Vega VEGAPULS 62 standpipeversion Operating Instructions Manual

Page 1
Operating Instructions
VEGAPULS 62 standpipe version
Foundation Fieldbus
Document ID:
36510
Radar
Page 2
1 About this document
1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4
1.3 Symbolism used. . . . . . . . . . . . . . . . . . . . . . . . . . . .
4
2 For your safety
2.1 Authorised personnel . . . . . . . . . . . . . . . . . . . . . . . .
5
2.2 Appropriate use . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5
2.3 Warning about misuse . . . . . . . . . . . . . . . . . . . . . . .
5
2.4 General safety instructions . . . . . . . . . . . . . . . . . . . .
5
2.5 CE conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6
2.6 NAMUR recommendations . . . . . . . . . . . . . . . . . . . .
6
2.7 Radio license for Europe. . . . . . . . . . . . . . . . . . . . . .
6
2.8 Radio license for USA/Canada . . . . . . . . . . . . . . . . .
6
2.9 Environmental instructions. . . . . . . . . . . . . . . . . . . . .
7
3 Product description
3.1 Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8
3.2 Principle of operation . . . . . . . . . . . . . . . . . . . . . . . .
9
3.3 Packaging, transport and storage . . . . . . . . . . . . . . .
10
3.4 Accessories and replacement parts . . . . . . . . . . . . . .
10
4 Mounting
4.1 General instructions . . . . . . . . . . . . . . . . . . . . . . . . .
12
4.2 Instructions for installation . . . . . . . . . . . . . . . . . . . . .
12
5 Connecting to the bus system
5.1 Preparing the connection . . . . . . . . . . . . . . . . . . . . .
13
5.2 Connect. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
14
5.3 Wiring plan, single chamber housing . . . . . . . . . . . . .
15
5.4 Wiring plan, double chamber housing . . . . . . . . . . . .
16
5.5 Wiring plan with double chamber housing Ex d . . . . .
17
5.6 Wiring plan - version IP 66/IP 68, 1 bar . . . . . . . . . . .
19
5.7 Switch-on phase. . . . . . . . . . . . . . . . . . . . . . . . . . . .
19
6 Set up with the indicating and adjustment module
6.1 Insert indicating and adjustment module. . . . . . . . . . .
20
6.2 Adjustment system . . . . . . . . . . . . . . . . . . . . . . . . . .
22
6.3 Parameter adjustment. . . . . . . . . . . . . . . . . . . . . . . .
23
6.4 Saving the parameter adjustment data . . . . . . . . . . . .
30
7 Setup with PACTware
7.1 Connect the PC . . . . . . . . . . . . . . . . . . . . . . . . . . . .
32
7.2 Parameter adjustment with PACTware . . . . . . . . . . . .
32
7.3 Saving the parameter adjustment data . . . . . . . . . . . .
33
8 Set up with other systems
8.1 DD adjustment programs . . . . . . . . . . . . . . . . . . . . .
34
2 VEGAPULS 62 standpipe version Foundation Fieldbus
Contents
36510-EN-120301
Page 3
8.2 Communicator 375, 475 . . . . . . . . . . . . . . . . . . . . . .
34
9 Diagnosis, Asset Management and service
9.1 Maintaining . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
35
9.2 Measured value and event memory . . . . . . . . . . . . . .
35
9.3 Asset Management function . . . . . . . . . . . . . . . . . . .
36
9.4 Remove interferences. . . . . . . . . . . . . . . . . . . . . . . .
40
9.5 Exchanging the electronics module . . . . . . . . . . . . . .
43
9.6 Software update . . . . . . . . . . . . . . . . . . . . . . . . . . . .
43
9.7 How to proceed in case of repair. . . . . . . . . . . . . . . .
44
10 Dismounting
10.1 Dismounting steps . . . . . . . . . . . . . . . . . . . . . . . . . .
45
10.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
45
11 Supplement
11.1 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
46
11.2 Supplementary information Foundation Fieldbus . . . . .
54
11.3 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
66
Safety instructions for Ex areas
Please note the Ex-specic safety information for installation and operation in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex-approved instruments.
Editing status: 2012-02-08
VEGAPULS 62 standpipe version Foundation Fieldbus 3
Contents
36510-EN-120301
Page 4
1 About this document
1.1 Function
This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectication. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained qualied personnel. The contents of this manual should be made available to these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result. Warning: If this warning is ignored, injury to persons and/or serious damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
l List
The dot set in front indicates a list with no implied sequence.
à Action
This arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
4 VEGAPULS 62 standpipe version Foundation Fieldbus
1 About this document
36510-EN-120301
Page 5
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.
2.2 Appropriate use
VEGAPULS 62 is a sensor for continuous level measurement.
You can nd detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly used according to the specications in the operating instructions manual as well as possible supplementary instructions.
2.3 Warning about misuse
Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.
2.4 General safety instructions
This is a state-of-the-art instrument complying with all prevailing regulations and guidelines. The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for the trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regulations.
The safety instructions in this operating instructions manual, the national installation standards as well as the valid safety regulations and accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modications are explicitly forbidden.
The safety approval markings and safety tips on the device must also be observed.
VEGAPULS 62 standpipe version Foundation Fieldbus 5
2 For your safety
36510-EN-120301
Page 6
Depending on the instrument version, the emitting frequencies are in the C or K band range. The low emitting frequencies are far below the internationally approved limit values. When used correctly, there is no danger to health.
2.5 CE conformity
The device fullls the legal requirements of the applicable EC guidelines. With the CE mark, we provide conrmation of successful testing.
You can nd the conformity certicate in the download section of our homepage.
Only with Ex-d-ia version
The instrument is designed for use in an industrial environment. Nevertheless, electromagnetic interference from electrical conductors
and radiated emissions must be taken into account, as is usual with a class A instrument according to EN 61326-1. If the instrument is used in a dierent environment, its electromagnetic compatibility with other devices must be ensured by suitable measures.
2.6 NAMUR recommendations
NAMUR is a user association for automation technology in the process industries in Germany. The published NAMUR recommendations are accepted as the standard in eld instrumentation.
The device fullls the requirements of the following NAMUR recommendations:
l NE 21 – Electromagnetic compatibility of equipment l NE 43 – Standardization of the signal level for malfunction
information from digital transmitters
l NE 53 – Compatibility of eld devices and indicating/adjustment
components
l NE 107 - Self-monitoring and diagnosis of eld devices
For further information see
www.namur.de.
2.7 Radio license for Europe
The instrument is approved according to EN 302372-1/2 (2006-04) for use in closed vessels.
2.8 Radio license for USA/Canada
The instrument is in conformity with part 15 of the FCC regulations. Take note of the following two regulations:
l The device must not generate interfering emissions, and
6 VEGAPULS 62 standpipe version Foundation Fieldbus
2 For your safety
36510-EN-120301
Page 7
l The device must be non-sensitive to interfering immissions,
including those that may cause undesirable operating conditions.
Modications not expressly approved by the manufacturer will lead to expiry of the operating licence according to FCC/IC.
The instrument is in conformity with RSS-210 of the IC regulations.
The instrument may only be operated in closed vessels made of metal,
concrete, or bre-reinforced plastic.
2.9 Environmental instructions
Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental protection.
The environment management system is certied according to DIN EN ISO 14001.
Please help us full this obligation by observing the environmental
instructions in this manual:
l Chapter "Packaging, transport and storage" l Chapter "Disposal"
VEGAPULS 62 standpipe version Foundation Fieldbus 7
2 For your safety
36510-EN-120301
Page 8
3 Product description
3.1 Structure
The type label contains the most important data for identication and use of the instrument:
2
1
13
14
15
16
12
11
5
3
6
4
7 8 9
10
Fig. 1: Structure of the type label (example)
1 Instrument type 2 Product code 3 Approvals 4 Power supply and signal output, electronics 5 Protection rating 6 Measuring range 7 Process and ambient temperature, process pressure 8 Material, wetted parts 9 Hardware and software version 10 Order number 11 Serial number of the instrument 12 Symbol of the device protection class 13 ID numbers, instrument documentation 14 Note to observe the instrument documentation 15 Notied authority for CE marking 16 Approval directive
With the serial number of the instrument on the type label you have access to the following data on our homepage:
l Article number of the instrument (HTML) l Delivery date (HTML) l Order-specic instrument features (HTML) l Operating instructions at the time of shipment (PDF) l Order-specic sensor data for an electronics exc hange (XML) l Test certicate "Measuring Accuracy" (PDF)
Go to
www.vega.com, "Service" "VEGA Tools" and "serial number
search".
Type label
Serial number
8 VEGAPULS 62 standpipe version Foundation Fieldbus
3 Product description
36510-EN-120301
Page 9
This operating instructions manual applies to the following instrument versions:
l Hardware from 2.1.1 l Software from 4.4.0
The instrument is available in two dierent electronics versions. Each version can be identied via the product code on the type label as well as on the electronics.
l Standard electronics type PS60FFC.- l Electronics with increased sensitivity type PS60FFS.-
The scope of delivery encompasses:
l Radar sensor l Documentation
- this operating instructions manual
- Test certicate measuring accuracy, depending on the
instrument VEGAPULS 62 (optional)
- Operating instructions manual "Indicating and adjustment module PLICSCOM" (optional)
- Supplementary instructions "GSM/GPRS radio module" (op- tional)
- Supplementary instructions manual "Heating for indicating and adjustment module" (optional)
- Supplementary instructions manual "Plug connector for con- tinuously measuring sensors" (optional)
- Ex-specic "Safety instructions" (with Ex versions)
- if necessary, further certicates
3.2 Principle of operation
The instrument is suitable for applications in liquids in storage tanks and process vessels under dicult process conditions. Application possibilities can be found in the chemical industry, in environmental and recycling technology as well as in the petrochemical industry.
The version with standpipe antenna is particularly suitable for measurement of solvents and liquid gases, for vessels with foam generation or for measurement of products with low dielectric gures
(DK > 1.6).
Measurement in a standpipe is not recommended for very adhesive
products.
The standard electronics enables the use of instruments in products with an ε
r
-Wert 1.8. The electronics version with increased sensitivity
enables the use of the instrument also in applications with very poor reective properties or products with an ε
r
value 1.5. The values that
can be actually reached depend on the measurement conditions, the antenna system or the standpipe or bypass tube.
Scope of the operating instructions manual
Versions
Scope of delivery
Application area
VEGAPULS 62 standpipe version Foundation Fieldbus 9
3 Product description
36510-EN-120301
Page 10
The antenna of the radar sensor emits short radar pulses with a duration of approx. 1 ns. These pulses are reected by the product and received by the antenna as echoes. The running time of the radar pulses from emission to reception is proportional to the distance and hence to the level. The determined level is converted into an appropriate output signal and outputted as measured value.
3.3 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN EN 24180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
l Not in the open l Dry and dust free l Not exposed to corrosive media l Protected against solar radiation l Avoiding mechanical shock and vibration
l Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
l Relative humidity 20 85 %
3.4 Accessories and replacement parts
The indicating and adjustment module PLICSCOM is used for measured value indication, adjustment and diagnosis. It can be inserted into the sensor and removed at any time.
You can nd further information in the operating instructions "Indicating and adjustment module PLICSCOM" (Document-ID 27835).
Functional principle
Packaging
Transport
Transport inspection
Storage
Storage and transport
temperature
Indicating and adjust­ment module
10 VEGAPULS 62 standpipe version Foundation Fieldbus
3 Product description
36510-EN-120301
Page 11
The interface adapter VEGACONNECT enables the connection of communication-capable instruments to the USB interface of a PC. For parameter adjustment of these instruments, an adjustment software such as PACTware with VEGA-DTM is required.
You can nd further information in the operating instructions "Interface adapter VEGACONNECT" (Document-ID 32628).
VEGADIS 61 is an external indicating and adjustment unit for sensors with single chamber housing and Ex-d double chamber housing.
It is suitable for measured value indication and adjustment of plics
®
sensors and is connected to the sensor with a four-wire, screened standard cable up to 50 m long.
You can nd further information in the operating instructions
"VEGADIS 61" (Document-ID 27720).
The PLICSMOBILE T61 is an external GSM/GPRS radio unit for
transmission of measured values and for remote parameter adjust­ment of plics
®
sensors. The adjustment is carried out via PACTware/
DTM by using the integrated USB connection.
You can nd further information in the supplementary instructions "PLICSMOBILE T61" (Document-ID 36849).
The protective cover protects the sensor housing against soiling and
intense heat from solar radiation.
You will nd additional information in the supplementary instructions manual "Protective cover" (Document-ID 34296).
Flanges are available in dierent versions according to the following standards: DIN 2501, EN 1092-1, ANSI B 16.5, JIS B 2210-1984,
GOST 12821-80.
You can nd additional information in the supplementary instructions manual "Flanges according to DIN-EN-ASME-JIS" (Document-ID
31088).
The electronics module VEGAPULS series 60 is a replacement part for radar sensors of VEGAPULS series 60. An own version is available for each type of signal output.
You can nd further information in the operating instructions "Electronics module VEGAPULS series 60" (Document-ID 36801).
Interface adapter
External indicating and
adjustment unit
External radio unit
Protective cap
Flanges
Electronics module
VEGAPULS 62 standpipe version Foundation Fieldbus 11
3 Product description
36510-EN-120301
Page 12
4 Mounting
4.1 General instructions
With instruments with threaded process tting, suitable tools must be applied for tightening the hexagon.
Warning:
The housing must not be used to screw the instrument in! Applying tightening force can damage internal parts of the housing.
Use the recommended cables (see chapter "Connecting to power supply") and tighten the cable gland.
You can give your instrument additional protection against moisture penetration by leading the connection cable downward in front of the cable entry. Rain and condensation water can thus drain off. This applies mainly to outdoor mounting as well as installation in areas where high humidity is expected (e.g. through cleaning processes) or on cooled or heated vessels.
Make sure that all parts of the instrument exposed to the process, in particular the active measuring component, process seal and process tting, are suitable for the existing process conditions. These include above all the process pressure, process temperature as well as the chemical properties of the medium.
You can nd the specications in chapter "Technical data" and on the type label.
4.2 Instructions for installation
By using the standpipe version, the inuence of turbulence and vessel installations, such as e.g. heating spirals or agitators, is excluded. If turbulence or vigorous product movement occurs in the vessel, long standpipe antennas should be fastened to the vessel wall.
The standpipe antenna must extend all the way down to the requested min. level, as measurement is only possible within the tube. If a good mixing of the product is important, you should use a version with perforated surge pipe.
Warning:
Remember that with the version with ¾" thread, a max. torque of 80 Nm (59 lbf ft) must not be exceeded.
Screwing in
Moisture
Suitability for the pro-
cess conditions
Installation
12 VEGAPULS 62 standpipe version Foundation Fieldbus
4 Mounting
36510-EN-120301
Page 13
5 Connecting to the bus system
5.1 Preparing the connection
Always keep in mind the following safety instructions:
l Connect only in the complete absence of line voltage l If voltage surges are expected, install overvoltage arresters
The instrument requires a operating voltage of 9 … 32 V DC. Operating voltage and the digital bus signal are carried on the same two-wire connection cable. Power is supplied via the H1 power supply.
Connection is carried out with screened cable according to Fieldbus
specication.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm (0.2 0.35 in) ensures the seal eect of the cable gland. If you are
using cable with a dierent diameter or cross-section, exchange the seal or use a suitable cable gland.
Make sure that the entire installation is carried out according to the Fieldbus specication. In particular, make sure that the bus is
terminated with suitable terminating resistors.
With plastic housing, the NPT cable gland or the Conduit steel tube must be screwed without grease into the threaded insert.
Max. torque for all housings see chapter "Technical data"
Make sure that the cable screening and ground is executed according
to the Fielbus specication. If electromagnetic interference is expected which is above the test values of EN 61326-1 for industrial areas, we recommend to connect the cable screen on both ends to ground potential.
In this case with systems with potential equalisation, connect the cable screen directly to ground potential at the power supply unit, in the connection box and at the sensor. The screen in the sensor must be connected directly to the internal ground terminal. The ground terminal outside on the housing must be connected to the potential equalisation (low impedance).
In systems without potential equalisation with cable screening on both sides, connect the cable screen directly to ground potential at the power supply unit and at the sensor. In the connection box or T- distributor, the screen of the short stub to the sensor must not be connected to ground potential or to another cable screen. The cable screens to the power supply unit and to the next distributor must be connected to each other and also connected to ground potential via a ceramic capacitor (e.g. 1 nF, 1500 V). Low-frequency potential equalisation currents are thus suppressed, but the protective eect against high frequency interference signals remains.
Safety instructions
Voltage supply
Connection cable
Cable gland ½ NPT
Cable screening and
grounding
VEGAPULS 62 standpipe version Foundation Fieldbus 13
5 Connecting to the bus system
36510-EN-120301
Page 14
5.2 Connect
The voltage supply and signal output are connected via the spring­loaded terminals in the housing.
The connection to the indicating and adjustment module or to the interface adapter is carried out via contact pins in the housing.
Proceed as follows:
1 Unscrew the housing cover 2 If an indicating and adjustment module is installed, remove it by
turning it slightly to the left.
3 Loosen compression nut of the cable entry 4 Remove approx. 10 cm (4 in) of the cable mantle, strip approx.
1 cm (0.4 in) of insulation from the ends of the individual wires
5 Insert the cable into the sensor through the cable entry
Fig. 2: Connection steps 5 and 6
6 Insert the wire ends into the terminals according to the wiring plan
Information:
Solid cores as well as exible cores with cable end sleeves are inserted directly into the terminal openings. In case of exible cores without end sleeves, press the terminal head with a small screwdriver; the terminal opening is freed. When the screwdriver is released, the terminal closes again.
7 Check the hold of the wires in the terminals by lightly pulling on
them
8 Connect the screen to the internal ground terminal, connect the
outer ground terminal to potential equalisation
Connection technology
Connection procedure
14 VEGAPULS 62 standpipe version Foundation Fieldbus
5 Connecting to the bus system
36510-EN-120301
Page 15
9 Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
10 Screw the housing cover back on
The electrical connection is nished.
Information:
The terminal block is pluggable and can be removed from the electronics. To do this, lift the terminal block with a small screwdriver and pull it out. When inserting the terminal block again, you should hear it snap in.
5.3 Wiring plan, single chamber housing
1
2
( )
(-)
1
5
0
1
0
1
+
6 7 8
Bus
2
3
4
5
Fig. 3: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output 2 Contact pins for the indicating and adjustment module or interface adapter 3 Simulation switch ("1" = mode for simulation release) 4 For external indicating and adjustment unit 5 Ground terminal for connection of the cable screen
Electronics and connec­tion compartment
VEGAPULS 62 standpipe version Foundation Fieldbus 15
5 Connecting to the bus system
36510-EN-120301
Page 16
5.4 Wiring plan, double chamber housing
5
0
1
0
1
+
6 7 8
Bus
2
3
1
2
( )
(-)
11
Fig. 4: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment 2 Contact pins for the indicating and adjustment module or interface adapter 3 Simulation switch ("on" = mode for simulation release)
+
Bus
2
3
1
2
( )
(-)
1
Fig. 5: Connection compartment, double chamber housing
1 Voltage supply, signal output 2 For indicating and adjustment module or interface adapter 3 Ground terminal for connection of the cable screen
Electronics compart­ment
Connection compart­ment
16 VEGAPULS 62 standpipe version Foundation Fieldbus
5 Connecting to the bus system
36510-EN-120301
Page 17
1
Bus
USB
Status
Test
SIM-Card
1
2
+
( )
(-)
Fig. 6: CConnection of the voltage supply of the radio module
1 Voltage supply
You can nd detailed information for connection in the supplementary instructions "PLICSMOBILE GSM/GPRS radio module".
5.5 Wiring plan with double chamber housing Ex d
5
0
1
0
1
+
6 7 8
Bus
2
3
1
2
( )
(-)
11
Fig. 7: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment 2 Contact pins for the indicating and adjustment module or interface adapter 3 Simulation switch ("on" = mode for simulation release)
Radio module PLICS­MOBILE integrated in
the connection compart­ment
Electronics compart­ment
VEGAPULS 62 standpipe version Foundation Fieldbus 17
5 Connecting to the bus system
36510-EN-120301
Page 18
+
Bus
2
1
2
( )
(-)
1
Fig. 8: Connection compartment, double chamber housing Ex d
1 Voltage supply, signal output 2 Ground terminal for connection of the cable screen
34
1
2
Fig. 9: Top view of the plug connector
1 Pin 1 2 Pin 2 3 Pin 3 4 Pin 4
Contact pin Colour connection
cable in the sensor
Terminal, electronics module
Pin 1 Brown 5
Pin 2 White 6
Pin 3 Blue 7
Pin 4 Black 8
Connection compart­ment
Plug M12 x 1 for external indicating and adjust­ment unit
18 VEGAPULS 62 standpipe version Foundation Fieldbus
5 Connecting to the bus system
36510-EN-120301
Page 19
5.6 Wiring plan - version IP 66/IP 68, 1 bar
1
2
Fig. 10: Wire assignment x-connected connection cable
1 brown (+) and blue (-) to power supply or to the processing system 2 Shielding
5.7 Switch-on phase
After VEGAPULS 62 is connected to the bus system, the instrument carries out a self-test for approx. 30 seconds. The following steps are carried out:
l Internal check of the electronics l Indication of the instrument type, hardware and software version,
measurement loop name on the display or PC
l Indication of the status message "F 105 Determine measured
value" on the display or PC
l Status byte goes briey to fault value
As soon as a plausible measured value is found, it is outputted to the signal cable. The value corresponds to the actual level as well as the settings already carried out, e.g. factory setting.
Wire assignment con­nection cable
VEGAPULS 62 standpipe version Foundation Fieldbus 19
5 Connecting to the bus system
36510-EN-120301
Page 20
6 Set up with the indicating and adjustment
module
6.1 Insert indicating and adjustment module
The indicating and adjustment module can be inserted into the sensor and removed any time. Four positions displaced by 90° can be selected. It is not necessary to interrupt the power supply.
Proceed as follows:
1 Unscrew the housing cover 2 Place the indicating and adjustment module in the requested
position onto the electronics and turn to the right until it snaps in
3 Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order.
The indicating and adjustment module is powered by the sensor, an
additional connection is not necessary.
Fig. 11: Insertion of the indicating and adjustment module with single chamber housing into the electronics compartment
20 VEGAPULS 62 standpipe version Foundation Fieldbus
6 Set up with the indicating and adjustment module
36510-EN-120301
Page 21
1 2
Fig. 12: Insertion of the indicating and adjustment module with double chamber housing
1 In the electronics compartment 2 In the connection compartment (with Ex d version not possible)
Note:
If you intend to retrot the instrument with an indicating and adjustment module for continuous measured value indication, a higher cover with an inspection glass is required.
VEGAPULS 62 standpipe version Foundation Fieldbus 21
6 Set up with the indicating and adjustment module
36510-EN-120301
Page 22
6.2 Adjustment system
1
2
Fig. 13: Indicating and adjustment elements
1 LC display 2 Adjustment keys
l [OK] key:
- Move to the menu overview
- Conrm selected menu
- Edit parameter
- Save value
l [->] key:
- Presentation change measured value
- Select list entry
- Select editing position
l [+] key:
- Change value of the parameter
l [ESC] key:
- interrupt input
- Return to higher-ranking menu
The sensor is adjusted via the four keys of the indicating and adjustment module. The LC display indicates the individual menu items. The functions of the individual keys are shown in the above illustration. Approx. 10 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not conrmed with [OK] will not be saved.
Key functions
Adjustment system
22 VEGAPULS 62 standpipe version Foundation Fieldbus
6 Set up with the indicating and adjustment module
36510-EN-120301
Page 23
6.3 Parameter adjustment
Through the parameter adjustment the instrument is adapted to the application conditions. The parameter adjustment is carried out via an adjustment menu.
The main menu is divided into ve sections with the following functions:
Setup: Settings, e.g., for measurement loop name, medium, applica- tion, vessel, adjustment, signal output
Display: Settings, e.g., for language, measured value display, lighting
Diagnosis: Information, e.g. on the instrument status, pointer,
reliability, simulation, echo curve
Further settings: Instrument unit, false signal suppression, linear- isation curve, reset, date/time, reset, copy function
Info: Instrument name, hardware and software version, calibration date, instrument features
Information:
In this operating instructions manual, the instrument-specic param­eters in the menu sections "Setup", "Diagnosis" and "Additional settings" are described. The general parameters in these menu section are described in the operating instructions manual "Indicating
and adjustment module".
You can nd in the operating instructions manual "Indicating and adjustment module" also the description of the menu sections "Display" and "Info".
In the main menu point "Setup", the individual submenu points should be selected subsequently and provided with the correct parameters to ensure the optimum adjustment of the measurement. The procedure is described in the following.
Each medium has dierent reection properties. With liquids, further interfering factors are uctuation product surface and foam generation. With bulk solids, these are dust generation, material cone and additional echoes from the vessel wall.
To adapt the sensor to these dierent measuring conditions, the selection "Liquid" or "Bulk solid" should be made in this menu item.
Main menu
Setup/Medium
VEGAPULS 62 standpipe version Foundation Fieldbus 23
6 Set up with the indicating and adjustment module
36510-EN-120301
Page 24
Through this selection, the sensor is adapted perfectly to the product and measurement reliability, particularly in products with poor reective properties, is considerably increased.
Enter the requested parameters via the appropriate keys, save your settings with [OK] and jump to the next menu item with the [ESC] and the [->] key.
Because a radar sensor is a distance measuring instrument, the distance from the sensor to the product surface is measured. To have the real product level displayed, an allocation of the measured distance to the percentage height must be made. To carry out this adjustment, the distance is entered with full and empty vessel. If these values are not known, an adjustment with the distance values, e.g. 10 % and 90 % is also possible. Starting point for these distance specications is always the seal surface of the thread or ange. By means of these settings, the real level is calculated.
The real product level during this adjustment is not important, because the min./max. adjustment is always carried out without changing the product level. These settings can be made ahead of time without the instrument having to be installed.
Proceed as follows:
1 Select the menu item "Setup" with [->] and conrm with [OK]. Now
select with [->] the menu item "Min. adjustment" and confirm with
[OK].
2 Edit the percentage value with [OK] and set the cursor to the
requested position with [->].
3 Set the requested percentage value with [+] and save with [OK].
The cursor jumps now to the distance value.
Setup/Adjustment
Setup/Min. adjustment
24 VEGAPULS 62 standpipe version Foundation Fieldbus
6 Set up with the indicating and adjustment module
36510-EN-120301
Page 25
4 Enter the suitable distance value in m for the empty vessel (e.g.
distance from the sensor to the vessel bottom) corresponding to the percentage value.
5 Save settings with [OK] and move with [ESC] and [->] to the max.
adjustment.
Proceed as follows:
1 Select with [->] the menu item max. adjustment and conrm with
[OK].
2 Prepare the percentage value for editing with [OK] and set the
cursor to the requested position with [->].
3 Set the requested percentage value with [+] and save with [OK].
The cursor jumps now to the distance value.
4 Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. Keep in mind that the max. level must lie below the min. distance to the antenna edge.
5 Save settings with [OK]
Min. and max. measured values are saved in the sensor. The values
are displayed in the menu item "Peak values".
When non-contact level sensors are used, the measurement can be inuenced by the respective process conditions. In this menu item, the measurement reliability of the level echo is displayed as dB value. The measurement reliability equals signal strength minus noise. The higher the value, the more reliable the measurement. With a functioning measurement, the values are > 10 dB.
Setup/Max. adjustment
Diagnosis/Peak value
Diagnosis/Measurement
reliability
VEGAPULS 62 standpipe version Foundation Fieldbus 25
6 Set up with the indicating and adjustment module
36510-EN-120301
Page 26
The "Echo curve" shows the signal strength of the echoes over the measuring range in dB. The signal strength enables an evaluation of the quality of the measurement.
The "False signal suppression" displays the saved false echoes (see menu "Additional settings") of the empty vessel with signal strength in
"dB" over the measuring range.
A comparison of echo curve and false signal suppression allows a
more accurate conclusion on measurement reliability. The selected curve is continuously updated. With the [OK] key, a submenu with zoom functions is opened:
l "X-Zoom": Zoom function for the meas. distance l "Y-Zoom": 1, 2, 5 and 10x signal magnication in "dB" l "Unzoom": Reset the presentation to the nominal measuring range
with single magnication
With the function "Echo curve memory" the echo curve can be saved at the time of setup. This is generally recommended; for using the Asset Management functions it is absolutely necessary. If possible, the curve should be saved with a low level in the vessel.
With the adjustment software PACTware and the PC, the high resolution echo curve can be displayed and used to recognize signal changes over the operating time. In addition, the echo curve of the setup can be also displayed in the echo curve window and compared with the actual echo curve.
Diagnoses/Curve indica­tion
Diagnosis/Echo curve memory
26 VEGAPULS 62 standpipe version Foundation Fieldbus
6 Set up with the indicating and adjustment module
36510-EN-120301
Page 27
The following circumstances cause interfering reections and can inuence the measurement:
l High sockets l Vessel installations such as struts l Agitators l Buildup or welded joints on vessel walls
Note:
A false signal suppression detects, marks and saves these false signals so that they are no longer taken into account for level measurement.
This should be done with a low level so that possible interfering reections can be detected.
Proceed as follows:
1 Select the menu item "Additional settins" with [->] and conrm with
[OK]. With [->] you have to select the menu item "False signal suppression" and conrm with [OK].
2 Conrm again with [OK].
3 Conrm again with [OK].
4 Conrm again with [OK] and enter the actual distance from the
sensor to the product surface.
5 All interfering signals in this section are detected by the sensor
and stored after conrming with [OK].
Further settings/False signal suppression
VEGAPULS 62 standpipe version Foundation Fieldbus 27
6 Set up with the indicating and adjustment module
36510-EN-120301
Page 28
Note:
Check the distance to the product surface, because if an incorrect (too large) value is entered, the existing level will be saved as false signal.
The lling level would then no longer be detectable in this area.
If a false signal suppression is already created in the sensor, then the
following menu window appears when selecting "False signal suppression":
The menu item "Delete" is used to completely delete an already created false signal suppression. This is useful if the saved false signal suppression no longer matches the metrological conditions in the vessel.
The menu item "Extend" is used to extend an already created false signal suppression. This is useful if a false signal suppression was carried out with a too high level and not all false signals could be detected. When selecting "Extend", the distance to the product surface of the created false signal suppression is displayed. This value can now be changed and the false signal suppression can be extended to this range.
A linearization is necessary for all vessels in which the vessel volume does not increase linearly with the level - e.g. in a horizontal cylindrical or spherical tank - and the indication or output of the volume is required. Corresponding linearization curves are preprogrammed for these vessels. They represent the correlation between the level percentage and vessel volume.
By activating the appropriate curve, the volume percentage of the vessel is displayed correctly. If the volume should not be displayed in percent but e.g. in l or kg, a scaling can be also set in the menu item
"Display".
Enter the requested parameters via the appropriate keys, save your settings and jump to the next menu item with the [ESC] and [->] key.
Further settings/Linear­ization curve
28 VEGAPULS 62 standpipe version Foundation Fieldbus
6 Set up with the indicating and adjustment module
36510-EN-120301
Page 29
Caution:
Note the following if the instrument with corresponding approval is used as part of an overll protection system according to WHG:
If a linearisation curve is selected, the measuring signal is no longer compulsorily linear proportional to the level. This must be taken into consideration by the user, particularly when adjusting the switching point on the level switch.
With radar sensors with standpipe antenna, the sensor length is already preset in this menu item. When shortening the standpipe antenna afterwards, this value must be corrected respectively.
When a reset is carried out, all settings (with only a few exceptions) are reset. The exceptions are: PIN, language and lighting.
The following reset functions are available:
Delivery status: Restoring the parameter settings at the time of shipment from the factory incl. the order-specic settings. A created false signal suppression, user-programmable linearization curve as well as the measured value memory will be deleted.
Basic settings: Resetting the parameter settings incl. special and laboratory parameters to the default values of the respective instru­ment. A created false signal suppression, user programmable linearization curve as well as the measured value memory will be deleted.
Setup: Resetting of the parameter settings to the default values of the respective instrument in the menu item Setup. Order-related settings remain but are not taken over into the current parameters. User­generated false signal suppression, user-programmed linearization curve, measured value memory as well as event memory remain untouched. The linearization is set to linear.
False signal suppression: Deleting a previously created false signal suppression. The false signal suppression created in the factory remains active.
Peak values distance: Resetting the measured min. and max. distances to the actual measured value.
Select the requested reset function [->] and conrm with [OK].
Additonal settings/Pro­be length
Additional settings/Re­set
VEGAPULS 62 standpipe version Foundation Fieldbus 29
6 Set up with the indicating and adjustment module
36510-EN-120301
Page 30
The following table shows the default values of VEGAPULS 62:
Menu section Menu item Default value
Setup Measurement
loop name
Sensor
Min. adjustment Length of the standpipe Ex factory
Damping 0.0 s
Display Language Like order
Displayed value Distance
Display unit m(d)
Scaling 0.00 %, 0 l
100.00 %, 100 l
Further settings Distance unit m
Temperature unit °C
Unit SV2 m
Probe length Length of the standpipe Ex factory
Linearisation cur-veLinear
6.4 Saving the parameter adjustment data
It is recommended noting the adjusted data, e.g. in this operating instructions manual and archive them afterwards. They are hence available for multiple use or service purposes.
If the instrument is equipped with an indicating and adjustment module, the data in the sensor can be saved in the indicating and adjustment module. The procedure is described in the operating instructions manual "Indicating and adjustment module" in the menu item "Copy sensor data". The data remain there permanently even if the sensor power supply fails.
The following data or settings for adjustment of the indicating and adjustment module are saved:
l All data of the menu "Setup" and "Display" l In the menu "Additional settings" the items "Sensor-specic units,
temperature unit and linearization"
l The values of the user programmable linearization curve
The function can also be used to transfer settings from one instrument to another instrument of the same type. If it is necessary to exchange a sensor, the indicating and adjustment module is inserted into the replacement instrument and the data are likewise written into the sensor via the menu item "Copy sensor data".
30 VEGAPULS 62 standpipe version Foundation Fieldbus
6 Set up with the indicating and adjustment module
36510-EN-120301
Page 31
VEGAPULS 62 standpipe version Foundation Fieldbus 31
6 Set up with the indicating and adjustment module
36510-EN-120301
Page 32
7 Setup with PACTware
7.1 Connect the PC
3
1
2
Fig. 14: Connection of the PC directly to the sensor via the interface adapter
1 USB cable to the PC 2 Interface adapter VEGACONNECT 4 3 Sensor
Information:
The interface adapter VEGACONNECT 3 or additional previous versions are not suitable for connection to the sensor.
7.2 Parameter adjustment with PACTware
For parameter adjustment of the sensor via a Windows PC, the conguration software PACTware and a suitable instrument driver
(DTM) according to FDT standard are required. The up-to-date PACTware version as well as all available DTMs are compiled in a DTM Collection. The DTMs can also be integrated into other frame applications according to FDT standard.
Note:
To ensure that all instrument functions are supported, you should always use the latest DTM Collection. Furthermore, not all described functions are included in older rmware versions. You can download the latest instrument software from our homepage. A description of the update procedure is also available in the Internet.
Via the interface adapter directly on the sensor
Prerequisites
32 VEGAPULS 62 standpipe version Foundation Fieldbus
7 Setup with PACTware
36510-EN-120301
Page 33
Further setup steps are described in the operating instructions manual
"DTM Collection/PACTware" attached to each DTM Collection and which can also be downloaded from the Internet. Detailed descriptions are available in the online help of PACTware and the DTMs.
Fig. 15: Example of a DTM view
All device DTMs are available as a free-of-charge standard version and as a full version that must be purchased. In the standard version, all functions for complete setup are already included. An assistant for simple project conguration simplies the adjustment considerably. Saving/printing the project as well as import/export functions are also part of the standard version.
In the full version there is also an extended print function for complete project documentation as well as a save function for measured value and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves.
The standard version is available as a free-of-charge download under http://www.vega.com. The full version is available on CD from the agency serving you.
7.3 Saving the parameter adjustment data
We recommend documenting or saving the parameter adjustment data via PACTware. That way the d ata are available for multiple use or service purposes.
Standard/Full version
VEGAPULS 62 standpipe version Foundation Fieldbus 33
7 Setup with PACTware
36510-EN-120301
Page 34
8 Set up with other systems
8.1 DD adjustment programs
Device descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS and PDM.
A free-of-charge download of these les is available via Internet. Move via
www.vega.com and "Downloads" to "Software".
8.2 Communicator 375, 475
Device descriptions for the instrument are available as DD or EDD for parameter adjustment with the Field Communicator 375 or 475.
A free-of-charge download of these les is available via Internet. Move via
www.vega.com and "Downloads" to "Software".
34 VEGAPULS 62 standpipe version Foundation Fieldbus
8 Set up with other systems
36510-EN-120301
Page 35
9 Diagnosis, Asset Management and service
9.1 Maintaining
If the device is used correctly, no maintenance is required in normal operation.
9.2 Measured value and event memory
The instrument has several memories which are available for diagnosis purposes. The data remain even with voltage interruption.
Up to 70,000 measured values can be stored in the sensor in a ring memory. Each entry contains date/time as well as the respective measured value.
Stored values are for example:
l Original sensor value l Filling height l Signal current l Meas. reliability l Electronics temperature
The requested values and recording conditions are set via a PC with PACTware/DTM or the control system with EDD. Data are thus read
out and also reset.
Up to 500 events are automatically stored with a time stamp in the sensor (non-deletable). Each entry contains date/time, event type, event description and value.
Event types are for example:
l Modication of a parameter l Switching on and otimes l Status messages l Error messages
The data are read out via a PC with PACTware/DTM or the control system with EDD.
The echo curves are stored with date and time and the corresponding echo data. The memory is divided into two sections:
Echo curve of the setup: This is used as reference echo curve for the measurement conditions during setup. Changes in the measurement conditions during operation or buildup on the antenna can thus be recognized.
The echo curve of setup is stored via:
l PC with PACTware/DTM l Control system with EDD l Indicating and adjustment module
Measured value memory
Event memory
Echo curve memory
VEGAPULS 62 standpipe version Foundation Fieldbus 35
9 Diagnosis, Asset Management and service
36510-EN-120301
Page 36
Further echo curves: Up to 10 echo curves can be stored in a ring buer in this memory section.
Additional echo curves are stored via:
l PC with PACTware/DTM l Control system with EDD
9.3 Asset Management function
The instrument features self-monitoring and diagnosis according to NE
107 and VDI/VDE 2650. In addition to the status messages in the following tables there are more detailed error messages available under the menu item "Diagnosis" via the indicating and adjustment module, PACTware/DTM and EDD.
The status messages are classied in the following categories:
l Failure l Function check l Out of specication l Maintenance requirement
and explained by pictographs:
41 2 3
Fig. 16: Pictograms of the status messages
1 Failure - red 2 Function check - orange 3 Out of specication - yellow 4 Maintenance - blue
Failure: Due to a malfunction in the instrument, a failure message is outputted.
This status message is always active. A deactivation by the user is not possible.
Function check: The instrument is in operation, the measured value is temporarily invalid (for example during simulation).
This status message is always active. A deactivation by the user is not possible.
Out of specication: The measured value is unstable because the instrument specication is exceeded (e.g. electronics temperature).
This status message is inactive by default. Activation by the user via PACTware/DTM or EDD is possible.
Status messages
36 VEGAPULS 62 standpipe version Foundation Fieldbus
9 Diagnosis, Asset Management and service
36510-EN-120301
Page 37
Maintenance: Due to external inuences, the instrument function is limited. The measurement is aected, but the measured value is still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup).
This status message is inactive by default. Activation by the user via PACTware/DTM or EDD is possible.
The following table shows the codes and text messages of the status
message "Failure" and provides information on causes as well as corrective measures.
Code Text mes-
sage
Cause Rectication PA Dev-
Spec Diagnosis
F013
no measu­red value available
l Sensor does not de-
tect an echo during operation
l Antenna system con-
taminated or defecti­ve
l Check or correct in-
stallation and/or pa­rameter adjustment
l Clean or exchange
process component or antenna
Bit 0
F017 Adjustment
span too small
l Adjustment not within
specication
l Change adjustment
according to the limit values (dierence between min. and max. 10 mm)
Bit 1
F025 Error in the
linearization table
l Index markers are not
continuously rising, for examle unlogical value pairs
l Check linearization
table
l Delete table/Create
new
Bit 2
F036 No operab-
le software
l Failed or interrupted
software update
l Repeat software up-
date
l Check electronics
version
l Exchange of the
electronics
l Send instrument for
repair
Bit 3
F040 Error in the
electronics
l Hardware defect l Exchange of the
electronics
l Send instrument for
repair
Bit 4
F080 l General software er-
ror
l Separate operating
voltage briefly
Bit 5
Failure (failure)
VEGAPULS 62 standpipe version Foundation Fieldbus 37
9 Diagnosis, Asset Management and service
36510-EN-120301
Page 38
Code Text mes-
sage
Cause Rectication PA Dev-
Spec Diagnosis
F105 Determine
measured value
l The instrument is still
in the start phase, the measured value could not yet be de­termined
l Wait for the warm-up
phase
l Duration depending
on the version and parameter adjust­ment up to approxi­mately 3 min.
Bit 6
F113 Communi-
cation error
l Error in the internal
instrument communi­cation
l Separate operating
voltage briefly
l Send instrument for
repair
Bit 7
F125 Unpermis-
sible elect­ronics tem­perature
l Temperature of the
electronics in the non-specied section
l Check ambient tem-
perature
l Isolate electronics l Use instrument with
higher temperature range
Bit 8
F260 Error in the
calibration
l Error in the calibra-
tion carried out in the factory
l Error in the EEPROM
l Exchange of the
electronics
l Send instrument for
repair
Bit 9
F261 Error in the
congurati­on
l Error during setup l False signal sup-
pression faulty
l Error when carrying
out a reset
l Repeat setup l Repeat reset
Bit 10
F264 Installation/
Setup error
l Adjustment not within
the vessel height/ measuring range
l Max. measuring ran-
ge of the instrument not sucient
l Check or correct in-
stallation and/or pa­rameter adjustment
l Use an instrument
with bigger measu­ring range
Bit 11
F265 Measure-
ment func­tion distur­bed
l Sensor does no lon-
ger carry out a mea­surement
l Operating voltage too
low
l Check operating vol-
tage
l Carry out a reset l Separate operating
voltage briefly
Bit 12
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes as well as corrective measures.
Function check
38 VEGAPULS 62 standpipe version Foundation Fieldbus
9 Diagnosis, Asset Management and service
36510-EN-120301
Page 39
Code Text mes-
sage
Cause Rectication
C700 Simulation
l Simulation active l Finish simulation
l Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the status message "Out of specication" and provides information on causes as well as corrective measures.
Code Text mes-
sage
Cause Rectication
S600 Unpermissible
electronics temperature
l Temperature of the elect-
ronics in the non-specied section
l Check ambient temperatu-
re
l Isolate electronics l Use instrument with higher
temperature range
S601 Overlling
l Danger of vessel overlling l Make sure that there is no
further lling
l Check level in the vessel
The following table shows the error codes and text messages in the
status message "Maintenance" and provides information on causes as well as corrective measures.
Code Text mes-
sage
Cause Rectication
M500 Error with the
reset delivery status
l With the reset to delivery
status, the data could not be restored
l Repeat reset l Load XML le with sensor
data into the sensor
M501 Error in the
non-active li­nearization table
l Hardware error EEPROM l Exchange of the electro-
nics
l Send instrument for repair
M502 Error in the
diagnosis me­mory
l Hardware error EEPROM l Exchange of the electro-
nics
l Send instrument for repair
Out of specication
Maintenance
VEGAPULS 62 standpipe version Foundation Fieldbus 39
9 Diagnosis, Asset Management and service
36510-EN-120301
Page 40
Code Text mes-
sage
Cause Rectication
M503 Reliability too
low
l The echot/noise ratio is the
small for a reliable measu­rement
l Check installation and pro-
cess conditions
l Clean the antenna l Change polarisation direc-
tion
l Use instrument with higher
sensitivity
M504 Error on an
device inter­face
l Hardware defect l Check connections
l Exchange of the electro-
nics
l Send instrument for repair
M505 No echo avai-
lable
l Level echo can no longer
be detected
l Clean the antenna l Use a more suitable an-
tenna/sensor
l Remove possible false
echoes
l Optimize sensor position
and orientation
9.4 Remove interferences
The operator of the system is responsible for taking suitable measures to rectify faults.
The rst measures are:
l Evaluation of fault messages via the indicating and adjustment
module
l Checking the 4 20 mA signal l Treatment of measurement errors
Further comprehensive diagnosis possibilities are oered by a PC with the software PACTware and suitable DTM.
In many case, the failure reasons can thus be determined and faults rectied.
The below tables show typical examples for application-relevant measurement errors with standpipe versions. There are two meas­urement errors:
l Constant level l Filling l Emptying
Reaction when malfunc­tions occur
Procedure for fault rec­tication
Treatment of measure­ment errors with stand­pipe versions
40 VEGAPULS 62 standpipe version Foundation Fieldbus
9 Diagnosis, Asset Management and service
36510-EN-120301
Page 41
The images in column "Error pattern" show the actual level with a broken line and the level displayed by the sensor as a continuous line.
1
2
Level
time
0
Fig. 17: The broken line 1 shows the real level, the continuous line 2 shows the level displayed by the sensor
Instructions:
l Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value"
l In case of a too low level indication, the reason could be a line
resistance that is too high
Measurement error with constant level
Fault description Error pattern Cause Rectication
1. Measured value shows a too low or too high level
Level
time
0
l Min./max. adjustment not cor-
rect
l Adapt min./max. adjustment
l Linearization curve incorrect l Adapt linearization curve
l Installation in a bypass tube or
standpipe, hence running time error (small measurement er­ror close to 100 %/large error close to 0 %)
l Check parameter "Application"
with respect to vessel form, adapt if necessary (bypass, standpipe, diameter)
2nd measured value jumps towards 100 %
Level
time
0
l Due to the process, the amp-
litude of the product echo sinks
l A false signal suppression was
not carried out
l Carry out false signal suppres-
sion
l Amplitude or position of a false
echo has changed (e.g. con­densation, buildup); false sig­nal suppression no longer matches
l Determine the reason for the
changed false echoes, carry out false signal suppression, e. g. with condensation
VEGAPULS 62 standpipe version Foundation Fieldbus 41
9 Diagnosis, Asset Management and service
36510-EN-120301
Page 42
Measurement error during lling
Fault description Error pattern Cause Rectication
3. Measured value re­mains in the bottom section during lling
Level
time
0
l The tank oor echo is larger
than the product echo, e.g. with products with ε
r
< 2.5 oil-
based, solvents
l Application parameter "Me-
dium", check Vessel height and Floor form, adapt if necessary
4. Measured value jumps to 100 % or 0 m distance
Level
time
0
l Level echo is no longer de-
tected in the close range due to foam generation or false echoes in the close range. The sensor goes into overll pro­tection mode. The max. level (0 m distance) as well as the status message "Overll pro­tection" are outputted.
l Check measuring site: Antenna
must protrude out of the socket
l Remove pollution on the an-
tenna
l Use a sensor with a more
suitable antenna
Measurement error during emptying
Fault description Error pattern Cause Rectication
5. measured value re- mains unchanged in the close range during emptying
Level
time
0
l False echo larger than the
level echo
l Level echo too small
l Remove false echoes in the
close range. Check: Antenna must protrude out of the socket
l Remove pollution on the an-
tenna
l Minimize interfering installa-
tions in the close range by changing the polarization di­rection
l After removing the false
echoes, the false signal sup­pression must be deleted. Carry out a new false signal suppression
6. Measured value jumps towards 0 % du­ring emptying
Level
time
0
l The tank oor echo is larger
than the product echo, e.g. with products with ε
r
< 2.5 oil-
based, solvents
l Application parameter "Me-
dium type", check Vessel height and Floor form, adapt if necessary
Depending on the reason for the fault and the measures taken, the steps described in chapter "Setup" must be carried out again or must be checked for plausibility and completeness.
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is also available outside the normal working hours on seven days a week around the clock.
Reaction after fault rec­tication
24 hour service hotline
42 VEGAPULS 62 standpipe version Foundation Fieldbus
9 Diagnosis, Asset Management and service
36510-EN-120301
Page 43
Since we oer this service worldwide, the support is in English language. The service is free-of-charge, only the standard telephone costs will be charge.
9.5 Exchanging the electronics module
If the electronics module is defective, it can be replaced by the user.
In Ex applications only one instrument and one electronics module with respective Ex approval may be used.
If there is no electronics module available on site, the electronics
module can be ordered through the agency serving you. The electronics modules are adapted to the respective sensor and dier in signal output or voltage supply.
The new electronics module must be loaded with the default settings of the sensor. These are the options:
l In the factory l Or on site by the user
In both cases, the serial number of the sensor is needed. The serial numbers are stated on the type label of the instrument, on the inside of the housing as well as on the delivery note.
When loading on site, rst of all the order data must be downloaded from the Internet (see operating instructions manual "Electronics
module").
9.6 Software update
The following components are required to update the sensor software:
l Sensor l Voltage supply l Interface adapter VEGACONNECT 4 l PC with PACTware l Current sensor software as le
Caution:
Keep in mind that a software update can lead to expiry of the approvals. You can nd detailed information on our homepage
www.
vega.com.
l Go under "
www.vega.com/downloads" to "Software"
l Select under "plics sensors and instruments", "Firmware updates"
your instrument series and software version
l Load the zip le via the right mouse key with "Save target as", for
example, on the desktop of your PC
l Go with the right mouse key to the folder and select "Extract all" l Store the extracted les, for example on the desktop
Load sensor software to PC
VEGAPULS 62 standpipe version Foundation Fieldbus 43
9 Diagnosis, Asset Management and service
36510-EN-120301
Page 44
l Connecting the sensor to voltage supply and via the interface
adapter to the PC
l Start PACTware and go via the menu "Project" to the VEGA
project assistant
l Select "USB", "Create new project" and "Set instrument online" l Activate the project assistant with "Start"; the connection to the
sensor is carried out automatically and the parameter window "Sensor # online parameter adjustment" opens
l Close the parameter window "Sensor # online parameter adjust-
ment" befor you carry out further steps
l Select the sensor in the project with the right mouse key and go to
"Additional functions"
l Click to "Software update", the window "Sensor # software update"
opens
PACTware now checks the sensor data and shows the actual hardware and software version of the sensor. This procedure lasts approx. 60 s.
l Clcik to "Continue", then to le "Selection" l Select the previously extracted XML le l Push button "Update software"
Then the software update can be started. The additional les are installed automatically. Depending on the sensor, this procedure can last up to 15 min. Then the message appears ""Software update
successfully executed".
9.7 How to proceed in case of repair
If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our homepage at
www.
vega.com under: "Downloads - Forms and certicates - Repair form".
By doing this you help us carry out the repair quickly and without having to call back for needed information.
l Print and ll out one form per instrument l Clean the instrument and pack it damage-proof l Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
l Please ask the agency serving you for the address of your return
shipment. You can nd the competent agency on our website www.vega.com.
Prepare update
Load software into sen-
sor
44 VEGAPULS 62 standpipe version Foundation Fieldbus
9 Diagnosis, Asset Management and service
36510-EN-120301
Page 45
10 Dismounting
10.1 Dismounting steps
Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.
10.2 Disposal
The instrument consists of materials which can be recycled by specialised recycling companies. We use recyclable materials and have designed the electronics to be easily separable.
Correct disposal avoids negative eects on humans and the environ­ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Pass the instrument directly on to a specialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
VEGAPULS 62 standpipe version Foundation Fieldbus 45
10 Dismounting
36510-EN-120301
Page 46
11 Supplement
11.1 Technical data
General data
316L corresponds to 1.4404 or 1.4435
Materials, wetted parts
- Process tting 316L, Hastelloy C22
- Process seal On site (instruments with thread: Klingersil C-4400
is attached)
- Antenna impedance cone PTFE, PP, PEEK, ceramic (99.7 % Al
2O3
)
- seal, antenna system FKM (SHS FPM 70C3 GLT), FFKM (Kalrez 6375), FFKM (Kalrez 2035), FFKM (Kalrez 6230 - FDA)
- Standpipe 316L, Hastelloy C22
Materials, non-wetted parts
- Plastic housing plastic PBT (Polyester)
- Aluminium die-casting housing Aluminium die-casting AlSi10Mg, powder-coated -
basis: Polyester
- Stainless steel housing 316L
- Seal between housing and housing
cover
NBR (stainless steel housing, precision casting), silicone (aluminium/plastic housing; stainless steel housng, electropolished)
- Inspection window in housing cover
(optional)
Polycarbonate
- Ground terminal 316L
Ohmic contact Between ground terminal, process tting and
standpipe
Process ttings
- Pipe thread, cylindrical (ISO 228 T1) G¾ A
- American pipe thread, tapered ¾ NPT, 1½ NPT
- Flanges DIN from DN 25, ANSI from 1"
Weights
- Instrument (depending on housing,
process tting and antenna)
approx. 2 15.4 kg (4.409 33.95 lbs)
- Antenna extension 1.6 kg/m (1.157 lbs/ft)
Length antenna extension max. 5.85 m (19.19 ft)
Torque for NPT cable glands and Conduit tubes
- Plastic housing max. 10 Nm (7.376 lbf ft)
- Aluminium/Stainless steel housing max. 50 Nm (36.88 lbf ft)
Input variable
Measured variable
46 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 47
The measured quantity is the distance between process tting of the sensor and product surface. The reference plane is the seal surface on the hexagon or the lower side of the ange.
3
4
2
1
Fig. 24: Data of the input variable
1 Reference plane 2 Measured variable, max. measuring range 3 Antenna length 4 Useful measuring range
Max. measuring range 35 m (114.83 ft)
Useful measuring range According to the respective standpipe length.
1)
Output variable
Output
- Signal digital output signal, Foundation Fieldbus protocol
- Physical layer according to IEC 61158-2
Damping (63 % of the input variable) 0 999 s, adjustable
Channel Numbers
- Channel 1 Primary value
- Channel 2 Secondary value 1
- Channel 3 Secondary value 2
Transmission rate 31.25 Kbit/s
Current value
1)
Standpipe supply Ex factory up to 5.85 m (19.19 ft)
VEGAPULS 62 standpipe version Foundation Fieldbus 47
11 Supplement
36510-EN-120301
Page 48
- Non-Ex and Ex ia instrument 10 mA, ±0.5 mA
- Ex d instruments 16 mA, ±0.5 mA
Resolution, digital > 1 mm (0.039 in)
Accuracy (according to DIN EN 60770-1)
Process reference conditions according to DIN EN 61298-1
- Temperature +18 … +30 °C (+64 +86 °F)
- Relative humidity 45 75 %
- Air pressure 860 … 1060 mbar/86 106 kPa (12.5 15.4 psig)
Deviation with liquids See following diagrams
1
2
3
0,5 m (1.6 ft)
10 mm (0.394 in)
2 mm (0.079 in)
- 2 mm (- 0.079 in)
- 10 mm (- 0.394 in)
0
Fig. 25: Deviation under reference conditions
1 Reference plane 2 Distance vent hole from reference plane 33 mm (1.30 in) 3 Standpipe end
Reproducibility ≤ ±1 mm
Variables inuencing measurement accuracy
Temperature drift - Digital output ±3 mm/10 K relating to the max. measuring range
or max. 10 mm
Additional deviation through strong, high frequency electromagnetic elds acc. to EN
61326
< ±50 mm
Inuence of the superimposed gas and pressure to the accuracy
The propagation speed of the radar impulses in gas or vapour above the medium is reduced by high pressure. This eect depends on the superimposed gas or vapour and is especially large at low temperatures.
The following table shows the resulting deviation for some typical gases and vapours. The specied values refer to the distance. Positive values mean that the measured distance is too large, negative values that the measured distance is too small.
48 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 49
Gas phase Temperature Pressure
1 bar/
14.5 psig
10 bar/ 145 psig
50 bar/ 725 psig
100 bar/ 1450 psig
200 bar/ 2900 psig
Air 20 °C/68 °F 0.00 % 0.22 % 1.2 % 2.4 % 4.9 %
200 °C/392 °F -0.01 % 0.13 % 0.74 % 1.5 % 3.0 %
400 °C/752 °F -0.02 % 0.08 % 0.52 % 1.1 % 2.1 %
Hydrogen 20 °C/68 °F -0.01 % 0.10 % 0.61 % 1.2 % 2.5 %
200 °C/392 °F -0.02 % 0.05 % 0.37 % 0.76 % 1.6 %
400 °C/752 °F -0.02 % 0.03 % 0.25 % 0.53 % 1.1 %
Steam (satura­ted steam)
100 °C/212 °F 0.26 % - - - -
180 °C/356 °F 0.17 % 2.1 % - - -
264 °C/507 °F 0.12 % 1.44 % 9.2 % - -
366 °C/691 °F 0.07 % 1.01 % 5.7 % 13.2 % 76.0 %
Characteristics and performance data
Frequency K-band (26 GHz technology)
Measuring cycle time
- Standard electronics approx. 450 ms
- Electronics with increased sensitivity
approx.
700 ms
Step response time
2)
3 s
Tracking speed of the measuring window max.
1 m/min
Emitted HF power (depending on the parameter adjustment)
3)
- Average spectral transmission power
density
-14 dBm/MHz EIRP
- Max. spectral transmission power den-
sity
+43 dBm/50 MHz EIRP
- Max. power density in a distance of 1 m < 1 µW/cm²
Ambient conditions
Ambient, storage and transport temperature -40 +80 °C (-40 +176 °F)
2)
Time span after a sudden measuring distance change by max. 0.5 m in liquid applications until the output signal has taken for the rst time 90 % of the nal value (IEC 61298-2).
3)
EIRP: Equivalent Isotropic Radiated Power
VEGAPULS 62 standpipe version Foundation Fieldbus 49
11 Supplement
36510-EN-120301
Page 50
Process conditions
For the process conditions, please also note the specications on the type label. The lower value always applies.
Seal Antenna impedance co-neProcess temperature (measured on the process
tting)
FKM (SHS FPM 70C3 GLT)
PFFE -40 +200 °C (-40 +392 °F)
FFKM (Kalrez 6375) PFFE -20 … +200 °C (-4 … +392 °F)
FFKM (Kalrez 2035) PFFE -15 … +200 °C (5 … +392 °F)
FFKM (Kalrez 6230) PFFE -15 … +200 °C (5 … +392 °F)
Vessel pressure relating to the antenna
system
-1 40 bar/-100 4000 kPa (-14.5 580 psi)
Vessel pressure relating to the ange nom-
inal stage
see supplementary instructions manual "Flanges
according to DIN-EN-ASME-JIS"
Vibration resistance 1 g at 5 … 200 Hz according to EN60068-2-6
(vibration with resonance)
Shock resistance 100 g, 6 ms according to EN60068-2-27 (mechan-
ical shock)
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug
4)
- Single chamber housing l 1 x cable gland M20 x 1.5 (cable: ø 5 9 mm),
1 x blind stopper M20 x 1.5
or:
l 1 x closing cap M20 x 1.5; 1 x blind stopper
M20 x 1.5
or:
l 1 x closing cap ½ NPT, 1 x blind plug ½ NPT
or:
l 1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5
- Double chamber housing l 1 x cable entry M20 x 1.5 (cable: ø 5 9 mm),
1 x blind stopper M20 x 1.5; 1 x blind stopper
M16 x 1.5 or optional5)1 x plug M12 x 1 for external indicating and adjustment unit
or:
4)
Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.
5)
Only with Ex d version
50 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 51
l 1 x closing cap ½ NPT, 1 x blind stopper
½ NPT, 1 x blind stopper M16 x 1.5 or
optionally6)1 x plug M12 x 1 for external indicating and adjustment unit
or:
l 1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5; 1 x blind stopper M16 x 1.5 or optionally7)1 x plug M12 x 1 for external indicating and adjustment unit
Spring-loaded terminals for wire cross-section
- Massive wire, cord 0.2 … 2.5 mm² (AWG 24 … 14)
- Stranded wire with end sleeve 0.2 … 1.5 mm² (AWG 24 … 16)
Electromechanical data - version IP 66/IP 68 (1 bar)
Cable entry
- Single chamber housing 1 x IP 68 cable gland M20 x 1.5; 1 x blind stopper M20 x 1.5
- Double chamber housing 1 x IP 68 cable gland M20 x 1.5; 1 x blind stopper M20 x 1.5; 1 x blind stopper M16 x 1.5
Connection cable
- Wire cross-section 0.5 mm² (AWG 20)
- Wire resistance < 0.036 Ω/m
- Tensile strength < 1200 N (270 lbf)
- Standard length 5 m (16.4 ft)
- Max. length 180 m (590.6 ft)
- Min. bending radius 25 mm (0.984 in) with 25 °C (77 °F)
- Diameter approx. 8 mm (0.315 in)
- Colour - standard PE Black
- Colour - standard PUR Blue
- Colour - Ex-version Blue
Indicating and adjustment module
Voltage supply and data transmission through the sensor
Display LC display in dot matrix
Measured value indication
- Number of digits 5
- Size of digits W x H = 7 x 13 mm (sensors with software 4.0.0,
hardware 2.0.0) W x H = 5 x 8 mm (sensors with software 3.99, hardware < 2.0.0)
6)
Only with Ex d version
7)
Only with Ex d version
VEGAPULS 62 standpipe version Foundation Fieldbus 51
11 Supplement
36510-EN-120301
Page 52
Adjustment elements 4 keys
Protection rating
- unassembled IP 20
- mounted into the sensor without cover IP 40
Materials
- Housing ABS
- Inspection window Polyester foil
Integrated clock
Date format Day.Month.Year Time format 12 h/24 h
Time zone Ex factory CET
Measurement electronics temerature
Resolution 1 °C (1.8 °F) Accuracy ±1 °C (1.8 °F)
Voltage supply
Operating voltage
- Non-Ex instrument 9 32 V DC
- Non-Ex instrument 9 32 V DC
- EEx-ia instrument - Power supply FISCO
model
9 17.5 V DC
- EEx-ia instrument - Power supply EN­TITY model
9 24 V DC
- EEx-d instrument 14 … 32 V DC
Power supply by/max. number of sensors
- H1 power supply max. 32 (max. 10 with Ex)
Electrical protective measures
Protection, depending on housing version
- Plastic housing IP 66/IP 67
- Aluminium housing, stainless steel
housing - investment casting, stainless steel housing - electro-polished
IP 66/IP 68 (0.2 bar)
8)
- Aluminium and stainless housing, in- vestment casting (optionally available)
IP 66/IP 68 (1 bar)
Overvoltage category III
8)
The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.
52 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 53
Protection class II
Approvals
Instruments with approvals can have dierent technical data depending on the version.
That's why the associated approval documents have to be noted with these instruments. They are
part of the delivery or can be downloaded under
www.vega.com via "VEGA Tools" and "serial
number search" as well as via "Downloads" and "Approvals".
VEGAPULS 62 standpipe version Foundation Fieldbus 53
11 Supplement
36510-EN-120301
Page 54
11.2 Supplementary information Foundation Fieldbus
The following table gives you an overview of the instrument versions and the corresponding device descriptions, the electrical characteristics of the bus system as well as the applied function blocks.
Revisions Data DD-Revision Rev_01
CFF-File 010101.c
Device Revision 0101.o
0101.sym
Cff-Revision xx xx 01
Device software revision > 4.4.0
ITK (Interoperability Test Kit) Number 5.0.2
Electricial Characteristics Physicial Layer Type Low-power signaling, bus-
powered, FISCO I.S.
Input Impedance > 3000 Ohms between 7.8 KHz
- 39 KHz
Unbalanced Capacitance < 250 pF to ground from either
input terminal
Output Amplitude 0.8 V P-P
Electrical Connection 2 Wire
Polarity Insensitive Yes
Max. Current Load 10 mA
Device minimum operating voltage 9 V
Transmitter Function Blocks Resource Block (RB) 1
Transducer Block (TB) 1
Standard Block (AI) 3
Execution Time 30 mS
Advanced Function Blocks Discret Input (DI) Yes
PID Control Yes
Output Splitter (OS) Yes
Signal Characterizer (SC) Yes
Integrator Yes
Input Selector (IS) Yes
Arithmetic (AR) Yes
Diagnostics Standard Yes
Advanced Yes
54 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 55
Performance No
Function Blocks Instantiable No
General Information LAS (Link Active Scheduler) Yes
Master Capable Yes
Number of VCRs (Virtual Communication Relationships)
24
Function blocks
Transducer Block (TB)
The Transducer Block "Analog Input (AI)" takes the original measured value (Secondary Value 2), carries out the min./max. adjustment (Secondary Value 1), carries out a linearization (Primary Value) and makes the values on its output available for further function blocks.
m(d)
% Lin%
Primary
Value
Secondary
Value 2
Secondary
Value 1
TB
Sensor_Value
min/max
adjustment
Linearization
Fig. 26: Schematic presentation Transducer Block (TB)
Function block Analog Input (AI)
VEGAPULS 62 standpipe version Foundation Fieldbus 55
11 Supplement
36510-EN-120301
Page 56
The function block "Analog Input (AI)" takes the original measured value selected by a Channel Number and makes it available to additional function blocks on its output.
Fig. 27: Schematic presentation function block Analog Input (AI)
Function block Discret Input (DI)
The function block "Discret Input (DI)" takes the original measured value selected by a Channel Number and makes it available to additional function blocks on its output.
Fig. 28: Schematic presentation function block Discret Input (DI)
Function block PID Control
56 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 57
The function block "PID Control " is a key component for various tasks in the process automation and is used universally. PID blocks can be cascaded if this is necessary or requested due to dierent time constants with the primary and secondary process measurement.
Fig. 29: Schematic presentation function block PID Control
Function block Output Splitter
VEGAPULS 62 standpipe version Foundation Fieldbus 57
11 Supplement
36510-EN-120301
Page 58
The function block "Output Splitter" generates two control outputs out of one input. Each output is a linear image of a part of the input. A retrograde calculation function is realised by using the linear imaging function inversely. A cascading of several Output Splitters is supported by an integrated decision table for the combinability of inputs and outputs.
Fig. 30: Schematic presentation function block Output Splitter
Function block Signal Characterizer
58 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 59
The function block "Signal Characterizer" has two channels the outputs of which are not in linear relation with the respective input. The non-linear relation is dened by a look-up table with individually selectable x/y-pairs. The respective input signal is imaged on the corresponding output, hence this function block can be used in a control loop or signal path. Optionally the function axis can be exchanged in channel 2 so that the block can be also used in a reverse control loop.
Fig. 31: Schematic presentation function block Signal Characterizer
Function block Integrator
VEGAPULS 62 standpipe version Foundation Fieldbus 59
11 Supplement
36510-EN-120301
Page 60
The function block "Integrator" integrates a continuous input signal over the time and sums the results of an impulse input block. It is used as a totalizer up to a reset or as a subtotalizer up to a reference point at which the integrated and accumulated value is compared with the default values. When these default values are reached, digital output signals will be outputted. The integration function is carried out upwardly starting with zero and downwards with a default value. Two ow values are also available so that the net ow volume can be calculated and integrated. This can be used for calculation of volume and mass changes in the vessel or for optimisation of ow controls.
Fig. 32: Schematic presentation function block Integrator
Function block Input Selector
60 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 61
The function block "Input Selector" oers selection possibilities for up to four inputs and generates an output signal according to the selection criteria. Typical input signals are AI blocks. Selection possibilities are maximum, minimum, mean value, average value and rst useful signal. Through parameter combination, the block can be used as rotary switch or as preselection switch for the rst useful value. Switch information can be received by other input blocks or the user. Mean value selection is also supported.
Fig. 33: Schematic presentation function block Input Selector
Function block Arithmetic
The function block "Arithmetic" allows the simple integration of usual metrological calculation functions. The user can select the requested algortihm according to the name without known the formula.
The following algorithms are available:
l Flow compensation, linear l Flow compensation, square root l Flow compensation, approximate l BTU ow l Traditional Multiply Divide l Average l Traditional Summer l Fourth order polynomial l Simple HTG compensated level
VEGAPULS 62 standpipe version Foundation Fieldbus 61
11 Supplement
36510-EN-120301
Page 62
l Fourth order Polynomial Based on PV
Fig. 34: Schematic presentation function block Arithmetic
Parameter list
The following table gives you an overview of the parameters used.
FF desciptor Description Unit
PRIMARY_VALUE PRIMARY_VALUE (Linearized value). This is the process value
after min/max adjustment and Linearization with the status of the transducer block. The unit is dened in "PRIMAR-
Y_VALUE_UNIT"
PRIMARY_VALUE_UNIT Selected unit code for "PRIMARY_VALUE"
SECONDARY_VALUE_1 This is the measured value after min/max adjustment with the
status of the transducer block. The unit is dened in
"SECONDARY_VALUE_1_UNIT"
SECONDARY_VALUE_1_U- NIT
Selected unit code for "SECONDARY_VALUE_1"
SECONDARY_VALUE_2 This is the distance value ("sensor_value") with the status of
the transducer block. The unit is dened in "SECONDA-
RY_VALUE_2_UNIT"
FILL_HEIGHT_VALUE Filling height. The unit is dened in "FILL_HEIGHT_VA-
LUE_UNIT"
FILL_HEIGHT_VALUE_UNIT Filling height unit
CONST_VALUE Constant value
SECONDARY_VALUE_1_- TYPE
Secondary value 1 type
62 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 63
FF desciptor Description Unit
SECONDARY_VALUE_2_- TYPE
Secondary value 2 type
FILL_HEIGHT_VALUE_Type Filling height value type
DIAGNOSIS AITB Diagnosis
DIAG_MASK_1
DIAG_OUT_1
DIAG_MASK_2
DIAG_OUT_2
DEVICE_IDENTIFICATION Manufacturer ID, device type, bus type ID, measurement
principle, serial number, DTM ID, device revision
DEVICE_NAME Device name
IS-SPARE_ELECTRONICS Device name
DEVICE_VERSION_INFO Hard- and software version for system, function and error
CALIBRATION_DATE Day, month and year
FIRMWARE_VERSION_ASCII Software version
HW_VERSION_ASCII Hardware version
ADJUSTMENT_DATA Min./max.-adjustment physical, percent and oset
FIRMWARE_VERSION_MAIN Firmware versions major, minor, revision and build
PHYSICAL_VALUES Distance, distance unit, distance status, level and status
DEVICE_UNITS Distance and temperature units of the instrument
APPLICATION_CONFIG Medium type, media, application type, vessel bottom, vessel
height
LINEARIZATION_TYPE_SEL Type of linearization
SIMULATION_PHYSCAL
INTEGRATION_DATA Physical oset and integration time
DEVICE_CONFIG_PULS_RA- DAR
Electronics variant, probe type, max. measuring range, antenna
extension length, adjustment propagation antenna extension lprapproval configuration
ADJUSTMENT_LIMITS_MIN Min. range min.-/max.- values physical, percent, oset
ADJUSTMENT_LIMITS_MAX Max. range min.-/max.- values physical, percent, oset %
FALSE_SIGNAL_COMMAND %
FALSE_SIGNAL_CMD_C­REATE_EXTEND
FALSE_SIGNAL_CMD_DE­LET_REGION
VEGAPULS 62 standpipe version Foundation Fieldbus 63
11 Supplement
36510-EN-120301
Page 64
FF desciptor Description Unit
FALSE_SIGNAL_CMD_STA-TEBusy, last command, errorcode
FALSE_SIGNAL_CMD_CON­FIGURATION1
Amplitude safety of the 0 % curve, safety of the false signal
suppression, position of the 0 % and 100 % curve in near and far range
FALSE_SIGNAL_CMD_CON­FIGURATION2
Gradient of the manual sectors, safety at the end of false echo
memory and depending on the import range gating out the false signals
ECP_CURVE_AVARAGING_- CONFIG
Averaging factor on increasing and decreasing amplitude
LEVEL_ECHO_MEASURE­MENT
Function measured value lter
ECHO_CURVE_STATUS
PACKET_COUNT
GU_ID_END
ECHO_CURVE_READ Echo curve data
ECHO_EVALUATOR Echo parameters, rst large echo, amplitude threshold rst
large echo
ECHO_DECIDER Echo selection criteria, fault signal on loss of echo, delay on
fault signal on loss of echo
DISPLAY_SETTINGS Indication value, menu language, lightning
SIL_MODE
EDENVELOPE_CURVE_FIL- TER
Parameters of envelope curve lter, activation of smooth raw
value curve
EDDETECTION_CURVE_FIL- TER
Parameters of the detection lter, oset threshold value curve
EDECHO_COMBINATION Parameters for echo combination, function combine echoes,
amplitude dierence of combined echoes, position dierence of combined echoes
LIN_TABLE_A…LIN_TABLE_-Q32 couples of percentage and lin. percentage values
ELECTRONICS_INFORMATI-ONElectronics version
APPLICATION_CONFIG_- SERVICE
Limitation measuring range begin, safety of measuring range
end
LEVEL_ECHO_INFO Level echo ID, amplitude, measurement safety
DEVICE_STATUS Device status
FALSE_SIGNAL_LIMITS False signal distance min./max.
USER_PEAK_ELEC_TEMP Min.-/max.- values of electronics temperature, date
64 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 65
FF desciptor Description Unit
USER_MIN_MAX_PHYSI­CAL_VALUE
Min.-/max.- distance values, date
RESET_PEAK_PHYSICAL_- VALUE
RESET_LINEARIZATION_- CURVE
DEVICE_STATUS_ASCII Device status
ECHO_CURVE_PLICSCOM_- REQUEST
Parameters as curve selection and resolution
ECHO_CURVE_PLICSCOM_- LIMITS
Parameters as start and end
APPROVAL_WHG Sensor acc. to WHG
DEVICE_STATE_CONFIG Function check, maintenance required, out of specication
ELECTRONIC_TEMPERATU-REElectronics temperature
RESET_PEAK_ELECTRO­NIC_TEMP
FOCUS_RANGE_CONFIG Width focusing range, time for opening the focusing range, min.
measurement reliability in and outside the focusing range
NOISE_DETECTION_INFO Increase of the system noise
NOISE_DETECTION_CON- FIG
System noise treatment
ECHO_MEM_SAVE_CUR­VE_TYPE
ECHO_MEM_STATE Busy, curve type, error code
VEGAPULS 62 standpipe version Foundation Fieldbus 65
11 Supplement
36510-EN-120301
Page 66
11.3 Dimensions
The following dimensional drawings represent only an extract of the possible versions. Detailed dimensional drawings can be downloaded on
www.vega.com under "Downloads" and "Drawings".
Plastic housing
~ 69 mm
(2.72")
ø 79 mm
(3.03")
112 mm (4.41") M20x1,5/ ½ NPT
~ 84 mm (3.31")
M16x1,5
112 mm (4.41")
M20x1,5/ ½ NPT
1
2
ø 79 mm
(3.31")
Fig. 35: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated indicating and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version 2 Double chamber version
Aluminium housing
21
ø 86 mm (3.39")
~ 116 mm (4.57")
116 mm (4.57")
M20x1,5
M20x1,5/ ½ NPT
~ 87 mm (3.43")
M16x1,5
ø 86 mm
(3.39")
120 mm (4.72")
M20x1,5/ ½ NPT
Fig. 36: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated indicating and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version 2 Double chamber version
66 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 67
Aluminium housing in protection rating IP 66/IP 68, 1 bar
120 mm (4.72")
~ 105 mm (4.13")
116 mm (4.57")
~ 150 mm (5.91")
ø 84 mm (3.31")
ø 84 mm
(3.31")
M20x1,5/ ½ NPT
M16x1,5
2
M20x1,5M20x1,5
1
Fig. 37: Housing versions in protection IP 66/IP 68 (1 bar) - with integrated indicating and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version 2 Double chamber version
Stainless steel housing
~ 69 mm
(2.72")
ø 79 mm
(3.11")
117 mm (4.61")
M20x1,5/ ½ NPT
~ 59 mm
(2.32")
ø 80 mm
(3.15")
112 mm (4.41")
M20x1,5/ ½ NPT
~ 87 mm (3.43")
ø 86 mm
(3.39")
120 mm (4.72")
M20x1,5/ ½ NPT
M16x1,5
321
Fig. 38: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated indicating and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version, electropolished 2 Single chamber version, precision casting 2 Double chamber version, precision casting
VEGAPULS 62 standpipe version Foundation Fieldbus 67
11 Supplement
36510-EN-120301
Page 68
Stainless steel housing in protection rating IP 66/IP 68, 1 bar
117 mm (4.61")
~ 103 mm
(4.06")
ø 77 mm
(3.03")
120 mm (4.72")
~ 105 mm (4.13")
ø 84 mm
(3.31")
~ 93 mm
(3.66")
ø 80 mm
(3.15")
112 mm (4.41")
M20x1,5/ ½ NPT
M20x1,5
3
M20x1,5/ ½ NPT
M16x1,5
21
Fig. 39: Housing versions in protection IP 66/IP 68 (1 bar) - with integrated indicating and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version, electropolished 2 Single chamber version, precision casting 2 Double chamber version, precision casting
68 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 69
VEGAPULS 62 - S tandpipe version with thread
ø 21,3 mm
(0.84")
ø 21,3 mm
(0.84")
ø 39,5 mm
(1.56")
ø 39,5 mm
(1.56")
224 mm (8.82")
16 mm (0.63")
32 mm (1.26")
32 mm (1.26")
ø 3,5 mm (0.14")
20 mm (0.79")
22 mm (0.87")
224 mm (8.82")
SW 46 mm
(1.81")
SW 32 mm
(1.26")
L
L
G¾A
¾ NPT
G1½A
1½ NPT
Fig. 40: VEGAPULS 62 - Standpipe version with thread
9)
9)
The plant operator must provide a suitable support for the standpipe de­pending on the length and the process conditions.
VEGAPULS 62 standpipe version Foundation Fieldbus 69
11 Supplement
36510-EN-120301
Page 70
VEGAPULS 62 - S tandpipe version with ange
ø 39,5 mm
(1.56")
ø 21,3 mm
(0.84")
203 mm (7.99")
33 mm (1.30")
L
Fig. 41: VEGAPULS 62 - Standpipe version with ange
10)
10)
The plant operator must provide a suitable support for the standpipe de­pending on the length and the process conditions.
70 VEGAPULS 62 standpipe version Foundation Fieldbus
11 Supplement
36510-EN-120301
Page 71
11.4 Industrial property rights
VEGA product lines are global protected by industrial property rights. Further information see http://www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter http://www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle. Pour plus d'informations, on pourra se référer au site http://www.vega.com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad industrial. Para mayor información revise la pagina web http://www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность. Дальнейшую информацию смотрите на сайте http://www.vega.com.
VEGA 站<http://www.vega.com>
11.5 Trademark
All the brands as well as trade and company names used are property of their lawful proprietor/originator.
VEGAPULS 62 standpipe version Foundation Fieldbus 71
11 Supplement
36510-EN-120301
Page 72
INDEX
A
Accessory
- External indicating and adjustment unit 11
- External radio unit 11
- Flanges 11
- Indicating and adjustment module 10
- Interface adapter 11
- Protective cap 11
Adjustment 24
- Max. adjustment 25
- Min. adjustment 24
Adjustment system 22
C
Cable entry 13 Connection procedure 14 Connection technology 14 Curve indication
- Echo curve 26
- False signal suppression 26
D
DD (Device Description) 34
E
Echo curve memory 35 Echo curve of the setup 26 EDD (Enhanced Device Description) 34 Error messages 36 Event memory 35
F
False signal suppression 27 Fault rectication 40 FF parameter 62 Function blocks
- Analog Input (AI) 55
- Arithmetic 61
- Discret Input (AI) 56
- Input Selector 60
- Integrator 59
- Output Splitter 57
- PID Control 56
- Signal Characterizer 58
- Transducer Block (TB) 55
Functional principle 10
K
Key function 22
L
Linearisation curve 28
M
Main menu 23 Meas. reliability 25 Measured value memory 35 Measurement error 40 Medium
- Bulk solid 23
Moisture 12
N
NAMUR NE 107
- Failure 37
- Function check 38
- Maintenance 39
- Out of specication 39
O
Overll protection according to WHG 29
P
Packaging 10 Peak value 25
R
Reection properties 24 Repair 44 Replacement parts
- Electronics module 11
S
Service hotline 42 Status messages 36 Storage 10
T
Type label 8
72 VEGAPULS 62 standpipe version Foundation Fieldbus
Index
36510-EN-120301
Page 73
VEGAPULS 62 standpipe version Foundation Fieldbus 73
Index
36510-EN-120301
Page 74
74 VEGAPULS 62 standpipe version • Foundation Fieldbus
Index
36510-EN-120301
Page 75
VEGAPULS 62 standpipe version Foundation Fieldbus 75
Index
36510-EN-120301
Page 76
VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com
www.vega.com
Printing date:
ISO 9001
All statements concerning scope of delivery, application,
practical use and operating conditions of the sensors and
processing systems correspond to the information avail-
able at the time of printing.
© VEGA Grieshaber KG, Schiltach/Germany 2012
Subject to change without prior notice 36510-EN-120301
Loading...