10.1 Menu schematic "Reduced menu" ................................. 49
10.2 Menu schematic "Extended menu" ................................ 50
Safety informatio n ...................................................................... 56
Note Ex area .............................................................................. 56
VEGAMET 614V3
Page 4
1 Product description
Product description
1.1 Function and configuration
VEGAMET 614 V is a signal conditioning
instrument for a number of different applications, e.g.:
- level measurement
- gauge measurement
- process pressure measurement
- additional temperature measurement (with
respective sensors)
Software configuration
Input component
- continuous current
input
Function
The signal conditioning instruments power
the connected sensors and process their
analogue measuring data. The conditioning is
done by a special software consisting of
functional components (FB), input components (EB), output components (AB) as well
as DCS components (PB).
Configuration
The VEGAMET 614 V signal conditioning
instrument is a module unit with plug-in
socket. The front plate contains an adjustment module with LC display and keys.
Output components
- current output
- voltage output
- DISBUS outputs
- failure output
- MET display
- relay outputs
AB
EBFB
PC/DCS components
- PC/DCS outputs
Functional component
- measurement loop
TA G 1
4VEGAMET 614V
PB
Page 5
Product description
1.2 Approvals
As an accessory, intrinsically safe instrument.
EC type approval certificate acc. to
CENELEC
II (1) G [EEx ia] II C
1.3 Features
- microcomputer-controlled signal conditioning instrument for continuous measurement
- adjustment module with LC display and 6
keys
- adjustable integration time
- two fixed and three user-programmable
linearisation curves
- fault monitoring
- fault signal and failure diagnosis via display
- temperature measurement loop (on pressure transmitters and ultrasonic sensors
delivering a temperature signal)
Input:
- 1 sensor input for digital transmission of
measured data (VBUS) (pressure transmitters, ultrasonic and radar sensors)
Outputs:
- 2 current outputs 0/4 … 20 mA
- 2 voltage outputs 0/2 … 10 V
- 2 relay outputs (spdt)
- 1 DISBUS output for digital wiring and for
connection of VEGADIS 174
- 1 fail safe relay
- adjustment also via PC with adjustment
software VVO; connection via the interface
adapter of the VEGACONNECT series
VEGAMET 614V5
Page 6
Product description
1.4 Technical data
Power supply
Operating voltage / power consumption U
VA
Optional instrument versionU
Power supply with Ex approved instrument
Operating voltage / power consumption U
Meas. data input
Number1 input
Type of inputactive two-wire input, digital (VBUS) (sensor
Sensorpressure transmitters, ultrasonic sensors and
Voltageapprox. 25 V DC
Power limitationat approx. 4.5 W, short-circuit proof
Min. adjustment delta2 % of sensor range (hydrostatic)
Connection cable2-wire standard cable (screening
max. cable length1000 m
max. resistance per wire15 Ω
= 90 … 250 V AC, 50/60 Hz / approx. 14
nom
= 90 … 250 V DC
= 20 … 72 V DC / 7 W
nom
= 20 … 53 V AC, 50/60 Hz, 14 VA
= 20 … 250 V AC, 50/60 Hz / 14 VA
nom
= 20 … 72 V DC / 5 W
powered by VEGAMET)
radar sensors
5 mm (ultrasonic, radar)
recommended)
Meas. data input with Ex approved instrument
Voltageapprox. 15 V at 5 mA
approx. 13.5 V at 25 mA
Current limitationat 32 mA (short-circuit proof)
further data of the meas. data input the same as the instrument without Ex approval
Current output
Number2 outputs
Functionanalogue output of the processing results
Rangeindividually adjustable in the range of 0 … 20 mA
Loadmax. 500 Ω
Resolution1 µA
Linearity error0.05 % (relating to 20 mA)
Temperatur e error0.05 %/10 K (relating to 20 mA)
Voltage output
Number2 outputs
Functionanalogue output of the processing results
Rangeindividually adjustable in the range of 0 … 10 V
Currentmax. 1 mA
Resolution0.5 mV
Linearity error0.05 % (relating to 10 V)
Temperature error0.06 %/10 K (relating to 10 V)
6VEGAMET 614V
Page 7
Product description
Relay outputs
Numbe r2 operating relays, 1 fail safe relay
Contactfloating spdt
Contact materialAGSNO
Tur n-on voltagemin. 10 mV DC
, hard gold plated
2
max. 250 V AC/DC
Switching currentmin. 10 µA
max. 3 A AC, 1 A DC
Turn-on voltagemax. 750 VA
max. 18 W at U > 40 V DC
max. 40 W at U ≤ 40 V DC
Min. switching hysteresis
(Low/High delta)0.5 %
DISBUS output
Functiondigital transmission for linking between signal
conditioning instruments and for connecting
digital indicating instruments
Connection cable2-wire standard cable (screening
recommended)
max. cable length1000 m
max. resistance per wire15 Ω
Indicating elements
Clear text indicationLC display
- 4-lines, 6 digits each
- background lighting
Analogue indicationLED chain consisting of:
- 11 segments 0 % … 100 %
- indicates the actual value of the selected
measurement loop
LEDs in front plategreen: operating voltage on
red: fault signal (LED lights with deenergised
relay)
yellow: relay switching condition (standard
adjustment: LED lights with energised relay)
Adjustment elements
Front plate6 keys for configuration and parameter setting
Upper housing siderotary switch for adjustment of the instrument
address
on DISBUS (in mounted status covered by
plug-in socket)
Ambient conditions
Permissible ambient temperature-20°C … +60°C
Storage and transport temperature-40°C … +80°C
- insulation resistance1.4 kV
Potential separationbetween DISBUS and outputs
- reference voltage50 V
- insulation resistance0.5 kV
Common reference potential atvoltage and current output
Mechanical data
Seriesmodule unit with plug-in socket, incl. transpa-
rent cover, sensor terminal cover, coded
pin, two connection jumpers
Mountingcarrier rail mounting acc. to DIN 46 277, Bl. 3
Dimensions, unmountedW = 72 mm, H = 118.5 mm, D = 134 mm
Weightapprox. 480 g
CE conformity
The VEGAMET 614V Ex signal conditioning instrument meets the protective regulations of
EMC (89/336/EWG) and NSR (73/23/EWG). Only for use in industrial areas. Conformity has
been judged acc. to the following standards:
EMCEmissionEN 50 081 - 2: 1993
SusceptibilityEN 50 082 - 2: 1995
NSREN 61 010 - 1: 1993
8VEGAMET 614V
Page 9
Product description
Ex technical data
Power supply
Operating voltageU
Reference voltageU
Meas. data input (intrinsically safe circuit)
Classification
Max. values
- voltageU
- currentI
- powerP
Characteristicslinear
Effective inner inductance L
Effective inner capacitance C
I
I
= 20 … 250 V AC
nom
= 20 … 72 V DC
= 253 V AC or 125 V DC
m
II (1) G [EEx ia] II C
= 18.2 V
o
= 270 mA
o
= 926 mW
o
negligible
negligible
EEx ia IIC
Max. permissible outer inductance L
O
0.28 mH
Max. permissible outer capacitance CO297 nF
The intrinsically safe circuits is reliably separated from the non-intrinsically safe circuits up
to a peak value of the nominal voltage of 375 V. The max. voltage on the non-intrinsically
safe circuits must not exceed 253 V
in case of failure.
eff
1.5 Dimensions
Carrier rail 35 x 7.5 or 35 x 15 acc.
to EN 50 022
Transparent cover
12
on
CONNECT
614
56 78
13 1415 16
23 24 25 26 27 28
72
%
100
+
-
118,5
OKESC
Ser. No.
12345678
134
54,5
12 34
910 1112
17 18 20 21 22
VEGAMET 614V9
Page 10
2 Mounting and installation instructions
Mounting
Each series 600 signal conditioning instrument consists of a plug-in socket for carrier
rail mounting DIN 46 277 and a module unit.
The supply voltage can be connected to
terminals 17 and 18.
For neighbouring series 600 signal conditioning instruments, it is possible to continue the
connection L1 and N directly via the supplied
jumpers. The same is valid for the connection
of VEGAMET voltage output 0 … 10 V (terminals 15 and 16) and for DISBUS output (terminals 9 and 10).
Note!
The jumpers must never be used with single
instruments or at the terminating end of an
instrument row. If this is not observed, there
is danger of coming in touch with the operating voltage or causing a short-circuit.
VEGAMET 614V Ex signal conditioning instrument is intrinsically safe and must not be
installed in hazardous areas.
Before setup, the Ex separating chamber
must be plugged into VEGAMET 614V Ex as
shown in the illustration.
Safe operation is ensured only if the operating instructions manual and the EC type
approval certificate are observed. VEGAMET
614V Ex must not be opened.
Separating
chambers
Ex coding
Direct supply for
voltage output 1
DISBUS and supply
voltage with jumpers
Coding
To avoid interchanging the various signal
conditioning instruments, the plug-in socket
is provided with pins and the signal conditioning instrument with corresponding gaps
(mechanical coding)
An Ex coding with inserted coding pins ensures that the non-Ex and Ex instruments are
not interchanged.
and Ex coding pin) must be inserted by the
user according to the below table.
Mounting and installation instructions
leftright
A o
B o
C o
1 o
2 o
3 o
o
0…10 V
o
DISBUS
o
7 o
8 o
9 o
o
o
12 o
A o
VEGA
-
+
N
L1
VEGA
B o
C o
1 o
2 o
3 o
4 o
0…10 V
5 o
DISBUS
6 o
7 o
8 o
9 o
N
10 o
L1
11 o
12 o
On VEGAMET 614V Ex, the supplied
coding pins (instrument coding pin
-
+
InstrumentEx
codingcoding
VEGAMET 614Vright 2––
VEGAMET 614V Exright 2left A
left B
right B
left B
right B
10VEGAMET 614V
Page 11
Mounting and installation instructions
Transparent cover
To protect the instrument against unauthorized adjustment, the front plate of VEGAMET
can be equipped with two lockable transparent covers after setup. To remove the transparent cover, see illustration.
VEGAMET 614V11
Page 12
3 Electrical connection
Electrical connection
3.1 Connection instructions
Note the following instructions for electrical
connection:
- The connection must be made acc. to the
local installation standards (e.g. in Germany acc. to the VDE-regulations).
- The wiring between VEGAMET 614 V and
sensor can be made with standard twowire cable.
- If strong electromagnetic interferences are
expected, we recommend the use of
screened cable. The screening must be
earthed on both ends.
- The line resistances stated in the technical
data must not be exceeded.
- If overvoltages are expected, we recommend a sensor electronics with integrated
overvoltage protection or the installation of
VEGA overvoltage arresters.
Connection instructions for Ex
approved applications
When wiring between VEGAMET 614V Ex
and an explosion protected sensor, the installation rules, in Germany acc. to VDE 165
and the EC type approval certificate must be
observed. The official documents are supplied with the respective instrument.
3.2 Wiring plan
Fail
safe
relay
Current
output 2
5678
Voltage
output 2
13 1415 16
23 24 25 26 27 28
Level
relay 1
Current
output 1
Voltage
output 1
Level
relay 2
VBUS
sensor
+-+-+-
1234
DISBUS
output
+-+-+-
9101112
17 18 19 20 21 22
+-
L1
N
Power supply
The connections „9, 10, 15, 16, 17 and 18“ can be
connected with the supplied jumpers to neighbouring
series 600 signal conditioning instruments.
Ex version
Before connecting Ex certified instruments,
please note the instructions in the attached
official documents as well as the valid installation regulations. Make sure that the Ex separating chamber is plugged in on the left
instrument side before setup (see illustration
Coding).
12VEGAMET 614V
Page 13
Adjustment
4 Adjustment
4.1 Indicating and adjustment
elements
A rotary switch is located on the upper side
of the housing. It is used to set the instrument
address on DISBUS. This setting is necessary if several VEGAMET signal conditioning
instruments are linked via the DISBUS or if
you control a VEGADIS 174 indicating instrument via the DISBUS.
The adjustment range 1 … F corresponds to
the DISBUS addresses 1 … 15.
Factory setting: 1
Note:
If address 0 is set, the VEGAMET does not
participate in the DISBUS communication.
Rotary switch
8
9
7
A
6
5
4
33
2
B
C
D
E
F
1
0
The set instrument address on the DISBUS
can be indicated via the LC display.
Note
If several VEGAMET are linked via the
DISBUS, each address must be assigned
only once!
10
11
12
13
14
15
Position of the rotary switch
Rotary switch
Front plate
1
%
100
+
4
5
6
7
1 LCD (4 lines with 6 figures each, illuminated) for clear text
indication
2 LED chain (yellow) for quasianalogue indication of the
measured value
3 Keys for menu adjustment
4 Connection socket for VEGACONNECT
5 LED (yellow) lights if the relay is energised (standard
setting)
6 LED (red), lights if fail safe relay is deenergised
7 LED (green), lights if operating voltage is on
CONNECT
12
on
614
-
OKESC
Ser.No.
12345678
2
3
VEGAMET 614V13
Page 14
4.2 Adjustment system
Indicating and adjustment module
Branching point, i.e. jump to the
next lower menu with [OK]
Adjustment
Indication of
- measured value
- menu item
- parameter
%
100
+
-
Depending on the menu item, change value
or choose out of a list
Choose menu item
Depending on the menu item, interrupt
adjustment or change to the next higher
menu
"Reduced menu" – "Extended menu"
Two menu systems are available: the „reduced menu“ and the „extended menu“.
Note:
VEGAMET 614 V is preset to „reduced menu“
mode!
In the majority of applications, you can carry
out the necessary settings in this menu mode
(see also chapter 4.5 Comparison of the
menu modes).
Adjustment structure
The adjustment is made via 6 keys in conjunction with the text display. The jump from
the measured value display to the next lower
menu item is made with [OK]. Use [–>] or
[<–] to change within this menu level from
one menu item to the other. A branching point
is indicated by the symbol ▼ and enables
with [OK] a jump to the next lower menu item.
Parameters are indicated by the missing
symbol ▼. The value of the parameters can
be modified with [+] or [–] or chosen out of a
list. The modified value can be saved with
[OK]. To interrupt a setting (without saving
the modification), you have to push [ESC]. A
jump up to the next higher menu item is
Depending on the menu item, save the set
value or change to the lower menu
OKESC
made with [ESC]. 15 minutes after pushing a
key for the last time, an automatic reset to the
measured value display is triggered. A jump
up to the next higher menu level with [ESC] is
possible from each menu item , even if no
connection is shown in the menu schematics.
Add’l
function
ESCESCESCESC
Menu
Language
English
TAG 1
to default
mode
Reduced
Password
off
If you are in the utmost right menu item of a
menu row and push the [–>] key, you will
immediately reach the utmost left menu item
of this row. You also reach immediately the
right menu item from the left one, if you push
the [<–] key.
The connections (shown with a broken line)
are not shown in the menu schematics in
order to provide a clearer overview.
Volt
output 1
prop.
to
Percent
14VEGAMET 614V
Unit
0,0%
Volt
output
0/10V
Failure
mode
0V
Volt
limitation
on
Page 15
Adjustment
Formatting examples of the menu
schematic
White letters indicate parameters
Language
which can be modified with the [+] or
Eng-
[–] key and saved with the [OK] key.
lish
Examples:
- In the menu item „Language“ you can
change from German to English.
Bolt print/italic inputs, e.g. the meas-
Param.
TAG-
urement loop name "TAG-No. 1" can
No. 1
show a different text if you have
▼
carried out a parameter adjustment
or configuration of the measurement loop. In
the menu schematics, you will see the factory
setting.
Light grey menu items are only dis-
Volt
at
played if necessary (depending on
0%
the adjustments in the other menu
0,000
items).
Changing the menu mode:
Move from the display of measured value
(TAG-No.1) to the menu item „Menu mode
reduced“. Now push [+] twice until „Menu
mode Extended“ appears, then push [OK].
You can now return to the measured value
indication by pushing [ESC] twice.
4.3 Adjustment via PC
Setup
- directly via the keys of the adjustment
module
- or with a PC, equipped with VVO software
(VEGA Visual Operating) and an interface
adapter VEGACONNECT.
%
100
+
-
CONNECT
12
on
614
VEGACONNECT
OKESC
Ser.No.
12345678
TAGNo. 1
%
xx,x
Add’l
Confi-
Param.
TAGNo. 1
guration
functions
Password
off
Language
English
TAG 1
to default
Reset
TAG 1
Menu
mode
Reduced
VEGAMET 614V15
Page 16
Adjustment
4.4 Configuration and parameter
adjustment
No matter whether you set up your
VEGAMET via the keys of the adjustment
module or via a PC with the software VVO,
the procedure is always the same.
Proceed in the following sequence:
- configuration (if not already configured with
a factory setting)
- parameter adjustment.
In this operating instructions manual, the
adjustment steps are described which are
directly carried out on the keys of VEGAMET.
The adjustment via the software VVO is described in a separate manual.
Configuration
Configuration means initially assigning or
setting functions. VEGAMET requires a
(mostly once) basic co-ordination that determines the applications and the assignments
of the inputs and outputs. Choices can be
made from existing functions and options.
This procedure is called configuration. The
signal conditioning instruments are delivered
with a configuration which only needs to be
changed in special cases.
The basic configuration includes the following
steps:
1 Configuration measurement loop
• Choose the kind of application (level,
gauge…)
• Choose sensor type (hydrostatic, ultra-
sonic…)
• Determine application (standard, level
difference…)
• Determine options (no option, correc-
tions…)
Note:
The measurement loop configuration can only
be modified, if a „Reset to single measurement“ has been carried out previously (under
„Additional functions“ in the extended menu).
If you only want to modify the sensor type,
you first have to carry out „Reset TAG 1“
(under „Additional functions“ in the reduced
menu).
2 Configuration of inputs
• Determine from where your VEGAMET
receives the input data (sensor, other
VEGAMET)
• Enter sensor characteristics values (meas.
range, current range)
3 Configuration of outputs
• All outputs (except fail safe relay) can been
assigned to one sensor or can be switched
off.
After basic configuration, VEGAMET goes
into operating mode and indicates the actual
measured value. The other configurations
should be carried out after the parameter
adjustment.
Parameter adjustment
Parameter adjustment means changing the
values. Signal conditioning instruments have
many parameters, the values of which can be
modified, such as for example, the integration
time of 0 … 600 s. The modification of these
values is called parameter adjustment. The
parameter adjustment does not influence the
configuration. Take note that parameter adjustment is only possible after a configuration
has been carried out (e.g. set values of the
current input after the current output has
been assigned).
16VEGAMET 614V
Page 17
Adjustment
4.5 Comparison: "Reduced menu" – "Extended menu"
In the table the most important functions of VEGAMET 614 V are listed – each with a mark
indicating in which menu mode the functions are available.
FunctionReduc. Ext.
Adjustment with mediumxx
Adjustment without mediumxx
Set integration timexx
Simulationxx
Switch over current output (4/20 mA, 0/20 mA, 20/4 mA, 20/0 mA)xx
Switch over voltage output (2/10 V, 0/10 V, 10/2 V, 10/0 V)xx
Switch over relay mode (overfill protection – dry run protection)xx
Scaling (display)xx
Change menu languagexx
Activate/deactivate passwordxx
Change sensor typexx
Change measurement loop namexx
Reset measurement loop (Reset TAG)xx
Reset measurement loop to default (TAG to default)xx
Configuration and parameter adjustment of temperature measurement loop–x
Edit linearisation curves–x
Change mode (e.g. level – gauge)–x
Select different reference values for outputs–x
Define output behaviour (current/voltage output) in case of failure–x
Assign outputs to measurement loop (current, Volt, DCS, DIS,
relay outputs)–x
Select further relay modes (switching window, tendency determination)–x
menu menu
VEGAMET 614V17
Page 18
5 Setup
• Set the VEGAMET instrument address on
the circuit board by means of the rotary
switch (only necessary, if several
VEGAMET are connected on the DISBUS).
• Install VEGAMET.
• Connect sensor and power supply.
• After switching on, the display indicates for
approx. 10 secs the instrument type and
the software version, e.g. „MET 614V
V1.12“.
• If a measurement loop has not yet been
configured, the display shows „Configuration“, the red failure LED also lights. In this
case, proceed as described under „6.1
Configuration of measurement loop“.
• If a measurement loop has been
configured, the display shows a value, e.g.
„TAG-No. 1 21.8 %“.You can now carry out
the settings as described in the chapter „6
Settings in the reduced menu“.
Setup
18VEGAMET 614V
Page 19
Settings in the „Reduced menu“
6 Settings in the „Reduced menu“
6.1 Configuration of measurement
loop
Choose sensor type
Meas.
TAG-
value
No. 1
%
indication
xx,x
Param.
TAGNo. 1
Configuration
Sensor
type
Hydrostatic
Pulse-Echo
Pulse-Radar
By default, VEGAMET 614V is configured for
connection to a hydrostatic pressure transmitter. If, for example, you use another sensor, you have to carr y out previously „Reset
TAG 1“ (in the menu item „Reset Now! OK?“
you confirm this with the [OK] key). Move to
the menu item „Sensor type Hydrostatic“,
there you choose with the [+] or [–] keys the
sensor type and confirm it with [OK].
Add’l
functions
Password
off
Language
English
TAG 1
to default
Reset
TAG 1
Delete
TAG
1?
Reset
OK ?
Reset
Now !
OK ?
Name measurement loop
TAGNo. 1
%
xx,x
Param.
TAGNo. 1
Configuration
Sensor
type
Hydrostatic
Sensor
optimize
TAGName
TAG No. 1
You can modify the measurement loop name
„TAG-No. 1“, if you want a more useful designation.
Two lines with six figures each are available.
To do this, move to the menu item „TAG name
…“.
Push [+] or [–] once. The first char-
TAGName
acter of the measurement loop name,
TAG -
in this example the letter T, flashes
No. 1
(editing mode). Continue pushing the
[+] or [–] key, by doing this you scroll
through the alphabet, through the row of
numbers and through a list of characters.
When you have reached the re-
TAGName
quested character, push [–>]. The
TAG -
next character is activated and be-
No. 1
gins flashing. Scroll again with [+]
and [–] to the requested character. With [–>]
you move to the next character.
If you have finished the last position
TAGName
of the name and you have "written"
Ves.4
the name of the measurement loop in
Hall3
this way, you have to push [OK] to
save the setting. If you quit the menu item
without saving, the name of the measurement
loop is again „TAG-No. 1“.
In the display of measured value and
Ves.4
Hall3
in all other menu items in which the
%
measurement loop name is shown,
27,8
you will now find the new designation.
VEGAMET 614V19
Page 20
Settings in the „Reduced menu“
6.2 Sensor optimization
TAGNo. 1
%
xx,x
Param.
Configura-
TAG-
tion
No. 1
Sensor
Sensor
type
opti-
Hydro-
mize
static
For radar/ultrasonic sensors: see
MINICOM menu schematic in the
VEGASON/VEGAPULS operating
instruction manual in chapter „Adjust-
ment with VEGAMET or MINICOM“.
Like the measurement loop name, the sensor
name can also be modified if you want to give
it another designation. To do this, move to the
menu item „Sensor-Tag…“ (first menu item in
the sensor menu). The procedure is the
same as the modification of the measurement
loop name. The sensor name is saved in the
memory component of the sensor.
When you move to the menu item „Sensor
menu“, a new menu item opens which is
independent of VEGAMET. It is provided by
the connected sensor and has different adjustment options (e.g. installation geometry,
meas. environment…) depending on the
sensor type. The appropriate description is
stated in the operating instructions manual of
the sensor. These data, like the sensor name,
are also saved in the sensor. If at a later date
the sensor is connected to another signal
conditioning instrument, its adjustment data
remain.
Normally, the sensor will have been adapted
to customer-specific conditions so that modifications in the sensor menu will only be necessary if the sensor is used for a
measurement other than originally planned.
6.3 Adjustment with medium
You can either carry out the adjustment with
medium or without medium. Adjustment with
medium means that you carry out the adjustment taking the actual level into account. To
carry out an adjustment with medium, it is
necessary to know the percentage value of
the actual vessel filling.
TAGNo. 1
%
xx,x
Param.
TAGNo. 1
Adjustment
with
medium
Min-
Max-
adjust
adjust
at %
at %
100,0
0,0
In this procedure, the percentage values
which correspond to the actual min. and max.
fillings are entered. It doesn’t matter if you
enter first the min. value or the max. value. If
you know, for example, that your filling is
actually 80 %, you can enter in the menu
“max. adjustment” the value „80“. Later, when
your filling is, for example, 10 %, you enter in
the menu “min. adjustment” the value „10“.
The bigger the difference between the two
adjustment points, the more precise the
measurement over the whole measuring
curve. An adjustment at 0 % and 100 %
would be ideal. For practical reasons, it is not
always possible to completely empty or fill a
vessel. The distance between the two adjustment points, however, should be at least
10 % of the sensor range.
20VEGAMET 614V
Page 21
Settings in the „Reduced menu“
Adjustment example
You know that your vessel is actually
Maxadjust
filled up to 80%. Move to the menu
at %
item „Max. adjust at % 100.0“. Push
100,0
the [–] key briefly, so that the figure
100.0 flashes (editing mode).
Continue pushing [–] until the figure
Maxadjust
80.0 appears on the display. Now
at %
push [OK] to save the adjustment.
80,0
Later, when your filling is, for exam-
Minadjust
ple, 10 %, you move to the menu item
at %
„Min. adjust at % 0.0“. Briefly push
0,0
[+], until the figure 0.0 flashes
(editing mode).
Continue pushing [+] until „10“ ap-
Minadjust
pears on the display. Now push [OK]
at %
to save the adjustment.
10,0
You can, of course, carry out the adjustment
with medium at 0 % and 100 % filling. Note:
The figures 0 % and 100 % must flash (editing mode), before they can be saved with the
[OK] key.
6.4 Adjustment without medium
Meas.
TAG-
value
No. 1
indication
%
xx,x
Param.
TAGNo. 1
Adjustment
with
without
medium
medium
Adjust-
0%
at
bar
0,000
100%
at
bar
1,000
ment
bar
m
For this adjustment procedure, you have to
enter two values (pressure or distance) corresponding to the levels 0 % and 100 %.
VEGAMET 614V21
Page 22
Settings in the „Reduced menu“
Adjustment example
You know the data of the vessel, i.e. you
know that at 0 % filling, the distance from the
liquid to the sensor is 3.5 m and at 100 %
filling, 0.3 m.
Radar sensor
d1
0.3 m
at 100 %
d2
3.5 m at 0 %
In this menu item, you choose the
Adjustment
adjustment unit, e.g. m (d), if you use
a radar sensor [m (d) = distance in
m(d)
m]. Use the [+] and [–] key to
choose the unit and [OK] to save the selected unit.
0%
Enter in these two menu items the
at
distance values (d1 and d2). With the
m (d)
[+] and [–] keys you set the values,
3,50
with [OK] you save them.
100%
at
Note that in this example, a lower
m (d)
distance value exists at 100 % filling
0,30
than with 0 % filling!
6.5 Scaling
TAG-
Meas.
No. 1
value
%
indication
xx,x
Param.
TAGNo. 1
Sig-
Ad-
nal
just-
condi-
ment
tioning
Integr
Scal-
ation
ing
time s
0
0%
100%
corres-
corres-
ponds
ponds
1000
0
The adjustments under scaling determine
which number values and which unit will be
actually displayed on the measured value
indication (upper menu item). In the menu
items „0 % corresponds…“ and „100 % corresponds…“ you can enter with [+] or [–]
and [OK] values which should be displayed
at 0 % and at 100 % filling. In the menu item
„Decimal point…“ you determine the position
of the decimal point. In the next menu item
„prop. to…“ you can select from a list of parameters: percent, mass, volume, pressure
etc. After pushing [+] or [–] save with [OK].
According to this adjustment, the menu item
„Unit…“ provides a new choice, e.g. m, dm,
cm, mm, ft, yd and in, if you have selected
the parameter „Height“ (compare chapter „8
Measured quantity and units“).
Decimal
point
888,8
prop.
to
undefined
Unit
– –
22VEGAMET 614V
Page 23
Settings in the „Reduced menu“
Scaling example
You have a raised tank and you want the
level to be indicated in meters with three
places after the decimal point.
100 %
h=4.5 m
h=1.5 m
2.341 m
0%
corresponds
1500
0 %
Enter the number 1500 (with keys
[+], [–] and [OK]) in this menu item.
This corresponds to the level at 0 %
(if you want to have a lower resolution
of the meas. value indication, you can also
enter 150 or 15).
Enter the number 4500 (with keys
100%
corres-
[+], [–] and [OK]) in this menu item.
ponds
This corresponds to the level at
4500
100 %.
6.6 Integration time
Meas.
TAG-
value
No. 1
%
indication
xx,x
Param.
TAGNo. 1
Sig-
Ad-
nal
just-
condi-
ment
tioning
Scaling
If the integration time is set to 0 s (factory
setting), each quick change of the product
surface (e.g. waves) will be immediately
detected and interpreted as level change. All
output values of VEGAMET will follow the
wave movements. To avoid this, the integration time can be increased (max.
600 s are possible with VEGAMET). The
higher the time setting, the slower the measured value display reacts.
Integr
ation
time s
0
Adjust the decimal point as shown
Decimal
(with keys [+], [–] and [OK]).
point
8,888
prop.
Choose the term „Height“ in this
to
menu item (with keys [+], [–] and
Height
[OK]).
Percent
Pressure
Mass
If you have chosen „Height“ in the
Unit
previous menu item, you get a choice
m
of several units (mm, dm, cm, m,
yard…). Choose m (with the keys
dm
cm
[+], [–] and [OK]).
….
TAG -
You can now see the scaled value in
No. 1
the measured value display.
m
2,341
VEGAMET 614V23
Page 24
6.7 Outputs
Meas.
TAG -
value
No. 1
indication
%
xx,x
Param.
TAGNo. 1
Adjustment
Signal
conditioning
Outputs
Settings in the „Reduced menu“
Current
output 1
4/20mA
0/20mA
20/4mA
20/0mA
Current
output 2
4/20mA
0/20mA
20/4mA
20/0mA
Volt
output
1
0/10V
10/2V
10/0V
2/10V
Volt
output
2
0/10V
10/2V
10/0V
2/10V
The current output of VEGAMET is set to
4 … 20 mA. By pushing [+] or [–] and [OK]
you can set the output to the following values:
0/20 mA, 20/4 mA and 20/0 mA.
Beside the preadjustment (0/10 V), the following options are available for the voltage output: 10/2 V, 10/0 V and 2/10 V.
As a factory setting, the relay outputs are set
to mode „Overfill protection“ (mode A). By
pushing [+] or [–] and [OK] you can assign
the mode „Dry run protection“ (mode B)
separately for each relay output. In the next
menu items you can determine the upper and
lower switching points for the set mode.
The graphics on the right show the behaviour
of the relay outputs and LEDs depending on
the level.
Relay
Relay
output
output
2
1
High
Low
%
0,0
High
%
100,0
%
100,0
Mode
Overfill
protec
Dry run
protec
Mode
Overfill
protec
Dry run
protec
Low
%
0,0
Overfill protection
Parameter
High
Low
Relay output
t
The relay of relay output 1 deenergises at the
switching point „High“ (safe switching condition).
Dry run protection
Parameter
High
Low
Relay output
t
The relay of relay output 1 deenergises at the
switching point „Low“ (safe switching condition).
24VEGAMET 614V
Page 25
Settings in the „Reduced menu“
6.8 Simulation
Meas.
TAG-
value
No. 1
%
indication
xx,x
Param.
TAGNo. 1
Adjustment
Signal
conditioning
Outputs
Simulation
Simulation
Now !
OK ?
Simulation
%
xx,x
To check outputs and connected instruments,
you can adjust any individual percentage
value with the keys [+] and [–] in this menu
item. Initial point is always the actual measured value. The indicated value flashes when
simulation is activated. 15 minutes after setting the simulated value, the simulation automatically terminates and resets to meas.
value indication.
6.9 Password, language, reset,
menu mode
Meas.
TAG-
value
No. 1
%
indication
xx,x
Configuration
Add’l
functions
Password
off
on
Language
English
German
French
etc.
TAG 1
to default
Reset
OK ?
Reset
TAG 1
Menu
mode
Reduced
Param.
TAGNo. 1
Password
You can also protect VEGAMET 614V against
unauthorised or incorrect adjustment. Activate in the menu „Password“ the option „on“
(with keys [+], [–] and [OK]). Now you can
only check the meas. value on the display.
You can still reach any menu item with the
arrow and OK keys, but as soon as you try
to modify an adjustment, you are asked for
the password. Enter the number „614“ (with
the keys [+], [–] and [OK]), to carry out the
adjustment. Entering the password once is
sufficient to open all protected menu items.
Language
In the menu item „Language“, you can
choose a language other than German, such
as English, French etc. as menu language
(with keys [+], [–] and [OK]).
Reset
In the menu item „TAG 1 to default“ you reset
all settings relating to measurement loop 1 to
default. If, for example, you have changed
the name of the measurement loop to „Silo 5“,
after the reset, the name is again „TAG-No.
1“. Also, all the adjustment values you have
entered for measurement loop 1 are now
reset to default. A measurement loop configuration carried out later on, however, remains.
Menu mode
The menu modes „Reduced“ and „Extended“
are available. With the keys [+], [–] and [OK]
you choose the required menu mode. The
menu schematics of the reduced menu and
the extended menu are shown later in this
operating instructions manual. The factory
Ex-
setting is „Reduced menu“.
tended
Reset
Now !
OK ?
VEGAMET 614V25
Page 26
7 Settings in the „Extended menu“
7.1 Configuration of measurement loop
TAG -
TAG -
No. 1
No. 2
%
°C
xx,x
xx,x
Confi-
Param.
Param.
TAGNo. 1
TAGNo. 2
guration
Config
inputs
Config
meas.
loop
TAG 1
Level
TAG 2
Temperature
Settings in the „Extended menu“
Application
Level
Gauge
Pressure
TAGname
TAG No. 2
Sensor
type
Hydrostatic
Pulse-Echo
Pulse-Radar
Failure
mode
on
off
Mode
Standard
The measurement loop configuration is usually provided as factory setting. Should a
fundamental modification of the measurement
loop configuration be necessary, a „Reset to
single measurement“ must be carried out first
(in the menu range „Additional functions“,
„Reset VEGAMET“). Only then can you
modify the adjustments in the menu items
Application, Sensor type, Mode and Option.
You should, therefore, work through the individual menu items in the given sequence. The
selection or adjustment options always depend on the adjustment in the previous menu
item. If, for example, in the menu item Application "Level" has been set, only a sensor
type that is suitable for level detection can be
selected in the menu "Sensor type".
As soon as you have carried out and confirmed an adjustment in the menu item „Option“, the following message appears on the
display: „TAG being created!“.
Option
no
options
Sensor
coordination
Position A
Input
no.1
TAGname
TAG No. 1
Failure
mode
on
off
In the menu item „TAG-Name“ you can assign
any name to the measurement loop (see
chapter „6.1 Configuration of measurement
loop in the Reduced menu“). If you chose
„Failure mode off“, the fail safe relay and the
failure indication on the front plate become
nonoperatable. In the following illustrations
further application modes are shown. In addition, you will see there which adjustments you
have to carry out in the menu items „Sensor
type“, „Mode“ and „Option“.
Mode:Standard
Option:none
Parameter: TAG1 process pressure
Pos. APos. A
Pos. A
Level measurement
Pos. APos. APos. A
Gauge measurement
Pos. APos. A
Distance measurement
7.2 Configuration of temperature measurement loop
With pressure transmitters and ultrasonic
sensors, it is possible to create an additional
measurement loop for temperature measurement. One requirement is that a sensor be
used which delivers a temperature signal.
After having selected a suitable sensor type
for configuration of measurement loop 1,
VEGAMET 614V creates automatically measurement loop 2. In the highest menu level
(measured value indication), the temperature
is displayed. Parameter adjustment of the
temperature measurement loop is like adjustment of measurement loop 1.
Here you can call another VEGAMET (e.g.
Met 05) and thereby specify that the input
signal should not be taken from a directly
connected sensor, but from a VEGAMET
connected via DISBUS (e.g. VEGAMET
614V). The called VEGAMET number corresponds to the DISBUS address of the instrument (see chapter 4.1).
Sensor
optimisation
Max.
meas.
range
5,00
For radar/ultrasonic sensors: see
MINICOM menu schematic in the
VEGASON/VEGAPULS operating instruc-
tion manual under chapter „Adjustment
with VEGAMET or MINICOM“.
If another VEGAMET has been selected, the
menu item „Input undefined" appears in
which you can set the input number of the
other VEGAMET (see illustration).
The sensor characteristics values are delivered by the connected sensor and cannot be
modified here. The sensor menu is delivered
by the connected sensor and enables configuration of the sensor from VEGAMET.
DISBUS
Ein-
gang
Ein-
gang1
%
100
+
-
OKESC
614
Ser.No.
12345678
CONNECT
12
on
TAGNo. 1
%
35,9
%
100
+
-
CONNECT
OKESC
MET 05
input 1
Ser.No.
12345678
614
12
on
„local MET“
28VEGAMET 614V
Page 29
Settings in the „Extended menu“
7.4 Configuration of outputs
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Param.
TAGNo. 2
Configuration
Config
inputs
Config
meas.
loop
Config
outputs
Config
current
output
Current
1 to
TAG No. 1
TAG 2
----
Config
Volt
output
Volt1
to
TAG No. 1
TAG 2
----
Current
2 to
TAG No. 2
TAG 1
----
PC/DCS output and DIS output:
The PC/DCS signal as well as the DIS signal
are transmitted via the DISBUS. The PC/DCS
output delivers output signals from
VEGAMET to the interface converter
VEGACOM 557. This interface converter
transfers the signals from VEGAMET via
different standard protocols to connected
systems (DCS/DCS). In the menu items „DCS
1 to“ up to „DCS 7 to…“ the measurement
loop can be assigned or switched off. In the
menu items „PC/DCS relay status“ and „PC/
DCS input status“ it can be specified that the
relay mode is transmitted via the DISBUS. Up
to seven external indicating instruments
(VEGADIS 174) can be connected to
DISBUS. In the menu items „DIS 1 to…“ up to
„DIS 7 to–“ the measurement loop can be
coordinated or switched off.
Config
Config
PC/
DCSoutput
PC/
DCS
meas.
values
DCS 1
to
TAG No. 1
----
VEGADIS
DIS 1
to
TAG No. 1
----
PC/DCS
relay
status
off
DCS 7
to
----
DIS 7
to
----
PC/DCS
input
status
off
V olt2
to
TAG No. 2
TAG 1
----
Config
relay
output
Operating
relay
Rel. 1
to
TAG No. 1
TAG 2
----
Fail
safe
relay
Rely
Standard
Rel. 2
to
TAG No. 2
TAG 1
----
The current, volt and relay outputs with
number 1 (current 1, Volt 1 and Rel. 1) are
assigned as factory setting to measurement
loop „TAG-No. 1“. If a second measurement
loop exists, as is the case when a sensor
with temperature output is connected, then
the outputs with number 2 are assigned to
measurement loop TAG-No. 2. All outputs can
be assigned individually to the other measurement loop or switched off (----), i.e. the
switched off outputs do not access a measurement loop.
A hold function or a reset of alarm functions
can be assigned to each operating relay. In
this case, the menu item „Input“ appears.
Here, the relay can be assigned to „Input 4“
(correction input). For a description of the
relay functions, see the following pages.
VEGAMET 614V29
Page 30
7.5 Adjustment
TAG No. 1
%
xx,x
Param.
Param.
TAG-
TAG-
No. 2
No. 1
Adjustment
like T A G- No. 1
Settings in the „Extended menu“
with
medium
Minadjust
at %
0,0
Max adjust
at %
100,0
without
medium
Adjustment
bar
Offset
correction
Sensor
unpressur’d?
OK ?
Offset
corr.
Now !
OK ?
0%
at
bar
0,000
100%
at
bar
5,000
The adjustment with medium corresponds
to the procedure in the reduced menu (see
appropriate chapter).
- if the medium is changed later on (medium
with another density value) and no new
adjustment can be made
or
- if the MET indication and the DIS outputs
relate to height in meters
note the following:
In applications with hydrostatic pressure
transmitters, the adjusted density value (see
menu „Parameter“) must correspond to the
value of the medium.
For the adjustment without medium, two
anticipated values for level, pressure etc.
corresponding to 0 % and 100 % must be
entered.
The menu item „Offset correction“ is activated
when hydrostatic pressure transmitters are
connected. The offset correction should be
carried out after installation of the sensor, as
the sensor values can differ slightly due to
different installation positions or strong tightening during mounting.
With the offset correction, the meas. data of
the unpressurised sensor (vessel empty) are
detected as correction values and taken into
account in all subsequent measurements.
30VEGAMET 614V
Page 31
Settings in the „Extended menu“
7.6 Conditioning
TAG No. 1
%
xx,x
Param.
Param.
TAG -
TAG-
No. 1
No. 2
like T A G- No. 1
Adjust-
Signal
ment
conditioning
Scaling
0%
corresponds
0
Integr
Lin.
ation
curve
time s
0
linear
cylindrical tank…
Deci-
100%
mal
corres-
point
ponds
888,8
1000
Density in
Kg/dm
1,000
prop.
to
undefined
3
Unit
– –
The menu items „Scaling“ and „Integration
time“ correspond to the menu items in the
„reduced menu“ (see appropriate chapter).
The menu item „Density…“ is displayed as
an option when under „Configuration measurement loop“ the appropriate sensor type
has been selected. The density value must,
before adjustment with medium, be so adjusted that it corresponds to the value of the
medium (water has a density of 1.0 kg/dm
3
The following linearisation curves are available:
- linear (factory setting)
- horiz. cylindrical tank
- spherical tank
- Lin-curve 1
- Lin-curve 2
- Lin-curve 3
Lin-curves 1 … 3 can be edited by the
user (see appropriate chapter).
Meas.
value
limitation
nega-
Failure
tive
mode
values
Stan
yes
dard
>110%
After scaling or linearisation, the required
output must be assigned to the scaled or
linearised value (menu item "prop. to..." and
"Unit...") in the menu range "Parameter adjustment outputs".
Under Meas. value limitation, you can exclude negative values.
Failure mode Standard
means that at a sensor current of 3.6 mA or
).
21 mA, the fail safe relay responds.
Failure mode >110%
also causes a fault signal if a meas. value of
more than 110 % or less than -10 % is
reached.
VEGAMET 614V31
Page 32
7.7 Parameter adjustment of outputs
Current outputs/Volt outputs
TAG No. 1
%
xx,x
Param.
Param.
TAG -
TAG-
No. 1
No. 2
like T A G- No. 1
Out-
Signal
Adjustment
conditioning
puts
Settings in the „Extended menu“
Current
outputs
Current
output
1
prop.
to
Percent
VolumeT ARE%
Current
output 2
Unit
0,0%
like current
output 1
Current
output
4/20mA
0/20mA
20/4mA
20/0mA
free
Volt
outputs
Volt
output
1
prop.
to
Percent
Volum eTAR E %
Current at
0%
4,000
0,0
..20,0
In the menu item „prop. to“, you can choose
between percent and volume. „Prop. to volume“ is then useful when under „Signal conditioning - Linearisation curves“, for example,
a spherical tank or an individually defined
linearisation curve is selected. The output
current or the output voltage is then proportional to the volume of the tank contents.
Unit:
Selection list dependent on the choice in the
previous menu item „prop. to“ (see chapter
„8 Measured quantity and units“).
Volt output/Current output:
If „free“ is chosen, the following (light grey)
menu items appear:
Volt
output 2
like Volt output 1
Unit
0,0%
Current
at
100%
20,000
0,0
..20,0
Volt
output
0/10V
10/2V
10/0V
2/10V
free
Failure
mode
0mA
22mA
- 0%
100%
Volt
at
0%
0,000
0,0
..10,0
Current
limitation
on
off
Volt
at
100%
10,000
0,0
..10,0
Failure
mode
0V
11V
- 0%
100%
- Volt at 0 %/100 %:
Individual voltage values between 0 V and
+10 V can be set.
- Current at 0 %/100 %:
Individual current values between 0 mA
and +20 mA can be set.
Failure mode:
The height of the output current/output voltage in case of failure is determined. If „--“ is
chosen, the actual value is maintained.
Volt limitation/Current limitation on:
Voltage/current remain within the values
which are given in the menu item “Voltage
output/current output”.
Volt
limitation
on
off
32VEGAMET 614V
Page 33
Settings in the „Extended menu“
Relay outputs
TAG No. 1
%
xx,x
Param.
Param.
TAG-
TAG-
No. 1
No. 2
like T A G- No. 1
Signal
conditioning
Outputs
Current
outputs
Volt
outputs
Relay
outputs
Relay
output
1
prop.
to
Percent
Volume
Height
Pressure
Adjustment
Relay
output
2
like relay output 1
Unit
Mode
Overfill
protec
0,0%
0,00 %
Tare %
Low
%
0,0
Dry run protec.
Switching window on
Switching window off
Tendency rising
Tendency falling
High
%
100,0
Deviation
s
1,0
Deviation
per
time
Cycle
time
s
1
Failure
mode
off
hold
Number
of
cycles
10
Add’l
functions
Switching
delay
t on
s
0
t off
s
0
In the menu item „prop. to“, you choose a
parameter that the relay will react to. You can
choose between percent and volume („prop.
to volume“ is useful when under „Conditioning
- Linearisation curves“, for example a spherical tank or a user-programmable lin. curve is
selected). In the next menu item you can
select the suitable Unit. In the menu item
„Mode“ the mode of the relay can be se-
With all other modes, the menu items „Low“
and „High“ appear. There you can determine
the switching points for the relay.
The menu item „Failure mode“ always ap-
pears. „Failure mode off“ causes the relay to
deenergise when a failure (e.g. sensor
shortcircuit) occurs. „Failure mode hold“
causes the actual relay status to be main-
tained when a failure occurs.
lected. The individual relay modes are described on the following pages.
Under „Add’l functions“ you can adjust a
switch on delay (t on) and a switch off delay
If mode „Tendency rising“ or „Tendency falling“ is selected, the menu items „Deviation“,
„Dev. per time“, „Cycle time“ and „Number of
(t off) of the relay of max. 600 s (with the relay
modes „Tendency rising“ or „Tendency fall-
ing“, no switching delay can be set).
cycles“ appear.
VEGAMET 614V33
Page 34
Settings in the „Extended menu“
Relay mode: overfill protection
The relay deenergises at switching point
„High“ (safe condition).
Relay mode: dry run protection
The relay deenergises at switching point
„Low“ (safe condition).
Relay mode: switching window on
The relay of relay output 1 is energised within
the switching window.
Parameter
High
Low
t
Relay output
Parameter
High
Low
t
Relay output
Parameter
High
Low
t
Relay output
Relay mode: switching window off
The relay of relay output 1 is deenergised
within the switching window.
Parameter
High
Low
t
Relay output
34VEGAMET 614V
Page 35
Settings in the „Extended menu“
Tendency determination
The level change within the cycle time (ta) is determined and after completion of all cycles (n)
an average value is generated from the sum of the level changes. If this average value exceeds a previously defined % value, the tendency determination responds, i.e. an energised
relay deenergises. If „Tendency rising“ or „Tendency falling“ are selected, the menu items
„Deviation in %“ and „Deviation per time“ are activated.
%
t ges.
t
a
–– Mode Tendency rising
The relay deenergises when the
rising average value of „t tot.“
exceeds the previously defined %
value.
Tendency falling
Relay output
ta = cycle time
n = number of cycles
%
4 n=4
321
t ges.
t
a
same dependence, however with
falling average value
4 n=4
321
Relay output
ta = cycle time
n = number of cycles
–– Deviation in %2If the level change exceeds the % value set
here, the tendency determination responds.
0 … 110 %
Example 2 % (change) per t tot.
–– Deviation per time
t
t
–– Cycle time
in s60Adjustment range 0 … 999 s
Example 60 s correspond to 1 minute
–– Number of
cycles4Adjustment range 0 … 99 cycles
Example: t tot = ta • n = 60 • 4 = 240 s = 4 min,
i.e. after every 4 min., the average value of the
deviation is generated and the result is compared with the given deviation in %, here in the
example = 2 %
VEGAMET 614V35
Page 36
MET indication/PC/DCS outputs/DIS outputs
TAG No. 1
%
xx,x
Param.
Param.
TAG-
TAG-
No. 1
No. 2
like T A G- No. 1
Signal
conditioning
Outputs
Adjustment
Settings in the „Extended menu“
Current
outputs
Volt
outputs
Relay
outputs
MET
display
prop.
to
Percent
scaled
Height
…
MET indication/Unit:
In the menu item „prop. to“, you choose the
parameter to which the VEGAMET display
reacts (if you have carried out a scaling under the menu item „Signal conditioning“,
„prop. to scaled“ will have been automatically
set, i.e. the display will indicate the scaled
value). „Prop. to volume“ is then useful if you
have selected, for example, a spherical tank
or a user-programmable lin. curve under
„Signal conditioning - Linearisation curves“.
The indicated values of the VEGAMET display then react proportionally to the volume of
the tank contents.
DCS
output 1
prop.
to
Percent
scaled
Height
…
DIS
outputs
DIS
output
1
prop.
to
Percent
scaled
Height
…
Unit
0,0%
Unit
0,0%
DIS
output 7
like DIS output 1
DCS
output
7
like DCS output 1
Unit
0,0%
PC/
DCS
outputs
PC/DCS outputs and DIS outputs:
Are adjusted according to the same procedure as the MET indication.
36VEGAMET 614V
Page 37
Settings in the „Extended menu“
7.8 Simulation7.9 Special function: Reset
TAG No. 1
%
xx,x
Param.
Param.
TAG-
TAG -
No. 2
No. 1
like TAG-No. 1
Out-
Adjustment
Signal
conditioning
puts
Simulation
Simulation
Now!
OK?
Simulation
%
XX,X
To check the outputs and connected instruments, you can adjust in this menu item any
individual percentage value with the keys [+]
and [–]. The starting point is always the actual measured value. The indicated value
flashes when the simulation is activated. 15
minutes after setting the simulated value, the
simulation is automatically terminated and the
display reset to measured value indication.
TAG No. 1
%
xx,x
Param.
Param.
TAG-
TAG-
No. 2
No. 1
like T A G- No. 1
Out-
Simu-
Adjustment
Signal
conditioning
puts
lation
Special
functions
Reset
Level
Reset
OK ?
Reset
Now!
OK ?
With this reset, all parameter values of the
measurement loop TAG-No. 1 are reset to
factory settings (empty and full adjustment,
relay outputs, current output, volt output and
scaling of the meas. value indication are
reset).
VEGAMET 614V37
Page 38
7.10 Password, language, menu mode
See chapter „6 Settings in the reduced menu“.
TAG No. 1
%
Settings in the „Extended menu“
Param.
TAGNo. 1
Param.
TAGNo. 2
Configuration
Add’l
functions
Password
off
lin.
curve
InfoEdit
Language
English
7.11 Special function: Manual correction
TAG Nr. 1
%
xx,x
Param.
TAGNr. 1
Adjustment
Param.
TAGNo. 2
like T A G- No. 1
Signal
conditioning
Outputs
Simulation
Special
functions
Reset
Level
Manual
corr.
Offset
correction
Manual offset correction
(only with hydrostatic pressure transmitters)
This correction can be carried out to compensate changes of the sensor values due to
ageing. In order to do this, the sensor must
be in an unpressurized condition (empty
vessel).
Manual real value correction
In the menu item „Correction % 0.0“ you can
enter a percentage value which corresponds
to the actual level.
Real
value
corr.
Reset
VEGAMET
Service
Menu
mode
Extended
Reduced
Offset
correc tion
OK?
Correction
now!
Real
value
corr.
OK?
Correction
%
0,0
Correction
now!
OK ?
38VEGAMET 614V
Page 39
Settings in the „Extended menu“
7.12 Linearisation curves
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Param.
TAGNo. 2
Configuration
Add’l
functions
Pass
word
off
Edit
lin.
curve
Lin.
curve1
Lin.
curve2
Lin.
curve3
Add
lin.
point
x %
0,0
y V%
0,0
like LIN-curve 1
Edit
lin.
curve 1
x 0 %
y 0 V%
0,0
Linearisation curve 1 … 3
A linearisation curve is generated by a certain number of index markers and their value
pairs. A value pair consists of a value for
level percent (X %) and a value for volume
percent (Y V %). Up to max. 32 index markers can be added. A linearisation curve can
be terminated after any individual number of
index markers. The signal conditioning instrument terminates the linearisation curve automatically with the values X = 100 % and Y
= 100 %.
The data for the value pairs can be determined by incremental filling or can be taken
from a table of values provided by the vessel
manufacturer.
An example of incremental filling
Initial situation:
- you have already carried out the 0 % and
100 % adjustment
- the total volume of the vessel is known to
you, for example 300 m
- the gauging volume in this example was
set at 15 m
3
3
Delete
lin.
point
x 1 %
0,0
100,0
y 1
V%100,0
Index marker number
(0…32)
x 2 %
100,0
y 2
V% 100,0
x 32 %
100,0
y 32V%
100,0
x 0,0
y 0,0
delete
Delete
Now?
1st index marker
- fill the vessel with the liquid quantity of the
gauging volume of 15 m
3
- the value for level percent is indicated on
VEGAMET, transfer this value into the linearity protocol, column X %
- the value for volume percent must be calculated acc. to the following formula
100 % x gauging volume
Y 1 V = ––––––––––––––––––––––––
total volume
100 % x 15 m
= ––––––––––––––– = 5 %
300 m
3
3
Transfer this value also into the linearity
protocol, column Y V %.
2nd index marker, etc.
See also the following illustration.
x 100,0
y 100,0
delete
Delete
Now?
VEGAMET 614V39
Page 40
Illustration of a linearisation curve
Settings in the „Extended menu“
X (in %)
100
80
60
40
20
Add
lin.
point
x %
0,0
y V%
0,0
With [OK] the new value pair
is automatically placed at the
correct position in the
linearisation curve. The already present index markers
are re-sorted.
0
X1 %
12,0
Y1 V %
5,0
Edit
lin.
curve 1
x 0 %
0,0
y 0 V%
0,0
x 1 %
100,0
y 1
V%100,0
x 2 %
100,0
y 2
V% 100,0
Select the index markers
1 … 32 with [–>] or [<–], edit
the Y value with [+] or [–],
accept with [OK].
406080100
Delete
lin.
point
x 0,0
y 0,0
delete
Delete
now?
X32 %
100,0
Y32V %
100,0
x100,0
y100,0
delete
Delete
now?
To delete the index markers,
select the value pairs with [–
>] or [<–]. With [OK] the
value pair and the corresponding index marker are
deleted from the lin. curve.
The remaining index markers
move up.
V (in %)20
40VEGAMET 614V
Page 41
Settings in the „Extended menu“
Linearity protocol
A linearisation curve is entered in the menu item "Add. lin. point“.
Linearisation curve 1
IndexValue pair
marker no. X %Y V %
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
29
30
31
32
Linearisation curve 2
IndexValue pair
marker no. X %Y V %
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
29
30
31
32
Linearisation curve 3
IndexValue pair
marker no. X %Y V %
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
29
30
31
32
Note
Date
Name
VEGAMET 614V41
Page 42
7.13 Info
TAG No. 1
%
xx,x
Param.
Param.
TAG-
TAG-
No. 1
No. 2
Configuration
Add’l
functions
Password
off
Edit
lin.
curve
Info
Input
info
VEGAMET
info
Type
MET614
Program
info
Program
info
xxxxxx
Serial
number
xxxx
xxxx
Settings in the „Extended menu“
Meas.
loop
info
TAG 1
TAG 1
TAG -
TAG -
No. 1
No. 2
like TAG-No. 1
Max -
Max -
Min-
Minadjust
at %
0,0
Instr.
addr.
adjust
bar
0,000
Softw.
vers.
V1.13
06
1
adjust
at %
100,0
Softw.
date
47/97
adjust
bar
1,000
Param.
vers.
5
Parameter
Level
In-
In-
put2
put1
VBUS
like input 1
For radar/ultrasonic sensors: see
MINICOM menu schematic in the
VEGASON/VEGAPULS operating
instruction manual under chapter „Adjust-
ment with VEGAMET or MINICOM“.
All menu items under „Info“ only have display
function.
42VEGAMET 614V
Page 43
Settings in the „Extended menu“
7.14 Reset VEGAMET
TAG No. 1
%
Param.
TAGNo. 1
Param.
TAGNo. 2
Configuration
Add’l
functions
Password
off
Lin.
curve
Info
Language
English
Reset
VEGAMET
Reset
configuration
TAG 1
to
default
Reset
OK?
Reset
Now!
OK?
TAG 2
to
default
Reset
OK?
Reset
Now!
OK?
Reset
sensor
values
to
single
meas.
Reset
TAG1
Delete
TAG 1?
Reset
OK?
Reset
Now!
OK?
Reset
TAG2
Delete
TAG2?
Reset
OK?
Reset
Now!
OK?
Reset
lin.curves
Reset
lin.
curve1
Reset
OK?
Reset
Now!
OK?
Curve 2
and 3
like 1
Reset
all
curves
Delete
all
curves?
Reset
OK?
Reset
Now!
OK?
With the function „TAG 1 to default“, you
reset the whole configuration and all parameter adjustments to factory setting. However,
With the function „Reset lin-curves“, you can
delete your manually edited linearisation
curves.
a measurement loop configuration carried out
later will remain.
With „Reset all curves“, you delete all to-
gether, with „Reset lin-curve 1“ you always
With the function „Reset T A G 1“, the basic
delete only one curve.
configuration is also deleted. Then, the measurement loop must be created completely
anew (config. meas. loop).
The three pre-programmed lin-curves „linear,
horiz. cylindrical tank, spherical tank“ cannot
be deleted.
With „Reset sensor values“, the adjusted
sensor data of the connected sensor are
deleted.
VEGAMET 614V43
Page 44
8 Measured quantity and units
Measuring results and units
With all outputs and with scaling, you find the
menu items „prop. to“
and „Unit“. Here you
can choose a measured quantity and a suitable unit. The list of the units depends on the
selected measured quantity. In the table, the
available units are listed.
Current/V olt output
prop. toUnit
Percent0.0 %
Vol umeV %
VEGADIS/DCS/MET indication
prop. toUnit
Percent0.0 %
0.00 %
Vol umeV %
Heightm
Pressurebar
scaled(like with
scaling)
Distancem (d)
Scaling
prop. toUnit
Percent%
Vol umem
Distancem (d)
Heightm
Pressurebar
Masskg
Densitykg/dm
3
hl
l
gal
V %
3
ft
3
in
dm
cm
mm
yd
ft
in
mbar
psi
hPa
t
lb
g/cm
3
3
Relay outputs
prop. toUnit
Percent0.0 %
Vol umeV %
Heightm
Pressurebar
Distancem (d)
44VEGAMET 614V
Page 45
Diagnosis
9 Diagnosis
9.1 Maintenance
The instrument is maintenance-free.
9.2 Simulation
Note
When the simulation is activated, the entire
display flashes. The measurement data delivered by the sensors are not released for
processing. You should, ther efore, quit this
menu item as soon as possible. After 15
minutes, there is an automatic reset to meas.
value indication.
9.3 Fault signal
The signal conditioning instrument and the
connected sensor are continuously monitored during operation. During parameter
adjustment, the adjusted values are checked
for plausibility. In both cases, a fault signal
can be triggered by irregularities.
In case of failure, an error code is displayed
(only within the meas. value indication), the
fail safe relay deenergises, the signal lamp
lights and the outputs react according to their
preset operational mode.
Configuration
For each individual measurement loop, you
can determine if a fault signal should be triggered or not. See menu range „Configuration
measurement loop“, menu item „Fault signal“.
Parameter adjustment
In the menu item „Failure mode“ you can
determine individually for each measurement
loop whether and how an overrun of the
measuring range should trigger a fault signal.
9.4 Repair
For safety and warranty reasons, any internal
work on the instruments, apart from that
involved in normal installation and electrical
connection, must be carried out only by
VEGA personnel.
In case of defect, please return the affected
instrument with a short description of the fault
to our repair department.
The error codes are listed in section „9.5
Error codes“ on the following pages.
The diagnosis info is transmitted as a scrolling text info.
VEGAMET 614V45
Page 46
9.5 Error codes
Error codeMeaning
Diagnosis
E004This operation is not supported by the instrument (different software versions).
E007Sensor type does not match.
E008VBUS sensor not found. Sensor cannot be found.
E010The serial number of the connected VBUS sensors are not yet assigned to an
E013VBUS sensor signals failure; however the communication with the sensor is
You have tried to activate a function with a newer software version of VEGA
Visual Operating in an instrument with older software version. However, the
instrument cannot support this operation.
Have the software of your instrument updated.
A VBUS sensor with the serial number given by you was found, however the
sensor type does not match the settings.
Check the entered serial number and modify the configuration.
Probable reasons:
- no sensor connected
- no sensor with this serial number connected
- line break (VBUS)
- short-circuit (VBUS)
- sensor is defective
- interfered communication (VBUS)
Also possible:
- failure on the VBUS transmitter/receiver part of VEGAMET
Check the serial number, the connection cable, the sensor and if necessary
VEGAMET.
input component.
okay.
Probable reasons:
- failure on the VBUS sensor
Check the VBUS sensor.
E016Empty/full adjustment reversed. The values for the empty adjustment are
E017Adjustment span too small. The adjustment values are too close together or are
E021Span of the scaling values too small or values reversed. The scaling values are
46VEGAMET 614V
above the values for the full adjustment. Carry out the adjustment again.
identical. Carry out the adjustment again.
too close together or are identical. Please check the scaling.
Page 47
Diagnosis
Error codeMeaning
E022Scaling span too big.
E027Actual measured value is less than -10 %.
E28Actual measured value is more than 110 %.
E031Same DISBUS address. At least one other instrument with the same DISBUS
address is connected to the DISBUS. Please check the DISBUS address.
E032Instrument address 0. Instrument address 0 is adjusted on VEGAMET. With this
address VEGAMET does not participate in the DISBUS communication, even if
the DISBUS is wired. If necessary, modify the instrument address.
E034Instrument error. A serious error was detected in the instrument (possibly de-
fective EEPROM). Disconnect voltage and then connect again (cold start). If the
error is not eliminated, please contact your VEGA agency.
E035Instrument error. A serious error was detected in the instrument (check sum
error in EPROM). Disconnect voltage and then connect again (cold start). If the
error is not eliminated, please contact your VEGA agency.
E050Remote VEGAMET not found. The VEGAMET from which a measured value is
expected cannot be found.
Reasons:
- the remote VEGAMET is not connected
- no VEGAMET with this instrument address connected
- line break (DISBUS)
- short circuit (DISBUS)
- the remote VEGAMET is defective
- interfered communication (DISBUS)
Also possible
- failure on the DISBUS transmitter/receiver part of VEGAMET
Please check the instrument address on both VEGAMETs, the connection cable, the remote VEGAMET and, if necessary, this VEGAMET.
VEGAMET 614V47
Page 48
Error codeMeaning
Diagnosis
E051Failure during transmission of measured values via DISBUS. It is not possible
to receive a valid measured value via DISBUS from the remote VEGAMET; the
communication with the remote VEGAMET, however, is okay.
Reason:
- failure on remote VEGAMET
- the input on the remote VEGAMET which should be transmitted is not yet
configured
E052The input number of the VEGAMET from which a measured value is expected
has not yet been assigned.
E060Span of the current output values too small. The values for the current output
are too close together or are identical. Please check the adjustment of the
current output.
E061Span of the voltage output values too small. The values for the voltage output
are too close together or are identical. Please check the adjustment of the
voltage output.
E063Incorrect sensor characteristics values. Please check the sensor characteris-
tics values
E102A reset of the configuration was carried out on this VEGAMET via the key-
board.
E110Switching hysteresis too small. The switching points are too close together or
are identical.
E111Relay switching points swapped. The relay switching points are reversed with
respect to the selected relay mode.
48VEGAMET 614V
Page 49
Menu schematic "Reduced menu"
10Menu schematics
10.1 Menu schematic "Reduced menu"
TAGNo. 1
%
xx,x
Param.
TAGNo. 1
Adjustment
TAGNo. 2
°C
xx,x
Configuration
Sensor
type
Hydrostatic
The sensor type can
only be modified,
when a „Reset TAG
1“ has been carried
out first.
VEGASON/VEGAPULS operating instruc-
tion manual under chapter „Adjustment
Scaling
TAG
1-2
xx,x
xx,x
Add’l
functions
Password
off
Sensor
optimisation
For radar/ultrasonic sensors: see
MINICOM menu schematic in the
with VEGAMET or MINICOM“.
Signal
conditioning
Integr
ation time
s
0
Display of
measured
value
Language
English
TAGname
TAG No. 1
Outputs
TAG 1
to de
fault
Reset
OK ?
Reset
Now !
OK ?
Reset
TAG 1
Delete
TAG
1?
Reset
OK ?
Reset
Now !
OK ?
Menu
mode
Reduced
Sensor
type
Hydrostatic
Language
English
Current
output 2
4/20mA
Bold print/italic is the
sensor or measured value
information which cannot
be modified in this position.
In white letters you can see
the parameters which can be
modified with the [+] or [–]
key and saved with the [OK]
key.
Light grey menu items are
only displayed if necessary (depending on the
adjustment in other menu
items).
Simulation
Simulation
Now !
OK ?
like relay output 1
Low
%
0,0
Simulation
%
xx,x
High
%
100,0
ESC
OK
With these keys you
move in the menu field
with
medium
Minadjust
at %
0,0
0%
corresponds
Maxadjust
at %
100,0
Volt
output
2
0/10V
Relay
output
1
Mode
Overfill
protec
Relay
output
2
prop.
to
undefined
100%
at
bar
1,000
Volt
output
1
0/10V
Unit
– –
Current
Current
output 2
output 1
4/20mA
4/20mA
100%
corresponds
1000
0
without
medium
Adjustment
bar
Decimal
point
888,8
0%
at
bar
0,000
VEGAMET 614V49
Page 50
TAG 12
xx,x
xx,x
TAG No. 1
%
xx,x
Param.
TAG No. 1
Adjustment
TAG No. 2
°C
xx,x
Param.
TAGNo. 2
like TAG-No. 1
Signal
conditioning
Outputs
Menu schematic "Extended menu"
10.2 Menu schematic "Extended menu"
Configuration
see following pages
Add’l
functions
34
with
medium
Current
outputs
Current
output
1
prop.
to
Percent
Scaling Lin. curve
linear
100%
0%
corres-
corres-
ponds
ponds
1000
0
Volt
outputs
Volt
output
1
prop.
to
Percent
Current
output
2
like current output 1
Unit
0,0%
Density
Integr
in
ation
Kg/dm
time s
0
1,000
prop.
Deci-
to
mal
unde-
point
fined
888,8
without
medium
Volt
output
2
like V olt output 1
Unit
0,0%
Current
output
4/20 mA
3
Volt
output
0/10V
Current
at
0%
4,000
Unit
– –
Volt
at
0%
0,000
Current
at
100%
20,000
Meas.
value
limitation
negative
values
yes
Volt
at
100%
10,000
Failure
mode
0mA
Failure
mode
Stan
dard
Failure
mode
0V
Current
limitation
on
Volt
limitation
on
Minadjust
at %
0,0
Maxadjust
at %
100,0
Adjustment
bar
Offset
correction
Sensor
unpressur’d ?
OK ?
Offset
corr.
now!
OK ?
0%
at
bar
0,000
100%
at
bar
1,000
ESC
OK
With these keys you move
in the menu field
50VEGAMET 614V
Page 51
Menu schematic "Extended menu"
Relay
outputs
Relay
output
1
prop.
to
Percent
1 23
Sensor
type
Hydrostatic
MET
display
prop.
to
Percent
Relay
output
2
like relay output 1
Unit
0,0%
… setup sequence
Unit
0,0%
Mode
Overfill
protec
PC/
DCS
outputs
PC/
DCS
output1
prop.
to
Percent
Low
%
0,0
Bold print/italic is the sensor or
measured value information
which cannot be modified in
this position.
DISoutputs
DIS
output1
prop.
to
Percent
PC/
DCS
output2
like DCS output 1
Unit
0,0%
High
%
100,0
DIS
output2
like DIS output1
Unit
0,00%
PC/DCS
output7
Deviation
%
1,0
DIS
output7
Deviation
per
time
Cycle
time
s
1
Simulation
Simulation
Now!
OK?
Simulation
%
XX,X
Failure
mode
off
Number
of
cycles
10
Special
functions
Reset
Level
Reset
OK ?
Reset
Now!
OK ?
Add’l
functions
Switching
delay
t on
s
0
Offset
corr.
Offset
corr.
OK?
t off
s
0
Manual
corr.
Correction
now !
OK ?
Real
value
corr.
Real
value
corr.
OK?
Correction
%
0,0
Correction
Now !
OK ?
Language
English
In white letters you can see the
parameters which can be modified with the [+] or [–] key and
saved with the [OK] key.
Current at
100%
20,000
Light grey menu items are only
displayed if necessary (depending on the adjustment in
other menu items).
VEGAMET 614V51
Page 52
TAG 12
xx,x
xx,x
TAG No. 1
%
xx,x
Param.
TAG No. 1
TAG No. 2
°C
xx,x
Param.
TAGNo. 2
Configuration
Add’l
functions
Menu schematic "Extended menu"
see previous pages
Config
inputs
2
Input
no. 1
1
Config
meas.
loop
TAG 1
Level
Application
Level
1)1)1)
1)
Input
no. 2
Config
outputs
TAG 2
Temperature
Sensor
type
Hydrostatic
Config
Volt
output
V olt1
to
TAG No. 1
Current
2to
TAG No. 2
Fault
signal
on
Mode
Standard
Config
current
output
Current
1 to
TAG No. 1
TAGname
TAG No. 1
The parameters in these menu items
can only be modified, when a „Reset
to single measurement“ has been
carried out first.
Volt2
to
TAG No. 2
Option
no
option
1)
see following pages
Config
relay
output
Fail
Op-
safe
era-
relay
ting
relay
Relay
Standard
Rel. 2
Rel. 1
to
to
TAG -
TAG -
No. 1
No. 1
Sensor
coordination
Position A
Input1
TAGname
TAG No. 1
Config
PC/
DCS
output
PC/
DCS
meas.
values
DCS 1
to
TAG No. 1
Fault
signal
on
PC/DCS
relay
status
off
Config
VEGADIS
DIS 1
to
TAG No. 1
DCS 7
to
----
DIS 7
to
----
like input no. 1
Input
from
local
Met
Serial
no.
xxxx
xxxx
Input
undefined
Sensor
characteristics
Min.
meas.
range
0,00
Max.
meas.
range
5,00
Sensor
optimisation
For radar/ultrasonic sensors: see MINICOM
menu schematic in the VEGASON/VEGAPULS
operating instruction manual under chapter
„Adjustment with VEGAMET or MINICOM“.
5 … 10
52VEGAMET 614V
Page 53
Menu schematic "Extended menu"
2
1
3
… setup sequence
Bold print/italic is the sensor or
measured value information
which cannot be modified in
this position.
In white letters you can see the
parameters which can be modified with the [+] or [–] key and
saved with the [OK] key.
Light grey menu items are only
displayed if necessary (depending on the adjustment in
other menu items).
ESC
OK
With these keys you move
Sensor
type
Hydrostatic
Language
English
Current at
100%
20,000
in the menu field
VEGAMET 614V53
Page 54
TAG 12
xx,x
xx,x
TAG No. 1
%
xx,x
Param.
TAGNo. 1
TAG No. 2
°C
xx,x
Param.
TAG No. 2
Configuration
Add’l
functions
Menu schematic "Extended menu"
see previous pages
Password
off
Edit
lin.
curve
see previous pages
InfoLang-
Meas.
Program
info
xxxxxx
Serial
number
xxxx
like input 1
Program
info
xxxx
loop
info
TAG 1
TAG No. 1
Minadjust
at %
0,0
Instr.
addr.
1
Input
info
Input1
VBUS
instruction manual under chapter „Adjust-
VEGAMET
info
Type
MET
614V
Input2
VBUS
For radar/ultrasonic sensors: see
MINICOM menu schematic in the
VEGASON/VEGAPULS operating
ment with VEGAMET or MINICOM“.
TAG 2
TAG No. 2
Minadjust
bar
0,000
Softw.
vers.
V1.13
06
uage
English
like T AG 1
Maxadjust
at %
100,0
Softw.
date
47/97
Maxadjust
bar
1,000
Param.
vers.
5
Parameter
Level
Lin.
curve1
Add
lin.
point
x %
0,0
y V%
0,0
Lin.
curve2
Lin.
curve3
like LIN-curve 1
Edit
lin.
curve 1
x
y 0 V%
0,0
0 %
0,0
x 1 %
y 1
V%100,0
x 2 %
100,0
100,0
y 2
V%100,0
Index marker number (0…32)
x32 %
100,0
y32
V%100,0
Delete
lin.
point
x 0,0
y 0,0
delete
Delete
now?
x100,0
y100,0
delete
Delete
now?
54VEGAMET 614V
Page 55
Menu schematic "Extended menu"
Reset
VEGAMET
Reset
configuration
TAG 1
to
default
Reset
OK?
Reset
Now!
OK?
Reset
sensor
values
TAG 2
to
default
Reset
OK?
Reset
Now!
OK?
Service
Reset
lin.curves
Reset
lin.
curve1
Reset
OK?
Reset
Now!
OK?
to
single
meas.
Reset
TAG1
Delete
TAG 1?
Reset
Menu
mode
Extended
Reset
lin.
curve2
Reset
OK?
Reset
Now!
OK?
Reset
TAG 2
Delete
TAG2?
Reset
Reset
lin.
curve3
Reset
OK?
Reset
Now!
OK?
Reset
all
curves
Delete
all
curves?
Reset
OK?
Reset
Now!
OK?
Sensor
type
Hydrostatic
Language
English
ESC
OK
With these keys you move
in the menu field
Bold print/italic is the sensor or
measured value information
which cannot be modified in
this position.
In white letters you can see the
parameters which can be modified with the [+] or [–] key and
saved with the [OK] key.
OK?
Reset
Now!
OK?
OK?
Reset
Now!
OK?
Current at
100%
20,000
Light grey menu items are only
displayed if necessary (depending on the adjustment in
other menu items).
VEGAMET 614V55
Page 56
VEGA Grieshaber KG
Am Hohenstein 113
D-77761 Schiltach
Phone(07836) 50-0
Fax(07836) 50-201
E-Mailinfo@de.vega.com
www.vega.com
Safety information
Please read this manual carefully, and also take
note of country-specific installation standards
(e.g. the VDE regulations in Germany) as well
as all prevailing safety regulations and accident prevention rules.
For safety and warranty reasons, any internal
work on the instruments, apart from that involved in normal installation and electrical connection, must be carried out only by qualified
VEGA personnel.
ISO 9001
Note Ex area
Please note the attached safety instructions
containing important information on installation
and operation in Ex areas.
These safety instructions are part of the operating instructions and come with the Ex approved instruments.
All statements concerning scope of delivery, application, practical
use and operating conditions of the sensors and processing systems correspond to the latest information at the time of printing.
Technical data to subject to alterations
2.21 866 / February 2001
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.