10 Menu schematic ....................................................................... 58
Safety i nformation ...................................................................... 64
Note Ex a rea .............................................................................. 64
VEGAMET 515V3
Page 4
1 Product description
Product description
1.1 Function and configuration
VEGAMET 515V is a signal conditioning instrument for a number of applications, e.g.:
- level measurement
- gauge measurement
- gauge difference measurement
- process pressure measurement
- density measurement
- measurement with automatic correction
-etc.
Software configuration
Input components
- continuous current
inputs
- contact inputs
Function
The signal conditioning instruments power
the connected sensors and process their
measuring data. The conditioning is carried
out by a special software consisting of functional components (FB), input components
(EB), output components (AB) as well as
DCS components (PB).
Configuration
The VEGAMET 515V signal conditioning
instrument is constructed as a module card
in European size with 5 TE (= 25.4 mm)
width. The front plate contains an adjustment
module with LC display and keys. The front
plate of VEGAMET 515VN does not contain
an adjustment module.
Output components
- current outputs
- voltage outputs
- DISBUS outputs
- failure output
- MET display
- relay outputs
AB
EBFB
PC/DCS components
- PC/DCS outputs
Functional components
- measurement loop
TAG 1
- measurement loop
TAG 2
- measurement loop
TAG 3
4VEGAMET 515V
PB
Page 5
Product description
1.2 Approvals
VEGAMET 515V is suitable for measurements
in Ex areas in conjunction with the following
safety barriers:
- Safety barrier type 146
Conformity certificate
PTB no. Ex-93.C.4025
- Safety barrier type 546
Conformity certificate
PTB no. Ex-97.D.2098 X
EC type approval certificate
TÜV 97 ATEX 1186 X
- Safety barrier VEGATRENN 547 V Ex
Conformity certificate
PTB no. Ex-97.D.2098 X
- Safety barrier VEGATRENN 548 V Ex
Conformity certificate
PTB no. Ex-96.D.2137
For such applications, please note the appropriate official documents (general type approval, test and conformity certificates). They
come with the respective instrument.
1.3 Features
- microcomputer-controlled signal conditioning instrument for continuous measurement
- adjustment module with LC display and 6
keys (not with VEGAMET 515VN)
- adjustable integration time
- two fixed and three user-programmable
linearisation curves
- fault monitoring
- fault signal and failure diagnosis via display
Inputs:
- 2 sensor inputs for digital transmission of
measured data (VBUS) (pressure transmitters, ultrasonic and radar sensors)
- 2 correction inputs (for key, switch or level
switch)
Outputs:
- 3 current output 0/4 … 20 mA
- 3 voltage outputs 0/2 … 10 V
- 2 relay outputs (spdt)
- 1 DISBUS output for digital wiring and for
connection of VEGADIS 174
- 1 fail safe relay
- adjustment also via PC with adjustment
software VVO; connection via the interface
adapter of the VEGACONNECT series
VEGAMET 515V5
Page 6
Product description
1.4 Technical data
Power supply
Operating voltageUnom = 20 … 53 V AC, 50/60 Hz
Power consumptionapprox. 20 VA or approx. 15 W
Meas. data input
Number2 inputs
Input typeactive two-wire input, digital (VBUS) (sensor is
Possible sensors (loop powered)pressure transmitters, ultrasonic sensors,
Voltageapprox. 25 V DC
Current limitationat approx. 7 W, short-circuit proof
Min. adjustment delta2 % of sensor range (hydrostatic)
Connection cable2-wire standard cable (screening recommen-
max. cable length1000 m
max. resistance per wire20 Ω
Correction signal input
Number2 inputs
Functionswitching signal to trigger corrections
Voltage5 V (from instrument)
Current5 mA
External cable resistance≤150 Ω
= 20 … 72 V DC
powered by VEGAMET)
radar sensors
5 mm (ultrasonic, radar)
ded)
or special relay modes via
external switching contact
Current output
Number3 outputs
Functionanalogue output of the processing results
Rangeindividually adjustable in the range of 0 … 20 mA
Loadmax. 500 Ω
Resolution1 µA
Linearity error0.05 % (relating to 20 mA)
Temperature error0.05 %/10 K (relating to 20 mA)
Voltage output
Number3 outputs
Functionanalogue output of the processing results
Rangeindividually adjustable in the range of 0 … 10 V
Currentmax. 1 mA
Resolution0.5 mV
Linearity error0.05 % (relating to 10 V)
Temperature error0.06 %/10 K (relating to 10 V)
6VEGAMET 515V
Page 7
Product description
Relay outputs
Number2 operating relays
1 fail safe relay
Contactfloating spdt
Contact materialAgNi, hard gold plated
Turn-on voltagemin. 10 mV DC
max. 250 V AC/DC
Switching currentmin. 10 µA
max. 3 A AC, 1 A DC
Breaking capacitancemax. 500 VA, 54 W
Min. switching hysteresis
(Low-/High-Delta)0.5 %
DISBUS output
Functionfor linking signal conditioning instruments with
another and for connecting digital indicating
instruments
Connection cable2-wire standard cable (screening is
recommended)
max. cable length1000 m
max. resistance per wire15 Ω
Indicating elements
Clear text indicationLC display
(not available with 515VN)- 4-line, 6 digits each
- background lighting
Analogue indicationLED chain consisting of:
(not available with 515VN)- 11 segments 0 % … 100 %
- shows the actual value of the measurement
loop
LEDs in front plategreen: operating voltage on
red: fault signal (LED lights with deenergised
relay)
yellow: relay status (standard setting:
LED lights with energised relay)
Adjustment elements
Front plate6 keys for configuration and parameter
adjustment (not available with 515VN)
Circuit boardrotary switch for adjustment of the instrument
address on DISBUS
Ambient conditions
Permissible ambient temperature-20°C … +60°C
Storage and transport temperature-40°C … +80°C
Electrical connection
Multiple plug componentacc. to DIN 41 612, series F, 33-pole, 3 rows
d, b, z (partly equipped)
Carrier BGT 596 (Ex)connection to the appropriate module
Housing type 505 or type 506connection to screw terminals (max. 2.5 mm2)
VEGAMET 515V7
Page 8
Product description
Electrical protective measures
Protection
- unassembledIP 00
- mounted into
carrier BGT 596
- front side completely equippedIP 40
- upper and lower sideIP 20
- wiring sideIP 00
- mounted into
housing type 505 or type 506
- terminal sideIP 20
- housing generalIP 30
Protection classII (in housing type 505)
Overvoltage categoryII
Electrical separating measures
Reliable separation acc. to
VDE 0106, part 1 betweenpower supply, fail safe and level relay
and measuring data inputs
- reference voltage250 V
- isolation resistance2.3 kV
Galvanic isolationbetween the relay outputs
- reference voltage250 V
- isolation resistance1.4 kV
Potential separationbetween DISBUS and outputs
- reference voltage50 V
- isolation resistance0.5 kV
Common reference potential atvoltage and current output,
correction signal output
Mechanical data
Seriesmodule unit for carrier BGT 596
or housing type 505
DimensionsB = 25.4 mm (5 TE), H = 128.4 mm, T = 162 mm
CE conformity
VEGAMET 515V signal conditioning instrument meets the protective regulations of EMC
(89/336/EWG) and NSR (73/23/EWG). Conformity has been judged acc. to the following
standards:
EMCEmissionEN 50 081 - 2: 1993
SusceptibilityEN 50 082 - 2: 1995
NSREN 61 010 - 1: 1993
8VEGAMET 515V
Page 9
Product description
1.5 Dimensions
Circuit board 100 x 160 x 1.5
European size
Multiple plug
5 TE
15
5,5
162
100
128,4
25,4
VEGAMET 515V9
Page 10
2 Mounting
Mounting
VEGAMET 515V signal conditioning instrument can be either mounted with a socket
module in a carrier BGT 596 or BGT
596 Ex.M or in a single housing type 505 or
506 .
Mounting in carrier
Mount the respective socket module (standard or Ex version) in your carrier. Wire the
connection of the multipoint connector according to the wiring plan.
Module
Multipoint connector DIN 41 612, series F, 33pole (d, b, z) with coded pins and mounting
material for mounting in carrier BGT 596 or
BGT 596 Ex.M.
The multipoint connector is available in the
following versions:
- Wire-Wrap standard connection
1.0 x 1.0 mm
- Plug connection 2.8 x 0.8 mm
- Termi-Point standard connection
1.6 x 0.8 mm
- Soldering connection
- Screw terminals 2 x 0.5 mm
2
For further mounting information, see the
operating instructions of the carrier.
Mounting in single housing
Either screw the housing socket directly to
the mounting plate or plug onto a carrier rail
(35 x 15 acc. to EN 50 022 or 32 acc. to
EN 50 035). Connect the terminals according
to the wiring plans on the following pages.
For further mounting information, see the
operating instructions of the housing.
Transparent cover
To protect the instrument against unauthorised adjustment, the front plate of VEGAMET
can be provided with a lockable transparent
cover (supplied with the instrument).
Coding
To prevent an interchanging of the different
signal conditioning instruments, the multipoint
connector of the carrier or the housing can
be equipped with coded pins.
The multiple plug of the signal conditioning
instrument is equipped with corresponding
holes (mechanical coding). An instrument
coding ensures that the different signal conditioning instruments are not interchanged.
The appropriate coded pins are supplied
with each module or housing. Equip the
multipoint connector with these coded pins
acc. to the following table.
Instrument coding
VEGAMET 515V(N)a7 / c5
z b d
a c
o 1 o
o 3 o
o 5 o
o 7 o
a7
o 9 o
o11o
o13o
o15o
o17o
o19o
o21o
o23o
o25o
o27o
o29o
o31o
c5
10VEGAMET 515V
Page 11
Electrical connection
3 Electrical connection
3.1 Connection instructions
Note the following instructions for electrical
connection:
- The connection must be made acc. to the
local installation standards (e.g. in Germany acc. to the VDE-regulations).
- The wiring between VEGAMET and sensor
can be done with standard two-wire cable.
- If strong electromagnetic interference is
expected, screened cable is recommended. The screening must be earthed
on both ends.
- The line resistances stated in the technical
data must not be exceeded.
- If overvoltages are expected, we recommend a sensor electronics with integrated
overvoltage protection or the installation of
VEGA overvoltage arresters.
- The voltage supply of VEGAMET must
provide extra-low voltage to ensure protection class II. When using VEGASTAB
593, a reliable separation from the mains
circuits acc. to DIN VDE 0106, part 101 is
ensured.
3.2 Wiring plan
d b z
+–
2
6
10
12
16
18
20
22
24
28
30
32
2
1
+
++
–
–
+
+
+
–
–+–
+
–
3
–
+
+
+
–
Fail safe relay
Level relay 1
Level relay 2
DISBUS output
+
–
Sensor 2
Sensor 1
L (+)
N (–)
+
–
Supply
voltage
Current outputs 1 … 3
Voltage outputs 1 … 3
Correction signal
Input 4
Correction signal
Input 5
Note:
Terminals d30 and z30 or d32 and z32 are
switched in parallel in the signal conditioning
instrument.
VEGAMET 515V11
Page 12
4 Adjustment
Adjustment
4.1 Indicating and adjustment
elements
A rotary switch is located on the circuit
board. It is used for adjustment of the instrument address on DISBUS. This setting is
necessary if measured values are to be
transmitted via the DISBUS cable to external
instruments, e.g. VEGADIS 174 indicating
instrument, VEGACOM 557 interface converter or other VEGAMETs. With these external instruments, a received measured value
can be associated with a specific measurement loop by means of the address.
The adjustment range 1 … F corresponds to
the DISBUS addresses 1 … 15.
Factory setting: 1
Note:
With address 0, the VEGAMET does not
participate in the DISBUS communication.
Rotary switch
8
9
7
6
5
4
33
The set instrument address on the DISBUS
can be indicated via the LC display.
Note
If several VEGAMET are linked via the
DISBUS, each address must be assigned
only once!
A
B
C
D
2
E
F
1
0
10
11
12
13
14
15
Position of the rotary switch
Rotary switch
Front plate VEGAMET 515V
without adjustmentwith adjustment
module 515VNmodule 515V
1
2
%
%
100
100
+
+
-
-
33
OKESC
CONNECT
12
on
515V
1 LCD (4 lines with 6 figures each, illuminated) for clear text
indication
2 LED chain (yellow) for quasianalogue indication of the
measured value
3 Keys for menu adjustment
4 Connection socket for VEGACONNECT
5 LED (yellow) lights if the relay is energised (standard
setting)
6 LED (red), lights if fail safe relay is deenergised
7 LED (green), lights if operating voltage is on
4
5
6
7
OKESC
CONNECT
CONNECT
12
12
on
on
514
515V
12VEGAMET 515V
Page 13
Adjustment
4.2 Adjustment system
Indicating and adjustment module
Branching point, i.e. jump to the next
lower menu with [OK]
Indication of
- measured value
- menu item
- parameter
Depending on the menu item, interrupt
adjustment or change to the next higher
menu
Adjustment structure
The adjustment is made via 6 keys in conjunction with the text display. The jump from
the measured value display to the lower
menu item is made with [OK]. Use [–>] or
[<–] to change within this menu level from
one menu item to the other. A branching point
is indicated by the symbol ▼ and enables
with [OK] a jump to the next lower menu item.
Parameters are indicated when the symbol ▼
is missing. The value of the parameters can
be modified with [+] or [–] or chosen out of a
list. The modified value can be saved with
[OK]. To interrupt a setting (without saving
the modification), you have to push [ESC]. A
jump to the next higher menu item is made
with [ESC]. 15 minutes after pushing the key
for the last time, an automatic reset to the
measured value display is triggered. A jump
to the next higher menu level with [ESC] is
possible from each menu item, even if no
connection is shown in the menu schematics.
Depending on the menu item, change
value or choose out of a list
Choose menu item
Depending on the menu item, save the
set value or change to the lower menu
ESC
ESC
Info
ESC
Language
English
ESC
Reset
VEGAMET
Service
-
100
Add’l
functions
ESC
Password
on
%
+
OKESC
Edit
lin.
curve
If you are in the utmost right menu item of a
menu row and push the [–>] key, you will
immediately reach the utmost left menu item
of this row. You also reach immediately the
utmost right menu item from the left one, if
you push the [<–] key.
The connections (shown with a broken line)
are not shown in the menu schematics in
order to provide a better overview.
Volt
output 1
ESC
prop.
to
Percent
Unit
0,0%
Voltoutput
0/10V
Failure
mode
0V
Volt
limitation
on
VEGAMET 515V13
Page 14
Adjustment
Illustrations in the menu schematic
Lang-
White letters represent the param-
uage
eters which can be modified with the
English
[+] or [–] key and saved with the
Examples:
- In the menu item "Language“ you can
Param.
TAGNo. 1
configuration of the measurement loop. In the
menu schematics you see the factory setting.
Volt
at
0%
0,000
[OK] key.
change from German to English.
Bolt print/italic entries, e.g. the measurement loop name "TAG-No. 1" can
differ if you have made a corre-
▼
sponding parameter adjustment or
Light grey menu items are only displayed when necessary (depending
on the adjustments in the other menu
items).
4.3 Adjustment via PC
Signal conditioning instrument with adjustment module
Setup
- directly via the keys of the adjustment
module
- or with a PC, equipped with VVO software
(VEGA Visual Operating) and the interface
adapter VEGACONNECT.
Signal conditioning instrument without
adjustment module
Setup
- with a PC, equipped with VVO software
(VEGA Visual Operating) and the interface
adapter VEGACONNECT.
%
100
+
-
OKESC
CONNECT
12
on
515
VEGACONNECT
PC with VVO software
14VEGAMET 515V
Page 15
Adjustment
4.4 Configuration and parameter adjustment
No matter whether you set up your
VEGAMET via the keys of the adjustment
module or via a PC with the software VVO,
the procedure is always the same.
Proceed in the following sequence:
- configuration (if not already configured as
factory setting)
- parameter adjustment.
In this operating instructions manual, the
adjustment steps are described which are
carried out directly on the keys of VEGAMET.
The adjustment via the software VVO is described in a separate manual.
Configuration
Configuring means assigning or setting functions (once). VEGAMET requires a basic
configuration (usually once) that determines
the applications and the assignments of the
inputs and outputs. Choices can be made
from existing functions and options. This
procedure is called configuration. The signal
conditioning instruments are delivered with a
configuration which only needs to be
changed in special cases.
The basic configuration includes the following
steps:
1 Configuration measurement loop
• Choose the kind of application (level,
gauge…)
• Choose sensor type (capacitive, hydro-
static…)
• Determine application (standard, level
difference…)
• Determine options (no option, correc-
tions…)
2 Configuration inputs
• Specify from where your VEGAMET receives the input data (sensor, other
VEGAMET)
• Enter sensor characteristics values (meas.
range, current range)
3 Configuration outputs
• All outputs (except fail safe relay) can be
assigned to one sensor or can be switched
off.
After basic configuration, VEGAMET goes
into operating mode and indicates the actual
measured value. The other configurations
should be carried out after the parameter
adjustment.
Parameter adjustment
Parameter adjustment means changing the
values. Signal conditioning instruments have
many parameters, the values of which can be
modified, such as, for example, the integration time of 0 … 600 s. The modification of
these values is called parameter adjustment.
The parameter adjustment does not influence
the configuration. Take notice that parameter
adjustment is only possible after a configuration has been carried out (e.g. set values of
the current input after the current output has
been assigned).
Note:
The measurement loop configuration can only
be modified if a "Reset to single measurement“ or "Reset to linked measurement“ has
been carried out before (under "Additional
functions“).
VEGAMET 515V15
Page 16
5 Setup
• Adjust the VEGAMET instrument address
with the rotary switch on the circuit board
(if measured values are to be transferred
via the DISBUS cable to external instruments, e.g. to VEGADIS 174 indicating
instrument, VEGACOM 557 interface converter or other VEGAMETs).
• Install VEGAMET.
• Connect sensor and power supply.
• After switching on, the display indicates for
approx. 10 secs the instrument type and
the software version, e.g. "MET 515V
V1.12“.
• If no measurement loop has yet been
configured, the display shows "Configuration“, the red fault indicator LED also lights.
In this case, proceed as described under
"6.1 Configuration measurement loop“.
• If a measurement loop has already been
configured, the display shows a value, e.g.
"TAG-No. 1 21.8 %“.You can now carry out
the settings, e.g. adjustment settings.
Setup
16VEGAMET 515V
Page 17
Configuration
6 Configuration
6.1 Configuration measurement loop, two independent single
measurements
TAG No. 1
%
xx,x
Param.
Confi-
TAG-
gura-
No. 1
tion
Config
inputs
Config
meas.
loop
TAG 1
Level
1)
Application
Level
Gauge
Distance
TAG 2
Level
like TAG1
1)
Sensor
type
Hydrostatic
Pulse-echo
Pulse-radar
Mode
Standard
1)
Option
no
option
Offset cor.
Real value
cor.
Density
cor.
1)
The measurement loop configuration is usually provided as a factory setting. Should a
general modification of the measurement loop
configuration be necessary, "Reset to single
measurement“ or "Reset to linked application“
should be carried out first (with "Reset to
single measurement“ the VEGAMET is prepared for configuration of two independent
single measurements. With "Reset to linked
application“ it is prepared for configuration of
linked applications). Only then can setting in
the menu items Application, Sensor type,
Mode and Option be modified. Proceed in
the menu items in the given sequence. The
adjustment or selection options always depend on what was set in the previous menu
item. When e.g. in the menu item Application
"Level“ is set, only one sensor type suitable
for level detection can be selected in the
menu item "Sensor type“.
1)
The settings in these menu items can only be
modified if a "Reset to single measurement“ was
carried out beforehand (under menu item "Additional functions“).
Sensor
coordination
Position A
Input
no.1
Input 2
TAGname
TAG No. 1
Position D
Input
no.4
Input5
Fault
signal
on
off
PositionE
Input
no.4
Input5
Tar e
Input
from
undefined
Input 4
Fault
signal
on
Input
from
undefined
Input 4
As soon as you have made an adjustment in
the menu item "Option“ and confirmed it, the
display shows the message: "TAG being
created!“.
In the following illustrations, different types of
applications are shown. In addition, you will
see which adjustments you have to carry out
in the menu items "Sensor type“ and "Mode“.
The menu item "Position D“ only appears
when in the menu item "Option“ either "Offset
correction“ or "Real value correction“ has
been set. The menu item "Position E“ only
appears when in the menu item "Option“
"Density correction“ has been set.
VEGAMET 515V17
Page 18
Configuration
In the menu items "Pos. D“ and "Pos. E“ you
can determine if "Correction input 4“ or "Correction input 5“ is to be used for one of the
stated corrections. You then have to enter
under "Parameter adjustment“ – "Special
functions“ in the menu items "Real value correction at…“ or "Density correction at…“ a
value at which the correction should be carried out.
Note!
For tare and monitoring it is also possible to
assign input 4. However, a correction input
must not be used for tare, monitoring and
e.g. density correction at the same time!
In the menu item "TAG-Name“ you can enter
an individual name for the measurement loop.
If you choose "Fault signal off“, the fail safe
relay and the fault indicator LED in the front
plate do not function. For a description of the
correction modes, see the following pages.
(note the explanations under "6.1 Configuration measurement loop, two independent single
measurements“)
Level measurement in pressurised vessels
Pos. B
Pos. A
Level measurement in pressurised
vessels
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Configuration
Config
inputs
Config
meas.
loop
Application
Level
1)1)1)
1)
The adjustments in these menu items can only be
modified if a "Reset to linked measurement“ was
carried out beforehand (under menu item "Additional functions“).
Sensor
type
Hydrostatic
Mode
Pressurized
1)
Option
no
option
Sensor
coordination
Position A
Input
no.1
TAG 1
Level
TAGname
TAG No. 1
Position B
Input
no.2
TAG 2
Gauge
pressure
Fault
signal
on
off
Position D
Input
no.4
like TAG 1
Tar e
Position E
Input
no.5
TAG 3
Total
pressure
Input
from
undefined
Monitoring
Input
from
undefined
20VEGAMET 515V
Page 21
Configuration
Product with variable density
Pos. C
Pos. A
Level measurement, product with
variable density (density compensation)
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Configuration
Config
inputs
1)
The adjustments in these menu items can only be
Config
meas.
loop
Application
Level
1)
Sensor
type
Hydrostatic
1)
Mode
variable
density
1)
modified if a "Reset to linked measurement“ was
carried out beforehand (under menu item "Additional functions“).
Option
no
option
1)
Offset cor.
Real value cor.
Density cor.
Sensor
coordination
Position A
Input
no.1
TAG 1
Level
TAGname
TAG No. 1
Position C
Input
no.2
TAG 2
Density
Fault
signal
on
off
Position D
Input
no.4
like TAG 1
Tar e
Position E
Input
no.5
TAG 3
incor.
level
Input
from
undefined
Monitoring
Input
from
undefined
VEGAMET 515V21
Page 22
Density measurement
Pos. C
Pos. A
Density measurement
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Configuration
Config
inputs
Config
meas.
loop
Configuration
Application
Density
1)
1)
The adjustments in these menu items can only be
modified if a "Reset to linked measurement“ was
carried out beforehand (under menu item "Additional functions“).
Sensor
type
Hydrostatic
1)
Mode
Standard
1)
1)
Option
no
option
Offset cor.
Sensor
coordination
Position A
Input
no.1
TAG 1
Density
TAGname
TAG No. 1
Position C
Input
no.2
TAG 2
Level
Fault
signal
on
off
Position D
undefined
like TAG 1
Tar e
Position E
undefined
TAG 3
incor.
level
Input
from
undefined
Monitoring
Input
from
undefined
22VEGAMET 515V
Page 23
Configuration
Gauge difference
Pos. A Pos. BPos. A Pos. B
Gauge difference
TAG No. 1
%
xx,x
Confi-
Param.
gura-
TAG-
tion
No. 1
Config
inputs
1)
The adjustments in these menu items can only be
Config
meas.
loop
Application
Gauge
1)
1)
Sensor
type
Hydrostatic
Pulse-echo
Pulse-radar
Mode
Difference
1)
modified if a "Reset to linked measurement“ was
carried out beforehand (under menu item "Additional functions“).
1)
Option
no
option
Real value
cor.
Sensor
coordination
Position A
Input
no.1
TAG 1
Upstream
water
TAGname
TAG No. 1
Position B
Input
no.2
TAG 2
Downstream
water
Fault
signal
on
off
Position D
Input
no.4
like TAG 1
TAG 3
Gauge
diff.
Tar e
Input
from
undefined
Position E
Input
no.5
Monitoring
Input
from
undefined
VEGAMET 515V23
Page 24
Level difference
Pos. A Pos. BPos. A Pos. B
Level measurement, level difference
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Configuration
Config
inputs
Config
meas.
loop
Configuration
Application
Level
1)
The adjustments in these menu items can only be
modified if a "Reset to linked measurement“ was
carried out beforehand (under menu item "Additional functions“).
1)
Sensor
type
Hydrostatic
Pulse-echo
Pulse-radar
Mode
Difference
1)1)
1)
Option
no
option
Real value
cor.
Sensor
coordination
Position A
Input
no.1
TAG 1
Level1
TAGname
TAG No. 1
Position B
Input
no.2
TAG 2
Level2
Fault
signal
on
off
Position D
Input
no.4
like TAG 1
Tar e
Position E
Input
no.5
TAG 3
Level
diff.
Input
from
undefined
Monitoring
Input
from
undefined
24VEGAMET 515V
Page 25
Configuration
6.3 Measurement loop name
Naming the measurement loop
You can modify the measurement loop name
"TAG-No. 1“, if you want a more meaningful
designation.
Two lines with six figures each are available.
To do this, move to the menu item "TAG name
…“.
TAGName
TAG -
No. 1
pushing the [+] or [–] key, by doing this you
scroll through the alphabet, through the row
of numbers and through a list of characters.
[–] to the requested character. With [–>] you
move to the next character.
Push [+] or [–] once. The first character of the measurement loop
name, in this example the letter T,
flashes (editing mode). Continue
TAG-
When you have reached the re-
Name
quested character, push [–>]. The
TAG No. 1
next character is activated and begins to flash. Scroll again with [+] and
6.4 Fault signal
If you choose "Fault signal off“ for a certain
measurement loop, the fail safe relay and the
fault indicator LED in the front plate will not
react when a failure occurs on this measurement loop, e.g. in case of short-circuit. Failures on other measurement loops, however,
will be signalled.
Only if on all measurement loops "Fault signal
off“ is selected will the fail safe relay and the
fault indicator LED be completely out of operation.
TAG-
When you have finished the last posi-
Name
tion of the name and "written" the
Ves.4
Hall3
name of the measurement loop in this
way, you have to push [OK] to save
the setting. If you quit the menu item without
saving, the name of the measurement loop
will be still "TAG-No. 1“.
Ves.4
In the display of measured value and
Hall3
all other menu items in which the
%
27,8
measurement loop name is stated,
you will now find the new designation.
VEGAMET 515V25
Page 26
Configuration
6.5 T a re
With the function "Tare“, a measured value
beginning at 0 % is created from the actual
measured value. The tare function can be
started via a key connected to correction
signal input 4.
Meas. values in %
100%
Taster
40
0
tared measured values
The following steps are necessary to activate
the tare function:
• Connect a key (closer) to input 4.
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Configuration
Config
inputs
Config
meas.
loop
tared course
actual course
40100%
actual measured values
• Choose in the menu item "Input from“ (un-
der the menu item "Tare“) input 4.
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Signal
conditioning
Outputs
Curr.
outputs
Volt
outputs
Volt
output
1
prop.
to
Percent
VolumeTare
Unit
Tare
%
V%
Adjustment
• Assign the tared value in the menu item
"Unit“ to the requested output (e.g. current,
volt, DIS, PC/DCS output, MET indication).
If you want to deactivate the tare function,
you first have to cancel taring in the menu
item "Unit“. Then you have to cancel the allocation of the correction input (input 4) under
the menu item "Tare“.
TAG 1
Level
Application
Level
Tar e
Input
from
Input
no.4
26VEGAMET 515V
Page 27
Configuration
6.6 Monitoring
To activate the monitoring, the following
prerequirements must be met:
TAG No. 1
%
xx,x
Param.
Confi-
TAG-
gura-
No. 1
tion
Config
inputs
• Choose in the menu item "Input from “ (un-
der the menu item "Monitoring“) input 4.
• Connect a key switch (opener) to input 4.
Config
meas.
loop
TAG 1
Level
Application
Level
Monitoring
Input
from
Input
no.4
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Adjust-
Signal
conditioning
Unit
0,0%
Outputs
Curr.
outputs
Mode
(monitoring)
Volt
outputs
Monit.
to
Low &
High
Relay
outputs
Low
%
5,0
High
%
5,0
ment
Relay
output
1
prop.
to
Percent
• Choose in the menu range "Parameter
adjustment relay outputs“ in the menu item
"Mode“ the mode "Monitoring“ for the selected relay.
The selected relay becomes energised. By
pushing the key switch (opener opens) the
actual measured value is first of all frozen
and, derived from this, the relay deenergises
if this value is underrun (Low) or exceeded
(High). After a reset and repeated push of
the key switch, the actual measured value is
again frozen.
Meas.
value 50 %
High 5 %
Low –5 %
>55 %
<45%
Relay
deenergises
If the monitoring function has to be deactivated, monitoring must first be cancelled in
the menu item "Mode“. Then the allocation of
the correction input (input 4) must be cancelled under the menu item "Monitoring“.
VEGAMET 515V27
Page 28
6.7 Configuration of inputs
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Configuration
Configuration
Config
inputs
Input
no. 1
Input
from
local
Met
MET02
MET03
MET04
….
Input
no. 2
like input no. 1
Serial
no.
XXXX
XXXX
Inputs no. 1 and 2: VBUS input (measured value)
Inputs no. 4 and 5: Correction input
Input
no. 4
Input
from
local
Met
Edit
ser. no.
0000
0000
Input
no. 5
like input no. 4
Channel
no.
K1
Input
undefined
Input1
Input 2
Input 3
Input local MET:
Here you can enquire another VEGAMET
(e.g. Met 04) and thereby specify that the
input signal will not be received from a directly connected sensor, but from a
VEGAMET connected via DISBUS. The enquired VEGAMET number corresponds to
the DISBUS address of the instrument (see
chapter 4.1).
Autom.
sensor
search
Input
undefined
Sensor
characteristics
Min.
meas.
range
0,00
Max.
meas.
range
1,00
Sensor
optimise
For radar/ultrasonic sensors: see
MINICOM menu schematic in
VEGASON/VEGAPULS operating
instruction manual under chapter
"Adjustment with VEGAMET or
MINICOM“.
Sensor
search
OK ?
Sensor
search
Now !
OK ?
The sensor characteristics values are provided by the connected sensor and cannot
be modified here.
The sensor menu is also provided by the
connected sensor. It allows the sensor to be
configured from VEGAMET.
If another VEGAMET has been chosen, the
menu item "Input undefined" appears in which
you can set the input number of the other
VEGAMET (see example "Weir control“).
28VEGAMET 515V
Page 29
Configuration
Example weir control
Sensor 1
Weir 2
Upstream water
Sensor 2B
In the example illustrating weir control, two
linked measurements are shown. The gauge
difference is measured on each weir with one
VEGAMET 515V. Normally, two sensors
would be required for each VEGAMET. However, since the gauge value of the upstream
water is valid for both VEGAMETs, one sensor is sufficient for the upstream water. The
input signal of sensor 1 (upstream water) is
transmitted via the DISBUS cable from one
VEGAMET (MET04) to the other VEGAMET.
The value of sensor 1 is therefore available to
both signal conditioning instruments.
Weir 1
Sensor 2A
%
100
+
-
OKESC
CONNECT
12
on
515V
"local MET“
Input2
DISBUS
%
100
+
-
OKESC
CONNECT
12
on
515V
"MET04“
Input1
VEGAMET 515V29
Page 30
6.8 Configuration of outputs
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Configuration
Configuration
Config
inputs
Config
meas.
loop
Config
outputs
Config
curr.
output
mA1
to
TAG No. 1
Config
volt
output
Volt1
to
TAG No. 1
mA2
to
TAG No. 2
Config
relay
output
Opera
ting
relay
Rel. 1
to
TAG No. 1
TAG 2
TAG 3
Volt2
to
TAG No. 2
mA3
to
TAG No. 3
Fail
safe
relay
Relay
Standard
Rel. 1
Standard
Hold
function
Reset of alarm
functions
Volt3
to
TAG No. 3
Config
PC/
DCS
output
PC/DCS
meas.
values
DCS 1
to
TAG No. 1
TAG 2/ 3
Input
undefined
Input 4
Input 5
Config
VEGADIS
DIS 1
to
TAG No. 1
TAG 2/ 3
PC/DCS
relay
status
off
Rel. 2
to
TAG No. 2
TAG 1
TAG 3
PC/DCS
input
status
off
DCS
7
to
----
Rel. 2
Standard
Hold
function
Reset of
alarm
functions
DIS 7
to
----
Input
undefined
Input4
Input5
30VEGAMET 515V
Page 31
Configuration
PC/PLC output and DIS output:
The PC/PLC signal as well as the DIS signal
are transmitted via the DISBUS. The PC/PLC
output delivers output signals from
VEGAMET to the interface converter
VEGACOM 557. This interface converter
transfers the signals of VEGAMET via different standard protocols to connected systems (DCS/PLC). In the menu items "DCS 1
to“ up to "DCS 7 to…“ the measurement loop
can either be allocated or switched off.
In the menu items "PC/DCS relay status“ and
"PC/DCS input status“ you can order the
switching conditions of the relays and the
conditions of the correction inputs to be
transmitted via the DISBUS.
Up to seven external indicating instruments
(VEGADIS 174) can be connected to
DISBUS. In the menu items "DIS 1 to…“ up to
"DIS 7 to–“ the measurement loop can either
be allocated or switched off.
Relay outputs
A hold function or a function for reset of alarm
functions can be allocated to each operating
relay. If this is done, the menu item "Input
undefined“ appears. There, "Input 4“ or "Input
5“ (correction inputs) can be assigned to the
relay. For a description of the relay functions,
see drawing.
Hold function
Parameter
High
Low
Taster
Relay output
Reset of alarm functions
Parameter
High
Low
Taster
Relay output
t
bz
d
t
bz
d
The relay modes "Hold function“ and "Reset
of alarm functions“ are configured as described above and activated each by an
external signal (correction signal input).
VEGAMET 515V31
Page 32
7 Parameter adjustment
Parameter adjustment
7.1 Adjustment with medium
You can either carry out the adjustment with
medium or without medium. To carry out an
adjustment with medium, it is necessary that
you know the percentage values of the vessel for the actual filling.
TAGNo. 1
%
xx,x
Param.
TAGNo. 1
Adjustment
with
medium
Minadjust
at %
0,0
For this procedure, percentage values must
be entered for min. and max. values that
correspond to the actual fillings. The sequence of the min. or max. value is not important. For example, if you know that your filling
is presently 80 %, you can enter in the menu
item Max. adjustment the value "80“. Then if
your filling is, for example, 10 %, you enter in
the menu item Min. adjustment the value "10“.
The larger the difference between the two
adjustment points, the more exact the measurement over the whole measuring curve. It
would be ideal if the adjustment could be
carried out at 0 % and 100 %. For practical
reasons, however, it is not always possible to
completely fill or empty a vessel. The distance between the two adjustment points
however, should be at least 10 % of the sensor range.
Signal
conditioning
Scaling
Maxadjust
at %
100,0
Lin.
curve
linear
Integr
ation
time s
0
Density
in
kg/dm3
1.000
Adjustment example
Max-
You know that your vessel is pres-
adjust
ently filled up to 80%. Move to the
at %
100,0
menu item "Max-adjust at % 100.0“.
Briefly push the [–] key, so that the
figure 100.0 flashes (editing mode).
Max-
Continue pushing [–] until the figure
adjust
80.0 appears on the display. Now
at %
80,0
push [OK] to save the adjustment.
Min-
Then, if your filling is for example
adjust
10 %, move to the menu item "Min-
at %
0,0
adjust at % 0.0“. Briefly push the [+]
key until the figure 0,0 flashes (edit-
ing mode).
Min-
Continue pushing [+] until "10“ ap-
adjust
pears on the display. Now push [OK]
at %
10,0
to save the adjustment.
Of course, you can carry out the adjustment
with medium also at 0 % and 100 % filling.
Please note: The figures 0 % and 100 % must
flash (editing mode) before they can be
saved with the [OK] key.
Note:
If the medium is later changed (medium with
another density value or dielectric constant)
and no new adjustment is possible or when
the MET indication and the DIS outputs refer
to height in metres, note the following: For
applications with hydrostatic pressure transmitters, the adjusted density value must
correspond to the value of the medium.
The menu item "Density…“ is displayed if the
appropriate sensor type was selected under
"Configuration measurement loop“.
32VEGAMET 515V
Page 33
Parameter adjustment
7.2 Adjustment without medium
Meas.
TAG-
value
No. 1
indication
%
xx,x
Param.
TAGNo. 1
Adjustment
with
w.o.
medium
medium
Adjustment
bar
m
For the "
Adjustment without medium
anticipated values (pressure or distance)
that correspond to 0 % and 100 % must be
entered.
Offset
correction
Sensor
unpres
sur’d?
OK?
Offset
corr.
Now!
OK?
0%
at
mA
4,000
100%
at
mA
20,000
“, two
Adjustment example
You know the data of the vessel, i.e. you
know that at 0 % filling the distance of the
liquid to the sensor is 3.5 m and at 100 %
filling 0.3 m.
Radar sensor
d1
0.3 m
at 100 %
d2
3.5 m at 0 %
In this menu item you choose the
Adjustment
adjustment unit, e.g. m (d), if you use
a radar sensor [m (d) = distance in
m(d)
m]. Use the [+] and [–] key, to
choose the unit and [OK] to save the selected unit.
Depending on the sensor type used, the
adjustment can be carried out in bar or m.
The menu item "Offset correction“ is activated
when connecting hydrostatic pressure transmitters. The offset correction should be carried out after installation of the sensors,
because the outputted sensor values can
differ slightly with various installation posi-
0%
Enter in these two menu items the
at
distance values (d1 and d2). With the
m (d)
[+] and [–] keys you set the values,
3,50
with [OK] you save them.
100%
at
Note that in this example, a lower
m (d)
distance value exists at 100 % filling
0,30
than at 0 % filling!
tions (vertical or horizontal mounting). With
offset correction, the measured data of an
unpressurized sensor (vessel empty) are
detected as a corrective value and taken into
account in all subsequent measurements.
VEGAMET 515V33
Page 34
Parameter adjustment
7.3 Adjustment with linked measurements
Linked measurements are possible with
VEGAMET 515V. For this, two sensors are
connected which carry out one measurement
each (TAG1 and TAG2). VEGAMET 515V
calculate from the two sensor values a third
value, e.g. a gauge difference (TAG3). When
adjusting linked measurements, factors other
than those associated with single measurements will possibly have to be taken into
account.
Adjustment gauge difference
TAG1 TAG2
TAG3
Upstream
water
Gauge difference
The adjustment of the individual measurement loops TAG1 and TAG3 is the same as
the adjustment of single measurement loops,
see under "7.1 Adjustment with medium“ and
"7.2 Adjustment without medium“. It is also
possible to carry out the adjustment only for
measurement loop TAG3. If you want to adjust all three measurement loops, the sequence is not important.
Downstream
water
Adjustment with medium
Adjustment
with
medium
Minadjust
at %
0,0
Maxadjust
at %
100,0
• Make sure the gauge (level) is the same
(difference = 0 %).
• Enter in the menu item "Min-adjust at %“ the
figure "0.0“ (push the "+“ or "-“ key until the
figure "0.0“ flashes, then push the "OK“
key).
• Mark the position of the sensor in the upstream water.
• Suspend the sensor 30 cm higher (if 30 cm
should correspond to 100 %). A 30 cm
lower gauge is now simulated.
• Enter in the menu item "Max-adjust at %“
the figure "100.0“ (push the "+“ or "-“ key
until the figure "100.0“ flashes, then push
the "OK“ key).
• Suspend the sensor in the original position.
Note:
You can also suspend the sensor in upstream water in a lower position in order to
carry out the max. adjustment.
Adjustment of the gauge difference
measurement loop (TAG3)
Suspend the sensors to the same level. Note
that a difference of 30 cm (e.g. with weir
control) can correspond to 100 %.
34VEGAMET 515V
Page 35
Parameter adjustment
Adjustment without medium
The easiest way to carry out the adjustment
without medium is to do it using the adjustment unit "m“.
Param.
TAGNo. 1
Adjustment
with
w.o.
medium
medium
Offset
Adjustment
m
bar
correction
Sensor
unpres
sur’d?
OK?
Offset
corr.
Now!
OK?
0%
at
m
0,00
100%
at
m
0,300
• Adjust in the menu item "Adjustment“ the
unit "m“.
• Enter in the menu item "0 % at m…“ and
"100 % at m…“ the values 0.00 and 0.30 (if
30 cm should correspond to 100 %).
Only carry out the offset correction at this
point if both sensors are unpressurised.
If you want to carry out the adjustment in bar,
you have to enter the differential values of the
pressures.
If you want that the gauge difference of 30
cm is shown in the meas. value indication in
cm, you can carry out a scaling (see "7.4
Scaling“).
Param.
TAGNo. 3
Ad-
Sig-
just-
nal
ment
conditioning
Scaling
0%
corres
ponds
0
100%
corres
ponds
300
Decimal
point
888,8
prop.
to
Height
Unit
cm
Adjustment with level difference and pressure difference
The adjustment of this linked measurement
corresponds to the adjustment of gauge
difference and is therefore not described
separately.
Adjustment with level measurement in
pressurized vessels
For the adjustment of measurement loop
TAG1 (level), you can proceed in the same
manner as with the adjustment of a single
measurement loop. The adjustment can be
either carried out with or without medium
(product).
Measurement loop TAG2 (gauge pressure) is
adjusted without medium, preferably in "bar“.
For level measurement (TAG1), an adjustment of TAG2 is not necessary.
Also, the adjustment of measurement loop
TAG3 (total pressure) must not be carried out
if you only want to know level of TAG1. Often,
the total pressure is the requested measured
value, e.g. if, in a filling system, the filling
pressure must be measured and controlled.
In this case, carry out the adjustment without
medium in "bar“.
VEGAMET 515V35
Page 36
Parameter adjustment
7.4 Scaling
TAG-
Meas.
No. 1
value
%
indication
xx,x
Param.
TAGNo. 1
Ad-
Sig-
just-
nal
ment
conditioning
Scaling
0%
corres
ponds
0
100%
corres
ponds
1000
Decimal
point
888,8
prop.
to
unde
fined
The adjustments under scaling determine
which number values and which unit will actually be displayed on the measured value
indication (upper menu item). In the menu
items "0 % corresponds…“ and "100 % corresponds…“ you can enter with [+] or [–]
and [OK] values which should be displayed
at 0 % and at 100 % filling. In the menu item
"Decimal point…“ you determine the position
of the decimal point. In the next menu item
"prop. to…“ you get a choice of parameters:
push [+] or [–] to select percent, mass, volume, pressure etc. and save with [OK]. The
menu item "Unit…“ then serves up a corresponding new choice, e.g. m, dm, cm, mm,
ft, yd and in, if you have selected the parameter "Height“ (compare chapter "8 Measuring
results and units“).
Unit
– –
Scaling example
You have a raised tank and you want the
level to be indicated in meters with three
places after the decimal point.
100 %
h=4.5 m
0 %
2.341 m
Set the number 1500 (with keys [+],
0%
corres
[–] and [OK]) in this menu item. This
ponds
corresponds to the level at 0 % (if
1500
h=1.5 m
you want to have a lower resolution of
the meas. value indication, you can also set
150 or 15).
100%
Set the number 4500 (with keys [+],
corres
[–] and [OK]) in this menu item. This
ponds
corresponds to the level at 100 %.
4500
Deci-
Adjust the decimal point as shown
mal
(with keys [+], [–] and [OK]).
point
8,888
prop.
Choose the term "Height“ in this
to
menu item (with keys [+], [–] and
Height
[OK]).
Percent
Pressure
Mass
After a scaling, the requested output can be
assigned to the scaled value in the menu
range "Parameter adjustment outputs“ (menu
items "prop. to…“ and "Unit…“).
If you have chosen "Height“ in the
Unit
previous menu item, you get a
m
choice of several units (mm, dm, cm,
m, yard…). Choose m (with the keys
dm
cm
[+], [–] and [OK]).
….
The scaled value is now shown in the
TAG No. 1
meas. value indication.
m
2,341
36VEGAMET 515V
Page 37
Parameter adjustment
7.5 Choose linearisation curve
Meas.
TAG-
value
No. 1
%
indication
xx,x
Param.
TAGNo. 1
Adjustment
VEGAMET's evaluation routines can be
adapted to different vessel forms. For this
purpose, you can choose between different
linearisation curves:
- linear (factory setting)
- cylindrical tank
- spherical tank
- Lin-curve 1
- Lin-curve 2
- Lin-curve 3
You can edit your modify lin. curves 1 … 3
yourself (see appropriate chapter).
Signal
conditioning
Scaling
Lin.
curve
linear
Cylindrical tank
Spherical tank
Lin. curve1
Lin. curve2
Lin. curve3
7.6 Integration time
Meas.
TAG-
value
No. 1
%
indication
xx,x
Param.
TAGNo. 1
Adjustment
If the integration time is set to 0 s (factory
setting), each quick change of the product
surface (e.g. waves) will be immediately
detected and interpreted as level change. All
output values of VEGAMET will follow the
wave movements. To avoid this, the integration time can be increased (a max. of
600 s is possible with VEGAMET). The higher
the time setting, the slower the measurement
reacts.
Signal
conditioning
Scaling
Lin.
curve
linear
Integr
ation
time s
0
After a linearisation, the requested output can
be assigned to the linearised value in the
menu range "Parameter adjustment outputs“
(menu items "prop. to…“ and "Unit…“).
VEGAMET 515V37
Page 38
7.7 Meas. value limitation
Meas.
TAG-
value
No. 1
%
indication
xx,x
Param.
TAGNo. 1
Parameter adjustment
Adjustment
Under
Signal
conditioning
Lin.
Scaling
curve
linear
Integr
ation
time s
0
"Meas. value limitation“
Density
in
kg/dm3
1.000
you can ex-
Meas.
value
limitation
negative
values
yes
no
clude negative values on the display (and on
the outputs).
Failure mode standard
means that the fail safe relay responds at a
sensor current of 3.6 mA or 21 mA.
Failure mode > 110 %
generates an additional fault signal at a
measured value of more than 110 % or less
than -10 %.
Failure
mode
Standard
>110 %
38VEGAMET 515V
Page 39
Parameter adjustment
7.8 Adjustment of output parameters
Current outputs/Volt outputs
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Signal
conditioning
Outputs
Adjustment
Curr.
outputs
Curr.
output
1
VolumeTARE%
prop.
to
Percent
Volt
outputs
Volt
output
1
prop.
to
Percent
VolumeTARE%
Curr.
output
2
like current output1
Unit
0,0%
Volt
output
2
like Volt output 1
Unit
0,0%
Curr.
output
3
Range
0%
+100%
Volt
output
3
Range
0%
+100%
Curr.
output
0/20mA
20/4mA
20/0mA
free
Volt
output
0/10V
10/2V
10/0V
2/10V
free
Curr.
at
0%
4,000
0,0
..20,0
Volt
at
0%
0,000
0,0
..10,0
Curr.
at
100%
20,000
0,0
..20,0
Volt
at
100%
10,000
0,0
..10,0
Failure
mode
0mA
22mA
- 0%
100%
Failure
mode
0V
11V
- 0%
100%
Curr.
limitation
on
off
Volt
limitation
on
off
VEGAMET 515V39
Page 40
Parameter adjustment
In the menu item "
prop. to“,
you can choose
between percent and volume. "Prop. to volume“ is then useful when you have chosen
spherical tank or user programmable curve
under "Conditioning - Linearisation curves“.
The output current or the output voltage then
assumes a value proportional to the volume
of the tank contents.
Unit:
Selection list dependent on the choice in the
previous menu item "prop. to“ (see chapter "8
Measuring results and units“).
The menu item range
"0%…100%“
only appears with measurement loop TAG-No. 3
(with gauge difference, level difference
measurement). The measured value can be
here -100 %.
Therefore, the range can be adjusted to
"
-100 % to +100 %“
. The correlation between
output current and output voltage can be
seen in the diagram below.
+100%
Range
0%
+100%
0%
4mA
0V
20mA
10V
Volt output/Current output:
If "free“ is chosen, the following (light grey)
menu items appear:
- Volt at 0 %/100 %:
Individual voltage values between 0 V and
+10 V can be set.
- Current at 0 %/100 %:
Individual current values between 0 mA
and +20 mA can be set.
In failure mode:
The height of the output current/output voltage in case of failure is set here. If "--“ is
chosen, the existing value is maintained.
Volt limitation/Current limitation on:
Voltage or current of the output remain within
the values which were set in the menu item
Voltage output/Current output, even if a level
of more than 100 % or less than 0 % is measured.
-100%
+100%
Range
-100%
0%
4mA
0V
20mA
10V
-100%
+100%
40VEGAMET 515V
Page 41
Parameter adjustment
Relay outputs
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Adjust-
Signal
conditioning
Relay
output
1
prop.
to
Percent
Volume
Height
Pressure
Outputs
Curr.
outputs
Relay
output
2
like relay output 1
Unit
0,0%
0,00 %
Tare %
ment
Volt
Relay
outputs
outputs
Monit.
Mode
on
Over-
Low &
fill
High
protec-
Dry run protec.
Switching window on
Switching window off
Rising tendency
Falling tendency
(monitoring)
Low
%
0,0
High
%
100,0
Deviation
s
1,0
Deviation
period
Scan
time
s
1
Failure
mode
off
hold
No. of
scans
10
Add’l
functions
Switching
delay
t on
s
0
t off
s
0
In the menu item "
parameter (meas. quantity) to which the relay
reacts. You can choose between percent and
volume ("Prop. to volume“ is then useful if
under "Conditioning - Linearisation curves“,
e.g. a spherical tank or a user-programmable
curve was chosen). In the next menu item
you can choose a suitable
item "
Mode
prop. to
“, you choose a
"Unit“
. In the menu
“, you can choose the mode of the
In all other modes, the menu items "Low“ and
"High“ appear. There, the switching points for
the relay can be set.
The menu item "
Failure mode
“ always appears. "Failure mode off“ causes the relay to
deenergize in case of failure (e.g. shortcircuit). "Failure mode hold“ causes the existing relay status to remain unchanged in case
of failure.
relay. The individual relay modes are described on the following pages.
Under "Additional functions“, you can adjust a
switch on (t on) and switch off delay (t off) of
If mode "Rising tendency“ or "Falling tendency“ was chosen, the menu items "Deviation“, "Deviation period“, "Scan time“ and "No.
the relay of max. 600 s (with the relay modes
"Rising tendency“ or "Falling tendency“, no
switching delay can be set).
of scans“ appear.
VEGAMET 515V41
Page 42
Parameter adjustment
d
t
bz
Relay mode "Overfill protection"
The relay deenergises at switching point
"High“ (safe condition).
Relay mode "Dry run protection"
The relay deenergises at switching point
"Low“ (safe condition).
Relay mode "Switching window on"
The relay is energised within the switching
window.
Relay mode "Switching window off"
High
Low
Relay output
High
Low
Relay output
High
Low
Relay output
Parameter
Parameter
Parameter
Parameter
bz
d
t
bz
d
t
The relay is deenergised within the switching
window.
High
Low
Relay output
d
t
Relay mode "Monitoring"
The enquired relay energises. By pushing
the key switch (opener), the actual measured
Meas.
value 50 %
High 5 %
Low –5 %
>55 %
<45%
Relay
deenergised
value is first of all frozen and, derived from
this, the relay deenergises if this value is
underrun (LOW) or exceeded (High). After a
reset and a repeated push of the key switch,
the actual measured value is frozen again.
42VEGAMET 515V
bz
Page 43
Parameter adjustment
Tendency determination
The level change within the cycle time (ta) is determined and after all cycles (n) an average
value is generated out of the sum of the level changes. If this average value exceeds a previously defined % value, the tendency determination responds, i.e. an energised relay
deenergises. After the selection of "Tendency rising“ or "Tendency falling“ the menu items
"Deviation in %“ and "Deviation per time“ are activated.
%
t
ges.
t
a
–– Mode Tendency rising
The relay deenergises when the
rising average value of "t tot.“
exceeds the previously defined %
value.
Tendency falling
Relay output
ta = cycle time
n = number of cycles
%
4 n=4
321
t ges.
t
a
same dependence, however with
falling average value
4 n=4
321
Relay output
ta = cycle time
n = number of cycles
–– Deviation in %2If the level change exceeds the % value set
here, the tendency determination responds.
0 … 110 %
Example 2 % (change) per t tot.
–– Deviation per time
t
bz
d
t
bz
d
–– Cycle time
in s60Adjustment range 0 … 999 s
Example 60 s correspond to 1 minute
–– Number of
cycles4Adjustment range 0 … 99 cycles
Example: t tot = ta • n = 60 • 4 = 240 s = 4 min,
i.e. after every 4 minutes, the average value of
the deviation is generated and the result is
compared with the given deviation in %, here in
the example = 2 %
VEGAMET 515V43
Page 44
MET indication/PC/DCS outputs/DIS outputs
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Adjust-
Signal
conditioning
Outputs
Curr.
outputs
Volt
outputs
Relay
outputs
MET
display
ment
PC/
DCS
outputs
DIS
outputs
Parameter adjustment
prop.
to
Percent
scaled
Height
…
MET indication/Unit:
In the menu item "prop. to“, you choose the
parameter to which the VEGAMET display
reacts (if you have carried out a scaling under the menu item "Signal conditioning“,
"prop. to scaled“ will have been automatically
adjusted, i.e. the display will indicate the
scaled value). "Prop. to volume“ is then useful
if you have selected, for example, a spherical
tank or a user-programmable lin. curve under
"Signal conditioning - Linearisation curves“.
The indicated value on the VEGAMET display
then correlates proportionally to the volume of
the tank contents.
Unit
0,0%
PC/
DCS
output 1
prop.
to
Percent
scaled
Height
…
DIS
output
1
prop.
to
Percent
scaled
Height
…
Unit
0,0%
DIS
output
2
like DIS output 1
Unit
0,0%
DCS
output
2
like DCS
output 1
DIS
output
7
DCS
output
7
PC/DCS outputs and DIS outputs:
Adjustment according to the same procedure
as the MET indication.
44VEGAMET 515V
Page 45
Parameter adjustment
7.9 Simulation7.10 Special function: Reset
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Adjust-
Signal
Out-
ment
conditioning
puts
Simulation
Simulation
Now!
OK?
Simulation
%
XX,X
To check the outputs and connected instruments, you can adjust in this menu item any
individual percentage value with the keys [+]
and [–]. The initial point is always the actual
measured value. The indicated value flashes
when the simulation is activated. 15 minutes
after adjustment of the simulated value, the
simulation is automatically terminated and the
display reset to measured value indication.
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Adjust-
Signal
Out-
Simu-
ment
conditioning
puts
lation
Special
funct.
Reset
Level
Reset
OK ?
Reset
Now!
OK ?
With this reset, all parameter values of the
measurement loop are reset to factory setting
(empty and full adjustment, relay outputs,
current output, volt output and scaling of the
meas. value indication are reset).
VEGAMET 515V45
Page 46
7.11 Special function: Manual and automatic correction
TAG Nr. 1
%
xx,x
Param.
TAGNr. 1
Adjustment
Signal
conditioning
Outputs
Simulation
Specal
funct.
Parameter adjustment
Simulation
Now!
OK?
Simulation
%
XX,X
Automatic correction:
1)
This menu item only appears if in the measurement
Reset
Level
Reset
OK ?
Reset
Now!
OK ?
Real
value
corr.
at % 0
1)
loop configuration a correction mode was set. You
then enter here the percentage value to which the
automatic correction should respond (triggered by
the level sensor on the correction input).
Manual offset correction
(only with hydrostatic pressure transmitters)
This correction can be carried out to compensate deviations of the sensor values due
to ageing. In order to do this, the sensor
must be in an unpressurized condition
(empty vessel).
Density
corr.
at %
50,0
Manual
corr.
1)
Offset
correction
Offset
correction
OK?
Correction
Now !
OK ?
Real
value
corr.
Real
value
corr.
OK?
Correction
%
0,0
Correction
Now !
OK ?
Manual real value correction
In the menu item "Correction % 0.0“ you can
enter a percentage value which corresponds
to the actual existing level.
46VEGAMET 515V
Page 47
Additional functions
8 Additional functions
8.1 Linearisation curves
TAG No. 1
%
xx,x
Param.
Configuration
Add’l
funct.
Password
off
Edit
lin.
curve
Lin.
curve1
Add
lin.
point
TAGNo. 1
Lin.
Lin.
curve2
curve3
like LIN curve 1
Edit lin.
curve 1
Delete
lin.
point
x %
0,0
y V%
0,0
x 0 %
0,0
y 0 V%
0,0
x 1 %
100,0
y 1 V%
100,0
Index marker number (0…32)
Linearisation curve 1 … 3
A linearisation curve is generated by a certain number of index markers and their value
pairs. A value pair consists of a value for
level percent (X %) and a value for volume
percent (Y V %). Up to max. 32 index markers can be included. A linearisation curve can
be terminated after any individual number of
index markers. The signal conditioning instrument terminates the linearisation curve automatically with the values X = 100 % and Y
= 100 %.
The data for the value pairs can be determined by incremental filling or can be taken
from value tables provided by vessel manufacturer.
Example of incremental filling
Initial situation:
- the adjustment 0 % and 100 % has already
been carried out by you
- the total volume of the vessel is known to
you, for example 300 m
- the gauging volume (increment) in this
example was set at 15 m³
3
x 2 %
100,0
y 2 V%
100,0
x 32 %
100,0
y 32V%
100,0
x 0,0
y 0,0
delete
Delete
Now?
x 100,0
y 100,0
delete
Delete
Now?
1st index marker
- fill the vessel with the liquid quantity of the
gauging volume (increment) of 15 m
3
- the value for level percent is indicated on
VEGAMET, transfer this value into the linearity protocol, column X %
- the value for volume percent must be calculated acc. to the following formula
100 % x gauging volume
Y 1 V = ––––––––––––––––––––––––
total volume
100 % x 15 m
= ––––––––––––––– = 5 %
300 m
3
3
Transfer this value also into the linearity
protocol, column Y V %.
2nd index marker etc.
See also the following illustration.
VEGAMET 515V47
Page 48
Illustration of a linearisation curve
Additional functions
X (in %)
100
80
60
40
20
Add
lin.
point
x %
0,0
y V%
0,0
With [OK] the new value pair
is automatically added to the
correct position in the
linearisation curve. The existing index markers are resorted.
0
X1 %
12,0
Y1 V %
5,0
Edit lin.
curve 1
x 0 %
x 1 %
100,0
y 1 V%
100,0
x 2 %
100,0
y 2 V%
100,0
0,0
y 0 V%
0,0
Selection of an index marker
1 … 32 with [–>] or [<–],
editing of the Y value with [+]
or [–], accept with [OK].
406080100
Delete
lin.
point
x 0,0
y 0,0
delete
Delete
Now?
X32 %
100,0
Y32V %
100,0
x100,0
y100,0
delete
Delete
Now?
To delete an index marker,
select the value pair with [–>]
or [<–]. With [OK] the value
pair and the respective index
marker are deleted in the lin.
curve. The remaining index
markers move up.
V (in %)20
48VEGAMET 515V
Page 49
Additional functions
Linearity protocol
A linearisation curve is entered in the menu item "Add. lin. point“.
For radar/ultrasonic sensors: see
MINICOM menu schematic in the
VEGASON/VEGAPULS operating
instruction manual under chapter
"Adjustment with VEGAMET or
Info
Input
info
Input1
MINICOM“.
VEGAMET
info
Type
MET515V
like input1
Program
info
Program
info
xxxxxx
Serial
number
xxxx
xxxx
Input 4
Sensor
type
Vibration
Meas.
loop
info
TAG 1
TAG No. 1
Minadjust
at %
0,0
Instr.
addr.
Input5Input2
like input 4
Input
from
local
MET
TAG 2
TAG 3
TAG -
TAG -
No. 2
No. 3
Min-
Max-
adjust
adjust
bar
at %
0,000
Softw.
Softw.
vers.
date
V1.13
1
06
47/97
Chan-
Input
nel no.
undef-
K1
ined
100,0
Act.
switch
status
Maxadjust
bar
1,000
Param.
vers.
5
Parameter
Level
All menu items under "Info“ have only a display (readout) function.
50VEGAMET 515V
Page 51
Additional functions
8.3 Password
TAGNo. 1
%
xx,x
Param.
TAGNo. 1
Param.
TAGNo. 2
Param.
TAGNo. 3
Configuration
You have the option of protecting VEGAMET
515V against unauthorised or incorrect adjustment. Activate in the menu "Password“ the
option "on“ (with keys [+], [–] and [OK]).
Now, you can only check the meas. value on
the display. You can still reach any menu item
with the arrow and OK keys, but as soon as
you want to modify a setting, you are asked
for the password. Enter the number "515“
(with the keys [+], [–] and [OK]), to carry out
the adjustment. Entering the password once
is sufficient to open all protected menu items.
Add’l
funct.
Password
on
off
8.4 Language
TAGNo. 1
%
xx,x
Confi-
Param.
Param.
TAGNo. 1
TAGNo. 2
Param.
TAGNo. 3
Password
on
off
In the menu item "Language“ you can choose
a language other than German, such as
English, French etc., as menu language (with
keys [+], [–] and [OK]).
guration
Edit
lin.
curve
Add’l
funct.
Info
Language
English
German
French
etc.
VEGAMET 515V51
Page 52
8.5 Reset VEGAMET
TAG No. 1
%
xx,x
Param.
Configuration
Add’l
funct.
Password
off
Lin.
curve
TAGNo. 1
Info
Language
English
Reset
VEGAMET
Additional functions
Reset
configuration
TAG 1
to
linked
application
Delete
all
TAGs?
Reset
OK?
Reset
Now!
OK?
to
single
meas.
Reset
TAG1
Delete
TAG1?
Reset
OK?
Reset
Now!
OK?
Reset
TAG2
Delete
TAG2?
Reset
OK?
Reset
Now!
OK?
“, you reset
to de
fault
Reset
OK?
Reset
Now!
OK?
With the function "
TAG 1 to default
the whole configuration and parameter adjustment to factory setting. However, the
basic configuration remains, e.g. "Level
measurement/Hydrostatic“.
With the function "
Reset TAG 1
“, the basic
configuration is also deleted. Then, the measurement loop must be completely re-created
(config. meas. loop).
With "
Reset sensor values
“ the adjusted
sensor data of the connected sensor are
deleted.
Reset
all
inputs
Delete
all sensor values
Reset
OK?
Reset
Now!
OK?
Reset
lin.
curves
Reset
Reset
lin.
curve2
Reset
OK?
Reset
Now!
OK?
Reset
lin.
curve3
Reset
OK?
Reset
Now!
OK?
lin.
curve1
Reset
OK?
Reset
Now!
OK?
Reset Lin-curves
Reset
sensor
values
Reset
Reset
input2
input1
Reset
Reset
OK?
OK?
Reset
Reset
Now!
Now!
OK?
OK?
With the function "
delete your manually-edited linearisation
curves.
With "Reset all curves“, you delete all together, with "Reset lin-curve 1“, you always
delete only one curve.
The three pre-programmed lin-curves "linear,
horiz. cylindrical tank, spherical tank“ cannot
be deleted.
Reset
all
curves
Delete
all
curves?
Reset
OK?
Reset
Now!
OK?
“, you can
52VEGAMET 515V
Page 53
Measuring results and units
8 Measuring results and units
With all outputs and with scaling you will find
the menu items "prop. to“ and "Unit“. Here
you can choose the meas. result and the
suitable unit. The list of the units depends on
the selected meas. result (physical quantity).
The available units of measurement are listed
in the table.
Current/Volt output
prop. toUnit
Percent0,0 %
Tare %
Volum eV %
Tare V %
VEGADIS/DCS/MET indication
prop. toUnit
Percent0,0 %
0,00 %
Tare %
Volum eV %
Tare V %
Heightm
Pressurebar
scaledscaled
Tare scaled
Relay outputs
prop. toUnit
Percent0,0 %
Volum eV %
Heightm
Pressurebar
Scaling
prop. toUnit
Percent%
Volum em
Heightm
Tare %
Tare V %
3
hl
l
gal
V %
3
ft
3
in
dm
cm
mm
yd
ft
in
Distancem (d)
Pressurebar
mbar
psi
hPa
Masskg
t
lb
Densitykg/dm
g/cm
3
3
Distancem (d)
VEGAMET 515V53
Page 54
9 Diagnosis
Diagnosis
9.1 Maintenance
The instrument is maintenance-free.
9.2 Simulation
Note
With activated simulation, the entire display
flashes. The measured data delivered by the
sensors are not released for processing you should, therefore, quit this menu item as
soon as possible. After 15 minutes, there is
an automatic reset to meas. value indication.
9.3 Fault signal
The signal conditioning instrument and the
connected sensor are continuously monitored during operation. During the parameter
adjustment, the set values are checked for
plausibility. In both cases, a fault signal can
be triggered by irregularities.
In case of failure, an error code is displayed
(only within the meas. value indication), the
fail safe relay deenergises, the signal lamp
lights and the outputs react according to their
preadjusted control behaviour.
Configuration
For each individual measurement loop, you
can determine if a fault signal should be triggered or not. See menu range "Configuration
measurement loop“, menu item "Fault signal“.
Parameter adjustment
In the menu item "Failure mode“, you can
determine individually for each measurement
loop whether and how a measuring range
overrun should trigger a fault signal.
9.4 Repair
For safety and warranty reasons repair on
the instrument must only be carried out by
VEGA staff.
In case of defect, please return the impaired
instrument with a short description of the fault
to our repair department.
The error codes are listed in section "9.5
Error codes“ on the following pages.
The diagnosis info is transmitted as a running
text tape.
54VEGAMET 515V
Page 55
Diagnosis
9.5 Error codes
Error codeMeaning
E004This operation is not supported by the instrument (different software versions).
E007Sensor type does not match.
E008VBUS sensor not found. Sensor cannot be found.
E010The serial numbers of the connected VBUS sensors are not yet assigned to an
E013VBUS sensor signals failure; however the communication with the sensor is
You have tried to activate a function with a newer software version of VEGA
Visual Operating in an instrument with an older software version. However, the
instrument cannot support this operation.
Have the software of your instrument updated.
A VBUS sensor with the serial number given by you was found, however the
sensor type does not match the settings.
Check the entered serial number and modify the configuration.
Probable reasons:
- no sensor connected
- no sensor with this serial number connected
- line break (VBUS)
- short-circuit (VBUS)
- sensor is defective
- interfered communication (VBUS)
Also possible:
- failure on the VBUS transmitter/receiver part of VEGAMET
Check the serial number, the connection cable, the sensor and if necessary
VEGAMET.
input component.
okay.
Probable reasons:
- failure on the VBUS sensor
Check the VBUS sensor.
E016Empty/full adjustment reversed. The values for the empty adjustment are
E017Adjustment span too small. The adjustment values are too close together or are
E021Span of the scaling values too small or values reversed. The scaling values are
E022Scaling span too big.
E027Actual measured value is less than -10 %.
E028Actual measured value is more than 110 %.
VEGAMET 515V55
above the values for the full adjustment. Carry out the adjustment again.
identical. Carry out the adjustment again.
Also possible:
- an incorrect measuring range has been set on the capacitive electrode.
Adjust a higher range.
too close together or are identical. Please check the scaling.
Page 56
Error codeMeaning
Diagnosis
E031Same DISBUS address. At least one other instrument with the same DISBUS
address is connected to the DISBUS. Please check the DISBUS address.
E032Instrument address 0. Instrument address 0 is adjusted on VEGAMET. With
this address, VEGAMET does not participate in the DISBUS communication,
even if the DISBUS is wired. If necessary, modify the instrument address.
E034Instrument error. A considerable error was detected on the instrument (possi-
bly defective EEPROM). Disconnect voltage and then connect again (cold
start). If the error is not eliminated, please contact your VEGA agency.
E035Instrument error. A considerable error was detected on the instrument (check
sum error in EPROM). Disconnect voltage and then connect again (cold start).
If the error is not eliminated, please contact your VEGA agency.
E050Remote VEGAMET not found. The VEGAMET from which a measured value is
expected cannot be found.
Reasons:
- the remote VEGAMET is not connected
- no VEGAMET with this instrument address connected
- line break (DISBUS)
- short-circuit (DISBUS)
- the remote VEGAMET is defective
- interfered communication (DISBUS)
Also possible:
- failure on the DISBUS transmitter/receiver part of VEGAMET
Please check the instrument address on both VEGAMETs, the connection
cable, the remote VEGAMET and, if necessary, this VEGAMET.
E051Failure during transmission of measured values via DISBUS. It is not possible
to receive a valid measured value via DISBUS from the remote VEGAMET; the
communication with the remote VEGAMET, however, is okay.
Reason:
- failure on remote VEGAMET
- the input on the remote VEGAMET which should be transmitted is not yet
configured
Please check the remote VEGAMET and, if necessary, its input configuration.
E052The input number of the VEGAMET from which a measured value is expected
has not yet been assigned.
56VEGAMET 515V
Page 57
Diagnosis
Error codeMeaning
E060Span of the current output values too small. The values for the current output
are too close together or are identical. Please check the adjustment of the
current output.
E061Span of the voltage output values too small. The values for the voltage output
are too close together or are identical. Please check the adjustment of the
voltage output.
E063Wrong sensor characteristics values. Please check the sensor characteristics
values
E102A reset of the configuration was carried out on this VEGAMET via the key-
board.
E110Switching hysteresis too small. The switching points are too close together or
are identical.
E111Relay switching points reversed. The relay switching points are reversed with
regard to the selected relay mode.
VEGAMET 515V57
Page 58
Menu schematic
TAG1-3
xx,x
xx,x
xx,x
3
TAG No. 1
%
xx,x
Param.
TAGNo. 1
Adjustment
TAG No. 2
%
xx,x
Param.
TAGNo. 2
like TAG- No .1
Signal
conditioning
4
TAG No. 3
%
xx,x
Param.
TAGNo. 3
Outputs
Curr.
outputs
Curr.
output
1
prop.
to
Percent
Meas. value
indication
Volt
outputs
Volt
output
1
prop.
to
Percent
Curr.
output
2
Unit
0,0%
Volt
output
2
like Volt output1
Unit
0,0%
Curr.
output
3
like current output 1
Range
0%
+100%
10Menu schematic
Volt
output
3
Range
0%
+100%
Curr.
output
4/20mA
Add’l
funct.
Volt
output
0/10V
Current
at
0%
4,000
Volt
at
0%
0,000
Current
at
100%
20,000
Configuration
see following pages
Volt
at
100%
10,000
Failure
mode
0mA
Failure
mode
0V
Curr.
limitation
on
Volt
limitation
on
with
medium
Minadjust
at %
0,0
Scaling Lin.
0%
corres
ponds
0
Maxadjust
at %
100,0
curve
linear
100%
corres
ponds
1000
Integr
ation
time s
0
Decimal
point
888,8
w.o.
medium
Adjustment
bar
Density
in
Kg/dm3
1,000
prop.
to
undefined
Offset
correction
Sensor
unpres
sur’d ?
OK ?
Offset
corr.
Now!
OK ?
Unit
– –
0%
at
bar
0,000
100%
at
bar
1,000
Meas.
value
limitation
negative
values
yes
Failure
mode
Standard
ESC
OK
With these keys you
move in the menu field
58VEGAMET 515V
Page 59
Menu schematic
Language
English
1
Relay
outputs
In white letters you can
see the parameters
which can be modified
with the [+] or [–] key
and saved with the
[OK] key.
2
3
… setup sequence
MET
display
prop.
Unit
to
Percent
0,0%
PC/DCS
outputs
DIS
outputs
DIS
output
1
prop.
to
Percent
Sensor
type
hydrostatic
Bold print/italic is the
sensor or measured
value information
which cannot be
modified in this position.
Simu-
Spe-
lation
cial
funct.
Reset
Simu-
Level
lation
Now!
OK?
Simulation
%
XX,X
Unit
0,0%
Reset
OK ?
Reset
Now!
OK ?
DIS
output
7
like DIS output 1
Real
value
corr.
at %
0
Density
corr.
at %
0
Curr.
at
100%
20,000
Light grey menu items
are only displayed if
necessary (depending
on the adjustment in
other menu items).
Manual
corr.
Offset
correction
Offset
correction
OK?
Correction
Now !
OK ?
Real
value
corr.
Real
value
corr.
OK?
Correction
%
0,0
Correction
Now !
OK ?
Unit
0,00%
PC/DCS
output 7
Low
%
0,0
like PC/DCS output 1
High
Deviation
%
%
100,0
1,0
Deviation
period
Scan
time
s
1
Failure
mode
off
No. of
scans
10
Add’l
funct.
Switching
delay
t on
s
0
t off
s
0
Relay
output
1
prop.
to
Percent
Relay
output
2
like relay output 1
Unit
0,0%
Mode
Overfill
protection
PC/DCS
output 1
prop.
to
Percent
Monit.
on
Low&
High
VEGAMET 515V59
Page 60
Menu schematic
TAG1-3
xx,x
xx,x
xx,x
Config
inputs
2
TAG No. 2
%
xx,x
Param.
TAGNo. 2
TAG -
Meas. value
No. 3
indication
%
xx,x
Param.
TAGNo. 3
with linked
applications
Application
Level
1)
1) The settings in these menu
TAG No. 1
%
xx,x
Param.
TAGNo. 1
see previous pages
Config
meas.
loop
1
with two
independent
single measurements
items can only be modified if a
"Reset to linked application“
was carried out beforehand
2) The settings in these menu
items can only be modified if a
TAG 1
Level
Application
Level
2)2)2)2)
TAG 2
Level
Sensor
type
Hydrostatic
"Reset to single measurement“
was carried out beforehand
like TAG 1
Mode
Standard
Sensor
type
Hydrostatic
1)
Option
no
option
Configuration
Mode
Pressurized
1)1)
Sensor
coordination
Position A
Input
no.1
Add’l
functions
see following pages
Option
no
option
TAGName
TAG No. 1
Position D
undefined
Sensor
coordination
Position A
Input
no.1
Fault
signal
on
Position E
undefined
TAG 1
Level
TAGName
TAG No. 1
Position B
Input
no.2
Tar e
Input
from
undefined
TAG 2
Gauge
pres.
Fault
signal
on
Position C
Input
no.3
TAG 3
Total
pres.
like TAG1
Monitoring
Input
from
undefined
Tar e
Input
from
undefined
Position D
Input
no.4
Monitoring
Input
from
undefined
Position E
Input
no.5
Input
no. 1
Input
from
local
Met
Input
no. 2
like input no.1
Serial
no.
XXXX
XXXX
Input
no. 4
Input
from
local
Met
Edit
ser.no.
0000
0000
Input
no. 5
like input no.4
Channel no.
K1
Input
undefined
Input
undefined
Sensor
characteristics
Min.
meas.
range
0,00
Max.
meas.
range
1,00
Autom.
sensor
search
Sensor
search
OK ?
Sensor
optimize
For radar/ultrasonic sensors: see MINICOM menu
schematic in the VEGASON/VEGAPULS operating
instruction manual under chapter "Adjustment with
VEGAMET or MINICOM“.
Sensor
search
Now !
OK ?
5 … 10
60VEGAMET 515V
Page 61
Menu schematic
Config
outputs
Config
curr.
output
Curr.1
to
TAG No. 1
Config
Volt
output
Volt 1
to
TAG No. 1
Curr.2
to
TAG No. 2
Config
relay
outp.
Operating
relay
Rel. 1
to
TAG No. 1
Volt 2
to
TAG No. 2
Curr.3
to
TAG No. 3
Fail
safe
relay
Relay
Standard
Rel. 1
Standard
Volt 3
to
TAG No. 3
ESC
OK
Config
PC/
DCS
output
PC/
DCS
meas.
values
DCS 1
to
TAG No. 1
Input
undefined
PC/DCS
relay
status
off
Rel. 2
to
TAG No. 2
DCS 2
to
----
PC/DCS
input
status
off
Rel. 2
Standard
Config
VEGADIS
DIS 1
to
TAG No. 1
Input
undefined
1
Language
English
Sensor
type
Hydrostatic
DIS 7
to
----
2
3
… setup sequence
In white letters you can see the
parameters which can be modified
with the [+] or [–] key and saved
with the [OK] key.
Bold print/italic is the sensor or
measured value information
which cannot be modified in this
position.
Light grey menu items are only
displayed if necessary (depending on the adjustment in other
menu items).
With these keys you
move in the menu field
Current at
100%
20,000
VEGAMET 515V61
Page 62
Menu schematic
TAG1-3
xx,x
xx,x
xx,x
Password
on
TAG No. 1
%
xx,x
Param.
TAGNo. 1
see previous pages
Edit
lin.
curve
For radar/ultrasonic sensors: see MINICOM
under chapter "Adjustment with VEGAMET
Lin.
curve1
TAG -
TAG No. 2
%
xx,x
Param.
TAGNo. 2
InfoLang-
Input
info
Input 1Typ e
menu schematic in the VEGASON/
VEGAPULS operating instruction manual
Lin.
curve2
Meas. value
No. 3
indication
%
xx,x
Param.
TAGNo. 3
Input 4
Input 2Input 5
like input1like input4
Sensor
type
Vibration
or MINICOM“.
Lin.
curve3
Configuration
see previous
pages
Input
from
local
MET
VEGAMET
info
MET515V
Channel no.
K1
Add’l
functions
Program
info
Program
info
xxxxxx
Serial
number
xxxx
xxxx
Input
undefined
Meas.
loop
info
TAG 1
TAG No. 1
Minadjust at
%
0,0
Instr.
addr.
1
Actual
switch
status
TAG 2
TAG No. 2
Minadjust
mA
4,000
Softw.
vers.
1.13
uage
English
like TAG 1
Maxadjust at
%
100,0
Softw.
date
47/97
TAG 3
TAG No. 3
Maxadjust
mA
20,000
Param.
vers.
5
Parameter
Level
Add
lin.
point
x %
0,0
y V%
0,0
like LIN-curve 1
Edit
lin.
curve 1
x 0 %
0,0
y 0 V%
0,0
x 1 %
x 2 %
100,0
y 1 V%
100,0
100,0
y 2 V%
100,0
Index marker number (0…32)
x32 %
100,0
y32V%
100,0
Delete
lin.
point
x 0,0
y 0,0
delete
Delete
Now?
x100,0
y100,0
delete
Delete
Now?
62VEGAMET 515V
Page 63
Menu schematic
Reset
VEGAMET
Reset
configuration
TAG1
to default
Reset
OK?
Reset
Now!
OK?
TAG2
to default
Reset
OK?
Reset
Now!
OK?
to
linked
application
Delete
all
TAGs ?
Reset
OK?
Reset
Now!
OK?
to
single
meas.
Reset
TAG 1
Delete
TAG 1
?
Reset
OK?
Reset
Now !
OK?
Reset
sensor
value
Reset
TAG 2
Delete
TAG 2
?
Reset
OK?
Reset
Now !
OK?
Service
only accessible with
password
Reset
lin.
curves
Reset
lin.
curve1
Reset
OK?
Reset
Now !
OK?
Reset
lin.
curve2
Reset
OK?
Reset
Now !
OK?
Reset
lin.
curve3
Reset
OK?
Reset
Now !
OK?
Reset
all
curves
Delete
all
curves?
Reset
OK?
Reset
Now !
OK?
In white letters you can see the
parameters which can be modified
with the [+] or [–] key and saved
with the [OK] key.
Bold print/italic is the sensor or
measured value information
which cannot be modified in this
position.
Light grey menu items are only
displayed if necessary (depending on the adjustment in other
menu items).
ESC
OK
With these keys you
move in the menu field
Language
English
Sensor
type
Hydrostatic
Current at
100%
20,000
VEGAMET 515V63
Page 64
VEGA Grieshaber KG
Am Hohenstein 113
D-77761 Schiltach
Phone(07836) 50-0
Fax(07836) 50-201
E-Mailinfo@de.vega.com
www.vega.com
Safety information
Please read this manual carefully, and also take
note of country-specific installation standards
(e.g. the VDE regulations in Germany) as well
as all prevailing safety regulations and accident prevention rules.
For safety and warranty reasons, any internal
work on the instruments, apart from that involved in normal installation and electrical connection, must be carried out only by qualified
VEGA personnel.
Note Ex area
Please note the attached safety instructions
containing important information on installation
and operation in Ex areas.
These safety instructions are part of the operating instructions and come with the Ex approved instruments.
ISO 9001
All statements concerning scope of delivery, application, practical
use and operating conditions of the sensors and processing systems correspond to the latest information at the time of printing.
Technical data to subject to alterations
2.22 709 / February 2001
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.