Welch Allyn 9020 User manual

Page 1
Thermometry
Service Manual
Page 2
The CE mark on this product indicates it has been tested to and conforms with the provisions noted within the 93/42/EEC Medical Device Directive.
European Regulatory Manager, Medical Division Welch Allyn Ltd., Navan, Co. Meath Republic of Ireland
FOR TECHNICAL SUPPORT AND CUSTOMER SERVICE
Welch Allyn, Inc. 7420 Carroll Road San Diego, CA 92121-2334 (619) 621-6600 (800) 854-2904
U.S. Patents 265,525; 4,863,211; 4,911,559; 4,993,424, 4,993,438; and 5,018,872; Australia 108527 and 108528; Belgium 19809 and 19694; Canada 2,001,196-3; 66391 and 66392; Denmark MR 0272 1991 and MR 0273 1991; Norwegian 68692 and 68.370; Sweden 48,348; 48, 347; Switzerland 117,864 and 117,865; Taiwan R.O.C. New Invention Pat. No. ND-25315; ND 25316; and 42337; United Kingdom 2001897 and 2001898; West Germany M 89 07 483.1 and M 89 07 556.0; Additional U.S. and Foreign Patents Issued and Pending.
Page 3
CONTENTS
INTRODUCTION.................................................................................... 1
Infrared Thermometry..................................................................... 1
Key Benefits ................................................................................... 1
FEATURES ............................................................................................ 2
WALL MOUNT (OPTION)...................................................................... 3
Removing the 9020 from the Wall Mount ....................................... 3
CONTROLS ........................................................................................... 4
DISPLAY INDICATORS......................................................................... 6
OPERATION MODES ............................................................................ 8
PROBE COVERS ................................................................................ 10
OPERATING PROCEDURES .............................................................. 11
Taking Temperatures in Oral, Core, Rectal Modes ...................... 11
Proper Technique ......................................................................... 14
Taking Temperatures in Surface Mode ........................................ 14
Changing Operation Modes.......................................................... 15
Self Test ....................................................................................... 15
MAINTENANCE ................................................................................... 16
Changing the Probe Cover Dispenser .......................................... 16
Cleaning the Probe Sensor Lens .................................................. 17
Method 1: Removing Light Dust........................................... 17
Method 2: Removing Smudges or Excessive Debris ........... 18
Cleaning the Thermometer ........................................................... 19
Changing the Probe Tip................................................................ 20
Replacing the Batteries................................................................. 21
i
Page 4
Welch Allyn 9020 Infrared Tympanic Thermometer
THEFT PROTECTION ......................................................................... 22
OPTIONS & ACCESSORIES .............................................................. 22
TROUBLESHOOTING......................................................................... 23
ERROR CODES .................................................................................. 27
SECURITY CABLE INSTALLATION (OPTION) .................................. 28
WALL MOUNT INSTALLATION (OPTION) ......................................... 29
SPECIFICATIONS ............................................................................... 30
LIMITED WARRANTY ......................................................................... 31
ii
Page 5
9020 Users’ Manual

INTRODUCTION

The Welch Allyn™ Model 9020 tympanic thermometer incorporates infrared optics with advanced microelectronics to quickly and efficiently register tympanic membrane-derived temperatures.

INFRARED THERMOMETRY

All matter emits thermally-generated electromagnetic waves. As the temperature of an object increases, it increased radiates infrared energy (heat rays) at amplitudes and wavelengths corresponding to its tempera­ture.
The infrared sensing system of the 9020, co-developed with NASA’s Jet Propulsion Laboratory, employs state-of-the art technology to measure infrared radiation within the ear canal. The thermometer calculates the patient’s temperature from the highest level of infrared energy detected.
Clinical studies indicate that the temperature of the tympanic membrane correlates well with core body temperature.

KEY BENEFITS

The ergonomically-designed 9020 offers the most noninvasive approach to determine body temperature.
Tympanic site measurements enhance patient comfort and avoid contact with mucous membranes, substantially minimizing the risk of cross-contamination.
In addition, tympanic temperatures are not affected by patient-specific variables such as respiration, eating, drinking, and smoking.
The 9020 is ideal for intubated patients and others who cannot have their temperatures taken orally.
NOTE: This manual provides important instructions. Be sure to read the “Operating Procedures” section carefully. We recommend that you practice taking a temperature to familiarize yourself with the 9020 and develop the proper technique to ensure safe, quick, and accurate temperatures.
1
Page 6
Welch Allyn 9020 Infrared Tympanic Thermometer

FEATURES

Approximately 1-second scan
time to register body temperature
Patented, fully-enclosed
probe cover dispenser holds 100 disposable probe covers
Automated placement of probe
covers to maintain cleanliness
Retractable probe tip protects
sensor when thermometer not in use
Probe cannot be extended
when probe cover dispenser is empty or missing
Choice of operating modes
includes oral, core, and rectal equivalency
Surface operating mode scans
skin surface temperatures
One-touch Fahrenheit – Celsius
selector switch
Simple, 1–2–3 step operation
Convenient and lightweight
one-piece design
Large, LCD display shows
temperature, mode, and other status indicators.
Visual and audible pulse/
respiration timer
Memory recall displays last
temperature taken
Thermometer continuously and
automatically self-calibrates before every use
Optional wall mount allows the
thermometer to be locked into place.
Field calibration can be
performed in minutes with the Welch Allyn 9600 Calibration Tester
2
Page 7
WALL MOUNT (OPTION)
9020 Users’ Manual
REMOVING THE 9020 FROM THE WALL MOUNT
Lift the 9020 out of the wall mount while holding the release key to the right (as shown here). This release key temporarily disen­gages the wall mount’s spring latch mechanism and allows the thermometer to be removed.
RELEASE KEY
3
Page 8
Welch Allyn 9020 Infrared Tympanic Thermometer

CONTROLS

SLIDE
Moves forward to extend the probe, automatically applying a probe cover. Moves backward to retract the probe, automatically removing the probe cover. Once the probe is partially advanced or partially retracted, you cannot reverse the direction of the slide.
The slide will not move unless a probe cover dispenser containing probe covers is in place.
PROBE COVER DISPENSER WINDOW
Vertical window in the handle of the 9020 allows you to check the supply of probe covers remaining in the dispenser.
BATTERY COMPARTMENT
Approximately 1 year of operation (depending on usage) from each set of 4 AAA batteries. Visual and audible signals indicate when batteries are low.
RECALL
BUTTON
When pressed, the
last temperature taken
will appear on the
display for 5 seconds.
Moving the slide forward
erases the previous
temperature from the
thermometer's memory.
MODE SWITCH
Cycles through Oral,
Rectal, Core, and
Surface to select the
operation mode.
Temperatures must
be taken in the mode
desired; they will
not be converted
between modes.
4
Page 9
9020 Users’ Manual
LCD DISPLAY
Display provides readout of temperature, status, operation mode, and error conditions. Ergonomically designed for visual access as you hold the thermometer.
PULSE TIMER BUTTON
Pressing this button activates a 30-second visual and audible timer to measure respiration and/or pulse.
°F/°C SWITCH
Toggles the thermometric scale between Fahrenheit and Celsius, at any time. When this button is pressed, the currently displayed temperature will be con­verted, and redisplayed.
PROBE
When inserted into the ear,
the infrared sensor in the
probe scans thermal energy
radiated within the ear canal.
SCAN
BUTTON
After inserting
the probe into the
ear canal, press this
button to take a
temperature. The
infrared sensor then
measures tympanic
temperatures, and
after 1 second with
no increase in
temperature, the
highest of these
temperatures is
displayed.
5
Page 10
Welch Allyn 9020 Infrared Tympanic Thermometer

DISPLAY INDICATORS

30-SECOND PULSE TIMER
Press the PULSE TIMER button to check pulse or respiration. The first of six 5-second segments blinks once per second for five seconds, and then remains steady. The other segments continue the countdown pattern until 30 seconds has elapsed. The 9020 beeps when the timer is first activated, after 15 seconds, and again after 30 seconds.
FAHRENHEIT/CELSIUS
This indicator shows the current thermometric scale, Fahrenheit (F) or Celsius (C). Press the °F/°C switch to toggle between these scales, and the currently displayed temperature is automatically converted.
LOW BATTERY
This symbol appears when the thermometer’s microprocessor detects a low battery condition. Once the Battery symbol is displayed, the 9020 will not function until the batteries have been replaced.
THERMOMETER READY
The Ready symbol (a thermometer in a circle) indicates that the 9020 has completed self­calibration procedures. The thermometer is ready for use in 2 – 15 seconds between temperature measurements.
6
Page 11
RETRACT SLIDE
This symbol appears, along with an audible double-beep, as a reminder when the probe has not been fully retracted. After taking a temperature, always move the slide back to its original position to protect the probe from damage and debris.
MALFUNCTION
If the 9020 requires service due to an internal malfunction, this symbol appears, along with a numeric error code indicating the nature of the problem. When this indicator is displayed, no temperatures can be taken.
UNDER/OVER TEMPERATURE RANGE INDICATORS
This Up Arrow/Down Arrow symbol blinks for five seconds, then remains steady if the 9020 detects a temperature that is beyond the normal operating range (75.0°F – 108.0°F). The Down Arrow indicates a temperature below 75°F. The Up Arrow indicates a temperature above 108°F.
9020 Users’ Manual
7
Page 12
Welch Allyn 9020 Infrared Tympanic Thermometer

OPERATION MODES

The 9020 measures, converts, and displays temperatures in the following modes:

Oral

Core
Rectal
Surface
The oral, core, and rectal modes display tympanic temperatures that have been converted to reflect the corresponding temperature of the traditional oral, core, or rectal sites.
Conversion factors used by the 9020 are derived from statistical mean values obtained through comparisons of tympanic membrane-derived temperatures with corresponding oral, core, and rectal temperatures.
The surface mode shows the actual, unadjusted radiated temperatures measured during a scan (up to 30 seconds of continuous scan time).
To change operation modes, see the “Operating Procedures” section of this manual.
The current operation mode of the 9020 appears in the top, right corner of the thermometer’s display:
ORAL MODE
This mode converts the tympanic temperature to display its “oral equivalent.” Use this mode where glass or electronic oral thermometers are now used. The normal range of oral temperature is 97.6°F to 99.6°F (36.5°C to 37.5°C).
8
Page 13
CORE MODE
This mode converts the tympanic temperature to display its “core equivalent.” Use this mode in areas where esophageal or pulmonary artery temperatures are now used. The normal range of core temperature is 98.2°F to 100.2°F (36.8°C to 37.9°C).
RECTAL MODE
This mode converts the tympanic temperature to display its “rectal equivalent.” Use this mode where glass or electronic rectal thermometers are now used. The normal range of rectal temperature is 98.6°F to 100.6°F (37.0°C to 38.1°C).
SURFACE MODE
This mode measures the temperatures of different surface areas. Pressing the SCAN button activates the infrared sensor and displays the continuous radiated temperatures. The thermometer shows the last temperature read when the SCAN button is released, or when 30 seconds has elapsed (a double beep sounds).
9020 Users’ Manual
TEST MODE (NOT AN OPERATION MODE)
This mode displays an unadjusted temperature. The test mode is used exclusively to test or calibrate the 9020 thermometer using the WELCH ALLYN‚ 9600 Calibration Tester. Please note that this function is accessed by a special procedure described in the 9600 Operation Manual.
9
Page 14
Welch Allyn 9020 Infrared Tympanic Thermometer

PROBE COVERS

A dispenser of 100 disposable probe covers fits conveniently into the thermometer’s handle. This one-piece design minimizes handling to protect both the patient and the probe covers from environmental con­taminants. The 9020 will not operate unless a probe cover is in place.
NOTE: Use
9020. The design and thermal properties of these probe covers ensure the most efficient measurement of tympanic temperature.
As you advance the slide to extend the probe, a cover is automatically positioned over the tip of the probe. The slide probe cover dispenser containing probe covers is in place.
When you retract the probe, the probe cover is ejected automatically. To avoid cross-contamination, probe covers cannot be reused.
Note that once the probe is partially advanced or partially retracted, you cannot reverse the direction of the slide.
only
WELCH ALLYN Model 050 probe covers with the
will not move
Ultra-thin, heat-energy transparent film stretches evenly over the entire probe tip.
Probe covers are attached to one another within the probe cover dispenser; as one cover is used, the next one advances into place, ready for the next temperature.
unless a
WARNING: A damaged probe cover could affect the accuracy of a temperature reading. If a probe cover is torn, retract the probe, wait for the Ready symbol, then advance the slide again to apply a new probe cover.
10
Page 15
9020 Users’ Manual

OPERATING PROCEDURES

TAKING TEMPERATURES IN ORAL, CORE, RECTAL MODES
1
Slowly move slide forward to extend probe, automatically applying probe cover.
Look for the Ready symbol on the display before you advance the slide. This can take from 2–15 seconds, depending on when the last temperature was taken.
Make sure that the tem­perature scale (°F, °C) and the operation mode (Oral, Core, Rectal) are those that you want to use to take the temperature.
READY SYMBOL
Check the condition of the patient’s ear. The 9020 should not be used if any discharge is present in or around the ear.
After extending the probe, inspect the probe cover. If it is damaged or misaligned, return the slide to its original position to discard the probe cover. Wait for the Ready symbol, then extend the probe again.
Take the temperature within 30 seconds after extending the probe. If the probe remains extended for more than 30 sec­onds, you will hear a double-beep every 5 seconds for 5 minutes or until you retract the probe. The 9020 will then recalibrate automatically and be ready for use in a few seconds.
11
Page 16
Welch Allyn 9020 Infrared Tympanic Thermometer
Insert probe snugly into ear, completely closing
2
off ear canal. Position probe towards patient’s opposite temple, press SCAN button, and wait for beep.
For optimum control, obtain a temperature in the patient’s right ear if you are right-handed. Use the patient’s left ear if you are left-handed.
After approximately 1 second, a single beep will sound to indicate that the temperature has been taken.
12
NOTE: Please be sure to read the next section, “Proper Technique,” for important information about the tech­nique required to obtain accurate temperatures.
Page 17
3
9020 Users’ Manual
Remove probe from ear. Retract slide, automati­cally removing probe cover. Read and record temperature.
After the beep, indicating that the temperature has been taken, withdraw the probe from the patient’s ear.
Removing the probe prior to the beep may result in a temperature measurement that is lower than the actual temperature.
Hold the 9020 over a waste container, then return the slide back to its original position to retract the probe. As you do, the probe cover is automatically removed. Read and record the temperature.
The temperature will remain displayed for five minutes, unless you take another temperature.
Before repeating a temperature measurement using the same ear of the patient, wait 2 minutes to allow tympanic tempera­tures to stabilize, or use the opposite ear. Note that slight, but insignificant, differences in temperature often exist between the right ear and the left ear.
13
Page 18
Welch Allyn 9020 Infrared Tympanic Thermometer

PROPER TECHNIQUE

Obtaining correct temperature measurements with the 9020, as with any thermometer, requires the proper technique.
There is a temperature gradient within the ear canal that causes variations in temperature of up to 5°F, as shown in this diagram.
Temperatures are lowest at the external opening of the ear canal due to ambient air temperature.
The highest temperatures occur at the deepest portion of the ear canal, at or around the tympanic membrane.
Measuring the tympanic membrane provides the most accurate tempera­ture. However, the 9020’s infrared sensor will register only the tympanic temperatures that it can “see” during a temperature measurement.
In some patients, anatomical characteristics of the ear canal may obstruct the sensor’s view of the tympanic membrane.
You should optimize the sensor’s field of view by consistently:
Positioning the probe snugly to close off the ear canal
Directing the probe toward the opposite temple.
TAKING TEMPERATURES IN SURFACE MODE
Follow these steps to measure skin surface temperatures with the 9020:
Make sure that the Ready symbol is displayed, and that the
thermometer is in the surface (S) operation mode. For informa­tion, see the “Operation Modes” and “Changing Operation Modes” sections of this manual.
14
Page 19
9020 Users’ Manual
Advance the slide to extend the probe tip, automatically applying
a probe cover.
Hold the probe approximately 1/4" above the skin surface, and
position the sensor lens to “see” the area to be scanned.
Press and
readings appear on the display approximately four times a second for up to 30 seconds. After 30 seconds, a double beep sounds, and the
Release the SCAN button. Slowly retract the slide to its original
position, automatically removing the probe cover. Return the 9020 to the wall mount.
hold
the SCAN button. Updated temperature
last
surface temperature read is displayed.

CHANGING OPERATION MODES

You can change the operating mode of the 9020 at any time to measure and display oral (O), core (C), rectal (R), or surface (S) temperatures. For more information, see the “Operation Modes” section of this manual.
Insert a pointed object, such as a paper clip, through the small
hole in the Mode button.
Press in and hold the mode switch to cycle through the operation
modes. When the desired mode appears in the upper, right corner of the display, release the Mode switch.
NOTE: Temperatures cannot be converted between modes. If you take a temperature, then change the operation mode, pressing the RECALL button will display that temperature in the previous mode.

SELF TEST

All segments of the display (LCD) are turned on for 1/2 second while the probe is being moved out of the thermometer case; this is to permit the verification of proper display operation.
The thermometer performs an internal self test continually while the unit is self calibrating. The results are displayed as a “Test Temperature”
98.6°F (37°C) nominal, for 1/2 second following the display test. The “Test Temperature” is also compared against internal limits, If the limit test fails, a malfunction is displayed.
15
Page 20
Welch Allyn 9020 Infrared Tympanic Thermometer

MAINTENANCE

CHANGING THE PROBE COVER DISPENSER
A vertical window in the handle of the 9020 allows you to see when the probe cover dispenser is empty. To replace the dispenser:
1. Ensure that the probe is retracted, then remove the probe cover dispenser from the thermometer’s handle by pulling on the tab at the end of the dispenser (Figure A). Discard the empty dispenser.
2. Check to see that the first probe cover in the new dispenser is positioned properly (Figure A).
Figure B
16
Figure A
3. Insert the new probe cover dispenser (flat end first) into the handle of the thermometer. Align the edges of the dispenser with the slots inside of the handle, then slide the dispenser upward until it snaps into place (Figure B). The 9020 is now ready for use.
You should routinely take a moment to inspect the sensor lens for clarity, and clean as needed. Refer to “Cleaning the Probe Sensor Lens” on the next page.
Page 21
9020 Users’ Manual
CLEANING THE PROBE SENSOR LENS
Dust and other minute particles of environmental debris can build up on the probe sensor lens during normal use of the 9020.
The probe sensor lens should always be shiny and clear. Inspect the lens frequently and clean as necessary. The lens should also be cleaned as routine maintenance according to the two cleaning methods below.
NOTE: A dirty lens will affect the accuracy of temperature readings.
METHOD 1: REMOVING LIGHT DUST
(RECOMMENDED DAILY)
1. Advance the slide to extend the probe tip.
2. Leave the probe cover in place. Using a dry cotton swab, gently press the probe cover against the sensor lens (Figure C). Loose debris on the lens will adhere to the probe cover.
Figure C
3. Move the slide back to its original position to remove the probe cover and retract the probe tip. The 9020 is now ready for use.
17
Page 22
Welch Allyn 9020 Infrared Tympanic Thermometer
METHOD 2: REMOVING SMUDGES OR EXCESSIVE DEBRIS
(RECOMMENDED AS NEEDED OR IF LOW TEMPERATURE READINGS PERSIST AFTER METHOD #1)
MATERIALS NEEDED Isopropyl Alcohol
Cotton Swab
1. Advance the slide to extend the probe tip.
2. Manually remove the probe cover to expose the probe sensor lens.
3. Using a cotton swab slightly moistened with isopropyl alcohol, gently wipe the surface of the lens to remove debris (Figure D).
4. Wipe the lens with a dry cotton swab to absorb moisture and pick up any remaining debris (Figure D).
5. Move the slide back to its original position to retract the probe tip.
6. Allow 30 minutes before using the thermometer.
18
Figure D
Page 23
9020 Users’ Manual
CLEANING THE THERMOMETER
Clean the exterior of both the 9020 and the optional wall mount as needed. Wipe all surfaces with a clean cloth, dampened with warm water and a mild detergent or germicidal cleaner (see the table below for acceptable cleaning agents).
As some solutions may stain, try a small test area first.
Make sure that the cloth is damp, but
not
wet.
Do not allow cleaning solution to run inside of the thermometer.
Never immerse the thermometer in the cleaning solution.

CLEANING AGENT LENS THERMOMETER WALL MOUNT

Ammonium ✔✔ Bleach (50%) ✔✔ Sproricidin® ✔✔ MetriSpray™ ✔✔ Cidex ✔✔ Detergent (mild) ✔✔ Isopropyl Alcohol ✔✔ Lysol ✔✔
Refer to the
Technical Manual
Sporicidin is a registered trademark of Sporicidin International (800) 424-3733. MetriSpray is a trademark of Metrex Research Corporation (800) 841-1428.
WELCH ALLYN 9020 Infrared Tympanic Thermometer
for the procedure to sterilize the thermometer.
19
Page 24
Welch Allyn 9020 Infrared Tympanic Thermometer
CHANGING THE PROBE TIP
A damaged probe tip can tear the ultra-thin film in the probe covers. A temperature taken with a torn probe cover may result in a measurement that is higher than the actual temperature.
Periodically inspect the plastic probe tip for nicks or other damage, and replace if necessary.
MATERIALS NEEDED Probe Tip Replacement Kit
(Part #20688-000)
1. Advance the slide to extend the probe tip.
2. Manually remove the probe cover to expose the probe tip.
3. Remove the probe tip by turning it counterclockwise (Figure E). Discard the damaged probe tip.
4. Install the new probe tip by turning it clockwise, then hand-tightening (5 inch-lbs.) until it is positioned snugly against the base of the probe. Do not over-tighten the tip.
5. As a precaution, clean the sensor lens to remove any dust or debris that may have adhered to the exposed lens during replacement of the probe tip.
Follow the steps for “Cleaning the Probe Sensor Lens – Method 2” described earlier in this section of the manual.
Figure E
20
Page 25
9020 Users’ Manual

REPLACING THE BATTERIES

Under normal use, a set of batteries will provide power to the 9020 for about 1 year, depending on usage. When battery power becomes low, a Battery symbol appears on the display and the thermometer cannot be used until they are replaced.
1. Remove the screw that secures the battery compartment door, then press the release notch to remove the door (see Figure F).
2. Remove the four depleted batteries and dispose of properly.
3. Press SCAN button to reset the thermometer.
4. Install four new AAA size Alkaline batteries in the position as shown in the diagram inside of the battery compartment.
The 9020 will automatically perform a diagnostic self-test of the electronics (indicated by a segment cycling through the display). If not, reinstall the batteries to ensure correct positioning.
exact
order and
Figure F
5. Replace the battery compartment door, then reinstall the screw to secure the door (Figure F).
21
Page 26
Welch Allyn 9020 Infrared Tympanic Thermometer

THEFT PROTECTION

The optional wall mount serves as a base for the thermometer when it is not in use. You can lock the 9020 into the wall mount by rotating the release key 90° to the right. Once in this position, the key can be re­moved by pulling downward.
An optional security cable (20796-000) may be added to your thermom­eter and wall mount for single area use.

OPTIONS & ACCESSORIES

WELCH ALLYN 050 Probe Covers
WELCH ALLYN Part #05051 Box of 2 dispensers (100 probe covers/dispenser) Packaged in quantities of 5,000 probe covers/carton
Probe Tip Replacement Kit
WELCH ALLYN Part #20688-000 12 replacement probe tips
Security Cable
WELCH ALLYN Part #20796-000 Lanyard cable attaches 9020 to wall mount for added security
WELCH ALLYN 9020 Wall Mount
WELCH ALLYN Part #20757-400 Additional or replacement wall mount for the 9020
Portable Stand
WELCH ALLYN Part #08071-000 Rolling stand provides additional security without sacrificing mobility
WELCH ALLYN 9600 Calibration Tester
WELCH ALLYN Part #01800-210 Temperature reference for calibration of WELCH ALLYN thermometers (both infrared and thermistor types)
22
Page 27

TROUBLESHOOTING

9020 Users’ Manual
PROBLEM
Slide Will Not Move
High Temperatures
SOLUTION
Probe cover dispenser missing or empty. Check position of dispenser in thermometer’s handle. If empty, install new dispenser.
Probe covers positioned improperly within probe cover dispenser. Remove dispenser from ther­mometer handle. Pull on first probe cover to remove any misaligned or bent covers. Re-position first probe cover to extend 1/8" beyond top of dispenser, then reinstall dispenser to thermometer handle. If problem persists, replace probe cover dispenser.
Probe tip not installed securely and may be lodged against reference arm (shields lens when probe retracted). Looking into front of 9020, press down on reference arm while moving slide forward. Hand tighten tip (clockwise) until snug against base of probe, then retract probe.
Check current operation mode of the 9020. Core and rectal modes obtain higher temperatures than oral mode (test mode is used for calibration only).
Detected warmer temperature prior to inserting probe into ear, or heat source (such as a heat lamp) nearby when taking temperature. Insert probe into ear, wait 1 second, then press SCAN button.
Check integrity of probe cover prior to taking tem­perature. Damaged (or missing) probe cover can result in temperature reading up to 1.5°F higher than actual temperature.
23
Page 28
Welch Allyn 9020 Infrared Tympanic Thermometer
PROBLEM
Low Temperatures
Inconsistent Results of Repeated Temperatures
SOLUTION
Improper technique. Insert probe snugly into ear, completely closing off ear canal.
If ear canal is curved significantly, sensor’s view of the tympanic membrane may be obstructed. Gently pull up and back on external ear to straighten ear canal before inserting probe.
Debris on probe sensor lens. Clean lens. See “Main­tenance” section of this manual for procedure.
Probe tip not installed securely; hand tighten tip (clockwise) until snug against base of probe. See “Maintenance” section of this manual for procedure.
Check current operation mode of the 9020. Oral mode obtains lower temperatures than core and rectal modes (test mode is used for calibration only).
Not enough time elapsed since last temperature taken (same ear). Insertion of probe results in slight “draw down” of ear canal temperature. Wait 2 min­utes to allow temperature to stabilize between re­peated temperature measurements in the same ear, or use patient’s opposite ear.
24
Improper scanning technique. Insert probe snugly into ear, completely closing off ear canal.
If ear canal is curved significantly, sensor’s view of the tympanic membrane may be obstructed. Gently pull up and back on external ear to straighten ear canal before inserting probe.
Page 29
9020 Users’ Manual
PROBLEM
Improper Delivery of Probe Cover
or Probe Cover
Positioned Incorrectly on Probe
SOLUTION
Probe covers positioned improperly within probe cover dispenser. Remove dispenser from ther­mometer handle. Pull on first probe cover to remove any misaligned or bent covers. Re-position first probe cover to extend 1/8" beyond top of dispenser, then reinstall dispenser to thermometer handle. If problem persists, replace probe cover dispenser.
Obstruction between probe and probe cover dis­penser. Remove dispenser, inspect thermometer handle and remove debris, if any. Ensure probe covers are properly positioned in probe cover dis­penser, then reinsert dispenser into handle.
Probe tip is damaged. Replace with new probe tip. See “Maintenance” section of this manual for proce­dure.
Damaged probe covers. Replace probe cover dis­penser.
Probe Cover Breakage
Probe tip is damaged. Replace with new probe tip. See “Maintenance” section of this manual for proce­dure.
Probe covers positioned improperly within probe cover dispenser. Remove dispenser from ther­mometer handle. Pull on first probe cover to remove any misaligned or bent covers. Re-position first probe cover to extend 1/8" beyond top of dispenser, then reinstall dispenser to thermometer handle. If problem persists, replace probe cover dispenser.
25
Page 30
Welch Allyn 9020 Infrared Tympanic Thermometer
PROBLEM
Continuous, Intermittent Audible Tone
Blank LCD Display, No Temperature on Display
Unable to Remove 9020 from Wall Mount
SOLUTION
Probe is partially extended. Ensure slide is fully retracted.
Improper operating condition. Check display for error status. See “Error Codes” section of this manual for explanation.
Probe extended for over 5 minutes. Retract probe.
Battery power depleted. Replace batteries. See “Maintenance” section of this manual for procedure.
Wall mount’s spring latch mechanism is engaged. Hold release key (underside of wall mount) to the right while removing thermometer.
Release key for spring latch mechanism (underside of wall mount) is missing or locked. Insert key into mechanism, then rotate key clockwise until tab is perpendicular to front panel of wall mount. Hold release key to the right while removing thermometer.
26
Page 31
9020 Users’ Manual

ERROR CODES

When the 9020 detects an improper operating condition, the malfunction symbol (X) will appear with one of the numeric error codes (Er) below.
Set the thermometer aside for one hour to allow the electronics to stabilize to room temperature.
If the 9020 is faulty, the error status will
remain to indicate that the thermometer needs service.
If the condition is transient in nature, the
error status will disappear. The 9020 will then self-calibrate, and be ready for use.
ERROR EXPLANATION
Er2 The ambient temperature is beyond the operating
range of 60°F – 104°F. Allow the thermometer to stabilize at normal room temperature before use.
Er4 The infrared sensing system may be detecting a
nearby temperature source that is greater than 110°F. If not, the thermometer requires service.
Er5 The ambient temperature is above 110°F. Er6 If not, the thermometer requires service.
Er0 Internal error. The thermometer requires service. Er1 Er3 Er7 Er8 Er9
27
Page 32
Welch Allyn 9020 Infrared Tympanic Thermometer
SECURITY CABLE INSTALLATION (OPTION)
For added security, an optional Security Cable may be installed that will connect the thermometer to the wall mount.
MATERIALS NEEDED Security Cable (Part #20796-000)
Flat-Blade and Phillips Screwdrivers
1. Insert screwdriver into the wall mount to release the clip.
2. Insert one end of the security cable through the slot of the clip, then reinsert the clip into the wall mount.
28
Page 33
9020 Users’ Manual
3. Remove the screw from the battery door. Spread the case of the thermometer just wide enough to insert the other end of the security cable. Close the battery door and reinstall the screw.
WALL MOUNT INSTALLATION
The following guidelines are provided to ensure proper placement and installation of your thermometer wall mount.
The top of the wall mount should be located about 60 inches from the floor. You will need at least a 4 inch clearance from the bottom and sides of the wall mount and 12 inches from the top.
Wall Mount Requires Three
Pan or Round Head Screws (Your Choice of #6, #8, or #10)
Wall Surface Fastener Dry wall E-Z Twist in anchor (Molly or toggle)*
Concrete/masonry Expanding anchors Wood Wood screws
*Available from Welch Allyn in quantities of 100, part number 20938-000
29
Page 34
Welch Allyn 9020 Infrared Tympanic Thermometer

SPECIFICATIONS

Thermometer Type:........................................................................... Infrared
Measurement Range:................................ 23.9°C to 42.2°C (75°F to 108°F)
Measurement Calibration Accuracy: ......................... + 0.1°C at 37°C - 39°C
+0.2°F at 98°F - 102°F
Other ranges meet ASTM Specification #E 1112 Table 1
Operating Environment: ............................ 15.6°C to 40.0°C (60°F to 104°F)
70% relative humidity
Storage Environment: ...................................-55°C to 65°C (-67°F to 149°F)
70% relative humidity
Case Dimensions: ........................................... 15.38cm x 15.38cm x 5.12cm
(6in x 6in x 2in)
Weight (including batteries and probe covers):...................... 258 gr (9.1 oz)
Case Material: .............................................................................ABS Plastic
Power Source: ................................................................... AAA batteries (4)
Battery Life: ................................................................. Approximately 1 Year
Display Type: .............................................................. Liquid Crystal Display
30
Page 35
9020 Users’ Manual
LIMITED WARRANTY
WELCH ALLYN LIMITED WARRANTY:NEW UNITS–1 YEAR
REFURBISHED UNITS–1 YEAR
Welch Allyn, Inc. (“Welch Allyn”) warrants that each new Welch Allyn Model 9020 Thermometer (the “Product”) is free from defects in material and workmanship under normal use and service for a period of one year for new units or 1 year for refurbished from the date of shipment from Welch Allyn. This Warranty shall be fulfilled by the repair or replacement by Welch Allyn or its authorized representative, at Welch Allyn’s discretion of any such defect, free of charge for parts and labor.
EXCLUSIONS FROM WARRANTY
This Welch Allyn Warranty shall be void, and Welch Allyn will not be responsible for any loss associated with, any of the following: (i) probe tips damaged or defective after delivery to Customer, (ii) not providing routine maintenance and calibration, (iii) missing, damaged, or discharged batteries after the first year, (iv) any product that (a) has had the serial number defaced, (b) has been repaired by anyone other than Welch Allyn or an authorized Welch Allyn service representative (c) has been altered or (d) has been used in any manner other than in accordance with instructions or shows evidence of abusive handling.
WARRANTY CLAIMS
If a Product requires service during the applicable warranty period, the Customer should immediately return the Product at Customer’s own expense to Welch Allyn, 7420 Carroll Road, San Diego, CA 92121. Any risk of loss or damage in shipment of the Product to Welch Allyn shall be borne exclusively by Customer. Repair or replacement will be made at Welch Allyn’s expense subject to the terms of this Warranty. Product repaired under warranty provisions will be returned to Customer at Welch Allyn's expense. Any damage in shipment from Welch Allyn is the responsibility of the delivery carrier.
LIMITATIONS
In no event shall Welch Allyn be liable for any incidental, indirect, or consequential damages in connection with the purchase or use of any Product.
This Warranty is in lieu of other obligations or liabilities on Welch Allyn’s part, and Welch Allyn neither assumes, nor authorizes, any representative or other person to assume for it any other liability in connection with the sale of the Product.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR OTHER WAR­RANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED, WELCH ALLYN WILL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES.
31
Page 36
70858-0000B
Loading...