VEGA FLEX52V User Manual

Operating Instruction
VEGAFLEX 51V and 52V
Level and Pressure
Safety information / Note Ex-area

Safety information

The described module must only be installed and operated as described in this operating instruction. Please note that other action can cause damage for which VEGA does not take responsibility.
2 VEGAFLEX 51V and 52V

Note Ex-area

Please note the attached approval documents (yellow binder) and especially the included safety data sheet.
Contents
Contents
Safety information ........................................................................ 2
Note Ex-area ............................................................................... 2
1 Product description
1.1 Function ................................................................................ 4
1.2 Application features ............................................................. 5
1.3 Types and versions............................................................. 7
1.4 Type code ............................................................................ 8
1.5 Technical data ...................................................................... 9
1.6 Approvals .......................................................................... 12
1.7 Dimensions ........................................................................ 12
2 Mounting
2.1 Installation instructions ...................................................... 14
3 Electrical connection
3.1 Connection and connection cable .................................... 18
3.2 Connection of the sensor .................................................. 20
3.3 Connection of the indicating instrument VEGADIS 50 .... 21
4 Set-up
4.1 Adjustment structure ......................................................... 22
4.2 Adjustment with adjustment module MINICOM ............... 23
4.3 Adjustment with PC ........................................................... 28
4.4 Adjustment with signal conditioning instrument .............. 31
5 Diagnosis
5.1 Simulation ........................................................................... 32
5.2 Error removal ..................................................................... 32
VEGAFLEX 51V and 52V 3

1 Product description

1.1 Function

High frequency microwave impulses are guided along a steel rope. When reaching the product surface, the microwave impulses are reflected. The running time of the im­pulses is processed by the integral electron­ics and provided as level.
The sensors measure levels of all kind of solids. Density, conductivity and dielectric constant of the medium do not influence the measurement. Also varying solid characteris­tics do not influence the measured value.
Lime, cement, cereals, plastic granules, flour, gravel - in solids VEGAFLEX microwave sensors are the ideal solution for arduous applications. The level is also reliably de­tected in products with varying humidity or fluctuating dielectric constant. High, narrow vessels in which non-contact measuring principles often do not deliver optimum mea­suring results, VEGAFLEX can be used with­out problems. VEGAFLEX 51 and 52 mainly differ in the thickness of the measuring cable. VEGAFLEX 51 has a cable with 4 mm diam­eter and is available up to a length of 10 m. The cable of VEGAFLEX 52 has a diameter of 8 mm and can have a length of up to 20 m. With the thicker cable it is also suitable for solids with higher extraction forces.
Product description / Function
• Measuring range up to 20␣ m with small min. distance
• Up to 15 sensors can be connected via one two-wire cable
• Two-wire sensor, supply and digital measuring signal via one two-wire cable
• Independent on application conditions such as:
- dust and noise generation
- humidity changes
- varying products
- build-up
• Adjustment without vessel filling or emp­tying
• Individual wiring, with 15 sensors on one two-wire line (digital output signal)
• Unaffected by vessel material, e.g. metal, concrete, plastic etc.
4 VEGAFLEX 51V and 52V
Product description / Application features

1.2 Application features

Application features
• Level measurement in solids
• Measurement also in vacuum
• All slightly conductive and all substances
with a dielectric constant εr > 1,7 can be measured
• Measuring ranges VEGAFLEX 51 0 … 10 m VEGAFLEX 52 0 ... 20 m
Two-wire technology
• Supply, signal transmission and output
signal on one two-wire cable
• digital output signal
• low wiring costs due to the use of bus
systems or two-wire technology
Rugged
• High resistance materials: PA, 1.4401
Exact and reliable
• Resolution 1 mm
• Unaffected by noise, vapours, dusts, gas
compositions and layering
• Unaffected by varying density
• Measurements under pressure up to
16 bar and medium temperatures up to 120°C
Communicative
• Integral measured value indication
• Indication either separated from the sensor
• Adjustment from the DCS-level
• Connection to all bus systems:
Interbus S, Modbus, Siemens 3964R, Profibus DP, Profibus FMS, ASCII
Adjustment with adjustment module MINICOM
With the 6-key-adjustment module MINICOM (3,2␣ cm␣ x␣ 6,7␣ cm) you can carry out the ad­justment in clear text dialogue. The adjustment module can be plugged into VEGAFLEX or into the external indicating instrument VEGADIS 50. VEGAFLEX can therefore also be adjusted from the external indicating instrument VEGADIS 50.
Pluggable adjustment module MINICOM
Adjustment with VEGA signal condi­tioning instruments
VEGAFLEX-series 50V sensors can be also operated with a VEGAMET signal condi­tioning instrument. The display leads you in clear text. With the 6-key adjustment field the parameter adjust­ment can be carried out.
Approvals
Dust-Ex - Zone 10
VEGAFLEX 51V and 52V 5
Adjustment with PC
The set-up and adjustment of the VEGAFLEX-sensors can be made via PC with the adjustment program VVO (VEGA Visual Operating) under Windows®. The software leads you quickly through the adjustment and parameter adjustment by means of pictures, graphics and process visualisations.
The PC can be connected to any individual position of the system or the signal line. It is therefore connected with the two-wire PC­interface converter VEGACONNECT 2 to the sensor or to the signal line.
The adjustment and parameter adjustment data can be saved with the adjustment soft­ware on the PC. If necessary the adjustments can also be transferred to other sensors.
On the following figures you see the different connection possibilities.
Product description / Application features
Adjustment with the PC on the digital signal and supply line to the processing system or on the sensor
Adjustment with the PC on the digital signal and supply line between sensor and VEGAMET signal conditioning instrument
Adjustment with the PC and the standard cable RS 232 on the processing system
Adjustment with the PC on the signal condi­tioning instrument
6 VEGAFLEX 51V and 52V
Product description / Types and versions

1.3 T ypes and versions

VEGAFLEX series 50 V
Generally all VEGAFLEX series 50 V sensors can be adjusted with the pluggable adjust­ment module MINICOM or with a VEGAMET signal conditioning instrument. The sensor can be adjusted with the software VEGA Visual Operating (VVO) also via PC.
VEGADIS 50
1
Configuration
A measuring system with a VEGAFLEX is possible in different ways. See the following figures. The external indicating instrument VEGADIS 50 can be mounted up to 25 m separated from the sensor.
Two-wire technology
1 Connection to a VEGAMET 514V, 515V,
614V signal conditioning instrument
2 Connection to a VEGALOG 571
4 … 20 mA
VEGAFLEX
e.g. VEGAMET 514V
2
4 … 20 mA passive
VEGALOG
VEGAFLEX 51V and 52V 7

1.4 Type code

The second figure of the type designation, e.g. VEGAFLEX 5[1]… distinguishes the instruments according to the stability of the cable.
The figure e.g. VEGAFLEX 51[V] character­izes the output signal: K stands for an analogue 4␣ …␣ 20␣ mA output signal (compact instrument) V stands for a digital output signal (VBUS).
VEGAFLEX XX X EXS.X X
B - 20…72 V DC, 20...250 V AC / 4...20 mA-output, HART D - Two-wire loop-powered / 4...20 mA-output, HART E - Supply via signal conditioning instrument / V-BUS-output
Dust-Ex approved
V - Digital output signal (two-wire technology) K - Analogue 0 … 20 mA output signal (two-wire or
four-wire technology)
Type 51: Instrument series with 4 mm-cable Type 52: Instrument series with 8 mm-cable
Product description / Type code
®
®
Meas. principle (FLEX for guided microwave)
8 VEGAFLEX 51V and 52V
Product description / Technical data

1.5 T echnical data

Power supply
Supply voltage from signal conditioning instrument VEGAMET
Current consumption max. 22,5 mA Power consumption max. 80 mW, 0,45 VA
Measuring range
VEGAFLEX 51 0,3 … 10 m VEGAFLEX 52 0,3 ... 20 m
Output signal
Digital measuring signal (VBUS)
Adjustment
- signal conditioning instruments VEGAMET 514V, 515V, 614V
- PC and adjustment software VEGA Visual Operating
- adjustment module MINICOM
Accuracy (under reference conditions acc. to IEC 770 - relating to the max. meas. range)
Linearity error < 0,1 % Temperature drift 0,015 %/10 K Resolution 1 mm
or from the processing system VEGALOG 571 (max. 36 v DC) fuse 0,5 A (slow-blow)
Characteristics
Min. span between full and empty adjustment
- digital output signal 10 mm
Ambient conditions
Vessel pressure -1 … 16 bar Ambient temperature on the housing -40°C … +60°C Process temperature -40°C … +120°C Storage and transport temperature -40°C … +80°C Protection IP 66/IP 67 (meets both protections) Protection class
- two-wire sensor II
Overvoltage category III Max. tensile load cable → 4 mm 5 KN
Ex-technical data
Dust-Ex Zone 10 The permissible operating data of the VEGAFLEX-sensors for dust-Ex-areas are stated in the certificate.
VEGAFLEX 51V and 52V 9
cable 8 mm 20 KN
Product description / Technical data
Materials
Housing PBT (Valox) or
Aluminium (powder coated)
Cable 1.4401 (stst; 316) or
galvanized steel - PA-coated
Process connections
VEGAFLEX 51, 52 G 11/2 A, 11/2“ NPT
steel chromated or of 1.4301 (stst)
Connection lines
- supply and signal via one two-wire cable, the line resistance depends on the supply voltage (see diagram).
Cross-section area of conductor generally 2,5 mm Earth connection max. 4 mm
2
2
Cable entry 2 x M20 x 1,5 (cable diameter 5 … 9 mm)
Voltage limit Ex-sensor
Voltage limit not-Ex-sensor
Supply voltage
CE-conformity
VEGAFLEX sensors meet the protective regulations of EMVG (89/336/EWG) and NSR (73/23/EWG). The conformity has been judged acc. to the following standards: EMVG Emission EN 50 081 - 1: 1992
Susceptibility EN 50 082 - 2: 1995
NSR EN 61 010 - 1: 1993
Display indication
Display scalable analogue and digital measured
value indication (option). An external measured value indication, supplied by the sensor, can be mounted up to 25 m separated from the sensor.
10 VEGAFLEX 51V and 52V
Product description / Technical data
Signal output VEGAFLEX 51V, 52V
Signal output digital output signal in two-wire technology
(VBUS)
Two-wire technology: The digital output signal (measuring signal) is modulated to the power supply and further processed in the signal conditioning instrument or in the processing system.
VEGAFLEX 51V and 52V 11
Product description / Approvals

1.6 Approvals

VEGAFLEX sensors are approved for dust­Ex-Zone 10. Note the attached approval documents when using a sensor in Ex-area.

1.7 Dimensions

VEGAFLEX 51
Plastic housing Aluminium housing Aluminium housing (ExD)
Cable - steel with gravity weight
12 VEGAFLEX 51V and 52V
Cable - plastic coated with gravity weight
Cable - steel with cable loop
Cable - plastic coated with cable loop
Product description / Dimensions
VEGAFLEX 52
Plastic housing Aluminium housing Aluminium housing (ExD)
Cable - steel with gravity weight
Cable - plastic coated with gravity weight
Cable - steel with cable loop
Cable - plastic coated with cable loop
Adjustment module MINICOM Cable loops
Adjustment module for plugging into series 50 sensors or into the external indicating instrument VEGADIS 50
Cable - steel
VEGAFLEX 51V and 52V 13
Cable ­plastic coated

2 Mounting

2.1 Installation instructions

Lateral load
Note that the measuring cable is not sub­jected to strong lateral forces. Mount VEGAFLEX to a position in the vessel where no interfering influences such as e.g. of stir­rers, filling openings etc. can occur (see fig.
2.1).
Fig. 2.1 Lateral load
Mounting / Installation instructions
Shortening of the measuring cable
The cable can be shortened afterwards. Loosen the three pins on the gravity weight (hexagon) and remove the pins. Pull the cable out of the gravity weight.
To avoid splicing of the steel cable during cutting, you have to tin the cable around the cutting position with a copper bit or strongly tighten the cable with a wire. Shorten the cable with a metal cutting saw or a cutting-off wheel by the requested length.
In case of isolated cables with gravity weight, the isolation should be removed on a certain length (see fig. 2.2). VEGAFLEX 51 45 mm VEGAFLEX 52 60 mm In case of cable loops the cable should re­main isolated.
Extraction forces
In case of strong extraction forces, e.g. dur­ing quick emptying or settling of solids, high tensile loads can be caused. When such an installation position is neces­sary, the cable should not be strained but only equipped with a gravity weight as the cable can more easily follow the product movements. Note that the cable does not touch the vessel wall.
Pressure
In case of gauge or low pressure in the ves­sel, the mounting boss should be sealed on the thread. Use the attached seal ring. Check if the seal ring is resistant against the me­dium.
14 VEGAFLEX 51V and 52V
Fig. 2.2 Shortening of the measuring cable
Mounting / Installation instructions
Cable entries
When mounting outside, on cooled vessels or in humid areas where cleaning is made e.g. with steam or high pressure, the seal of the cable entry is very important.
Use cable with a round cross-section area of conductor and tighten the cable entry. The cable entry is suitable for cable diameters of 5␣ mm to 9␣ mm.
Humidity from outside
To avoid humidity ingress, the connection line to the housing should be looped directly behind the cable entry to the bottom so that rain and condensation water can drain off. This is mainly valid for mounting outside, in areas where humidity is expected (e.g. by cleaning processes) or on cooled or heated vessels (see fig. 2.3).
Fig. 2.3 Humidity
VEGAFLEX 51V and 52V 15
Mounting / Installation instructions
Material cone
Note when installing VEGAFLEX into the ves­sel that material cones can be caused with solids which can change the measured value. We recommend to choose an installa­tion place where the cable detects an aver­age value of the material cone.
Dependent on the position of the filling and emptying opening in the vessels, VEGAFLEX should be installed appropriately. For com­pensation of measuring errors caused by the material cone, you should install the cable at a distance of d/6 from the vessel wall or vessel installation. In any case keep a min. distance of 300 mm.
3
1 Emptying 2 Filling opening 3 VEGAFLEX
16 VEGAFLEX 51V and 52V
1
2
Mounting / Installation instructions
Straining
Generally the measuring result is only slightly influenced when the gravity weight is in con­tact with the vessel wall. Dependent on the kind of solid, the installa­tion or the kind of filling, the cable can "float" (fig. 2.4) despite of the gravity weight. The solid can push the cable to the vessel wall or to the top and wrong measured values are caused. In this case fasten the cable. When straining the cable avoid strong cable forces. In our pricelist you will find a spring as accessory avoiding overloading of the cable.
Build-up
Generally build-up on the cable causes changes of the measured value. If strong build-up on the cable has to be expected, it is recommended to use an isolated cable. The higher surface quality avoids build-up.
Abrasion
In very abrasive products, e.g. the measure­ment of grit, use a measuring cable without isolation.
Socket
If possible avoid long sockets.
Fig. 2.4 Cable electrode in solids
VEGAFLEX 51V and 52V 17
Electrical connection / Connection and connection cable

3 Electrical connection

3.1 Connection and connection cable

Safety information
Ensure that the instrument is in currentless condition. Always switch off the power sup­ply before you carry out terminal work on the radar sensors. Protect yourself and the in­strument, especially when you use sensors which do not work with low voltage.
Skilled staff
Instruments which are not operated with a protective low voltage must only be con­nected by skilled staff.
Connection
For connection a standard two or four-wire cable (sensors with separate supply) with max. 2,5 mm2 can be used. Very often "Elec­tromagnetic pollution" by electronic actuators, energy lines and emitting systems is so considerable that the two or four-wire cable should be screened.
We recommend to use screened cable. This screening prevents against future interfer­ences (figure 3.1 a). It is favourable to earth the screens on both ends. However it must be noted that no earth compensation currents flow via the sensor cable screens (figure 3.1 b). Earth compen­sation currents can be avoided by connect­ing the cable screen in case of earthing on both sides on one earthing side (e.g. in the switching cabinet) via a capacitor (e.g. 0,1␣ µF; 250␣ V AC) to earth potential. Use a very low impedance earth connection (foun­dation, plate or mains earth). In dust-Ex-applications the screen should be on one end.
Ex-protection
If an instrument is used in dust-Ex areas, the appropriate regulations as well as the dust­Ex-certificate of VEGAFLEX for systems in dust-Ex-areas should be noted.
Connection cable
Note that the connection cables must be speci­fied for the expected operating conditions in your systems. The cable must have an outer diameter of 5␣ …␣ 9␣ mm (1/5 to 1/3␣ inch). Other­wise the seal effect of the cable entry will not be ensured.
Cables for intrinsically safe circuits must be marked blue and must not be used for other circuits.
Earth conductor terminal
On VEGAFLEX sensors the earth conductor terminal is galvanically connected with the metal thread.
18 VEGAFLEX 51V and 52V
Electrical connection / Screening
Earthing on one sensor side
Processing
Earthing on both ends (on the signal conditioning instrument via the potential separating capacitor)
Fig. 3.1 a
0,1 µF 250 V AC
Processing
In dust-Ex-applications the screening should only be made on one side. Due to the potential difference earthing on both sides is not al-
Fig. 3.1 b
lowed.
VEGAFLEX 51V and 52V 19

3.2 Connection of the sensor

• Loosen the closing screw on the upper
side of the sensor.
• Open the sensor cover.
• Remove the compression nut of the Pg
and shift the nut a little bit over the con­nection cable.
• Remove the rubber seal out of the Pg and
shift the seal over the connection cable.
• Remove the cover of the connection cable
over a length of approx. 10 cm.
• Loop the cable through the Pg into the
sensor.
Electrical connection / Connection of the sensor
• Connect the cable. Push the white buckets of the spring terminals with a small screwdriver and insert the copper core of the connection line into the terminal opening. Check the position of the cables in the terminal by slightly pulling.
• Fasten the compression nut on the Pg.
VBUS
To the indicating instrument in the sensor cover or to the external indicat­ing instrument VEGADIS 50
Two-wire sensor
M20 x 1,5 (diameter of the connection cable 5…9␣ mm)
20 VEGAFLEX 51V and 52V
Set-up / Connection of the indicating instrument VEGADIS 50

3.3 Connection of the indicating instrument VEGADIS 50

Loosen the 4 screws of the housing cover on VEGADIS 50. You can facilitate the connection procedure by fastening the housing cover during con­nection on the housing by means of the screws (figure).
Adjustment module
VEGADIS 50
Screws
VBUS
to VEGADIS 50 (use screened cable. See fig. 3.1 a or fig. 3. 1 b)
T wo-wire sensor
M20x1,5
VEGAFLEX 51V and 52V 21

4 Set-up

4.1 Adjustment structure

Series 50 sensors can be adjusted
- with PC (adjustment program VVO)
- with detachable adjustment module MINICOM
- with a signal conditioning instrument VEGAMET (V)
The sensors can only be adjusted with one of the possibilities at the same time. Independent whether you set-up a measur­ing system (unit of sensor and signal condi­tioning instrument VEGAMET or sensor and processing system VEGALOG) with the adjustment software VVO, with the signal conditioning instrument or with the adjustment module MINICOM, the adjustment procedure is always the same:
- first of all configure a measuring system
in the menu "
- carry out the parameter adjustment of the
sensors in the menu "
Adjustment module MINICOM
With the adjustment module MINICOM you adjust directly in the sensor or in the external indicating instrument VEGADIS 50. The ad­justment module MINICOM enables with the 6-key adjustment field with text display the adjustment of the sensor in the same function volume as with the adjustment program VVO.
Configuration
Instrument data
“ and then
“.
Set-up / Adjustment structure
Adjustment program VVO
With the adjustment program VVO (VEGA Visual Operating System) on the PC you adjust the sensors very comfortably. The PC communicates via the interface adapter VEGACONNECT␣ 2 with the sensor. The si­gnal and supply line is therefore superim­posed by a digital adjustment signal. The PC can be connected to any position on the cable or directly to the sensor.
Signal conditioning instrument VEGAMET (V)
The signal conditioning instrument VEGAMET enables with the 6-key adjustment field with text display the parameter adjustment in the same function volume as with the adjustment program VVO on the PC.
22 VEGAFLEX 51V and 52V
Set-up / Adjustment module MINICOM
4.2 Adjustment with adjustment module MINICOM
As with the PC, the sensor can also be ad­justed with the small detachable adjustment module MINICOM. The adjustment module is therefore inserted into the sensor or the external indicating instrument (optionally).
For the adjustment with adjustment module also all adjustment options are available as on the PC with adjustment program VVO. All adjustment steps can be carried out with the 6 keys of the adjustment module. A small display gives you beside the measured value a short information on the menu point or the figure of a menu adjustment.
Adjustment elements
The adjustment module MINICOM is menu orientated. The clear text indications on the display lead through the menu. The functions of the keys are described in the following.
OK-key (4)
With the OK-key adjustments can be con­firmed. When the symbol or is displayed, you can move with the OK-key to the next lower menu stage. With the symbol there is no branch below the menu point but only another menu point of the appropriate function.
ESC-key (3)
With the Escape-key (ESC) you can depen­dent on the menu point interrupt an adjust­ment or a function or change to the next higher menu stage. E.g. to reach the top menu stage, push the ESC-key several times.
+ and - key (2)
With the keys + and - you can modify the values of the parameters or choose out of several possibilities. After pushing the first time, the value to be adjusted flashes. The value will be modified by further pushing the key.
Arrow keys (5)
With the keys > and < you can move within the menu stage from one menu point to the other.
Digital indication (1)
On the digital indication the actual measured value is displayed during operation. When you adjust the instrument, the clear text indication displays the appropriate function.
Branch out of which you can change to a
lower menu.
This symbol informs you about a follow-
ing safety enquiry.
▼ ▼
▼ ▼
2
1
5
VEGAFLEX 51V and 52V 23
3
4
Set-up / Adjustment module MINICOM
Adjustment steps
On the following pages you see the menu plan of the adjustment module MINICOM.
Set-up the sensor in the following sequence. The numbers correspond to the sequence of a set-up. The numbers are stated near the appropriate menu points in the menu plan on the following pages.
1. Adjustment
2. Conditioning
3.Outputs
4. Operating range
5. Meas. conditions
6. Indication of the useful and noise level
1. Adjustment
Under the menu point " form the sensor with which span it should work.
Max.
Adjustment
100 % (distance to the medium 0,300 m) correspond to 1200 l, e.g. 20 mA
“ you in-
Span
Adjustment without medium
• Enter with the "+“ and "–“ key the dis­tance to the medium your sensor has at 0 % filling (example: 5,850 m) If you do not know the distance, you have to sound.
• Enter the distance to the medium which your sensor has at 100 % filling.
Key Display
Sensor
m(d)
4.700
Para-
OK
OK
OK
OK
meter adjust­ment
Ad­just­ment
without medium
Ad­just­ment in
(Min-adjustment)
m(d)
0.0 %
at
m (d)
XX.XXX
1)
1)
In case of two values which can be modified you can move with the "OK“-key to the sec­ond value.
Min.
0 % (distance to the medium 5,850 m) correspond to 456 l, e.g. 4 mA)
Adjustment with medium
Fill the vessel e.g. to 10 %, enter in the menu "
Min-adjustment
“ with the "+“ and "–“ keys 10 % and confirm with the "OK“-key. Then fill the vessel e.g. to 80 % or 100 %, enter in the menu "
You can carry out the adjustment without or with medium. Generally you will carry out the
keys 80 % or 100 % and confirm with the "OK“-key.
adjustment without medium as you can adjust without vessel filling.
24 VEGAFLEX 51V and 52V
Max-adjustment
“ with the "+“ and "–“
with medium
Max.
Min.
adjust
adjust
at %
at %
XXX.X
XXX.X
Set-up / Adjustment module MINICOM
2. Conditioning
Under the menu point " choose the distance at 0 % and at 100 % filling. Then you enter the parameter and the appropriate unit as well as the decimal point.
• Enter in the menu window "
sponds
“ the value of the 0 %-filling. This
will be e.g. "80“ for 80 l.
• Confirm with "OK“.
• With the arrow key you change to the 100␣ % menu. Enter here the value of your parameter corresponding to a 100 %­filling. E.g. "1200“ for 1200 l.
• Confirm with "OK“.
• Choose if necessary a decimal point. Note that max. 4 digits can be shown.
• In the menu " parameter (mass, volume, distance…).
• In the menu " cal unit (kg, l, ft3, gal, m3 …).
• With the ESC-key you change to the superimposed menu stage. With the arrow key you choose the next menu point.
• In the menu between three standard linearisation curves. Preadjusted is a linear dependence be­tween percentage value of the product distance and the percentage value of the filling volume. You can choose between linear, spherical tank and cylindrical tank. The adjustment of a user programmable curve is only possible with the PC and the adjustment program VVO.
• In the menu point can adjust a delay for the signal output.
Conditioning
0 % corre-
prop. to
“ you choose the
Unit
“ you choose the physi-
"Lin. curve“
you can choose
"Integration time“
“ you
you
3. Outputs
Under the menu " display and determine which parameter should be provided by the sensor display.
Outputs
“ you can scale the
4. Operating range
Without special adjustment, the operating range corresponds to the measuring range. The measuring range was already adjusted with the Min/Max-adjustment. It is generally suitable to choose a slightly bigger operating range (approx. 5 %) than the measuring range (span). Example: Set Min/Max-adjustment: 0,300 … 5,850 m; operating range to approx. 0,250 … real cable length. When you shorten the cable, you have to define the cable length again.
5. Meas. conditions
With these functions you can enter the ambi­ent conditions in the vessel. (see menu plan)
6. Useful and noise level
In the menu tion on the signal quality of the product echo. The higher the " the measurement.
Ampl.: Means amplitude of the product echo
S-N: Means Signal-Noise (useful level
The higher the S-N-value (useful level minus noise level), the better the reliability: > 50 dB Measurement very good 40 … 50 dB Measurement good 20 … 40 dB Measurement satisfactory 15 … 20 dB Measurement sufficient < 15 dB Measurement bad
"Info“
you get important informa-
S-N
“-value, the more reliable
in dB (useful level)
minus noise level)
VEGAFLEX 51V and 52V 25
Set-up / Menu plan of the adjustment module MINICOM
Menu plan of the adjustment module MINICOM
Sensor
m(d)
4.700
Para­meter adjust­ment
Sensor opti­mize
Sensor Tag
Sensor
Meas. environ ment
Opera­ting range
Begin
m (d)
0.50
FLEX51
When switching on, the sensor type and the
V
software version are displayed for a few seconds.
0.50
4.
Cable length
m (d)
6.00
Meas. condit ions
Condi­tion
Solid
5.
fast change
No
agitated sur face
No
Low DK value
No
Empty vessel pro file
Create new
Meas. dist.
m (d)
X.XX
Create new Now!
OK ?
Learn­ing !
Update
Meas. dist.
m (d)
X.XX
Update now!
OK ?
Learn­ing !
Delete
Delete Now!
OK ?
Dele­ting !
Adjust­ment
Adjust­ment in
m(d)
1.
0%
at
m (d)
XX.XXX
100 %
at m (d)
XX.XXX
Note:
Set-up the sensor in the sequence of the numbers.
Signal condit­ioning
Sca­ling
0 % corres­ponds
XXXX
2.
Lin. curve
Linear
100 % corres­ponds
XXXX
Integr ation time
Deci­mal­point
888.8
0 s
prop.
Unit
to
Mass
kg
26 VEGAFLEX 51V and 52V
Set-up / Menu plan of the adjustment module MINICOM
Add’l func­tions
With these keys you move in the menu field to the left, the right, top and bottom.
ESC
OK
Reset to de­fault
Reset Now !
OK ?
Reset runs!
Out­puts
Sensor displ.
act. dist.
m (d)
4.700
Ampl.:
XX dB
S-N:
XX
dB
Info
Reset to de fault
Reset Now !
OK ?
Reset runs !
Lang­uage
Eng­lish
Meas. unit
m (d)
6.
Sensor Tag
Sensor
Sensor type
FLEX51 V
Serial number
1094 0213
Softw. Vers.
1.00
Softw. date
15.09. 1997
dist.
m (d)
4.700
XX dB
S-N:
XX
dB
Ampl.:
act.
3.
Bolt print figures are sensor or measured value information and cannot be modified in this posi­tion.
The menu points in white letters stand for figures which can be modified with the or key and saved with the key.
OK
prop. to
Di­stance
act. dist. m
X,XX
fast change
Yes
VEGAFLEX 51V and 52V 27

4.3 Adjustment with PC

Connection
The PC with the adjustment program VVO (VEGA Visual Operating from version 2.50) can be connected
- to the sensor or
- to the signal line.
PC on the sensor
For connection of the PC to the sensor you require the interface adapter VEGACON­NECT 2. Insert VEGACONNECT 2 into the provided CONNECT-sockets in the sensor.
PC on the signal line
Connect the two-wire line of VEGACONNECT 2 to the signal or supply line of the sensor. When the resistors of the systems (DCS, current source etc.) connected to the signal/ supply line are less than 250 Ω, a resistor of 250 … 350 should be looped into the si­gnal/supply line during adjustment. The digital signals modulated to the signal line would be extremely damped via the low system resistors (i.e. short-circuited) so that the communication with the PC would be interfered.
Set-up / Adjustment with PC
When you have connected the PC with the operating software VVO to your measuring system,
• switch on the power supply of the con­nected sensor.
In the first 10 … 15 seconds the sensor takes a current of approx. 22 mA (self check) and takes then a level proportional or distance proportional current.
• Start the adjustment software VVO (VEGA Visual Operating) on your PC.
• In the entrance window you choose with the arrow keys or the mouse the point
Planning
You should only choose you are entitled to modify instrument parameters. Otherwise choose or In the window asked for the name and the password.
• For set-up ( VEGA and under password also: VEGA. It does not matter if you use capital or small letters.
VVO determines automatically the type of the connected sensor and shows shortly with which sensor a connection exists.
and click to OK.
Maintenance
planning
Planning,
.
User identification
) enter under name:
when
Operator
you are
When you get no sensor connection, check the following:
- The supply voltage should be at least 20 V.
- When connecting VEGACONNECT 2 directly to the sensor line, the load resis­tor should be 250 … 350 Ohm.
- You have to use a VEGACONNECT 2.
VEGACONNECT 2
28 VEGAFLEX 51V and 52V
Older versions of VEGACONNECT are not compatible.
Further information on the adjustment with PC and the adjustment software VVO is stated in the operating instruction of the adjustment software VEGA Visual Operating (VVO).
Set-up / Adjustment with PC
Special functions
When a VEGAFLEX-sensor is determined, you can choose with the adjustment software VVO some special functions of VEGAFLEX.
Under the menu point Instrument data/Pa­rameter adjustment you can choose the sen­sor optimisation. Here you find all special functions of VEGAFLEX:
- meas. environment
- echo curve
Measuring environment
Measuring conditions
Choose in this menu if you want to measure in solids or liquids.
When you want to measure a product the DK-value of which is below 2, you can in­crease the sensitivity of the sensor.
Echo curve
The echo curve shows all reflections of the guided microwave signal. This means not only the product signal is shown but also false signals caused, e.g. by vessel installations. Generally the sensor determines the biggest amplitude as useful echo. You see a black arrow above the highest amplitude. When you click to "Display info“, a small window is displayed, listing the detailed values of the selected echo.
Operating range
With this command you can limit the operat­ing range of the sensor or enter the cable length again after shortening the cable. Carry out the adjustment before modifying the operating range as a later adjustment can overwrite the values of the operating range. The right part of the graphics shows the operating range, the left part the adjustment. When e.g. you cannot fill your vessel com­pletely, you can limit the operating range.
VEGAFLEX 51V and 52V 29
Zoom / Unzoom
After activation of these two functions, you can either zoom or unzoom the curve with the left mouse key. When you activate "Zoom“, you can choose the requested part of the figure by pushing the left mouse key and drawing a frame.
The following curves can be displayed:
Raw value curve (red)
The red curve means the absolute signal which the receiver of VEGAFLEX detects. Beside the useful echo this curve also con­tains false signals. The highest amplitude is detected as useful echo. Check by sounding if the value of the useful echo corresponds to the actual dis­tance to the level. When both values do not correspond, you have to gate out the dominant false echoes with the function "Empty vessel profile“.
Set-up / Adjustment with PC
Start
When clicking to "Start“, the echo curve will be updated continuously. With "Stop“ you can terminate the update.
Vessel profile (blue)
This curve shows in normal condition the actual empty vessel profile. For gating out of false echoes you can modify the blue curve of the vessel profile with the function "Empty vessel profile".
Reference line (green)
All amplitudes of the red curve which are below the green reference line, are sup­pressed (ringing, noise, false echoes etc.). All amplitudes above this line are possible echoes which are processed by the soft­ware.
Logarithmic raw value curve (black)
The curve shows the difference between raw value curve (red) and reference line (green). For this curve the scale on the right page (dB) is valid.
Documentation
When clicking to the button "Documentation“ the actual echo curve will be saved.
Echo data
When clicking to the button "Echo data“, a window is displayed where all echoes de­tected by the sensor with DB-information and a probability processing are listed.
Empty vessel profile
With the functions in this window you can gate out false echoes. The blue empty vessel profile curve in the echo curve window shows the actual empty vessel profile.
Determine the level by sounding. When the sensor has determined instead of the sounded level a false echo as most probable echo, you can gate the echo out. Choose the function "Create new“. Enter the measured distance to the medium. All false echoes, in the range to the adjusted distance, are auto­matically gated out. Note that then also the blue curve of the empty vessel profile changes. .
With "Update“ you can extend the empty vessel profile when new false echoes occur with reducing level. When you want to delete the existing empty vessel profile, click to "Cancel“.
30 VEGAFLEX 51V and 52V
Set-up / Adjustment with signal conditioning instrument

4.4 Adjustment with signal conditioning instrument

VEGAMET (V)
VEGAFLEX 51V and 52V sensors can be adjusted with the following signal condi­tioning instruments:
- VEGAMET 514V
- VEGAMET 515V
- VEGAMET 614V
For adjustment the digital signal conditioning instruments VEGAMET (V) are provided with a 6-key adjustment field with display. The parameter adjustment can be therefore car­ried out in clear text dialogue. The adjustment structure corresponds to the adjustment on the adjustment module MINI­COM. The complete menu plan of the signal condi­tioning instrument is included in the appropri­ate operating instruction of the signal conditioning instrument used.
VEGALOG 571
VEGAFLEX sensors can be also adjusted with the processing system:
- VEGALOG 571
VEGALOG 571 processing system can only be adjusted with the PC in conjunction with the processing software VVO (VEGA Visual Operating). The adjustment of VEGALOG is described in the operating instruction of the adjustment software VVO.
VEGAFLEX 51V and 52V 31

5 Diagnosis

5.1 Simulation

For simulation of a certain filling, you can enquire the function "Simulation" on the ad­justment module MINICOM, in the software program VVO or in the signal conditioning instrument VEGAMET (V). You simulate a certain measured value. Note that connected instruments such as e.g. a DCS react according to their adjustments and probably activate alarm signals or sys­tem functions.
5.2 Error removal
Error Error removal
E 013 Sensor does not find - Message is displayed during feeding
level echo phase.
- When the message remains, the DK-value of the medium can be too low.
Check the useful and noise level. See 4.2 Adjustment with adjustment module MINICOM / 6. Useful and noise level. When the message still remains, carry out a new adjustment.
Diagnosis
E 017 Adjustment span too small Carry out a new adjustment.
E 036 Software-Update incorrect Return the instrument for repair.
E 040 Hardware failure / Check all connection lines.
electronics defect Contact our service department.
32 VEGAFLEX 51V and 52V
Note that the difference between min. and max.-adjustment is at least 10 mm.
Notes
VEGAFLEX 51V and 52V 33
Notes
34 VEGAFLEX 51V and 52V
Notes
VEGAFLEX 51V and 52V 35
VEGA Grieshaber KG Am Hohenstein 113 D-77761 Schiltach Phone (0 78 36) 50 - 0 Fax (0 78 36) 50 - 201 e-mail info@vega-g.de
The statements on types, application, use and operating conditions of the sensors and processing systems correspond to the actual know­ledge at the date of printing.
Technical data subject to alteration.
2.23 227 / March ’99
Loading...