come with the delivery. You can find them listed in chapter
"Product description".
Instructions manuals for accessories and replacement
parts
Tip:
ensure reliable setup and operation of your VEGAFLEX 66,
To
we offer accessories and replacement parts. The associated
documents are:
l 27720 - VEGADIS 61
l 30207 - Electronics
l 31088 - Flanges
l 30391 - Spacer
T3
module VEGAFLEX series 60
according to DIN-EN-ASME-JIS-GOST
1 Abou
t this document
1 Abo
ut this document
1.1 Function
This
operating instructions manual provides all the information
you need for mounting, connection and setup as well as
important instructions for maintenance and fault rectification.
Please read this information before putting the instrument into
operation and keep this manual accessible in the immediate
vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained
personnel. The contents of this manual should be made
available to these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, no
This symbol indicates helpful additional information.
Caution: If
tions can result.
Warning: If this warning is ignored, injury to persons and/or
serious damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons
and/or destruction of the instrument can result.
this warning is ignored, faults or malfunc-
te
applications
Ex
This symbol indicates special instructions for Ex applications.
lList
The dot set in front indicates a list with no implied sequence.
àAction
This arrow
1Sequence
Numbers set in front indicate successive steps in a procedure.
4VE
indicates a single action.
32762-EN-080716
GAFLEX 66 • 4 … 20 mA/HART
2 For
your safety
2 For
your safety
2.1 Authorised personnel
All
operations described in this operating instructions manual
must be carried out only by trained specialist personnel
authorised by the plant operator.
During work on and with the device the required personal
protection equipment must always be worn.
2.2 Appropriate use
VEGAFLEX 66 is a sensor for continuous level measurement.
You can find detailed information on the application range in
chapter "Product description".
Operational reliability is ensured only if the instrument is
properly used according to the specifications in the operating
instructions manual as well as possible supplementary
instructions.
For safety and warranty reasons, any invasive work on the
device beyond that described in the operating instructions
manual may be carried out only by personnel authorised by the
manufacturer. Arbitrary conversions or modifications are
explicitly forbidden.
32762-EN-080716
VEGAFLEX 66 • 4 … 20 mA/HAR
2.3 Warning about misuse
Inappropriate or incorrect use of the instrument can give rise to
application-specific hazards, e.g. vessel overfill or damage to
system components through incorrect mounting or adjustment.
2.4 General safety instructions
This is a high-tech instrument requiring the strict observance of
standard regulations and guidelines. The user must take note
of the safety instructions in this operating instructions manual,
the country-specific installation standards as well as all
prevailing safety regulations and accident prevention rules.
The instrument must only be operated in a technically flawless
and reliable condition. The operator is responsible for troublefree operation of the instrument.
T5
2 For
your safety
the entire duration of use, the user is obliged to
During
determine the compliance of the required occupational safety
measures with the current valid rules and regulations and also
take note of new regulations.
2.5 Safety approval markings and safety tips
The safety approval markings and safety tips on the device
must be observed.
2.6 CE conformity
The protection goals of the EMC Directive 2004/108/EC (EMC)
and the Low Voltage Directive 2006/95/EC (LVD) are fulfilled.
Conformity has been judged according to the following
standards:
EMC: EN 61326-1: 2006
(electrical instruments for control technology and laboratory
use - EMC requirements)
l Emission: Class B
l Susceptibil
ity: Industrial
LVD: EN 61010-1: 2001
(safety regulations for electrical measurement, control and
laboratory instruments - part 1: General requirements)
areas
2.7 Fulfilling of NAMUR recommendations
With respect to interference resistance and emitted interference, the NAMUR recommendation NE 21 is fulfilled.
With respect to compatibility, the NAMUR recommendation
NE 53 is fulfilled. This applies also to the corresponding
indicating and adjustment components. VEGA instruments are
generally upward and downward compatible.
l Sensor
l DTM VEGAFLEX 66 for adjustment software PACTware
l Indicating and adjustment module for sensor software
software for DTM
The parameter adjustment of the basic sensor functions is
independent of the software version. The range of available
functions depends on the respective software version of the
individual components.
The software version of VEGAFLEX 66 can be determined as
follows:
6VE
VEGAFLEX 66 HART, PA or FF
32762-EN-080716
GAFLEX 66 • 4 … 20 mA/HART
your safety
2 For
l via PACTware
the type label of the electronics
l on
l via the indicating and adjustment module
You can
view all software histories on our website www.vega.
com. Make use of this advantage and get registered for update
information via e-mail.
2.8 SIL conformity
VEGAFLEX 66 fulfills the requirements for functional safety
according to IEC 61508/IEC 61511. You can find further
information in the Safety Manual "VEGAFLEX series 60 -
4 … 20 mA/HART".
2.9 Safety instructions for Ex areas
Please note the Ex-specific safety information for installation
and operation in Ex areas. These safety instructions are part of
the operating instructions manual and come with the Exapproved instruments.
2.10 Manufacturer declaration for zone 2
In conformity with DIN EN 60079-15/2005 VEGAFLEX 66 is
suitable for use in zone 2.
The operator must use the instrument as it was intended to be
used and follow the specifications of the following documents:
l this operating instructions manual
l the manufacturer declaration 32905 (download under
"www.vega.com")
applicable installation regulations
l the
32762-EN-080716
VEGAFLEX 66 • 4 … 20 mA/HAR
2.11 Environmental instructions
Protection of the environment is one of our most important
duties. That is why we have introduced an environment
management system with the goal of continuously improving
company environmental protection. The environment management system is certified according to DIN EN ISO 14001.
Please help us fulfil this obligation by observing the environmental instructions in this manual:
and
l Chapter "Packaging, transport
l Chapter "Disposal"
T7
storage"
3 Produc
t description
Scope of delivery
Components
3 Prod
3.1 Configurat
uct description
ion
The scope of delivery encompasses:
l VEGAFLEX 66 level sensor
l Documentation
- this
operating instructions manual
- Safety Manual 31339 "VEGAFLEX series 60 -
4 … 20 mA/HART"
- Operating instructions manual 27835 "Indicating and
adjustment module PLICSCOM" (optional)
- Supplementary instructions manual 31708 "Heating for
indicating and adjustment module" (optional)
- Supplementary instructions manual "Plug connector for
l Process fitting
l Housing with electronics
l Housing cover, optionally available with indicating and
probe
adjustment module
8VE
32762-EN-080716
GAFLEX 66 • 4 … 20 mA/HART
1
2
3
3 Produc
t description
Type label
Application range
32762-EN-080716
VEGAFLEX 66 • 4 … 20 mA/HAR
Fig. 1: VEGAFLEX 66 - cable
1Housing cover with integrated indicating and adjustment module (optional)
2Housing with electronics
3Process fitting
The type label contains the most important data for identi-
fication and use of the instrument:
l Article number
l Serial number
l Technical data
l Article numbers, document
version with plastic housing
ation
With the serial number, you can access the delivery data of the
instrument via "
www.vega.com", "VEGA
Tools" and "serial
number search". In addition to the type label outside, you can
also find the serial number on the inside of the instrument.
3.2 Principle of operation
VEGAFLEX 66 is a level sensor with coax, rod or cable probe
for continuous level measurement in non-conductive products
with temperatures of -110 … +400 °C (-166 … +752 °F).
T9
3 Produc
t description
is designed for industrial use in all areas of process
It
technology and can be used in non-conductive liquids.
Functional principle
Voltage supply
High frequency microwave pulses are guided along a steel
rope or a rod. Upon reaching the product surface, the
microwave pulses are reflected. The running time is evaluated
by the instrument and outputted as distance.
Two-wire electronics 4 … 20 mA/HART for power supply and
measured value transmission over the same cable.
The supply voltage range can differ depending on the
instrument version.
Data for power supply are specified in chapter "Technical
data".
The backlight of the indicating and adjustment module is
powered by the sensor. The prerequisite for this is a supply
voltage at a certain level. The exact voltage specifications are
stated in chapter "Technical data".
The optional heating requires its own power supply. You can
find further details in the supplementary instructions manual
"Heating for indicating and adjustment module".
This function is generally not available for approved instruments.
3.3 Operation
VEGAFLEX 66 can be adjusted with different adjustment
media:
l with indicating and adjustment module
l with the suitable VEGA
DTM in
adjustment software according to the FDT/DTM standard,
e.g. PACTware and PC
l with manufacturer-specific
PDM
l With a HART handheld
The entered parameters are generally saved in VEGAFLEX
66, optionally also in the indicating and adjustment module or
in PACTware.
conjunction with an
adjustment
programs AMS™ or
32762-EN-080716
10VE
GAFLEX 66 • 4 … 20 mA/HART
3 Produc
t description
Packaging
Transport
Transport inspection
Storage
Storage and transport
temperature
3.4 Pack
aging, transport and storage
Your instrument was protected by packaging during transport.
Its capacity to handle normal loads during transport is assuredby a test according to DIN EN 24180.
The packaging of standard instruments consists of environ-
ment-friendly, recyclable cardboard. For special versions, PE
foam or PE foil is also used. Dispose of the packaging material
via specialised recycling companies.
Transport must be carried out under consideration of the notes
on the transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible
transit damage immediately at receipt. Ascertained transit
damage or concealed defects must be appropriately dealt
with.
Up to the time of installation, the packages must be left closed
and stored according to the orientation and storage markings
on the outside.
Unless otherwise indicated, the packages must be stored only
under the following conditions:
l Not
l Dry and dust free
l Not exposed to corrosive media
l P
l Avoiding mechanical shock and vibration
l Storage and transport temperature see "Supplement -
l Relative humidity 20 … 85 %
in the open
rotected against solar radiation
Technical
data - Ambient conditions"
32762-EN-080716
VEGAFLEX 66 • 4 … 20 mA/HAR
T11
4 M
ounting
Mounting position
Welding work
Handling
Moisture
4 Mou
nting
4.1 General instructions
Select
an installation position you can easily reach for
mounting and connecting as well as later retrofitting of an
indicating and adjustment module. The housing can be rotated
by 330° without the use of any tools. You can also install the
indicating and adjustment module in four different positions
(each displaced by 90°).
Before beginning the welding work, remove the oscillator
(electronics) from the sensor. By doing this, you avoid damageto the electronics through inductive coupling.
With threaded versions, the housing must not be used to screw
in the instrument! Applying tightening forces on the housing
can damage its internal parts.
Use the hexagon for screwing in.
Use the recommended cables (see chapter "Connecting to
power supply") and tighten the cable gland.
You can give your instrument additional protection against
moisture penetration by leading the connection cable downward in front of the cable entry. Rain and condensation water
can thus drain off. This applies mainly to outdoor mounting as
well as installation in areas where high humidity is expected (e.
g. through cleaning processes) or on cooled or heated
vessels.
Measuring range
Fig. 2: Measures
The reference plane for the measuring range of the sensors is
against moisture penetration
the sealing surface of the thread or flange.
12VE
32762-EN-080716
GAFLEX 66 • 4 … 20 mA/HART
ounting
4 M
in mind that a min. distance must be maintained below
Keep
the reference plane and possibly also at the end of the probe measurement in these areas is not possible (dead band). Keep
in mind that the cable length cannot be used all the way to the
end because measurement in the area of the gravity weight is
not possible. These min. distances (dead bands) are listed in
chapter "Technical data".
Pressure
Mounting position
The process fitting must be sealed if there is gauge or low
pressure in the vessel. Before use, check if the seal material is
resistant against the measured product and the process
temperature.
The max. permissible pressure is specified in chapter
"Technical data" or on the type label of the sensor.
4.2 Mounting instructions
Mount VEGAFLEX 66 in such a way that the distance to vessel
installations or to the vessel wall is at least 300 mm (12 in).
During operation, the probe must not touch any installations or
the vessel wall. If necessary, fasten the probe end.
In vessels with conical bottom it can be advantageous to
mount the sensor in the center of the vessel, as measurement
is then possible down to the lowest point of the bottom. When
using the cable version, keep in mind that measurement down
to the tip of the probe is not possible. The exact value of the
32762-EN-080716
VEGAFLEX 66 • 4 … 20 mA/HAR
T13
4 M
ounting
Standpipes or bypass
tubes
min. distance (lower
dead band) is stated in chapter "Technical
data".
Fig. 3: Vessel
with conical bottom
Standpipes or bypass tubes are normally metal tubes with a
diameter of 30 … 200 mm (1.18 … 7.87 in). In measurement
technology such a tube corresponds to a coax probe. It does
not matter if the standpipe is perforated or slotted for better
mixing. Lateral inlets with bypass tubes also do not influence
the measurement.
Measuring probes can be mounted in bypass tubes up to
DN 200.
The max. temperature is 250 °C (482 °F).
If VEGAFLEX 66 is used in standpipes or bypass tubes,
contact with the tube wall should be avoided. We offer spacers
as accessories for fastening the probe in the middle of the
tube.
Depending on the tube diameter or tube length, one or several
spacers can be mounted. With cable probes, the cable can
also be strained to avoid contact with the tube.
The max. temperature for the spacers is 250 °C (482 °F).
A special high temperature version of VEGAFLEX 66 with
integrated bypass tube can be used in applications with
temperatures up to 400 °C (752 °F).
32762-EN-080716
14VE
GAFLEX 66 • 4 … 20 mA/HART
in mind that buildup can form on the spacers. Strong
1
Keep
buildup can influence the measurement.
4 M
ounting
Inflowing medium
32762-EN-080716
VEGAFLEX 66 • 4 … 20 mA/HAR
Fig. 4: Position
1Spacer
of the spacer
Note:
Measureme
nt in a standpipe is not recommended for very
adhesive products.
Make sure that the probe is not subjected to strong lateral
forces. Mount VEGAFLEX 66 at a position in the vessel where
no disturbances, e.g. from filling openings, agitators, etc., can
occur.
T15
4 M
ounting
Fixing
Heat insulation
Fig. 5: Lateral
load
Excessive system vibration or shocks, e.g. caused by agitators
or turbulence in the vessel (e.g. from fluidisation) can cause
the coax probe of VEGAFLEX 66 to vibrate in resonance.
Should a coax probe of more than 1 m (3.281 in) length should
be used, you can provide a suitable support or guy directly
above the end of the probe to stabilise it.
If there is a danger of the probe touching the vessel wall during
operation due to product movements or agitators etc., the
measuring probe should be securely fixed.
In the gravity weight there is a thread (M12), e.g. for a ring bolt
(article no. 2.27424).
Make sure that the probe cable is not completely taut. Avoid
tensile loads on the cable.
Avoid undefined cable-vessel connections, i.e. the connection
must be either grounded reliably or isolated reliably. Any
deviation from this requirement can lead to measurement
errors.
VEGAFLEX 66 should be integrated in the vessel installation.
This prevents the electronics from strong heating by heat
radiation.
32762-EN-080716
16VE
GAFLEX 66 • 4 … 20 mA/HART
1
2
ounting
4 M
sure that with heated vessels the permissible ambient
Make
temperature on the housing is not exceeded. The permissible
ambient temperature is stated in chapter "Technical data"
under "Ambient conditions".
32762-EN-080716
VEGAFLEX 66 • 4 … 20 mA/HAR
Fig. 6: Vessel
1Heat insulation
2Ambient temperature on the housing
with heat insulation
T17
ecting to power supply
5 Conn
Note safety instructions
Take note of
safety instructions for Ex applications
Select power supply
Selecting connection
cable
5 Con
5.1 Preparin
necting to power supply
g the connection
Always keep in mind the following safety instructions:
l Connect only in the complete absence of line voltage
l If overvoltage surges are expected, overvoltage
arresters
should be installed
Tip:
We
recommend using VEGA overvoltage arresters B63-48
and ÜSB 62-36G.X.
In
hazardous areas you should take note of the appropriate
regulations, conformity and type approval certificates of the
sensors and power supply units.
Power supply and current signal are carried on the same twowire cable. The voltage supply range can differ depending on
the instrument version.
Data for power supply are specified in chapter "Technical
data".
Provide a reliable separation between the supply circuit and
the mains circuits according to DIN VDE 0106 part 101. The
VEGA power supply units VEGATRENN 149A Ex, VEGASTAB 690 as well as all VEGAMETs meet this requirement.
Bear in mind the following factors regarding supply voltage:
l Outpu
t voltage of the power supply unit can be lower under
nominal
load (with a sensor current of 20.5 mA or 22 mA in
case of fault message)
l Influence
of additional instruments in the circuit (see load
values in chapter "Technical data")
The instrument is connected with standard two-wire cable
without screen. If electromagnetic interference is expected
which is above the test values of EN 61326 for industrial areas,
screened cable should be used.
Use cable with round cross-section. A cable outer diameter of
5 … 9 mm (0.2 … 0.35 in) ensures the seal effect of the cable
gland. If you are using cable with other diameter or crosssection, you have to exchange the seal or use a suitable cable
gland.
32762-EN-080716
18VE
GAFLEX 66 • 4 … 20 mA/HART
ecting to power supply
5 Conn
generally recommend the use of screened cable for HART
We
multidrop mode.
Cable gland ½ NP T
Cable screening and
grounding
Select connection cable for Ex
applications
On the instrument with cable entry ½ NPT and plastic housing
there is a metallic ½" threaded insert moulded into the plastic
housing.
Caution:
No
grease should be used when screwing the NPT cable
gland or steel tube into the threaded insert. Standard grease
can contain additives that corrode the connection between
threaded insert and housing. This would influence the stability
of the connection and the tightness of the housing.
If screened cable is necessary, connect the cable screen on
both ends to ground potential. In the sensor the screen must
be connected directly to the internal ground terminal. The
ground terminal on the outside of the housing must be
connected to the potential equalisation (low impedance).
If potential equalisation currents are expected, the connection
on the processing side must be made via a ceramic capacitor
(e. g. 1 nF, 1500 V). The low frequency potential equalisation
currents are thus suppressed, but the protective effect against
high frequency interference signals remains.
note of the corresponding installation regulations for Ex
Take
applications. In particular, make sure that no potential equalisation currents flow over the cable screen. In case of
grounding on both sides this can be achieved by the use of a
capacitor or a separate potential equalisation.
32762-EN-080716
VEGAFLEX 66 • 4 … 20 mA/HAR
5.2 Connection steps - Instrument housing
Proceed as follows:
1 Unscrew the housing cover
2 If an indicating and adjustment module is installed, remove
it by turning it slightly to the left.
3 Loosen compression nut of the cable entry
4 Remove approx. 10 cm (4 in) of the cable mantle, strip
approx. 1 cm (0.4 in) insulation from the ends of the
individual wires
5 Insert the cable through the cable gland into the sensor
T19
ecting to power supply
5 Conn
the opening levers of the terminals with a screwdriver
6 Lift
(see following illustration)
7 Insert the wire ends into the open terminals according to
the wiring plan
Fig. 7: Connection
steps 6 and 7
8 Press down the opening levers of the terminals, you will
hear the terminal spring closing
9 Check the hold of the wires in the terminals by lightly
pulling on them
10 Connect the screen to the internal ground terminal, connect
the outer ground terminal with potential equalisation
11 Tighten the compression nut of the cable entry. The seal
ring must completely encircle the cable
12 Screw the housing cover on
The electrical connection is finished.
5.3 Wiring plan, single chamber housing
following illustrations apply to the non-Ex as well as to the
The
Ex-ia version.
20VE
32762-EN-080716
GAFLEX 66 • 4 … 20 mA/HART
Loading...
+ 44 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.