VEGA FLEX54K User Manual

Operating Instructions
VEGAFLEX 54K
Level and Pressure
Safety information / Note Ex area

Safety information

Please read this manual carefully, and also take note of country-specific installation standards (e.g. the VDE regulations in Germany) as well as all prevailing safety regulations and acci­dent prevention rules. For safety and warranty reasons, any internal work on the instruments, apart from that in­volved in normal installation and electrical con­nection, must be carried out only by qualified VEGA personnel.
2 VEGAFLEX 54K

Note Ex area

Please note the attached safety instructions containing important information on installation and operation in Ex areas. These safety instructions are part of the oper­ating instructions and come with the Ex ap­proved instruments.
Contents
Contents
Safety information ........................................................................ 2
Note Ex area ................................................................................ 2
1 Product description
1.1 Function ................................................................................. 4
1.2 Application features ............................................................. 4
1.3 Types and versions ............................................................. 5
1.4 Type code ............................................................................. 7
1.5 Technical data ....................................................................... 8
1.6 Dimensions ......................................................................... 11
2 Mounting
2.1 Installation instructions ...................................................... 14
3 Electrical connection
3.1 Connection and connection cable .................................... 16
3.2 Connection of the sensor .................................................. 18
3.3 Connection of the indicating instrument VEGADIS 50 .... 20
4Setup
4.1 Adjustment media .............................................................. 21
4.2 Adjustment with the adjustment module MINICOM ........ 21
4.3 Adjustment with the PC ...................................................... 28
4.4 Adjustment with HART® handheld ..................................... 31
5 Diagnosis
5.1 Simulation ............................................................................ 32
5.2 Failure rectification ............................................................. 32
VEGAFLEX 54K 3

1 Product description

Product description

1.1 Function

High frequency microwave impulses are guided along a steel rope or a rod. The microwave impulses are reflected when they reach the product surface. The impulse running time is processed by the integrated electronics and outputted as level.
The sensors detect levels in all types of liq­uids. Density, conductivity and dielectric value of the product do not influence the measurement. Continuously changing prop­erties of the medium also do not influence the measured value.
In many applications, VEGAFLEX microwave sensors are the solution to difficult technical problems. Even in products with varying or fluctuating dielectric constant, the level is reliably detected. In high, narrow vessels, where non-contact measurement technolo­gies often deliver less than optimal results, VEGAFLEX carries out problem-free meas­urements. VEGAFLEX 54 is optionally equipped with a 4 mm diameter cable up to a length of 20 m, or an 8 mm diameter rod up to a length of 4 m.
• Adjustment without filling or emptying the vessel.
• 4 … 20 mA two-wire sensors, power sup­ply and measuring signal via one two-wire cable (loop powered).
• Up to 15 sensors can be connected via one two-wire cable.
• Measuring range up to 20 m, very small minimum measuring distance.
• Unaffected by application conditions such as:
- varying products
• Unaffected by the vessel material, e.g. metal, concrete, plastic etc.
• As an option, the display with adjustment functions can be mounted separate from the sensor.
• Low wiring costs through the use of two­wire technology.

1.2 Application features

Applications
• Level measurement of liquids.
• Measurement also in vacuum.
• All slightly conductive substances and all substances with a dielectric value ε can be measured.
• Measuring range 0 ... 20 m.
Two-wire technology
• Power supply, signal transmission and output signal on one two-wire cable.
• 4 … 20 mA output signal or digital output signal.
Rugged
• Highly resistant materials: PCTFE, 1.4435, Hastelloy C22.
Precise and reliable
• Resolution 1 mm.
• Unaffected by noise, steam, gas composi­tions and inert gas stratification.
• Unaffected by varying density.
• Measurements under pressure up to 40 bar and with product temperatures up to 150°C.
Communicative
• Integrated measured value display.
• As an option, display can be mounted separate from the sensor.
• Adjustment from the PLC level.
Approvals
• ATEX II 1 G or II 1/2 G or II 2 G EEx ia II C T6
• ATEX II 1/2 G or II 2 G EEx d ia II C T6
> 1.7
r
4 VEGAFLEX 54K
Product description
Adjustment with PC
The setup and adjustment of VEGAFLEX sensors is generally done on the PC with the adjustment program VVO (VEGA Visual Op­erating) under Windows®. The program leads quickly through adjust­ment and parameter setting by means of pictures, graphics and process visualisations.
The PC can be connected at any individual location in the system or directly to the signal cable. This is done by connecting the two­wire PC interface converter VEGACONNECT 2 or 3 to the sensor or to the signal cable.
The adjustment and parameter data can be saved with the adjustment software on the PC and can be protected by passwords. On request, the adjustments can be quickly transferred to other sensors.
Adjustment with adjustment module MINICOM
With the small (3.2 cm x 6.7 cm) 6-key ad­justment module MINICOM, the adjustment can be carried out in clear text dialogue. The adjustment module can be plugged into VEGAFLEX or the external indicating instru­ment VEGADIS 50. Thus, VEGAFLEX can also be adjusted ffrom the external indicating instrument VEGADIS
50.
Tank 1 m (d)
12.345
ESC
+
-
OK
Adjustment with HART® handheld
Series 50 sensors with 4 … 20 mA output signal can also be adjusted with the HART handheld or via a PC with HART® software (e.g. Smart version). A special DDD (Data Device Description) is not necessary. The sensors can be adjusted with the HART standard menus of the handheld.
HART Communicator
HART® handheld
®
®

1.3 Types and versions

VEGAFLEX series 50 K
In general, all VEGAFLEX series 50 sensors can be adjusted with the pluggable adjust­ment module MINICOM or with any standard HART® handheld. With the software VEGA Visual Operating (VVO), it is also possible to adjust the sensor with a PC.
4 … 20 mA sensors
Two-wire sensors for connection to a power supply unit or a PLC.
Detachable adjustment module MINICOM
VEGAFLEX 54K 5
Configuration
A measuring system with a VEGAFLEX can be realised in different ways (see following illustration). The external indicating instrument VEGADIS 50 can be mounted up to 25 m away from the sensor.
Two-wire technology
1
4 ... 20 mA sensor, power supply and measuring signal via one two-wire cable (loop-powered), indicating instrument VEGADIS 11 only possible with four-wire technology.
2
Connection to a PLC (active).
3
Connection to indicating instrument VEGADIS 371 with up to four relay outputs.
VEGADIS 50
1
VEGAFLEX
4
2
Product description
1)
2 2
4 … 20 mA
VEGADIS 11
1)
4 … 20 mA passive
2
4 … 20 mA passive
+
-
2
PLC
+
-
VEGADIS 371 Ex
Relay (4 x)
0/4 … 20 mA
1)
only with HART
2
3
®
6 VEGAFLEX 54K
Product description

1.4 Type code

The second figure of the type name, e.g. VEGAFLEX 5[4]… distinguishes the instru­ments according to the type and version of the sensor component.
The letter, e.g. VEGAFLEX 54[K] character­ises the output signal: K stands for an analogue 4 … 20 mA output signal (compact instrument) P stands for a digital signal transmission for connection to Profibus PA
VEGAFLEX 54 K .CX D
D - Two-wire loop-powered / 4 … 20 mA output, HART
®
G - Profibus PA
CX - ATEX II 1 G or II 1/2 G or II 2 G EEx ia II C T6
DX - ATEX II 1/2 G or II 2 G EEx d ia II C T6
P - Profibus PA K - Analogue 0 … 20 mA output signal (two-wire technology)
Instrument series with 4 mm cable / 8 mm rod
Measuring principle (FLEX for guided microwave)
VEGAFLEX 54K 7

1.5 Technical data

Power supply
Supply voltage
- two-wire sensor 24 V DC (15 … 36 V DC) Current consumption
- two-wire sensor max. 22.5 mA Power consumption
- two-wire sensor max. 810 mW
1000
680
500
250
HART® load
Voltage limit Ex sensor
Product description
Voltage limit Non-Ex sensor
0
20 22 24 26 28 30 32 34 3615 16 18
V
Supply voltage
The min. supply voltage depends on the sensor current.
V
36
15 14
04 22
Range of the required terminal voltage on the sensor
Sensor current
mA
Measuring range
VEGAFLEX 54 rod 0.15 ... 4 m (with DK value > 4: 0.1 … 4 m)
cable 0.15 … 20 m (with DK value < 4: 0.3 … 20 m)
Output signal
4 … 20 mA current signal, load max. 500
Adjustment
- PC and adjustment software VEGA Visual Operating
- adjustment module MINICOM
- HART® handheld
8 VEGAFLEX 54K
Product description
Accuracy (under reference conditions acc. to IEC 770 - relating to the max. meas. range)
Linearity error < 0.1 % Temperature drift 0.015 %/10 K Resolution of the 4 … 20 mA signal 0.025 % of range (DA converter) Resolution 1 mm
Characteristics
Min. span between full and empty adjustment
- analogue output signal 10 mm
Ambient conditions
Vessel pressure -1 … 40 bar Ambient temperature on the housing -40°C … +60°C Process temperature -40°C … +150°C Storage and transport temperature -40°C … +80°C Protection IP 66/IP 67 (meets both protection standards) Protection class
- two-wire sensor II
- four-wire sensor I Overvoltage category III Max. tensile load cable ø 4 mm: 1 KN
rod ø 8 mm: 1 KN
Min. dielectric constant ε
> 1.7
r
Process fittings
VEGAFLEX 54 G1 A, 1" NPT, G 11/2 A, 11/2" NPT
of 1.4435 (stst; 316L) or Hastelloy C22 plated
Ex technical data
Ex-Zone 0 acc. to ATEX II 1 G or II 1/2 G or II 2 G EEx ia II C T6, ATEX II 1/2 G or II 2 G EEx d ia II C T6 and ATEX II 1/2 G The permissible operating data of the VEGAFLEX sensors for Ex areas are listed in the certificate.
Materials
Housing PBT (Valox) or
Aluminium (powder-coated) Rod 1.4435 (stst; 316L) or Hastelloy Cable 1.4401 (stst; 316L)
VEGAFLEX 54K 9
Product description
gravity weight 1.4435
Connection cables
Two-wire sensor
- power supply and signal via one two-wire cable. The cable resistance depends on the supply voltage (see diagram).
Conductor cross-section generally 2.5 mm Ground connection max. 4 mm
2
2
Cable entry 2 x M20 x 1.5 (cable diameter 5 … 9 mm)
2 x 1/2" NPT with ExD housing
CE conformity
VEGAFLEX sensors meet the protective regulations of EMC (89/336/EWG) and NSR (73/23/EWG). Conformity has been judged acc. to the following standards: EMC Emission EN 50 081 - 1: 1992
Susceptibility EN 50 082 - 2: 1995
NSR EN 61 010 - 1: 1993
Display
Display scalable, analogue and digital display of
measured values (option).
An external measured value display, powered by the sensor, can be mounted at a distance of up to 25 m from the sensor.
Signal output
Signal output
- two-wire technology 4 … 20 mA (see diagram)
Resolution of the 20 mA signal 0.025 % of range Load 0 … 500
Two-wire technology: The analogue 4 … 20 mA output signal (measuring signal) is transmitted together with the power supply via one two-wire cable.
10 VEGAFLEX 54K
Product description

1.6 Dimensions

Adjustment module MINICOM
Tank 1 m (d)
12.345
67,5
74
ESC
+
-
OK
Adjustment mod­ule for insertion
32,5
into series 50 sensors or into the external indi­cating instrument VEGADIS 50
VEGAFLEX 54K 11
Plastic housing A luminium housing
201
165
Product description
322
182
Thread Thread G 1A / 1" NPT G 11/2A / 11/2" NPT
SW41
G1A 1NPT
ø 8 ø 8ø 20
10˚
91
ø 4
M20x1,5
80
70
215
101
0 7 3
5 0 2
55
22
L
185
SW60
G1 1/2A 1 1/2NPT
25
M20x1,5
ø 4
ø 20
116
55
26
80
L
70
Flange
53
24
88
L
70
ø 8
ø 4
ø 20
ø 200
12 VEGAFLEX 54K
Product description
Aluminium housing with ExD terminal compartment
ø 8
215
185
370
205
Flange Thread Thread (Hastelloy C22) G 1A / 1" NPT G 11/2A / 11/2" NPT
24
4
L 53
ø 138
ø 200
SW41
G1A
1NPT
25
½" NPT
ø 8
116
(Hastelloy C22) (Hastelloy C22)
SW60
55
22
G1 1/2A
1 1/2NPT
L
ø 8
55
33, 5
L
VEGAFLEX 54K 13

2 Mounting

Mounting

2.1 Installation instructions

Lateral load
Make sure that the cable is not subjected to strong lateral forces. Mount VEGAFLEX at a location where no interference caused by, e.g., stirrers, filling inlets etc. can occur.
Lateral load
Pressure
In case of gauge or low pressure in the ves­sel, the mounting boss must be sealed on the thread. Make sure that the seal ring is resist­ant to the medium.
Cable entries
When mounting outdoors, on cooled vessels or in humid areas where cleaning is done, e.g. with steam or high pressure, the seal of the electrical cable entry is very important.
Use electrical cable with a round cross-sec­tion and tighten the cable entry firmly. The cable entry is suitable for cable diameters of 5 mm to 9 mm.
Moisture from outside
To avoid moisture ingress, the connection cable leading to the housing should be looped downward (directly in front of the cable entry), so that rain and condensation water can drain off. This mainly applies to outdoor mounting, in areas where moisture is expected (e.g. by cleaning processes) or to cooled or heated vessels.
Moisture
14 VEGAFLEX 54K
Mounting
Measuring range
The electrode (rod / cable) cannot be used over its entire length. Keep especially in mind that the cable length cannot be utilised down to the very tip. A meas­urement in the area of the gravity weight is not possible. Rod 0.15 ... 4 m (with DK value > 4: 0.1 … 4 m) Cable 0.15 … 20 m (with DK value < 4: 0.3 … 20 m)
x
100%
L
Measuring range
L
0%
Sockets
Avoid long sockets. Mount the sensor flush and use short sockets with small diameters. Note under 4.3 Adjustment with PC / Empty vessel profile.
x
100%
0%
VEGAFLEX 54K 15

3 Electrical connection

Electrical connection
3.1 Connection and connection ca­ble
Safety information
As a rule, do all work in the complete ab­sence of line voltage. Always switch off the power supply before you carry out connect­ing work on the sensors. Protect yourself and the instrument, especially when you use sensors which do not work with low voltage.
Qualified personnel
Instruments which are not operated with a protective low voltage must only be con­nected by qualified personnel.
Connection
For connection, a standard two-wire cable with max. 2.5 mm2 can be used. Quite often the "Electromagnetic pollution" from electronic actuators, energy lines and transmitting stations is so considerable that the two-wire cable should be shielded.
We recommend the use of shielded cable. Shielding is also a good preventative meas­ure against future sources of interference. Ground the shielding only on one sensor end (fig. 3.1 a). It is advantageous to ground the shielding on both ends. However, you must make sure that no earth equalisation currents flow through the cable shielding (fig. 3.1 b). Earth equalisation currents (when grounding at both ends) can be prevented by connect­ing the cable shielding on one end (e.g. in the switching cabinet) via a capacitor (e.g.
0.1 µF; 250 V AC) to the earth potential. Use
a very low-resistance earth connection (foun­dation, plate or mains earth). In Ex applica­tions, the shielding should be grounded only at one end. Potential transfer can be caused when grounding at both ends.
Ex protection
If an instrument is used in hazardous areas, the necessary regulations as well as the Ex certificate of VEGAFLEX for systems in Ex areas should be noted.
Connection cable
Make sure that the connection cables are specified for the expected operating tempera­tures in your systems. The cable must have an outer diameter of 5 … 9 mm (1/5 ... 1/3 inch). Otherwise, the seal effect of the cable entry will not be ensured.
Cables for intrinsically safe circuits must be marked blue and must not be used for other circuits. Note the special wiring regulations for Profibus PA sensors (P), see "1.5 Techni­cal data connection cables".
Earth conductor terminal
On VEGAFLEX sensors, the earth conductor terminal is galvanically connected to the metal thread.
16 VEGAFLEX 54K
Electrical connection
Grounding at one end (next to the sensor)
Processing
Fig. 3.1 a
Grounding at both ends (on the signal conditioning instrument via the potential sepa­rating capacitor)
0.1 µF 250 V AC
Processing
Fig. 3.1 b
VEGAFLEX 54K 17
Electrical connection

3.2 Connection of the sensor

• Loosen the closing screws on the upper
side of the sensor.
• Open the sensor cover.
• Remove the sleeve nut from the cable
fitting and slide it a short distance up the connection cable.
• Remove the rubber seal from the cable
fitting and slide it also a short distance up the connection cable.
• Remove the outer insulation of the connec-
tion cable over a length of approx. 10 cm.
4 … 20 mA passive
+
• Insert the cable through the cable fitting into the sensor.
• Connect the cables. Push the white tabs of the spring-loaded terminals with a small screwdriver and insert the copper core of the connection cable into the terminal open­ing. Check the hold of the individual wires in the terminals by pulling lightly on them.
• Screw the sleeve nut back onto the cable fitting and screw it down tightly.
1)
To the indicating instrument in the
-
sensor cover or to the external indicat­ing instrument VEGADIS 50
M20 x 1.5 (diameter of the connection cable 5 … 9 mm)
+ 1 4-20mA
2 -
Communication
5678
-+
Display
ESC
OK
2.23272
Two-wire technology (analogue)
(loop powered)
1)
4 … 20 mA passive means that the sensor consumes a level-dependent current of 4 … 20 mA (consumer).
18 VEGAFLEX 54K
Electrical connection
Exd version (loop-powered with pressure-tight encapsulated terminal compartment)
EEx d terminal compartment
(opening in Ex area not allowed)
4 … 20 mA passive
1)
-+
Locking of the cover
Exd terminal box
Supply: 20...36V DC/4...20mA HART
Shield
-
+
2
1
2
1
1)
4 … 20 mA passive means that the sensor consumes a level-dependent current of 4 … 20 mA (consumer).
1
/2" NPT EEx d (diameter of the connection cable
3.5 … 8.7 mm)
Adjustment module and indication termi­nal compartment
(opening in Ex area allowed)
1
/2" NPT EEx d (diameter of the connection cable to the Exd terminal box
3.5 … 8.7 mm)
Exd safe channel through to the Exd terminal box
1
+-2
4-20mA
Communication
-+
ESC
Display
OK
8765
2.23272
VEGAFLEX 54K 19
3.3 Connection of the indicating in­strument VEGADIS 50
Loosen the 4 screws of the housing cover on VEGADIS 50. The connection procedure can be facilitated during connecting work by fastening the cover to one side of the housing with the afore-mentioned screws.
Adjustment module
OUTPUT (to the sensor)
SENSOR
DISPLAY (in the cover of the indicating instrument)
DISPLAY1234 5678
Tank 1 m (d)
12.345
ESC
+
-
OK
Electrical connection
VEGADIS 50
Screws
4 … 20 mA passive
+
-
to VEGADIS 50 (use shielded cable. See fig. 3.1 a or fig. 3.1 b)
M20x1.5
Communication
-+
5678
Display
ESC
OK
2.23272
+ 1 4-20mA
2 -
Two-wire sensor
(loop powered)
20 VEGAFLEX 54K
Setup

4 Setup

4.1 Adjustment media

Series 50 sensors can be adjusted with the
- PC (adjustment program VVO)
- detachable adjustment module MINICOM
- HART® handheld. It is only possible to use one adjustment device at a time.
Adjustment module MINICOM
With the adjustment module MINICOM, you can adjust directly in the sensor or in the external indicating instrument VEGADIS 50. With a dialogue text display and 6 keys, the module offers the same adjustment function­ality as the adjustment software VVO.
Adjustment program VVO
With the adjustment program VVO (VEGA Visual Operating System) on the PC, the sensors can be adjusted in a very conven­ient, user-friendly manner. The PC communi­cates with the sensor via the interface converter VEGACONNECT 2 or 3. A digital adjustment signal is superimposed on the signal and supply cables. The PC can be connected at any location along the cable or directly to the sensor.
HART® handheld
VEGAFLEX 50 K sensors, like other HART protocol-compatible instruments, can be adjusted with the HART® handheld. A manu­facturer-specific DDD (Data Device Descrip­tion) is not required. The sensors communicate with the HART® standard menus, through which all sensor functions are accessible. Functions that are rarely used, such as, for example, the scaling of the A/D converter for the signal output or the adjustment with medium, are not possible or are blocked with the HART® handheld. These functions must be carried out with the PC or the adjustment module MINICOM.
®

4.2 Adjustment with the adjustment module MINICOM

Just as with the PC, you can adjust the sen­sor with the small, detachable adjustment module MINICOM. The adjustment module is plugged into the sensor or into the external indicating instrument (optional).
All the adjustment options provided by a PC with the adjustment program VVO are also available with the adjustment module MINICOM. You can carry out all adjustment steps with the 6 keys of the adjustment mod­ule. A small display shows you, apart from the measured value, a short message on the menu item or an entered numerical value.
ESC
+
-
Tank 1 m (d)
12.345
OK
2
4 ... 20 mA
ESC
+
-
Tank 1 m (d)
12.345
OK
4
VEGAFLEX 54K 21
Setup
Adjustment elements
The adjustment module MINICOM is menu­orientated. The clear text indications on the display lead through the menu. The functions of the keys are described below.
2
1
Tank 1
-
m (d)
12.345
5
"OK" key (4)
With the "OK" key you can confirm the set­tings. When the symbol or is shown in the digital indication, you can switch to the lower menu level with the "OK" key. The symbol indicates there is no branching point below the menu item, but only a further
▼ ▼
menu item of the respective function.
"ESC" key (3)
With the "ESC" key (Escape) you can inter­rupt an adjustment or a current function or switch to the next higher menu level. To reach e.g. the uppermost menu level, push the "ESC" key several times.
"+" and "–" key (2)
With the "+" and "–" keys you can modify the values of the parameters or choose from a number of options. After an initial push, the value to be adjusted flashes. Each additional push changes the value.
3
ESC
+
OK
4
▼ ▼
Arrow keys (5)
With the keys "–>" and "<–", you can move within the menu level from one menu item to the next.
Digital indication (1)
During operation, the digital display shows the actual measured value. When adjusting the instrument, the clear text display shows the respective function.
Branching point, from where you can
move to the next lower menu.
This symbol informs you of a subsequent
safety enquiry.
Adjustment steps
On the following pages, you will see the menu schematic of the adjustment module MINICOM.
Set up the sensor in the following sequence. The numbers correspond to the sequence of the setup. You will find the numbers with the respective menu items in the menu schematic on the following pages.
1. Adjustment
2. Conditioning
3. Outputs
4. Operating range
5. Measuring conditions
6. Indication of the useful signal and noise level
22 VEGAFLEX 54K
Setup
1. Adjustment
Under the menu item " the sensor of the measuring range it should operate in. The instruments are already adjusted to the respective probe length. You only have to carry out a max. adjustment and enter the max. level.
Max.
Min.
You can carry out the adjustment with or without medium. Generally, you will carry out the adjustment without medium, because you can adjust without having to fill the vessel.
Adjustment without medium
• Enter the distance of your sensor to the medium at 0 % (example: 5.850 m) with the "+" and "–" key. If you do not know the distance, you have to do a sounding.
• Enter the distance from the sensor to the medium at 100 % filling.
Adjustment
100 % (distance to the product
0.300 m) corresponds to 1200 liters, e.g. 20 mA
0 % (distance to the product
5.850 m) corresponds to 456
liters, e.g. 4 mA)
", you inform
Meas. range
Key Display indication
Sensor
m(d)
4.700
Para-
OK
OK
OK
OK
1)
In case of two values which can be modified, you
meter adjust­ment
Adjust­ment
w.o. medium
Ad­just­ment in
(min. adjustment)
m(d)
0.0 %
at
m (d)
XX.XXX
1)
can move with the "+" key to the second value (confirm with the "OK" key).
Adjustment with medium
Fill the vessel e.g. to 10 %, enter in the menu "
Min-adjustment
" with the "+" and "–" keys 10 % and confirm with the "OK" key. Then fill the vessel e.g. to 80 % or 100 %, enter in the menu "
Max-adjustment
" with the "+" and "–" keys 80 % or 100 % and confirm with the "OK" key.
with medium
Max.
Min.
adjust
adjust
at %
at %
XXX.X
XXX.X
VEGAFLEX 54K 23
Setup
2. Conditioning
Under the menu item " choose the distance at 0 % and at 100 % filling. Then you enter the physical quantity and the unit as well as the decimal point.
• Enter in the menu window "
spond
" the value of the 0 % filling. This will
be e.g. "80" for 80 liters.
• Confirm with "OK".
• With the arrow key you switch to the 100 % menu. Enter here the value of your param­eter corresponding to a 100 % filling. E.g. "1200" for 1200 liters.
• Confirm with "OK".
• If necessary, choose a decimal point. Note that max. 4 digits can be shown.
• In the menu " parameter (mass, volume, distance…),
• In the menu " unit (kg, l, ft3, gal, m3 …).
• With the "ESC" key, you go back to the next higher menu level. With the arrow key, you choose the next menu item.
• In the menu between three standard linearisation curves. A linear correlation between percentage value of the product distance and the per­centage value of the filling volume has been preset. You can choose between linear, spherical and cylindrical tank. Entering a user-pro­grammable curve is only possible with the PC and the adjustment program VVO.
• In the menu item set a delay period for the signal output.
Conditioning
0 % corre-
prop. to
", you choose the
Unit
", you choose the physical
"Lin. curve",
you can choose
"Integration time",
" you
you can
3. Outputs
Under the menu " e.g. if the current output should be inverted, or which parameter should be indicated on the sensor display.
Outputs
", you determine
4. Operating range
Without special adjustment, the operating range corresponds to the measuring range. The measuring range was already adjusted with the min./max. adjustment. It is generally better to choose a slightly wider span for operating range (approx. 5 %) than for the measuring range (span).
Example: Min./Max. adjustment: 0.300 … 5.850 m; set operating range to approx. 0.250 … actual cable length.
5. Meas. conditions
With these functions, you can enter the ambi­ent conditions in the vessel (see menu sche­matic).
24 VEGAFLEX 54K
Setup
6. Useful level and noise level
In the menu tion on the signal quality of the product echo. The higher the " the measurement.
Ampl.: Means amplitude of the product echo
S-N: Means Signal-Noise (useful signal
The higher the S-N value (useful signal level minus noise level), the higher the reliabil­ity: > 30 dB Measurement good 10 … 30 dB Measurement satisfactory < 10 dB Measurement bad
"Info"
you get important informa-
S-N
" value, the more reliable
in dB (useful signal level)
level minus noise level)
VEGAFLEX 54K 25
Menu schematic of the adjustment module MINICOM
Setup
Sensor
m(d)
4.700
Para­meter adjust­ment
Sensor optimize
Sensor Tag
Sensor
Meas. enviro­nment
Opera­ting range
Begin
m (d)
0.50
FLEX54
After switching on, the sensor type and the
K
software version are displayed for a few seconds.
0.50
4.
Cable length
m (d)
6.00
Meas. condit­ions
Low DK value No
5.
Empty vessel profile
Create new
Meas. dist.
m (d)
X.XX
Create new Now !
OK ?
Learn­ing!
Update
Meas. dist.
m (d)
X.XX
Update Now !
OK ?
Learn­ing!
Delete
Delete Now !
OK ?
Dele­ting !
Adjust­ment
w.o. medium
Adjust­ment in
m(d)
1.
with medium
Min. adjust at %
XXX.X
0.0 %
at
m (d)
XX.XXX
Max. adjust at %
XXX.X
100.0%
at
m (d)
XX.XXX
Note:
Set up the sensor in the sequence of the numbers.
Signal condit­ioning
Scaling
0 % corres ponds
XXXX
2.
Lin. curve
Linear
100 % corres ponds
XXXX
Integr ation time
Deci­mal point
888.8
0 s
prop.
Unit
to
Mass
kg
26 VEGAFLEX 54K
Setup
Add’l func­tions
With these keys you move in the menu field to the left, the right, top and bottom.
ESC
OK
Outputs
mA output
act. dist.
m (d)
4.700
Ampl.:
XX dB
S-N:
XX
dB
Info
Reset to de­fault
Reset Now!
OK ?
Reset runs!
Lan­guage
Eng­lish
Meas. unit
m (d)
6.
Sensor Tag
Sensor
Sensor type
FLEX54 K
Serial number
1094 0213
Softw. vers.
1.00
Softw. date
15.09. 1997
dist.
m (d)
4.700
S-N:
XX dB
XX
dB
Ampl.:
act.
3.
Simu­lation
Sensor displ.
Simu­lation Now!
OK ?
Simu­lation
XXX.X
%
act. dist. m
X,XX
Meas. unit
m (d)
Bolt print figures are sensor or measured value information and cannot be modified in this posi­tion.
The menu items in white letters stand for figures which can be modified with the "+" or "–" key and saved with the "OK" key.
mA output
4-20mA
prop. to
Di­stance
Failure mode
22mA
VEGAFLEX 54K 27
Setup

4.3 Adjustment with the PC

Connection
The PC with the adjustment program VVO (VEGA Visual Operating from version 2.80) can be connected to the
- sensor or
- signal cable.
PC on the sensor
To connect the PC to the sensor, the interface converter VEGACONNECT 2 or 3 is required. Plug it into the provided CONNECT sockets in the sensor.
PC on the signal cable
Connect the two-wire cable of VEGACONNECT 2 / 3 to the supply cable (two-wire sensor) of the sensor. If the resist­ance of the systems (PLC, current source etc.) connected to the supply cable is less than 250 , a resistor of 250 … 350 must be connected to the supply cable during adjustment. The digital signals modulated onto the signal cable would otherwise be considerably damped (short-circuited) through insufficient resistance, and communi­cation with the PC would not be ensured.
+
250 ... 350
-
You have connected the PC with the adjust­ment software VVO to your measuring sys­tem.
• Now switch on the power supply of the connected sensor.
In the first 10 … 15 seconds the sensor starts to draw a current of approx. 22 mA (selftest) and then takes on a level-proportional or dis­tance-proportional current of 4 … 20 mA.
• Start the adjustment software VVO on your PC.
• In the entrance window, you choose with the arrow keys or the mouse the item
ning"
and click to
You should choose
"OK"
.
"Planning"
"Plan-
only if you are authorised to modify instrument param­eters. Otherwise choose
"Maintenance"
In the window
.
"User identification",
"Operator"
or
you are
asked for the name and the password.
• For setup ( name:
"VEGA"
Planning
"VEGA"
), you enter under
and under password also:
. It does not matter if the words are
entered in small or capital letters.
VVO determines then automatically the type of the connected sensor and indicates a moment later with which sensor a connection exists.
If you get no communication with the sensor, check the following:
- The supply voltage must be at least 15 V at 4 mA or 14 V at 22 mA.
- In case VEGACONNECT 2 / 3 is directly connected to the sensor cable, the load resistance must be 250 … 350 Ohms.
- You have to use a VEGACONNECT 2 or 3. Older versions of VEGACONNECT are not compatible.
VEGACONNECT 2/3
You will find further information on adjustment with the PC and the adjustment software VVO in the operating instructions manual of the adjustment software VEGA Visual Operating (VVO).
28 VEGAFLEX 54K
Setup
Special function
If a VEGAFLEX sensor is found, several special functions of VEGAFLEX can be cho­sen with the adjustment software VVO.
Under the menu item "Instrument data/Pa­rameter adjustment", you can select sensor optimisation. Here you find all special functions of VEGAFLEX:
- Meas. environment
- Echo curve
Meas. environment
Meas. conditions
Select in this menu which medium you want to measure in.
If you want to measure a product with a DK value below 4, you can increase the sensitiv­ity of the sensor here.
Operating range
With this command, you can limit the operat­ing range of the sensor. Carry out the adjustment before modifying the operating range, as a later adjustment will overwrite the values of the operating range.
Echo curve
The echo curve shows all reflections of the guided microwave signal. This means that not only the level signal is shown, but also interference signals, e.g. caused by vessel installations. The sensor generally interprets the highest amplitude as the level echo.
A black arrow is shown above the highest amplitude peak. When clicking to "Display info", a small window will be displayed, listing the detailed information on the selected echo.
The right part of the graphics shows the operating range, the left part the adjustment. If e.g. you do not want to fill your vessel com­pletely, you can limit the operating range.
VEGAFLEX 54K 29
Zoom/Unzoom
After activation of one of these two functions, you can either zoom or unzoom the curve with the left mouse key.
When activating "Zoom", you can choose the requested part of the picture by pressing the left mouse key and drawing a frame.
The following curves can be displayed:
Raw value curve (red)
The red curve represents the absolute sig­nal, detected by the receiver of VEGAFLEX. Beside the useful echo, this curve also con­tains interference signals. The highest ampli­tude is interpreted as the level echo. Check by sounding if the value of the useful echo corresponds to the real distance to the me­dium. If the two values do not correspond, you have to filter out the dominant false ech­oes with the function "Empty vessel profile".
Setup
Start
If you click "Start", the echo curve is updated continuously. With "Stop" you can terminate the update.
Vessel profile (blue)
In normal condition, this curve represents the beginning and end of the vessel. To filter out false echoes, you can modify the blue curve of the vessel profile with the func­tion "Empty vessel profile".
Reference line (green)
All amplitudes of the red curve which are below the green reference line will be sup­pressed (ringing, noise, false echoes, etc.). All amplitudes above this line are possible echoes that will be evaluated by the software.
Logarithmic raw value curve (black)
This curve represents the difference between raw value curve (red) and reference line (green). The scale on the right (dB) applies to this curve.
Documentation
If you click the button "Documentation", the immediate echo curve will be saved.
Echo data
If you click the button "Echo data", a window is displayed in which all echoes detected by the sensor are listed with dB information and a probability evaluation.
Empty vessel profile
With the functions in this window, you can filter out false echoes. The blue empty vessel profile curve in the echo curve window represents the actual empty vessel profile.
Determine the level by sounding. It is possible to filter out the echo if the sensor has inter­preted a false echo as the probable level echo instead of the sounded level. Choose the func­tion "Create new". Enter the sounded distance to the medium. All false echoes within the range of the adjusted distance are gated out auto­matically. Please note that the blue curve of the empty vessel profile also changes.
With "Update" you can extend the empty vessel profile, in case new false echoes appear when the level falls. If you want to delete the existing empty vessel profile, click "Delete".
If VEGAFLEX is mounted on a socket or a mounting boss, the measurement may be inaccurate at close range. If the socket or mounting boss exceeds the following lengths, you should carry out an empty vessel profile storage: Thread 1", 11/2" - length > 50 mm Flange DN 50 - length > 50 mm Flange DN 80 - length > 80 mm Flange DN 100 - length > 100 mm
30 VEGAFLEX 54K
Setup

4.4 Adjustment with HART® handheld

The VEGAFLEX sensors can be set up with any HART® handheld. A special DDD (Data Device Description) is not necessary. Just connect the HART® handheld to the signal cable after having connected the sensor to power supply.
Note
If the resistance of the signal current circuit is less than 200 , a resistor of 250 … 350 must be connected to the signal/connection cable during adjustment. Simply loop the resistor into the sensor connection cable (see figure).
+
250 ... 350
The most important adjustment steps
You will find the complete adjustment proce­dure via HART® protocol in the operating instructions of the HART® handheld.
To enter parameters, first press "
Generic: SENSOR PV LRV
5.850 m
0.300 m
HELP DEL ABORT EN TER
The entered value will be saved in the handheld, but not in the sensor itself. Press "
SEND
" to transmit the entered value to
the sensor.
ENTER
".
-
SPS
PLC
Generic: SENSOR
1 LRV 5.850 m 2 PV URV 0.300 m
HELP SEND HOME
Press "OK" to acknowledge the following warning. Follow the instructions on the dis­play.
Press again "OK", and the adjustment that was just carried out is displayed.
Generic: SENSOR
1 LRV 5.850 m 2 PV URV 0.300 m
HELP HOME
VEGAFLEX 54K 31
Diagnosis

5 Diagnosis

5.1 Simulation

To simulate a certain filling, you can call up the function "Simulation" on the adjustment module MINICOM, in the software program VVO or in the HART® handheld. This function simulates a certain current value. Please keep in mind that connected devices, such as e.g. a PLC, react according to their settings and will probably activate alarms or system functions.

5.2 Failure rectification

Error Corrective measure
E 013 Sensor does not find a - Message is displayed during the warm-up
E 017 Adjustment span too small Carry out the adjustment again. Make sure that
E 036 Software update incorrect Return the instrument for repair.
E 040 Hardware failure/ Check all connection cables.
level echo phase.
- If the message remains, the DK value of the medium might be too low.
Check the useful level and noise level. See "4.2 Adjustment with adjustment module MINICOM; 6. Useful level and noise level". If the message still remains, carry out a new adjustment.
the difference between min. and max. adjust­ment is at least 10 mm.
Electronics defective Contact our service department.
32 VEGAFLEX 54K
Notes
VEGAFLEX 54K 33
Notes
34 VEGAFLEX 54K
Notes
VEGAFLEX 54K 35
VEGA Grieshaber KG Am Hohenstein 113 D-77761 Schiltach Phone (07836) 50-0 Fax (07836) 50-201 E-Mail info@de.vega.com
www.vega.com
ISO 9001
All statements concerning scope of delivery, application, practical use and operating conditions of the sensors and processing sys­tems correspond to the latest information at the time of printing.
Technical data to subject to alterations
2.26 362 / November 2001
Loading...