VEGA MET514D User Manual

Operating Instruction
VEGAMET 514… D
Level and Pressure
-
ESC
%
100
+
OK
CONNECT
2
1
on
out
in
Contents
Safety information ....................................................................................... 2
1 Product description
1.1 Function and configuration ................................................................ 4
1.2 Types and versions ............................................................................ 5
1.3 Applications ....................................................................................... 6
1.4 Technical data .................................................................................... 7
1.5 Dimensions ....................................................................................... 10
1.6 Ex-technical data ............................................................................. 11
1.7 Approvals ......................................................................................... 11
2 Mounting
2.1 Mounting versions ............................................................................ 12
2.2 Instrument address .......................................................................... 13
3 Electrical connection......................................................................... 14
Contents

Safety information

The described module must only be inser ted and operated as described in this operating instruction. Please note that other action can cause damage for which VEGA does not take responsibility.
2 VEGAMET 514… D
Contents
4 Set-up
4.1 Indicating and operating elements ................................................. 15
4.2 Adjustment system ........................................................................... 16
4.3 Menu list ........................................................................................... 17
4.4 Adjustment via PC ............................................................................ 18
4.5 Set-up sequence .............................................................................. 18
4.6 Set-up, configuration ....................................................................... 19
4.6.1 Configuration inputs............................................................ 19
4.6.2 Configuration meas. point .................................................. 21
4.6.3 Configuration output ........................................................... 23
4.7 Set-up, parameter adjustment ......................................................... 26
4.7.1 Parameter adjustment MST1, adjustment .......................... 26
4.7.2 Flow measurement .............................................................. 28
4.7.3 Parameter adjustment MST1 and MST2, processing ........ 33
4.7.4 Parameter adjustment MST1 and MST2, outputs .............. 35
4.7.5 Parameter adjustment MST1 and MST2, simulation .......... 42
4.7.6 Parameter adjustment MST1 and MST2, special functions43
4.8 Set-up, additional functions ............................................................. 43
4.8.1 Password ............................................................................. 43
4.8.2 Linearization curve 1 … 3 ................................................... 43
4.8.3 Info ....................................................................................... 44
4.8.4 Language ............................................................................ 44
4.8.5 VEGAMET Reset ................................................................. 45
4.8.6 Service................................................................................. 45
5 Diagnosis
5.1 Detailled menu list ............................................................................ 46
5.2 Maintenance ..................................................................................... 48
5.3 Repair ............................................................................................... 48
5.4 Fault signal ....................................................................................... 48
5.5 Error codes ....................................................................................... 49
VEGAMET 514… D 3

1 Product description

Product description

1.1 Function and configuration

VEGAMET 514 D and 514 DV signal conditioning instruments are designed for flow measurement.
Function
Flow measurement in flumes
Flumes or their throat cause a reflux of the measured product. The flow can be determined out of the appropriate levels.
Sensors detect levels and deliver analog or digital measured data to connected VEGAMET.
In the signal conditioning instruments the parameter "flow" is generated out of the measured data.
The output result is provided as current, volt and relay output as well as digital signal output (DISBUS).
The current and relay outputs are sometimes provided with a flow specific preconfiguration.
Sketch
AB
Function components
Parameter for
- meas. point 1 (MST1)
- meas. point 2 (MST2)
Application
- Sensor
- Kind
- Option
- Sensor position A … D
FB
1
2
EB
1
AB
AB
Display
AB
EB
mA
V
PC
DIS
4
Configuration
The signal conditioning instruments are microcomputer controlled and consist of software components.
Mechanically the signal conditioning instruments are designed in 19"-technology (width 5 TE = 24,5 mm).
Output components
Current outputs 0 … 20 mA Voltage outputs 0 … 10 V
Relay outputs 1 … 2
Display
Digital signal output
- PC/DCS-output
- DISBUS-output
Input components
Correction signal input
- input 4
- channel no. K1
Sensor input
- input 1
- channel no. S1 or
- series no of sensor 1
4 VEGAMET 514… D
Product description
Examplary application
- flow measurement in a flume
- with offset correction (by additional sensor)
A
no. 1 S1
no. 4 K1
Inputs
- input 1 and 4
- channel no. S1 and K1
D
Sensor
- offset correction signal in pos. D
- sensor 1 in pos. A
Meas. points Parameter
- MST1 = A - level
- MST2 = A - flow

1.2 Types and versions

According to the application and the appropriate sensor the type and version of the signal conditioning instrument can be determined.
VEGAMET for sensors with analog measured data transmission
- hydrostatic pressure transmitter
VEGAMET Number of Number of outpouts
514 D… 1 1 3 3 2 1 1
… Ex = Signal conditioning instrument for meas. systems
inputs
Sensor Correct. mA V DIS
signal
in hazardous areas, certificate to CENELEC
VEGAMET for sensors with digital measured data transmission (VBUS)
- hydrostatic pressure transmitter
- ultrasonic sensor
- radar sensor
VEGAMET Number of Number of outputs
514 DV 1 1 3 3 2 1 1
inputs
Sensor Correct. mA V DIS
signal
VEGAMET 514… D 5

1.3 Applications

Product description
Flow measurement
Standard
Application: Flow Sensor: Hydrostatic pressure transmitter
Kind: Level Option: Offset correction (only with hydr. pressure transmitter) Parameter: MST1 level
Option
A key connected to the correction input corrects the output results when being set to zero.
Offset correction
- in empty flume
- preferably manually with key
- parallel shifting of the adjustment characteristics
- effects MST1 and MST2
Ultrasonic sensors Radar sensors
MST2 flow
actual measured value in %
A
D
Fig. 1.13 Correction in the point
100 %
0
0
corrected measured value e.g. 0 % (m h)
A
Fig. 1.2 Flow measurement
AA
incorrect course of measured values
measured value too low
measured value too high
100 %
3
/
6 VEGAMET 514… D
Product description

1.4 Technical data

Power supply
Operating voltage U
Power consumption
- VEGAMET 514 D appr. 7 VA or appr. 5 W
- VEGAMET 514 VD appr. 17 VA or appr. 12 W Fuse 630 mA slow-blow
Measuring data input VEGAMET 514 D
Number 1 input Kind of input active two-wire input, analog Range 4 … 20 mA
Voltage at 4 mA appr. 18 V DC, at 20 mA appr. 15 V DC Current limitation at appr. 26 mA, shortcircuit proof Detection-line interruption < 2 mA Detection-line shortcircuit > 23 mA Min. adjustment delta 2 % of the adjusted sensor values Connection line 2-wire (standard line) Resistance per conductor max. 35 Resolution 1 µA Linearity error 0,025 % at 4 … 20 mA Temperature error 0,04 %/10 K at 4 … 20 mA
Measuring data input VEGAMET 514 VD
Number 1 input Data transmission digital (VBUS) Sensor hydrostatic pressure transmitters,
Voltage appr. 25 V DC Power limitation at appr. 7 W, shortcircuit proof Connection line 2-wire (screened) Line length max. 1000 m Resistance per conductor max. 200 for hydrostatic pressure transmitters
Correction signal input
Number 1 input Function switching signal to trigger and offset correction Voltage 5 V (from instrument) Current 5 mA External line resistor 150
= 24 V AC (20 … 53 V), 50/60 Hz
nenn
= 24 V DC (20 … 72 V)
hydrostatic pressure transmitters
ultrasonic or radar sensor
max. 20 for ultrasonic or radar sensors
>>
VEGAMET 514… D 7
Product description
Current output
Number 3 outputs Function current output 2 and 3 can be configured and
Output values
- level / flow proportional
Load max. 500 Range 0/4 … 20 mA Resolution 0,05 % of range Linearity error 0,05 % of range Temperature error 0,05 %/10 k of range
- pulse output
Load min. 600 Voltage pulse 12 V DC at 20 mA Pulse length 200 ms
Voltage output
Number 3 outputs Function analog output of the output results Range 0/2 … 10 V Current max. 1 mA Resolution 0,05 % of range Linearity error 0,05 % of range Temperature error 0,06 %/10 k of range
Relay outputs
Number 2 switching relays
Function can be configured and parameter adjusted as:
Contact 1 floating spdt
Turn-on voltage min. 10 mV DC Switching current min. 10 µA
Breaking capacity max. 500 VA, 54 W Min. switching hysteresis (Low/High-Delta) 0,5 %
parameter adjusted as:
- level proportional
- flow proportional
- flow pulses
- sampling pulse
1 fail dafe relay
- switching relay for level or flow
- pulse relay-flow quantity
- pulse relay-sampling
AgNiO and hard gold plated
max. 250 V AC/DC
max. 3 A AC, 1 A DC
DISBUS-output
Function digital transmission of output results and
Connection line 2-wire Line length max. 1000 m
system information to VEGADIS 174 or VEGACOM 557
>>
8 VEGAMET 514… D
Product description
Indicating element
Clear text indication LC-display
Analog indication 11 segments 0 % … 100 %
LED in front plate green on: operating voltage on
Operating elements
Front plate 6 keys for configuration and parameter adjustment
Ambient conditions
Permissible ambient temperature -20°C … +60°C Storage and transport temperature -40°C … +80°C
Electrical connection
Multipoint connector component acc. to DIN 41 612, series F, 36-pole (three row) d, b, z
Carrier BGT 596 (Ex) connection to the respective module Housing type 505 connection to screw terminals (max. 2,5 mm
Electrical protective measures
Protection
- unassembled IP 00
- mounted into carrier BGT 596
- mounted into housing
- terminal housing IP 20
- housing general IP 30 Protection II (in housing type 505 or type 506) Overvoltage category II
Electrical separating measures
Reliable separation acc. to VDE 0106, part 1 between power supply, fail safe and level relay,
- limit voltage 250 V
- isolation resistance 2,3 kV
- 4 line, 6 figures each
- background lightning
acc. to MST-enquiry
- %-value of level
- %-value of flow
red: fault signal yellow (per relay): relay switching condition
(partly equipped)
front side completely equipped: IP 40 upper and lower side: IP 20 wiring side: IP 00
meas. data inputs and correction inputs
2
)
Galvanic separation between relay outputs
- limit voltage 250 V
- isolation resistance 1,4 kV
Potential separation between DISBUS, transistor outputs and correction inputs
- limit voltage 50 V
- isolation resistance 0,5 kV Common reference potential at voltage/current outputs, correction inputs
>>
VEGAMET 514… D 9
Product description
Mechanical data
Series module instrument for carrier BGT 596
or housing type 505 or type 506
Dimensions unassembled W = 25, 4 mm (5 TE), H = 128,4 mm, D = 162 mm
CE-approval, conformity judgement
VEGAMET signal conditioning instruments meets the protective regulations of EMVG (89/336/EWG) and NSR (73/23/EWG). The conformity has been judged by means of a typical configuration acc. to the following standards: EMVG Emission EN 50 081 - 2: 1993
Immission EN 50 082 - 2: 1995
NSR EN 61 010 - 1: 1993

1.5 Dimensions

14,5
Circuit board 100 x 160 x 1,5 European size Multiple plug
9
3
162
5 TE
%
100
+
-
100
ESC
Ser.No.12345678
12
128,4
OK
CONNECT
on
514 D
25
>>
10 VEGAMET 514… D
Product description

1.6 Ex-technical data

Power supply
Operating voltage U Limit voltage U
corresponds to the not-Ex-version
nenn
= 250 V AC or 125 V DC
nenn
Measuring data input (intrinsically safe circuit)
Classification [EEx ia] IIC, [EEx ia] IIB, [EEx ib] IIC or [EEx ib] IIB Max. values
- Voltage U
- Current I
- Power P Characteristics linear Effective inner inductance L Effective inner capacitance C
i
i
= 20 V
O
= 128 mA
O
= 640 mW
O
unimportant unimportant
EEx ia IIC EEx ia IIB EEx ib IIC EEx ib IIB
Max. permissible outer inductance LO (mH) 0,5 1 1,5 2 2 9
Max. permissible out capacitance CO (nF) 97 78 68 486 200 1000
The intrinsically safe circuits are reliably separated from the not-intrinsically safe circuits up to a peak value of the nominal voltage of 375 V. The max. voltage on the not-intrinsically safe circuits must not exceed 250 V
in case of failure.
eff

1.7 Approvals

VEGAMET are available with the following approvals:
- Explosions protection: appropriate operating instrument with intrinsically safe(s)
- Overfill protection to WHG and VbF applied
Observe with these applications the appropriate legal documents (test certificates, type approvals and conformity certificates). These are supplied with the respective instrument.
WHG-approval
Signal conditioning instrument as part of an overfill protection to WHG or VbF (applied).
Ex-approval
For hazardous areas, certificate to CENELEC defined in the conformity certificate PTB-no. Ex-95.D.2161 X.
VEGAMET 514… D 11

2 Mounting

Mounting

2.1 Mounting versions

VEGAMET 514 D … 514 VD signal conditioning instruments can be either mounted with a module in a carrier BGT 596 or BGT 596 Ex.M or in a single housing type 505 or type 506.
Module
Multipoint connector DIN 41 612, series F, 33-pole (d, b, z) with coded pins and mounting material for mounting in carrier BGT 596 or BGT 596 Ex.M.
Module Ex
Multipoint connector DIN 41 612, series F, 33-pole (d, b, z) with coded pins, Ex-separating chamber and mounting material for mounting in carrier BGT 596 Ex.M.
Single housing
Plastic housing type 505 or type 506 for single mounting of signal conditioning instruments with a width of 5 TE (25,4 mm).
Mounting carrier
Mount the respective module (standard or Ex­version) in the carrier. Wire the connections of the multipoint connector according to the wiring plan on page 14.
The multipoint connector is available as follows:
- Wire-Wrap standard connection 1,0 x 1,0 mm
- plug connection 2,8 x 0,8 mm
- Termi-Point Standard connection 1,6 x 0,8 mm
- soldering connection
- screw terminals 2 x 0,5 mm
For further information concerning mounting see operating instruction of the carrier.
Mounting single housing
You can either screw the housing socket directly to the mounting plate or plug on a carrier rail (TS 35 x 7,5 acc. to EN 50 022 or TS 32 acc. toEN 50 035). Connect the terminals according to the wiring plans on the following pages. For further information to the mounting see operating instruction of the housing.
2
Coding
To avoid interchanging of the various signal conditioning instruments, the multipoint connector of the carrier or the housing is provided with pins and the multiple plug of the signal conditioning instrument with holes (mechanical coding).
An Ex-coding with fixed coded pin ensures that not­Ex and Ex-instruments are not interchanged.
An instrument coding ensures that the various signal conditioning instruments are not interchanged. The respective coded pins are attached to each module or housing. Provide the multipoint connector with these coded pins acc. to the following schedule and figure.
Instrument Ex-
coding coding
VEGAMET 514 D a5 / c5
VEGAMET 514 D Ex a5 / c5 c23
VEGAMET 514 VD a7 / c5
Fig. 2.1 Coding schedule
Instrument coding Function coding
514 D
514 VD
z b d
a c
o 1 o o 3 o
a5 a7
o 5 o o 7 o o 9 o o11o o13o o15o o17o o19o o21o o23o o25o o27o o29o o31o
c5
c23
VEGAMET
Ex-coding
Transparent cover
To protect the housing against unauthorized operation, the front plate of VEGAMET can be provided with a lockable transparent cover.
12 VEGAMET 514… D
Fig. 2.2 Module coding, VEGAMET
Mounting
Ex-version
Note!
Signal conditioning instruments must be generally mounted outside hazardous areas or special Ex­protective measures must be taken.
Ex-separating chamber
Mounting in carriers
If you mount your VEGAMET with Ex-approval in a carrier, you have to use a VEGA-Ex-module. Keep a distance of at least 10 mm (2 TE) to modules of other manufacturers. If you want to mount the VEGAMET in the extreme left position in the carrier, you have to mount a blind cover of at least 20 mm (4 TE) in front of the module of the instrument.
To ensure sufficient "air and creeping distances", an Ex-separating chamber must be mounted to the connections of VEGAMET. Lead the wires through the Ex-separating chamber and connect them. Fasten the Ex-separating chamber with the lower screw. Observe the operating instrumention of carrier BGT 596 Ex.M.
Protection for Ex-applications
In Ex-applications a protection of IP 20 must be kept. Cover the spaces or not used modules from the front by appropriate blind covers.
%
100
+
-
OK
ESC
CONNECT
on
VEGAMET
515 N
Blind cover

2.2 Instrument address

Adjust the instrument address on the appropriate switch on the circuit board. Address range 1 ...15. Observe that no address is used twice. Factory setting = instrument address 0.
514 VD Switch
8
9
7
A
6
5
4
33
2
F
1
0
corresp.
B
C
D
E
514 D Ex Switch on bottom, visible through
an appropriate opening 10 11 12 13 14 15
Under the menu point "Instrument address" ´the adjustment carried out on the switch can be indicated.
VEGAMET 514… D 13

3 Electrical connection

VEGAMET 514 D, 514 D Ex VEGAMET 514 VD
d b z
+-
Supply voltage
L (+) N (-)
Current outputs 1 … 3
Voltage output 1 … 3
Correction signal input 4
+
-
Sensor
2
6
10 12
1 2 3
+
16
-
18
+
20
+
22
--
24
28
+
30
-
32
+
Fail safe relay
Relay output 1
Relay output 2
DISBUS-output
Electrical connection
Ex-version
For connection of Ex-certified instruments observe the instructions in the attached legal documents as well as in the valid mounting instructions. Note that the Ex-separating chamber is mounted on the multipoint connector.Always lead the wires through the Ex-separating chamber. Observe the operating instruction of the carrier BGT 596 Ex.M and the Ex­instructions.
14 VEGAMET 514… D
Set-up

4 Set-up

4.1 Indicating and operating elements

Signal conditioning instrument
with adjustment module
1
2
%
3
5
7 10 11 12 13 14
1
100
-
ESC
CONNECT
2
1
!
on
514
Ser.No.12345678
+
OK
4 6 8
9
16
18
1 Lockable fixing screw 2 Display (4 lines with 6 figures each,
illuminated) 3 … 8 See following page 9 Analog indication 11-digit 10 Connection for VEGACONNECT
connection cable 11 LED relay output 1 and 2 12 LED fault signal
[17]
15
13 LED supply voltage 14 Ledge 15 Multiple plug with wiring plan 16 Position of the switch for adjustment
of the instrument address [17] on VEGAMET 514 D Ex
18 Transparent cover
Fig. 4.1
VEGAMET 514… D 15

4.2 Adjustment system

Set-up
Indicating and adjustment module
Branch, i.e. jump to the lower menu
[OK]
Acc. to the menu point delete the
adjustment or change to the
upper menu
Menu tree
Indication of measured values
Main
ESC
menu
Sub
OK
menu
MST1
36.9
Param. MST1
Ad­just­ment
MST2
%
Param. MST2
cess­ing
%
Indication of
- measured value
- menu points
- parameters
- values
Acc. to the parameter, change value or
%
100
+
-
ESC
OK
OutputPro-
lation
select from list
Select menu point
Acc. to the menu point store the adjusted value or change to the lower menu
Analog indication
Confi­gura­tion
InputsSimu-
Meas. point
Out­puts
add. functions
Pass­word off
Sca­ling
Lin. curve
The adjustment is menu oriented and is made via six keys in conjunction with clear text indication. The jump from indication of measured values to the main menu is made with [OK]. Use [→] or [←] to change within the main menu from one menu point to the other.
A branch is indicated by the symbol
and ensures with [OK] a jump to the lower menu. In each menu are thematically combined parameters (possibly in further sub menus).
Integra­tion time
Menu point
Parameter
10
value
Parameters are indicated by the missing symbol The value of the parameters can be modified or
selected out of a list by [+] or [–]. The modified value can be stored with [OK]. Push [ESC] to interrupt an adjustment (without storing the modfication).
Certain parameters can be only indicated, however their value cannot be modified.
Reset to the upper menu is made with [ESC]. 60 minutes after a key is pushed for the last time, an automatic reset to the indication of measured values.
16 VEGAMET 514… D
.
Set-up

4.3 Menu list

Values must be adjusted or modified to determine the meas. points. This determination is made separately for each meas. point by the following parameter adjustments.
Parameter adjustment MST1
–– Adjustment
–– with level –– without level
–– Processing
–– Scaling (level) –– Lin. curves –– Integration time
–– Outputs
–– Current outputs –– Volt outputs –– Relay outputs –– MET-indication –– PC- / PLC-outputs –– DIS-outputs
–– Simulation
–– Special functions
Adjustments and coordinations must be provided to determine the instrument functions. This determination is made by the following configurations.
Configuration
–– Configuration
Inputs, Sensor connections, mark
–– Input 1
–– Input 4
–– Configuration meas. points
–– Application –– Sensor –– Kind –– Options –– Sensor coordination
–– Instrument address
–– Configuration outputs
–– Current outputs –– Volt outputs –– Relay outputs –– PC- / PLC-output –– VEGADIS-output
Parameter adjustment MST2
–– Adjustment
–– with flow –– without flow
–– Processing
–– Scaling (flow) –– Lin. curves –– Integration time
–– Outputs
–– Current outputs –– Volt outputs –– Relay outputs –– Display –– PC / PLC –– DISBUS-output
–– Simulation
–– Special functions
VEGAMET 514… D 17
Under the section "Additional functions" all superset instrument functions, i.e. their adjustment and coordination possibilities are composed.
Additional functions
–– Password
–– Edit lin. curves
–– Info
–– Input info –– VEGAMET-Info –– Program info –– Meas. point info
–– Language
–– Reset VEGAMET
–– Reset configuration –– Reset sensor characteristic values –– Reset lin. curves
–– Service

4.4 Adjustment via PC

Signal conditioning instrument with operating module
Set-up
%
100
-
ESC
2
1
!
on
Ser.No.12345678
CONNECT
515
+
OK
- directly via the keys of the adjustment module (see following descriptions)
- or with a PC, provided with VVO-Software (VEGA Visual Operating), and a VEGACONNECT connection cable
VEGACONNECT connection cable
PC with VVO-Software
Set-up

4.5 Set-up sequence

Configuration
- adjust instrument address
- configure inputs
- prepare meas. points
- coordinate outputs
Parameter adjustment
If you want to detect the level and the flow as parameter, it is recommended, to carry out the adjustment for MST1 and MST2 at the same time i.e. immediately one after the other.
- Parameter adjustment MST1
- carry out adjustment
- determine the processing
- define outputs
- Parameter adjustment MST2 as described above
18 VEGAMET 514… D
Additional functions
- edit linearization curves
- change language (if necessary)
- activate password (if necessary)
Set-up, configuration

4.6 Set-up, configuration

4.6.1 Configuration inputs

The inputs of the signal conditioning instruments are configured according to the ordered instrument version. Therefore no activities from your side are necessary. However should there be a modification necessary, you will find some explanations and connection examples in the following.
The signal conditioning instruments are either connected directly to the sensors or they receive measured data from other sensors via the DISBUS­connection.
Connection example 1
VEGAMET 514 D
- Sensors connected directly to VEGAMET Marking: The sensor must be marked via its sensor characteristics values (only hydrostatic pressure transmitters).
A
Sensor characteris­tics values
D
S1
K1
Connection example 2
VEGAMET 514 VD
- Sensors connected directly to VEGAMET Marking: The sensor must be marked with its series number.
A
Sensor series number
Pos. A – input 1, series number 8888.8888 Pos. D – input 4, channel K1
D
K1
Pos. A – input 1, channel S1 Pos. D – input 4, channel K1
>>
VEGAMET 514… D 19
Set-up, configuration
Connection example 1
Configuration, inputs
–– Input 1 Sensor with analog transmission of measured data
–– Input of this MET
–– Channel no. S1
–– Sensor characteristics values The menu point "Sensor characteristics values" is
–– Min. meas. range 0.00 Meas. range and sensor values are stated in the –– Max. meas. range 1.00 test certificate of the hydrostatic pressure –– Min. sensor value 4.000 transmitter. –– Max. sensor value 20.000
–– Input 4 Correction signal (key)
–– Input of this MET –– Channel no. K1
Connection example 2
Configuration, inputs
–– Input 1 Sensor with digital transmission of measured data
–– Input of this MET –– Sensor adaption can be necessary for ultrasonic / radar sensors (see
–– Input 4 Correction signal (key)
–– Input of this MET –– Channel no. K1
only visible with hydrostatic pressure transmitters.
BA of the appropriate sensor)
20 VEGAMET 514… D
Set-up, configuration
A

4.6.2 Configuration meas. point

VEGAMET 514 D and 514 VD signal conditioning instruments in their basic configuration are designed for combi application, i.e.
- MST1, level
- MST2, flow measurement Supplementary the following modifications of the configuration can be carried out.
Configuration meas. point
–– Combi application If a general modification of the basic configuration
–– Application flow should be necessary, a "Reset to combi application" –– Sensor Hydrostatic should be carried out, see respective menu point on –– Kind Standard page 44. A new basic configuration can be then
programmed acc. to the product description 1.3 (see page 6).
–– Options no
offset corr.
(for MST1 and MST2)
–– Sensor coordination
–– Position A Input 1 –– Position D Input 4
Input 1
Channel no. S1
Input 4
Channel no. K1
–– MST1 level
–– MST name MST Providing of a special MST
name MST name e.g. name
level
MST1
–– Fault signal on on means that a failure of MST1 is transmitted
off to the failure processing off means no
transmission
–– Tare see page 21
–– Monitoring see page 21
–– MST2 flow
–– MST name MST Providing of a special MST
name MST name, e.g. name
flow
MST2
–– Fault signal as above, however MST2
–– Monitoring see page 21
VEGAMET 514… D 21
Tare
By means of the tare function an additional measured value exists from the actual value beginning at 0%. The tare function can be triggered via a key connected to correction signal input 4.
Each current, voltage or relay output as well as the display can be afterwards coordinated to the tared measured value in the menu point "relating to". With the allocation or allocations the tared measured value is activated.
–– MST1
–– Tare
Measured values in %
100 %
Key
40
Set-up, configuration
tared course
0
0 40 100 %
actual measured values
tared measured values
actual course
–– Input not defined
Input 4 Determination of the correction signal input (key)
Monitoring
For the activation of the monitoring the following pre­requirements must be met:
- Determine in the menu range "Configuration meas. point" input 4
- Connect the key switch (opener) to input 4 (see page 13).
Meas.value
50 %
High 7 %
Low –5 %
>57 %
<45%
- Select in the menu range "Parameter adjustment relay outputs" the monitoring mode for the concerned relay (see page 37 and 40).
Relay energized
Relay de­energized
Relay de­energized
The selected relay is energized.
By pushing the key switch (opener opens) the respectively actual measured value is frozen and derived from this, in case the value is decreased (low) or exceeded (high) the relay energizes.
After a reset and pushing the key switch again the actual measured value is frozen again etc.
–– MST1
–– Monitoring
–– Input not defined
Input 4 with input 4 determination of the correction signal
input (key switch)
–– MST2 as described above
22 VEGAMET 514… D
Set-up, configuration

4.6.3 Configuration output

Dependent on the basic configuration the outputs are automatically coordinated to the measuring points MST1 and MST2. This coordination can be modified separately for each output. In a list either the measuring point numbers MST1 and MST2 or the previously adjusted measuring point names are provided as choice. In addition each output can be switched off (--), i.e. the switched off output does not access to any measuring point.
Configuration, outputs
–– Configuration current outputs
–– Current 1 to MST1 ––– 0 … 20 mA level proportional
MST2
--
–– Current 2 to MST2 ––– 0 … 20 mA flow proportional
-­MST1
–– Kind 2 Standard ––
Pulses
–– Current 3 to MST2 ––– Flow pulses
-­MST1
–– Kind 3 Standard
Pulsed–
–– Configuration volt outputs
–– Volt 1 to MST1 ––– 0 … 10 V level proportional
MST2
--
–– Volt 2 to MST2 ––– 0 … 10 V flow proportional
-­MST1
–– Volt 3 to -- ––– not coordinated to a MST
MST1 MST2
Coordination acc. to basic configuration:
>>
VEGAMET 514… D 23
Configuration, outputs
–– Configuration relay outputs
–– Operating relay
–– Relay 1 to MST1 acc. to basic configuration coordinated to level
–– Relay 1 kind Standard
Set-up, configuration
Hold function
Parameter
High
Low
Taster
Relay outp.
t
d b z
Reset of alarm funct.
Parameter
High
Low
Taster
Relay outp.
t
d b z
The relay modes "Hold function" and "Reset of alarm functions" are configured as described above and activated with one external signal each (correction signal input).
–– Input not defined
Input 4 Determination of the correction signal input
(key, see above)
–– Relay 2 to MST2 acc. to basic configuration coordinated to flow
–– Relay 2 kind Standard
Pulses see Parameter adjustment page 41
–– Fail safe relay kind Standard The fail safe relay can be operated like the
Hold funct. operating relay, i.e. as described above. Reset of alarm funct.
–– Input not defined
Input 4 as scribed above
>>
24 VEGAMET 514… D
Set-up, configuration
Configuration, outputs
–– Configuration PC/PLC Dependent on the basic configuration outputs
–– PC-/PLC-measured values
–– PLC 1 to MST1 –– PLC 2 to MST2 This coordination can be modified separately –– PLC 3 to -- for each output. In a list either the measuring –– PLC 4 to -- point numbers MST1 and MST2 or the –– PLC 5 to -- previously adjusted measuring point names –– PLC 6 to -- are available as a choice. –– PLC 7 to --
are automatically coordinated to measuring points MST1 and MST2.
Example acc. to configuration measuring point
MST name level
MST name flow
–– PC/PLC relay status off In addition each output can be switched off
–– PC/PLC Inp. status off
–– Configuration VEGADIS
–– PC-/PLC-measured values
–– DIS 1 to MST1 Configuration as abobe –– DIS 2 to MST2 –– DIS 3 to -­–– DIS 4 to -­–– DIS 5 to -­–– DIS 6 to -­–– DIS 7 to --
Parameter adjustment of outputs see page 41
on (– –), i.e. the switched off output does not access to
any measuring point.
on
VEGAMET 514… D 25
Set-up, parameter adjustment

4.7 Set-up, parameter adjustment

The menu described in the following is available for parameter adjustment. This menu has the same function for VEGAMET 514… for each measuring point number, i.e. parameter adjustment MST1 and MST2.

4.7.1 Parameter adjustment MST1, adjustment

The menu range "Parameter adjustment MST1, adjustment" adapts automatically to the measuring point configuration, i.e. acc. to the "Application" and "Sensor" different menu points are displayed.
Adjustment
- Application Flow measurement
- Sensor Hydrostatic pressure transmitters
Parameter adjustment MST1, adjustment
e.g. rectangular weir
= 0,60 m
h
max.
100 % ^ 0,06 m (h)
0 % ^ 0,00 m (h)
–– with measured product With this adjustment procedure %-values each must
be adjusted for min. and max. which correspond to the actually measured levels/flow etc.
–– Min. adjustment at 0.0 % Range 0 … 80 % –– Max. adjustment at 100.0 % Range 20 … 100 %
a of 20 % must be observed between min. and max.
Please observe that for all these applications with hydrostatic pressure transmitters and adjustment without level/flow first of all a position correction must be carried out.
–– without measured product For this adjustment procedure two future levels/flows
must be adjusted which correspond to 0% and 100%.
–– Adjustment in m (h) etc. modify unit if necessary
–– Position correction The menu points "Position correction" etc. are
released when connecting hydrostatic pressure transmitters.
–– Sensor unpressurized? OK? With the position correction the measured data of
the unpressurized sensor (flume empty) are detected
–– Position correction now? OK? as correction size and is considered for all following
measurements
–– 0 % at 0.00 m (h) the unit is displayed acc. to the menu point
"Adjustment in"
–– 100 % at 0.60 m (h) 0.00 and 0,60 correponds e.g. to an adjustment in m
Parameter adjustment MST2, adjustment as described above.
If you want to detect the level and the flow as parameter, it is recommended to carry out the adjustment for MST1 and MST2 at the same time, i.e. immediately one of the other.
>>
26 VEGAMET 514… D
Set-up, parameter adjustment
Adjustment
- Application Flow measurement
- Sensor ultrasonic sensors
e.g. rectangular weir h
= 0,60 m
max.
radar sensors
100 %
^ 0,80 m (d)
^ 1,40 m (d)
0 %
Parameter adjustment MST1, adjustment
–– with measured product With this adjustment procedure %-values have to be
adjusted each for min. and max. which correspond to the actual measured distances between sensor and level/flow.
–– Min. adjustment at 0.0 % Range 0 … 80 % –– Max. adjustment at 100.0 % Range 20 … 100 %
a of 20 % must be observed between min. and max.
–– without measured product For this adjustment procedure two future distances
(sensor … level/flow) must be adjusted which correspond to 0 % and 100 %.
–– Adjustment in m (d) etc. if necessary modify unit, the unit is displayed
acc. to the menu point "Adjustment in"
–– 0 % at 0.80 m(d) 0.80 and 1.40 –– 100 % at 1.40 m(d) corresponds e.g. to adjustment distances in m
Parameter adjustment MST2, adjustment as described above
If you want to detect the level and the flow as parameter, it is recommended to carry out the adjustment for MST1 and MST2 at the same time, i.e. immediately one of the other.
VEGAMET 514… D 27
Set-up, parameter adjustment

4.7.2 Flow measurement

General information
In the following section elementary information to dimensions in front of flumes, walls and sensor installation positions is given.
Flume with meas. section (e.g. Khafagi-Venturi flume)
Top view
3 - 4 x h
max
Sensor
90°
Sensor
min. distance of the sensor
h
max
Inlet side
Lateral view
Outlet side
Fth meas. wall (e.g. rectangular weir)
k 3 cm
Sensor
90°
Sensor
min. distance of the sensor
h
max
2 x h
max
3 - 4 x h
90°
Upstream water Downstream water
max
45°
k 3 cm
5 cm
Note
General
is the max. damming or overflow height acc. to the meas. section or meas. wall.
-h
max
- All other dimensions are nearly derived from h
(see above drawings)
max
Sensor mounting position
- Mount sensor on the inlet /upstream water side.
- Observe distance 3 - 4 x h
- Mount the ultrasonic and radar sensors centred to the flume and vertically to the flow medium.
from sensor to the throat or meas. wall.
max
- Consider with ultrasonic sensors the min. distance to hmax.
Meas. wall
- Dimension dependent on h
- Thickness on the meas. wall (horizontal part) k 3 mm.
.
max
- Swelling of the meas. wall with a transition of 45° (see above drawing).
- Distance from the opening of the meas. wall to downstream water 5 cm.
>>
28 VEGAMET 514… D
Set-up, parameter adjustment
Max. flow, Q
In the term Q
1000) the following flume specific factors are
max
(menu point 100 % corresponds to
max
considered relating to the adjustment at 100 %
- geometry of the flume
- material of the flume
- flow speed of the flow medium.
The respectively valid values for Q determined by the manufacturer of the flume.
If the value for Q this value can taken as a proximity out of the
of an existing flume is not known
max
are normally
max
following schedule or can be calculated.
In addition it is recommended to determine and observe the frame conditions defined in the literature to the flow measurement (surface structure of the flume, flow speed etc.).
Some of the usual flumes are stated in the following in form of schedules and calculation examples.
Explanations Q
= maximum flow in m3/h
max
Q
= minimum flow in m3/h
min
h
= maximum damming or overflow height in mm
max
or m
h
= minimum damming or overflow height in mm
min
or m B = flume or overflow width in mm or m K = flume or overflow specific factor k = thickness of the meas. wall in mm D = tube diameter in inch
Calculation information
- Liter/sec x 3,6 = m
1
3
-m
/h x ––– = liter/sec
3,6
- Example 10 liter/sec
3
/h
^ 36 m3/h
Khafagi-Venturi flume
With rectangular cross section and even canal bottom
B
h
max
Values for Q
BK Q in mm (Factor) in m3/h in mm in m3/h in mm
max
maxhmax
Q
minhmin
45 605 80 260 4,8 40 120 838 124 280 9,3 50 160 1106 210 330 16,2 60 200 1383 363 410 25,6 70 240 1654 532 470 37,4 80 320 2178 864 540 58,8 90 640 4280 3776 920 383 200
Calculation of Q
max
The following formula can be used for calculation of the actual flow of liquid levels < h
max…hmin
.
Presumed values for a calculation example B = 0,16 m K = 1106 (derived from B) h = 0,25 m
Q
max
1,5
=K x h = 1106 x 0,25
1,5
= 138,2 m3/h
>>
VEGAMET 514… D 29
Set-up, parameter adjustment
Trapezoidal weir (Cipolletti)
With sharp overflow edge
1
/4 h
Values for Q
BK Q in mm (Factor) in m3/h in mm in m3/h in mm
B
max
h
max
max
Q
maxhmax
h
min
min
300 2011 116,8 150 29,5 60 450 3014 354,4 240 44,3 60 600 4011 760,4 330 58,9 60 800 5353 1457 420 78,7 60 1000 6688 2581 530 93,3 60 1500 10043 6523 750 147,6 60 2000 13381 14397 1050 196,7 60 3000 20082 36893 1500 295,1 60
Calculation of Q
max
The following formula can be used for calculation of the actual flow of liquid levels < h
max…hmin
.
Presumed values for a calculation example B = 0,6 m K = 4011 (derived from B) h = 0,27 m
Q
max
1,5
=K x h = 4011 x 0,27
1,5
= 562,73 m3/h
Rectangular overflow without contraction
With sharp overflow edge without throat
B
h
max
K = 6593
Values for Q
BQ in mm in m3/h in mm in m3/h in mm
max
h
Q
max
max
h
min
min
200 54,81 120 19,4 60 400 310 240 38,7 60 600 650 300 58,1 60 800 1592 450 77,5 60 1000 3064 600 96,9 60 1500 6423 750 145,3 60
Calculation of Q
max
The following formula can be used for calculation of the actual flow of liquid levels < h
max…hmin
.
Presumed values for a calculation example B = 1,5 m K = 6593 h = 0,3 m
= B x K x h
Q
max
= 1,5 x 6593 x 0,3
1,5
1,5
= 1625 m3/h
>>
30 VEGAMET 514… D
Set-up, parameter adjustment
Rectangular overflow without contraction
With sharp overflow edge and throat
B
h
max
K = 6290
Values for Q
BQ in mm in m3/h in mm in m3/h in mm
max
h
Q
max
max
h
min
min
200 52,3 120 18,5 60 300 109,6 150 27,7 60 400 295,8 240 37,0 60 500 441,2 270 46,2 60 600 620 300 55,5 60 800 1519 450 73,9 60 1000 2923 600 92,4 60 1500 6128 750 138,7 60 2000 13535 1050 184,9 60 3000 34666 1500 277,3 60
Calculation of Q
max
The following formula can be used for calculation of the actual flow of liquid levels < h
max…hmin
.
Presumed values for a calculation example B = 0,4 m K = 6290 h = 0,13 m
= B x K x h
Q
max
= 0,4 x 6290 x 0,13
1,5
1,5
= 117,9 m3/h
Rectangular flume (V-Notch)
With sharp overflow edge
α
h
max
Values for Q α KQ
in ° (Factor) in m
max
maxhmax
3
/h in mm in m3/h in mm
Q
minhmin
90 4966 1385,0 600 4,4 60 60 2869 800,0 600 2,5 60 45 2056 573,3 600 1,8 60 30 1334 372,0 600 1,2 60 22,5 989 275,8 600 0,87 60
Calculation of Q
max
The following formula can be used for calculation of the actual flow of liquid levels < h
max…hmin
.
Presumed values for a calculation example α = 45° K = 2056 (derived from α) h = 0,33 m
max
= 2056 x 0,33
2,5
2,5
=K x h
Q
= 128,6 m3/h
>>
VEGAMET 514… D 31
Palmer-Bowlus-Flume
Set-up, parameter adjustment
D
D
D
/
4
30°
Values for Q
DK Q in Zoll (Factor) in m3/h in mm in m3/h in mm
D
/
/
2
4
30°
h
max
D
/
6
max
maxhmax
D
/
2
Q
h
min
min
6" 1952 37,82 120 2,9 30 8" 2338 68,60 150 7,3 45 10" 2712 148,8 210 12,3 55 12" 3046 214,3 240 18,9 65 15" 3519 374,8 300 32,1 80 18" 3943 501,4 330 54,4 100 24" 4745 1074,0 450 114,5 135 30" 5488 2122,0 600 192,3 165
Note: The data of above schedule relate to Palmer-Bowlus­Flumes of Messrs. Plasti-Fab.
Calculation of Q
max
The following formula can be used for calculation of the actual flow of liquid levels < h
max…hmin
.
Presumed values for a calculation example D = 12" K = 3046 (derived from D 12") h = 0,2 m
max
= 3046 x 0,2
1,86
1,86
=K x h
Q
= 152,63 m3/h
32 VEGAMET 514… D
Set-up, parameter adjustment

4.7.3 Parameter adjustment MST1 and MST2, processing

Parameter adjustment MST1, processing (level)
–– Scaling
–– 0 % corresponds to 0.0 both values can be additionally influenced –– 100 % corresponds to 100.0 with the determination of the decimal point
–– Decimal point 888.8
–– relating to Percent etc. choice out of list (scaled)
–– Unit % etc. choice adapted to above determination
.8888
–– Linearization curve linear
After a scaling or linearization the requested output must be coordinated to the scaled or linearized value in the menu range "Parameter adjustment outputs", see menu points "relating to" and "Unit".
–– Integration time 0 s Range 0 … 600 sec.
Lin.curve 1 enquiry of three curves to be edited Lin.curve 2 Editing see page 42 Lin.curve 3
>>
VEGAMET 514… D 33
Set-up, parameter adjustment
Parameter adjustment MST2, processing (flow)
With the scaling of MST2 the conditions for the signal conditioning instrument is provided to calculate the
Example relating to a rectangular flume (see page
31):
- Max. flow 1385 m
- Min. flow 4,4 m
3
/h
3
/h flow volume/time. The respective values for min. and max. flow are generally stated by the flume manufacturers or can be calculated as proximity
100 %
values (see page 29 … 32).
In case of calculation the value of min. flow is unimportant.
Parameter adjustment MST2, processing (flow)
–– Scaling
–– 0 % corresponds to 0 Range 0 … (4) … 9999 (Min. flow) –– 100 % corresponds to 1000 Range 0 … (1385) … 9999 (Max. flow)
This value (1385) is automatically transferred to the menu point "Pulse parameter volume", see page 36 and 41.
–– Decimal point 888.8
.8888
–– relating to flow
–– Unit L/s etc. choice adapted to flow measurement
–– Linearization curve linear Given curves for flumes and
Root funct. series Venturi Palmer-Bowlus Quadrant V-Notch Trapezoidal weir Rectangular overflow Lin.curve 1 Enquiry of three curves to be edited Lin.curve 2 Editing see page 43 Lin.curve 3
^ 1385 m3/h
4,4 m
3
/h
After a scaling or linearization the requested output must be coordinated to the scaled or linearized value in the menu range "Parameter adjustment outputs", see menu points "relating to" and "Unit".
–– Integration time 0 s Range 0 … 600 sec.
34 VEGAMET 514… D
Set-up, parameter adjustment

4.7.4 Parameter adjustment MST1 and MST2, outputs

Parameter adjustment MST1 and MST2, outputs
–– Current outputs
–– Current output 1 acc. to basic configuration coordinated to MST1, i.e. 0 … 20 mA
relating to level
–– relating to Percent etc. choice acc. to list –– Unit % etc. choice adapted to above determination
–– Current output 4/20 mA
0/20 mA 20/4 mA 20/0 mA
-free- activated the menu points "Current at 0% and 100%"
–– Current at 0 % 4 mA adjustable range 0.0 … 20.0 mA –– Current at 100 % 20 mA adjustable range 0.0 … 20.0 mA
reaction of the current output in case of failure
–– at failure 0 mA 0 mA
22 mA 22 mA
-- actual current value is kept 0 % current value relating to 0 % 100 % current value relating to 100 %
–– Current output 2 acc. to basic configuration coordinated to MST2, i.e. 0 … 20 mA
relating to flow, otherwise like current output 1. The output can be also configured for pulses, see menu range of current output 3
–– Current output 3 acc. to basic configuration coordianted as pulse output MST2
–– Kind Pulse output
Parameter, flow
t
20 mA
0 mA
Sampling output
Parameter, flow
t
20 mA
0 mA
>>
VEGAMET 514… D 35
Parameter adjustment MST1 and MST2, outputs
Current output 3 (continuation)
–– Pulse parameter
–– Volume 1000 Range 0 … 9999 –– Quantity x factor 1 Multiplication factor x1, x10, x100, x1000
–– Creeping quantity
suppression 0 % Range 0 % … 100 %
–– Sampling h 1 Range 1 … 100 h (hours)
This menu point is only displayed in mode "Sampling output".
–– Volt outputs
Set-up, parameter adjustment
–– Volt output 1 acc. to basic configuration coordinated to MST1, i.e. 0 … 10 V relating
–– relating to Percent etc. choice acc. to list
–– Current output 2/10 V
–– Volt at 0 % 2 V adjustable range 0.0 … 10.0 V –– Volt at 100 % 10 V adjustable range 0.0 … 10.0 V
–– at failure 0 V 0 V
–– Volt output 2 as volt output 1, however acc. to basic configuration
–– Volt output 3 as basic configuration coordinated to no MST (– –).
to level
0/10 V 10/2 V 10/0 V
-free- activates the menu points "Current at 0% and 100%"
Reaction of the volt output in case of failure
11 V 11 V
-- Actual value is kept 0 % Volt value relating to 0 % 100 % Volt value relating to 100 %
coordinated to MST2, i.e. 0 … 10 V relating to flow. Volt output 2 can be configured to level like volt output 1.
36 VEGAMET 514… D
Set-up, parameter adjustment
Parameter adjustment MST1, outputs
–– Relay outputs
–– Relay output 1 acc. to basic configuration coordinated to MST1, i.e. level
–– relating to Percent etc. choice acc. to list
–– Unit % etc. choice adapted to above determination
–– Kind Overfill protection
–– Low 0.0 % In these menu points the switch points can be –– High 100.0 % determined for above relay modes (except
–– at failure off - Relay de-energizes
–– additional functions
–– Switching delay
–– t on 0 s Switch on delay –– t off 0 s Switch off delay
–– Relay output 2 see page 41
Prot. against dry running of pumps Switching window Off Switching window Off see additionally page 38 … 40 Tendency rising Tendency falling Monitoring
"Tendency rising" and "Tendency falling").
Range and unit acc. to selection in menu point "relating to" and "Unit", see above
Reaction of the relay output in case of failure
-- - Actual switching condition is kept
>>
VEGAMET 514… D 37
–– Relay output 1 (continuation 1)
–– Kind Over fill protection
The relay of relay output 1 de-energizes at switch point "High" (reliable switching condition)
Set-up, parameter adjustment
Parameter
High
Low
Relay outp.
d b z
t
Protection against dry running of pumps
The relay of relay output 1 de-energizes at switch point "Low" (reliable switch point)
Switching window On
The relay of relay output 1 is energized within the switching window
Switching window Off
The relay of relay output 1 is de­energized within the switching window
Parameter
High
Low
Relay outp.
Parameter
High
Low
Relay outp.
Parameter
High
Low
Relay outp.
d b z
t
d b z
t
d b z
t
>>
38 VEGAMET 514… D
Set-up, parameter adjustment
Tendency determination
The level change within the scanning time (ta) is determined and after all scannings (n) an average value is generated out of the sum of the level changes.
If this average value exceeds a previously defined %-value, the tendency determination responds, i.e. an energized relay de-energizes.
With the enquiry of "Tendency rising" or "Tendency falling" the menu points "Deviation in %" and "Deviation per time" are released.
–– Relay output 1 (continuation 2)
–– Kind Tendency rising
%
t ges.
ta
The relay of relay output 1 de-energizes, when the increasing average value of "t ges." exceeds the previously defined %-value
Tendency falling
same function however with falling
2
1
Relay outp.
ta = Scanning time n = Number of scannings
%
t ges.
34
ta
n = 4
t
d b z
average value
t
2
1
Relay outp.
ta = Scanning time n = Number of scannings
34
n = 4
d b z
–– Deviation in % 2 The level change exceeds the adjusted %-value, the
tendency determination responds. 0 … 110 % Example 2 % (modification) per t ges.
–– Deviation per time
–– Scanning time in s60 Adjustment range 0 … 999 s
Example 60 s corresponds to 1 minute
–– Number of
scannings 4 Adjustment range 0 … 99 scannings
Example: t ges = ta • n = 60 • 4 = 240 s = 4 min, i.e. after every 4 mins the average value of the deviation is formed and the result is compared with the given deviation in %, here in the example = 2%
>>
VEGAMET 514… D 39
Set-up, parameter adjustment
Monitoring
The following requirements must be met for activation of the monitoring:
- Determine input 4 in the menu range "Configuration meas. point".
- Connect a key word (opener) to input 4 (see page 14).
- Select for the concerned relay the mode monitoring in the menu range "Parameter adjustment relay outputs"
(see page 37).
The selected relay is engergized. By pushing the key switch (opener opens) the respective actual measured value is first of all frozen and derived thereof the relay de-energizes if the value is decreased (low) or exceeded (high).
See additional diagram "Configuration meas. points" page 21.
–– Relay output 1 (continuation 3)
–– Kind Monitoring With the enquiry "Monitoring" the menu point
"Monitoring on" is released.
%
–– Monitoring on High
The relay of relay output 1 de-energizes when the %-value stated under High is exceeded.
Low
The relay of relay output 1 de-energizes when the %-value stated under Low is decreased.
High and Low
The relay of relay output 1 de-energizes when the %-value stated under High and Low is exceeded or decreased.
High
frozen meas.value
Key switch
Relay outp.
%
frozen meas.value
Low
Key switch
Relay outp.
%
High frozen
meas.value Low
Key switch
Relay outp.
t
d b z
t
d b z
t
d b z
–– Low -5 % Range and unit acc. to the selection in the menu –– High 7 % points "relating to" and "Unit", see page 34.
–5 % and 7 % are exemplary values, relating to above diagrams.
>>
40 VEGAMET 514… D
Set-up, parameter adjustment
Parameter adjustment MST2, outputs
–– Relay outputs
–– Relay output 2 acc. to basic configuration coordinated to MST2, i.e. flow
Relay output
Parameter, flow
–– Kind Pulse output
d b z
t
Relay output
Parameter, flow
Sampling output
–– Pulse parameter
–– Volume 0 Range 0 … 9999 –– Quantity x factor 1 Multiplication factor x1, x10, x100, x1000
–– Creeping quantity
suppression 0 % Range 0 % … 100 %
–– Sampling h 1 Range 1 … 100 h (hours)
This menu point is only displayed in the mode "Sampling output".
d b z
t
VEGAMET 514… D 41
Parameter adjustment MST1 and MST2, outputs
–– MET-indication
Set-up, parameter adjustment
–– relating to Percent
–– Unit %
–– PC/PLC-outputs
–– PC/PLC output 1
–– relating to Percent
–– Unit %
–– PC/PLC output 2 … 7 as described above
–– DIS-outputs
–– DIS-output 1
–– relating to Percent
–– Unit %
–– DIS-output 2 … 7 as described above
scaled etc. choice acc. to MST-enquiry
scaled etc. choice adapted to above determination
scaled etc. choice acc. to MST-enquiry
scaled etc. choice adapted to above determination
scaled etc. choice acc. to MST-enquiry
scaled etc. choice adapted to above determination

4.7.5 Parameter adjustment MST1 and MST2, simulation

Note
The whole indication flashes with activated simulation. The measured data delivered from the sensors are not transmitted for processing, therefore this menu point should be quit as soon as possible. Reset to the indication of measured values automatically after 60 minutes.
Parameter adjustment MST1, simulation
–– Simulation now! OK?
–– Simulation with [+] or [–]
42 VEGAMET 514… D
Set-up, parameter adjustment

4.7.6 Parameter adjustment MST1 and MST2, special functions

Parameter adjustment MST1 and MST2, special functions
–– Reset meas. point (e.g. level or flow) Reset of all parameter values of the meas. point
–– Reset OK? MST1 or MST2 with this reset to factory
–– Reset now! OK? setting.
–– Failure mode Standard no fault signal
–– Manual corrections
–– Offset correction Only visible with hydrostat. pressure transmitters.
–– Offset correction OK?
–– Correction now! OK?
<–10% Fault signal with decreasing and/or exceeding <110% of the adjusted measuring range >110% <-10a.>110% <-110a.>110%
Carry out correction only with unpressurized sensor.

4.8 Set-up, additional functions

4.8.1 Password

Additional functions
–– Password off each menu point can be enquired and modified
on the password can be enquire menu point with [+]
and [–]. With the adjustment of password the whole menu is released again. Password for VEGAMET 514 D … and 514 VD = -0514-

4.8.2 Linearization curve 1 … 3

A linearization curve is generated by a certain number of index markers and their value pairs. A value pair consists of a value for level percent (X %) and a value for volume percent ( Lin %). Up to max. 32 index markers can be added.
A linearization curve can be terminated after any individual number of index markers. The signal conditioning instrument terminates automatically the linearization curve with the values X = 100 % and Y = 100 %.
The input of a linearization curve is made in the menu point "Add index marker".
>>
VEGAMET 514… D 43
Additional functions
–– Edit linearization curve
–– Linearization curve 1
–– Lin. curve 1
Set-up, additional functions
–– Edit lin.curve 1 Number of index markers 1 … 32 with [] or []
–– Y … V % Edit the Y-value with [+] or [–], take over
–– Add index marker The new value pair is added automatically with [OK]
–– X …… at the correct position of the linearization curve.
Y …… The already available index markers are sorted
–– Delete index markers Enquiry of the value pairs with [] or []
–– X ……
Y … delete? With [OK] the value pair and the appropriate index
–– Delete marker is deleted out of the linearization curve.
now? The remaining index markers and their value pairs
X…
%
Y… V%
X%
12,0
YV%
5,0
X…… Y……
… lösch- en
with [OK].
again.
move up
–– Lin. curve 2 as described above –– Lin. curve 3

4.8.3 Info

All following menu points have display function.
Additional functions
–– Info
–– Input Info - for input configuration –– VEGAMET Info - for VEGAMET configuration –– Program Info - to program –– Meas. point Info - to the individual meas. points
detailled information

4.8.4 Language

Additional functions
–– Language
– Deutsch –– English Preadjustment and acc. to request –– Francais modification, without data loss –– Italiano –– Espanol
44 VEGAMET 514… D
Set-up, additional functions

4.8.5 VEGAMET Reset

Additional functions
–– VEGAMET Reset
–– Reset configuration
–– to default With reset to default the whole configuration
–– Reset OK? and the whole parameter adjustment
–– Reset now! OK? (if necessary all MST) are reset to factory setting
–– to combi application The combi application available at the moment
–– delete all MST? is deleted and in their position a preconfiguration
–– etc. to the new combi application to be adjusted is
–– Reset sensor values
–– Reset Input 1 The sensor data of the sensor connected to input 1
–– Reset OK? are deleted
–– Reset now! OK?
–– Reset linearization curves
–– Reset lin. curve 1 With this reset you delete the index markers
–– Reset OK? of linearization curve 1
–– Reset now! OK?
The basic configuration however remains, e.g. pressurized level measurement.
generated, see page 6.
–– Reset lin. curve 2 as above, however linearization curve 2 –– Reset lin. curve 3 as above, however linearization curve 3 –– Reset all curves as above, however all linearization curves

4.8.6 Service

Additional functions
–– Service
–– Current output 1...3 the actual currents and voltages are –– Volt output 1...3 indicated
VEGAMET 514… D 45

5 Diagnosis

Diagnosis

5.1 Detailled menu list

Configuration
–– Configuration, inputs
–– Input 1 (analog)
–– Input of –– Channel no.
–– Sensor characteristic values
or
–– Input 1 (VBUS)
–– Input von –– Series no.
–– Sensor adaption
–– Input 4
–– Input of –– Channel no.
–– Configuration meas. points
–– Application –– Sensor –– Kind –– Options –– Sensor coordination
–– MST1
–– MST-name, MST1 –– Fault signal –– Tare –– Monitoring
–– MST2, as MST1 (without tare)
–– Instrument address
(deleted up from Software-Version 6.0)
–– Configuration outputs
–– Configuration current outputs
–– Current 1 to (coordination) –– Current 2 to (coordination) –– Current 2 Standard (Kind) –– Current 3 to (coordination) –– Current 3 Standard (Kind)
–– Configuration volt outputs
–– Volt 1 to (coordination) –– Volt 2 to (coordination) –– Volt 3 to (coordination)
–– Configuration relay outputs
–– Operating relay
–– Rel. 1 to (coordination) –– Rel. 1 Standard (Kind) –– Rel. 2 to (coordination) –– Rel. 2 Standard (Kind)
–– Fail safe relay
–– Relay Standard
–– Configuration PC / PLC-outputs
–– PC / PLC-output
–– PLC 1 to (coordination)
bis
–– PLC 7 to (coordination)
–– PC / PLC-relay status –– PC / PLC-input status
–– Configuration VEGADIS
–– DIS 1 to (coordination)
bis
–– DIS 7 to (coordination)
46 VEGAMET 514… D
Diagnosis
Parameter adjustment
–– Adjustment
–– with level / flow
–– Min. adjustment –– Max. adjustment
–– without level / flow
–– Adjustment in –– Position correction –– 0 % at –– 100 % at
–– Processing
–– Scaling (flow)
–– 0 % corresp. (Min.-flow) –– 100 % corresp. (Max.-flow) –– Decimal –– relating to –– Unit
–– Lin. curves –– Integration time
–– Outputs
–– Current outputs
–– Current output 1
–– relating to –– Unit
–– Volt outputs
–– Volt output 1
–– relating to –– Unit
–– Relay outputs
–– Relay output 1
–– relating to –– Unit –– Kind –– Low –– High
Additional functions
–– Password
–– Edit lin. curves
–– Lin. curve 1 –– Lin. curve 2 –– Lin. curve 3
–– Info
–– Input info –– VEGAMET-Info –– Program info –– Meas. point info
–– Language
–– Reset VEGAMET
–– Reset configuration –– Reset sensor values –– Reset lin. curves
–– Service
–– MET-indication –– PC / PLC-outputs –– DIS-outputs –– Simulation
–– Special functions
–– Reset meas. point –– Failure mode –– Manual corrections
VEGAMET 514… D 47
Diagnosis

5.2 Maintenance

The instrument is maintenance free.

5.3 Repair

Due to safety and guarantee reasons repair work beside the wiring must only be made by VEGA-staff.
In case of a defect please return the respective instrument with a short description of the error to our repair department.

5.4 Fault signal

The signal conditioning instrument and the connected sensor are permanently monitored during operation. During the parameter adjustments to adjusted values are checked on plausibility. In both cases fault signals can be caused by irregularities.
In case of failure an error code is indicared on the display (only in the range indication of measured values), the fail safe relay de-energizes, the failure LED lights and the outputs react acc. to their control condition.
The error codes are listed in section 5.5 on the following pages.
The diagnosis info is transferred by means of a current text indication.
Configuration
You can determine separately for each meas. point, if a fault signal should be transferred or not. See menu range "Configuration meas. point" menu point "Fault signal", see page 21.
Parameter adjustment
In the menu point "Failure mode" can be determined separately for each measuring point how a decreasing of the measuring range should trigger a fault signal.
48 VEGAMET 514… D
Diagnosis

5.5 Error codes

Error code Meaning
E004 This operation is not supported by the instrument (different software version). You tried to
E007 Sensor type not suitable.
E008 VBUS-sensor not found. The sensor cannot be found.
E010 The serial numbers of the connected VBUS-sensor has not been coordinated.
E013 VBUS-sensor signals failure. The VBUS-sensor signals failure; the communication with the
E014 Suspected shortcircuit
activate a function with a newer software version of VEGA Visual Operating in an instrument with older software version. However the instrument cannot support this operation. Please update the software of your instrument.
A VBUS-sensor with the serial number stated by you was found, however the sensor does not correspond. Please check the adjusted serial number or change the configuration.
Reasons:
- sensor is not connected
- no sensor connected with this serial number
- line break (VBUS)
- shortcircuit (VBUS)
- the sensor is defect
- interfered communication (VBUS) probably
- interference in the VBUS-transmitter/receiver part of VEGAMET Please check the serial number, the connection line, the sensor and if necessary the VEGAMET.
sensor however is okay. Reasons:
- failure on the VBUS-sensor Please check the VBUS-sensor.
Reasons:
- shortcircuit
- sensor is defect Please check the connection line and if necessary the sensor.
E015 Suspected line break
E016 Empty/full adjustment exchanged. The values for the empty adjustment are higher than the
E017 Adjustment span too small. The adjustment values are too close togther or are identical.
E021 Span of the sensor characterstic values too small or values exchanged. The sensor
VEGAMET 514… D 49
Reasons:
- no sensor connected
- line break
- sensor is defect Please check the connection line and if necessary the sensor.
values for the full adjustment. Please carry out adjustment again.
Please carry out adjustment again.
characteristic values are too close together or are identical or the min. sensor characteristic value is higher than the max. sensor characteristic value. Please check the sensor charactersistic values.
Diagnosis
Error code Meaning
E029 Measured valuated. The measured value is generated by simulation and does not reflect the
real condition.
E031 Same DISBUS-address. At least another instrument with the same DISBUS-address is
E032 Instrument address 0. Instrument address 0 is adjusted on VEGAMET. With this address
E034 Instrument error. A considerable error was detected on the instrument (possibly defect
E035 Instrument error. A considerable error was detected on the instrument (possibly check sum
E050 Remote VEGAMET not found. The VEGAMET from which a measured value is expected,
E051 Failure during the transmission of measured values via DISBUS. It is not possible to receive
E052 Input number not yet coordinated.
connected to this instrument. Please check the DISBUS-address.
VEGAMET does not participate in the DISBUS-communication, even the DISBUS is wired. If necessary modify the instrument address.
EEPROM). Please disconnect voltage and then connect again (cold start). If the error does not extinguish, please contact your VEGA-agency.
error in EEPROM). Please disconnect voltage and then connect again (cold start). If the error does not extinguish, please contact your VEGA-agency.
cannot be found. Reasons:
- the remote VEGAMET is not connected
- no VEGAMET with this instrument address connected
- line break (DISBUS)
- shortcircuit (DISBUS)
- the remote VEGAMET is defect
- interfered communication (DISBUS) probably
- interference in the DISBUS-transmitter/receiver part of VEGAMET Please check the instrument address on both VEGAMET, the connection line, the remote VEGAMET and if necessary this VEGAMET.
a valid measured value via DISBUS from the remote VEGAMET; the communication to the remote VEGAMET however is okay. Reasons:
- failure on the remote VEGAMET
- the input of the remote VEGAMET which should be transmitted, is not yet configured Please check the remote VEGAMET and if necessary its input configuration.
E053 Span of the sensor characteristic values too small.
E054 Sensor characteristic values exchanged.
E055 The hysteresis on the correction signal input is too small.
E060 Span of the current output values too small. The values for the current output are too close
E061 Span of the voltage output values too small. The values for the voltage output are too close
E062 Pulse parameters too small. The pulse parameters in relation to the max. flow are too small,
50 VEGAMET 514… D
together or are identical. Please check the adjustment of the current output.
together or are identical. Please check the adjustment of the voltage output.
i.e. with this adjustment the time between two pulses would be less than one second. Please check the relation of pulse parameters to max. flow.
Diagnosis
Error code Meaning
E070 Scaling span too small. The values of the scaling are too close together or are identical.
E102 On this VEGAMET a reset of the configuration has been carried out via the keyboard.
E110 Switching hysteresis too small. The switch points are too close together or are identical.
E111 Relay switch points exchanged. The values for the relay switch points relating to the
E4.50 Wrong DISBUS-address adjusted.
E4.51 Wrong VEGADIS-number adjusted.
E9.10 No communication on DISBUS.
Please check the adjustments of the scaling.
selected relay mode are exchanged.
VEGAMET 514… D 51
VEGA Grieshaber KG Am Hohenstein 113 D-77761 Schiltach Phone (0 78 36) 50 - 0 Fax (0 78 36) 50 - 201 Fax (0 78 36) 50 - 203
ISO 9001
52 VEGAMET 514… D
Technical data subject to alterations 2.20 164 / April '96
Loading...