VEGA FLEX66 User Manual

Ope
rating Instructions
VEGAFLEX 66
-110 +400 °C 4 20 mA/HART
Guided Micro
wave

Content

Con
1 About this document
1.1 Function .
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . .
1.3 Symbolism used . . . . . . . . . . . . . . . . . . . . . . .
2 For your safety
2.1 Authorised personnel. . . . . . . . . . . . . . . . . . . .
2.2 Appropriate use. . . . . . . . . . . . . . . . . . . . . . . .
2.3 Warning about misuse . . . . . . . . . . . . . . . . . . .
2.4 General safety instructions . . . . . . . . . . . . . . . .
2.5 Safety approval markings and safety tips . . . . .
2.6 CE conformity . . . . . . . . . . . . . . . . . . . . . . . . .
2.7 Fullling of NAMUR recommendations . . . . . . .
2.8 SIL conformity . . . . . . . . . . . . . . . . . . . . . . . . .
2.9 Safety instructions for Ex areas . . . . . . . . . . . .
2.10 Manufacturer declaration for zone 2 . . . . . . . . .
2.11 Environmental instructions . . . . . . . . . . . . . . . .
3 Product description
3.1 Conguration. . . . . . . . . . . . . . . . . . . . . . . . . .
3.2 Principle of operation . . . . . . . . . . . . . . . . . . . .
3.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.4 Packaging, transport and storage . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 4 4
5 5 5 5 6 6 6 7 7 7 7
8
9 10 11
4 Mounting
4.1 General instructions. . . . . . . . . . . . . . . . . . . . .
4.2 Mounting instructions. . . . . . . . . . . . . . . . . . . .
5 Connecting to power supply
5.1 Preparing the connection . . . . . . . . . . . . . . . . .
5.2 Connection steps - Instrument housing . . . . . . .
5.3 Wiring plan, single chamber housing. . . . . . . . .
5.4 Wiring plan, double chamber housing . . . . . . . .
5.5 Wiring plan double chamber housing Ex d . . . .
5.6 Wiring plan - version IP 66/IP 68, 1 bar. . . . . . .
6 Set up with the indicating and adjustment module
PLICSCOM
6.1 Short description . . . . . . . . . . . . . . . . . . . . . . .
6.2 Insert indicating and adjustment module . . . . . .
6.3 Adjustment system . . . . . . . . . . . . . . . . . . . . .
6.4 Setup procedure . . . . . . . . . . . . . . . . . . . . . . .
6.5 Menu schematic . . . . . . . . . . . . . . . . . . . . . . .
2 VE
12 13
18 19 20 22 24 26
27 27 29 30 37
GAFLEX 66 4 20 mA/HART
32762-EN-080716
Content
7 Setup
with PACTware and other adjustment programs
7.1 Connecting the PC . . . . . . . . . . . . . . . . . . . . .
7.2 Parameter adjustment with PACTware . . . . . . .
7.3 Parameter adjustment with AMSand PDM . .
7.4 Saving the parameter adjustment data . . . . . . .
8 Maintenance and fault rectication
8.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . .
8.2 Remove interferences . . . . . . . . . . . . . . . . . . .
8.3 Exchanging the electronics module. . . . . . . . . .
8.4 Instrument repair . . . . . . . . . . . . . . . . . . . . . . .
9 Dismounting
9.1 Dismounting steps . . . . . . . . . . . . . . . . . . . . . .
9.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10 Supplement
10.1 Technical data. . . . . . . . . . . . . . . . . . . . . . . . .
10.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.3 Industrial property rights. . . . . . . . . . . . . . . . . .
10.4 Trademark . . . . . . . . . . . . . . . . . . . . . . . . . . .
39 40 41 41
42 42 44 44
46 46
47 58 62 62
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
Supplementary documentation
Information:
Supplemen
tary documents appropriate to the ordered version
come with the delivery. You can nd them listed in chapter
"Product description".
Instructions manuals for accessories and replacement parts
Tip:
ensure reliable setup and operation of your VEGAFLEX 66,
To
we oer accessories and replacement parts. The associated documents are:
l 27720 - VEGADIS 61 l 30207 - Electronics l 31088 - Flanges l 30391 - Spacer
T 3
module VEGAFLEX series 60
according to DIN-EN-ASME-JIS-GOST
1 Abou
t this document
1 Abo
ut this document

1.1 Function

This
operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectication. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.

1.2 Target group

This operating instructions manual is directed to trained personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, no
This symbol indicates helpful additional information.
Caution: If tions can result. Warning: If this warning is ignored, injury to persons and/or serious damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or destruction of the instrument can result.
this warning is ignored, faults or malfunc-
te
applications
Ex
This symbol indicates special instructions for Ex applications.
l List
The dot set in front indicates a list with no implied sequence.
à Action
This arrow
1 Sequence
Numbers set in front indicate successive steps in a procedure.
4 VE
indicates a single action.
32762-EN-080716
GAFLEX 66 4 20 mA/HART
2 For
your safety
2 For
your safety

2.1 Authorised personnel

All
operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protection equipment must always be worn.

2.2 Appropriate use

VEGAFLEX 66 is a sensor for continuous level measurement.
You can nd detailed information on the application range in
chapter "Product description".
Operational reliability is ensured only if the instrument is properly used according to the specications in the operating instructions manual as well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modications are explicitly forbidden.
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR

2.3 Warning about misuse

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

This is a high-tech instrument requiring the strict observance of standard regulations and guidelines. The user must take note of the safety instructions in this operating instructions manual, the country-specic installation standards as well as all prevailing safety regulations and accident prevention rules.
The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for trouble­free operation of the instrument.
T 5
2 For
your safety
the entire duration of use, the user is obliged to
During determine the compliance of the required occupational safety measures with the current valid rules and regulations and also take note of new regulations.

2.5 Safety approval markings and safety tips

The safety approval markings and safety tips on the device must be observed.

2.6 CE conformity

The protection goals of the EMC Directive 2004/108/EC (EMC) and the Low Voltage Directive 2006/95/EC (LVD) are fullled.
Conformity has been judged according to the following standards:
EMC: EN 61326-1: 2006
(electrical instruments for control technology and laboratory use - EMC requirements)
l Emission: Class B l Susceptibil
ity: Industrial
LVD: EN 61010-1: 2001
(safety regulations for electrical measurement, control and laboratory instruments - part 1: General requirements)
areas
2.7 Fullling of NAMUR recommendations
With respect to interference resistance and emitted interfer­ence, the NAMUR recommendation NE 21 is fullled.
With respect to compatibility, the NAMUR recommendation NE 53 is fullled. This applies also to the corresponding
indicating and adjustment components. VEGA instruments are generally upward and downward compatible.
l Sensor l DTM VEGAFLEX 66 for adjustment software PACTware l Indicating and adjustment module for sensor software
software for DTM
The parameter adjustment of the basic sensor functions is
independent of the software version. The range of available functions depends on the respective software version of the individual components.
The software version of VEGAFLEX 66 can be determined as follows:
6 VE
VEGAFLEX 66 HART, PA or FF
32762-EN-080716
GAFLEX 66 4 20 mA/HART
your safety
2 For
l via PACTware
the type label of the electronics
l on l via the indicating and adjustment module
You can
view all software histories on our website www.vega. com. Make use of this advantage and get registered for update information via e-mail.

2.8 SIL conformity

VEGAFLEX 66 fullls the requirements for functional safety according to IEC 61508/IEC 61511. You can nd further information in the Safety Manual "VEGAFLEX series 60 -
4 20 mA/HART".

2.9 Safety instructions for Ex areas

Please note the Ex-specic safety information for installation and operation in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex­approved instruments.

2.10 Manufacturer declaration for zone 2

In conformity with DIN EN 60079-15/2005 VEGAFLEX 66 is suitable for use in zone 2.
The operator must use the instrument as it was intended to be used and follow the specications of the following documents:
l this operating instructions manual l the manufacturer declaration 32905 (download under
"www.vega.com")
applicable installation regulations
l the
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR

2.11 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental protection. The environment man­agement system is certied according to DIN EN ISO 14001.
Please help us full this obligation by observing the environ­mental instructions in this manual:
and
l Chapter "Packaging, transport l Chapter "Disposal"
T 7
storage"
3 Produc
t description
Scope of delivery
Components
3 Prod
3.1 Congurat
uct description
ion
The scope of delivery encompasses:
l VEGAFLEX 66 level sensor l Documentation
- this
operating instructions manual
- Safety Manual 31339 "VEGAFLEX series 60 -
4 20 mA/HART"
- Operating instructions manual 27835 "Indicating and
adjustment module PLICSCOM" (optional)
- Supplementary instructions manual 31708 "Heating for
indicating and adjustment module" (optional)
- Supplementary instructions manual "Plug connector for
continuously measuring sensors" (optional)
- Ex-specic "Safety instructions" (with Ex-versions)
VEGAFLEX 66 consists of the following components:
with
l Process tting l Housing with electronics l Housing cover, optionally available with indicating and
probe
adjustment module
8 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
1 2
3
3 Produc
t description
Type label
Application range
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
Fig. 1: VEGAFLEX 66 - cable 1 Housing cover with integrated indicating and adjustment module (optional) 2 Housing with electronics 3 Process tting
The type label contains the most important data for identi-
cation and use of the instrument:
l Article number l Serial number l Technical data l Article numbers, document
version with plastic housing
ation
With the serial number, you can access the delivery data of the instrument via "
www.vega.com", "VEGA
Tools" and "serial
number search". In addition to the type label outside, you can
also nd the serial number on the inside of the instrument.

3.2 Principle of operation

VEGAFLEX 66 is a level sensor with coax, rod or cable probe for continuous level measurement in non-conductive products with temperatures of -110 +400 °C (-166 +752 °F).
T 9
3 Produc
t description
is designed for industrial use in all areas of process
It technology and can be used in non-conductive liquids.
Functional principle
Voltage supply
High frequency microwave pulses are guided along a steel rope or a rod. Upon reaching the product surface, the microwave pulses are reected. The running time is evaluated by the instrument and outputted as distance.
Two-wire electronics 4 … 20 mA/HART for power supply and measured value transmission over the same cable.
The supply voltage range can dier depending on the instrument version.
Data for power supply are specied in chapter "Technical data".
The backlight of the indicating and adjustment module is powered by the sensor. The prerequisite for this is a supply voltage at a certain level. The exact voltage specications are stated in chapter "Technical data".
The optional heating requires its own power supply. You can nd further details in the supplementary instructions manual
"Heating for indicating and adjustment module".
This function is generally not available for approved instru­ments.

3.3 Operation

VEGAFLEX 66 can be adjusted with dierent adjustment media:
l with indicating and adjustment module l with the suitable VEGA
DTM in adjustment software according to the FDT/DTM standard, e.g. PACTware and PC
l with manufacturer-specic
PDM
l With a HART handheld
The entered parameters are generally saved in VEGAFLEX
66, optionally also in the indicating and adjustment module or in PACTware.
conjunction with an
adjustment
programs AMSor
32762-EN-080716
10 VE
GAFLEX 66 4 20 mA/HART
3 Produc
t description
Packaging
Transport
Transport inspection
Storage
Storage and transport
temperature
3.4 Pack
aging, transport and storage
Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN EN 24180.
The packaging of standard instruments consists of environ-
ment-friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instruc­tions can cause damage to the device.
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
l Not l Dry and dust free l Not exposed to corrosive media l P l Avoiding mechanical shock and vibration
l Storage and transport temperature see "Supplement -
l Relative humidity 20 85 %
in the open
rotected against solar radiation
Technical
data - Ambient conditions"
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 11
4 M
ounting
Mounting position
Welding work
Handling
Moisture
4 Mou
nting

4.1 General instructions

Select
an installation position you can easily reach for mounting and connecting as well as later retrotting of an indicating and adjustment module. The housing can be rotated by 330° without the use of any tools. You can also install the indicating and adjustment module in four dierent positions (each displaced by 90°).
Before beginning the welding work, remove the oscillator (electronics) from the sensor. By doing this, you avoid damage to the electronics through inductive coupling.
With threaded versions, the housing must not be used to screw in the instrument! Applying tightening forces on the housing can damage its internal parts.
Use the hexagon for screwing in.
Use the recommended cables (see chapter "Connecting to
power supply") and tighten the cable gland.
You can give your instrument additional protection against moisture penetration by leading the connection cable down­ward in front of the cable entry. Rain and condensation water can thus drain off. This applies mainly to outdoor mounting as well as installation in areas where high humidity is expected (e. g. through cleaning processes) or on cooled or heated vessels.
Measuring range
Fig. 2: Measures
The reference plane for the measuring range of the sensors is
against moisture penetration
the sealing surface of the thread or ange.
12 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
ounting
4 M
in mind that a min. distance must be maintained below
Keep the reference plane and possibly also at the end of the probe ­measurement in these areas is not possible (dead band). Keep in mind that the cable length cannot be used all the way to the end because measurement in the area of the gravity weight is not possible. These min. distances (dead bands) are listed in chapter "Technical data".
Pressure
Mounting position
The process tting must be sealed if there is gauge or low pressure in the vessel. Before use, check if the seal material is resistant against the measured product and the process temperature.
The max. permissible pressure is specied in chapter "Technical data" or on the type label of the sensor.

4.2 Mounting instructions

Mount VEGAFLEX 66 in such a way that the distance to vessel installations or to the vessel wall is at least 300 mm (12 in).
During operation, the probe must not touch any installations or the vessel wall. If necessary, fasten the probe end.
In vessels with conical bottom it can be advantageous to mount the sensor in the center of the vessel, as measurement is then possible down to the lowest point of the bottom. When using the cable version, keep in mind that measurement down to the tip of the probe is not possible. The exact value of the
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 13
4 M
ounting
Standpipes or bypass tubes
min. distance (lower
dead band) is stated in chapter "Technical
data".
Fig. 3: Vessel
with conical bottom
Standpipes or bypass tubes are normally metal tubes with a diameter of 30 … 200 mm (1.18 7.87 in). In measurement technology such a tube corresponds to a coax probe. It does not matter if the standpipe is perforated or slotted for better mixing. Lateral inlets with bypass tubes also do not inuence the measurement.
Measuring probes can be mounted in bypass tubes up to DN 200.
The max. temperature is 250 °C (482 °F).
If VEGAFLEX 66 is used in standpipes or bypass tubes,
contact with the tube wall should be avoided. We oer spacers as accessories for fastening the probe in the middle of the tube.
Depending on the tube diameter or tube length, one or several spacers can be mounted. With cable probes, the cable can also be strained to avoid contact with the tube.
The max. temperature for the spacers is 250 °C (482 °F).
A special high temperature version of VEGAFLEX 66 with
integrated bypass tube can be used in applications with temperatures up to 400 °C (752 °F).
32762-EN-080716
14 VE
GAFLEX 66 4 20 mA/HART
in mind that buildup can form on the spacers. Strong
1
Keep buildup can inuence the measurement.
4 M
ounting
Inowing medium
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
Fig. 4: Position 1 Spacer
of the spacer
Note:
Measureme
nt in a standpipe is not recommended for very
adhesive products.
Make sure that the probe is not subjected to strong lateral forces. Mount VEGAFLEX 66 at a position in the vessel where no disturbances, e.g. from lling openings, agitators, etc., can occur.
T 15
4 M
ounting
Fixing
Heat insulation
Fig. 5: Lateral
load
Excessive system vibration or shocks, e.g. caused by agitators or turbulence in the vessel (e.g. from uidisation) can cause the coax probe of VEGAFLEX 66 to vibrate in resonance. Should a coax probe of more than 1 m (3.281 in) length should be used, you can provide a suitable support or guy directly above the end of the probe to stabilise it.
If there is a danger of the probe touching the vessel wall during operation due to product movements or agitators etc., the measuring probe should be securely xed.
In the gravity weight there is a thread (M12), e.g. for a ring bolt (article no. 2.27424).
Make sure that the probe cable is not completely taut. Avoid tensile loads on the cable.
Avoid undened cable-vessel connections, i.e. the connection must be either grounded reliably or isolated reliably. Any deviation from this requirement can lead to measurement errors.
VEGAFLEX 66 should be integrated in the vessel installation. This prevents the electronics from strong heating by heat
radiation.
32762-EN-080716
16 VE
GAFLEX 66 4 20 mA/HART
1
2
ounting
4 M
sure that with heated vessels the permissible ambient
Make temperature on the housing is not exceeded. The permissible ambient temperature is stated in chapter "Technical data" under "Ambient conditions".
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
Fig. 6: Vessel 1 Heat insulation 2 Ambient temperature on the housing
with heat insulation
T 17
ecting to power supply
5 Conn
Note safety instructions
Take note of
safety instruc­tions for Ex ap­plications
Select power supply
Selecting connection
cable
5 Con
5.1 Preparin
necting to power supply
g the connection
Always keep in mind the following safety instructions:
l Connect only in the complete absence of line voltage l If overvoltage surges are expected, overvoltage
arresters
should be installed
Tip:
We
recommend using VEGA overvoltage arresters B63-48
and ÜSB 62-36G.X.
In
hazardous areas you should take note of the appropriate regulations, conformity and type approval certicates of the sensors and power supply units.
Power supply and current signal are carried on the same two­wire cable. The voltage supply range can dier depending on the instrument version.
Data for power supply are specied in chapter "Technical data".
Provide a reliable separation between the supply circuit and the mains circuits according to DIN VDE 0106 part 101. The
VEGA power supply units VEGATRENN 149A Ex, VEGAS­TAB 690 as well as all VEGAMETs meet this requirement.
Bear in mind the following factors regarding supply voltage:
l Outpu
t voltage of the power supply unit can be lower under
nominal
load (with a sensor current of 20.5 mA or 22 mA in
case of fault message)
l Inuence
of additional instruments in the circuit (see load
values in chapter "Technical data")
The instrument is connected with standard two-wire cable without screen. If electromagnetic interference is expected which is above the test values of EN 61326 for industrial areas, screened cable should be used.
Use cable with round cross-section. A cable outer diameter of 5 9 mm (0.2 0.35 in) ensures the seal eect of the cable gland. If you are using cable with other diameter or cross­section, you have to exchange the seal or use a suitable cable gland.
32762-EN-080716
18 VE
GAFLEX 66 4 20 mA/HART
ecting to power supply
5 Conn
generally recommend the use of screened cable for HART
We multidrop mode.
Cable gland ½ NP T
Cable screening and
grounding
Select connec­tion cable for Ex applications
On the instrument with cable entry ½ NPT and plastic housing there is a metallic ½" threaded insert moulded into the plastic housing.
Caution:
No
grease should be used when screwing the NPT cable gland or steel tube into the threaded insert. Standard grease can contain additives that corrode the connection between threaded insert and housing. This would inuence the stability of the connection and the tightness of the housing.
If screened cable is necessary, connect the cable screen on both ends to ground potential. In the sensor the screen must be connected directly to the internal ground terminal. The ground terminal on the outside of the housing must be connected to the potential equalisation (low impedance).
If potential equalisation currents are expected, the connection on the processing side must be made via a ceramic capacitor (e. g. 1 nF, 1500 V). The low frequency potential equalisation currents are thus suppressed, but the protective eect against high frequency interference signals remains.
note of the corresponding installation regulations for Ex
Take applications. In particular, make sure that no potential equal­isation currents ow over the cable screen. In case of grounding on both sides this can be achieved by the use of a capacitor or a separate potential equalisation.
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
5.2 Connection steps - Instrument housing
Proceed as follows:
1 Unscrew the housing cover 2 If an indicating and adjustment module is installed, remove
it by turning it slightly to the left.
3 Loosen compression nut of the cable entry 4 Remove approx. 10 cm (4 in) of the cable mantle, strip
approx. 1 cm (0.4 in) insulation from the ends of the
individual wires
5 Insert the cable through the cable gland into the sensor
T 19
ecting to power supply
5 Conn
the opening levers of the terminals with a screwdriver
6 Lift
(see following illustration)
7 Insert the wire ends into the open terminals according to
the wiring plan
Fig. 7: Connection
steps 6 and 7
8 Press down the opening levers of the terminals, you will
hear the terminal spring closing
9 Check the hold of the wires in the terminals by lightly
pulling on them
10 Connect the screen to the internal ground terminal, connect
the outer ground terminal with potential equalisation
11 Tighten the compression nut of the cable entry. The seal
ring must completely encircle the cable
12 Screw the housing cover on The electrical connection is nished.

5.3 Wiring plan, single chamber housing

following illustrations apply to the non-Ex as well as to the
The Ex-ia version.
20 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
Loading...
+ 44 hidden pages