version 6.31.01
DeutschAktualisierung für Softwareversion 6.31.01
EspañolActualización de la versión de software 6.31.01
FrançaisMise à jour pour la version logicielle 6.31.01
ItalianoAggiornamento per la versione software 6.31.01
iiCopyright InformationWelch Allyn Acuity Central Monitoring System
Copyright 2007, Welch Allyn Inc. All rights are reserved. No one is permitted to reproduce or duplicate, in
any form, this manual or any part thereof without permission from Welch Allyn.
Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the
product, that may result from failure to use this product in accordance with the instructions, cautions,
warnings, or statement of intended use published in this manual.
®
Welch Allyn
, Propaq®, Acuity®, Micropaq®, Flexible Monitoring®, and FlexNet® are registered trademarks
of Welch Allyn.
Software in this product is copyright by Welch Allyn Inc. or its vendors. All rights are reserved. The
software is protected by United States of America copyright laws and international treaty provisions
applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software
incorporated with this instrument as intended in the operation of the product in which it is embedded. The
software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to
human-perceivable form. This is not a sale of the software or any copy of the software; all right, title and
ownership of the software remains with Welch Allyn or its vendors.
For information about any Welch Allyn product, call Welch Allyn Technical Support:
USA 800 289 2501
503 530 7500
Canada 800 561 8797China 86 21 6327 9631
European Call Center 353 46 9067790France 33 1 6009 3366
Germany 49 7477 9271 86Japan 81 3 3219 0071
Latin America 305 669 9003Netherlands 31 157 505 000
This document describes changes in operation for the Acuity Central Monitoring System
(software version 6.31.01). This document replaces information in the following manuals:
•810-1598-02 B Acuity System Directions For Use (software version 6.3X)
•810-1600-02 B Acuity LT and Mobile Acuity LT Wireless System Directions For Use
(software version 6.3X)
•810-1608-02 B Acuity LT and Mobile Acuity LT Hardwired System Directions For Use
(software version 6.3X)
For other operating information not contained in this document, refer to the preceding
manuals.
Heart Rate Calculation
Understanding Differences Between Acuity System and Monitor
It is possible for the Acuity Central Station with Arrhythmia Analysis Option to display a
different heart rate than the bedside patient monitor.
•The Acuity System with the Arrhythmia Analysis Option calculates the patient heart
rate by using the Mortara arrhythmia library analysis algorithms that are integrated
into the Acuity System software. This approach analyzes beat morphology using
multiple leads (when available). The result is a more sophisticated assessment of
heart rate that is less affected by noise and other artifact. Noise and other artifact can
affect the single-lead heart rate calculation performed by the bedside monitor. The
Acuity System heart rate is also in greater harmony with the ECG waveform,
arrhythmia detection, and arrhythmia event alarms.
•The bedside monitor displays the heart rate that it calculates using the single-lead
heart rate analysis. It does not use arrhythmia analysis. It is possible that this heart
rate may be different from the heart rate at the Acuity Central Station. The differences
are most likely to occur with rare ECG morphologies, such as “tall ‘T’ waves,” certain
paced rhythms, and noise or artifact.
2Chapter 1 English Update for Acuity SystemWelch Allyn Acuity System
Understanding Differences in Heart Rate Alarms
The possible differences between ECG-based heart rates at the Acuity Central Station and
the bedside ECG monitor can affect the frequency of heart rate alarms. For example,
when heart rate alarm limit settings are the same at the Acuity Central Station and the
bedside monitor, the bedside monitor could display a higher heart rate (due, perhaps, to
noise or artifact) and present a heart rate alarm event at bedside.
For patients monitored using a Propaq CS or Encore monitor, if a high-quality pulse
oximetry waveform is available for the patient, you can try selecting pulse rate oximetry
as the heart rate source. This may result in a more accurate heart rate, thereby reducing
erroneous or nuisance heart rate alarm events at the bedside monitor.
To select pulse rate oximetry as the heart rate source:
•From Main Menu, press SETUP, MORE to access Setup Menu 2.
•Press NEXT as needed to highlight the HR/PR selection for SELECTED SOURCE
(typically set to ECG).
•Press CHANGE as needed to highlight SpO2.
•Press Home or MAIN MENU key to exit.
Acuity Message for Monitor Heart Rate Alarm
If a heart rate alarm occurs at the bedside monitor but does not occur at the Acuity
Central Station for a defined period of time (see table below), the Acuity Central Station
displays an equipment alert to inform the operator of this condition:
When the heart rate alarm is no longer occurring at the bedside monitor, the Acuity
system dismisses this message. The operator can also dismiss the message.
Acuity System (Software Version 6.31)Time Delay Between Monitor HR Alarm Occurrence and
Display of Acuity Message
Acuity Central Monitoring SystemConfigurable by system administrator: 5 to 120 seconds, or off (no
You can change the following Acuity LT system parameters according to the needs of your
facility.
•Time zone
•Acuity language
•Micropaq language
•Printout paper size
•EN1060 compliance
•Temperature units
•Enable or disable the optional parallel printer
•Set Facility Name (appears on printouts)
•Set Unit Name (appears in upper right corner of Acuity LT system display)
Chapter 1 English Update for Acuity System3
These parameters have been added for Acuity LT system software version 6.31:
Set Facility Name, and Set Unit Name.
To change any of these parameters:
1.Ac uit y LT:
Hold down the “diamond” key (next to the spacebar), move the cursor to the
upper right corner of the screen, and click the right mouse button.
Mobile Acuity LT (Tadpole Computer laptop):
Hold down the “ * ” key (between “Fn” and “Alt” keys), move the cursor to the
upper right corner of the screen, and click the right mouse button.
2. Click “Quit Application”, then click Confirm.
3. Wait for the login screen to appear (within 1 to 2 minutes), then type “iupdate” and
press Enter.
4. Follow the prompts to make the desired changes, then press q to quit.
5. If the change requires a system reboot (Acuity will inform you), choose 2 to quit and
reboot the system. The system will reboot and startup Acuity LT application.
If the change does not require a system reboot (Acuity will inform you), choose 1 to
quit, then wait for the login prompt, type “Acuity” and press Enter to return to the
Acuity LT application.
4Chapter 1 English Update for Acuity SystemWelch Allyn Acuity System
Note
Acuity LT Hardwired System - Sort Windows Button
This feature is now available for the 28-monitor Acuity LT Hardwired System.
It is not required for the 12-monitor Acuity LT Hardwired System.
Figure 1. Sort Windows Button
This sorts all Waveform Windows by room number in alpha-numeric order. Waveform
Windows for monitors without confirmed patient IDs are located at the end.
Changes in Acuity Messages
In the following displayed buttons or messages, “Acuity” has been removed or replaced
by “Central”.
Table 1. Summary of Changes in Acuity Messages and Buttons
New ButtonPrevious Button
PresetsAcuity Presets
New MessagePrevious Message
PRESETS: USE CENTRAL SETTINGSPRESETS: USE ACUITY SETTINGS
Central Print RequestsAcuity Print Requests
Device Connecting to CentralDevice Connecting to Acuity
Device Connected to CentralDevice Connected to Acuity
Device Disconnecting from CentralDevice Disconnecting from Acuity
Chapter 1 English Update for Acuity System5
Monitor Disconnected from Central at time XXMonitor Disconnected from Acuity at time XX
Room is not valid for this Central SystemRoom is not valid for this Acuity System
Incompatible with CentralIncompatible with Acuity
MONITOR CO2 UNITS CHANGED TO MATCH CENTRAL
UNITS.
CAN’T CHANGE CO2 UNITS WHEN PROPAQ USED
WITH CENTRAL.
MONITOR CO2 UNITS CHANGED TO MATCH ACUITY UNITS.
CAN’T CHANGE CO2 UNITS WHEN PROPAQ USED WITH
ACUITY.
6Chapter 1 English Update for Acuity SystemWelch Allyn Acuity System
Specifications
Heart Rate and Arrhythmia Analysis Option
Method for Calculating Heart Rate
Micropaq monitor
Propaq CS/
Propaq Encore
monitor
Acuity System with
Arrhythmia Option
ST AnalysisST Analysis can be performed for any or all of seven leads, depending on the operator selection.
Heart RateHeart rate information is available in the trend data which can be viewed on the display or printed.
Determined by monitor (displayed at monitor)
Heart rate = 60 / latest average interval in seconds.
For higher heart rates, latest average interval = 7/8 of previous average interval + 1/8 of latest
interval
For lower heart rates, latest average interval = 3/4 (previous average interval) + 1/4 latest interval.
Transition rates for choice of formula include hysteresis and are 70 and 80 beats per minute.
Determined by Acuity Arrhythmia Option software (displayed at Acuity Central Station)
The beat-to-beat heart rate (HR) value is calculated as follows:
HR = 60000/actual RR (bpm).
Actual RR = time between last detected QRS complex and previously detected QRS complex
(ms).
Average HR is calculated on the basis of the mean RR interval in the last 6 seconds or 8 RR
intervals (whichever is shorter).
Arrhythmia Analysis Option
The operator can select a measurement offset.
ST segment shifts are recorded in continuous trend data every second. The operator can inspect
trend data to see the duration and elevation or depression for each episode for any time period
recorded. The operator can also inspect a summary of ST segment shift data within tabular trends.
The operator can inspect the trend data to see the lowest, highest, and median (averaged) heart
rates. Trend data also includes the total beats per range of time.
Definition of Pause
Arrhythmia Event
A pause is defined as the R-R interval which is greater than or equal to two times the average R-R.
EMC Compliance
Acuity System:
•Acuity Central Monitoring System (Hardwired and Wireless)
•Acuity LT Wireless System
•Acuity LT Hardwired System
The Acuity System complies with all applicable and required standards for
electromagnetic interference.
•It does not normally affect nearby equipment and devices.
•It is not normally affected by nearby equipment and devices.
•It is safe to operate the Acuity System in the presence of high-frequency surgical
•However, it is good practice to avoid using the Acuity System in extremely close
Chapter 1 English Update for Acuity System7
equipment.
proximity to other equipment.
Guidance and manufacturer’s declaration—electromagnetic emissions
The Acuity System is intended for use in the electromagnetic environment specified below. The customer or the user of the
Acuity System should assure that it is used in such an environment.
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Group 1The Acuity System uses RF energy only for its internal
Class AThe Acuity System is suitable for use in all establishments,
Class A
Complies
function.Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
other than domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
8Chapter 1 English Update for Acuity SystemWelch Allyn Acuity System
Guidance and manufacturer’s declaration—electromagnetic immunity
The Acuity System is intended for use in the electromagnetic environment specified below. The customer or the user of the
Acuity System should assure that it is used in such an environment.
Immunity testIEC 60601 test levelCompliance levelElectromagnetic environment—
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations on
power-supply input
lines
IEC 61000-4-11
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV differential
mode
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
±2 kV common mode
<5% U
t
(>95% dip in Ut)
for 0.5 cycle
40% U
t
(60% dip in Ut)
for 5 cycles
70% U
t
(30% dip in Ut)
for 25 cycles
<5% U
t
(>95% dip in Ut)
for 5 sec
<5% U
t
(>95% dip in Ut)
for 0.5 cycle
40% U
t
(60% dip in Ut)
for 5 cycles
70% U
t
(30% dip in Ut)
for 25 cycles
<5% U
t
(>95% dip in Ut)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Acuity System requires
continued operation during power mains
interruption, it is recommended that the
monitor be powered from an uninterruptible
power supply or a battery.
3 A/m3 A/mPower frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
NoteU
is the AC mains voltage prior to application of the test level.
t
Chapter 1 English Update for Acuity System9
P
P
P
Guidance and manufacturer’s declaration—electromagnetic immunity
The Acuity System is intended for use in the electromagnetic environment specified below. The customer or the user of the
Acuity System should assure that it is used in such an environment.
Immunity testIEC 60601 test levelCompliance
level
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
rms
3 V/md = 1.2 80 MHz to 800 MHz
Electromagnetic environment—guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Acuity
System, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2
d = 2.3 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts according to the transmitter
manufacturer and d is the recommended separation
distance in meters.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
be less than the compliance level in each frequency
b
range
.
a
, should
Interference may occur in the vicinity of equipment
marked with the following symbol:
Note 1At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
and reflection from structures, objects and people.
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in
which the Acuity System is used exceeds the applicable RF compliance level above, the Acuity System
should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the Acuity System.
10Chapter 1 English Update for Acuity SystemWelch Allyn Acuity System
PPP
Recommended separation distances between portable and mobile RF communications equipment and the
Acuity System
The Acuity System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the Acuity System can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the Acuity System as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
m
power of transmitter
Separation distance according to frequency of transmitter
W
150 kHz to 80 MHz
d = 1.2
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5 GHz
d = 2.3
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (w) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Warranty
Note
Chapter 1 English Update for Acuity System11
The warranty for Acuity LT Monitoring System software and hardware
components is 12 months as described below.
LIMITED WARRANTY: ACUITY LT MONITORING SYSTEM AND ACCESSORIES
This product is sold by Welch Allyn under the warranties set forth in the following
paragraphs. These warranties are extended only to the end user with respect to the
purchase of this product directly from Welch Allyn or Welch Allyn’s authorized distributors
as new merchandise.
For a period of 12 months from the date of original delivery to the buyer, the Acuity LT
Monitoring System software and hardware components are warranted to be free from
functional defects in materials and workmanship and to conform in all material respects to
the description of the product contained in the Directions For Use and accompanying
labels and/or inserts. For a period of 90 days this same warranty is made for accessories
provided by Welch Allyn. Warranty of accessories purchased separately from listed
suppliers will be the responsibility of the listed suppliers.
This warranty is valid only if (a) all network equipment and cabling are approved for use
with the Acuity LT system by Welch Allyn and are installed according to instructions
provided by Welch Allyn or its authorized distributors; (b) the product is properly operated
under conditions of normal use in accordance with applicable safety and regulatory
requirements; (c) replacements and repairs are made in accordance with the instructions
provided by Welch Allyn; (d) only the Acuity LT system or other software authorized by
Welch Allyn is used on the workstation; (e) the product has not been configured,
modified, adjusted or repaired other than by Welch Allyn or by persons expressly
authorized by Welch Allyn, or in accordance with written instructions provided by Welch
Allyn; (f) the product has not been subject to misuse, negligence or accident.
Welch Allyn’s sole and exclusive obligation, and buyer’s sole and exclusive remedy under
the above warranties, is limited to repairing or replacing, free of charge, a product which is
reported to Welch Allyn at one of the telephone numbers listed below not later than seven
days after the expiration of the warranty:
Welch Allyn
8500 S.W. Creekside Place
Beaverton, Oregon 97008-7107 USA
Telephone: (503) 530-7500 or (800) 289-2500
Facsimile: (503) 526-4200
WELCH ALLYN SHALL NOT BE OTHERWISE LIABLE FOR ANY DAMAGES INCLUDING,
BUT NOT LIMITED TO, INCIDENTAL, CONSEQUENTIAL, OR SPECIAL DAMAGES.
THERE ARE NO EXPRESS OR IMPLIED WARRANTIES WHICH EXTEND BEYOND THE
WARRANTIES IN THIS DOCUMENT. WELCH ALLYN MAKES NO WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
12Chapter 1 English Update for Acuity SystemWelch Allyn Acuity System
Deutsches Update für das Acuity-
13
2
System
Dieses Dokument beschreibt Änderungen hinsichtlich des Betriebs des zentralen AcuityÜberwachungssystems (Software-Version 6.31.01). Die Angaben in diesem Dokument
ersetzen die entsprechenden Informationen in den folgenden Handbüchern:
•810-1413-02 B Acuity-System Gebrauchsanweisung (Software-Version 6.3X)
•810-1563-01 B Acuity LT- und Mobile Acuity LT-Funksystem Gebrauchsanweisung
(Software-Version 6.3X)
•810-1621-01 B Acuity LT- und Mobile Acuity LT-Kabelsystem Gebrauchsanweisung
(Software-Version 6.3X)
Alle sonstigen Informationen zum Betrieb, die nicht im vorliegenden Dokument enthalten
sind, finden Sie in den genannten Handbüchern.
Herzfrequenzberechnung
Unterschiede zwischen Acuity-System und Monitor
An einer Acuity-Zentralstation mit Arrhythmieanalyseoption kann eine von der am
bettseitigen Monitor des Patienten abweichende Herzfrequenz angezeigt werden.
•Das mit der Arrhythmieanalyseoption ausgestattete Acuity-System berechnet die
Herzfrequenz des Patienten mit Hilfe von Analyse-Algorithmen, die in die Software
des Acuity-Systems integriert sind und die die Mortara-Arrhythmie-Bibliothek
verwenden. Bei diesem Verfahren wird die Herzschlagmorphologie unter Einsatz
mehrerer Ableitungen (sofern verfügbar) analysiert. Hieraus resultiert eine
differenziertere Bewertung der Herzfrequenz, die weniger stark von Störungen und
anderen Artefakten verfälscht ist. Störungen und andere Artefakte können die am
bettseitigen Monitor mittels nur einer Ableitung durchgeführte
Herzfrequenzberechnung nachteilig beeinflussen. Die vom Acuity-System
bestimmte Herzfrequenz befindet sich außerdem in höherer Übereinstimmung mit
der EKG-Kurve, der Arrhythmie-Erkennung und den Arrhythmie-Ereignisalarmen.
•Am bettseitigen Monitor wird die Herzfrequenz angezeigt, die der Monitor anhand der
Herzfrequenzanalyse mit nur einer Ableitung berechnet. Hierbei wird keine
Arrhythmieanalyse eingesetzt. Es ist möglich, dass diese Herzfrequenz von der an der
Acuity-Zentralstation angezeigten Frequenz abweicht. Ein Auftreten solcher
Abweichungen ist bei seltenen EKG-Morphologien wie „hohen 'T'-Wellen“,
bestimmten stimulierten Rhythmen oder Störungen bzw. Artefakten am
wahrscheinlichsten.
14Kapitel 2 Deutsches Update für das Acuity-SystemWelch Allyn Acuity-System
Unterschiede zwischen Herzfrequenz-Alarmen
Die möglichen Unterschiede zwischen EKG-basierten Herzfrequenzen an der AcuityZentralstation und Herzfrequenzen am bettseitigen EKG-Monitor können sich auf die
Häufigkeit der Auslösung von Herzfrequenz-Alarmen auswirken. Wenn beispielsweise die
Einstellungen der Herzfrequenz-Alarmgrenzen an der Acuity-Zentralstation und dem
bettseitigen Monitor identisch sind, könnte am bettseitigen Monitor eine höhere
Herzfrequenz angezeigt werden (möglicherweise durch Störungen oder Artefakte
verursacht) und somit bettseitig ein Herzfrequenz-Alarmereignis darstellen.
Bei Patienten, die mit einem Propaq CS- oder Encore-Monitor überwacht werden, können
Sie versuchen, die Pulsfrequenzoximetrie als Herzfrequenzquelle einzustellen, wenn für
diesen Patienten eine Pulsoximetriekurve von hoher Qualität zur Verfügung steht. Dieses
Vorgehen kann zu einer exakter erkannten Herzfrequenz führen, wodurch sich die
Häufigkeit von Herzfrequenz-Fehlalarmen am bettseitigen Monitor verringern würde.
So stellen Sie Pulsfrequenzoximetrie als Herzfrequenzquelle ein:
•Im Hauptmenü drücken Sie SETUP, WEITER, um ins Setup-Menü 2 zu gelangen:
•Drücken Sie so oft WAHL wie nötig, um den Eintrag „HF/PF“ für GEWÄHLTE
QUELLE (normalerweise auf EKG eingestellt) zu markieren.
•Drücken Sie so oft ÄNDERN wie nötig, um „SpO2“ zu markieren.
•Drücken Sie die Taste „Pos 1“ oder HAUPTMENÜ, um das Menü zu verlassen.
Acuity-Meldung für Monitor-Herzfrequenz-Alarm
Tritt über einen festgelegten Zeitraum (siehe folgende Tabelle) am bettseitigen Monitor
ein Herzfrequenz-Alarm auf, an der Acuity-Zentralstation aber nicht, dann zeigt die AcuityZentralstation einen Gerätealarm an, um den Bediener über diesen Zustand zu
informieren:
Wenn der Herzfrequenz-Alarm am bettseitigen Monitor nicht mehr auftritt, hebt das
Acuity-System diese Meldung auf. Die Meldung kann ebenfalls vom Bediener aufgehoben
werden.
Acuity-System (Software-Version 6.31)Zeitverzögerung zwischen dem Auftreten eines
Monitor-HF-Alarms und der Anzeige der Acuity-Meldung
Acuity-ZentralüberwachungssystemVom Systemadministrator konfigurierbar: 5 bis 120 Sekunden, bzw.
Kapitel 2 Deutsches Update für das Acuity-System15
Hinweis
Hinweis
Einstellen von länderspezifischen Acuity LT-Parametern
Diese Funktion ist nur für folgende Systeme verfügbar:
•Acuity LT- und Mobile Acuity LT-Funksystem
•Acuity LT- und Mobile Acuity LT-Kabelsystem
Die folgenden Parameter des Acuity LT-Systems können je nach Bedarf Ihrer Einrichtung
festgelegt werden.
•Zeitzone
•Acuity-Sprache
•Micropaq-Sprache
•Druckerpapierformat
•EN1060-Konformität
•Temperatureinheiten
•Aktivieren oder Deaktivieren des optionalen Paralleldruckers
•Einstellen des Namens der Einrichtung (Anzeige auf Ausdrucken)
•Einstellen des Namens der Abteilung (Anzeige in der rechten oberen Ecke der
Acuity LT-Systemanzeige)
Die folgenden Parameter wurden für das Acuity LT-System, SoftwareVersion 6.31, hinzugefügt: „Einstellen des Namens der Einrichtung“ und
„Einstellen des Namens der Abteilung“.
So ändern Sie diese Parameter:
1.Ac u i t y LT:
Halten Sie die „Diamanten“-Taste (neben der Leertaste) gedrückt, bewegen Sie
den Cursor in die obere rechte Bildschirmecke und klicken Sie mit der rechten
Maustaste.
Mobile Acuity LT (Tadpole Laptop-Computer):
Halten Sie die „*“-Taste (zwischen den Tasten „Fn“ und „Alt“) gedrückt,
bewegen Sie den Cursor in die obere rechte Bildschirmecke und klicken Sie mit
der rechten Maustaste.
2. Klicken Sie zuerst auf Anwendung verlassen und dann auf Bestätigen.
3. Warten Sie, bis der Anmeldebildschirm angezeigt wird (nach 1 bis 2 Minuten), geben
Sie dann iupdate ein und drücken Sie die Eingabetaste.
4. Folgen Sie den Aufforderungen, um die gewünschten Änderungen vorzunehmen, und
drücken Sie dann q, um den Einstellungsmodus zu beenden.
5. Wenn die erfolgten Änderungen einen Systemneustart erfordern (Acuity meldet, ob
ein Neustart erforderlich ist), wählen Sie die Option 2 zum Beenden und zum
Neustarten des Systems. Das System wird neu gestartet und startet im Anschluss
die Acuity LT-Anwendung.
Wenn die erfolgte Änderung keinen Systemneustart erfordert (Acuity meldet, ob ein
Neustart erforderlich ist), wählen Sie die Option 1 zum Beenden. Warten Sie dann auf
die Anmeldeaufforderung, geben Sie Acuity ein und drücken Sie die Eingabetaste, um
zur Acuity LT-Anwendung zurückzukehren.
16Kapitel 2 Deutsches Update für das Acuity-SystemWelch Allyn Acuity-System
Hinweis
Acuity LT-Kabelsystem - Sort Windows Button
Diese Funktion steht jetzt für das 28-Monitor-Acuity LT-Kabelsystem zur
Verfügung. Sie wird für das 12-Monitor-Acuity LT-Kabelsystem nicht
benötigt.
Figure 2. Bildschaltfläche Fenster sortieren
Hiermit werden alle Kurvenfenster in alphanumerischer Reihenfolge nach Zimmernummer
sortiert. Kurvenfenster für Monitore ohne bestätigte Patienten-IDs befinden sich am
Ende.
Loading...
+ 44 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.