10.1 Menu schematic "reduced menu" .................................. 56
10.2 Menu schematic "extended menu" ................................. 58
Safety informatio n ...................................................................... 64
Note Ex area .............................................................................. 64
VEGAMET 5143
1 Product description
Product description
1.1 Function and configuration
VEGAMET 514 is a signal conditioning instrument for a number of applications, e.g.:
- level measurement
- gauge measurement
- process pressure measurement
- measurement with automatic correction
-etc.
Software configuration
Input component
- continuous current
input
- contact input
Function
The signal conditioning instruments power
the connected sensors and process their
analogue measuring data. The conditioning is
done via a special software consisting of
functional components (FB), input components (EB), output components (AB) as well
as DCS components (PB).
Configuration
The VEGAMET 514 signal conditioning instrument is constructed as module card in European size with 5 TE (= 25.4 mm) width. The
front plate contains an adjustment module
with LC display and keys. The front plate of
VEGAMET 514 N does not contain an adjustment module.
Output components
- current output
- voltage output
- DISBUS outputs
- failure output
- MET display
- relay outputs
AB
EBFB
PC/DCS components
- PC/DCS outputs
Functional component
- measurement loop
TAG 1
4VEGAMET 514
PB
Product description
1.2 Approvals
VEGAMET are available with the following
approvals:
For such applications, please note the appropriate official documents (general type approval, test and conformity certificates). They
come with the respective instrument
Ex approval
For hazardous areas, certification acc. to
CENELEC defined in the conformity certificate
PTB no. Ex-95.D.2145 X for VEGAMET 514
Ex.
WHG approval
Signal conditioning instrument as part of an
overfill protection system acc. to WHG (test
certificate number see table).
1.3 Features
- microcomputer-controlled signal conditioning instrument for continuous measurement
- adjustment module with LC display and 6
keys (not with VEGAMET 514 N)
- adjustable integration time
- two fixed and three user-programmable
linearisation curves
- fault monitoring
- fault signal and failure diagnosis via display
Inputs:
- 1 sensor input (capacitive electrode, pressure transmitter or other 4 … 20 mA
sensor)
- 1 correction input (for key, switch or level
switch)
Outputs:
- 1 current output 0/4 … 20 mA
- 1 voltage output 0/2 … 10 V
- 2 relay outputs (spdt)
- 1 DISBUS output for digital wiring and for
connection of VEGADIS 174
- 1 fail-safe relay
- adjustment also via PC with adjustment
software VVO; connection via the interface
adapter of the VEGACONNECT series
Number1 input
Input typeactive two-wire input, analogue (sensor is
Range4 … 20 mA
Sensorcapacitive electrodes, pressure transmitters,
Voltage
- at 4 mAapprox. 18 V DC
- at 20 mAapprox. 15 V DC
Current limitationat approx. 26 mA, short-circuit-proof
Detection line break< 3.6 mA
Detection short-circuit> 21 mA
Min. adjustment delta2 % of the entered sensor values
Connection cable2-wire standard cable (screening
max. resistance per wire35 Ω
Resolution1 µA
Linearity error0.025 % at 4 … 20 mA
Temperature error0.04 %/10 K at 4 … 20 mA
= 24 V AC (20 … 53 V), 50/60 Hz
nom
= 24 V DC (20 … 72 V)
powered by VEGAMET)
process pressure transmitters, differential
pressure transmitters
recommended)
Correction signal input
Number1 input
Functionswitching signal for triggering corrections
or special relay modes via
external switching contact
Voltage5 V (from instrument)
Current5 mA
External cable resistance≤€150 Ω
Current output
Number1 output
Functionanalogue output of the processing results
Rangeadjustable between 0 … 20 mA
Loadmax. 500 Ω
Resolution1 µA
Linearity error0.05 % (relating to 20 mA)
Temperature error0.05 %/10 K (relating to 20 mA)
6VEGAMET 514
Product description
Voltage output
Number1 output
Functionanalogue output of the processing results
Rangeadjustable between 0 … 10 V
Currentmax. 1 mA
Resolution0.5 mV
Linearity error0.05 % (relating to 10 V)
Temperature error0.06 %/10 K (relating to 10 V)
Relay output
Number2 switching relays
1 fail-safe relay
Contactfloating spdt
Contact materialAgNi, hard gold plated
Turn-on voltagemin. 10 mV DC
max. 250 V AC/DC
Switching currentmin. 10 µA
max. 3 A AC, 1 A DC
Breaking capacitancemax. 500 VA, 54 W
Min. switching hysteresis
(Low-/High delta)0.5 %
DISBUS output
Functionfor linking the signal conditioning instruments
and for connection of digital indicating
instruments
Connection cable2-wire standard cable (screening is
recommended)
max. cable length1000 m
Indicating elements
Clear text indicationLC display
(not available with VEGAMET 514N)- 4-line, 6 digits each
- background lighting
Analogue indicationLED chain consisting of:
(not available with VEGAMET 514N)- 11 segments 0 % … 100 %
- shows the actual value of the measurement
loop
LEDs in front plategreen: operating voltage on
red: fault signal (LED lights with deenergised
relay)
yellow: relay status (standard setting:
LED lights with energised relay)
Adjustment elements
Front plate6 keys for configuration and parameter
adjustment
(not available with VEGAMET 514N)
Circuit boardrotary switch for adjustment of the instrument
address on DISBUS
VEGAMET 5147
Product description
Ambient conditions
Permissible ambient temperature-20°C … +60°C
Storage and transport temperature-40°C … +80°C
Electrical connection
Multiple plug componentacc. to DIN 41 612, series F, 33-pole, 3 lines
d, b, z (partly equipped)
Carrier BGT 596 (Ex)connection to the appropriate module
Housing type 505connection to screw terminals (max. 2.5 mm
Electrical protective measures
Protection
- unassembledIP 00
- mounted into
carrier BGT 596
- front side completely equippedIP 40
- upper and lower sideIP 20
- wiring sideIP 00
- mounted into
housing type 505 or type 506
- terminal sideIP 20
- housing generalIP 30
Protection classII (in housing type 505)
Overvoltage categoryII
Electrical separating measures
Reliable separation acc. to
VDE 0106, part 1 betweenpower supply, fail safe and level relay
and measuring data inputs
- reference voltage250 V
- isolation resistance2.3 kV
Galvanic isolationbetween the relay outputs
- reference voltage250 V
- isolation resistance1.4 kV
Potential separationbetween DISBUS and outputs
- reference voltage50 V
- isolation resistance0.5 kV
Common reference potential atvoltage and current output,
correction signal output
2
)
Mechanical data
Seriesmodule unit for carrier BGT 596
or housing type 505
DimensionsB = 25.4 mm (5 TE), H = 128.4 mm, T = 162 mm
CE conformity
VEGAMET 514 and VEGAMET 514 Ex signal conditioning instruments meet the protective
regulations of EMC (89/336/EWG) and NSR (73/23/EWG). Conformity has been judged
acc. to the following standards:
EMCSusceptibilityEN 50 081 - 2: 1993
ImmissionEN 50 082 - 2: 1995
NSREN 61 010 - 1: 1993
8VEGAMET 514
Product description
Ex technical data
Power supply
Operating voltageU
Reference voltageU
Meas. data input (intrinsically safe circuit) on VEGAMET 513 Ex 2-times available
- powerP
Characteristicslinear
Effective inner inductance L
Effective inner capacitance C
Max. permissible outer inductance LO (mH)0.5 11.5229
Max. permissible outer capacitance CO (nF)97 78 684862001000
I
I
The intrinsically safe circuits are reliably separated from the non-intrinsically safe circuits up
to a peak value of the nominal voltage of 375 V. The max. voltage on the non-intrinsically
safe circuits should not exceed 250 V
eff
corresponds to non-Ex version
nom
= 250 V AC or 125 V DC
m
[EEx ib] IIB
= 20 V
O
= 128 mA
O
= 640 mW
O
negligible
negligible
EEx ia IICEEx ia IIBEEx ib IICEEx ib IIB
in case of failure.
1.5 Dimensions
Circuit board 100 x 160 x 1.5
European size
15
5,5
VEGAMET 5149
162
Multipoint
connector
100
5 TE
128,4
25,4
2 Mounting
Mounting
VEGAMET 514 signal conditioning
instruments can be either mounted with a
module in a carrier BGT 596 or BGT
596 Ex.M or in a single housing type 505.
Mounting in carrier
Mount the respective module (standard or Ex
version) in your carrier. Wire the terminals of
the multipoint connector according to the
wiring plan.
Module
Multipoint connector DIN 41 612, series F, 33pole (d, b, z) with coded pins and mounting
material for mounting in carrier BGT 596 or
BGT 596 Ex.M.
Module Ex
Multipoint connector DIN 41 612, series F, 33pole (d, b, z) with coded pins, Ex separating
chamber and mounting material for mounting
in carrier BGT 596 Ex.M.
The multipoint connector is available in the
following versions:
- Wire-Wrap standard connection
1.0 x 1.0 mm
- Plug connection 2.8 x 0.8 mm
- Termi-Point standard connection
1.6 x 0.8 mm
- Soldering connection
- Screw terminals 2 x 0.5 mm
2
For further mounting information, see the
operating instructions of the carrier.
Mounting in single housing
Either screw the housing socket directly to
the mounting plate or plug it onto a carrier rail
(35 x 15 acc. to EN 50 022 or 32 acc. to
EN 50 035). Connect the terminals according
to the wiring plans on the following pages.
For further mounting information see the operating instructions of the housing.
Transparent cover
To protect the instrument against unauthorised adjustment, the front plate of VEGAMET
can be provided with a lockable transparent
cover (supplied with the instrment).
Coding
To prevent an interchanging of the different
signal conditioning instruments, the multipoint
connector of carrier or the housing can be
provided with coded pins.
Note:
The coding is absolutely necessary for Ex
instruments!
The multiple plug of the signal conditioning
instrument is equipped with respective holes
(mechanical coding). Instrument coding
ensures that the different signal conditioning
instruments cannot be interchanged. The
appropriate coded pins are supplied with
each module or housing. Equip the multipoint
connector with these coded pins acc. to the
following table.
InstrumentEx
codingcoding
VEGAMET 514a3 / c5––
VEGAMET 514 Exa3 / c5c23
z b d
a c
a3
o 1 o
o 3 o
o 5 o
o 7 o
o 9 o
o11o
o13o
o15o
o17o
o19o
o21o
o23o
o25o
o27o
o29o
o31o
c5
c23
10VEGAMET 514
Electrical connection
3 Electrical connection
3.1 Connection instructions
Note the following instructions for electrical
connection:
- The connection must be made acc. to the
local installation standards (e.g. in Germany acc. to the VDE-regulations).
- The wiring between VEGAMET and sensor
can be done with standard two-wire cable.
- If strong electromagnetic interferences are
expected, screened cable is recommended. The screening must be earthed
on both ends.
- The line resistances stated in the technical
data must not be exceeded.
- If overvoltages are expected, we recom-
mend a sensor electronics with integrated
overvoltage protection or the installation of
VEGA overvoltage arresters.
- The voltage supply of VEGAMET must
provide with extra-low voltage to ensure
protection class II. When using VEGASTAB
593 a reliable separation from the mains
circuits acc. to DIN VDE 0106, part 101 is
ensured.
3.2 Connection instructions
for Ex approved applications
The signal conditioning instruments must be
generally mounted outside the hazardous
areas, or special Ex protective measures
must be taken.
Ex separating chamber
To ensure sufficient "air and creep distance",
an Ex separating chamber must be mounted
to the connections of VEGAMET. Lead the
cables through the Ex separating chamber
and connect them. Fasten the Ex separating
chamber with the lower screw. Note the
operating instructions of the carrier BGT 596
Ex.M.
Protection in Ex applications
In Ex applications a protection class of IP 20
must be maintained. To accomplish this,
cover the gaps or free modules from the front
with suitable blind covers.
Mounting in carriers
If you mount your VEGAMET with Ex approval in a carrier, you have to use a VEGAEx module. Keep a distance of at least 10
mm (2 TE) to module cards of other manufacturers. If you want to mount VEGAMET at the
utmost left position in the carrier, you have to
mount a blind cover with at least 20 mm
(4 TE) in front of the module of the instrument.
%
100
+
-
OKESC
CONNECT
12
on
Blind cover (4 TE)
VEGAMET 51411
3.3 Wiring plan
Supply
voltage
L (+)
N (-)
d b z
+-
2
6
Electrical connection
Fail safe relay
Current output
Voltage output
Correction signal
(input 4)
Sensor
(input 1)
10
12
+
16
-
18
+
20
+
22
--
24
28
+
-
30
32
Level relay 1
Level relay 2
+
+
DISBUS output
+
-
Ex version
For connection of Ex certified instruments,
please note the instructions in the attached
official documents as well as the valid mounting regulations. Make sure that the Ex separating chamber is mounted to the multipoint
connector. Always lead the cables through
the Ex separating chamber. Also note the
operating instructions of the carrier
BGT 596 Ex.M and the Ex instructions.
12VEGAMET 514
Adjustment
4 Adjustment
4.1 Indicating and adjustment elements
Circuit board
A rotary switch is located on the circuit
board. It is used for adjustment of the instrument address on DISBUS. This setting is
necessary if several VEGAMET signal conditioning instruments are linked via the DISBUS
or if you control a VEGADIS 174 indicating
instrument via the DISBUS.
The adjustment range 1 … F corresponds to
the DISBUS addresses 1 … 15.
Factory setting: 1
Note:
With address 0 the VEGAMET does not participate in the DISBUS communication.
Rotary switch
8
9
7
6
5
4
33
The set instrument address on the DISBUS
can be indicated via the LC display.
Note
If several VEGAMET are linked via the
DISBUS, each address must only be
assigned once!
A
B
C
D
2
E
F
1
0
10
11
12
13
14
15
Position of the rotary switch
Rotary switch
Front plate VEGAMET 514
without adjustmentwith adjustment
modulemodule
1
2
%
%
100
100
+
+
-
-
33
OKESC
OKESC
CONNECT
CONNECT
12
12
on
on
514
514
CONNECT
12
on
4
5
6
514
7
1 LCD (4 lines with 6 figures each, illuminated) for clear text
indication
2 LED chain (yellow) for quasi-analogue indication of the
measured value
3 Keys for menu adjustment
4 Connection socket for VEGACONNECT
5 LED (yellow) lights if the relay is energised (standard
setting)
6 LED (red), lights if fail safe relay is deenergised
7 LED (green), lights if operating voltage is on
VEGAMET 51413
4.2 Adjustment system
Adjustment
Indicating and adjustment module
Branching point, i.e. jump to the
next lower menu with [OK]
Depending on the menu item, interrupt
adjustment or change to the next higher
menu
Reduced menu - extended menu
Two menu modes are available: the mode
"reduced menu" and the mode "extended
menu".
Note:
VEGAMET 514 is preset to automatically
place in the menu mode "reduced menu"!
In the majority of all applications, you can
carry out the necessary settings in this menu
mode (see also chapter 4.5 Comparison of
the menu modes).
Adjustment structure
The adjustment is made via 6 keys in conjunction with the text display. The jump from
the measured value display to the underlying
menu is made with [OK]. Use [–>] or [<–] to
change within this menu level from one menu
item to the other. A branching point is
indicated by the symbol ▼ and enables with
[OK] a jump to the underlying menu item.
Parameters can be recognised by the
absence of the symbol ▼. The value of the
parameters can be modified with [+] or [–] or
chosen out of a list. The modified value can
be saved with [OK]. To interrupt a setting
(without saving the modification), you have to
push [ESC]. A jump back to the next higher
menu item is made with [ESC]. 15 minutes
Indication of
- measured value
- menu item
- parameter
%
100
+
-
OKESC
Depending on the menu item, change value
or choose out of a list
Choose menu item
Depending on the menu item, save the set
value or change to the underlying menu
after pushing a key for the last time, an automatic reset to the measured value display is
triggered. A jump back to the next higher
menu level with [ESC] is possible from each
menu item , even if no connecting line is
shown in the menu schematics.
Add'l
functions
Password
off
ESCESC
Language
English
ESC
TAG 1
to de
fault
Reset
TAG 1
ESC
ESC
Menu
mode
Reduced
If you are in the utmost right menu item of a
menu row and push the [–>] key, you will
immediately reach the utmost left menu item
of this row. You also reach immediately the
right menu item from the left one, if you push
the [<–] key.
These shortcuts (shown with a broken line)
are not shown in the menu schematics in
order to give a better overview.
Volt
output 1
Volt
prop.
to
Percent
Unit
0,0%
Volt
output
0/10V
Failure
mode
0V
limitation
on
14VEGAMET 514
Adjustment
Format examples of the menu schematic
Lang-
In white letters you see the param-
uage
eters which can be modified with the
English
[+] or [–] key and saved with the
[OK] key.
Examples:
- In the menu item "Language" you can
change from German to English.
Bold print/italic entries, e.g. the
Param.
TAG-
measurement loop name "TAG-No. 1"
No. 1
can differ if you have made a corres-
▼
ponding parameter adjustment or
configuration of the measurement loop. In the
menu schematics you will see the factory
settings.
Light grey menu items are only dis-
Volt
at
played if necessary (depending on
0%
the adjustments made in the other
0,000
menu items).
Change of the menu mode:
Move from the display of measured value
(TAG-No.1) to the menu item "Menu mode
reduced". Now push [+] twice until "Menu
mode Extended" appears, then push [OK].
TAGNo. 1
%
xx,x
Add'l
Confi-
Param.
TAGNo. 1
guration
functions
4.3 Adjustment via PC
Signal conditioning instrument with adjustment module
Setup
- directly via the keys of the adjustment
module
- or with a PC, equipped with VVO software
(VEGA Visual Operating) and an interface
adapter VEGACONNECT.
Signal conditioning instrument without
adjustment module
Setup
- with a PC, equipped with VVO software
(VEGA Visual Operating) and an interface
adapter VEGACONNECT.
%
100
+
-
OKESC
CONNECT
12
on
515
VEGACONNECT
PC with VVO software
Password
off
Language
English
TAG 1
to default
Reset
TAG 1
Menu
mode
Reduced
Extended
VEGAMET 51415
Adjustment
4.4 Configuration and parameter
adjustment
Independent of whether you set up your
VEGAMET via the keys of the adjustment
module or via a PC with the software VVO,
the procedure is always the same.
Proceed in the following sequence:
- configuration (if not already configured as
factory setting)
- parameter adjustment.
In this operating instruction manual, the adjustment steps are described which are
directly carried out on the keys of VEGAMET.
The adjustment via the software VVO is described in a separate manual.
Configuration
Configuring means assigning or setting
(once) function. VEGAMET requires a (mostly
once) basic co-ordination, determining the
applications and the assignments of the
inputs and outputs. Choices can be made
from existing functions and options. This
procedure is called configuration. The signal
conditioning instruments are delivered with a
configuration which only needs to be
changed in special cases.
The basic configuration includes the following
steps:
1 Configuration measurement loop
• Choose the kind of application (level,
gauge…)
• Choose sensor type (capacitive, hydro-
static…)
• Specify application (standard, level
difference…)
• Determine options (no option, correc-
tions…)
Note:
The measurement loop configuration can only
be modified if a "Reset to single measurement" has been carried out previosly (under
"Additional functions" in the extended menu).
If you only want to modify the sensor type,
you first have to carry out "Reset TAG 1"
(under "Additional functions" in the reduced
menu).
2 Configuration, inputs
• Determine from where your VEGAMET
receives the input data (sensor, other
VEGAMET)
• Enter sensor characteristics values (meas.
range, current range)
3 Configuration, outputs
• All outputs (except fail-safe relay) can be
assigned to a sensor or can be switched
off.
After basic configuration, VEGAMET starts
operating and indicates the actual measured
value. The other configurations should be
carried out after the parameter adjustment.
Parameter adjustment
Parameter adjustment means changing the
values. Signal conditioning instruments have
many parameters, the values of which can be
modified, such as for example the integration
time of 0 … 600 s. The modification of these
values is called parameter adjustment. The
parameter adjustment does not influence the
configuration. Take notice that parameter
adjustment is only possible after a configuration has been done (e.g. set values of the
current input when the current output has
been assigned).
16VEGAMET 514
Adjustment
4.5 Comparison: reduced menu – extended menu
In the table the most important functions of VEGAMET 514 are listed – each with a mark
indicating in which menu mode the functions are available.
FunctionReduc. Ext.
Adjustment with mediumxx
Adjustment without mediumxx
Set integration timexx
Simulationxx
Switch over current output (4/20 mA, 0/20 mA, 20/4 mA, 20/0 mA)xx
Switch over voltage output (2/10 V, 0/10 V, 10/2 V, 10/0 V)xx
Changeover relay mode (overfill protection – dry run protection)xx
Scaling (display)xx
Change menu mode (reduced – extended)xx
Change menu languagexx
Activate/deactivate passwordxx
Change sensor typexx
Change measurement loop namexx
Reset measurement loop (Reset TAG)xx
Reset measurement loop to default (TAG to default)xx
Use correction input–x
Edit linearisation curves–x
Change mode (e.g. level – gauge)–x
Select different reference values for outputs–x
Define output behaviour (current/voltage output) in case of failure–x
Adapt VEGAMET to sensor characteristics values–x
Offset correction during adjustment–x
Adjust relay switching delay–x
Reset level–x
Assign inputs (from another VEGAMET)–x
Switch on/off current/voltage limitation–x
Individual setting of current/voltage output values between 0 … 20 mA/0 … 10 V–x
Manual offset correction–x
Manual real value correction–x
Info indication–x
menu menu
VEGAMET 51417
5 Setup
• Adjust the VEGAMET instrument address
on the circuit board by means of the rotary
switch (only necessary, if several
VEGAMET are connected by means of
DISBUS).
• Install VEGAMET.
• Connect sensor and power supply.
• After switching on, the display indicates for
approx. 10 secs the instrument type and
the software version, e.g. "MET 514
V1.12".
• If a measurement loop has not yet been
configured, the display shows "Configuration", the red failure LED also lights. In this
case, proceed as described under "6.1
Configuration of measurement loop".
• If a measurement loop is already
configured, the display shows a value, e.g.
"TAG-No. 1 21.8 %".You can now carry out
the settings, as described in the chapter "6
Settings in the reduced menu".
Setup
18VEGAMET 514
Settings in the "reduced menu"
6 Settings in the "reduced menu"
6.1 Configuration of measurement loop
Choose sensor type
Meas.
TAG-
value
No. 1
%
indication
xx,x
Param.
TAGNo. 1
By default, VEGAMET 514 is configured for
connection to a hydrostatic sensor. If, for
example, you use a capacitive electrode, you
have to proceed as follows:
Move to the menu item "Sensor type
Hydrostatic".
Push [+] or [–] once. The word "Hydrostatic"
flashes. You are now in the editing mode.
Continue pushing [+] or [–] until the display
shows "Sensor type capacitive". You are still
in the editing mode.
Now push the [OK] key. By doing this, you
save the settings and quit the editing mode.
Add'l
Confi-
func-
gura-
tions
tion
Password
off
Sensor
type
It is only possible to
Hydro-
modify the sensor type,
static
when a "Reset TAG 1" was
capa-
carried out before.
citive
VEGADIF
Language
English
TAG 1
to default
Reset
TAG 1
Delete
TAG
1?
Reset
OK ?
Reset
Now!
OK ?
Name measurement loop
TAGNo. 1
%
xx,x
You can modify the measurement loop name "TAGNo. 1", if you want a more
useful designation.
Param.
TAGNo. 1
Configuration
Two lines with six figures
each are available. To do
this, move to the menu item
"TAG name …".
TAG
name
TAG No. 1
Push [+] or [–] once. The first char-
TAG
name
acter of the measurement loop name,
TAG -
in this example the letter T, flashes
No. 1
(editing mode). Continue pushing the
[+] or [–] key, by doing this you scroll
through the alphabet, through the row of
numbers and through a list of characters.
When you have reached the re-
TAG
name
quested character, push [–>]. The
TAG -
next character is activated and be-
No. 1
gins flashing. Scroll again with [+]
and [–] to the requested character. With [–>]
you again move to the next character.
When you have finished the last
TAG
name
position of the name and you have
Ves.4
"written" the name of the measure-
Hall 3
ment loop in this way, you have to
push [OK] to save the setting. If you quit the
menu item without saving, the name of the
measurement loop is again "TAG-No. 1".
In the display of measured value and
Ves.4
all other menu items in which the
Hall 3
%
measurement loop name is shown,
27,8
you will now see the new designation.
VEGAMET 51419
Settings in the "reduced menu"
6.2 Adjustment with medium
You can either carry out the adjustment with
medium or without medium. Adjustment with
medium means that you carry out the adjustment taking the actual level into account. To
carry out an adjustment with medium, it is
necessary to know the percentage value of
the actual vessel filling.
TAGNo. 1
%
xx,x
Param.
TAGNo. 1
Adjustment
with
medium
Min-
Max-
adjust
adjust
at %
at %
100,0
0,0
In this procedure, the percentage values
which correspond to the actual min. and max
fillings are entered. The adjustment sequence
of the min. and max. value is unimportant.
When you know, for example, that your filling
is actually 80 %, you can enter in the menu
max. adjustment the value "80". Later when
your filling is, for example, 10 %, you enter in
the menu min. adjustment the value "10".
Adjustment example
You know that your vessel is actually
Maxadjust
filled up to 80%. Move to the menu
at %
item "Max. adjust at % 100.0". Push
100,0
the [–] key briefly, so that the figure
100.0 flashes (editing mode).
Continue pushing [–] until the figure
Maxadjust
80.0 appears on the display. Now
at %
push [OK] to save the adjustment.
80,0
Later, when your filling is, for exam-
Minadjust
ple, 10 %, go to the menu item
at %
"Min. adjust at % 0.0". Briefly push
0,0
[+], until the figure 0.0 flashes
(editing mode).
Continue pushing [+] until "10" ap-
Minadjust
pears on the display. Now push [OK]
at %
to save the adjustment.
10,0
You can of course carry out the adjustment
with medium at 0 % and 100 % filling. Note:
The figures 0 % and 100 % must flash (editing mode), before they can be saved with the
[OK] key.
The bigger the difference between the two
adjustment points, the more precise the
measurement over the whole measuring
curve. An adjustment at 0 % and 100 %
would be ideal. For practical reasons, it is not
always possible to completely empty or fill a
vessel. The distance between the two adjustment points, however, should be at least
10 % of the sensor range.
20VEGAMET 514
Loading...
+ 44 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.