Welch Allyn 9600 Plus User Manual

9600 Plus Calibration Tester

© 2017 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn. No other use, reproduction, or distribution of the publication, or any part of it, is permitted without written permission from Welch Allyn.
Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.
Welch Allyn is a registered trademark of Welch Allyn. Braun ThermoScan is a registered trademarks of Braun Corporation.
Printed in Mexico
DIR 80020333 Ver A Revised: 2017-06
Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Symbols and Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Unpacking the 9600 Plus Calibration Tester . . . . . . . . . . . . . 4
Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . 5
LCD Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Power On Sequence. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Front of the 9600 Plus Calibration Tester . . . . . . . . . . . . . . . 7
Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
9600 Plus Calibration Tester Setup. . . . . . . . . . . . . . . . . . . 11
Changing the Calibration Set Point Temperature . . . . . . . . 11
Welch Allyn SureTemp and SureTemp Plus Thermometers 13
Braun ThermoScan Pro 4000 Model. . . . . . . . . . . . . . . . . . 15
Braun ThermoScan Pro 6000 model. . . . . . . . . . . . . . . . . . 18
Braun ThermoScan Pro 3000 Model. . . . . . . . . . . . . . . . . . 23
Troubleshooting inaccurate readings . . . . . . . . . . . . . . . . . 25
i
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Problem Solving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Electromagnetic Compatibility . . . . . . . . . . . . . . . . . . . . . . 32
ii
Warranty and Service . . . . . . . . . . . . . . . . . . . . . . . . . 33
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Symboles et descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Mises en garde . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Déballage du testeur d’étalonnage 9600 Plus . . . . . . . . . . . 4
Commandes et indicateurs . . . . . . . . . . . . . . . . . . . . . . 5
Affichage ACL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Séquence de mise en marche . . . . . . . . . . . . . . . . . . . . . . . 7
Avant du testeur d’étalonnage 9600 Plus . . . . . . . . . . . . . . . 7
Fonctionnement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Installation du testeur d’étalonnage 9600 Plus. . . . . . . . . . 11
Modification de la température du point d'étalonnage . . . . 11
Thermomètres Welch Allyn SureTemp et SureTemp Plus . 13
Modèle Braun ThermoScan Pro 4000. . . . . . . . . . . . . . . . . 15
Modèle Braun ThermoScan Pro 6000. . . . . . . . . . . . . . . . . 19
Modèle Braun ThermoScan Pro 3000. . . . . . . . . . . . . . . . . 24
Résolution des problèmes de relevés incorrects . . . . . . . . 26
Entretien . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Nettoyage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Dépannage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Spécifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Compatibilité électromagnétique . . . . . . . . . . . . . . . . . . . . 32
Directions for Use Contents iii
Garantie et entretien . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Garantie limitée. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Entretien . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Einführung. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Symbole und Beschreibungen . . . . . . . . . . . . . . . . . . . . . . . 2
Vorsicht . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Auspacken des Kalibriertesters 9600 Plus . . . . . . . . . . . . . . 4
Bedienelemente und Anzeigen . . . . . . . . . . . . . . . . . . . 5
Flüssigkristallanzeige . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Einschaltreihenfolge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Vorderseite des Kalibriertesters 9600 Plus . . . . . . . . . . . . . . 7
Betrieb. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Konfiguration des Kalibriertesters 9600 Plus . . . . . . . . . . . 11
Ändern der Kalibrier-Sollwerttemperatur. . . . . . . . . . . . . . . 11
Die Thermometer SureTemp und SureTemp Plus von
Welch Allyn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Das Modell ThermoScan Pro 4000 von Braun . . . . . . . . . . 15
Modell Braun ThermoScan Pro 6000 . . . . . . . . . . . . . . . . . 18
Das Modell ThermoScan Pro 3000 von Braun . . . . . . . . . . 23
Fehlerbehebung bei ungenauen Werten . . . . . . . . . . . . . . 25
Wartung. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Reinigung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Fehlerbehebung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Technische Daten . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Elektromagnetische Verträglichkeit. . . . . . . . . . . . . . . . . . . 32
Gewährleistung und Service . . . . . . . . . . . . . . . . . . . . 33
Beschränkte Gewährleistung . . . . . . . . . . . . . . . . . . . . . . . 33
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
iv
1
1

Introduction

The 9600 Plus Calibration Tester verifies calibration of the following thermometers:
®
• Welch Allyn SureTemp
986)
• Welch Allyn SureTemp OEM)
• Braun ThermoScan
The 9600 Plus Calibration Tester provides a tightly regulated temperature source that eliminates the setup, precision calibration, and maintenance requirements of water baths and other general purpose calibration equipment. The user takes the “temperature” of the 9600 Plus Calibration Tester with the knowledge and assurance that precise readings are obtained because the unit is factory calibrated to National Institute of Standards and Technology (NIST) or equivalent devices. User adjustments are not required or recommended.
The “blackbody” of the 9600 Plus Calibration System acts as a heat reservoir for the thermistor probe and as a heat radiator for the infrared ear thermometers.
models (678, 678 OEM, 679 and
®
Plus models (690, 692 and 692
®
Pro 3000, Pro 4000, and Pro 6000
The unit contains solid-state circuitry for long-term stability with minimal maintenance. Welch Allyn uses only precision components in the construction of the 9600 Plus Calibration Tester. To maintain traceability to NIST or equivalent, return the 9600 Plus Calibration Tester to the factory for annual calibration.
2 Chapter 1 Introduction Welch Allyn 9600 Plus Calibration Tester

Symbols and Descriptions

Double Insulation.
Direct Current.
Handle with Care.
Attention! Caution; Consult Directions for Use for more information.
Transport/Storage Temperature
95%
15%
Relative Humidity
Separate collection for electronic equipment
Recyclable
Directions for Use Chapter 1 Introduction 3

Cautions

A caution statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately, could lead to equipment failure, equipment damage, or data loss.
Caution Do not immerse or soak the 9600 Plus Calibration Tester in any type of fluid.
Caution Only trained Biomedical personnel should operate the 9600 Plus Calibration Tester and conduct calibration verification of thermometers.
Caution When not using the Thermistor or Ear Device Ports, keep them covered with the Dust Cover.
Caution Store thermometers for testing in the same room as the 9600 Plus Calibration Tester for approximately 30 minutes prior to testing to allow the thermometer to reach ambient room temperature.
Caution Do not clean the Thermistor or Ear Device Ports. Doing so may damage the surface. Only a Welch Allyn authorized Service Center should clean the Device Ports.
4 Chapter 1 Introduction Welch Allyn 9600 Plus Calibration Tester

Unpacking the 9600 Plus Calibration Tester

Unpack the 9600 Plus Calibration Tester and inspect for missing items. Retain the shipping materials to return the device to Welch Allyn for recalibration.
The following items are enclosed:
• A 9600 Plus Calibration Tester
• A package containing a power supply and four mains plug inserts
• The Directions for Use
• A Carrying Case
If any of these items are missing after opening, please contact Welch Allyn Customer Service at the number provided on
page 34.
5
Current Cavity Temperature
Temperature Units
Calibration Set
Point Temperature
Set Point
Mode
Celsius/Fahrenheit
Selection Button
Temperature Selection Button
2

Controls and Indicators

LCD Display

Figure 1. LCD Display and Front Buttons

Calibration Set Point Temperature

There are three Calibration Set Point temperatures used to test the calibration of the thermometer. The three temperatures are 36° C (96.8° F), 38.5° C (101.3° F), and 41°
6 Chapter 2 Controls and Indicators Welch Allyn 9600 Plus Calibration Tester
C (105.8° F). The LCD displays the current Calibration Set Point Temperature in the upper left corner of the LCD.

Current Cavity Temperature

The current temperature of the 9600 Plus Calibration Tester is indicated by the large numbers in the center of the display
(Figure 1).
As the device is transitioning to the current Calibration Set Point Temperature, the actual temperature flashes on the display. When the device reaches the Calibration Set Point Temperature, the temperature in the center of the display stops flashing.

Temperature Units

The 9600 Plus Calibration Tester is capable of displaying the temperature in degrees Fahrenheit or Celsius.

Set Point Mode

The LCD will display either a "D" (Default) or "L" (Last Used) in the upper left hand corner (Figure 1). The 9600 Plus Calibration Tester has two set point modes of operation. The Default mode (from the factory) causes the 9600 Plus Calibration Tester to heat up to the lowest set point (36.0 Used mode causes the 9600 Plus Calibration Tester to heat up to the last temperature setting that was used.
°
C / 96.8°F) upon startup. The Last
To change the Set Point Mode, press and hold the C/F and Arrow buttons below the LCD display simultaneously until a single audio beep is heard. A “D” or an “L” will appear in the upper left corner of the display for the new mode.
Directions for Use Chapter 2 Controls and Indicators 7

Power On Sequence

The 9600 Plus Calibration Tester receives power from a wall­mount power pack that plugs into the back of the unit. There is no ON/OFF switch. On power up, the 9600 Plus Calibration Tester performs internal self-checks, displays an all-segments test pattern, beeps and displays the device’s current temperature. The 9600 Plus Calibration Tester immediately applies heat, if needed, to reach the selected Calibration Set Point Temperature.

Front of the 9600 Plus Calibration Tester

Temperature Selection Button

This button allows the user to select one of three set point temperatures: 36.0
°
(105.8
To scroll from one set point to the next, press and hold the Temperature Selection button until a beep is heard. The newly selected set point appears in the upper left corner of the LCD display. The device’s current temperature is displayed, will start to flash, and continue flashing until the cavity reaches the equilibrium at the new set point.
F).
°
C (96.8°F), 38.5°C (101.3°F), or 41.0°C

Celsius/Fahrenheit Selection Button

This button allows the user to select degrees Celsius or degrees Fahrenheit. To change the temperature units, press and hold the C/F button until a beep is heard. The temperatures will appear in either “°C” or “°F”, see (Figure 1).
8 Chapter 2 Controls and Indicators Welch Allyn 9600 Plus Calibration Tester
Thermistor Device Port
Ear Device Port
Dust Cover
Power Jack
Caution Verify the thermometer being tested and the 9600 Plus are set to use the same temperature scale before running a test. Reference the thermometers directions for use to change its temperature scale
Figure 2. Front of the 9600 Plus Calibration Tester
Directions for Use Chapter 2 Controls and Indicators 9

Dust Cover

This black, plastic cover keeps dust from entering the Thermistor and Ear Device Ports. To expose the Thermistor and Ear Device Ports, rotate the Dust Cover counter clockwise. To cover the ports, rotate the Dust Cover clockwise.
Caution When not using the Thermistor or Ear Device Ports, keep them covered with the Dust Cover.

Thermistor and Ear Device Ports

The Ear Device Port is the larger of the two ports located under the dust cover and is for the Braun ThermoScan Pro 3000 and Pro 4000 ear thermometers (Figure 2).
The Thermistor Device Port is the smaller port located under the dust cover and is for the SureTemp and SureTemp Plus thermometers.
The surface texture inside the Ear Device Port is painted with a special dull, black paint to enable Ear device readings. You can only see its surface under bright light. The Thermistor Device Port is not painted black. Keep both clean and dry.
The coating in the device ports of the unit is fragile, so please follow the cleaning instructions carefully. Do not touch the surface, and do not allow any foreign objects to be inserted into the ports. Avoid spilling liquids onto the unit. If liquid or other contaminants enter the unit, Welch Allyn can not assure proper calibration and you must return the unit to Welch Allyn for recalibration.
10 Chapter 2 Controls and Indicators Welch Allyn 9600 Plus Calibration Tester
11
3

Operation

9600 Plus Calibration Tester Setup

Place the 9600 Plus Calibration Tester on a level surface away from sunlight, drafts, and other sources of heat or cold. The 9600 Plus Calibration Tester takes approximately 20 minutes to heat to the lowest setting. When testing several thermometers at all three temperatures, it is recommended to test all probes at one Calibration Set Point Temperature before proceeding to the next Calibration Set Point Temperature. To further expedite testing start at the lowest Calibration Set Point Temperature. The 9600 Plus Calibration Tester does not have an internal fan, causing the cool down time to be greater than the time it takes to heat up.

Changing the Calibration Set Point Temperature

To scroll from one set point to the next, press and hold the Temperature Selection button until a beep is heard. The newly selected set point appears in the upper left corner of the LCD display. The device’s current temperature is displayed, starts to flash, and continues flashing until the cavity reaches the equilibrium at the new set point.
12 Chapter 3 Operation Welch Allyn 9600 Plus Calibration Tester
UKUS EUAUS
Two prongs in the power adapter
Figure 3. Power Adapter and Mains Plug Inserts
For the setup:
1. Choose the proper mains plug insert and slide it over the two prongs in the power converter (Figure 3).
2. Plug the power adapter into the 9600 Plus Calibration Tester
3. Observe the Set Point Mode in the upper left hand corner of
(Figure 2) and the opposite end into a wall outlet.
the LCD display (Figure 1). If the unit displays a “D”, it is in Default Mode and will heat to the lowest Set Point Temperature. If you do not want to conduct testing at this Set Point Temperature, press and hold the Temperature Selection button to select the desired setting. The temperature display will flash before staying on continuously to indicate the 9600 Plus Calibration Tester has stabilized and is ready for use.
Directions for Use Chapter 3 Operation 13
Note

Welch Allyn SureTemp and SureTemp Plus Thermometers

Caution Store thermometers for testing in the same room as the 9600 Plus Calibration Tester for approximately 30 minutes prior to testing to allow the thermometer to reach ambient room temperature.
Prior to testing visually inspect the probe for any physical damage. Damaged probes must be replaced.
Figure 4. 9600 Plus Calibration Tester with the Welch Allyn SureTemp Plus Thermometer
To begin calibration verification of the thermometers:
1. Remove the probe from the probe well and clean it with either a 70% isopropyl alcohol solution, a 10% chlorine bleach
14 Chapter 3 Operation Welch Allyn 9600 Plus Calibration Tester
solution, or a non-staining disinfectant. Let the probe air dry. Do not apply a probe cover.
2. Place the thermometer in Monitor Mode, refer to the thermometer’s Operator's Manual.
3. Insert the probe into the Thermistor Device Port (Figure 4).
4. Wait for approximately one minute or until temperature on the thermometer is stable for ten seconds. Compare the thermometer's temperature reading to the 9600 Plus' Calibration Set Point Temperature. If the temperatures are within ±0.1° C (±0.2° F), the thermometer is within calibration.
5. Test all available thermometers for calibration verification at the current Calibration Set Point Temperature. Proceed to the next Calibration Set Point Temperature, see "Changing the
Calibration Set Point Temperature" (page 11).
Directions for Use Chapter 3 Operation 15

Braun ThermoScan Pro 4000 Model

Caution Store thermometers for testing in the same room as the 9600 Plus Calibration Tester for approximately 30 minutes prior to testing to allow the thermometer to reach ambient room temperature.
Caution The ambient temperature must be stable and within the range of 18.3°C (65.0°F) – 26.7°C (80.0°F).
Caution Prolonged body contact with the thermometer may impact measurements. When instructed to wait before performing the next step place the thermometer on the work surface.
Figure 5. 9600 Plus Calibration Tester with the Braun ThermoScan Pro 4000
To begin calibration verification of the thermometers:
16 Chapter 3 Operation Welch Allyn 9600 Plus Calibration Tester
1. Check the probe tip lens window for displacement. There shouldn't be any gap in the seam formed between bezel and the lens window. If a gap exists the unit is damaged and must be replaced.
2. Gently wipe the probe tip with a cotton swab slightly moistened with isopropyl alcohol using care not to displace the lens window; remove excess alcohol with a clean cotton swab, and let air dry for 5 minutes. Do not use any chemical other than alcohol to clean the probe window.
3. Place the Braun ThermoScan Pro 4000 in Calibration Mode using the following steps.
a. Push and release the I/O MEM button to turn the product
on. The display will show symbols and functions. The Pro 4000 will perform an automatic self check. After a few seconds you will hear a beep and see three dashed lines, a sound symbol, and ° C or ° F on the display.
b. Push the I/O MEM button again and keep it pressed for the
following steps:
• After approximately. 3 seconds the “OFF” symbol will flash on the display (keep the button depressed).
• When you hear a beep, release the button immediately.
• The Pro 4000 is now in calibration check mode and the display is flashing and showing the “CAL” symbol.
4. After entering calibration mode (CAL), wait 1 minute before taking the first temperature.
Caution Prolonged body contact with the thermometer may impact measurements. When instructed to wait before performing the next step place the thermometer on the work surface.
Directions for Use Chapter 3 Operation 17
5. Apply a new probe cover.
6. Place the probe firmly into the Ear Device Port (Figure 5).
7. Wait approximately 3 seconds, press the thermometer Start button, and watch for the “Exac Temp” light to flash.
8. Leave the thermometer in the 9600 Plus Calibration Tester until you hear a beep.
9. Remove the Pro 4000 Thermometer from the 9600 Plus Calibration Tester and read the temperature in the thermometer’s display. If the temperatures are within ±0.2° C (±0.4° F), the thermometer is within calibration.
10.Wait 1 full minute before taking another reading with the same thermometer. Repeated measurements in short sequence may cause higher readings.
Caution Prolonged body contact with the thermometer may impact measurements. When instructed to wait before performing the next step place the thermometer on the work surface.
11. The probe cover must be replaced with a new probe cover before each reading, to remove any residual heat absorbed from the 9600 Plus Calibration Tester
12.The device will exit CAL mode after four minutes.
13.When using only one 9600 Plus Calibration Tester, test all available thermometers for calibration verification at the current Calibration Set Point Temperature. Proceed to the next Calibration Set Point Temperature, see "Changing the
Calibration Set Point Temperature" (page 11).
18 Chapter 3 Operation Welch Allyn 9600 Plus Calibration Tester

Braun ThermoScan Pro 6000 model

Caution Store thermometers for testing in the same room as the 9600 Plus Calibration Tester for approximately 30 minutes prior to testing to allow the thermometer to reach ambient room temperature.
Caution The ambient temperature must be stable and within the range of 18.3°C (65.0°F) – 26.7°C (80.0°F)
Caution Set the temperature scales on the 9600 Plus to match the Braun Thermoscan PRO 6000’s default temperature scales.
Caution Prolonged body contact with the thermometer may impact measurements. When instructed to wait before performing the next step place the thermometer on the work surface.
Caution On older versions of the 9600 Plus, the dust cover screw can interfere with the proper positioning of the Braun 6000 in the ear device port. Use the Braun 6000 test fixture (Welch Allyn part number 411690) to ensure proper positioning.
Directions for Use Chapter 3 Operation 19
Figure 6. 9600 Plus Calibration Tester with the Braun ThermoScan Pro 6000
A Braun Pro 6000 Calibration Fixture (material number 411690) is available for use with older 9600 Plus calibration testers where the dust cover screw might interfere with the proper alignment of the Braun 6000 in the ear device port.
The older models have a dust cover screw that rises above the O­ring washer. Newer models have a dust cover screw that is recessed into the washer.
When using the calibration fixture, follow the instructions provided with the fixture to install it to the calibration tester.
To begin calibration verification of the thermometers:
1. Check the probe tip lens window for displacement. There should not be any gap in the seam formed between the bezel and the lens window. If there is a gap, the unit is damaged and must be replaced.
20 Chapter 3 Operation Welch Allyn 9600 Plus Calibration Tester
Note
2. Gently wipe the probe tip with a cotton swab slightly moistened with isopropyl or ethyl alcohol. Use care not to displace the lens window. Remove excess alcohol with a clean cotton swab, and let air dry for 5 minutes. Do not use any chemical other than alcohol to clean the probe window.
3. Place the Braun ThermoScan Pro 6000 in Calibration Mode using the following steps.
a. With the thermometer in sleep mode, press and release
C/F.
b. IMMEDIATELY (while all segments of the LCD are
illuminated during self-test) press and hold C/F and Mem.
c. After 5 seconds the thermometer will emit a long beep.
During the beep release Mem and C/F. The thermometer is now in Calibration Check Mode, indicated by CAL displayed on the screen for 0.25 seconds every second
4. After entering calibration mode (CAL), wait 1 minute before taking the first temperature.
Do not hold the thermometer while waiting. This allows the thermometer to remain at ambient room temperature.
5. Apply a new probe cover. If applied properly a short beep sounds and the flashing probe cover icon will disappear from the display.
6. Wait for the screen to flash three dashes (- - -) alternating with CAL.
7. If you are using the Braun 6000 test fixture, place the Braun Pro 6000 in the fixture as shown in the assembly instructions (material number 721533). If the Braun Pro 6000 adapter is
Directions for Use Chapter 3 Operation 21
Note
Note
not used, place the probe firmly into the ear device port
(Figure 6). It is important to ensure that the device is
perpendicular in the port, directly aligned with the calibration tester sensor and not at an angle or tilted. Firmly push the device down into the well to insure perpendicularity and that the probe is fully seated in device port. Visually check to ensure proper alignment. Do not apply too much pressure once the device is firmly seated, as this may cause your hand to waiver and tilt or move the device during testing.
8. Wait 5 seconds, press the thermometer Measure button, and watch for the green ExacTemp light to flash.
9. Leave the thermometer in the 9600 Plus Calibration Tester until the ExacTemp light stops flashing and you hear a beep.
10.Remove the Pro 6000 Thermometer from the 9600 Plus Calibration Tester and read the temperature in the thermometer’s display. If the temperatures are within ±0.2 °C (±0.4 °F), the thermometer is within calibration.
If temperatures are outside ±0.2 °C, verify the test procedure was strictly observed and repeat the calibration check two more times at this temperature. If any two subsequent test are within the acceptable limits the test may be considered a pass. If all three readings are out of calibration, call Welch Allyn for service.
11. Replace the probe cover to reset the thermometer before taking the next measurement. Wait for the thermometer screen to flash three dashes (- - -) alternating with CAL
Do not hold the thermometer while waiting. This allows the thermometer to remain at ambient room temperature.
22 Chapter 3 Operation Welch Allyn 9600 Plus Calibration Tester
Note
Note
The probe cover must be replaced with a new probe cover before each reading, to remove any residual heat absorbed from the 9600 Plus Calibration Tester.
12.Wait one full minute before taking another reading with the same thermometer. Repeated measurements in short sequence may cause higher readings.
Do not hold the thermometer while waiting. This allows the thermometer to remain at ambient room temperature.
13.When using only one 9600 Plus Calibration Tester, test all available thermometers for calibration verification at the current calibration set point temperature before proceeding to the next calibration set point temperature. See "Changing the
Calibration Set Point Temperature" (page 11).
14.The device will exit CAL mode after 10 minutes of inactivity, or can be forced to exit by holding both C/F and Mem for 5 seconds. After exiting Calibration Check Mode the thermometer will return to Sleep Mode.
Directions for Use Chapter 3 Operation 23

Braun ThermoScan Pro 3000 Model

Caution Store thermometers for testing in the same room as the 9600 Plus Calibration Tester for approximately 30 minutes prior to testing to the thermometer to reach ambient room temperature.
Caution The ambient temperature must be stable and within the range of 18.3°C (65.0°F) – 26.7°C (80.0°F).
Caution Prolonged body contact with the thermometer may impact measurements. When instructed to wait before performing the next step place the thermometer on the work surface.
Figure 7. 9600 Plus Calibration Tester with Braun ThermoScan Pro 3000
1. Clean the probe tip with a cotton swab slightly moistened with isopropyl alcohol, remove excess alcohol with a clean cotton swab, and let air dry for 5 minutes. Do not use any chemical other than alcohol to clean the probe window.
24 Chapter 3 Operation Welch Allyn 9600 Plus Calibration Tester
2. Place the Braun ThermoScan Pro 3000 in calibration check mode using the following steps.
a. Wait until the thermometer is in Standby mode (the screen
is blank).
b. Press down and hold the Activation button, do not release
until Step d.
c. Press the /Mem button.
d. The LCD will flash “CAL” after the self check sequence.
The display will flash “CAL”, even if you have taken a temperature.
e. Release both buttons.
3. Apply a new probe cover. Place the device firmly into the Ear Device Port (Figure 7).
4. Wait approximately three seconds, press the thermometer Activation button, and listen for the beep.
5. Remove the Pro 3000 Thermometer from the 9600 Plus Calibration Tester and read the temperature in the thermometer’s display. If the temperatures are within ±0.2° C (±0.4° F), the thermometer is within calibration.
6. Wait one full minute before taking another reading with the same thermometer. Repeated measurements in short sequence may cause higher readings.
7. The device will exit Cal mode after approximately 4 minutes.
8. Test all available thermometers for calibration verification at the current Calibration Set Point Temperature. Proceed to the next Calibration Set Point Temperature, see "Changing the
Calibration Set Point Temperature" (page 11).
Loading...
+ 88 hidden pages