1.1Contents of the instruction manual .....................................................................................................................4
1.2Safety information ...............................................................................................................................................4
2 Product description
2.1Function and configuration ..................................................................................................................................4
2.2Types and versions .............................................................................................................................................5
2.3Configuration of a measuring system, sensor in not-Ex-area .............................................................................8
2.4Configuration of a measuring system, sensor in Ex-area ................................................................................... 9
2.5Order code for sensors .....................................................................................................................................10
2.6Technical data of the sensors............................................................................................................................11
2.7Configuration with compact instrument .............................................................................................................14
2.8Order code for compact instrument...................................................................................................................15
2.9Technical data of compact instruments .............................................................................................................16
3.1Installation recommendations relating to liquid tank / vessels ..........................................................................23
3.2Installation recommendations relating to solid silos ..........................................................................................25
4 Electrical connection
4.1Connection example of sensors ........................................................................................................................26
4.2Connection example of Ex-sensors ..................................................................................................................27
4.3Connection example of compact instruments ...................................................................................................29
5 Set-up
5.1Sensor and signal conditioning instrument (e.g. VEGAMET) ...........................................................................30
5.2Sensor and processing system (VEGALOG 571) .............................................................................................30
The Technical Information / Operating Instructions is
called TIB. It contains all necessary information for correct
- installation
- connection
- set-up
- optimization
of the pulse-echo-sensors VEGASON …. V… or pulseecho compact instrument VEGASON … K….
VEGA regularly revises the contents of TIBs as technical
improvements are made to the instruments.
1.2Safety information
The described module must only be inserted and operated
as described in this TIB. Please note that other action can
cause damage for which VEGA does not take
responsibility.
2Product description
2.1Function and configuration
Function
The VEGASON … sensor systems or compact
instruments are used for continuous and non-contact level
measurement.
Short sound impulse packets are emitted by the
transducer of the pulse-echo sensor. The combined
emitter and receiver system detects the pulses reflected
by the product.
The running period of the ultrasonic pulses is directly proportional to the distance between sensor and measured
product.
The signal processing ECHOFOX
of signal analysis to process the reflected ultrasonic
pulses.
®
uses modern methods
The signal-to noise ratio is optimized by digital signal
processing methods with mathematic algorithms (DSP).
The automatic configuration of a user specific data base
enables an optimum adaption to the respective
application.
The use of fuzzy-logic for echo evaluation ensures a
reliable determination of the level. False echoes, e.g. from
stirrers or other installations are detected and are not
included in the generation of the measured value.
These measuring data are transmitted digitally to
respective signal conditioning instruments or processing
systems and processed.
The operation of the sensors is made via the signal
conditioning instrument VEGAMET or via a PC
(VEGACONNECT necessary).
Compact instruments are additionally equipped with an
integral output of the measuring data and make a current
output 0 … 20 mA available.
The operation of the compact instrument is only possible
with PC via VEGACONNECT.
Configuration
Flange version
The sensor or compact instruments consist of a housing,
integrated therein the above electronics and all required
terminals, a mounting flange and a transducer.
Housing, flange and transducer are designed acc. to the
type.
Separate version
The separate version consists of a housing with integral
electronics and all necessary terminal and mounting plate.
Transducer with mounting tube and connection cable as
separate modular unit. The transducers differ acc. to the
various types.
4
VEGASON 83 … 87
2 Product description
2.2Types and versions
VEGASON 83 FK
VEGASON 83 FV
Flange version with flange size DN 100, transducer integral in flange, plastic housing with integral ECHOFOX
sensor electronics.
Suitable for a number of applications and mounting
positions,
- mounting directly on the vessel ceiling (tank, silo) or on
existing sockets
- positioning on fix points above the vessels etc.
Measuring range 0,7 m … 10 m
(see section 3 Evironment).
VEGASON 83 FV Ex, 83 FV Ex 0
®
-
VEGASON 84 FK
VEGASON 84 FV
Flange version with flange size DN 150, transducer below
the flanges, plastic housing with integral ECHOFOX
sensor electronics.
Suitable for a number of applications and mounting
positions,
- mounting directly on the vessel ceiling (tank, silo) or on
existing sockets
- positioning on fix points above the vessel etc.
Measuring range 0,7 m … 20 m
(see section 3 Environment).
®
-
VEGASON 84 FK StEx
VEGASON 84 FV StEx
as VEGASON 83 FV, however housing of sensor
electronics and transducer made of stainless steel 1.4305
Suitable for the use in hazardous areas (see section 2.11
Approvals).
VEGASON 83 FV Ex B, 83 FV Ex0 B
as VEGASON 83 FV Ex or FV Ex 0, however with integral
overvoltage arrester of sensor electronics.
as VEGASON 84 FK or FV , however additionally approved
for the use in dust-Ex areas StEx (see section 2.11
Approvals).
VEGASON 84 FV Ex, 84 FV Ex 0
Flange version with flange size DN 150, transducer integral in flange, housing of sensor electronics and
transducer made of stainless steel 1.4305.
Suitable for the use in hazardous areas (see section 2.11
Approvals).
VEGASON 83 … 87
5
2 Product description
VEGASON 84 FV Ex B, 84 FV Ex0 B
as VEGASON 84 FV Ex or FV Ex 0, however with integral
overvoltage arrester of sensor electronics.
Compact instrument
VEGASON 85 FK
VEGASON 85 FV
VEGASON 87 FK
VEGASON 87 FV
Flange version, flange size ø 450 mm, triple transducer
divided into emission and receipt unit, reset as described
under VEGASON 85 FV .
Measuring range 0,4 m … 60 m
(see section 3 Environment).
VEGASON 87 FK StEx
VEGASON 87 FV StEx
Flange version with flange size DN 150, transducer below
the flanges, plastic housing with integral ECHOFOX
sensor electronics.
Suitable for a number of applications and mounting
positions,
- mounting directly on the vessel ceiling (tank, silo) or on
existing sockets
- positioning on fix points above the vessels etc.
Measuring range 1,0 m … 30 m
(see section 3 Environment).
®
-
VEGASON 85 FK StEx
VEGASON 85 FV StEx
as VEGASON 85 FK or FV , however additionally approved
for the use in dust-Ex areas StEx (see section 2.11
Approvals).
as VEGASON 87 FK or FV , however additionally approved
for the use in dust-Ex areas StEx (see section 2.11
Approvals).
6
VEGASON 83 … 87
2 Product description
VEGASON 83 GK
VEGASON 83 GV
Separate version, transducer and housing are separated.
An ECHOFOX
housing. The transducer is provided with a fixing tube
(thread G 1). It can be mounted in a small hole (ø 35 mm)
at the vessel ceiling (tank/silo).
As option the transducer can be screwed with a swivelling
holder (DN 150) and can be optimally adapted to the
installation conditions.
Measuring range 0,7 m … 10 m
(see section 3 Environment).
®
-sensor electronics is integrated in the
VEGASON 84 GK
VEGASON 84 GV
as VEGASON 83 GKor GV, however
measuring range 0,7 m … 20 m
VEGASON 84 GK StEx
VEGASON 84 GV StEx
as above, however additionally approved for the use in
dust-Ex areas StEx (see section 2.11 Approvals).
VEGASON 85 GK
VEGASON 85 GV
as VEGASON 83 GK or GV , however
measuring range 1,0 m … 30 m
VEGASON 85 GK StEx
VEGASON 85 GV StEx
as above, however additionally approved for the use in
dust-Ex areas StEx (see section 2.11 Approvals).
VEGASON 87 GK
VEGASON 87 GV
as VEGASON 83 GK or GV, however triple transducer
divided into emission and receipt unit, rest as described
under
measuring range 0,4 m … 60 m
VEGASON 87 GK StEx
VEGASON 87 GV StEx
as above, however additionally approved for the use in
dust-Ex areas StEx (see section 2.11 Approvals).
6
VEGASON 83 … 87
7
2 Product description
11
11 … 15
6 … 10
1 … 5
4
2.3Configuration of a measuring system, sensor in not-Ex-area
Generally a measuring system consists of two components
- a local sensor and
- a signal conditioning instrument or processing system
The kind of signal conditioning instrument or processing system depends on the installation conditions.
Sensor(s)
VEGASON 83 … 87
Signal conditioning
instrument e.g.
VEGAMET 514 V
(VEGAMET 407 V)
Processing system
VEGALOG 571
VEGACONNECT
RS
232
PC
EV -input card
(via additionally EV-input cards max. 3 x 5 = 15 sensors
each can be connected)
CPU-central unit
Sensor(s) and signal conditioning instrument or processing system are connected via a two-wire line. The power supply
of the sensor or the sensors is ensured via this line and the measuring data are transmitted digitally to the signal
conditioning instrument or processing system.
The indicating and operating surface VEGA Visual Operating, installed on a PC ensures a simple and comfortable
configuration and parameter adjustment of the respective measuring system.
VEGACONNECT can be connected to respective sockets of the signal conditioning instrument or the sensor. If necessary, it is possible to connect VEGACONNECT directly to the two-wire line (signal conditioning instrument … sensor).
8
VEGASON 83 … 87
2 Product description
2.4Configuration of a measuring system, sensor in Ex-area
In this application a measuring system consists of three components
- a local Ex-sensor in the Ex-area
- a safety barrier and
- a signal conditioning instrument or processing system in the not-Ex-area
The kind of signal conditioning instrument or processing system mainly depends on the control and process technical
requirements.
Sensor(s)
VEGASON 83 und 84
Separator Type 146 or
VEGATRENN 546
Signal conditioning
instrument e.g.
VEGAMET 514 V
(VEGAMET 407 V)
4
1
1
1
Not-Ex-area
Ex-area
Processing system
VEGALOG 571
RS
232
1 … 5
4
1 … 5
6 … 10
4
44
6 … 10
usw.
etc.
etc.
1. EV-input card2. EV-input card
VEGACONNECT
(via additionally EV-input card 5 sensors each can be connected)
CPU-central unit
4
PC
Each sensor is connected to the respective signal conditioning instrument or processing system via a separator type 146
or VEGATRENN 546. The separator provides two intrinsically safe circuits. These circuits provide the power supply of the
sensor and the digital transmission of measuring data to the signal conditioning instrument or processing system.
The indicating and operating surface VEGA Visual Operating, installed on a PC ensures a simple and comfortable
configuration and parameter adjustment of the respective measuring system.
VEGASON 83 … 87
9
2.5Order code for sensors
2.5.1 VEGASON …V / …V StEx
TypeMeasuring range
8 3.................. 0,7 … 10 m
8 4.................. 0,7 … 20 m
8 5.................. 1,0 … 30 m
8 7.................. 0,4 … 60 m
Version
F Flange version
G Separate version
Approvals
. X without
EXS. XStEx for Zone 10 (only for VEGASON 84 … 87)
2 Product description
Flange size
E DN 100 PN 16 (for VEGASON 83)
F DN 150 PN 16 (for VEGASON 84 and 85)
U ø 450 (for VEGASON 87)
Y Others
Flange material
A Aluminium (for StEx-version)
P PPh
Y Others
S O NVE ZOrder no. for VEGASON … V
2.5.2 VEGASON … FV Ex…
TypeMeasuring range
8 3 F V E XCENELEC EEx d e ia IIB T6 .......... 0,7 … 10 m
8 3 F V E X 0 PTB Zone 0 EEx d e ia IIB T6 ........ 0,7 … 10 m
8 4 F V E XCENELEC EEx d e ia IIB T6 .......... 1,1 … 20 m
8 4 F V E X 0 PTB Zone 0 EEx d e ia IIB T6 ........ 1,1 … 20 m
Options
X without
B with overvoltage arrester
S O N. X V V VOrder no. for VEGASON … FV Ex…
10
VEGASON 83 … 87
2 Product description
2.6Technical data of the sensors
2.6.1 VEGASON 83 … 87 FV, 84 … 87 FV StEx
VEGASON 83 … 87 GV , 84 … 87 GV StEx
Approvals
see section 2.11 Approvals
Power supply
via signal conditioning instrumentVEGAMET … with digital transmission of measuring data
via processing systemVEGALOG 571 with EV-input cards
Operating voltagemax. 36 V DC
Power consumptionmax. 200 mA
Measuring range
VEGASON83 FV83 GV84 …85 …87 …
Min. distance forliquids and
solidsgranulation ≥ 5 mm 0,7 m0,5 m0,7 m1,0 m0,4 m
granulation ≤ 5 mm 0,8 m0,6 m1,0 m1,2 m0,4 m
Max. distance dependent on product and process10 m10 m20 m30 m60 m
Measuring data
Measuring resolution1 mm1 mm1 mm1 mm1 mm
Measuring frequency33 kHz33 kHz22 kHz16 kHz16 kHz
Measuring rate0,6 sec.0,6 sec.0,7 sec.1,0 sec.1,5 sec.
Radiation angle related to –3 dB12°12°12°15°3,5°
Error limits
Linearity error relating to the adjustment< 0,1 % of measuring range
Temperature error of the electronics0,015 %/10 K of measuring range
Materials
VEGASONFVGVStEx
Housing and cover of electronicsPBTPBTPBT
FlangePPh–––St 37 galv. or Alu
Transducer housingPVDFPVDFPVDF
Impedance adapterPEPEPE
Transducer fixing tube
- Type83–––PVDFSt 37 galvanized
- T ype 84 and 85–––RCH 1000St 37 galvanized
- T ype 87–––St 37 galvanizedSt 37 galvanized
Dimensions and weights
Flange size- Type 83DN 100 PN 16–––DN 100 PN 16
- Type 84 and 85DN 150 PN 16–––DN 150 PN 16
- Type 87ø 450–––ø 450
Thread of the fixing tube–––G 1 A–––
Total weight of sensors
- Type 83 … 85appr. 6 kgappr. 6 kgappr. 6 kg
- Type 87appr. 12 kgappr. 12 kgappr. 12 kg
Temperature reaction
VEGASON … StEx-version
At an ambient temperature40°C the max. temperature to be adjusted
- on the transducer in Zone 1045°C is reached and
- on the housing (electronics) in Zone 1155°C is reached
VEGASON 83 … 87
11
2 Product description
Ambient conditions
VEGASONFVGVStEx
Ambient temperature related to
- transducer in Zone 10-20°C … +80°C-20°C … +80°C-20°C … +75°C
- housing (electronics) in Zone 11-20°C … +60°C-20°C … +60°C-20°C … +60°C
Storage and transport temperature-20°C … +80°C-20°C … +80°C-20°C … +80°C
Protection
- of transducerIP 67IP 67IP 65
- the housing (electronics) generallyIP 67IP 67IP 65
certified
- transducer––––––IP 65
- housing––––––IP 54
Overvoltage classIIIIIIIII
Protection classIIIIII
Max. vessel pressure related to
- Type 830,5 bar1,0 bar–––
- T ype 84 and 850,5 bar0,5 bar0,5 bar
- Type 870,3 bar0,5 bar0,5 bar
Connection line
VEGASON … FV-version
Sensor to signal conditioning instrument /
processing system2 pole, maximal 20 Ohm/line
VEGASON … GV-version
Transducer to sensor electronics
- Type 83 … 85standard coax cable type RG 58
standard length 5 m
max. length 300 m
- Type 874 x 0,38 mm
2
, 1 screened wire
standard length 5 m, max. length 35 m
cable diameter 8 mm
Sensor electronics to signal conditioning
instrument / processing system2 pole, max. 20 Ohm/line
Temperature sensoreach integrated in the transmitter
with GV-version via connected connection line
above connection line
Electrical connection
Terminals for the connection linesfor max. 1,5 mm
Terminals for the earth connectionfor max. 4,0 mm
2
2
Cable entry1 x Pg 13,5 for all FV-versions
2 x Pg 13,5 for all GV-versions
12
VEGASON 83 … 87
2 Product description
2.6.2 VEGASON 83 and 84 Ex, Ex0
VEGASON 83 and 84 Ex B, Ex0 B
Approvals
see section 2.11 Approvals
Power supply
via separatorType 146 or VEGATRENN 546
Supply via2 intrinsically safe circuits of category ib IIB
Flame proofingEEx de ia IIB T6 (see also section 2.11)
Measuring range
VEGASON8384
Min. distance of liquids0,7 m1,1 m
Max. distance dependent on product and process10 m20 m
Measuring data
Measuring resolution1 mm1 mm
Measuring frequency33 kHz22 kHz
Measuring rate1,1 sec.1,5 sec
Radiation angle related to –3 dB12°12°
Error limits
Linearity error relating to the adjustment< 0,1 % of measuring range
Temperature error of the electronics0,015 %/10 K of measuring range
Materials
Housing of electronics1.4305
Flange1.4305
Diaphragm1.4531
Flange sealingViton
Dimensions and weights
Flange size- Type 83DN 100 PN 16
- Type 84DN 150 PN 16
Total weight of the sensorsat DN 100 approx. 17 kg
at DN 150 approx. 21 kg
Ambient conditions
Max. permissible ambient temperature
- on the transducer+80°C
- on the housing (electronics)+60°C
Storage and transport temperature-20°C … +80°C
ProtectionIP 67
Max. vessel pressure
- Type 832 bar
- Type 841 bar
Vessel pressures and ambient temperature
under Ex-conditionssee respective conformity certificate
Connection line
Sensor … separator2 x 2 pole
Resistance per conductormax. 7,5 Ohm
Temperature sensor and its connectionintegrated int he respective sensor
Electrical connection
Terminals for the connection linemax. 1,5 mm
Terminals for earth connectionmax. 4,0 mm
Cable entry1 x Pg 13,5
VEGASON 83 … 87
2
2
13
2 Product description
2.7Configuration with compact instrument
As described under "2 Product description" the compact instruments are equipped with an integral processing of
measuring data. A current output 0 … 20 mA is available.
First of all each compact instrument must have a voltage supply. The current output can be directly connected to
indicators or process control systems.
VEGASON 83 … 87
Power supply
VEGACONNECT
11
2
Ri
during the
adjustment
Ri = min. 100 Ohm
34
usw.
Processing
one current output 0 … 20 mA each
(e.g. for consumers with low inner resistance)
PC
The indicating and operating surface VEGA Visual Operating, installed on a PC ensures a simple and comfortable
configuration as well as operator supporting parameter adjustment of the compact instruments.
Therefore it is necessary to connect the computer via VEGACONNECT directly with the respective compact instruments.
A digital signal is superimposed to the current output. The adjustment of the sensor can be also carried out via the
current output. For indication and processing systems with an inner resistance of less than 100 Ohm a respective series
resistor must be connected for the time of adjustment.
14
VEGASON 83 … 87
2 Product description
2.8Order code for compact instrument
2.8.1 VEGASON …K / …K StEx
TypeMeasuring range
8 3.................. 0,7 … 10 m
8 4.................. 0,7 … 20 m
8 5.................. 1,0 … 30 m
8 7.................. 0,4 … 60 m
Version
F Flange version
G Separate version
Approvals
. X without
E X S . X StEx for Zone 10 (only for VEGASON 84 … 87)
Power supply
E 110 V AC
A 230 V AC
I 24 V AC, 16 … 36 V DC
X Others
Flange size
E DN 100 PN 16 (for VEGASON 83)
F DN 150 PN 16 (for VEGASON 84 and 85)
U ø 450 (for VEGASON 87)
Y Others
VEGASON83 FK83 GK84 …85 …87 …
Min. distance of liquids
and solidsgranulation ≥ 5 mm0,7 m0,5 m0,7 m1,0 m0,4 m
granulation ≤ 5 mm0,8 m0,6 m1,0 m1,2 m0,4 m
Max. distance dependent on product and process10 m10 m20 m30 m60 m
Measuring data
Measuring resolution1 mm1 mm1 mm1 mm1 mm
Measuring frequency33 kHz33 kHz22 kHz16 kHz16 kHz
Measuring rate0,6 sec.0,6 sec.0,7 sec.1,0 sec.1,5 sec.
Radiation angle related to –3 dB12°12°12°15°3,5°
24 V AC (–15 %, +10 %), 50/60 Hz, 10 VA
nenn
24 V DC (16 … 36 V), 8 W
230 V AC, (–10 %, +5 %), 50/60 Hz, 10 VA
nenn
110 V, 130 V, 240 V AC (–15 %, +10 %),
50/60 Hz, 10 VA
Current output
Range0 … 20 mA
Loadmax. 400 Ohm
Resolution of D/A-conversion0,025 % of adjustment
Error limits
Linearity error relating to the adjustment< 0,1 % of measuring range
Temperature error of the electronics0,015 %/10 K of measuring range
Materials
VEGASONFKGKStEx
Housing and cover of the electronicsPBTPPBTPPBTP
FlangePPh–––St 37 galv. or Alu
Transducer housingPVDFPVDFPVDF
Impedance adapterPEPEPE
Transducer fixing tube
- Type 83–––PVDFSt 37 galvanized
- T ype 84 and 85–––RCH 1000St 37 galvanized
- Type 87–––St 37 galvanizedSt 37 galvanized
Dimensions and weights
Flange size- Type 83DN 100 PN 16–––DN 100 PN 16
- Type 84 and 85DN 150 PN 16–––DN 150 PN 16
- Type 87ø 450–––ø 450
Thread of the fixing tube–––G 1 A–––
Total weight of the sensor
- Type 83 … 85appr. 6 kgappr. 6 kgappr. 6 kg
- Type 87appr. 12 kgappr. 12 kgappr. 12 kg
Temperature reaction
VEGASON … StEx-version
At an ambient temperature of40°C the max. temperature to be adjusted
- transducer in Zone 10 of45°C is reached and
- housing (electronics) in Zone 11 of55°C is reached
16
VEGASON 83 … 87
2 Product description
Ambient conditions
VEGASONFKGKStEx
Ambient temperatures related to
- transducer in Zone 10-20°C … +80°C-20°C … +80°C-20°C … +75°C
- housing (electronics) in Zone 11-20°C … +60°C-20°C … +60°C-20°C … +60°C
Storage and transport temperature-20°C … +80°C-20°C … +80°C-20°C … +80°C
Protection
- of the transducerIP 67IP 67IP 65
- of the housing (electronics) generallyIP 67IP 67IP 65
certified
- transducer––––––IP 65
- housing––––––IP 54
Overvoltage classIIIIIIIII
Protection classIIIIII
Max. vessel pressure related to
- Type 830,5 bar1,0 bar–––
- T ype 84 and 850,5 bar0,5 bar0,5 bar
- Type 870,3 bar0,5 bar0,5 bar
Connection line
VEGASON … FK-versions
Power supply2/3 pole, standard cable
VEGASON … GK-versions
Transducer to sensor electronics
- Type 83 … 85standard coax cable type RG 58
standard length 5 m
max. length 300 m
- Type 874 x 0,38 mm
2
, 1 screened wire
standard length 5 m, max. length 35 m
cable diameter 8 mm
Temperature sensoreach integrated in the transmitter
with GK-version connected via above
connection line
Electrical connection
Terminals for the connection linesfor max. 1,5 mm
Terminals for earth connectionfor max. 4,0 mm
2
2
Cable entry2 x Pg 13,5 for all FK-versions
3 x Pg 13,5 for all GK-versions
VEGASON 83 … 87
17
2.10 Dimensional drawings
Standard- and StEx-versions
(Dimensions in mm)
2 Product description
VEGASON 83 FV
VEGASON 83 FK
Min. distanceFlange acc. to DIN - DN 100 PN 16 (ANSI 31/2" 150 psi)
VEGASON 84 and 85 FV…
VEGASON 84 and 85 FK…
0,7…0,8 m
260 x 150
ø110
ø220
260 x 150
45˚
ø18
24730
2
ø22
247
ø180
45˚
Min. distance
VEGASON 87 FV…
VEGASON 87 FK…
0,4 m
0,7…1,2 m
20
95
ø145
ø285
Flange acc. to DIN - DN 150 PN 16 (ANSI 6" 150 psi)
260 x 150
272
20
95
ø240
45˚
ø22
Min. distance
18
ø450
Installation approx. 320
ø410
VEGASON 83 … 87
2 Product description
Standard- and StEx-versions
(Dimensions in mm)
Electronics unit
260 x 150
90
253
VEGASON 83 GV
VEGASON 83 GK
28
~155
138
ø162
Swivelling holder
Flange
DN 150 PN 16
Standard version
Thread G 1 A
45
ø140
ø6,4
VEGASON 83 … 87
ø95
Transducer
SW 73
SW 73 R Ex
90
Min. distance
0,5…0,6 m
19
Standard- or StEx-versions
(Dimensions in mm)
VEGASON 84 and 85 GV…,
84 and 85 GK…
28
2 Product description
Swivelling holder
VEGASON 87 GV…
VEGASON 87 GK…
~155
138
28
~155
138
Flange
DN 150 PN 16
Swivelling holder
Flange
DN 150 PN 16
Standard version
Thread G 1 A
ø145
Transducer
SW 74
SW 75
Standard version
Thread G 1 A
StEx-version
Thread G 1 A
28
82
138
Transducer
SW 74 R Ex
SW 75 R Ex
Min. distance
0,7…1,2 m
StEx-version
Thread G 1 A
28
60
20
Installation approx. 320
176
Min. distance
0,4 m
VEGASON 83 … 87
2 Product description
Ex-version
(Dimensions in mm)
VEGASON 83 FV Ex, Ex B
VEGASON 83 FV Ex 0, Ex 0 B
Flange
DN 100 PN 16
0,7 m
Min. distance
VEGASON 84 FV Ex, Ex B
VEGASON 84 FV Ex 0, Ex 0 B
ø154
ø220
Pg 16
335
80
20
ø18
45˚
ø180
Flange
DN 150 PN 16
1,1 m
Min. distance
ø154
ø285
Pg 16
335
60
22
ø22
45˚
ø240
VEGASON 83 … 87
21
2 Product description
2.11 Approvals
If measuring systems acc. to the following approvals are installed, the respective legal documents have to be observed.
The documents are enclosed to the respecitve measuring system.
StEx-approval
For measuring systems in dust-Ex areas.
- pulse-echo sensorVEGASON 84 … 87 FK StEx and GK StEx
VEGASON 84 … 87 FV StEx and GV StEx
defined in the type approval BVS 93.Y.8005
- signal conditioning instrumentVEGAMET … with digital transmission of measuring data
- processing systemVEGALOG 571 with EV-input cards
- auxiliary level switchall VEGASEL with current input 0/4 … 20 mA
Ex-approval
For measuring system in hazardous areas,
certificated acc. to- CENELEC
- ElexV Zone 0 (Germany)
- pulse-echo sensorVEGASON 83 and 84 FV Ex
VEGASON 83 and 84 FV Ex B
- power supply viaseparator type 146 or VEGATRENN 546
acc. to conformity certificate PTB-no. Ex-93.C.4025
defined in the conformity certificate PTB-no. Ex-93.C.4092 (CENELEC-certificate)
- signal conditioning instrumentVEGAMET … with digital transmission of measuring data
- processing systemVEGALOG 571 with EV-input cards
- auxiliary level switchall VEGASEL with current input 0/4 … 20 mA
- pulse-echo sensorVEGASON 83 and 84 FV Ex 0
VEGASON 83 and 84 FV Ex 0 B
- power supply viaseparator type 146 or VEGATRENN 546
acc. to conformity certificate PTB-no. Ex-93.C.4025
defined in the conformity certificate PTB-no. Ex-94.C.4041 (CENELEC and ElexV, Zone 0, Germany)
- signal conditioning instrumentVEGAMET … with digital transmission of measuring data
- processing systemVEGALOG 571 with EV-input cards
- auxiliary level switchall VEGASEL with current input 0/4 … 20 mA
Ex-technical data of sensors
Signal current circuitin flame proofing intrinsic safe EEx ia IIB
Terminals1 (+) und 2 (–) bzw. 3 (+) und 4 (–)
Effective inner inductance
each current circuit L
Effective inner capacitancenot important
22
i
VEGASON 8. FV Ex, Ex0not important
VEGASON 8. FV Ex B, Ex 0 B65 µH
VEGASON 83 … 87
3 Environment
3Environment
3.1Installation recommendations
relating to liquid tank / vessels
Installation under normal conditions
Flange version
Mounting on a very short, generally on a DIN-socket piece
available on the tank.
The max. filling of the tank depends on the min. distance
of the respective sensor or compact instrument.
Min. distance
Separate version
Mounting of the transducer in a very small hole (ø 35 mm)
in the tank ceiling. The mounting procedure can be
provided via an available access hatch.
Installation under special conditions
If the min. distance of the sensor is decreased, mounting
on a socket piece is necessary .
Generally the diameter of the socket piece should be as
large as possible and the socket length should be as short
as possible.
~45°
Min. distance
See the following schedule as general tube relating to the
dimensioning of the socket piece length to socket piece
diameter.
Tube length L to tube diameter ø
VEGASON83848587
The max. filling of the tank depends on the min. distance
of the respective sensor or compact instrument.
In most applications a separate version is recommended
for open vessels, i.e. mounting of the transducer to a
fastening bracket or in another fix position above the
vessel opening.
The max. filling of the vessel depends on the min. distance
of the respective sensor or compact instrument.
Due to variation possibilities of the transducer position, the
requested filling height can be ensured.
Information to socket piece / mounting
- In case of round top the sensor or the compact
instrument should be mounted on a socket piece
outside the center (mounting on half the range).
Radius
1
/
2
- The emitted sound impulse packets must not be
influenced by the filling stream.
Min. distance
Min. distance
- The socket piece must not be flooded by the measured
product.
24
- Chamfer the edges of the socket piece.
- There must be no weldment joints inside the socket
piece.
VEGASON 83 … 87
3 Environment
3.2Installation recommendations
relating to solid silos
Installation under normal conditions
Flange version
Mounting on a very short, generally on a DIN-socket piece.
The socket piece must be directed to the center of the silo
outlet to ensure sufficient reflection in case of different
material angles of repose. The max. filling on the silo
depends on the min. distance of the respective sensor or
compact instrument.
Min. distance
Installation under special conditions
If the min. distance of the sensor is decreased, mounting
on a conical socket piece extension is necessary.
Information to socket piece / mounting
- Conical socket pieces acc. to drawing, i.e. conus
≥ 2x15°.
≥15° ≥15°
Min. distance
Separate version
In case of horizontal silo top the transducer of the separate
version in conjunction with a swivelling holder can be
easily and optimally directed to the center of the silo outlet.
The max. filling of the silo depends on the min. distance of
the respective sensor or compact instrument.
Min. distance
- Cylindrical socket pieces acc. to the values in the
schedule (see page 23) should only be used in silos
after consultance.
- Chamfer the edges of the socket piece.
- There must be no weldment joints inside the socket
piece.
Important:
The distance from the mounting position to the filling and
suction openings should be as large as possible.
VEGASON 83 … 87
25
4Electrical connection
4 Electrical connection
4.1Connection example of sensors
VEGASON 83 … 87 FV und 84 … 87 FV StEx
Connection for
VEGACONNECT
R
ECHOFOX
VEGASON 8 FV
VBUS
4
+ -
5 6
+
VEGASON 83 … 85 GV
VEGASON 84 and 85 GV StEx
R
ECHOFOX
VEGASON 8 GV
VBUS
4
+ -
5 6
+
VEGACONNECT
Synchronisation
8 9
7
in
out
Synchronization
VEGACONNECT
Synchronisation
8 9
7
inout
10
SW
10
The connection between echo sensor and signal
conditioning instrument or processing system can be
made with standard two-wire cable.
For installation with strong interference, screened cable is
recommended. Screening must only be earthed at oneend.
The two terminals marked with "" are connected in the
housing. Connect screening as marked and carry out
screening on the upper large terminal. Potential
equalization to the outer terminals.
VEGAMET 407 V
18
17
16
15
14
13
9
8
7
6
5
-
-
2
+
+
1
VEGAMET 514 V
d b z
COM 02
t
!
MOD
+
-
STO
VEGAMET
407 V
SIM
Connection
as above
Connection cable of
transducers SW 73 … SW 75
(max. line length 300 m)
VEGASON 87 GV und GV StEx
R
ECHOFOX
VEGASON 8 GV
VBUS
+
-
synchronisation
in
4
5
out
7
6
+
Connection
as above
Connection cable of
transducers SW 7
(max. line length 35 m)
26
VEGACONNECT
SW7
8
1013
9
ws
gn
+
30
32
+
-
+
-
+
-
d b z
16
26
32
11
12
ge
br
Further connection
possibilities.
Max. 3 x 5 =
15 sensors
VEGALOG 571
EV-input card
VEGASON 83 … 87
4 Electrical connection
4.2Connection example of Ex-sensors
VEGASON 83 and 84 FV Ex, FV Ex B
VEGASON 83 and 84 FV Ex 0, FV Ex 0 B
2
3
1
4
Separator
Type 146
11
Ex-area
Not-Ex-area
12
Information concerning safety and installation see
following page.
In case of strong interferences, screened cable is
recommended. Screening must only be earthed at oneend.
18
17
16
15
14
13
9
8
7
6
5
2
-
1
+
COM 02
t
!
MOD
+
-
STO
VEGAMET
407 V
SIM
VEGAMET 407 V
1
1
2
+
+
2
-
3
3
4
4
-
VEGATRENN 546
d b z
20
+
22
-
+
30
-
+
32
-
Max. 5 sensors can be
connected via 5
VEGATRENN 546 to
the EV-input card
+
-
+
-
d b z
30
32
d b z
16
24
32
VEGAMET 514 V
VEGALOG 571
EV -input card
VEGASON 83 … 87
27
Connection example of Ex-sensors
4 Electrical connection
V essel with cathodic protection
External earth
terminal
Zone 1
VEGASON 83 a. 84 Ex, ExB
Zone 0
VEGASON 83 a. 84 Ex0, Ex0B
Cathodic tank protection
2 … 24 V (object voltage)
V essel without cathodic protection
External earth
terminal
Zone 1
VEGASON 83 a. 84 Ex, ExB
Zone 0
VEGASON 83 a.. 84 Ex0, Ex0B
to potential equalization line (PAL)
Furthermore all installations must be carried out acc. to
local regulations.
Furthermore all installations must be carried out acc. to
local regulations.
28
VEGASON 83 … 87
4 Electrical connection
4.3Connection example of compact instruments
VEGASON 83 … 87 FK and 84 … 87 FK StEx
Anschluß für
VEGACONNECT
R
ECHOFOX
+(L1) -(N)
power supply
VEGASON 8 FK
I out
3
4
+ -
0/4…20 mA
5 6
in
VEGACONNECT
Synchronisation
7
out
Synchronization
8 9
10
The two terminals marked "" are connected in the
housing.
Potential equalization at the outer terminals.
Evaluation, current output 0 … 20 mA
The current output can be directly connected to the
indicators or the processors.
Power supply ,
Supply voltage see 2.9.1 Technical data
VEGASON 83 … 85 GK and 84 and 85 GK StEx
R
ECHOFOX
+(L1) -(N)
power supply
Connection
as above
VEGASON 8 GK
4
3
+ -
0/4…20 mA
5 6
in
SynchronisationI out
VEGACONNECT
SW
7
out
8 9
10
Connection cable of
transducers SW 73 … SW 75
(max. line length 300 m)
VEGASON 87 GK and GK StEx
R
ECHOFOX
+(L1) -(N)
power supply
Connection
as above
VEGASON 8 GV
synchronisationSW7
I out
IN
3
2
+ -
0/4…20 mA
VEGACONNECT
5
6
8
out
7
1013
9
ws
gn
Connection cable of
transducers SW 7
(max. line length 35 m)
11
12
ge
br
VEGASON 83 … 87
29
5Set-up
5 Set-up
Set-up can be realized in different ways. Observe the
separate instruction for ECHOFOX
®
-sensors.
5.1Sensor and signal conditioning
instrument (e.g. VEGAMET)
The set-up can be made directly via the operating
elements of the signal conditioning instrument or via a PC
(equipped with VVO-Software).
In this case the computer can be connected via
VEGACONNECT with the signal conditioning instrument
or directly with the sensor.
Attention, in case of Ex-sensors in the Ex-area,
VEGACONNECT must only be connected in the Ex-area
(e.g. signal conditioning instrument).
5.2Sensor and processing system
(VEGALOG 571)
The set-up can be only made via PC (equipped with VVOSoftware).
The PC must be connected with a suitable central unit
(CPU) with the processing system VEGALOG 571
(interface RS 232).
5.3Compact instrument
The PC (equipped with VVO-Software) must be directly
connected via VEGACONNECT with the compact
instrument or the current output.