Welch Allyn, Inc.
7420 Carroll Road
San Diego, CA 92121-2334
(619) 621-6600
(800) 854-2904
U.S. Patents 265,525; 4,863,211; 4,911,559; 4,993,424, 4,993,438; and 5,018,872; Australia 108527 and 108528;
Belgium 19809 and 19694; Canada 2,001,196-3; 66391 and 66392; Denmark MR 0272 1991 and MR 0273 1991;
Norwegian 68692 and 68.370; Sweden 48,348; 48, 347; Switzerland 117,864 and 117,865; Taiwan R.O.C. New
Invention Pat. No. ND-25315; ND 25316; and 42337; United Kingdom 2001897 and 2001898; West Germany
M 89 07 483.1 and M 89 07 556.0; Additional U.S. and Foreign Patents Issued and Pending.
The Welch Allyn™ Model 9020 tympanic thermometer incorporates
infrared optics with advanced microelectronics to quickly and efficiently
register tympanic membrane-derived temperatures.
INFRARED THERMOMETRY
All matter emits thermally-generated electromagnetic waves. As the
temperature of an object increases, it increased radiates infrared energy
(heat rays) at amplitudes and wavelengths corresponding to its temperature.
The infrared sensing system of the 9020, co-developed with NASA’s Jet
Propulsion Laboratory, employs state-of-the art technology to measure
infrared radiation within the ear canal. The thermometer calculates the
patient’s temperature from the highest level of infrared energy detected.
Clinical studies indicate that the temperature of the tympanic membrane
correlates well with core body temperature.
KEY BENEFITS
The ergonomically-designed 9020 offers the most noninvasive approach
to determine body temperature.
Tympanic site measurements enhance patient comfort and avoid
contact with mucous membranes, substantially minimizing the risk of
cross-contamination.
In addition, tympanic temperatures are not affected by patient-specific
variables such as respiration, eating, drinking, and smoking.
The 9020 is ideal for intubated patients and others who cannot have their
temperatures taken orally.
NOTE: This manual provides important instructions. Be sure to
read the “Operating Procedures” section carefully. We recommend
that you practice taking a temperature to familiarize yourself with
the 9020 and develop the proper technique to ensure safe, quick,
and accurate temperatures.
performed in minutes with the
Welch Allyn 9600 Calibration
Tester
2
Page 7
WALL MOUNT (OPTION)
9020 Users’ Manual
REMOVING THE 9020 FROM THE WALL MOUNT
Lift the 9020 out of the wall mount
while holding the release key to
the right (as shown here). This
release key temporarily disengages the wall mount’s spring
latch mechanism and allows the
thermometer to be removed.
RELEASE KEY
3
Page 8
Welch Allyn 9020 Infrared Tympanic Thermometer
CONTROLS
SLIDE
Moves forward to extend the probe, automatically
applying a probe cover. Moves backward to retract
the probe, automatically removing the probe cover.
Once the probe is partially advanced or partially
retracted, you cannot reverse the direction of the
slide.
The slide will not move unless a probe
cover dispenser containing probe covers is in
place.
PROBE
COVER
DISPENSER
WINDOW
Vertical window in
the handle of the
9020 allows you to
check the supply
of probe covers
remaining in the
dispenser.
BATTERY
COMPARTMENT
Approximately 1 year of
operation (depending on
usage) from each set of
4 AAA batteries. Visual
and audible signals
indicate when batteries
are low.
RECALL
BUTTON
When pressed, the
last temperature taken
will appear on the
display for 5 seconds.
Moving the slide forward
erases the previous
temperature from the
thermometer's memory.
MODE SWITCH
Cycles through Oral,
Rectal, Core, and
Surface to select the
operation mode.
Temperatures must
be taken in the mode
desired; they will
not be converted
between modes.
4
Page 9
9020 Users’ Manual
LCD DISPLAY
Display provides
readout of
temperature,
status, operation
mode, and error
conditions.
Ergonomically
designed for visual
access as you hold
the thermometer.
PULSE TIMER
BUTTON
Pressing this
button activates a
30-second visual
and audible timer to
measure respiration
and/or pulse.
°F/°C SWITCH
Toggles the thermometric
scale between Fahrenheit
and Celsius, at any time.
When this button is pressed,
the currently displayed
temperature will be converted, and redisplayed.
PROBE
When inserted into the ear,
the infrared sensor in the
probe scans thermal energy
radiated within the ear canal.
SCAN
BUTTON
After inserting
the probe into the
ear canal, press this
button to take a
temperature. The
infrared sensor then
measures tympanic
temperatures, and
after 1 second with
no increase in
temperature, the
highest of these
temperatures is
displayed.
5
Page 10
Welch Allyn 9020 Infrared Tympanic Thermometer
DISPLAY INDICATORS
30-SECOND PULSE TIMER
Press the PULSE TIMER button to check pulse
or respiration. The first of six 5-second segments
blinks once per second for five seconds, and
then remains steady. The other segments
continue the countdown pattern until 30 seconds
has elapsed. The 9020 beeps when the timer is
first activated, after 15 seconds, and again after
30 seconds.
FAHRENHEIT/CELSIUS
This indicator shows the current thermometric
scale, Fahrenheit (F) or Celsius (C). Press the
°F/°C switch to toggle between these scales,
and the currently displayed temperature is
automatically converted.
LOW BATTERY
This symbol appears when the thermometer’s
microprocessor detects a low battery condition.
Once the Battery symbol is displayed, the 9020
will not function until the batteries have been
replaced.
THERMOMETER READY
The Ready symbol (a thermometer in a circle)
indicates that the 9020 has completed selfcalibration procedures. The thermometer is
ready for use in 2 – 15 seconds between
temperature measurements.
6
Page 11
RETRACT SLIDE
This symbol appears, along with an audible
double-beep, as a reminder when the probe
has not been fully retracted. After taking a
temperature, always move the slide back to
its original position to protect the probe from
damage and debris.
MALFUNCTION
If the 9020 requires service due to an internal
malfunction, this symbol appears, along with a
numeric error code indicating the nature of the
problem. When this indicator is displayed, no
temperatures can be taken.
UNDER/OVER TEMPERATURE
RANGE INDICATORS
This Up Arrow/Down Arrow symbol blinks for
five seconds, then remains steady if the 9020
detects a temperature that is beyond the normal
operating range (75.0°F – 108.0°F). The Down
Arrow indicates a temperature below 75°F. The
Up Arrow indicates a temperature above 108°F.
9020 Users’ Manual
7
Page 12
Welch Allyn 9020 Infrared Tympanic Thermometer
OPERATION MODES
The 9020 measures, converts, and displays temperatures in the
following modes:
•Oral
•Core
•Rectal
•Surface
The oral, core, and rectal modes display tympanic temperatures that
have been converted to reflect the corresponding temperature of the
traditional oral, core, or rectal sites.
Conversion factors used by the 9020 are derived from statistical mean
values obtained through comparisons of tympanic membrane-derived
temperatures with corresponding oral, core, and rectal temperatures.
The surface mode shows the actual, unadjusted radiated temperatures
measured during a scan (up to 30 seconds of continuous scan time).
To change operation modes, see the “Operating Procedures” section of
this manual.
The current operation mode of the 9020 appears in the top, right corner
of the thermometer’s display:
ORAL MODE
This mode converts the tympanic temperature to
display its “oral equivalent.” Use this mode
where glass or electronic oral thermometers are
now used. The normal range of oral temperature
is 97.6°F to 99.6°F (36.5°C to 37.5°C).
8
Page 13
CORE MODE
This mode converts the tympanic temperature to
display its “core equivalent.” Use this mode in
areas where esophageal or pulmonary artery
temperatures are now used. The normal range
of core temperature is 98.2°F to 100.2°F
(36.8°C to 37.9°C).
RECTAL MODE
This mode converts the tympanic temperature to
display its “rectal equivalent.” Use this mode
where glass or electronic rectal thermometers
are now used. The normal range of rectal
temperature is 98.6°F to 100.6°F
(37.0°C to 38.1°C).
SURFACE MODE
This mode measures the temperatures of
different surface areas. Pressing the SCAN
button activates the infrared sensor and displays
the continuous radiated temperatures. The
thermometer shows the last temperature read
when the SCAN button is released, or when 30
seconds has elapsed (a double beep sounds).
9020 Users’ Manual
TEST MODE (NOT AN OPERATION MODE)
This mode displays an unadjusted temperature.
The test mode is used exclusively to test or
calibrate the 9020 thermometer using the
WELCH ALLYN‚ 9600 Calibration Tester.
Please note that this function is accessed by a
special procedure described in the 9600
Operation Manual.
9
Page 14
Welch Allyn 9020 Infrared Tympanic Thermometer
PROBE COVERS
A dispenser of 100 disposable probe covers fits conveniently into the
thermometer’s handle. This one-piece design minimizes handling to
protect both the patient and the probe covers from environmental contaminants. The 9020 will not operate unless a probe cover is in place.
NOTE: Use
9020. The design and thermal properties of these probe covers
ensure the most efficient measurement of tympanic temperature.
As you advance the slide to extend the probe, a cover is automatically
positioned over the tip of the probe. The slide
probe cover dispenser containing probe covers is in place.
When you retract the probe, the probe cover is ejected automatically. To
avoid cross-contamination, probe covers cannot be reused.
Note that once the probe is partially advanced or partially retracted, you
cannot reverse the direction of the slide.
only
WELCH ALLYN Model 050 probe covers with the
will not move
Ultra-thin, heat-energy transparent film
stretches evenly over the entire probe tip.
Probe covers are attached to one another
within the probe cover dispenser; as one
cover is used, the next one advances into
place, ready for the next temperature.
unless a
WARNING: A damaged probe cover could affect the accuracy of
a temperature reading. If a probe cover is torn, retract the probe,
wait for the Ready symbol, then advance the slide again to apply
a new probe cover.
Look for the Ready symbol
on the display before you
advance the slide. This can
take from 2–15 seconds,
depending on when the last
temperature was taken.
Make sure that the temperature scale (°F, °C)
and the operation mode
(Oral, Core, Rectal) are
those that you want to use
to take the temperature.
READY
SYMBOL
Check the condition of the patient’s ear. The 9020 should not
be used if any discharge is present in or around the ear.
After extending the probe, inspect the probe cover. If it is
damaged or misaligned, return the slide to its original position
to discard the probe cover. Wait for the Ready symbol, then
extend the probe again.
Take the temperature within 30 seconds after extending the
probe. If the probe remains extended for more than 30 seconds, you will hear a double-beep every 5 seconds for 5
minutes or until you retract the probe. The 9020 will then
recalibrate automatically and be ready for use in a few
seconds.
11
Page 16
Welch Allyn 9020 Infrared Tympanic Thermometer
Insert probe snugly into ear, completely closing
2
off ear canal. Position probe towards patient’s
opposite temple, press SCAN button, and wait
for beep.
For optimum control,
obtain a temperature in the
patient’s right ear if you are
right-handed. Use the
patient’s left ear if you
are left-handed.
After approximately
1 second, a single beep
will sound to indicate that
the temperature has been
taken.
12
NOTE: Please be sure to read the next section, “Proper
Technique,” for important information about the technique required to obtain accurate temperatures.
Page 17
3
9020 Users’ Manual
Remove probe from ear. Retract slide, automatically removing probe cover. Read and record
temperature.
After the beep, indicating
that the temperature has
been taken, withdraw the
probe from the patient’s
ear.
Removing the probe prior
to the beep may result in a
temperature measurement
that is lower than the
actual temperature.
Hold the 9020 over a
waste container, then
return the slide back to its original position to retract the probe.
As you do, the probe cover is automatically removed. Read
and record the temperature.
The temperature will remain displayed for five minutes, unless
you take another temperature.
Before repeating a temperature measurement using the same
ear of the patient, wait 2 minutes to allow tympanic temperatures to stabilize, or use the opposite ear. Note that slight, but
insignificant, differences in temperature often exist between the
right ear and the left ear.
13
Page 18
Welch Allyn 9020 Infrared Tympanic Thermometer
PROPER TECHNIQUE
Obtaining correct temperature measurements with the 9020, as with any
thermometer, requires the proper technique.
There is a temperature
gradient within the ear canal
that causes variations in
temperature of up to 5°F,
as shown in this diagram.
Temperatures are lowest at
the external opening of the
ear canal due to ambient air
temperature.
The highest temperatures
occur at the deepest portion
of the ear canal, at or around
the tympanic membrane.
Measuring the tympanic membrane provides the most accurate temperature. However, the 9020’s infrared sensor will register only the tympanic
temperatures that it can “see” during a temperature measurement.
In some patients, anatomical characteristics of the ear canal may obstruct
the sensor’s view of the tympanic membrane.
You should optimize the sensor’s field of view by consistently:
•Positioning the probe snugly to close off the ear canal
•Directing the probe toward the opposite temple.
TAKING TEMPERATURES IN SURFACE MODE
Follow these steps to measure skin surface temperatures with the 9020:
•Make sure that the Ready symbol is displayed, and that the
thermometer is in the surface (S) operation mode. For information, see the “Operation Modes” and “Changing Operation Modes”
sections of this manual.
14
Page 19
9020 Users’ Manual
•Advance the slide to extend the probe tip, automatically applying
a probe cover.
•Hold the probe approximately 1/4" above the skin surface, and
position the sensor lens to “see” the area to be scanned.
•Press and
readings appear on the display approximately four times a
second for up to 30 seconds. After 30 seconds, a double beep
sounds, and the
•Release the SCAN button. Slowly retract the slide to its original
position, automatically removing the probe cover. Return the
9020 to the wall mount.
hold
the SCAN button. Updated temperature
last
surface temperature read is displayed.
CHANGING OPERATION MODES
You can change the operating mode of the 9020 at any time to measure
and display oral (O), core (C), rectal (R), or surface (S) temperatures.
For more information, see the “Operation Modes” section of this manual.
•Insert a pointed object, such as a paper clip, through the small
hole in the Mode button.
•Press in and hold the mode switch to cycle through the operation
modes. When the desired mode appears in the upper, right
corner of the display, release the Mode switch.
NOTE: Temperatures cannot be converted between modes. If you take
a temperature, then change the operation mode, pressing the RECALL
button will display that temperature in the previous mode.
SELF TEST
All segments of the display (LCD) are turned on for 1/2 second while the
probe is being moved out of the thermometer case; this is to permit the
verification of proper display operation.
The thermometer performs an internal self test continually while the unit
is self calibrating. The results are displayed as a “Test Temperature”
98.6°F (37°C) nominal, for 1/2 second following the display test. The
“Test Temperature” is also compared against internal limits, If the limit
test fails, a malfunction is displayed.
15
Page 20
Welch Allyn 9020 Infrared Tympanic Thermometer
MAINTENANCE
CHANGING THE PROBE COVER DISPENSER
A vertical window in the handle of
the 9020 allows you to see when
the probe cover dispenser is
empty. To replace the dispenser:
1. Ensure that the probe is
retracted, then remove the
probe cover dispenser from
the thermometer’s handle by
pulling on the tab at the end
of the dispenser (Figure A).
Discard the empty dispenser.
2. Check to see that the first
probe cover in the new
dispenser is positioned
properly (Figure A).
Figure B
16
Figure A
3. Insert the new probe cover
dispenser (flat end first) into the
handle of the thermometer. Align
the edges of the dispenser with
the slots inside of the handle, then
slide the dispenser upward until
it snaps into place (Figure B).
The 9020 is now ready for use.
You should routinely take a
moment to inspect the sensor lens
for clarity, and clean as needed.
Refer to “Cleaning the Probe
Sensor Lens” on the next page.
Page 21
9020 Users’ Manual
CLEANING THE PROBE SENSOR LENS
Dust and other minute particles of environmental debris can build up on
the probe sensor lens during normal use of the 9020.
The probe sensor lens should always be shiny and clear. Inspect the
lens frequently and clean as necessary. The lens should also be cleaned
as routine maintenance according to the two cleaning methods below.
NOTE: A dirty lens will affect the accuracy of temperature readings.
METHOD 1: REMOVING LIGHT DUST
(RECOMMENDED DAILY)
1. Advance the slide to extend the probe tip.
2. Leave the probe cover in place. Using a dry cotton swab, gently press
the probe cover against the sensor lens (Figure C). Loose debris on
the lens will adhere to the probe cover.
Figure C
3. Move the slide back to its original position to remove the probe cover
and retract the probe tip. The 9020 is now ready for use.
17
Page 22
Welch Allyn 9020 Infrared Tympanic Thermometer
METHOD 2: REMOVING SMUDGES OR EXCESSIVE DEBRIS
(RECOMMENDED AS NEEDED OR IF LOW TEMPERATURE
READINGS PERSIST AFTER METHOD #1)
MATERIALS NEEDED•Isopropyl Alcohol
•Cotton Swab
1. Advance the slide to extend the probe tip.
2. Manually remove the probe cover to expose the probe sensor lens.
3. Using a cotton swab slightly moistened with isopropyl alcohol, gently
wipe the surface of the lens to remove debris (Figure D).
4. Wipe the lens with a dry cotton swab to absorb moisture and pick up
any remaining debris (Figure D).
5. Move the slide back to its original position to retract the probe tip.
6. Allow 30 minutes before using the thermometer.
18
Figure D
Page 23
9020 Users’ Manual
CLEANING THE THERMOMETER
Clean the exterior of both the 9020 and the optional wall mount as
needed. Wipe all surfaces with a clean cloth, dampened with warm water
and a mild detergent or germicidal cleaner (see the table below for
acceptable cleaning agents).
•As some solutions may stain, try a small test area first.
•Make sure that the cloth is damp, but
not
wet.
•Do not allow cleaning solution to run inside of the thermometer.
•Never immerse the thermometer in the cleaning solution.
Sporicidin is a registered trademark of Sporicidin International (800) 424-3733. MetriSpray is a trademark
of Metrex Research Corporation (800) 841-1428.
WELCH ALLYN 9020 Infrared Tympanic Thermometer
for the procedure to sterilize the thermometer.
19
Page 24
Welch Allyn 9020 Infrared Tympanic Thermometer
CHANGING THE PROBE TIP
A damaged probe tip can tear the ultra-thin film in the probe covers. A
temperature taken with a torn probe cover may result in a measurement
that is higher than the actual temperature.
Periodically inspect the plastic probe tip for nicks or other damage, and
replace if necessary.
MATERIALS NEEDED•Probe Tip Replacement Kit
(Part #20688-000)
1. Advance the slide to extend the probe tip.
2. Manually remove the probe cover to expose the probe tip.
3. Remove the probe tip by turning it counterclockwise (Figure E).
Discard the damaged probe tip.
4. Install the new probe tip by turning it clockwise, then hand-tightening
(5 inch-lbs.) until it is positioned snugly against the base of the probe.
Do not over-tighten the tip.
5. As a precaution, clean the sensor lens to remove any dust or debris
that may have adhered to the exposed lens during replacement of the
probe tip.
Follow the steps for “Cleaning the Probe Sensor Lens – Method 2”
described earlier in this section of the manual.
Figure E
20
Page 25
9020 Users’ Manual
REPLACING THE BATTERIES
Under normal use, a set of batteries will provide power to the 9020 for
about 1 year, depending on usage. When battery power becomes low, a
Battery symbol appears on the display and the thermometer cannot be
used until they are replaced.
1. Remove the screw that secures the battery compartment door, then
press the release notch to remove the door (see Figure F).
2. Remove the four depleted batteries and dispose of properly.
3. Press SCAN button to reset the thermometer.
4. Install four new AAA size Alkaline batteries in the
position as shown in the diagram inside of the battery compartment.
The 9020 will automatically perform a diagnostic self-test of the
electronics (indicated by a segment cycling through the display).
If not, reinstall the batteries to ensure correct positioning.
exact
order and
Figure F
5. Replace the battery compartment door, then reinstall the screw to
secure the door (Figure F).
21
Page 26
Welch Allyn 9020 Infrared Tympanic Thermometer
THEFT PROTECTION
The optional wall mount serves as a base for the thermometer when it is
not in use. You can lock the 9020 into the wall mount by rotating the
release key 90° to the right. Once in this position, the key can be removed by pulling downward.
An optional security cable (20796-000) may be added to your thermometer and wall mount for single area use.
OPTIONS & ACCESSORIES
WELCH ALLYN 050 Probe Covers
WELCH ALLYN Part #05051
Box of 2 dispensers (100 probe covers/dispenser)
Packaged in quantities of 5,000 probe covers/carton
Probe Tip Replacement Kit
WELCH ALLYN Part #20688-000
12 replacement probe tips
Security Cable
WELCH ALLYN Part #20796-000
Lanyard cable attaches 9020 to wall mount for added security
WELCH ALLYN 9020 Wall Mount
WELCH ALLYN Part #20757-400
Additional or replacement wall mount for the 9020
Portable Stand
WELCH ALLYN Part #08071-000
Rolling stand provides additional security without sacrificing mobility
WELCH ALLYN 9600 Calibration Tester
WELCH ALLYN Part #01800-210
Temperature reference for calibration of WELCH ALLYN
thermometers (both infrared and thermistor types)
22
Page 27
TROUBLESHOOTING
9020 Users’ Manual
PROBLEM
Slide Will Not
Move
High
Temperatures
SOLUTION
Probe cover dispenser missing or empty. Check
position of dispenser in thermometer’s handle. If
empty, install new dispenser.
Probe covers positioned improperly within probe
cover dispenser. Remove dispenser from thermometer handle. Pull on first probe cover to remove
any misaligned or bent covers. Re-position first
probe cover to extend 1/8" beyond top of dispenser,
then reinstall dispenser to thermometer handle. If
problem persists, replace probe cover dispenser.
Probe tip not installed securely and may be lodged
against reference arm (shields lens when probe
retracted). Looking into front of 9020, press down on
reference arm while moving slide forward. Hand
tighten tip (clockwise) until snug against base of
probe, then retract probe.
Check current operation mode of the 9020. Core
and rectal modes obtain higher temperatures than
oral mode (test mode is used for calibration only).
Detected warmer temperature prior to inserting probe
into ear, or heat source (such as a heat lamp) nearby
when taking temperature. Insert probe into ear, wait
1 second, then press SCAN button.
Check integrity of probe cover prior to taking temperature. Damaged (or missing) probe cover can
result in temperature reading up to 1.5°F higher than
actual temperature.
23
Page 28
Welch Allyn 9020 Infrared Tympanic Thermometer
PROBLEM
Low
Temperatures
Inconsistent
Results of
Repeated
Temperatures
SOLUTION
Improper technique. Insert probe snugly into ear,
completely closing off ear canal.
If ear canal is curved significantly, sensor’s view of
the tympanic membrane may be obstructed. Gently
pull up and back on external ear to straighten ear
canal before inserting probe.
Debris on probe sensor lens. Clean lens. See “Maintenance” section of this manual for procedure.
Probe tip not installed securely; hand tighten tip
(clockwise) until snug against base of probe. See
“Maintenance” section of this manual for procedure.
Check current operation mode of the 9020. Oral
mode obtains lower temperatures than core and
rectal modes (test mode is used for calibration only).
Not enough time elapsed since last temperature
taken (same ear). Insertion of probe results in slight
“draw down” of ear canal temperature. Wait 2 minutes to allow temperature to stabilize between repeated temperature measurements in the same ear,
or use patient’s opposite ear.
24
Improper scanning technique. Insert probe snugly
into ear, completely closing off ear canal.
If ear canal is curved significantly, sensor’s view of
the tympanic membrane may be obstructed. Gently
pull up and back on external ear to straighten ear
canal before inserting probe.
Page 29
9020 Users’ Manual
PROBLEM
Improper
Delivery of
Probe Cover
or
Probe Cover
Positioned
Incorrectly
on Probe
SOLUTION
Probe covers positioned improperly within probe
cover dispenser. Remove dispenser from thermometer handle. Pull on first probe cover to remove
any misaligned or bent covers. Re-position first
probe cover to extend 1/8" beyond top of dispenser,
then reinstall dispenser to thermometer handle. If
problem persists, replace probe cover dispenser.
Obstruction between probe and probe cover dispenser. Remove dispenser, inspect thermometer
handle and remove debris, if any. Ensure probe
covers are properly positioned in probe cover dispenser, then reinsert dispenser into handle.
Probe tip is damaged. Replace with new probe tip.
See “Maintenance” section of this manual for procedure.
Probe tip is damaged. Replace with new probe tip.
See “Maintenance” section of this manual for procedure.
Probe covers positioned improperly within probe
cover dispenser. Remove dispenser from thermometer handle. Pull on first probe cover to remove
any misaligned or bent covers. Re-position first
probe cover to extend 1/8" beyond top of dispenser,
then reinstall dispenser to thermometer handle. If
problem persists, replace probe cover dispenser.
25
Page 30
Welch Allyn 9020 Infrared Tympanic Thermometer
PROBLEM
Continuous,
Intermittent
Audible Tone
Blank LCD
Display, No
Temperature
on Display
Unable to
Remove
9020 from
Wall Mount
SOLUTION
Probe is partially extended. Ensure slide is fully
retracted.
Improper operating condition. Check display for
error status. See “Error Codes” section of this manual
for explanation.
Probe extended for over 5 minutes. Retract probe.
Battery power depleted. Replace batteries. See
“Maintenance” section of this manual for procedure.
Wall mount’s spring latch mechanism is engaged.
Hold release key (underside of wall mount) to the
right while removing thermometer.
Release key for spring latch mechanism (underside
of wall mount) is missing or locked. Insert key into
mechanism, then rotate key clockwise until tab is
perpendicular to front panel of wall mount. Hold
release key to the right while removing thermometer.
26
Page 31
9020 Users’ Manual
ERROR CODES
When the 9020 detects an improper operating condition, the malfunction
symbol (X) will appear with one of the numeric error codes (Er) below.
Set the thermometer aside for one hour to allow
the electronics to stabilize to room temperature.
•If the 9020 is faulty, the error status will
remain to indicate that the thermometer
needs service.
•If the condition is transient in nature, the
error status will disappear. The 9020 will
then self-calibrate, and be ready for use.
ERROREXPLANATION
Er2The ambient temperature is beyond the operating
range of 60°F – 104°F. Allow the thermometer to
stabilize at normal room temperature before use.
Er4The infrared sensing system may be detecting a
nearby temperature source that is greater than
110°F. If not, the thermometer requires service.
Er5The ambient temperature is above 110°F.
Er6If not, the thermometer requires service.
For added security, an optional Security Cable may be installed that will
connect the thermometer to the wall mount.
MATERIALS NEEDED•Security Cable (Part #20796-000)
•Flat-Blade and Phillips Screwdrivers
1. Insert screwdriver into the wall mount to release the clip.
2. Insert one end of the security cable through the slot of the clip, then
reinsert the clip into the wall mount.
28
Page 33
9020 Users’ Manual
3. Remove the screw from the battery door. Spread the case of the
thermometer just wide enough to insert the other end of the security
cable. Close the battery door and reinstall the screw.
WALL MOUNT INSTALLATION
The following guidelines are provided to ensure proper placement and
installation of your thermometer wall mount.
The top of the wall mount should be located about 60 inches from the floor.
You will need at least a 4 inch clearance from the bottom and sides of the
wall mount and 12 inches from the top.
Wall Mount Requires Three
Pan or Round Head Screws (Your Choice of #6, #8, or #10)
Wall SurfaceFastener
Dry wallE-Z Twist in anchor (Molly or toggle)*
Welch Allyn, Inc. (“Welch Allyn”) warrants that each new Welch Allyn Model 9020
Thermometer (the “Product”) is free from defects in material and workmanship under
normal use and service for a period of one year for new units or 1 year for refurbished from
the date of shipment from Welch Allyn. This Warranty shall be fulfilled by the repair or
replacement by Welch Allyn or its authorized representative, at Welch Allyn’s discretion of
any such defect, free of charge for parts and labor.
EXCLUSIONS FROM WARRANTY
This Welch Allyn Warranty shall be void, and Welch Allyn will not be responsible for any
loss associated with, any of the following: (i) probe tips damaged or defective after
delivery to Customer, (ii) not providing routine maintenance and calibration, (iii) missing,
damaged, or discharged batteries after the first year, (iv) any product that (a) has had the
serial number defaced, (b) has been repaired by anyone other than Welch Allyn or an
authorized Welch Allyn service representative (c) has been altered or (d) has been used
in any manner other than in accordance with instructions or shows evidence of abusive
handling.
WARRANTY CLAIMS
If a Product requires service during the applicable warranty period, the Customer should
immediately return the Product at Customer’s own expense to Welch Allyn, 7420 Carroll
Road, San Diego, CA 92121. Any risk of loss or damage in shipment of the Product to
Welch Allyn shall be borne exclusively by Customer. Repair or replacement will be made
at Welch Allyn’s expense subject to the terms of this Warranty. Product repaired under
warranty provisions will be returned to Customer at Welch Allyn's expense. Any damage
in shipment from Welch Allyn is the responsibility of the delivery carrier.
LIMITATIONS
In no event shall Welch Allyn be liable for any incidental, indirect, or consequential
damages in connection with the purchase or use of any Product.
This Warranty is in lieu of other obligations or liabilities on Welch Allyn’s part, and Welch
Allyn neither assumes, nor authorizes, any representative or other person to assume for
it any other liability in connection with the sale of the Product.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTY
OF MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED, WELCH ALLYN WILL NOT
BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES.
31
Page 36
70858-0000B
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.