VEGA FLEX66 User Manual

Ope
rating Instructions
VEGAFLEX 66
-110 +400 °C 4 20 mA/HART
Guided Micro
wave

Content

Con
1 About this document
1.1 Function .
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . .
1.3 Symbolism used . . . . . . . . . . . . . . . . . . . . . . .
2 For your safety
2.1 Authorised personnel. . . . . . . . . . . . . . . . . . . .
2.2 Appropriate use. . . . . . . . . . . . . . . . . . . . . . . .
2.3 Warning about misuse . . . . . . . . . . . . . . . . . . .
2.4 General safety instructions . . . . . . . . . . . . . . . .
2.5 Safety approval markings and safety tips . . . . .
2.6 CE conformity . . . . . . . . . . . . . . . . . . . . . . . . .
2.7 Fullling of NAMUR recommendations . . . . . . .
2.8 SIL conformity . . . . . . . . . . . . . . . . . . . . . . . . .
2.9 Safety instructions for Ex areas . . . . . . . . . . . .
2.10 Manufacturer declaration for zone 2 . . . . . . . . .
2.11 Environmental instructions . . . . . . . . . . . . . . . .
3 Product description
3.1 Conguration. . . . . . . . . . . . . . . . . . . . . . . . . .
3.2 Principle of operation . . . . . . . . . . . . . . . . . . . .
3.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.4 Packaging, transport and storage . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 4 4
5 5 5 5 6 6 6 7 7 7 7
8
9 10 11
4 Mounting
4.1 General instructions. . . . . . . . . . . . . . . . . . . . .
4.2 Mounting instructions. . . . . . . . . . . . . . . . . . . .
5 Connecting to power supply
5.1 Preparing the connection . . . . . . . . . . . . . . . . .
5.2 Connection steps - Instrument housing . . . . . . .
5.3 Wiring plan, single chamber housing. . . . . . . . .
5.4 Wiring plan, double chamber housing . . . . . . . .
5.5 Wiring plan double chamber housing Ex d . . . .
5.6 Wiring plan - version IP 66/IP 68, 1 bar. . . . . . .
6 Set up with the indicating and adjustment module
PLICSCOM
6.1 Short description . . . . . . . . . . . . . . . . . . . . . . .
6.2 Insert indicating and adjustment module . . . . . .
6.3 Adjustment system . . . . . . . . . . . . . . . . . . . . .
6.4 Setup procedure . . . . . . . . . . . . . . . . . . . . . . .
6.5 Menu schematic . . . . . . . . . . . . . . . . . . . . . . .
2 VE
12 13
18 19 20 22 24 26
27 27 29 30 37
GAFLEX 66 4 20 mA/HART
32762-EN-080716
Content
7 Setup
with PACTware and other adjustment programs
7.1 Connecting the PC . . . . . . . . . . . . . . . . . . . . .
7.2 Parameter adjustment with PACTware . . . . . . .
7.3 Parameter adjustment with AMSand PDM . .
7.4 Saving the parameter adjustment data . . . . . . .
8 Maintenance and fault rectication
8.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . .
8.2 Remove interferences . . . . . . . . . . . . . . . . . . .
8.3 Exchanging the electronics module. . . . . . . . . .
8.4 Instrument repair . . . . . . . . . . . . . . . . . . . . . . .
9 Dismounting
9.1 Dismounting steps . . . . . . . . . . . . . . . . . . . . . .
9.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10 Supplement
10.1 Technical data. . . . . . . . . . . . . . . . . . . . . . . . .
10.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.3 Industrial property rights. . . . . . . . . . . . . . . . . .
10.4 Trademark . . . . . . . . . . . . . . . . . . . . . . . . . . .
39 40 41 41
42 42 44 44
46 46
47 58 62 62
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
Supplementary documentation
Information:
Supplemen
tary documents appropriate to the ordered version
come with the delivery. You can nd them listed in chapter
"Product description".
Instructions manuals for accessories and replacement parts
Tip:
ensure reliable setup and operation of your VEGAFLEX 66,
To
we oer accessories and replacement parts. The associated documents are:
l 27720 - VEGADIS 61 l 30207 - Electronics l 31088 - Flanges l 30391 - Spacer
T 3
module VEGAFLEX series 60
according to DIN-EN-ASME-JIS-GOST
1 Abou
t this document
1 Abo
ut this document

1.1 Function

This
operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectication. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.

1.2 Target group

This operating instructions manual is directed to trained personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, no
This symbol indicates helpful additional information.
Caution: If tions can result. Warning: If this warning is ignored, injury to persons and/or serious damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or destruction of the instrument can result.
this warning is ignored, faults or malfunc-
te
applications
Ex
This symbol indicates special instructions for Ex applications.
l List
The dot set in front indicates a list with no implied sequence.
à Action
This arrow
1 Sequence
Numbers set in front indicate successive steps in a procedure.
4 VE
indicates a single action.
32762-EN-080716
GAFLEX 66 4 20 mA/HART
2 For
your safety
2 For
your safety

2.1 Authorised personnel

All
operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protection equipment must always be worn.

2.2 Appropriate use

VEGAFLEX 66 is a sensor for continuous level measurement.
You can nd detailed information on the application range in
chapter "Product description".
Operational reliability is ensured only if the instrument is properly used according to the specications in the operating instructions manual as well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modications are explicitly forbidden.
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR

2.3 Warning about misuse

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

This is a high-tech instrument requiring the strict observance of standard regulations and guidelines. The user must take note of the safety instructions in this operating instructions manual, the country-specic installation standards as well as all prevailing safety regulations and accident prevention rules.
The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for trouble­free operation of the instrument.
T 5
2 For
your safety
the entire duration of use, the user is obliged to
During determine the compliance of the required occupational safety measures with the current valid rules and regulations and also take note of new regulations.

2.5 Safety approval markings and safety tips

The safety approval markings and safety tips on the device must be observed.

2.6 CE conformity

The protection goals of the EMC Directive 2004/108/EC (EMC) and the Low Voltage Directive 2006/95/EC (LVD) are fullled.
Conformity has been judged according to the following standards:
EMC: EN 61326-1: 2006
(electrical instruments for control technology and laboratory use - EMC requirements)
l Emission: Class B l Susceptibil
ity: Industrial
LVD: EN 61010-1: 2001
(safety regulations for electrical measurement, control and laboratory instruments - part 1: General requirements)
areas
2.7 Fullling of NAMUR recommendations
With respect to interference resistance and emitted interfer­ence, the NAMUR recommendation NE 21 is fullled.
With respect to compatibility, the NAMUR recommendation NE 53 is fullled. This applies also to the corresponding
indicating and adjustment components. VEGA instruments are generally upward and downward compatible.
l Sensor l DTM VEGAFLEX 66 for adjustment software PACTware l Indicating and adjustment module for sensor software
software for DTM
The parameter adjustment of the basic sensor functions is
independent of the software version. The range of available functions depends on the respective software version of the individual components.
The software version of VEGAFLEX 66 can be determined as follows:
6 VE
VEGAFLEX 66 HART, PA or FF
32762-EN-080716
GAFLEX 66 4 20 mA/HART
your safety
2 For
l via PACTware
the type label of the electronics
l on l via the indicating and adjustment module
You can
view all software histories on our website www.vega. com. Make use of this advantage and get registered for update information via e-mail.

2.8 SIL conformity

VEGAFLEX 66 fullls the requirements for functional safety according to IEC 61508/IEC 61511. You can nd further information in the Safety Manual "VEGAFLEX series 60 -
4 20 mA/HART".

2.9 Safety instructions for Ex areas

Please note the Ex-specic safety information for installation and operation in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex­approved instruments.

2.10 Manufacturer declaration for zone 2

In conformity with DIN EN 60079-15/2005 VEGAFLEX 66 is suitable for use in zone 2.
The operator must use the instrument as it was intended to be used and follow the specications of the following documents:
l this operating instructions manual l the manufacturer declaration 32905 (download under
"www.vega.com")
applicable installation regulations
l the
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR

2.11 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental protection. The environment man­agement system is certied according to DIN EN ISO 14001.
Please help us full this obligation by observing the environ­mental instructions in this manual:
and
l Chapter "Packaging, transport l Chapter "Disposal"
T 7
storage"
3 Produc
t description
Scope of delivery
Components
3 Prod
3.1 Congurat
uct description
ion
The scope of delivery encompasses:
l VEGAFLEX 66 level sensor l Documentation
- this
operating instructions manual
- Safety Manual 31339 "VEGAFLEX series 60 -
4 20 mA/HART"
- Operating instructions manual 27835 "Indicating and
adjustment module PLICSCOM" (optional)
- Supplementary instructions manual 31708 "Heating for
indicating and adjustment module" (optional)
- Supplementary instructions manual "Plug connector for
continuously measuring sensors" (optional)
- Ex-specic "Safety instructions" (with Ex-versions)
VEGAFLEX 66 consists of the following components:
with
l Process tting l Housing with electronics l Housing cover, optionally available with indicating and
probe
adjustment module
8 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
1 2
3
3 Produc
t description
Type label
Application range
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
Fig. 1: VEGAFLEX 66 - cable 1 Housing cover with integrated indicating and adjustment module (optional) 2 Housing with electronics 3 Process tting
The type label contains the most important data for identi-
cation and use of the instrument:
l Article number l Serial number l Technical data l Article numbers, document
version with plastic housing
ation
With the serial number, you can access the delivery data of the instrument via "
www.vega.com", "VEGA
Tools" and "serial
number search". In addition to the type label outside, you can
also nd the serial number on the inside of the instrument.

3.2 Principle of operation

VEGAFLEX 66 is a level sensor with coax, rod or cable probe for continuous level measurement in non-conductive products with temperatures of -110 +400 °C (-166 +752 °F).
T 9
3 Produc
t description
is designed for industrial use in all areas of process
It technology and can be used in non-conductive liquids.
Functional principle
Voltage supply
High frequency microwave pulses are guided along a steel rope or a rod. Upon reaching the product surface, the microwave pulses are reected. The running time is evaluated by the instrument and outputted as distance.
Two-wire electronics 4 … 20 mA/HART for power supply and measured value transmission over the same cable.
The supply voltage range can dier depending on the instrument version.
Data for power supply are specied in chapter "Technical data".
The backlight of the indicating and adjustment module is powered by the sensor. The prerequisite for this is a supply voltage at a certain level. The exact voltage specications are stated in chapter "Technical data".
The optional heating requires its own power supply. You can nd further details in the supplementary instructions manual
"Heating for indicating and adjustment module".
This function is generally not available for approved instru­ments.

3.3 Operation

VEGAFLEX 66 can be adjusted with dierent adjustment media:
l with indicating and adjustment module l with the suitable VEGA
DTM in adjustment software according to the FDT/DTM standard, e.g. PACTware and PC
l with manufacturer-specic
PDM
l With a HART handheld
The entered parameters are generally saved in VEGAFLEX
66, optionally also in the indicating and adjustment module or in PACTware.
conjunction with an
adjustment
programs AMSor
32762-EN-080716
10 VE
GAFLEX 66 4 20 mA/HART
3 Produc
t description
Packaging
Transport
Transport inspection
Storage
Storage and transport
temperature
3.4 Pack
aging, transport and storage
Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN EN 24180.
The packaging of standard instruments consists of environ-
ment-friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instruc­tions can cause damage to the device.
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
l Not l Dry and dust free l Not exposed to corrosive media l P l Avoiding mechanical shock and vibration
l Storage and transport temperature see "Supplement -
l Relative humidity 20 85 %
in the open
rotected against solar radiation
Technical
data - Ambient conditions"
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 11
4 M
ounting
Mounting position
Welding work
Handling
Moisture
4 Mou
nting

4.1 General instructions

Select
an installation position you can easily reach for mounting and connecting as well as later retrotting of an indicating and adjustment module. The housing can be rotated by 330° without the use of any tools. You can also install the indicating and adjustment module in four dierent positions (each displaced by 90°).
Before beginning the welding work, remove the oscillator (electronics) from the sensor. By doing this, you avoid damage to the electronics through inductive coupling.
With threaded versions, the housing must not be used to screw in the instrument! Applying tightening forces on the housing can damage its internal parts.
Use the hexagon for screwing in.
Use the recommended cables (see chapter "Connecting to
power supply") and tighten the cable gland.
You can give your instrument additional protection against moisture penetration by leading the connection cable down­ward in front of the cable entry. Rain and condensation water can thus drain off. This applies mainly to outdoor mounting as well as installation in areas where high humidity is expected (e. g. through cleaning processes) or on cooled or heated vessels.
Measuring range
Fig. 2: Measures
The reference plane for the measuring range of the sensors is
against moisture penetration
the sealing surface of the thread or ange.
12 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
ounting
4 M
in mind that a min. distance must be maintained below
Keep the reference plane and possibly also at the end of the probe ­measurement in these areas is not possible (dead band). Keep in mind that the cable length cannot be used all the way to the end because measurement in the area of the gravity weight is not possible. These min. distances (dead bands) are listed in chapter "Technical data".
Pressure
Mounting position
The process tting must be sealed if there is gauge or low pressure in the vessel. Before use, check if the seal material is resistant against the measured product and the process temperature.
The max. permissible pressure is specied in chapter "Technical data" or on the type label of the sensor.

4.2 Mounting instructions

Mount VEGAFLEX 66 in such a way that the distance to vessel installations or to the vessel wall is at least 300 mm (12 in).
During operation, the probe must not touch any installations or the vessel wall. If necessary, fasten the probe end.
In vessels with conical bottom it can be advantageous to mount the sensor in the center of the vessel, as measurement is then possible down to the lowest point of the bottom. When using the cable version, keep in mind that measurement down to the tip of the probe is not possible. The exact value of the
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 13
4 M
ounting
Standpipes or bypass tubes
min. distance (lower
dead band) is stated in chapter "Technical
data".
Fig. 3: Vessel
with conical bottom
Standpipes or bypass tubes are normally metal tubes with a diameter of 30 … 200 mm (1.18 7.87 in). In measurement technology such a tube corresponds to a coax probe. It does not matter if the standpipe is perforated or slotted for better mixing. Lateral inlets with bypass tubes also do not inuence the measurement.
Measuring probes can be mounted in bypass tubes up to DN 200.
The max. temperature is 250 °C (482 °F).
If VEGAFLEX 66 is used in standpipes or bypass tubes,
contact with the tube wall should be avoided. We oer spacers as accessories for fastening the probe in the middle of the tube.
Depending on the tube diameter or tube length, one or several spacers can be mounted. With cable probes, the cable can also be strained to avoid contact with the tube.
The max. temperature for the spacers is 250 °C (482 °F).
A special high temperature version of VEGAFLEX 66 with
integrated bypass tube can be used in applications with temperatures up to 400 °C (752 °F).
32762-EN-080716
14 VE
GAFLEX 66 4 20 mA/HART
in mind that buildup can form on the spacers. Strong
1
Keep buildup can inuence the measurement.
4 M
ounting
Inowing medium
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
Fig. 4: Position 1 Spacer
of the spacer
Note:
Measureme
nt in a standpipe is not recommended for very
adhesive products.
Make sure that the probe is not subjected to strong lateral forces. Mount VEGAFLEX 66 at a position in the vessel where no disturbances, e.g. from lling openings, agitators, etc., can occur.
T 15
4 M
ounting
Fixing
Heat insulation
Fig. 5: Lateral
load
Excessive system vibration or shocks, e.g. caused by agitators or turbulence in the vessel (e.g. from uidisation) can cause the coax probe of VEGAFLEX 66 to vibrate in resonance. Should a coax probe of more than 1 m (3.281 in) length should be used, you can provide a suitable support or guy directly above the end of the probe to stabilise it.
If there is a danger of the probe touching the vessel wall during operation due to product movements or agitators etc., the measuring probe should be securely xed.
In the gravity weight there is a thread (M12), e.g. for a ring bolt (article no. 2.27424).
Make sure that the probe cable is not completely taut. Avoid tensile loads on the cable.
Avoid undened cable-vessel connections, i.e. the connection must be either grounded reliably or isolated reliably. Any deviation from this requirement can lead to measurement errors.
VEGAFLEX 66 should be integrated in the vessel installation. This prevents the electronics from strong heating by heat
radiation.
32762-EN-080716
16 VE
GAFLEX 66 4 20 mA/HART
1
2
ounting
4 M
sure that with heated vessels the permissible ambient
Make temperature on the housing is not exceeded. The permissible ambient temperature is stated in chapter "Technical data" under "Ambient conditions".
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
Fig. 6: Vessel 1 Heat insulation 2 Ambient temperature on the housing
with heat insulation
T 17
ecting to power supply
5 Conn
Note safety instructions
Take note of
safety instruc­tions for Ex ap­plications
Select power supply
Selecting connection
cable
5 Con
5.1 Preparin
necting to power supply
g the connection
Always keep in mind the following safety instructions:
l Connect only in the complete absence of line voltage l If overvoltage surges are expected, overvoltage
arresters
should be installed
Tip:
We
recommend using VEGA overvoltage arresters B63-48
and ÜSB 62-36G.X.
In
hazardous areas you should take note of the appropriate regulations, conformity and type approval certicates of the sensors and power supply units.
Power supply and current signal are carried on the same two­wire cable. The voltage supply range can dier depending on the instrument version.
Data for power supply are specied in chapter "Technical data".
Provide a reliable separation between the supply circuit and the mains circuits according to DIN VDE 0106 part 101. The
VEGA power supply units VEGATRENN 149A Ex, VEGAS­TAB 690 as well as all VEGAMETs meet this requirement.
Bear in mind the following factors regarding supply voltage:
l Outpu
t voltage of the power supply unit can be lower under
nominal
load (with a sensor current of 20.5 mA or 22 mA in
case of fault message)
l Inuence
of additional instruments in the circuit (see load
values in chapter "Technical data")
The instrument is connected with standard two-wire cable without screen. If electromagnetic interference is expected which is above the test values of EN 61326 for industrial areas, screened cable should be used.
Use cable with round cross-section. A cable outer diameter of 5 9 mm (0.2 0.35 in) ensures the seal eect of the cable gland. If you are using cable with other diameter or cross­section, you have to exchange the seal or use a suitable cable gland.
32762-EN-080716
18 VE
GAFLEX 66 4 20 mA/HART
ecting to power supply
5 Conn
generally recommend the use of screened cable for HART
We multidrop mode.
Cable gland ½ NP T
Cable screening and
grounding
Select connec­tion cable for Ex applications
On the instrument with cable entry ½ NPT and plastic housing there is a metallic ½" threaded insert moulded into the plastic housing.
Caution:
No
grease should be used when screwing the NPT cable gland or steel tube into the threaded insert. Standard grease can contain additives that corrode the connection between threaded insert and housing. This would inuence the stability of the connection and the tightness of the housing.
If screened cable is necessary, connect the cable screen on both ends to ground potential. In the sensor the screen must be connected directly to the internal ground terminal. The ground terminal on the outside of the housing must be connected to the potential equalisation (low impedance).
If potential equalisation currents are expected, the connection on the processing side must be made via a ceramic capacitor (e. g. 1 nF, 1500 V). The low frequency potential equalisation currents are thus suppressed, but the protective eect against high frequency interference signals remains.
note of the corresponding installation regulations for Ex
Take applications. In particular, make sure that no potential equal­isation currents ow over the cable screen. In case of grounding on both sides this can be achieved by the use of a capacitor or a separate potential equalisation.
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
5.2 Connection steps - Instrument housing
Proceed as follows:
1 Unscrew the housing cover 2 If an indicating and adjustment module is installed, remove
it by turning it slightly to the left.
3 Loosen compression nut of the cable entry 4 Remove approx. 10 cm (4 in) of the cable mantle, strip
approx. 1 cm (0.4 in) insulation from the ends of the
individual wires
5 Insert the cable through the cable gland into the sensor
T 19
ecting to power supply
5 Conn
the opening levers of the terminals with a screwdriver
6 Lift
(see following illustration)
7 Insert the wire ends into the open terminals according to
the wiring plan
Fig. 7: Connection
steps 6 and 7
8 Press down the opening levers of the terminals, you will
hear the terminal spring closing
9 Check the hold of the wires in the terminals by lightly
pulling on them
10 Connect the screen to the internal ground terminal, connect
the outer ground terminal with potential equalisation
11 Tighten the compression nut of the cable entry. The seal
ring must completely encircle the cable
12 Screw the housing cover on The electrical connection is nished.

5.3 Wiring plan, single chamber housing

following illustrations apply to the non-Ex as well as to the
The Ex-ia version.
20 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
5
5
5
5
1 2
4
3
I²C
Display
1 2 5 6 7 8
3
4
1
2
Housing overview
ecting to power supply
5 Conn
Electronics and connec­tion compartment
Fig. 8: Material 1 Plastic 2 Aluminium 3 Stainless steel, investment casting 4 Stainless steel, electro-polished 5 Filter element for air pressure compensation of all material versions. Blind
stopper with version IP 66/IP 68, 1 bar for Aluminium and stainless steel
Fig. 9: Electronics
1 Plug connector for VEGACONNECT (I²C interface) 2 Spring-loaded terminals for connection of the external indication VEGADIS
61 3 Ground terminal for connection of the cable screen 4 Spring-loaded terminals for voltage supply
versions, single chamber housing
and connection compartment, single chamber housing
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 21
I2C
Display
1
1 2 5
6 7 8
1 2 3
4
5
ecting to power supply
5 Conn
Wiring plan
Housing overview
Fig. 10: Wiring 1 Voltage supply/Signal output
plan, single chamber housing

5.4 Wiring plan, double chamber housing

The
following illustration apply to non-Ex as well as Ex ia
versions. The Exd version is described in the next subchapter.
Fig. 11: Double 1 Housing cover, connection compartment 2 Blind stopper or plug M12 x 1 for VEGADIS 61 (optional) 3 Housing cover, electronics compartment 4 Filter element for air pressure compensation 5 Cable gland
chamber housing
22 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
1
3
2
Display
1 2 5 6 7 8
I2C
3
1
2
Display
1 2
I²C
Electronics compart­ment
ecting to power supply
5 Conn
Connection compart­ment
Fig. 12: Electronics 1 Plug connector for VEGACONNECT (I²C interface) 2 Internal connection cable to the connection compartment 3 Terminals for VEGADIS 61
Fig. 13: Connection 1 Plug connector for VEGACONNECT (I²C interface) 2 Ground terminal for connection of the cable screen 3 Spring-loaded terminals for voltage supply
compartment, double chamber housing
compartment, double chamber housing
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 23
I2C
1
1 2
1 2 3
4
5
ecting to power supply
5 Conn
Wiring plan
Housing overview
Fig. 14: Wiring 1 Voltage supply/Signal output
plan, double chamber housing
5.5 Wiring plan double chamber housing Ex d
Fig. 15: Double 1 Housing cover, connection compartment 2 Blind stopper or plug M12 x 1 for VEGADIS 61 (optional) 3 Housing cover, electronics compartment 4 Filter element for air pressure compensation 5 Cable gland
chamber housing
24 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
1
3
2
Display
1 2 5 6 7 8
I2C
1
2
1 2
Electronics compart­ment
ecting to power supply
5 Conn
Connection compart­ment
Fig. 16: Electronics 1 Plug connector for VEGACONNECT (I²C interface) 2 Internal connection cable to the connection compartment 3 Terminals for VEGADIS 61
Fig. 17: Connection 1 Spring-loaded terminals for power supply and cable screen 2 Ground terminal for connection of the cable screen
compartment, double chamber housing
compartment double chamber housing Ex d
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 25
1
1 2
+
-
1
2
ecting to power supply
5 Conn
Wiring plan
Wire assignment, con-
nection cable
Fig. 18: Wiring 1 Voltage supply/Signal output
plan double chamber housing Ex d
5.6 Wiring plan - version IP 66/IP 68, 1 bar
Fig. 19: Wire 1 brown (+) and blue (-) to power supply or to the processing system 2 Shielding
assignment, connection cable
26 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
up with the indicating and adjustment module PLICSCOM
6 Set
Function/Conguration
6 Set
up with the indicating and
adjustment module PLICSCOM

6.1 Short description

The
indicating and adjustment module is used for measured value display, adjustment and diagnosis. It can be mounted in the following housing versions and instruments:
sensors of the plics
l All
well as in the double chamber housing (optionally in the electronics or connection compartment)
l External
From
indicating and adjustment unit VEGADIS 61
a hardware version …- 01 or higher of the indicating and adjustment module as well as of the corresponding sensor, an integrated backlight can be switched on via the adjustment menu. The hardware version is stated on the type label of the indicating and adjustment module or the sensor electronics.
Information:
This
function is available at a later date for instruments with StEx, WHG or ship approval as well as country-specic approvals such as those according to FM or CSA.
Note:
can nd detailed information on the adjustment in the
You operating instructions manual "Indicating and adjustment
module".
®
instrument family, in the single as
Mount/Dismount indicat­ing and adjustment module
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR

6.2 Insert indicating and adjustment module

The indicating and adjustment module can be inserted into the sensor and removed again at any time. It is not necessary to interrupt the power supply.
Proceed as follows:
1 Unscrew the housing cover 2 Place the indicating and adjustment module in the desired
position on the electronics (you can choose any one of four dierent positions - each displaced by 90°)
3 Press the indicating and adjustment module onto the
electronics and turn it to the right until it snaps in.
4 Screw housing cover with inspection window tightly back
on
T 27
up with the indicating and adjustment module PLICSCOM
6 Set
Removal
is carried out in reverse order.
The indicating and adjustment module is powered by the sensor, an additional connection is not necessary.
Fig. 20: Installation
of the indicating and adjustment module
Note:
you intend to retrot the instrument with an indicating and
If adjustment module for continuous measured value indication, a higher cover with an inspection glass is required.
28 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
1.1
2
3
1
up with the indicating and adjustment module PLICSCOM
6 Set
Key functions

Adjustment system

6.3 Adj
Fig. 21: Indicating 1 LC display 2 Indication of the menu item number 3 Adjustment keys
l [OK] key:
ustment system
and adjustment elements
- Move
to
the menu overview
- Conrm selected menu
- Edit parameter
- Save
l [->] key to select:
- menu
value
change
- list entry
- Select
l [+] key:
editing position
- Change value of the parameter
l [ESC] key:
- interrupt input
- jump
The sensor is adjusted via the four keys of the indicating and adjustment module. The LC display indicates the individual menu items. The functions of the individual keys are shown in the above illustration. Approx. 10 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not conrmed with [OK] will not be saved.
to the next higher menu
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 29
up with the indicating and adjustment module PLICSCOM
6 Set
Switch on phase
Address setting HART­Multidrop
Parameter adjustment
6.4 Setup
procedure
After connecting VEGAFLEX 66 to power supply or after a voltage recurrence, the instrument carries out a self-check for approx. 30 seconds:
l Internal l Indication of the instrument type, the rmware
check of the electronics
as well as
the sensor TAGs (sensor designation)
l Output signal jumps briey (approx. 10 seconds) to the set
fault current
Then the corresponding current is outputted to the cable (the value corresponds to the actual level as well as the settings already carried out, e.g. factory setting).
In HART-Multidrop mode (several sensors on one input) the address must be set before continuing with the parameter adjustment. You will nd a detailed description in the operating instructions manual "Indicating and adjustment module" or in the online help of PACTware or DTM.
HART mode
Standard
Address 0
As VEGAFLEX 66 is a distance measuring instrument, the distance from the sensor to the product surface is measured. To have the real product level displayed, an allocation of the measured distance to the percentage height must be made. To carry out this adjustment, the distance is entered with full and empty vessel. If these values are not known, an adjustment with the distance values, e.g. 10 % and 90 % is also possible. Starting point for these distance specications is always the seal surface of the thread or ange. With these settings, the real level is calculated. Furthermore the operating range of the sensor is limited from maximum to the required range.
The real product level during this adjustment is not important, because the min./max. adjustment is always carried out without changing the product level. These settings can be made ahead of time without the instrument having to be installed.
In the main menu item "Basic adjustment", the individual submenu items should be selected one after the other and provided with the correct parameter values.
32762-EN-080716
30 VE
GAFLEX 66 4 20 mA/HART
up with the indicating and adjustment module PLICSCOM
6 Set
Caution:
there is a separation of dierent liquids in the vessel, e.g. by
If condensation, VEGAFLEX 66 will always detect the medium with the higher dielectric gure
).
r
Keep in mind that interfaces can cause faulty measurements.
If you want to measure the total height of both liquids reliably,
please contact our service department or use an instrument specially designed for interface measurement.
Start your parameter adjustment with the following menu items of the basic adjustment:
Carrying out min. ad­justment
Carrying out max. ad­justment
Proceed as follows:
1 Move from the measured value display to the main menu
by pushing [OK].
Basic adjustment Display Diagnostics Service Info
2 Select the menu item "Basic adjustment" with [->] and
conrm with [OK]. Now the menu item "Min. adjustment" is displayed.
Min. adjustment
0.00 %
=
10.000 m(d)
8.000 m(d)
3 Prepare the % value for editing with [OK] and set the
cursor to the requested position with [->]. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
4 Enter the suitable distance value in m for the empty vessel
(e.g. distance from the sensor to the vessel bottom) corresponding to the percentage value.
5 Save the settings with [OK] and move to "Max. adjustment"
with [->].
Proceed as follows:
Max. adjustment
100.00 %
=
1.000 m(d)
2.000 m(d)
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 31
up with the indicating and adjustment module PLICSCOM
6 Set
1 Prepare the % value for editing with [OK] and set the
cursor to the requested position with [->]. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
2 Enter the appropriate distance value in m (corresponding
to the percentage value) for the full vessel. Keep in mind that the max. level must lie below the dead band.
3 Save the settings with [OK].
Application
Each product has dierent reective properties. In addition, there are various interfering factors which have to be taken into account: agitated product surfaces and foam generation (with liquids); dust generation, material cones and echoes from the vessel wall (with solids). To adapt the sensor to these dierent conditions, you should rst select in this menu item under "Medium" the selection "Liquid" or "Solid".
With coax versions, this menu item must be set to "Liquid".
Application
Liquid
Standard (DK 2)
Depending on the dielectric gure (dielectri value or εr), measured products can have a dierent reective property.
Therefore an additional selection possibility is available.
Under "Sensitivity" you can select "Standard (DK 2)" or
"Increased sensitivity (DK < 2)".
Through this, the sensor is adapted perfectly to the product and measurement reliability, particularly in products with bad reective properties, is considerably increased.
Enter the requested parameter via the appropriate keys, save your settings and jump to the next menu item with the [->] key.
Damping
To suppress uctuations in the measured value display, e. g. caused by an agitated product surface, an integration time can be set. This time can be between 0 and 999 seconds. Keep in mind that the reaction time of the entire measurement will then be longer and the sensor will react to measured value changes with a delay. In general, a period of a few seconds is sucient to smooth the measured value display.
32 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
up with the indicating and adjustment module PLICSCOM
6 Set
0 s
Enter the requested parameter via the appropriate keys, save your settings and jump to the next menu item with the [->] key.
Linearisation curve
A linearization is necessary for all vessels in which the vessel volume does not increase linearly with the level - e. g. with a cylindrical or spherical tank - and the indication or output of the volume is required. Corresponding linearization curves are preprogrammed for these vessels. They represent the correlation between the level percentage and vessel volume. By activating the appropriate curve, the volume percentage of the vessel is displayed correctly. If the volume should not be displayed in percent but e.g. in l or kg, a scaling can be also set in the menu item "Display".
Linearisation curve
linear
Enter the requested parameter via the appropriate keys, save your settings and jump to the next menu item with the [->] key.
Caution:
Note
the following, if VEGAFLEX 66 is used as part of an
overll protection system according to WHG:
If a linearisation curve is selected, the measuring signal is no longer compulsorily linear proportional to the level. This must be taken into consideration by the user, particularly when adjusting the switching point on the level switch.
Sensor-TAG
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
In this menu item you can enter an unambiguous designation
for the sensor, e.g. the measurement loop name or the tank or product designation. In digital systems and in the documen­tation of larger plants, a singular designation should be entered for exact identication of individual measuring sites.
Sensor-TAG
Sensor
T 33
up with the indicating and adjustment module PLICSCOM
6 Set
this menu item, the Basic adjustment is nished and you
With can now jump to the main menu with the [ESC] key.
Gatin
g out of false sig-
nals
High sockets or vessel installations, such as e. g. struts or agitators as well as buildup and weld joints on the vessel walls cause interfering reections which can impair the measure­ment. A false echo storage detects and marks these false echoes, so that they are no longer taken into account for the level measurement. A false echo memory should be created with empty vessel so that all potential interfering reections will be detected.
Probes in coax version require no gating out of false echoes since they are not inuenced by false reections.
Gating out of false signals
Change now?
Proceed as follows:
1 Move from the measured value display to the main menu
by pushing [OK].
2 Select the menu item "Service" with [->] and conrm with
[OK]. Now the menu item "False signal suppression" is
displayed.
3 Conrm "False signal suppression - Change now" with
[OK] and select in the below menu "Create new". Enter the actual distance from the sensor to the product surface. All false signals in this area are detected by the sensor and saved after conrming with [OK].
Note:
the distance to the product surface, because if an
Check incorrect (too large) value is entered, the existing level will be saved as false signal. The lling level would then no longer be detectable in this area.
Copy sensor data
This function enables reading out parameter adjustment data as well as writing parameter adjustment data into the sensor via the indicating and adjustment module. A description of the function is available in the operating instructions manual
"Indicating and adjustment module".
The following data are read out or written with this function:
l Measured
34 VE
value presentation
32762-EN-080716
GAFLEX 66 4 20 mA/HART
up with the indicating and adjustment module PLICSCOM
6 Set
l Adjustment l Medium l Vessel form l Damping l Linearisation curve l Sensor-TAG l Displayed value l Display unit l Scaling l Current output l Unit of measurement l Language l Sensitivity
The following
l HART mode l PIN l SIL l Sensor length/Sens l Gating out of false signals
Copy sensor data
Copy sensor data?
safety-relevant data are not read out or written:
or type
Reset
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
Basic adjustment
If the function "Reset" is carried out, the sensor resets the values of the following menu items to the reset values (see
1)
chart):
following values will be reset:
The
Function Reset
Max. adjustment Distance, upper dead zone Min. adjustment - Rod/Coax version Distance, supplied sensor length Min. adjustment - Cable version Distance, lower dead zone Integration time ti 0 s Linearisation linear Sensor-TAG Sensor Display Distance Current output - characteristics 4 … 20 mA Current output - max. current 20 mA Current output - min. current 4 mA
1)
Sensor-specic basic adjustment.
T 35
value
up with the indicating and adjustment module PLICSCOM
6 Set
Optional settings
Function Reset
Current output - failure < 3.6 mA Application - rod/coax version Liquid Application - Cable version Solid
value
The values of the following menu items are not reset to the reset values (see chart) with "Reset":
Menu item Reset value
Lighting no reset Language no reset SIL no reset HART mode no reset
Factory setting
Like basic adjustment, furthermore special parameters are reset to default values.
r
Pointe
The
min. and max. values are reset to the actual value.
2)
Additional adjustment and diagnosis options such as e.g. scaling, simulation or trend curve presentation are shown in the following menu schematic. You will nd a detailed description of these menu items in the operating instructions manual "Indicating and adjustment module".
2)
Special parameters are parameters which are set customer-specically on the service level with the adjustment software PACTware.
36 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
up with the indicating and adjustment module PLICSCOM
6 Set
Basic adjustment
Basic adjustment Display Diagnostics Service Info
Min. adjustment
0.00 %
=
10.000 m(d)
8.000 m(d)
Sensor-TAG
Sensor
Display
Basic adjustment
Display Diagnostics Service Info
Displayed value
Scaled
1.1
1.5
2.1
1
2
6.5 Menu
Max. adjustment
100.00 %
=
1.000 m(d)
2.000 m(d)
Unit
Volume
hl
schematic
1.2
2.2
Damping
Scaling
0 s
0 % = 000.5 hl
100 % = 005.0 hl
1.3
2.3
Linearisation curve
linear
Lighting
Switched on
1.4
2.4
Diagnostics
Basic adjustment Display
Diagnostics Service Info
Pointer
Distance min.: 0.234 m(d) Distance max.: 5.385 m(d)
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
3
3.1
Sensor status
OK
3.2
Curve selection
Echo curve
3.3
Echo curve
Presentation of the echo
T 37
3.4
curve
up with the indicating and adjustment module PLICSCOM
6 Set
Service
Basic adjustment Display Diagnostics
Service Info
4
Sensor
5.00 m(d)
Rod
Simulation
HART mode
Info
Basic adjustment Display Diagnostics Service
Info
Sensor type
Serial number
12345678
Start simulation?
Standard
Address 0
4.1
4.5
4.10
5.1
5
Application
Liquid
Standard
(DK 2)
Reset
Select reset?
Copy sensor data
Copy sensor data?
Date of manufacture
12. Dec. 2005
Software version
3.22
4.2
4.6
4.11
5.2
Gating out of false signals
Change now?
Unit of measurement
m(d)
select?
PIN
Enable?
Last change using PC
04. March 2004
4.3
4.7
4.12
5.3
Current output
Output mode: 4-20 mA Failure mode: 20.5 mA
min. current: 4 mA max. current: 20.5 mA
Language
Deutsch
Sensor characteristics
Display now?
4.4
▼ ▼
4.8
5.4
38 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
~
=
Power supply
VEGACONNECT 3
TM
>PA<
2
3
1
7 Setup
with PACTware and other adjustment programs
Connection via I²C inter­face
7 Setup
with PACTware and other
adjustment programs
7.1 Connect
Fig. 22: Connection 1 RS232 connection 2 VEGAFLEX 66 3 I²C adapter cable for VEGACONNECT 3
Necessary components:
l VEGAFLEX 66 l PC with PACTware and l VEGACONNECT 3 with I²C adapter cable (article no.
2.27323)
l Power supply unit
ing the PC
of the PC directly to the sensor via I²C interface
suitable VEGA DTM
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 39
2
3
1
4
~
=
Power supply
VEGACONNECT 3
TM
7 Setup
Connection via HART
with PACTware and other adjustment programs
Fig. 23: Connecting 1 RS232 connection 2 VEGAFLEX 66 3 HART adapter cable for VEGACONNECT 3 4 HART resistor 250 Ω
the PC via HART to the signal cable
Necessary components:
l VEGAFLEX 66
and
l PC with PACTware l VEGACONNECT 3 with HART adapter
suitable VEGA DTM
cable (art. no.
2.25397)
l HART resistor approx. 250 Ω l Power supply unit
Note:
power supply units with integrated HART resistance
With
(internal resistance approx. 250 Ω), an additional external resistance is not necessary. This applies, e. g. to the VEGA instruments VEGATRENN 149A, VEGADIS 371, VEGAMET
381). Also usual Ex separators are most of the time equipped with a sucient current limitation resistor. In such cases, VEGACONNECT 3 can be connectedparallel to the 4 20 mA cable.

7.2 Parameter adjustment with PACTware

Further setup steps are described in the operating instructions manual "DTM Collection/PACTware" attached to each CD and
32762-EN-080716
which can also be downloaded from our homepage. A detailed description is available in the online help of PACTware and the VEGA DTMs.
40 VE
GAFLEX 66 4 20 mA/HART
7 Setup
with PACTware and other adjustment programs
Note:
in mind that for setup of VEGAFLEX 66, DTM-Collection
Keep
in the actual version must be used.
All currently available VEGA DTMs are provided in the DTM Collection on CD and can be obtained from the responsible VEGA agency for a token fee. This CD includes also the up-to- date PACTware version. The basic version of this DTM Collection incl. PACTware is also available as a free-of-charge download from the Internet.
Go via
www.vega.com
and "Downloads" to the item "Soft-
ware".

7.3 Parameter adjustment with AMSand PDM

For VEGA sensors, instrument descriptions for the adjustment programs AMSand PDM are available as DD or EDD. The instrument descriptions are already implemented in the current versions of AMSand PDM. For older versions of AMSand
PDM, a free-of-charge download is available via Internet.
Go via
www.vega.com
and "Downloads" to the item "Soft-
ware".

7.4 Saving the parameter adjustment data

It is recommended to document or save the parameter adjustment data. They are hence available for multiple use or service purposes.
The VEGA DTM Collection and PACTware in the licensed, professional version provide suitable tools for systematic project documentation and storage.
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 41
aintenance and fault rectication
8 M
Reaction when malfunc­tions occur
Causes of malfunction
Fault rectication
8 Main
tenance and fault rectication

8.1 Maintenance

When
used in the correct way, no special maintenance is
required in normal operation.

8.2 Remove interferences

The operator of the system is responsible for taken suitable measures to remove interferences.
A maximum of reliability is ensured. Nevertheless, faults can occur during operation. These may be caused by the following, e.g.:
l Sensor l Process l Voltage supply l Signal processing
The rst as well as to evaluate the error messages via the indicating and adjustment module. The procedure is described below. Further comprehensive diagnostics can be carried out on a PC with the software PACTware and the suitable DTM. In many cases, the causes can be determined in this way and faults can be rectied.
measures to be taken are to check the output signals
24 hour service hotline
However, should these measures not be successful, call the VEGA service hotline in urgent cases under the phone no. +49
1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since we oer this service world-wide, the support is only
available in the English language. The service is free of charge, only the standard telephone costs will be charged.
Checking the 4 20 mA signal
Connect a handheld multimeter in the suitable measuring range according to the wiring plan.
? 4 20 mA signal not stable
l Level uctuations
à Set
integration
time via the indicating/adjustment
module
42 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
aintenance and fault rectication
8 M
Fault messages via the indicating/adjustment module
? 4 20 mA signal
l Wrong connection
à Check
connection according to chapter "Connection
missing
steps" and if necessary, correct according to chapter "Wiring plan"
voltage supply
l No
à Check
l supply voltage too low or load resistance too high
cables for breaks; repair if necessary
à Check, adapt if necessary
? Current signal greater than 22 mA or less than 3.6 mA
l Electronics
module
defective
à Exchange instrument or return instrument for repair
In Ex
applications, the regulations for the wiring of intrinsically
safe circuits must be observed.
? E013
l no measured value available
in
à sensor
boot phase
à Sensor does not nd an echo, e.g. due to faulty
installation or wrong parameter adjustment
à Wrong sensor length entered
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
? E017
l Adjustment span
too small
à Carry out a fresh adjustment and increase the distance
between min. and max. adjustment
? E036
l no operable sensor software
à Carry out
a software update or send the instrument for
repair
? E042/E043
l Hardware error, electro
nics defective
à Exchange instrument or return instrument for repair
T 43
aintenance and fault rectication
8 M
Reaction after fault rec­tication
Preparations
Dependin
g on the failure reason and measures taken, the steps described in chapter "Set up" must be carried out again, if necessary.

8.3 Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user.
In Ex
applications only one oscillator with respective Ex
approval may be used.
If there is no electronics module available on site, one can be ordered from the agency serving you.
The new oscillator must contain the order data of the sensor. These can be loaded as follows:
the factory by VEGA
l At l Or on site by the user
Information:
When
loading on site, rst of all the order data must be downloaded from the Internet (see operating instructions manual "Oscillator").
In both cases, the serial number of VEGAFLEX 66 is required. The serial numbers are stated on the type label of VEGAFLEX
66, on the inner wall of the housing or on the delivery note.
Assignment
The oscillators are adapted to the respective sensor and dier in their signal output or in their power supply. You can nd a suitable oscillator in the following overview.
4 20 mA/HART
Electronics module FX-E.60H suitable for VEGAFLEX 61, 62, 63, 65, 66 - 4 … 20mA/HART:
l FX-E.60HX (X = without l FX-E.60HA (A = approvals CA, DA acc
approvals)
list)
l FX-E.60HC (C = approvals XM, CX, CM, CK, CI, DX, DM,
DI, GX, UX, UF according

8.4 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our Internet
homepage
44 VE
www.vega.com
under: "Downloads - Forms and
ording to product
to product list)
32762-EN-080716
GAFLEX 66 4 20 mA/HART
aintenance and fault rectication
8 M
Return of rod versions
certicates - Repair
form".
By doing this you help us carry out the repair quickly and without having to call back for needed information.
l Print and ll l Clean the instrument and pack it damage-proof l Attach the completed form and probably the safety data
sheet
l Please ask the agency serving you for the address of your
out one form per instrument
outside on the packaging
return shipment. You can nd the respective agency on our website
www.vega.com
under: "Company - VEGA world-
wide"
On instruments with exchangeable rod, the rod must be unscrewed for transport to avoid damages.
Return the parts separately for repair.
To loosen to rod, you require a fork spanner with wrench size
8.
1 Loosen the rod on the at surfaces with a fork spanner
(SW 8), provide counterforce with another fork spanner (SW 8)
2 Twist othe loosened rod manually
See also chapter "Maintenance and fault rectication"/
"Exchange cable/rod"
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 45
9 Dismoun
ting
9 Dism
ounting

9.1 Dismounting steps

Warning:
Before dismounting, be aware of dangerous process con­ditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.

9.2 Disposal

The instrument consists of materials which can be recycled by specialised recycling companies. We use recyclable materials and have designed the electronics to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/ EG and the respective national laws. Pass the instrument
directly on to a specialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
Correct disposal avoids negative eects to persons and environment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no possibility to dispose of the old instrument
professionally, please contact us concerning return and disposal.
46 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
10 Sup
plement
10 Sup
plement

10.1 Technical data

General
Material 316L corresponds to 1.4404 or 1.4435 Materials, wetted parts
- Process tting - coax version 316L and Aluminium oxide-ceramic 99.7 %
- Process tting - rod version 316L and Aluminium oxide-ceramic 99.7 %
- Process tting - cable version 316L and Aluminium oxide-ceramic 99.7 %
- Tube: ø 21.3 mm (0.839 in) 316L or Hastelloy C22 (2.4602)
- Rod: ø 6 mm (0.236 in) 316L or Hastelloy C22 (2.4602)
- Cable: ø 4 mm (0.157 in) 316 (1.4401)
- Process seal on the instrument side
- Process seal On site (instruments with thread: Klingersil
Materials, non-wetted parts
- Plastic housing plastic PBT (Polyester)
- Aluminium die-casting housing Aluminium die-casting AlSi10Mg, powder-
- Stainless steel housing - precision
- Stainless steel housing, electropol-
- Seal between housing and housing
- Inspection window in housing cover
- Ground terminal 316L
data
(cable/rod leadthrough)
casting
ished
cover
(optional)
(AlO), Hastelloy C22 (2.4602) and Alumi­nium oxide-ceramic 99.7 % (AlO)
(AlO), Hastelloy C22 (2.4602) and Alumi­nium oxide-ceramic 99.7 % (AlO)
(AlO)
graphite
C-4400 is attached)
coated - basis: Polyester
316L
316L
NBR (stainless steel housing, investment casting), silicone (Aluminium/plastic housing, stainless steel housing, electro-polished)
Polycarbonate
Process ttings
- Pipe thread, cylindrical (ISO 228 T1) G¾ A, G1 A, G1½ A
- American pipe thread, tapered ¾ NPT, 1 NPT, 1½ NPT
- Flanges DIN from DN 25, ANSI from 1"
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 47
10 Suppl
ement
Weight
- Instrument
weight (depending on
approx. 0.8 8 kg (0.176 17.64 lbs)
process tting)
- Tube: ø 21.3 mm (0.839 in) approx. 920 g/m (9.9 oz/ft)
- Rod: ø 6 mm (0.236 in) approx. 220 g/m (2.365 oz/ft)
- Cable: ø 4 mm (0.157 in) approx. 80 g/m (0.86 oz/ft)
- Gravity weight approx. 325 g (11.5 oz)
Sensor length L (from seal surface)
- Tube: ø 21.3 mm (0.839 in) 0.3 6 m (0.984 19.69 ft)
- Trimming accuracy - tube < 1 mm (0.039 in)
- Rod: ø 6 mm (0.236 in) 0.3 4 m (0.984 13.12 ft)
- Trimming accuracy - rod < 1 mm (0.039 in)
- Cable: ø 4 mm (0.157 in) 1 32 m (3.281 105 ft)
- Trimming accuracy - cable ±0.05 %
Lateral load
- Tube: ø 21.3 mm (0.839 in) 60 Nm (44 lbf ft)
- Rod: ø 6 mm (0.236 in) 4 Nm (3 lbf ft)
Max. tensile load with cable: ø 4 mm
2.5 KN (562 lbf)
(0.157 in)
Input variable
Measured value Level of non-conductive liquids (no steam
pressure applications)
Min. dielectric gure of the medium -
ε
> 1.4
r
coax version
48 VE
GAFLEX 66 4 20 mA/HART
32762-EN-080716
band - coax version
2
4
3
1
Dead
- top 30 mm (1.181 in)
- bottom 0 mm
10 Sup
plement
Fig. 24: Measuring 1 Reference plane
2 Probe length 3 Measuring range 4 Upper dead band
Min. dielectric gure of the medium ­rod, cable version
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 49
range of VEGAFLEX 66 - coax version
ε
> 1.7
r
2
4
5
3
1
2
4
3
1
10 Suppl
Dead
ement
band - rod version
- top 80 mm (3.15 in)
- bottom 0 mm
Dead band - cable version
- top 150 mm (5.91 in)
- bottom 250 mm (9.843 in) gravity weight + 100 mm
Fig. 25: Measuring 1 Reference plane
2 Probe length 3 Measuring range 4 Upper dead band 5 Lower dead band (only with cable version)
50 VE
ranges of the VEGAFLEX 66 - rod and cable version
32762-EN-080716
GAFLEX 66 4 20 mA/HART
10 Sup
plement
Output
variable
Output signal 4 20 mA/HART HART output values
- HART value (Primary Value) Distance to the level
- HART value (Secondary Value) Distance to the level - scaled
Signal resolution 1.6 µA Failure signal current output (adjust-
able)
mA-value unchanged 20.5 mA, 22 mA,
< 3.6 mA (adjustable) Max. output current 22 mA Load see load diagram under Power supply Damping (63 % of the input variable) 0 999 s, adjustable Fullled NAMUR recommendations NE 43
Accuracy (similar to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
- Temperature +18 … +30 °C (+64 +86 °F)
- Relative humidity 45 75 %
- Air pressure 860 1060 mbar/86 106 kPa
(12.5 15.4 psig)
Deviation in characteristics and characteristics
Reference installation conditions
- Flange size > DN 100
- min. distance to installations > 500 mm (19.69 in)
- Vessel diameter 1 m (3.281 ft) centrical installaton
- Medium Water
Temperature drift (current output) 0.06 %/10 K relating to the max. measuring
range Accuracy see diagrams
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 51
3mm
(1/8")
-3mm (-1/8")
10mm (25/64")
-10mm
(-25/64")
0,5 m
(1.6 ft)
0,08 m
(0.26 ft)
6 m
(20 ft)
-3mm (-1/8")
3mm
(1/8")
10mm (25/64")
-10mm
(-25/64")
0,5 m
(1.6 ft)
0,15 m (0.5 ft)
32 m/60 m
(105 ft/197 ft)
3mm
(1/8")
-3mm (-1/8")
10mm (25/64")
-10mm
(-25/64")
0,03 m
(0.1 ft)
6 m
(20 ft)
10 Suppl
ement
Fig. 26: Accuracy - Rod
Fig. 27: Accuracy - Cable
version -110 +400 °C (-166 +752 °F)
version -110 +400 °C (-166 +752 °F)
Fig. 28: Accuracy - Coax
52 VE
version -110 +400 °C (-166 +752 °F)
GAFLEX 66 4 20 mA/HART
32762-EN-080716
2
1
-110°C
(-166°F)
0°C
(32°F)
-40°C
(-40°F)
60°C
(140°F)
80°C
(176°F)
100°C
(212°F)
200°C
(392°F)
100°C
(212°F)
400°C
(752°F)
300°C
(572°F)
10 Sup
plement
Ambient
Ambient, storage and transport tem-
conditions
-40 +80 °C (-40 +176 °F)
perature
Process conditions
Process pressure -1 160 bar/-100 16000 kPa (-
-14.5 2321 psig), depending on the
process tting
Process temperature (graphite seal) -110 +400 °C (-166 +752 °F)
Fig. 29: Ambient 1 Process temperature
2 Ambient temperature
temperature - process temperature (version -110 +400 °C/-166 +752 °F)
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 53
10 Suppl
ement
Electrome
Cable entry/plug
chanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
3)
- Single chamber housing l 1 x cable gland M20 x 1.5 (cable:
ø 5 9 mm), 1 x
blind stopper M20 x 1.5
or:
closing cap M20 x 1.5; 1 x
l 1 x
blind
stopper M20 x 1.5
or:
closing cap ½ NPT, 1 x blind plug
l 1 x
½ NPT
or:
l 1 x plug (depending
on the version), 1 x
blind stopper M20 x 1.5
- Double chamber housing l 1 x cable entry M20 x 1.5 (cable: ø 5 9 mm), 1 x
blind stopper M20 x 1.5; 1 x blind stopper M16 x 1.5 or optionally available with 1 x plug M12 x 1 for
VEGADIS 61
or:
closing cap ½ NPT, 1 x
l 1 x
blind stopper ½ NPT, 1 x blind stopper M16 x 1.5 or optionally 1 x plug M12 x 1 for VEGADIS
61
or:
l 1 x
plug (depending
on the version), 1 x blind stopper M20 x 1.5; 1 x blind stopper M16 x 1.5 or optionally available with 1 x plug M12 x 1 for VEGADIS 61
Spring-loaded terminals for wire cross-
> 2.5 mm² (AWG 14)
section
Electromechanical data - version IP 66/IP 68, 1 bar
Cable entry
- Single chamber housing 1 x IP 68 cable gland M20 x 1.5; 1 x blind stopper M20 x 1.5
- Double chamber housing 1 x IP 68 cable gland M20 x 1.5; 1 x blind stopper M20 x 1.5; 1 x blind stopper M16 x 1.5
3)
Depending on the version M12 x 1, according to DIN 43650, Harting, Am­phenol-Tuchel, 7/8" FF.
54 VE
GAFLEX 66 4 20 mA/HART
32762-EN-080716
10 Sup
plement
Connectio
n cable
- Wire cross-section 0.5 mm² (AWG 20)
- Wire resistance < 0.036 Ω/m
- Tensile strength < 1200 N (270 lbf)
- Standard length 5 m (16.4 ft)
- Max. length 1000 m (3280 ft)
- Min. bending radius 25 mm (0.984 in) with 25 °C (77 °F)
- Diameter approx. 8 mm (0.315 in)
- Colour - standard PE Black
- Colour - standard PUR Blue
- Colour - Ex-version Blue
Indicating and adjustment module
Voltage supply and data transmission through the sensor Indication LC display in Dot matrix Adjustment elements 4 keys Protection
- unassembled IP 20
- mounted into the sensor without
IP 40
cover
Materials
- Housing ABS
- Inspection window Polyester foil
Voltage supply
Supply voltage
- Non-Ex instrument 14 36 V DC
- EEx-ia instrument 14 30 V DC
- EEx-d-ia instrument 20 … 36 V DC
Supply voltage with lighted indicating and adjustment module
- Non-Ex instrument 20 36 V DC
- EEx-ia instrument 20 30 V DC
- EEx-d-ia instrument 20 … 36 V DC
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 55
1000
750
500
250
14
1816 20 22 24 26 28 30 32 34 36
V
4
1
2
3
10 Suppl
ement
Permissible residual ripple
- < 100 Hz Uss< 1 V
- 100 Hz 10 kHz Uss< 10 mV
Load see diagram
Fig. 30: Voltage 1 HART load
2 Voltage limit EEx-ia instrument 3 Voltage limit non-Ex/Exd instrument 4 Supply voltage
diagram
Electrical protective measures
Protection, depending on housing version
- Plastic housing IP 66/IP 67
- Aluminium housing, stainless steel
IP 66/IP 68 (0.2 bar)
housing - investment casting, stain­less steel housing - electro-polished
- Aluminium
and
stainless housing,
IP 66/IP 68 (1 bar)
investment casting (optionally avail­able)
Overvoltage category III Protection class II
Functional safety (SIL)
Functional safety according to IEC 61508/IEC 61511
- Single channel architecture (1oo1D) up to SIL2
- double channel diversitary redun-
dant architecture (1oo2D)
56 VE
4)
A suitable cable is the prerequisite for maintaining the protection class.
up to SIL3
4)
GAFLEX 66 4 20 mA/HART
32762-EN-080716
10 Sup
plement
Approvals
5)
Approvals
- ATEX ia ATEX II 1G, 1/2G, 2G EEx ia IIC T6
- ATEX ia+d ATEX II 1/2G, 2G EExd ia IIC T6
- IEC ia IEC Ex ia IIC T6
- IEC ia+d IEC Ex d ia IIC T6
- FM FM CI.I, Div 2 (NI)+CI.II, III, Div 1 (DIP); FM
CI.I-III, Div 1 (IS); FM CI.I-III, Div 1 (IS)+Cl.I-III, Div 1 Gr.C-G(XP)
- CSA CSA CI.I, Div 2 (NI)+CI.II, III, Div 1 (DIP); FM CI.I-III, Div 1 (IS); FM CI.I-III, Div 1 (IS)+Cl.I-III, Div 1 Gr.C-G(XP)
- Miscellaneous WHG
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
5)
Deviating data in Ex applications: see separate safety instructions.
T 57
~ 69 mm
(2 23/32")
ø 77 mm
(3 1/32")
112 mm (4
13
/
32
")
M20x1,5/ ½ NPT
~ 69 mm (2 23/32")
ø 77 mm (3
1
/32")
117 mm (4
39
/
64
")
M20x1,5/ ½ NPT
~ 87 mm (3 27/64")
M16x1,5
ø 84 mm
(3 5/16")
120 mm (4
23
/
32
")
M20x1,5/ ½ NPT
~ 116 mm (4 9/16")
ø 84 mm (3 5/16")
116 mm (4
9
/
16
")
M20x1,5M20x1,5/
½ NPT
~ 59 mm (2 21/64")
ø 80 mm
(3
5
/32")
112 mm (4
13
/
32
")
M20x1,5/ ½ NPT
1 3 4
5
2
10 Suppl
ement
10.2 Dim
ensions
Housing in protection IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Fig. 31: Housing adjustment module the housing is 9 mm/0.35 in higher)
1 Plastic housing 2 Stainless steel housing, electropolished 3 Stainless steel housing - precision casting 4 Aluminium double chamber housing 5 Aluminium housing
versions in protection IP 66/IP 67 and IP 66/IP 68, 0.2 bar (with integrated indicating and
58 VE
GAFLEX 66 4 20 mA/HART
32762-EN-080716
117mm (4
39
/
64
")
120mm (4
23
/
32
")
~ 103mm
(4 1/16")
~ 105mm (4 9/64")
ø 77mm (3 1/32")
116mm (4
9
/
16
")
~ 150mm (5 29/32")
ø 84mm (3 5/16")
ø 84mm (3 5/16")
~ 93 mm (3 21/32")
ø 80 mm
(3
5
/32")
112 mm (4
13
/
32
")
M20x1,5/ ½ NPT
M20x1,5
M20x1,5/ ½ NPT
M16x1,5
1
2 3
M20x1,5M20x1,5
4
10 Sup
plement
Housing
Fig. 32: Housing
in protection IP 66/IP 68, 1 bar
versions in protection IP 66/IP 68, 1 bar (with integrated indicating and adjustment module the
housing is 9 mm/0.35 in higher)
1 Stainless steel housing - precision casting 2 Aluminium double chamber housing 3 Stainless steel housing, electropolished 4 Aluminium housing
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 59
L
150 mm (5.91")
53 mm
(2.09")
ø 6 mm (0.24")
12 mm (0.47")
ø 54 mm
(2.13")
ø 20 mm
(0.79")
ø 4 mm
(0.16")
ø 30 mm
(1.18")
271 mm (10.67")
22 mm
(0.87")
SW 50 (
1.97")
G1 / 1”NPT, G1½ / 1½”NPT
10 Suppl
ement
VEGAFLEX 66 - cable, rod
version (-110 +400 °C/-166 +752 °F)
Fig. 33: VEGAFLEX 66 - cable, rod
version with thread (-110 +400 °C/-166 +752 °F)
L Sensor length, see chapter "Technical data"
60 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
L
271 mm (10
5
/
8
")
18 mm (
45
/
64
")
29 mm (1
9
/
64
")
ø 21,1 mm
(53/64")
SW 50
(1
31
/32")
G1 / 1”NPT, G1½ / 1½”NPT
10 Sup
plement
VEGAFLEX 66 - coax
version (-110 +400 °C/-166 +752 °F)
Fig. 34: VEGAFLEX 66 - coax
version with thread (-110 +400 °C/-166 +752 °F)
L Sensor length, see chapter "Technical data"
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 61
10 Suppl
ement
10.3 In
dustrial property rights

10.4 Trademark

VEGA product lines are global protected by industrial property rights. Further information see http://www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter http://www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle. Pour plus d'informations, on pourra se référer au site http://www.vega.com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad industrial. Para mayor información revise la pagina web http://www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность. Дальнейшую информацию смотрите на сайте http://www.
vega.com.
VEGA 站<http://www.vega.com>
All brands used as well as trade and company names are property of their lawful proprietor/originator.
62 VE
32762-EN-080716
GAFLEX 66 4 20 mA/HART
10 Sup
plement
32762-EN-080716
VEGAFLEX 66 4 20 mA/HAR
T 63
VEGA Grieshaber KG
ISO 9001
Am Hohenstein 113 77761 Schiltach
Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info@de.vega.com
www.vega.com
Printing date:
statements concerning scope of delivery, application,
All
practical use and operating conditions of the sensors and
processing systems correspond to the information avail-
able at the time of printing.
© VEGA Grieshaber KG, Schiltach/Germany 2008
Subject to change without prior notice 32762-EN-080716
Loading...