Supplementary documents appropriate to the ordered version come
with the delivery. You can find them listed in chapter "Product
description".
Instructions manuals for accessories and replacement parts
Tip:
To ensure reliable setup and operation of your VEGAFLEX 65, we
offer accessories and replacement parts. The corresponding documentations are:
l 27720 - VEGADIS 61
l 30207 - Electronics module VEGAFLEX series 60
l 34296 - Protective cover
l 31088 - Flanges according to DIN-EN-ASME-JIS-GOST
l 30391 - Spacer
Editing status: 2012-03-19
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus3
1 About this document
1 About this document
1.1 Function
This operating instructions manual provides all the information you
need for mounting, connection and setup as well as important
instructions for maintenance and fault rectification. Please read this
information before putting the instrument into operation and keep this
manual accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained qualified
personnel. The contents of this manual should be made available to
these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can
result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
lList
The dot set in front indicates a list with no implied sequence.
àAction
This arrow indicates a single action.
1Sequence
Numbers set in front indicate successive steps in a procedure.
4VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
2 For your safety
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual must be
carried out only by trained specialist personnel authorised by the plant
operator.
During work on and with the device the required personal protective
equipment must always be worn.
2.2 Appropriate use
VEGAFLEX 65 is a sensor for continuous level measurement.
You can find detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly used
according to the specifications in the operating instructions manual as
well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbitrary
conversions or modifications are explicitly forbidden.
2.3 Warning about misuse
Inappropriate or incorrect use of the instrument can give rise to
application-specific hazards, e.g. vessel overfill or damage to system
components through incorrect mounting or adjustment.
2.4 General safety instructions
This is a high-tech instrument requiring the strict observance of
standard regulations and guidelines. The user must take note of the
safety instructions in this operating instructions manual, the countryspecific installation standards as well as all prevailing safety
regulations and accident prevention rules.
The instrument must only be operated in a technically flawless and
reliable condition. The operator is responsible for trouble-free
operation of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new
regulations.
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus5
2 For your safety
2.5 Safety label on the instrument
The safety approval markings and safety tips on the device must be
observed.
2.6 CE conformity
This device fulfills the legal requirements of the applicable EC
guidelines. By attaching the CE mark, VEGA provides a confirmation
of successful testing. You can find the CE conformity declaration in the
download area of "
www.vega.com".
2.7 Fulfillment of NAMUR recommendations
The device fulfills the requirements of the applicable NAMUR
recommendations.
2.8 Safety instructions for Ex areas
Please note the Ex-specific safety information for installation and
operation in Ex areas. These safety instructions are part of the
operating instructions manual and come with the Ex-approved
instruments.
2.9 Environmental instructions
Protection of the environment is one of our most important duties. That
is why we have introduced an environment management system with
the goal of continuously improving company environmental protection.
The environment management system is certified according to DIN
EN ISO 14001.
Please help us fulfil this obligation by observing the environmental
instructions in this manual:
l Chapter "Packaging, transport and storage"
l Chapter "Disposal"
6VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
3 Product description
1
2
3
3.1 Structure
3 Product description
Scope of delivery
Constituent parts
The scope of delivery encompasses:
l Level sensor VEGAFLEX 65
l Documentation
- this operating instructions manual
- Operating instructions manual 27835 "Indicating and adjust-
- Ex-specific "Safety instructions" (with Ex versions)
- if necessary, further certificates
The VEGAFLEX 65 consists of the components:
l Process fitting with probe
l Housing with electronics
l Housing cover, optionally available with indicating and adjustment
module
Fig. 1: VEGAFLEX 65 with single chamber housing
1Housing cover with integrated indicating and adjustment module (optional)
2Housing with electronics
3Process fitting
Type label
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus7
The type label contains the most important data for identification and
use of the instrument:
l Article number
l Serial number
l Technical data
3 Product description
l Article numbers, documentation
With the serial number, you can access the delivery data of the
instrument via
www.vega.com, "VEGA Tools" and "serial numbersearch". In addition to the type label outside, you can also find the
serial number on the inside of the instrument.
3.2 Principle of operation
Application range
Functional principle
Power supply and bus
communication
DD/CFF
VEGAFLEX 65 is a level sensor with pipe probe for continuous level
measurement.
It is designed for industrial use in all areas of process technology and
can be used in liquids.
High frequency microwave pulses are guided along a steel rope or a
rod. Upon reaching the product surface, the microwave pulses are
reflected. The running time is evaluated by the instrument and
outputted as distance.
Power is supplied via the H1 Fieldbus. A two-wire cable according to
Fieldbus specification serves as carrier of both power and digital datafor multiple sensors. This cable can be operated in two versions:
l via an H1 interface card in the control system and additional power
supply
l via a Linking device with HSE (High speed Ethernet) and
additional power supply according to IEC 61158-2
The DD (Device Descriptions) and CFF (capability files) necessary for
planning and configuration of your FF (Foundation Fieldbus) communication network are available in the download area of the VEGA
homepage
www.vega.com under "Services - Downloads - Software -Foundation Fieldbus". The appropriate certificates are also available
there. A CD with the appropriate files and certificates can be ordered
via e-mail under info@de.vega.com or by phone from one of the VEGA
agencies under the order number "DRIVER.S".
The optional heating requires its own operating voltage. You can find
details in the supplementary instructions manual "Heating for indicat-
ing and adjustment module".
3.3 Operation
The instrument can be adjusted with the following adjustment media:
l With indicating and adjustment module
l with the suitable VEGA DTM in conjunction with an adjustment
software according to the FDT/DTM standard, e.g. PACTware and
PC
l a configuration tool
8VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
3 Product description
3.4 Packaging, transport and storage
Packaging
Transport
Transport inspection
Storage
Storage and transport
temperature
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
according to DIN EN 24180.
The packaging of standard instruments consists of environmentfriendly, recyclable cardboard. For special versions, PE foam or PE foil
is also used. Dispose of the packaging material via specialised
recycling companies.
Transport must be carried out under consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or
concealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
l Not in the open
l Dry and dust free
l Not exposed to corrosive media
l Protected against solar radiation
l Avoiding mechanical shock and vibration
l Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
l Relative humidity 20 … 85 %
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus9
4 Mounting
4 Mounting
4.1 General instructions
Suitability for the process conditions
Mounting position
Welding work
Handling
Moisture
Make sure that all parts of the instrument exposed to the process, in
particular the sensor element, process seal and process fitting, are
suitable for the existing process conditions. These include above all
the process pressure, process temperature as well as the chemical
properties of the medium.
You can find the specifications in chapter "Technical data" and on the
type label.
Select an installation position you can easily reach for mounting and
connecting as well as later retrofitting of an indicating and adjustment
module. The housing can be rotated by 330° without the use of any
tools. You can also install the indicating and adjustment module in four
different positions (each displaced by 90°).
Before beginning the welding work, remove the electronics module
from the sensor. By doing this, you avoid damage to the electronics
through inductive coupling.
With threaded versions, the housing must not be used to screw in the
instrument! Applying tightening forces on the housing can damage its
internal parts.
Use the hexagon for screwing in.
Use the recommended cables (see chapter "Connecting to power
supply") and tighten the cable gland.
You can give your instrument additional protection against moisture
penetration by leading the connection cable downward in front of the
cable entry. Rain and condensation water can thus drain off. This
applies mainly to outdoor mounting as well as installation in areas
where high humidity is expected (e.g. through cleaning processes) or
on cooled or heated vessels.
10VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
Fig. 2: Measures against moisture penetration
4 Mounting
Measuring range
Pressure
Mounting position
The reference plane for the measuring range of the sensors is the seal
surface of the thread.
Keep in mind that a min. distance must be maintained below the
reference plane and possibly also at the end of the probe measurement in these areas is not possible. These min. distances are
listed in chapter "Technical data" in the "Supplement". Keep in mind
for the adjustment that the default setting for the measuring range
refers to water.
The process fitting must be sealed if there is gauge or low pressure in
the vessel. Before use, check if the seal material is resistant against
the measured product and the process temperature.
The max. permissible pressure is specified in chapter "Technical data"
or on the type label of the sensor.
4.2 Instructions for installation
Mount VEGAFLEX 65 in such a way that the probe does not touch any
installations or the vessel wall during installation.
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus11
4 Mounting
In vessels with conical bottom it can be advantageous to mount the
sensor in the center of the vessel, as measurement is then possible
down to the lowest point of the vessel bottom.
Fig. 3: Vessel with conical bottom
Inflowing medium
Make sure that the probe is not subjected to strong lateral forces.
Mount VEGAFLEX 65 at a position in the vessel where no
disturbances, e.g. from filling openings, agitators, etc., can occur.
Fig. 4: Lateral load
Extreme vibration caused by the system, e.g. due to agitators or
turbulence in the vessel from inflowing medium, can cause the coax
probe of VEGAFLEX 65 to vibrate in resonance. With coax probes of
12VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
4 Mounting
more than 1 m (3.281 in) length, you must secure the probe by
fastening a suitable isolated brace or guy directly above the end of the
rod.
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus13
5 Connecting to power supply
5 Connecting to power supply
5.1 Preparing the connection
Safety instructions
Voltage supply
Connection cable
Cable gland ½ NPT
Always keep in mind the following safety instructions:
l Connect only in the complete absence of line voltage
l If overvoltage surges are expected, overvoltage arresters should
be installed according to Foundation Fieldbus specification
Tip:
We recommend VEGA overvoltage arrester B63-32.
In hazardous areas you must take note of the respective regulations,
conformity and type approval certificates of the sensors and power
supply units.
The instrument requires a operating voltage of 9 … 32 V DC.
Operating voltage and the digital bus signal are carried on the sametwo-wire connection cable. Power is supplied via the H1 power supply.
Connection is carried out with screened cable according to Fieldbus
specification.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm
(0.2 … 0.35 in) ensures the seal effect of the cable gland. If you are
using cable with a different diameter or cross-section, exchange the
seal or use a suitable cable gland.
Make sure that the entire installation is carried out according to the
Fieldbus specification. In particular, make sure that the bus is
terminated with suitable terminating resistors.
On the instrument with cable entry ½ NPT and plastic housing there is
a metallic ½" threaded insert moulded into the plastic housing.
Caution:
No grease should be used when screwing the NPT cable gland or
steel tube into the threaded insert. Standard grease can contain
additives that corrode the connection between threaded insert and
housing. This would influence the stability of the connection and the
tightness of the housing.
Cable screening and
grounding
14VEGAFLEX 65 • Foundation Fieldbus
With systems with potential equalisation, connect the cable screen
directly to ground potential at the power supply unit, in the connection
box and at the sensor. The screen in the sensor must be connected
directly to the internal ground terminal. The ground terminal outside on
the housing must be connected to the potential equalisation (low
impedance).
31848-EN-120329
5 Connecting to power supply
In systems without potential equalisation, connect the cable screen
directly to ground potential at the power supply unit and at the sensor.
In the connection box or T-distributor, the screen of the short stub to
the sensor must not be connected to ground potential or to another
cable screen. The cable screens to the power supply unit and to the
next distributor must be connected to each other and also connected
to ground potential via a ceramic capacitor (e.g. 1 nF, 1500 V). Lowfrequency potential equalisation currents are thus suppressed, but the
protective effect against high frequency interference signals remains.
The total capacitance of the cable and of all capacitors must not exceed
10 nF in Ex applications.
Take note of the corresponding installation regulations for Ex
applications. In particular, make sure that no potential equalisation
currents flow over the cable screen. In case of grounding on both sides
this can be achieved by the use of a capacitor or a separate potential
equalisation.
5.2 Connection procedure
Proceed as follows:
1 Unscrew the housing cover
2 If an indicating and adjustment module is installed, remove it by
turning it to the left.
3 Loosen compression nut of the cable entry
4 Remove approx. 10 cm (4 in) of the cable mantle, strip approx.
1 cm (0.4 in) of insulation from the ends of the individual wires
5 Insert the cable into the sensor through the cable entry
6 Lift the opening levers of the terminals with a screwdriver (see
following illustration)
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus15
5 Connecting to power supply
7 Insert the wire ends into the open terminals according to the wiring
plan
Fig. 5: Connection steps 6 and 7
8 Press down the opening levers of the terminals, you will hear the
terminal spring closing
9 Check the hold of the wires in the terminals by lightly pulling on
them
10 Connect the screen to the internal ground terminal, connect the
outer ground terminal to potential equalisation
11 Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
12 Screw the housing cover back on
The electrical connection is finished.
5.3 Wiring plan, single chamber housing
The following illustrations apply to the non-Ex as well as to the Ex-ia
version.
16VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
I²C
3
4
5
1
2
Display
Ty p:
1 25 6 7 8
Bus
Sim.
I2C
Display
1
1 25
6 7 8
Electronics and connection compartment
Wiring plan
5 Connecting to power supply
Fig. 6: Electronics and connection compartment, single chamber housing
1Plug connector for VEGACONNECT (I²C interface)
2Spring-loaded terminals for connection of the external indication VEGADIS
61
3Ground terminal for connection of the cable screen
4Spring-loaded terminals for Foundation Fieldbus connection
5Simulation switch ("on" = mode for simulation release)
Fig. 7: Wiring plan, single chamber housing
1Voltage supply, signal output
5.4 Wiring plan, double chamber housing
The following illustrations apply to the non-Ex as well as to the Ex-ia
Fig. 11: Wire assignment connection cable
1brown (+) and blue (-) to power supply or to the processing system
2Shielding
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus19
6 Set up with the indicating and adjustment module PLICSCOM
6 Set up with the indicating and adjustment
module PLICSCOM
6.1 Short description
Function/Configuration
Mount/Dismount indica-
ting and adjustment module
The indicating and adjustment module is used for measured value
display, adjustment and diagnosis. It can be mounted in the following
housing versions and instruments:
®
l All sensors of the plics
in the double chamber housing (optionally in the electronics or
connection compartment)
l External indicating and adjustment unit VEGADIS 61
From a hardware version …- 01 or higher of PLICSCOM as well as a
hardware version …- 01, 03 or higher of the corresponding sensor, an
integrated backlight can be switched on via the adjustment menu. The
hardware version is stated on the type label of the PLICSCOM or the
sensor electronics.
Note:
You can find detailed information on the adjustment in the operating
instructions manual "Indicating and adjustment module".
instrument family, in the single as well as
6.2 Insert indicating and adjustment module
The indicating and adjustment module can be inserted into the sensor
and removed again at any time. It is not necessary to interrupt the
power supply.
Proceed as follows:
1 Unscrew the housing cover
2 Place the indicating and adjustment module in the desired position
on the electronics (you can choose any one of four different
positions - each displaced by 90°)
3 Press the indicating and adjustment module onto the electronics
and turn it to the right until it snaps in.
4 Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order.
The indicating and adjustment module is powered by the sensor, an
additional connection is not necessary.
20VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
6 Set up with the indicating and adjustment module PLICSCOM
Fig. 12: Insert indicating and adjustment module
Note:
If you intend to retrofit the instrument with an indicating and adjustment
module for continuous measured value indication, a higher cover with
an inspection glass is required.
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus21
2
3
1
1.1
6 Set up with the indicating and adjustment module PLICSCOM
6.3 Adjustment system
Fig. 13: Indicating and adjustment elements
1LC display
2Indication of the menu item number
3Adjustment keys
Key functions
l [OK] key:
- Move to the menu overview
- Confirm selected menu
- Edit parameter
- Save value
l [->] key to select:
- Menu change
- Select list entry
- Select editing position
l [+] key:
- Change value of the parameter
l [ESC] key:
- interrupt input
- Return to higher-ranking menu
Adjustment system
The sensor is adjusted via the four keys of the indicating and
adjustment module. The LC display indicates the individual menu
items. The functions of the individual keys are shown in the above
illustration. Approx. 10 minutes after the last pressing of a key, an
automatic reset to measured value indication is triggered. Any values
not confirmed with [OK] will not be saved.
22VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
6 Set up with the indicating and adjustment module PLICSCOM
6.4 Setup steps
Switch-on phase
Parameter adjustment
After VEGAFLEX 65 is connected to voltage supply or after voltage
recurrence, the instrument carries out a self-check for approx. 30
seconds. The following steps are carried out:
l Internal check of the electronics
l Indication of the instrument type, the firmware as well as the
sensor TAGs (sensor designation)
l Status byte goes briefly to fault value
Then the current measured value will be displayed and the
corresponding digital output signal will be outputted to the cable.
1)
As VEGAFLEX 65 is a distance measuring instrument, the distance
from the sensor to the product surface is measured. To have the real
product level displayed, an allocation of the measured distance to the
percentage height must be made. To carry out this adjustment, the
distance is entered with full and empty vessel. If these values are not
known, an adjustment with the distance values, e.g. 10 % and 90 % is
also possible. Starting point for these distance specifications is always
the seal surface of the thread or flange. With these settings, the real
level is calculated. Furthermore the operating range of the sensor is
limited from maximum to the required range.
The real product level during this adjustment is not important, because
the min./max. adjustment is always carried out without changing the
product level. These settings can be made ahead of time without the
instrument having to be installed.
In the main menu item "Basic adjustment", the individual submenu
items should be selected one after the other and provided with the
correct parameter values.
Caution:
If there is a separation of different liquids in the vessel, e.g. by
condensation, VEGAFLEX 65 will always detect the medium with the
higher dielectric figure (ε
).
r
Keep in mind that interfaces can cause faulty measurements.
If you want to measure the total height of both liquids reli ably, please
contact our service department or use an instrument specially
designed for interface measurement.
Start your parameter adjustment with the following menu items of the
basic adjustment:
Carry out min. adjust-
Proceed as follows:
ment
1)
The values correspond to the actual measured level as well as to the set-
tings already carried out, e.g. default setting.
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus23
6 Set up with the indicating and adjustment module PLICSCOM
1 Move from the measured value display to the main menu by
pushing [OK].
▶
Basic adjustment
Display
Diagnostics
Service
Info
2 Select the menu item "Basic adjustment" with [->] and confirm with
[OK]. Now the menu item "Min. adjustment" is displayed.
Min. adjustment
0.00 %
=
10.000 m(d)
8.000 m(d)
3 Prepare the % value for editing with [OK] and set the cursor to the
requested position with [->]. Set the requested percentage value
with [+] and save with [OK]. The cursor jumps now to the distance
value.
4 Enter the suitable distance value in m for the empty vessel (e.g.
distance from the sensor to the vessel bottom) corresponding to
the percentage value.
5 Save the settings with [OK] and move to "Max. adjustment" with [-
>].
Carry out max. adjust-
Proceed as follows:
ment
Max. adjustment
100.00 %
=
1.000 m(d)
2.000 m(d)
1 Prepare the % value for editing with [OK] and set the cursor to the
requested position with [->]. Set the requested percentage value
with [+] and save with [OK]. The cursor jumps now to the distance
value.
2 Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. Keep in mind that the max.
level must lie below the dead band.
3 Save the settings with [OK].
Application
Each product has different reflective properties. In addition, there are
various interfering factors which have to be taken into account:
agitated product surfaces and foam generation (with liquids); dust
generation, material cones and echoes from the vessel wall (with
solids). To adapt the sensor to these different conditions, you should
first select in this menu item under "Medium" either "Liquid" or "Solid".
With coax versions, this menu item must be set to "Liquid".
24VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
6 Set up with the indicating and adjustment module PLICSCOM
Application
Liquid
Standard
(DK ≥ 2)
Measured products have different reflective properties depending on
their respective permittivity (dielectric constant). That's why there is an
additional selection option.
Under "Sensitivity" you can select "Standard (DK ≥ 2)" or "Increased
sensitivity (DK < 2)".
Through this the sensor is optimally adapted to the product and
measurement reliability, particularly in products with poor reflective
properties, is considerably increased.
Enter the requested parameters via the appropriate keys, save your
settings and jump to the next menu item with the [->] key.
Damping
Linearisation curve
To suppress fluctuations in the measured value display, e. g. caused
by an agitated product surface, a damping can be set. This time can
be between 0 and 999 seconds. Keep in mind that the reaction time of
the entire measurement will then be longer and the sensor will react to
measured value changes with a delay. In general, a period of a few
seconds is sufficient to smooth the measured value display.
Damping
0 s
Enter the requested parameters via the appropriate keys, save your
settings and jump to the next menu item with the [->] key.
A linearization is necessary for all vessels in which the vessel volume
does not increase linearly with the level - e.g. in a horizontal cylindrical
or spherical tank - and the indication or output of the volume is
required. Corresponding linearization curves are preprogrammed for
these vessels. They represent the correlation between the level
percentage and vessel volume. By activating the appropriate curve,
the volume percentage of the vessel is displayed correctly. If the
volume should not be displayed in percent but e.g. in l or kg, a scaling
can be also set in the menu item "Display".
Linearisation curve
Linear
Enter the requested parameters via the appropriate keys, save your
settings and jump to the next menu item with the [->] key.
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus25
6 Set up with the indicating and adjustment module PLICSCOM
Copy sensor data
Reset
This function enables reading out parameter adjustment data as well
as writing parameter adjustment data into the sensor via the indicating
and adjustment module. A description of the function is available in the
operating instructions manual "Indicating and adjustment module".
The following data are read out or written with this function:
l Measured value presentation
l Adjustment
l Medium
l Vessel form
l Damping
l Linearisation curve
l Sensor-TAG
l Displayed value
l Unit of measurement
l Language
l Sensitivity
The following safety-relevant data are not read out or written:
l HART mode
l PIN
l Sensor length/Sensor type
l False signal suppression
Copy sensor data
Copy sensor data?
Basic adjustment
If the function "Reset" is carried out, the sensor resets the values of the
following menu items to the reset values (see chart):
2)
The following values will be reset:
FunctionReset value
Max. adjustmentDistance, upper dead zone
Min. adjustmentDistance, supplied sensor length
Min. adjustment - Cable versionDistance, lower dead zone
Damping ti0 s
LinearisationLinear
Sensor-TAGSensor
DisplayAI-Out 1
The values of the following menu items are not reset to the reset
values (see chart) with "Reset":
2)
Sensor-specific basic adjustment.
26VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
6 Set up with the indicating and adjustment module PLICSCOM
FunctionReset value
LanguageNo reset
Factory setting
Like basic adjustment, but in addition, special parameters are reset to
default values.
3)
Peak value
The min. and max. distance values are reset to the actual value.
Optional settings
Additional adjustment and diagnosis options such as e.g. scaling,
simulation or trend curve presentation are shown in the following menu
schematic. You will find a detailed description of these menu items in
the operating instructions manual "Indicating and adjustment module".
3)
Special parameters are parameters which are set customer-specifically on
the service level with the adjustment software PACTware.
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus27
6 Set up with the indicating and adjustment module PLICSCOM
6.5 Menu schematic
Basic adjustment
▶
Basic adjustment
Display
Diagnostics
Service
Info
1
Min. adjustment
000.0 %
=
10.000 m(d)
1.245 m(d)
Display
Basic adjustment
▶
Display
Diagnostics
Service
Info
Displayed value
AI-Out
Diagnostics
Basic adjustment
Display
▶
Diagnostics
Service
Info
Peak value
Distance min.: 0.580 m(d)
Dist. max.: 16.785 m(d)
1.1
2.1
3.1
2
3
Max. adjustment
100.0 %
=
0.000 m(d)
6.789 m(d)
Lighting
Switched off
▼
Sensor status
OK
1.2
2.4
3.2
Damping
1 s
Curve selection
Echo curve
1.3
3.3
Linearisation curve
Linear
Echo curve
Presentation of the echo
curve
1.4
3.4
28VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
Service
Basic adjustment
Display
Diagnostics
▶
Service
Info
6 Set up with the indicating and adjustment module PLICSCOM
4
Application
Liquid
Standard
(DK ≥ 2)
Unit of measurement
Info
Basic adjustment
Display
Diagnostics
Service
▶
Info
Device-ID
< max. 32 characters >
Sensor-TAG (PD_TAG)
< max. 32 characters >
Sensor characteristics
Version:
m(d)
4.2
4.6
5.1
5.5
5
False signal suppression
Change now?
Language
German
Instrument type
VEGAFLEX 6x
Serial number
12345678
4.3
4.7
5.2
Simulation
Start simulation?
Copy sensor data
Copy sensor data?
Calibration date
12. Dec. 2005
Software version
3.22
4.4
4.8
5.3
Reset
Select reset?
PIN
Enable?
Last change using PC
19. August 2003
17:48 h
4.5
4.9
5.4
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus29
6 Set up with the indicating and adjustment module PLICSCOM
6.6 Saving the parameter adjustment data
We recommended noting the adjusted data, e.g. in this operating
instructions manual, and archiving them afterwards. They are thus
available for multiple use or service purposes.
If VEGAFLEX 65 is equipped with an indicating and adjustment
module, the most important data can be read out of the sensor into the
indicating and adjustment module. The procedure is described in the
operating instructions manual "Indicating and adjustment module" in
the menu item "Copy sensor data". The data remain there permanently
even if the sensor power supply fails.
If it is necessary to exchange the sensor, the indicating and adjustment
module is inserted into the replacement instrument and the data are
written into the sensor under the menu item "Copy sensor data".
30VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
3
1
2
1
2
3
4
OPEN
TWIST
USB
LOCK
VEGACONNECT directly
on the sensor
7 Set up with PACTware and other adjustment programs
7 Set up with PACTware and other adjustment
programs
7.1 Connect the PC
Fig. 14: Connection of the PC via VEGACONNECT directly to the sensor
1USB cable to the PC
2VEGACONNECT
3Sensor
VEGACONNECT externally
Fig. 15: Connection via VEGACONNECT externally
1I²C bus (com.) interface on the sensor
2I²C connection cable of VEGACONNECT
3VEGACONNECT
4USB cable to the PC
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus31
7 Set up with PACTware and other adjustment programs
Necessary components:
l VEGAFLEX 65
l PC with PACTware and suitable VEGA DTM
l VEGACONNECT
l Power supply unit or processing system
7.2 Parameter adjustment with PACTware
Further setup steps are described in the operating instructions manual
"DTM Collection/PACTware" attached to each CD and which can also
be downloaded from our homepage. A detailed description is available
in the online help of PACTware and the VEGA DTMs.
Note:
Keep in mind that for setup of VEGAFLEX 65, DTM-Collection in the
actual version must be used.
All currently available VEGA DTMs are included as a DTM Collection
on a CD. They can be purchased for a token fee from the responsible
VEGA agency. In addition, the actual PACTware version is also
available on this CD.
In addition, this DTM Collection incl. the basic version of PACTware
can be downloaded free of charge from the Internet. Move via
vega.com and "Downloads" to "Software".
7.3 Parameter adjustment with AMS™
For VEGA sensors, instrument descriptions for the adjustment
program AMS™ are available as DD. The instrument descriptions are
already implemented in the current version of AMS™. For older
versions of AMS™, a free-of-charge download is available via Internet.
www.vega.com and "Downloads" to the item "Software".
Go via
www.
7.4 Saving the parameter adjustment data
It is recommended to document or save the parameter adjustment
data. That way they are available for multiple use or ser vice purposes.
The VEGA DTM Collection and PACTware in the licensed, professional version provide suitable tools for systematic project documentation and storage.
32VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
8 Maintenance and fault rectification
8 Maintenance and fault rectification
8.1 Maintenance
If the instrument is used properly, no special maintenance is required
in normal operation.
8.2 Remove interferences
Reaction when malfunc-
tions occur
Failure reasons
Fault rectification
24 hour service hotline
Checking Foundation
Fieldbus
The operator of the system is responsible for taking suitable measures
to rectify faults.
VEGAFLEX 65 offers maximum reliability. Nevertheless, faults can
occur during operation. These may be caused by the following, e.g.:
l Sensor
l Process
l Voltage supply
l Signal processing
The first measures to be taken are to check the output signals as well
as to evaluate the error messages via the indicating and adjustment
module. The procedure is described below. Further comprehensive
diagnostics can be carried out on a PC with the software PACTware
and the suitable DTM. In many cases, the causes can be determined
and the faults rectified this way.
Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since
we offer this service world-wide, the support is only available in the
English language. The service is free of charge, only the standard
telephone costs will be charged.
The following table describes possible errors and helps to remove
them:
ErrorCauseRectification
When an additional instrument is
connected, the H1
segment fails.
Measured value
on the indicating
and adjustment
module does not
correspond to the
value in the PLC
Max. supply current of the segment coupler
exceeded
The menu item
"Display - Dis-
play value" is
not set to "AI-
Out"
Measure the current consumption,
reduce size of segment
Check values and correct, if necessary
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus33
8 Maintenance and fault rectification
ErrorCauseRectification
Instrument does
not appear during
connection setup
In Ex applications, the regulations for the wiring of intrinsically safe
circuits must be observed.
Profibus DP cable pole-reversed
Incorrect termination
Instrument not
connected to the
segment
Check cable and correct, if necessary
Check termination at the beginning
and end points of the bus and terminate, if necessary, according to the
specification
Check and correct, if necessary
Error messages via the
indicating and adjustment module
The indicating and adjustment modules indicates faults via error codes
and text messages. The following table describes the error codes with
status according to NE 107 and gives information on the causes of
failure and their removal:
Status according to NE 107
FailureE013no measured value availableSensor in boot phase
Reaction after fault rectification
Error codeText messageCause/Rectification
no measured value availableSensor does not find an echo, e.g.
no measured value availableWrong sensor length entered
E017Adjustment span too smallAdjustment not within the specifica-
E036No operable softwareFailed or interrupted software up-
E042Hardware error, electronics defec-
tive
E043Hardware error, electronics defec-
tive
E113Communication conflictExchange the instrument or send it
due to faulty installation or wrong
parameter adjustment
tion. Carry out the adjustment again,
increasing the distance between
min. and max. adjustment
date/Repeat software update
Exchange the instrument or send it
in for repair
Exchange the instrument or send it
in for repair
in for repair
Depending on the reason for the fault and the measures taken, the
steps described in chapter "Set up" may have to be carried out again.
31848-EN-120329
34VEGAFLEX 65 • Foundation Fieldbus
8 Maintenance and fault rectification
8.3 Exchanging the electronics module
If the electronics module is defective, it can be replaced by the user.
In Ex applications only one instrument and one electronics module withrespective Ex approval may be used.
If there is no electronics module available on site, one can be orderedfrom the VEGA agency serving you.
Sensor serial number
Assignment
Foundation Fieldbus
The new electronics module must be loaded with the settings of the
sensor. These are the options:
l At the factory by VEGA
l Or on site by the user
In both cases, the sensor serial number is necessary. The serial
numbers are stated on the type label of the instrument, inside the
housing or on the delivery note.
Information:
When loading on site, first of all the order data must be downloaded
from the Internet (see operating instructions manual "Electronics
module").
The electronics modules are adapted to the respective sensor and
distinguish also in the signal output or power supply.
Electronics module FX-E.60F suitable for VEGAFLEX 61, 62, 63, 65,
66 - Foundation Fieldbus:
l FX-E.60FX (X = without approvals)
l FX-E.60FC (C = approvals CX, CK, GX according to product list)
8.4 Software update
The software version of VEGAFLEX 65 can be determined as follows:
l on the type label of the electronics
l via the indicating and adjustment module
l via PACTware
You can view all software histories on our website
Make use of this advantage and get registered for update information
via e-mail.
The following components are required to update the sensor software:
l Sensor
l Voltage supply
l VEGACONNECT
l PC with PACTware
l Current sensor software as file
www.vega.com.
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus35
8 Maintenance and fault rectification
Load sensor software to
PC
Prepare update
Load software into sen-
sor
At "www.vega.com/downloads" go to "Software". Select under "plics
sensors and instruments", "Firmware updates" the respective instru-
ment series and software version. Load the zip file via the right mouse
key with "Save target as" e.g. on the desktop of your PC. Move with the
right mouse key to the folder and select "Extract all". Save the
extracted files, for example on the desktop.
Connect the signal conditioning instrument to power supply and
provide the connection from the PC to the instrument via the interface
adapter. Start PACTware and go via the menu "Project" to the VEGAproject assistant. Select "USB" and "Set instruments online". Activate
the project assistant with "Start". The assistant establishes the
connection automatically and opens the parameter adjustment window
"Sensor # online parameter adjustment". Connect this parameter
adjustment window before you carry out further steps.
Select with the right mouse key the sensor in the project and go to
"Additional function". Then click to "Software update". The window
"Sensor # software update" opens. PACTware checks now the sensor
data and displays the actual hardware and software version of the
sensor. This takes approximately 60 s.
Push the button "Update software" and select the previously extracted
hex file. Then the software update can be started. The additional files
are installed automatically. Depending on the sensor, this procedure
lasts up to 1 h. Then the message appears ""Software update
successfully executed".
8.5 Instrument repair
If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our homepage at
www.
vega.com under: "Downloads - Forms and certificates - Repair form".
By doing this you help us carry out the repair qu ickly and without
having to call back for needed information.
l Print and fill out one form per instrument
l Clean the instrument and pack it damage-proof
l Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
l Please ask the agency serving you for the address of your return
shipment. You can find the respective agency on our website
www.vega.com under: "Company - VEGA worldwide"
36VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
9 Dismounting
9 Dismounting
9.1 Dismounting steps
Warning:
Before dismounting, be aware of dangerous process conditions such
as e.g. pressure in the vessel, high temperatures, corrosive or toxic
products etc.
Take note of chapters "Mounting" and "Connecting to power supply"
and carry out the listed steps in reverse order.
9.2 Disposal
The instrument consists of materials which can be recycled by
specialised recycling companies. We use recyclable materials and
have designed the electronics to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and
the respective national laws. Pass the instrument directly on to a
specialised recycling company and do not use the municipal collecting
points. These may be used only for privately used products according
to the WEEE directive.
Correct disposal avoids negative effects on humans and the environment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus37
10 Supplement
10 Supplement
10.1 Technical data
General data
Material 316L corresponds to 1.4404 or 1.4435
Materials, wetted parts
- Process fitting316L and TFM PCTFE +25 %GF, Hastelloy C22
- Process seal on the instrument side
(tube leadthrough)
- Process sealOn site (instruments with thread: Klingersil C-4400
- Tube: ø 21.3 mm (0.839 in)316L or Hastelloy C22 (2.4602)
Fig. 16: Measuring ranges of VEGAFLEX 65
1Reference plane
2Probe length L
3Measuring range (default setting refers to the measuring range in water)
4Upper dead band (see diagrams under Accuracy - grey section)
5Lower dead band (see diagrams under Accuracy - grey section)
Output variable
Output
- Signaldigital output signal, Foundation Fieldbus protocol
- Physical layeraccording to IEC 61158-2
Cycle timemin. 1 s (dependent on the parameter setting)
- Damping (63 % of the input variable)0 … 999 s, adjustable
- Fulfilled NAMUR recommendationsNE 43
Channel Numbers
- Channel 1Primary value
- Channel 2Secondary value 1
- Channel 3Secondary value 2
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus39
L
0
-2 mm (-0.079")
2 mm (0.079")
-10 mm (0.394")
5 mm (0.197")
0,15 m (5.906")
0,03 m (1.181")
0,05 m
(1.969")
1
10 Supplement
Transmission rate31.25 Kbit/s
Current value10 mA, ±0.5 mA
Resolution, digital> 1 mm (0.039 in)
Accuracy (according to DIN EN 60770-1)
Process reference conditions according to DIN EN 61298-1
- Temperature+18 … +30 °C (+64 … +86 °F)
- Relative humidity45 … 75 %
- Air pressure+860 … +1060 mbar/+86 … +106 kPa
(+12.5 … +15.4 psig)
Installation reference conditions
- Min. distance to installations> 500 mm (19.69 in)
- Vesselmetallic, ø 1 m (3.281 ft), centric installation,
process fitting flush with the vessel ceiling
- MediumWater/Oil (dielectric figure ~2.0)
- InstallationProbe end does not touch the vessel bottom
Sensor parameter adjustmentGating out of false signals carried out
Deviationsee diagrams
Depending on the installation conditions, there can
be deviations which can be rectified with an
adaptation of the adjustment or a change of the
measured value offset in the DTM service mode.
Fig. 17: Deviation VEGAFLEX 65 in coax version in water
1Dead zone - no measurement possible in this area
LProbe length
40VEGAFLEX 65 • Foundation Fieldbus
31848-EN-120329
L
0
-2 mm (-0.079")
2 mm (0.079")
10 mm (0.394")
5 mm (0.197")
-5 mm (-0.197")
0,15 m (5.906")
0,1 m (3.937")
0,15 m
(5.906")
0,1 m
(3.937")
11
10 Supplement
Fig. 18: Deviation VEGAFLEX 65 in coax version in oil
1Dead zone - no measurement possible in this area
LProbe length
Influence of the ambient temperature to the sensor electronics
Temperature drift0.03 %/10 K relating to the max. measuring range
or max. 0.3 %
Temperature drift - Digital output3 mm/10 K relating to the max. measuring range or
max. 10 mm
Ambient conditions
Ambient, storage and transport temperature-40 … +80 °C (-40 … +176 °F)
Process conditions
Process pressure-1 … +40 bar/-100 … +4000 kPa (-
-14.5 … +580 psig), depending on the process
fitting
Process temperature (thread or flange temperature)
- FKM (Viton)-40 … +150 °C (-40 … +302 °F)
- EPDM-40 … +150 °C (-40 … +302 °F)
- FFKM (Kalrez 6375)-20 … +150 °C (-4 … +302 °F)
The measurement error from the process conditions is in the specified pressure and temperature
range of below 1 %.
31848-EN-120329
VEGAFLEX 65 • Foundation Fieldbus41
0°C
(32°F)
40°C
(104°F)
20°C
(58°F)
60°C
(140°F)
80°C
(176°F)
50°C
(122°F)
-40°C
(-40°F)
100°C
(212°F)
150°C
(302°F)
1
2
10 Supplement
Fig. 19: Ambient temperature - Process temperature
1Ambient temperature
2Process temperature (depending on the seal material)
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug
4)
- Single chamber housingl 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5
or:
l 1 x closing cap M20 x 1.5; 1 x blind stopper
M20 x 1.5
or:
l 1 x closing cap ½ NPT, 1 x blind plug ½ NPT
or:
l 1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5
- Double chamber housingl 1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5; 1 x blind stopper
M16 x 1.5 or optionally available with 1 x plug
M12 x 1 for external indicating and adjustment
unit
or:
l 1 x closing cap ½ NPT, 1 x blind stopper
½ NPT, 1 x blind stopper M16 x 1.5 or
optionally 1 x plug M12 x 1 for external
indicating and adjustment unit
or:
l 1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5; 1 x blind stopper M16 x 1.5
or optionally available with 1 x plug M12 x 1 for
external indicating and adjustment unit
4)
Depending on the version M12 x 1, according to DIN 43650, Harting, 7/
8" FF.
31848-EN-120329
42VEGAFLEX 65 • Foundation Fieldbus
10 Supplement
Spring-loaded terminals for wire cross-section
Electromechanical data - version IP 66/IP 68 (1 bar)
Cable entry
- Single chamber housing1 x IP 68 cable gland M20 x 1.5; 1 x blind stopper
- Double chamber housing1 x IP 68 cable gland M20 x 1.5; 1 x blind stopper
Connection cable
- Wire cross-section0.5 mm² (AWG 20)
- Wire resistance< 0.036 Ω/m
- Tensile strength< 1200 N (270 lbf)
- Standard length5 m (16.4 ft)
- Max. length1000 m (3280 ft)
- Min. bending radius25 mm (0.984 in) with 25 °C (77 °F)
- Diameter approx.8 mm (0.315 in)
- Colour - standard PEBlack
- Colour - standard PURBlue
- Colour - Ex-versionBlue
Indicating and adjustment module
Voltage supply and data transmissionthrough the sensor
IndicationLC display in dot matrix
Adjustment elements4 keys
Protection rating
- unassembledIP 20
- mounted into the sensor without coverIP 40
Materials
- HousingABS
- Inspection windowPolyester foil
< 2.5 mm² (AWG 14)
M20 x 1.5
M20 x 1.5; 1 x blind stopper M16 x 1.5
Voltage supply
Operating voltage
- Non-Ex instrument9 … 32 V DC
- EEx-ia instrument9 … 24 V DC
- EEx-d instrument16 … 32 V DC
Operating voltage with illuminated indicating and adjustment module
VEGA product lines are global protected by industrial property rights.
Further information see http://www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor
housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche
Schutzrechte.
Nähere Informationen unter http://www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des
droits de propriété intellectuelle.
Pour plus d'informations, on pourra se référer au site http://www.vega.
com.
VEGA lineas de productos están protegidas por los derechos en el
campo de la propiedad industrial.
Para mayor información revise la pagina web http://www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру
правами на интеллектуальную собственность.
Дальнейшую информацию смотрите на сайте http://www.vega.com.