The symbol on the patient valve is a reminder
that the lip membrane must be changed immediately if it becomes crinkled, sticky or misshapen. The patient must not be used again until the membrane has
been changed, as this could cause malfunction (see
“8.3 Checking patient hose system” on page 48).
The symbol indicates the correct position for
insertion of the lip membrane.
When connecting the patient valve, take care to ensure that the direction of respiratory gas flow is correct (see arrows).
6Overview
Page 7
EN
MEDUMAT Easy
5678
3
2
1
4
+–
1Inlet 2,7 - 6 bar O
2
2Languages available on this device
MEDUMAT Easy device information plate
SN
Serial number of device
Yea r of m an ufa ct ur e
3.6 V lithium battery
Direct voltage
3
IP54
CE symbol (confirms that the product conforms to the
applicable European directives)
Protection against ingress of water
Protection class BF
Do not dispose of device in domestic waste
Overview7
Page 8
EN
4Follow instructions for use
5Tube sy st em c on ne ct io n
6Maximum pressure 100 mbar
Safety check and servicing label
Safety check label: (in Germany only) marks when the next
7
8Servicing label: indicates when the next service is due.
safety check as per §6 of the German law relating to users of
medical devices is required.
8Overview
Page 9
EN
2. Legend
A: MEDUMAT Easy control panel
1Mask/tube ventilation switch with indicator LEDs
2Ventilation pressure gauge
3Alarm panel
4Alarm mute button
5Colour code
6Regulator knob, ventilation parameters
7Stop notch
8LED Demandflow
9ON/OFF switch
B: MEDUMAT Easy connections
10Pressurized gas connection
11Battery compartment
12Speaker
13Pressure gauge hose connection
14Ventilation hose connection
15Relief outlet valve
C: MEDUMAT Easy device combinations
16Ventilation mask
17Ventilation hose
18Filter
19Patient valve
20PEEP valve
21Tube
Legend9
Page 10
EN
3. Description
3.1 Intended use
MEDUMAT Easy is an automatic oxygen respiration
device (short-term ventilator) with additional inhalation facility.
You can use MEDUMAT Easy:
•to revive patients at the site of the emergency;
•for longer periods in more protracted
emergencies;
•for short-term O
mask.
You can use MEDUMAT Easy while transporting
patients:
•between the various rooms and departments of
a hospital;
•between the hospital and other premises;
•in emergencies;
•when transport over considerable distances is
planned.
MEDUMAT Easy:
•is designed to provide controlled ventilation to
persons of 10 kg body weight or more;
•is used to treat respiratory arrest;
•can be preset to parameters that ensure evenly
balanced ventilation, provided that the selected
maximum ventilation pressure P
exceeded.
inhalation using a respiration
2
is not
max
10Description
Page 11
EN
•permits respiration-controlled oxygen inhalation
16
15
13
14
14
12
12
11
10
10
9
11
16
7
5
30
3
MV (l/min)
Freq.(min
-1
)
Demandflow
10
0
in Demand mode.
3.2 Owner/operator and user qualification
As an owner/operator or user, you must be familiar
with the operation of this medical device. Observe
the legal requirements for operation and use (in Germany, the regulations governing owner/operators of
medical devices apply in particular). Basic recommendation: get a person authorized by WEINMANN
Emergency to provide you with proper instruction
about the handling, use and operation of this medical device.
3.3 Ventilation function
MEDUMAT Easy operates within a pressure range of
2.7 to 6 bar and at a flow rate of not less than
70 l/min O
It uses high-pressure, medicinal-grade oxygen. An
external pressure reducer brings this down to the required operating pressure. The oxygen supply is fed
in at input valve.
The ventilation settings are continuously variable.
These settings (frequency and volume per minute
are coupled) and the inspiration/expiration ratio of
1:1.67 are regulated by internal electronic control
mechanisms.
The gas for inspiration flows along the hose and
through the patient valve and either the mask or
tube into the patient’s airways. The patient valve is
. It has a built-in power pack.
2
Description11
Page 12
EN
fitted with a lip membrane that enables expired gas
9
60
mbar
50
40
30
20
10
0
16
15
13
14
14
12
12
11
10
10
9
11
16
7
5
30
3
MV (l/min)
Freq.(min
-1
)
Demandflow
10
0
Demandflow setting
Respiration hose connection
Expiration
tube
Connection
for pressure
gauge tube
Mask/tube
connection
Spontaneous
breathing tube
to be conducted away through the expiration tube.
You can check the course of ventilation at ventilation
pressure gauge.
3.4 Demandflow function
The Demandflow setting switches the
MEDUMAT Easy breathing-controlled O
Such inhalation must be carried out with the respiration mask. A small inspiration (trigger) pulse causes
oxygen to continue flowing until slight overpressure
interrupts the flow. Expiration then takes place via
the patient valve as in ventilation.
inh alation.
2
3.5 Patient hose system with patient valve
The respiratory gas is routed to the patient via the
patient hose system with patient valve. The patient
hose system with patient valve is designed so that
even if the ventilator fails, spontaneous breathing is
possible, regardless of which ventilation mode you
have selected.
12Description
Page 13
EN
3.6 Audio response
The device has an audio response facility that can be
switched on for user guidance, especially for users
who have little practice.
If audio response is not required, a key combination
can be used to switch it off (see “6.10 Audio response for user guidance” on page 32).
Description13
Page 14
EN
4. Safety instructions
4.1 Safety regulations
For your own safety, the safety of your patients, and
to comply with the requirements of EU Directive 93/
42/EEC, please observe the following points:
General
•Please read the directions for use carefully. They
are an integral part of the ventilator and must be
kept available for reference at all times.
•Use the MEDUMAT Easy for the described purpose only (see “3.1 Intended use” on page 10).
•The user must subject the patient hose system to
a functional check and a visual inspection before
use (see “8.3 Checking patient hose system” on
page 48).
14Safety instructions
Note:•Do not use the MEDUMAT Easy in toxic environ-
ments or where there is a risk of explosion.
•MEDUMAT Easy is not suitable for hyperbaric
use (pressure chamber).
•Do not use MEDUMAT Easy with flammable
anesthetics.
•A back-up ventilator should always be available
in case of technical failure.
•Before starting to work with MEDUMAT Easy,
you must understand how to operate it.
•To prevent infection or bacterial contamination,
please observe section “7. Hygienic preparation” on page 40.
Page 15
EN
•MEDUMAT Easy should be used only by medically qualified personnel who have had training
in ventilation techniques. Incorrect use can cause
severe physical injury.
•Please note that a safe distance must be maintained between MEDUMAT Easy and equipment
that emits HF radiation (e.g. mobile phones),
otherwise malfunctions may occur (see “ Recommended safe distances between portable or
mobile HF telecommunication devices (e.g. mobile phones) and MEDUMAT Easy” on page 64).
•We recommend that maintenance work such as
inspections and repairs be performed only by the
manufacturer, WEINMANN Emergency, or by
qualified technicians expressly authorized by
WEINMANN Emergency.
•Malfunctions and a lack of biocompatibility may
result if third-party articles are used. Please note
that in such cases all warranty entitlement and liability claims shall be void where items other
than the accessories recommended in the instructions for use or original spare parts are
used.
•Design changes to the device are not permitted
and may put patients and users at risk.
Oxygen
Spontaneous explosive reactions can occur if highpressure oxygen comes into contact with flammable
substances (fat, oil, alcohol etc.):
•Keep the equipment and all screw connections
absolutely free from oil and grease.
•Always wash your hands before starting to work
on the oxygen supply.
Safety instructions15
Page 16
EN
•Smoking and open flames are strictly prohibited
in the vicinity of all fittings containing or transporting oxygen.
•During assembly and when cha ngin g the oxyge n
cylinder, only hand pressure should be used
when tightening the screw connections to the
cylinder and to the pressure reducer. Never use
tools for this purpose. Excessive tightening damages the screw threads and seals, and can cause
leaks.
•Protect oxygen cylinders from accidental falls. If
a cylinder falls over, the pressure reducer or the
valve may break off and cause a violent explosion.
Important•Always open the valve of the oxygen cylinder
slowly to prevent pressure damage to the other
fittings.
•The oxygen cylinder should never be completely
emptied, as this may allow air containing moisture to enter the cylinder and cause corrosion.
Ventilation/Operation
16Safety instructions
•Both patient and ventilator must be kept under
constant observation during ventilation.
•When connecting the patient valve, please ensure that the direction of respiratory gas flow
(> patient >) is correct. Make sure that neither
the expiration tube nor the spontaneous breathing tube on the patient valve are blocked or their
function impeded in any other way, e.g. by the
patient’s position.
Note: •Disposable hose system WM 28110 is intended
to be used only once.
Page 17
EN
Software
•Extensive validation tests have been performed
to minimize risks arising from software errors.
Accessories
•Please protect the silicone and rubber components from UV radiation and prolonged exposure to direct sunlight, as this can make them
brittle and friable.
Safety instructions17
Page 18
EN
5. Installation
A permanent mounting is usually necessary only
when MEDUMAT Easy is installed as a fixture in rescue vehicles, helicopters or aircraft.
If MEDUMAT Easy is supplied complete on a portable system or in an emergency rucksack, it is ready
for use and requires no further installation. Separate
directions for use are supplied for portable systems
and emergency rucksacks.
In order to ensure safe and reliable operation, functional tests must be carried out after installation (see
“8. Functional checks” on page 45).
5.1 Connecting the oxygen cylinder
Always wash your hands thoroughly before starting
any work on the oxygen supply. Products containing
hydrocarbons (e.g. oils, greases, alcohols, hand
creams, sticking plasters) may cause explosive reactions if they come into contact with high-pressure oxygen.
18Installation
Never use wrenches or similar tools to tighten or
loosen the screw connections.
Removing the empty cylinder
1. Close the valve of the oxygen cylinder.
Switch on MEDUMAT Easy with ON/OFF switch.
This exhausts any residual oxygen and depressurizes the ventilator. Wait until the pressure
gauge on the pressure reducer shows an oxygen
content of zero before undoing the screw connection by hand.
Page 19
EN
2. First switch off MEDUMAT Easy again.
3. Then loosen the screw connection to the
cylinder.
Connecting the new cylinder
1. First briefly open and close the valve of the new
oxygen cylinder. This should blow out any particulate matter.
Keep the valve opening away from the body,
making sure that neither yourself nor other persons can be injured by escaping particles.
2. Next use the fluted connecting nut to couple the
pressure reducer to the valve on the oxygen cylinder. Tighten the connecting nut by hand.
3. If the pressure hose is not already connected to
the exit from the pressure reducer, make this
connection with the G 3/8 connecting nut.
4. Screw the other end of the pressure hose on to
pressure gas connection on the MEDUMAT Easy
if this has not yet been done.
5.2 Ventilation hose
1. Slide the pressure gauge tube onto connection.
2. Slide the ventilation hose onto connection.
Make sure that this does not cause any kinks in
the pressure gauge tube already connected. If
necessary, turn the ventilation hose while sliding
on as appropriate.
Installation19
Page 20
EN
Always grasp the ventilation hose and
P
o
s
i
ti
on
>
P
S
U
< 13
4
°
pressure gauge tube by their end only,
otherwise they may be damaged or split.
3. Connect the patient valve to the other end of the
ventilation hose and pressure gauge tube.
4. If a mask is being used for ventilation, attach the
mask connection to the patient valve (identical
with tube connection),
or
if the patient is intubated, attach the patient
valve to the tube.
20Installation
Filter
If a filter is to be used, this should be installed
between the patient side connector on the patient
valve and the mask or tube.
Always follow the instructions supplied by the
manufacturer of the filter.
Page 21
EN
Note:Please note that the respiratory resistance of the
entire system is increased when an HME filter or bacteria filter is used and under certain circumstances,
may exceed the value permitted by EN 794-3.
PEEP valve
If a PEEP valve is to be used, this should be inserted
into the expiration tube on the patient valve.
To adjust the PEEP valve, always follow the instructions supplied by its manufacturer.
5.3 Wall mounting set
A wall mounting set (see “11.2 Accessories” on
page 60) is available for permanent fixing, e.g. on a
vertical surface inside a vehicle.
Please refer to the sheet enclosed with the wall
mounting set for details of dimensions and installation procedure.
Installation21
Page 22
EN
6. Using the ventilator
6.1 Switching on / self test
1. Open the valve of the oxygen cylinder slowly.
The pressure gauge will now show the pressure
in the cylinder.
2. Where appropriate, calculate how long the remaining oxygen will last (see “6.11 Calculation
of oxygen content/remaining operating time”
on page 37). Always change the cylinder in good
time, e.g. when the pressure is lower than
50 bar, to ensure that oxygen is available for an
adequate period.
3. Select the desired ventilation settings (see “6.2
Selecting the ventilation settings” on page 23).
4. Switch on the MEDUMAT Easy with ON/OFF
switch. The ventilator will then run a self test
lasting approx. 2 seconds.
If audio response is enabled, you will hear before
the self test the sentence “Open oxygen cylinder”.
22Using the ventilator
Page 23
EN
During this test, the four LEDs in alarm panel 3
flash on and off and a short alarm tone sounds.
Stenosis
Disconnection
< 2,7 bar O
2
If an error is found, all the LEDs in alarm panel
continue flashing and an alarm sounds. If this happens, MEDUMAT Easy must not be used for ventilation.
If audio response is enabled, you will hear the
message “Device malfunction! Administer alternative ventilation”.
After the self test, the ventilator repeatedly
checks the oxygen cylinder pressure until adequate pressure is detected. Otherwise an alarm is
sounded.
The MEDUMAT Easy will then start to function
with the selected ventilation settings.
6.2 Selecting the ventilation settings
We recommend selection before switching on, to
prevent unnecessary waste of oxygen.
3
Respiratory frequency and minute
volume
12
11
9
10
14
12
10
15
13
7
11
14
30
3
Freq.(min-1)
MV (l/min)
16
Demandflow
16
5
1314
1,1
0,6
0,3
1. Set the respiratory frequency and the minute
volume with regulator knob.
Recommended ventilation settings:
yellow orangebrown
Body weight 10 kg30 kg60 kg80 kg110 kg
Respiratory
frequency
Minute
volume
-1
30 min
16 min-111 min-110 min-110 min
3 l/min 5 l/min7 l/min 9 l/min11 l/min
Using the ventilator23
-1
Page 24
EN
The figures shown in the table are only recommendations. Different settings may be required
in cases of pulmonary damage or for special indications.
To see the relationship between the values, see
diagram “12.2 Relationship between ventilation
parameters” on page 66.
Maximum ventilation pressure
1. Use the mask/tube switch to set the maximum
ventilation pressure. The LEDs light up in active
mode.
Recommended maximum ventilation pressure:
Device
Audio response up
to serial number
7999
From serial number
8000 and following
repair/servicing/
firmware update
If the maximum level is reached, e.g. in cases where
compliance is inadequate, MEDUMAT Easy sets off a
stenosis alarm (see “ Stenosis alarm” on page 30).
Note:The mask/tube switch can only be operated when
the ventilator is switched on.
Page 25
EN
6.3 Performing ventilation
P
o
s
i
ti
on
>
P
S
U
< 13
4
°
9
60
mbar
50
40
30
20
10
0
Ventilation mask
1. Attach the mask to the patient valve.
2. If necessary, insert a Guedel tube to keep the
airways open before putting on the mask.
3. Place the mask over the patient's mouth and
nose.
4. Tilt the head back and use the E-C technique
ensure the mask is hermetically sealed.
Intubation
The patient will normally be intubated before the
tube is connected to the patient valve.
1. Attach the patient valve to the connector of the
tracheal tube.
2. Monitor the respiratory parameters during ventilation. This will indicate whether the tube is correctly positioned and ventilation is adequate.
1)
to
6.4 Monitoring ventilation
The patient must be monitored constantly during
ventilation.
You can read off ventilation pressure on the ventilation pressure gauge.
High airway resistance, as a result of obstructions or
external cardiac massage for example, will change
1. See also: Resuscitation, Volume 46 (2000), Special Issue International Guidelines 2000 for CPR and ECC -
A Consensus on Science; p. 115-125.
Using the ventilator25
Page 26
EN
the tidal volume set. Use suitable volume measuring
Example of ventilation sequence before and after
lung compliance diminishes
devices to check the tidal volume actually administered.
Check respiratory parameters during ventilation.
If lung compliance diminishes during ventilation, the
ventilator will react by increasing the ventilation
pressure at constant volume.
6.5 Ventilation with PEEP valve
A PEEP valve can be fitted to the expiration tube on
the patient valve with an adapter.
6.6 Ventilation with filter
26Using the ventilator
This valve makes it possible to use positive end-expiratory pressure (PEEP).
Please see the PEEP valve instructions for details of
settings.
A conventional filter with standard 15/
22 connectors can be fitted on the inspiration tube
of the patient valve for hygiene purposes and to condition the air for inspiration. This will increase both
Page 27
EN
1314
0,3
0,6
1,1
16
15
13
14
14
12
12
11
10
10
9
11
16
7
5
30
3
MV (l/min)
Freq.(min
-1
)
Demandflow
6.7 Demandflow
inspiratory and expiratory resistance. You should
therefore monitor ventilation pressure and ventilation volume very carefully.
A close watch must be kept for any increase in dead
space, especially in children.
Always be sure to follow the instructions supplied by
the manufacturer of the filter.
Note:A PEEP valve must not be fitted when using the
ventilator in Demandflow mode.
Demandflow must be switched on for O
inhalation.
2
To switch the MEDUMAT Easy to Demandflow
mode, turn the ventilation settings regulator knob
until it engages in the fixed point marked by the
white triangle. The green LED indicates that it is
ready for operation. If audio response is enabled, the
ventilator announces “Demandflow mode”.
Attach the mask to the patient valve and place it over
the patient’s mouth and nose. Hold the mask to ensure a firm seal. The flow is switched on by the patient breathing in (triggering the device). When the
patient starts to breathe out, the flow stops and the
expired air is removed via the patient valve. The patient should breathe calmly and evenly. The Demandflow cannot be altered. At higher breath rates, fresh
air is automatically mixed in with the oxygen. This is
done via the spontaneous breathing tube of the
patient valve.
The Demandflow mode is ended by turning the regulator knob back to ventilation mode from the index
Using the ventilator27
Page 28
EN
position marked by the white triangle, or by switching off the ventilator.
If audio response is activated, the ventilator confirms
the return to ventilation mode by announcing:
“Mask ventilation mode. Tilt bac k head , sec ure m ask
tightly” (from serial number 8000 and following repair/servicing/firmware update: “Ventilation pressure limit 20 mbar”).
6.8 Terminating ventilation or Demandflow
Important
Never empty the oxygen
cylinder completely. Return the cylinder for filling
while it still contains residual pressure. This prevents
entry of moist atmospheric
air that can cause
corrosion.
1. Check the oxygen supply on the pressure reducer gauge. Always change the cylinder in good
time, e.g. when the pressure is lower than
50 bar, to ensure that oxygen is available for an
adequate period.
2. Close the valve of the oxygen cylinder.
3. Switch off the MEDUMAT Easy. To prevent the
ventilator being switched off unintentionally, ON/
OFF switch must be kept pressed down for at least
2 seconds until the LEDs in the alarm panel
3 light
up. If audio response is enabled, the ventilator announces: “Close oxygen cylinder”.
28Using the ventilator
Page 29
EN
6.9 Alarm signals
Stenosis
Disconnection
< 2,7 bar O
2
+–
Alarm panel signals the following alarms:
Stenosis:Stenosis, or maximum ventilation
pressure P
reached in two succes-
max
sive inspiration phases
Disconnection:Disconnection between
MEDUMAT Easy and patient in two
successive inspiration phases
< 2.7 bar:Drop in oxygen pressure to below
2.7 bar
:Battery charge inadequate
All the visual alarms are accompanied by an acoustic
alarm.
If the ventilator detects a malfunction during the self
test after switching on or during continuous operation, all LEDs in the alarm panel stay flashing on and
off, and an alarm tone sounds. If audio response is
enabled, you will hear the message “Device malfunction! Administer alternative ventilation”.
In this case you must not use the MEDUMAT Easy.
The failure alarm can be confirmed by pressing the
ON/OFF switch.
The patient valve is designed to enable spontaneous
breathing in the event of equipment failure.
When is the alarm set off?
An alarm signal is given as soon as any one of the
functional problems mentioned above occurs. The
relevant LED starts flashing and an acoustic signal
sounds. If audio response is enabled, the user also
hears additional information about the individual
alarm.
Using the ventilator29
Page 30
EN
Simultaneous disconnection and drop in oxygen
pressure will initially set off only the < 2.7 bar alarm.
Stenosis alarm
Actual ventilation pressure exceeds the maximum
ventilation pressure (20 or 45 mbar).
MEDUMAT Easy briefly switches to expiration if the
maximum ventilation pressure is exceeded, but then
tries to continue inspiration in the same inspiration
phase.
If the maximum ventilation pressure is exceeded for
a second time during the same inspiration phase, the
unit finally switches to expiration and vents the patient tube system completely. The next inspiration
begins with the following ventilation stroke according to the frequency selected. This does not affect
the set frequency.
The alarm is set off if airway resistance is exceeded
in
two successive inspira tion phases. This is intended
to prevent false alarms, e.g. due to coughing.
If audio response is enabled, the unit announces
“Check airways and minute volume” (from serial
number 8000 and following repair/servicing/firmware update: “Check airways and settings”).
30Using the ventilator
Disconnection alarm
As a rule this alarm is due to interruption of the
breathing system.
The alarm is set off when the rise in pressure fails to
reach at least 8 mbar in
phases.
If audio response is enabled, the unit announces
“Check ventilation system and settings”.
two successive inspiration
Page 31
EN
Disconnection alarm in Demandflow
+–
mode
If the patient does not trigger MEDUMAT Easy within 15 seconds, the “Disconnection” alarm is given. If
audio response is enabled, the unit announces “Rule
out respiratory arrest and check mask fit”.
< 2.7 bar O2 alarm
Oxygen pressure at the pressure connection to the
MEDUMAT Easy has dropped to less than 2.7 bar. The
reason is usually an almost empty oxygen cylinder.
In this case MEDUMAT Easy can no longer function
correctly because the operating parameters are no
longer within the permissible limits.
If audio response is enabled, the ventilator announces:
“Check pressure hose system and gas supply”.
alarm
The battery is failing. Failure of the automatic ventilation function must be expected. Immediate steps
must therefore be taken to provide alternative ventilation (see “6.12 Alternative ventilation procedures” on page 39).
If audio response is enabled, you will hear the message: “Device malfunction! Administer alternative
ventilation”.
The ventilator must be switched off before the battery can be changed (see “ Changing the main battery” on page 57).
Cancelling acoustic alarm
If there is an alarm, you can temporarily cancel the
acoustic alarm by pressing the alarm mute button:
Using the ventilator31
Page 32
EN
Stenosis:30 seconds
+–
Disconnection:30 seconds
< 2.7 bar:30 seconds
:120 seconds
Even if audio response is enabled, no messages will
be output for the periods stated. The visual alarm
will continue to flash.
If the cause of the alarm is not el iminat ed, the acoustic alarm will start to sound again after a short interval. Audio response will also resume automatically.
Both the visual and acoustic alarms are cancelled automatically as soon as the malfunction is eliminated.
6.10 Audio response for user guidance
Selecting language / Switching off audio
guidance
The language setting can only be selected if the unit
is switched off.
To select a language or to switch off the audio response facility, proceed as follows:
1. Hold down the mask/tube switch. Switch on the
ventilator at the ON/OFF switch.
2. Then release the mask/tube switch. The unit is
now in the language selection menu. The ventilation pressure gauge displays the most recent
language setting. The following languages are
assigned to the individual LEDs:
3. Press mask/tube switch as many times as necessary until the LED for the desired language lights
up and a corresponding message is heard (example: LED 10 mbar, language: English, message:
“Selected language: English”). After five
seconds the new selection is stored.
Tip:
By pressing the ON/OFF switch briefly you can
store the language selection without waiting for
five seconds.
As there are more languages than LEDs on the
ventilation pressure gauge, a new cycle starts on
Level 2 once the 60 mbar LED has been reached.
Level 2 is displayed by the stenosis and disconnection alarm LEDs. Once the last language on
Level 2 has been reached, the cycle starts again
at 0 mbar on Level 1 and the stenosis and disconnection alarm LEDs go out.
Select the setting 0 (0 mbar) if you want to
switch off the audio response facility. You will
then hear the message: “Audio response is off!”
in the language most recently selected.
34Using the ventilator
After approx. 5 seconds the new setting is automatically stored. The LED for the selected language goes out.
Audio response messages
The following is a list of the individual audio response messages with notes on what they mean.
Note the differences between the audio responses.
The audio responses change from serial number
8000 and following repair/servicing/firmware update (see Table 2).
“Audio response is off”Confirmation that audio response is deactivated.
Depending on whether or not the patient is breathing, set
MEDUMAT Easy to one of the modes Demandflow (Page 27), mask
ventilation or intubation (Page 24).
Depending on patient weight, set respiration frequency and minute
volume (Page 23).
Connect patient to ventilator via ventilation hose and patient valve
using the patient mask or the connector of the tracheal tube.
Mask ventilation mode is selected.
While tilting the head back, use the E-C technique to seal the mask.
Tube v en ti lati on m ode is selected.
Maximum ventilation pressure for tube ventilation.
MEDUMAT Easy has measured excessive airway resistance. Check the
airways or adjust respiratory frequency and minute volume settings to
suit the patient (Page 23).
The device is faulty or the battery is failing. The device can no longer be
used for ventilation. You must therefore use another ventilation method
(Page 39).
MEDUMAT Easy has measured low pressure on the inlet side. Check
whether the O
oxygen hose is not leaking, kinked or jammed.
MEDUMAT Easy can no longer detect a breathing pulse (trigger) in
Demandflow mode. Check the patient’s breathing, and if necessary
switch to a different ventilation mode. Check the connections and
mask fit.
After switching off the ventilator, turn off the O2 cylinder or the external
O
supply.
2
Disconnection: a pressure rise of 8 mbar is not achieved during the
inspiration phase under controlled ventilation. This is usually due to an
interruption of the ventilation system or to a low minute volume setting.
Check the connections or adjust the minute volume to suit the patient.
When selecting the language for the audio response, press the mask/
tube switch as many times as necessary until the desired language is
announced.
cylinder still contains sufficient oxygen and that the
Breathing volume (BV) = inhalation flow x inhalation time
For the above example:
Breathing volume = 45 l/min x 0.033 min= 1,5 l
Minute volume (MV) = respiratory frequency x BV
For the above example:
Minute volume (MV) = 10 min-1 x 1.5 l = 15 l/min
Real Demandflow time (min) =
oxygen content (l)
MV (l/min)
Real Demandflow time =
2000 l
15 l/min
= 133 min = 2 h 13 min
Example:
content = 2000 l, MV = 15 l/min.
O
2
This gives the following equation:
38Using the ventilator
Page 39
EN
6.12 Alternative ventilation procedures
If MEDUMAT Easy ceases to function during a ventilation procedure, the following alternatives can be
applied:
Ventilation bags
1. Remove the patient valve from the tube or the
mask.
2. Replace it with the ventilation bag, e.g. a
WEINMANN Emergency COMBIBAG
WM 11000, and perform manual ventilation.
Exhaustion of oxygen supply
In emergency situations when the oxygen supply
runs out, MEDUMAT Easy can also be operated with
respiratory air.
Using the ventilator39
Page 40
EN
7. Hygienic preparation
After every use the MEDUMAT Easy and any acces-
sories used must undergo hygienic preparation.
•Be sure to carry out a functional check after every hygienic preparation (see “8. Functional
checks” on page 45).
•
This product may contain disposable items. Dis-
posable items are intended to be used only once.
So use these items only once and do
cess them. Reprocessing disposable items may
impair the functionality and safety of the product and lead to unforeseeable reactions as a result of ageing, embrittlement, wear, thermal
load, the effects of chemical processes, etc.
7.1 MEDUMAT Easy
You can keep MEDUMAT Easy clean by simply wiping with disinfectant as described in section 7.6 .
Never immerse MEDUMAT Easy in disinfectants or
other liquids. Otherwise damage may be caused to
the unit, thus endangering users and patients.
not repro-
7.2 Patient valve
40Hygienic preparation
1. Disconnect the patient valve from the hoses.
Always grasp the hoses by their ends. Otherwise you
might damage or tear them.
Page 41
EN
2. Dismantle the patient valve as shown in the ad-
Lip
membrane
Disc membrane
Expiration tube
Disc membrane
Spontaneous
breathing tube
jacent diagram. It is neither necessary nor permissible to remove the membrane in the
spontaneous breathing tube for cleaning and
disinfection.
3. Crinkled, misshapen and sticky lip and valve
membranes must be replaced.
4. Perform the hygienic preparation as described in
section 7.6.
5. Reassemble the patient valve. When reassembling, make sure that the lip membrane is correctly positioned.
6. Always perform a functional check before using
the valve again (see “8.3 Checking patient hose
system” on page 48).
Hygienic preparation41
Page 42
EN
7.3 Ventilation hose
Caution!Only reusable hose system WM 22520 (scope of
supply) is suitable for the hygienic preparation described here. Do
WM 28110 available as an accessory to hygiene
preparation. Replace it with a new one.
1. Take the ventilation hose and the pressure
2. Perform the hygienic preparation as described in
3. For reassembling, see “5.2 Ventilation hose” on
7.4 Masks
not subject disposable hose system
gauge hose off both connection ports.
Caution! Always grasp the hoses at the end, as
shown in the drawing, otherwise the hoses may
be damaged or torn off. Close both ends of the
pressure gauge hose.
section 7.6.
page 19.
7.5 Fittings
42Hygienic preparation
Perform the hygienic preparation of the masks as
described in section 7.6.
Ensure that no liquids get into the connections for
the oxygen fittings, otherwise there is a risk of explosion, especially if you are using products for disinfecting by wiping which contain alcohol.
For external cleaning of fittings (e.g. pressure reducer, valve), use only a clean cloth. The cloth may be dry
or moistened with clean water.
Page 43
EN
Never immerse the fittings in disinfectants or other
liquids. You may only disinfect them by wiping. On
no account may liquid get into the pressure reducer,
as this could cause explosions.
7.6 Cleaning and disinfection procedure
Hygienic preparation of the MEDUMAT Easy and the
accessories used should be performed as described
in the following table.
Observe the instructions for the disinfectant used.
We recommend gigasept
disinfection and terralin
tion. You are recommended to wear suitable gloves
(e.g. household or disposable gloves) during disinfection procedures.
®
FF (new) for immersion
®
protect for wipe disinfec-
Description of
component
MEDUMAT Easy
Patient valve
Silicone ventilation
mask
Venti lation hose
Hose casing, reusable
Oxygen fittings
CleaningDisinfection
With a dry or
damp cloth
in warm water
with a mild
household
detergent
Wipe with a dry
or damp cloth
With a dry or
damp cloth
WipingNot permissible
Immerse in dilute
solution so that all
surfaces, inside and out,
are thoroughly wetted
without bubbles. Wait
until the full exposure
time has elapsed. After
disinfection, rinse all
parts thoroughly inside
and out with distilled
water and leave to dry.
Rinse cycle 30°C,
no spin
WipingNot permissible
(1) To disinfect the pressure gauge tube of the
ventilation hose, proceed as follows:
Thermal
disinfector
Rinsing
programme up to
95 °C (thermal
disinfection in
automatic
cleansing unit)
(1)
Possible during
cycle
Sterilization
Not
permissible
Steam
sterilization at
134 °C in
devices to
EN 285,
residence time
at least 5 - 18
minutes.
Not permitted
Not
permissible
Hygienic preparation43
Page 44
EN
1. Connect one end of the pressure gauge tube to
a sterile disposable 20-ml syringe.
2. Immerse the other end in the dilute disinfectant
solution (for gigasept
®
FF (new): residence time
15 minutes).
3. Draw the disinfectant solution through the pressure gauge tube into the syringe until the latter
is full. Do not flush through the pressure gauge
tube in the opposite direction!
4. Detach the syringe from the pressure gauge
tube and empty it out completely.
5. Repeat the procedure 5 more times.
6. After disinfection, the pressure gauge tube must
be rinsed with distilled water at least 8 times using the same principle.
You can support the subsequent drying process
with medical compressed air or medical oxygen.
Then allow the components to dry thoroughly. If any
water is left in the patient valve or the pressure
gauge tube of the ventilation hose, the unit may not
function correctly.
44Hygienic preparation
Page 45
EN
8. Functional checks
Before each use, after each dismantling and reassembly, and at the very least every 6 months, the user must
carry out functional checks on the ventilator.
Note:Before carrying out the functional check on
MEDUMAT Easy, you must connect the ventilation
hose and the patient valve.
MEDUMAT Easy must not be used if the functional
checks reveal defects or deviations from the specified parameters.
First try to correct the error with the help of the information provided in section “9. Troubleshooting”
on page 54. If this is not possible, have the unit repaired by the manufacturer – WEINMANN
Emergency – or by specialists explicitly authorized by
WEINMANN Emergency.
A complete functional check includes:
•“8.2 Checking for leaks in the system” on
page 47;
•“8.3 Checking patient hose system” on
page 48;
•“8.4 Checking the minute volume” on page 49;
•“ Checking the breath volume” on page 49;
•“8.5 Checking maximum ventilation pressure”
on page 50;
•“8.6 Checking Demandflow” on page 51;
•“8.7 Checking the alarm systems” on page 52;
We recommend that you always hold reserve stocks
of the following items:
•replacement washers for the connections;
Functional checks45
Page 46
EN
8.1 Intervals
•lip membranes for the patient valve.
Note:Ensure that the test bag is not damaged and test its
function regularly, e.g. in the context of device
servicing.
Before each use:
•Perform a functional check.
After each use or dismantling:
•Clean, disinfect or sterilize the ventilator and its
components (see “7. Hygienic preparation” on
page 40);
•Check the lip membrane in the patient valve (see
“8.3 Checking patient hose system” on
page 48). It must not be crinkled, sticky or misshapen.
•Perform a functional check.
46Functional checks
At least every 6 months, if the ventilator
has not been used in the meantime:
•Perform a functional check.
Page 47
EN
8.2 Checking for leaks in the system
1. Open the valve of the oxygen cylinder slowly.
You can now read the pressure in the cylinder
from the gauge on the pressure reducer. For example, a reading of 200 bar means that the cylinder is full, whereas 100 bar means it is half full.
Always change the cylinder in good time, e.g.
when the pressure is lower than 50 bar, to ensure that oxygen is available for an adequate period.
2. Close the cylinder valve again.
3. Watch the needle of the gauge on the pressure
reducer for approx. 1 minute. If it stays in the
same place, the system is free of leaks. If the
needle drops steadily, there is a leak somewhere.
Repairing leaks
Always keep a stock of
washers for the connections available.
Important
The screw connections on
the oxygen lines must be
tightened by hand only.
1. Prepare a soap/water solution using non-perfumed soap.
2. Wet all the screw and hose connections with the
solution. Bubbles will form at the site of the leak.
3. Depressurize the system:
To do this, first close the oxygen cylinder. Switch
on MEDUMAT Easy briefly until the pressure
gauge on the O
cylinder reads “0”. Then switch
2
off MEDUMAT Easy again.
4. If leaks are discovered, the defective components must be changed.
5. After changing, make a fresh check for leaks.
6. If it proves impossible to eliminate the leak, the
ventilator will have to be repaired.
Functional checks47
Page 48
EN
8.3 Checking patient hose system
Checking reusable hose system
1. Dismantle the patient valve.
2. Carry out a visual check of all components for
cracks or other physical damage.
The lip membrane must be replaced if it is crinkled, sticky or misshapen. It must no longer be
used for ventilation as it could cause serious
functional problems.
Also perform a visual check of the valve membranes in the expiration and spontaneous
breathing arms. To do so, there is no need to dismantle the valve membranes. Crinkled, misshapen or sticky valve membranes must be replaced,
however, as they can lead to considerable malfunctions.
3. Reassemble the patient valve.
When reassembling, make sure that the lip
membrane is correctly positioned.
48Functional checks
Checking disposable hose system
Visual inspection
Check the following items by inspecting the patient
hose system:
•the patient valve and the connectors may not
exhibit any external damage, cracks or dirt.
Page 49
EN
•the hose connections must be located firmly and
securely on the connecting pieces.
•the patient valve and emergency air membranes
may not exhibit any damage or deformation.
8.4 Checking the minute volume
Checking ventilation frequency
1. Open the valve of the oxygen cylinder slowly.
2. Switch on MEDUMAT Easy.
3. Select the following settings:
–Frequency: 30 min
–Mask/tube switch:
4. Count the number of inspiration phases over a
period of exactly one minute. The figure should
lie between 28 and 32.
5. Turn up the frequency to 14 min
before the index point).
6. Count the number of inspiration phases over a
period of exactly one minute. The figure should
lie between 12 and 16.
-1
(left limit)
(P
: 45 mbar)
max
-1
(right limit
Inspiration stroke=
MV/frequency = 10/10 = 1
Checking the breath volume
1. MEDUMAT Easy must be switched off and the
oxygen cylinder must be open.
2. Attach the test bag to the patient valve with the
adapter from test set WM15323.
3. Select the following settings:
–MV: 10 l/min (between 9 and 11) /
Frequency: 10 min
–P
: (45 mbar)
max
4. Switch on MEDUMAT Easy. The test bag must
become fully inflated during inspiration. This
-1
Functional checks49
Page 50
EN
ensures a breath volume of 1 lit re per inspi ratio n
stroke. At all events the test bag is not sufficiently inflated if a disconnection alarm is set off.
Note:During the expiration phase, you must simulate
the expiration stroke of the test bag by hand. To
do so, place the test bag on a firm base. During
the expiration phase, press the test bag with the
flat of your hand until the volume is completely expelled through the patient valve.
5. Switch off MEDUMAT Easy again.
6. Detach the test bag from the patient valve.
7. Select the following settings:
–MV 3 L/min / frequency 30 min
–P
: (45 mbar)
max
-1
8. Switch on MEDUMAT Easy and close the patient
connection at the patient valve. A stenosis alarm
should be set off.
9. Switch off MEDUMAT Easy again.
8.5 Checking maximum ventilation pressure
Important
Be sure to use the test bag.
If you hold the tube connection closed by hand, the
needle swings across and
correct reading of the precise figure is impossible.
50Functional checks
1. MEDUMAT Easy must be switched off and the
oxygen cylinder must be open.
2. Attach the test bag to the patient valve with the
adapter from test set WM15323.
3. Select the following settings:
–MV: 7 l/min / frequency: 11 min
–P
: (20 mbar)
max
-1
4. Switch on MEDUMAT Easy.
Check that the needle of the pressure gauge on
the MEDUMAT Easy is standing at “0”.
During this test you must not assist the expiration stroke. This ensures that the pressure builds
up slowly. Between 15 and 25 mbar
Page 51
EN
MEDUMAT Easy must set off the stenosis alar m.
Usually this occurs after the second inspiration
stroke.
5. Switch the mask/tube switch to .
6. Repeat the check for intubation ventilation with
the settings:
–MV: 9 l/min / frequency: 10 min
–P
: (45 mbar)
max
-1
If audio response is enabled, the ventilator must
announce “Tube ventilation mode, ventilation
pressure limit 45 mbar” (from serial number
8000 and following repair/servicing/ firmware
update: “Ventilation pressure limit 45 mbar”).
During this test you must not assist the expiration stroke. This ensures that the pressure builds
up slowly. Between 40 and 50 mbar
MEDUMAT Easy must set off the
stenosis alarm.
Usually this occurs after the second inspiration
stroke.
7. Switch off MEDUMAT Easy again.
8.6 Checking Demandflow
1. Make sure that MEDUMAT Easy is switched off
and the oxygen cylinder is open.
2. Attach the test bag to the patient valve with the
adapter from test set WM 15323.
3. Select the setting “Demandflow”.
4. Switch on MEDUMAT Easy. The green “Demandflow” LED lights up.
If audio response is enabled, the ventilator announces “Demandflow mode”.
5. Simulate an inspiration pulse by hand by firmly
squeezing and quickly releasing the test bag.
Functional checks51
Page 52
EN
6. The MEDUMAT Easy switches the flow on and
then immediately off again. This test can be repeated several times.
7. Switch off MEDUMAT Easy again.
8.7 Checking the alarm systems
Important
In the case of the stenosis
alarm and the disconnection alarm, the alarm signal
(or message) is only set off
when the cause of the
alarm is repeated in two
successive inspiration
phases. This prevents the
alarm being triggered by a
very short-lived dysfunction.
Important
In this test the rise in pressure is so strong that the
pressure gauge needle may
overswing into the red
zone. There are technical
reasons for this, and it does
not indicate any malfunction.
Stenosis
1. Open the oxygen cylinder.
2. Remove the tube or the ventilation mask from
the patient valve.
3. Switch on MEDUMAT Easy.
4. Switch the mask/tube switch to .
5. Keep the ventilation connector on the patient
valve closed with the flat of your hand during
two successive inspiration phases. The
stenosis
alarm should be set off.
If audio response is enabled, the ventilator announces “Check airways and minute volume”
(from serial number 8000 and following repair/
servicing/ firmware update: “Check airways and
settings”).
Disconnection (interruption of breathing system)
52Functional checks
1. First proceed as for the stenosis alarm.
2. Then remove your hand. The
stenosis alarm
should now cease (LED goes out, acoustic alarm
stops sounding).
Page 53
EN
After two successive inspiration phases the dis-
+–
connection
alarm should be set off. If audio response is enabled, the ventilator announces
“Check ventilation system and settings”.
Drop in O2 pressure (<2.7 bar O2)
1. Open the valve of the oxygen cylinder slowly.
2. Switch on MEDUMAT Easy.
3. Close the valve on the oxygen cylinder. When
the oxygen pressure in the system has fallen below 2.7 bar, the
<2.7 bar O
alarm should be set
2
off.
If audio response is enabled, the ventilator announces “Check pressure hose system and gas
supply”.
Power supply ()
The alarm that indicates a failing battery is checked
automatically in the self test that runs when
MEDUMAT Easy is switched on.
The power supply is in order if no alarm is set off
when the valve on the oxygen cylinder is opened and
MEDUMAT Easy is switched on and starts to
function.
Functional checks53
Page 54
EN
9. Troubleshooting
+–
FaultCauseRemedy
MEDUMAT Easy defective.Arrange for repair.
Replace battery in battery compartment
MEDUMAT Easy will not
switch on.
Stenosis alarm
(excessive airway
resistance)
Disconnection alarm
(breathing system
interrupted).
< 2.7 bar alarm (oxygen
pressure too low).
alarm.Battery failing or fuse defective.
Battery failing.
Airways obstructed.Remove obstruction.
Kink or obstruction in ventilation
hose/mask/tube.
Tube incorre ct ly position ed.Correct tube position.
Replace battery in battery compartment
(10.1, page 57).
If ventilator still refuses to switch on,
have internal auxiliary battery replaced
by manufacturer or authorized
specialists.
54Troubleshooting
Page 55
EN
FaultCauseRemedy
Visual alarms flashing, but
no acoustic alarm and no
audio response.
Acoustic alarm, but no
visual alarm flashing.
Acoustic alarm sounds
and all visual alarms
flashing.
Alarm: Device
malfunction.
No audio response.Audio response deactivated.
MEDUMAT Easy is
functioning, but without
any displays.
MV too low.
Unusually high oxygen
consumption.
MEDUMAT Easy will not
switch off.
Pressure gauge not
reading “0”.
Short-term electronic problem or
defect in electronic system.
Device defective.
Pressure gauge hose on
MEDUMAT Easy or on patient
valve slipped off.
Kink in pressure gauge hose.
Venti lation paramet er(s)
incorrectly set.
MEDUMAT Easy defective.Arrange for repair.
Leak in oxygen supply.Find and eliminate leak (8.2, page 47).
Operating error.
MEDUMAT Easy defective.Arrange for repair.
Switch off, then on again.
If error recurs, arrange for repairs.
Activate audio response (6.10,
page 32).
Check pressure gauge hose.
Check ventilation parameter(s).
Keep button pressed for at least 3
seconds.
Troubleshooting55
Page 56
EN
10. Servicing
Have the cleaned and disinfected device serviced at
regular intervals. Servicing, safety checks ([sicherheitstechnische Kontrollen or STKs] in accordance
with §6 of the German law governing the owners/
operators of medical devices - only applies to Germany) and maintenance measures such as servicing and
repairs may only be performed by the manufacturer
or by specialists expressly so authorized by the manufacturer.
Maintain the following intervals:
IntervalParts affected
– System components: e.g. carrying
systems, tube connections*
Every 2 years (service
and safety check)
Every 4 years
Every 10 yearsSteel and aluminum oxygen cylinders
–Accessories
–Test bag
–Oxygen fittings
– Specified wear parts relevant to
safety
–Oxygen fittings
– Specified wear parts relevant to
safety
*Disposable hose system WM 28110 does not
require any maintenance.
Person to carry
out
Manufacturer or
specialists expressly
so authorized by
the manufacturer
56Servicing
Page 57
EN
10.1 Batteries
MEDUMAT Easy is equipped with two batteries:
•Main battery (lithium battery 3.6 V) for main
power supply. This may be changed by the operator.
•A CR2430 button cell. This can only be changed
by specialist personnel. It supplies auxiliary power to the electronic system if the main battery
fails. This makes it possible to set off an alarm
even if the main battery fails.
The battery capacity is calculated to be sufficient for
power requirements under normal operating conditions for the full period between the two-yearly services. The batteries are completely replaced every
two years during servicing.
We recommend having the batteries changed only
by the manufacturer – WEINMANN Emergency – or
by qualified specialists expressly authorized by the
manufacturer. Special precautions need to be taken
during the change in order to protect the electronic
system.
In an emergency, proceed as follows:
Changing the main battery
1. Make sure the ventilator is switched off.
2. Open the battery compartment on the side of
the MEDUMAT Easy (e.g. with a coin).
3. Remove the old 3.6 V lithium battery.
4. Insert a new battery. Make sure it is inserted the
right way round.
5. Close the battery compartment again.
Important
The 3.6 V lithium battery is a special battery for this
Servicing57
Page 58
EN
unit. Use only batteries supplied by
Spontaneous
breathing
arm
Expirationarm
WEINMANN Emergency.
10.2 Change valve membrane in patient valve
If one of the valve membranes in the expiration or
spontaneous breathing arms of the patient valve is
crinkled, sticky or misshapen, it must be changed.
Spontaneous breathing arm
1. Take the spontaneous breathing insert out of
the patient valve. To do so, push the two locking lugs out of their seat, using a small screwdriver, for example.
2. Pull the defective valve membrane out of the
spontaneous breathing insert using pointed
tweezers.
3. Put in a new valve membrane.
4. Push the spontaneous breathing insert back into
the patient valve.
10.3 Storage
58Servicing
Expiration arm
1. Use pointed tweezers to pull the defective valve
membrane out of the expiration arm.
2. Insert a new valve membrane.
If you do not intend to use MEDUMAT Easy for a
long period, we recommend the following storage
precautions:
Page 59
EN
ImportantRemember that the ventilator still requires servicing
10.4 Disposal
1. Clean and disinfect the ventilator (see “7. Hygienic preparation” on page 40).
2. Store MEDUMAT Easy in a dry place.
3. The battery can remain inside the unit even for
lengthy periods.
at the specified intervals even when in storage, otherwise it cannot be used when removed from
storage.
NoteFor disposable hose system WM 28110, observe the
storage temperature of -40 °C to 70 °C at a relative
humidity of 15 % to 95 %. This product can be
stored for a maximum of 2 years.
Do not dispose of the unit with domestic waste. For
proper disposal of the device and its components,
please contact a certified waste disposal site for electronic goods. Ask your Environmental Officer or local
council for the address. The device packaging (cardboard and inserts) can be disposed of in paper recycling facilities.
Disposing of batteries/rechargeable
batteries
Used batteries/rechargeable batteries may not be
disposed of in domestic waste. Contact WEINMANN
Emergency or your local authority waste disposal department.
Disposing of the patient hose system
After use, the patient hose system should be disposed of properly with plastics.
Servicing59
Page 60
EN
11. Product and accessories
11.1 Standard product
MEDUMAT Easy, completeWM28000
comprising:
–MEDUMAT Easy, single unitWM 28010
–Operating instructionsWM 16862
–Summary of operating instructionsWM 16844
–Set of mounting attachmentsWM15007
–Ventilation hose and patient valve
with spontaneous breathing facility (reusable)WM22520
–Ventilation mask for adults, size 5WM 5074
–Test kit for functional checksWM15323
11.2 Accessories
The following accessories are not included with the standard product:
11. Disposable ventilation mask
–size 3, for children and juvenilesWM 10563
–size 5, for adultsWM 10565
12. Set tracheal tubes (disposable):
–Set tracheal tubesWM 15075
–Set tracheal tubes, DIN 13232-N WM 15076
–Set tracheal tubes, DIN 13232-KWM 15077
13. Pressure hoses:
–1m, straight connection nozzle with nut at both ends WM22301
–1m, straight connection nozzle with nut and angled
connection nozzle with nut WM22302
–3m, sealing nipple and plug acc. to DIN 13260WM 22303
–3m, sealing nipple and bayonetWM22304
–3m, sealing nipple and straight connection nozzle
with nut WM22306
–3m, sealing nipple and angled connection nozzle
with nut WM22307
–3m, DIN 13260 plug and
straight connection nozzle with nut WM22308
–3m, sealing nipple (AGA) and straight connection
nozzle with nut WM22309
–3m, bayonet (male) and
straight connection nozzle with nut WM22311
–3m with O
plug acc. to DIN 13260 and
2
straight connection nozzle with nut WM22312
–3m, sealing nipple (AGA) and angled connection
nozzle with nut WM22313
–3m, bayonet (male) and angled connection nozzle
with nut WM22314
–1m, straight connection nozzle with nut and
screw nozzle WM22316
–3m, sealing nipple (Walther) and sealing nipple (AGA) WM 22288
–3m, bayonet seal at both ends WM 22371
14. Set, adapter G3/8-NIST, set of 5WM15554
15. O
plug DIN 13260-S-O2 for GCS socket WM2057
2
16. Angle adapter for French couplingWM22910
17. Hose casing WM8297
Product and accessories61
Page 62
EN
11.3 Spare parts
1. Washer for pressure hoseWM 1145/31
2. Set of mounting attachmentsWM15007
3. Battery, Li 3.6 VWM 28045
4. Ventilation hose and patient valve
with spontaneous breathing facility (reusable)WM22520
comprising:
–connection nozzle to patient hoseWM 3213
–upper control elementWM 3181
–Lip membraneWM 3211
–Bottom control element fittedWM 3285
consisting of:
–Bottom control element for spontaneous breathing WM 3281
–insert, spontaneous breathing arm WM 3282
–valve membrane for spontaneous breathing arm WM 3284
–valve membrane for expiration arm WM 3212
–O-ring 15/1,5 WM 1145/118
62Product and accessories
Page 63
EN
12. Technical data
MEDUMAT Easy
Dimensions
LxHxB in mm
100x145x90
incl. connections
Weight incl.
accessories
approx. 0.6 kg
Device category
(93/42/EEC)
II b
Operation:
Te mp e ra t ur e r an g e
Humidity
Air pressure
–18 °C to +60 °C
max. 95 %
70 kPa
(1)
to 110 kPa
Storage–40 °C to +70 °C
Electromagnetic
compatibility (EMC)
according to
EN 60601-1-2
and EN 794-3:
– Radio interference
suppression
– Radio interference
resistance
Te st pa r am e te r s an d l i mi t
values can be requested
from the manufacturer
(WEINMANN Emergency
Medical Technology
GmbH + Co. KG,
Frohboesestraße 12,
22525 Hamburg,
Germany).
EN 55011 B
EN 61000-4 Parts 2 to 6,
Part 11
Control
timing pulse, volume
constant
Gas inputMedical oxygen
Operating pressure2.7 to 6.0 bar
(2)
Minimum gas
volume required
70 l/min O
2
Insp-exp. ratio (I:E)1:1.67
Venti la tion freq uency
continuously variable
from 10 to 30 min
-1
Minute volume (MV)
continuously variable
from 3 to 16 l/min
MV tolerances:
Room temp. (20 °C)
–18 °C to +60 °C
for 3 l/min = ±20 %
for >3 l/min = ±15 %
for 3 l/min = ±35%
for >3 l/min = ±20%
Recommended safe distances between portable or mobile HF
telecommunication devices (e.g. mobile phones) and MEDUMAT Easy
Rated output of HF
device
in W
0.010.120.23
0.10.340.73
11.202.30
103.407.30
10012.0023.00
64Technical data
Safe distance depending on transmission frequency
in m
80 MHz - 800 MHz800 MHz – 2.5 GHz
Page 65
EN
12.1 Pneumatic / electronic systems
V6
V5
V4
V2
V1
V3
Relief valve
3/2-way valve
Function
Pneumatic system
Pressure regulator
Booster valve with
Flow
Coupling element
Respiratory frequency
Position
Demandflow
Battery
Sound
source
M/T
O/I
LED bar graph
4bar
Electronic
system
2,7-6bar2,5 bar
RF/MV
100mbar
Patient valve
Exp.
Insp.
metering facility
100mbar
10mbar
PpEPpEPp
E
MPU speech module
Stenosis
Disconnection
< 2,7
Batterie
P
Ventilation
hose
Pressure
gauge tube
The input pressure at p is max. 6 bar. This is reduced
by V1 to 2.5 bar dyn. This is the input pressure at V2,
V3 and V4.
Inspiration
An electrical impulse to V2 opens V3 and closes V4.
Oxygen flows through the ventilation hose to the patient valve. If the ventilation pressure in the patient
valve reaches >100 mbar, the relief valve V5 will
open.
Expiration
A fresh electrical impulse closes V2. The relief valve
V4 opens and vents the ventilation hose. The patient
breathes out through the patient valve.
Technical data65
Page 66
EN
Demandflow
Minute volume (MV) in l/min
Frequency in l/min
Minute volume (V
t
) in l
1,10,10,30,60,91,11,1
An inspiration impulse (trigger) at V2 opens valves
V3 and V4.
Electronic system
The microprocessor-controlled electronic system sets
the ventilation parameters and monitors ventilation,
and also O
visual and acoustic alarm is given. The ventilator has
an audio response facility that can be switched on
for user guidance.
supply and power supply. If necessary, a
2
12.2 Relationship between ventilation parameters
The following diagram shows the relationship between the ventilation parameters “minute volume”
and “respiratory frequency”:
66Technical data
Page 67
EN
13. Warranty
WEINMANN Emergency gives the customer a limited
manufacturer warranty on new original WEINMANN
Emergency products and any replacement part fitted
by WEINMANN Emergency in accordance with the
warranty conditions applicable to the product in
question and in accordance with the warranty
periods from date of purchase as listed below. The
warranty conditions can be downloaded from
www.weinmann-emergency.de on the Internet.
We can also send you the warranty conditions on request.
In the event of a claim under warranty, contact your
specialist dealer.
Product
WEINMANN Emergency devices including accessories (except
masks) for oxygen medicine and emergency medicine
Masks including accessories, rechargeable batteries, batteries
(unless quoted differently in the technical documentation),
sensors, tube systems
Disposable productsNone
Warranty
period
2 years
6 months
14. Declaration of conformity
WEINMANN Emergency Medical Technology GmbH
+ Co. KG declares herewith that the product complies fully with the respective regulations of the Medical Device Directive 93/42/EEC. The unabridged text
of the Declaration of Conformity can be found on
our website at www.weinmann-emergency.de