VEGA MET614 User Manual

Operating Instructions
VEGAMET 614
Level and Pressure
CONNECT
12
on
614
100
+
-
OKESC
Ser.No. 12345678
out
in
Contents
1 Product description
1.1 Function and configuration ................................................ 4
1.2 Approvals ........................................................................... 5
1.3 Features............................................................................. 5
1.4 Technical data .................................................................... 6
1.5 Dimensions......................................................................... 9
2 Mounting and installation instructions .............................. 10
3 Electrical connection
3.1 Connection instructions ................................................... 12
3.2 Connection instructions for Ex approved applications . 12
3.3 Wiring plan ........................................................................ 12
4 Adjustment
4.1 Indicating and adjustment elements ............................... 13
4.2 Adjustment system .......................................................... 14
4.3 Adj ustmen t via PC ........................................................... 15
4.4 Configuration and parameter adjustment ....................... 16
4.5 Comparison: reduced menu – extended menu .............. 17
Contents
5Setup.......................................................................................... 18
6 Settings in the "Reduced menu“
6.1 Con figurati on of mea surement loop ............................... 19
6.2 Adj ust men t wi th med ium.................................................. 20
6.3 Adjustment without medium ............................................ 21
6.4 Scaling .............................................................................. 22
6.5 Inte gration ti me ................................................................ 23
6.6 Outputs............................................................................. 24
6.8 Password, language, default, reset, menu mode .......... 25
6.7 Simulation ......................................................................... 25
2 VEGAMET 614
Contents
7 Settings in the "Extended menu“
7.1 Config uration o f measur ement loo p ............................... 26
7.2 Configuration of inputs ..................................................... 28
7.3 Configu ration of ou tputs .................................................. 2 9
7.4 Adjustment........................................................................ 30
7.5 Conditioning...................................................................... 31
7.6 Paramet er adjustm ent of outp uts ................................... 32
7.7 Simulation ......................................................................... 37
7.8 Special function: Reset .................................................... 37
7.9 Password, language, menu mode .................................. 38
7.10 Special function: Manual correction ................................ 38
7.11 Li nearisa tion cur ves ........................................................ 39
7.12 Info .................................................................................... 42
7.13 Reset VEGAMET ............................................................. 43
8 Measured quantity and units ................................................ 44
9 Diagnosis
9.1 Maintenance..................................................................... 45
9.2 Simulation ......................................................................... 45
9.3 Fault signal........................................................................ 45
9.4 Repair............................................................................... 45
9.5 Error codes ...................................................................... 46
10 Menu schematics
10.1 Menu schematic “reduced menu” .................................. 49
10.2 Menu schematic “extended menu” ................................. 50
Safety informatio n ...................................................................... 56
Note Ex area .............................................................................. 56
VEGAMET 614 3

1 Product description

Product description

1.1 Function and configuration

VEGAMET 614 is a signal conditioning instru­ment for a number of different applications, e.g.:
- level measurement
- gauge measurement
- process pressure measurement
-etc.
Software configuration
Input component
- continuous current input
- contact input
Function
The signal conditioning instruments power the connected sensors and process their analogue measuring data. The conditioning is done by a special software consisting of functional components (FB), input compo­nents (EB), output components (AB) as well as DCS components (PB).
Configuration
The VEGAMET 614 signal conditioning instru­ment is a module unit with plug-in socket. The front plate contains an adjustment module with LC display and keys.
Output components
- current output
- voltage output
- DISBUS outputs
- failure output
- MET display
- relay outputs
AB
EB FB
PC/DCS components
- PC/DCS outputs
Functional component
- measurement loop TAG 1
4 VEGAMET 614
PB
Product description

1.2 Approvals

Ex approval
As an accessory, intrinsically safe instrument. EC type approval certificate acc. to CENELEC
II (1) G [EEx ia] IIC
WHG approval
Signal conditioning instrument as part of an overfill protection system acc. to WHG ap­plied for.

1.3 Features

- microcomputer-controlled signal condition­ing instrument for continuous measurement
- adjustment module with LC display and 6 keys
- adjustable integration time
- two fixed and three user-programmable linearisation curves
- fault monitoring
- fault signal and failure diagnosis via dis­play
Input:
- 1 sensor input (capacitive electrode, pres­sure transmitter or other 4 … 20 mA sen­sor)
- 1 correction input (for key, switch or level switch)
Outputs:
- 1 current output 0/4 … 20 mA
- 1 voltage output 0/2 … 10 V
- 2 relay outputs (spdt)
- 1 DISBUS output for digital wiring and for connection of VEGADIS 174
- 1 fail safe relay
- adjustment also via PC with adjustment software VVO; connection via the interface adapter of the VEGACONNECT series
VEGAMET 614 5
Product description

1.4 Technical data

Power supply
Operating voltage U
Power consumption approx. 10 VA or approx. 4 W
Meas. data input
Number 1 input Input type active two-wire input, analogue (sensor is
Range 4 … 20 mA Applicable sensors (loop powered) capacitive electrodes, pressure transmitters,
Voltage
- at 4 mA approx. 19.5 V DC
- at 20 mA approx. 16 V DC Current limitation at approx. 26 mA, short-circuit-proof Detection line break < 3.6 mA Detection short-circuit > 21 mA Min. adjustment delta 2 % of the entered sensor values Connection cable 2-wire standard cable (screening
max. resistance per wire 35 Resolution 1 µA Linearity error 0.025 % at 4 … 20 mA Temperatur e error 0.04 %/10 K at 4 … 20 mA
= 20 … 250 V AC, 50/60 Hz
nom
= 20 … 72 V DC
powered by VEGAMET)
process pressure transmitters, differential pressure transmitters, ultrasonic sensor (4 … 20 mA sensor), radar sensor (4 … 20 mA sensor)
recommended)
Current output
Number 1 output Function analogue output of the processing results Range adjustable between 0 … 20 mA Load max. 500 Resolution 1 µA Linearity error 0.05 % (relating to 20 mA) Temperatur e error 0.05 %/10 K (relating to 20 mA)
Voltage output
Number 1 output Function analogue output of the processing results Range adjustable between 0 … 10 V Current max. 1 mA Resolution 0.5 mV Linearity error 0.05 % (relating to 10 V) Temperature error 0.06 %/10 K (relating to 10 V)
6 VEGAMET 614
Product description
Relay outputs
Number 2 switching relays
1 fail safe relay Contact floating spdt Contact material AGSNO Turn-on voltage min. 10 mV DC
, hard gold plated
2
max. 250 V AC/DC Switching current min. 10 µA
max. 3 A AC, 1 A DC Breaking capacitance max. 750 VA,
18 W at U = 60 V DC; 40 W at U 40 V DC Min. switching hysteresis (Low/High delta) 0.5 %
DISBUS output
Function for linking the signal conditioning instruments
and for connecting digital indicating
instruments Connection cable 2-wire standard cable (screening is
recommended) max. cable length 1000 m
Indicating elements
Clear text indication LC display
- 4-line, 6 digits each
- background lighting
Analogue indication LED chain consisting of:
- 11 segments 0 % … 100 %
- shows the actual value of the measurement loop
LEDs in front plate green: operating voltage on
red: fault signal (LED lights with deenergised relay) yellow: relay status (standard setting: LED lights with energised relay)
Adjustment elements
Front plate 6 keys for configuration and parameter
adjustment
Upper housing side rotary switch for adjustment of the instrument
address on DISBUS (in mounted status, covered by plug-in socket)
Ambient conditions
Permissible ambient temperature -20°C … +60°C Storage and transport temperature -40°C … +80°C
Electrical connection
Screw terminal max. 1.5 mm
VEGAMET 614 7
2
Product description
Electrical protective measures
Instrument IP 30 Plug-in socket IP 20 Protection class II Overvoltage category II
Electrical separating measures
Reliable separation acc. to VDE 0106, part 1 between power supply, fail safe and level relay
and measuring data inputs
- reference voltage 250 V
- isolation resistance 2.3 kV Galvanic isolation between relay outputs
- reference voltage 250 V
- isolation resistance 1.4 kV Potential separation between DISBUS and outputs
- reference voltage 50 V
- isolation resistance 0.5 kV Common reference potential at voltage and current output
Mechanical data
Series module unit with plug-in socket, including
transparent cover, sensor terminal cover,
coded pin, two jumpers Mounting carrier rail mounting acc. to DIN 46 277, Bl. 3 Dimensions, unmounted W = 72 mm, H = 118.5 mm, D = 134 mm Weight approx. 480 g
CE conformity
VEGAMET 614 and VEGAMET 614 Ex signal conditioning instruments meet the protective regulations of EMC (89/336/EWG) and NSR (73/23/EWG). The instruments must only be used in industrial areas. Conformity has been judged acc. to the following standards: EMC Emission EN 50 081 - 2: 1993
Susceptibility EN 50 082 - 2: 1995
NSR EN 61 010 - 1: 1993
8 VEGAMET 614
Product description
Ex technical data
Power supply
Operating voltage U Reference voltage U
Meas. data input (intrinsically safe circuit)
Flame proofing II(1) G [EEx ia] IIC, [EEx ia] IIB Max. values
- voltage U
- current I
- power P Characteristics linear Effective inner inductance L Effective inner capacitance C
I
I
corresponds to non-Ex version
nom
= 250 V AC or 125 V DC
m
= 22.2 V
O
= 112 mA
O
= 620 mW
O
negligible negligible
EEx ia IIC EEx ia IIB
Max. permissible outer inductance L
Max. permissible outer capacitance C
O
O
2 mH 10 mH
160 nF 1100 nF
The intrinsically safe circuits are reliably separated from the non-intrinsically safe circuits up to a peak value of the nominal voltage of 375 V. The max. voltage on the non-intrinsically safe circuits should not exceed 253 V
in case of failure.
eff

1.5 Dimensions

Carrier rail 35 x 7.5 or 35 x 15 acc. to EN 50 022
Transparent cover
12
on
CONNECT
614
56 78
13 14 15 16
23 24 25 26 27 28
72
%
100
+
-
118,5
OKESC
Ser. No. 12345678
134
54,5
12 34
910 1112
17 18 20 21 22
VEGAMET 614 9

2 Mounting and installation instructions

Mounting and installation instructions
Mounting
Each series 600 signal conditioning instru­ment consists of a plug-in socket for carrier rail mounting DIN 46 277 and a module unit.
The supply voltage can be connected to terminals 17 and 18.
For neighbouring series 600 signal condition­ing instruments, it is possible to continue the connection L1 and N directly via the supplied jumpers. The same is valid for the connection of VEGAMET voltage output 0 … 10 V (termi­nals 15 and 16) and for DISBUS output (ter­minals 9 and 10).
Note!
The jumpers must never be used with single instruments or at the terminating end of an instrument row. If this is not observed, there is danger of coming in touch with the operat­ing voltage or causing a short-circuit.
VEGAMET 614 Ex signal conditioning instru­ment is intrinsically safe and must not be installed in hazardous areas.
Before setup, the Ex separating chamber must be plugged into VEGAMET 614 Ex as shown in the illustration.
Safe operation is ensured only if the operat­ing instructions manual and the EC type approval certificate are observed. VEGAMET 614 Ex must not be opened.
Coding
To avoid interchanging the various signal conditioning instruments, the plug-in socket is provided with pins and the signal condi­tioning instruments with corresponding gaps (mechanical coding).
Instrument coding ensures through differ­ently positioned coding pins that the various series 600 signal conditioning instruments are not interchanged.
An Ex coding with inserted coding pins en­sures that the non-Ex and Ex instruments are not interchanged.
On VEGAMET 614 Ex, the supplied coding pins (instrument coding pin and Ex coding pin) must be inserted
by the user according to the below table.
Instrument Ex
coding coding
VEGAMET 614 right 1 ––
VEGAMET 614 Ex right 1 left A
Separating chamber (must always by plugged in on the left with Ex instruments)
Ex coding
Jumpers
Direct supply for voltage output 0 … 10 V and DISBUS
Direct connection for supply voltage
left B
right B
left B
right B
VEGA
right
A o
VEGA
B o C o 1 o 2 o 3 o
-
4 o
+
0…10 V
5 o
DISBUS
6 o
7 o 8 o 9 o
N
10 o
L1
11 o 12 o
left
A o B o C o
1 o 2 o 3 o
o
0…10 V
o
DISBUS
o
7 o 8 o 9 o
N
o o
L1
12 o
-
+
10 VEGAMET 614
Mounting and installation instructions
Transparent cover
To protect the instrument against unauthor­ized adjustment, the front plate of VEGAMET can be equipped with two lockable transpar­ent covers after setup. To remove the trans­parent cover, see illustration.
VEGAMET 614 11
17 18 19 20 21 22
910 1112
12 34
23 24 25 26 27 28
13 14 15 16
56 78
L1
N
+- +-
+- +-
+-

3 Electrical connection

Electrical connection

3.1 Connection instructions

Note the following instructions for electrical connection:
- The connection must be made acc. to the local installation standards (e.g. in Ger­many acc. to the VDE-regulations).
- The wiring between VEGAMET 614 and sensor can be made with standard two­wire cable.
- If strong electromagnetic interferences are expected, we recommend the use of screened cable. The screening must be earthed on both ends.
- The line resistances stated in the technical data must not be exceeded.
- If overvoltages are expected, we recom­mend a sensor electronics with integrated overvoltage protection or the installation of VEGA overvoltage arresters.
3.2 Connection instructions for
Ex approved applications
When wiring between VEGAMET 614 Ex and an explosion protected sensor, the installation rules, in Germany acc. to VDE 165 and the EC type approval certificate, must be ob­served. The official documents are supplied with the respective instrument.

3.3 Wiring plan

Sensor
DISBUS output
Power supply
Fail safe relay
The connections "9, 10, 15, 16, 17 and 18“ can be connected with the supplied jumpers to neighbouring series 600 signal conditioning instruments.
Level relay 1
Current output
Voltage output
Level relay 2
Ex version
Before connecting Ex certified instruments, please note the instructions in the attached official documents as well as the valid installa­tion regulations. Make sure that the Ex sepa­rating chamber is plugged in on the left instrument side before setup (see illustration Coding).
12 VEGAMET 614
Adjustment

4 Adjustment

4.1 Indicating and adjustment elements

A rotary switch is located on the upper side of the housing. It is used to set the instrument address on DISBUS. This setting is neces­sary if several VEGAMET signal conditioning instruments are linked via the DISBUS or if you control a VEGADIS 174 indicating instru­ment via the DISBUS.
The adjustment range 1 … F corresponds to the DISBUS addresses 1 … 15. Factory setting: 1
Note:
If address 0 is set, the VEGAMET does not participate in the DISBUS communication.
Rotary switch
8
9
7
6
5
4
33
A
B
C
D
E
2
F
1
0
10 11 12
13 14 15
Position of the rotary switch
Rotary switch
Front plate
%
100
+
4 5 6
7
CONNECT
12
on
614
-
OKESC
Ser.No. 12345678
1 2
3
The set instrument address on the DISBUS can be indicated via the LC display.
Note
If several VEGAMET are linked via the DISBUS, each address must be assigned only once!
VEGAMET 614 13
1 LCD (4 lines with 6 figures each, illuminated) for clear text
indication
2 LED chain (yellow) for quasianalogue indication of the
measured value 3 Keys for menu adjustment 4 Connection socket for VEGACONNECT 5 LED (yellow) lights if the relay is energised (standard
setting) 6 LED (red), lights if fail safe relay is deenergised 7 LED (green), lights if operating voltage is on

4.2 Adjustment system

Indicating and adjustment module
Branching point, i.e. jump to the next lower menu with [OK]
Adjustment
Indication of
- measured value
- menu item
- parameter
%
100
+
-
Depending on the menu item, change value or choose out of a list
Choose menu item
Depending on the menu item, interrupt adjustment or change to the next higher menu
Reduced menu - extended menu
Two menu systems are available: the "re­duced menu“ and the "extended menu“.
Note:
VEGAMET 614 is preset to "reduced menu“ mode!
In the majority of applications, you can carry out the necessary settings in this menu mode (see also chapter 4.5 Comparison of the menu modes).
Adjustment structure
The adjustment is made via 6 keys in con­junction with the text display. The jump from the measured value display to the next lower menu item is made with [OK]. Use [–>] or [<–] to change within this menu level from one menu item to the other. A branching point is indicated by the symbol and enables with [OK] a jump to the next lower menu item. Parameters are indicated by the missing symbol . The value of the parameters can be modified with [+] or [–] or chosen out of a list. The modified value can be saved with [OK]. To interrupt a setting (without saving the modification), you have to push [ESC]. A jump up to the next higher menu item is made with [ESC]. 15 minutes after pushing a key
Depending on the menu item, save the set value or change to the lower menu
OKESC
for the last time, an automatic reset to the measured value display is triggered. A jump up to the next higher menu level with [ESC] is possible from each menu item, even if no connection is shown in the menu schematics.
Add’l func­tion
ESC ESC
Pass­word
off
ESC
Langu­age
Eng­lish
TAG 1 to de fault
ESC
Menu mode
Redu­ced
If you are in the utmost right menu item of a menu row and push the [–>] key, you will immediately reach the utmost left menu item of this row. You also reach immediately the right menu item from the left one, if you push the [<–] key. The connections (shown with a broken line) are not shown in the menu schematics in order to provide a clearer overview.
Volt output 1
prop. to
Per­cent
Unit
0,0%
Volt output
0/10V
Failure mode
0V
Volt limi­tation
on
14 VEGAMET 614
Adjustment
Formatting examples the menu schematic
White letters indicate parameters
Langu­age
which can be modified with the [+] or
Eng-
[–] key and saved with the [OK] key.
lish
Examples:
- In the menu item "Language“ you can change from German to English.
Param.
Bolt print/italic inputs, e.g. the meas-
TAG -
urement loop name "TAG-No. 1" can
No. 1
show a different text if you have
carried out a parameter adjustment or configuration of the measurement loop. In the menu schematics, you will see the factory setting.
Volt
Light grey menu items are only dis-
at
played if necessary (depending on
0%
0,000
the adjustments in the other menu items).
Changing the menu mode:
Move from the display of measured value (TAG-No.1) to the menu item "Menu mode reduced“. Now push [+] twice until "Menu mode Extended“ appears, then push [OK]. You can now return to the measured value indication by pushing [ESC] twice.

4.3 Adjustment via PC

Setup
- directly via the keys of the adjustment module
- or with a PC, equipped with VVO software (VEGA Visual Operating) and an interface adapter VEGACONNECT.
%
100
+
-
CONNECT
12
on
614
VEGACONNECT
OKESC
Ser.No. 12345678
TAG­No. 1 % xx,x
Add’l
Confi-
Param.
TAG­No. 1
gura­tion
func­tions
Pass­word
off
Langu­age
Eng­lish
TAG 1 to de fault
Reset TAG 1
Menu mode
Redu­ced
VEGAMET 614 15
Adjustment

4.4 Configuration and parameter adjustment

No matter whether you set up your VEGAMET via the keys of the adjustment module or via a PC with the software VVO, the procedure is always the same.
Proceed in the following sequence:
- configuration (if not already configured with
a factory setting)
- parameter adjustment.
In this operating instructions manual, the adjustment steps are described which are directly carried out on the keys of VEGAMET. The adjustment via the software VVO is de­scribed in a separate manual.
Configuration
Configuration means initial assigning or set­ting functions. VEGAMET requires a (mostly once) basic co-ordination that determines the applications and the assignments of the inputs and outputs. Choices can be made from existing functions and options. This procedure is called configuration. The signal conditioning instruments are delivered with a configuration which only needs to be changed in special cases.
The basic configuration includes the following steps:
1 Configuration measurement loop
• Choose the kind of application (level,
gauge…)
• Choose sensor type (capacitive, hydro-
static…)
• Determine application (standard, level
difference…)
• Determine options (no option, correc-
tions…)
Note:
The measurement loop configuration can only be modified, if a "Reset to single measure­ment“ has been carried out previously (under "Additional functions“ in the extended menu). If you only want to modify the sensor type, you first have to carry out "Reset TAG 1“ (under "Additional functions“ in the reduced menu).
2 Configuration of inputs
• Determine from where your VEGAMET receives the input data (sensor, other VEGAMET)
• Enter sensor characteristics values (meas. range, current range)
3 Configuration of outputs
• All outputs (except fail safe relay) can been assigned to one sensor or can be switched off.
After basic configuration, VEGAMET goes into operating mode and indicates the actual measured value. The other configurations should be carried out after the parameter adjustment.
Parameter adjustment
Parameter adjustment means changing the values. Signal conditioning instruments have many parameters, the values of which can be modified, such as for example, the integration time of 0 … 600 s. The modification of these values is called parameter adjustment. The parameter adjustment does not influence the configuration. Take note that parameter ad­justment is only possible after a configuration has been carried out (e.g. set values of the current input after the current output has been assigned).
16 VEGAMET 614
Adjustment
4.5 Comparison: reduced menu – extended menu
In the table the most important functions of VEGAMET 614 are listed – each with a mark indi­cating in which menu mode the functions are available.
Function Reduc. Ext.
Adjustment with medium x x
Adjustment without medium x x
Set integration time x x
Simulation xx
Switch over current output (4/20 mA, 0/20 mA, 20/4 mA, 20/0 mA) x x
Switch over voltage output (2/10 V, 0/10 V, 10/2 V, 10/0 V) x x
Switch over relay mode (overfill protection – dry run protection) x x
Scaling (display) x x
Change menu mode (reduced – extended) x x
Change menu language x x
Activate/deactivate password x x
Change sensor type x x
Change measurement loop name x x
Reset measurement loop (Reset TAG) x x
Reset measurement loop to default (TAG to default) x x
Edit linearisation curves x
Change mode (e.g. level – gauge) x
Select different reference values for outputs x
Define output behaviour (current/voltage output) in case of failure x
Adapt VEGAMET to sensor characteristics values x
Offset correction during adjustment x
Adjust relay switching delay x
Reset level –x
Assign inputs (from another VEGAMET) x
Switch on/off current/Volt limitation x
Individual setting of current/voltage output values betw. 0 … 20 mA/0 … 10 V x
Manual offset correction x
Manual real value correction x
Info indication x
menu menu
VEGAMET 614 17
Loading...
+ 39 hidden pages