www.vaisala.com
∆
°C
°C
0.5
0.4
0.3
0.2
0.1
-0.1
-0.2
-0.3
-0.4
-0.5
0
0-40 -20 20 40 60
PTU300 Combined Pressure, Humidity and
Temperature Transmitter
Technical data
Performance
BAROMETRIC PRESSURE
Pressure range 500 ... 1100 hPa, 50 ... 1100 hPa
Accuracy 500 ... 1100 hPa 500 ... 1100 hPa 50 ... 1100 hPa
CLASS A CLASS B
Linearity ±0.05 hPa ±0.10 hPa ±0.20 hPa
Hysteresis* ±0.03 hPa ±0.03 hPa ±0.08 hPa
Repeatability* ±0.03 hPa ±0.03 hPa ±0.08 hPa
Calibration
uncertainty**
Accuracy at +20 °C*** ±0.10 hPa ±0.20 hPa ±0.30 hPa
Temperature
dependence****
Total accuracy
(-40 ... +60 °C/
-40 ... +140 °F)
Long-term stability/year ±0.1 hPa ±0.1 hPa ±0.2 hPa
Response time
(100 % response)
one sensor 2 s
Pressure units
* Defined as ±2 standard deviation limits of endpoint non-linearity,
hysteresis error or repeatability error and calibration.
** Defined as ±2 standard deviation limits of accuracy of the working
standard including traceability to NIST.
*** Defined as the root sum of the squares (RSS) of endpoint non-linearity,
hysteresis error, repeatability error and calibration uncertainty at room
temperature.
**** Defined as ±2 standard deviation limits of temperature dependence over
the operating temperature range.
±0.07 hPa ±0.15 hPa ±0.20 hPa
±0.1 hPa ±0.1 hPa ±0.3 hPa
±0.15 hPa ±0.25 hPa ±0.45 hPa
•
1 s
•
1 s
hPa, mbar, kPa, Pa, inHg, mmH20, mmHg, torr, psia
The Vaisala PTU300 Combined Pressure, Humidity and
Temperature Transmitter is a versatile, multi-purpose
instrument.
Response time (90 %) at +20 °C (+68 °F) in still air
with grid filter 8 s / 17 s*
with grid + steel netting filter 20 s / 50 s*
with sintered filter 40 s / 60 s*
* with HUMICAP
•
TEMPERATURE
Measurement range, all probes -40 ... +60 °C (-40 ... +140 °F)
Accuracy at +20 °C (+68 °F) ± 0.2 °C (± 0.4 °F)
Temperature units °C, °F
ACCURACY OVER TEMPERATURE RANGE
®
180R or 180RC sensor
RELATIVE HUMIDITY
Measurement range 0 ... 100 % RH
Accuracy (including non-linearity,
hysteresis, and repeatability at
+15 ... +25 °C ±1 %RH (0 ... 90 % RH)
±1.7 %RH (90 ... 100 %RH)
at -20 ... +40 °C ±(1.0 + 0.008 x reading) %RH
at -40 ... +60 °C ±(1.5 + 0.015 x reading) %RH
Factory calibration uncertainty (+20 °C )
(Defined as ±2 standard deviation
limits. Small variations possible,
± 0.6 % RH (0 ... 40 %RH)
± 1.0 % RH (40 ... 97 %RH)
see also calibration certificate.)
Sensor
®
for typical applications Vaisala HUMICAP
for applications with chemical
purge/warmed probe Vaisala HUMICAP
180 or 180R*
®
180C or 180RC*
Temperature sensor PT100 RTD 1/3 Class B IEC 751
Operating Environment
Operating temperature -40 ... +60 °C (-40 ... +140 °F)
with display 0 ... +60 °C (+32 ... +140 °F)
Humidity range non-condensing
Electromagnetic compatibility EN61326-1:1997 + Am1:1998
+Am2:2001; Industrial Environment
Technical data
Inputs and outputs
Operating voltage 10 ... 35 VDC, 24 VAC
with optional power supply module 100 ... 240 VAC, 50/60 Hz
Power consumption at +20 °C (U
RS-232 max. 28 mA
3 x 0 ... 1 V/0 ... 5 V/0 ... 10 V max. 33 mA
U
out
3 x 0 ... 20 mA max. 63 mA
I
out
display and backlight +20 mA
during chemical purge max. +110 mA
during probe heating +120 mA
Settling time at power-up (one sensor)
class A 4 s
class B 3 s
External loads
current outputs R
0 ... 1 V output R
0 ... 5 V and 0 ... 10 V outputs R
Recommended wire size 0.5 mm
Digital outputs RS-232, RS-485 (optional)
Service connection RS-232, USB
Relay outputs (optional) 0.5 A, 250 VAC
Ethernet interface (optional)
Supported standards 10/100Base-T
Connector RJ45
Protocols Telnet
Software support Vaisala MI70 link
WLAN interface (optional)
Supported standards 802.11b
Antenna connector type RP-SMA
Protocols Telnet
Security WEP 64/128,WPA
Software support Vaisala MI70 link
Authentication / Encryption (WLAN)
Open / no encryption
Open / WEP
WPA Pre shared key / TKIP
WPA Pre shared key / CCMP (a.k.a. WPA2)
Optional data logger with real-time clock
Logged parameters max. three with trend/min/max values
Logging interval 10 sec (fixed)
Max. logging period 4 years 5 months
Logged points 13.7 million points per parameter
Battery lifetime min. 5 years
Display LCD with backlight, graphic trend display
24 VDC)
in
< 500 ohm
L
> 2 kohm
L
> 10 kohm
2
(AWG 20) stranded wires
L
of any parameter
Menu languages English, Finnish, French, German, Japanese,
Chinese, Spanish, Swedish, Russian
Analog outputs (optional)
current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0 ... 1 V, 0 ... 5 V, 0 ... 10 V
Humidity and temperature
accuracy at +20 °C ±0.05% full scale
temperature dependence ±0.005%/°C full scale
Pressure 500 ... 1100 hPa 50 ... 1100 hPa
accuracy at +20 °C ±0.30 hPa ±0.40 hPa
accuracy at -40 ... +60 °C ±0.60 hPa ±0.75 hPa
Mechanics
Cable bushing M20 x 1.5 for cable diameter
8 ... 11 mm/0.31 ... 0.43"
Conduit fitting 1/2" NPT
User cable connector (optional) M12 series 8-pin (male)
option 1 female plug with 5 m (16.4 ft) black cable
option 2 female plug with screw terminals
Probe cable diameter
PTU303 6.0 mm
other probes 5.5 mm
Housing material G-AlSi 10 Mg (DIN 1725)
Housing classification IP 65 (NEMA 4)
Weight
depending on selected probe 1.5 ... 2.0 Kg
Accessories
PC software and cable 215005
USB-RJ45 Serial Connection Cable 219685
Connection cable for HM70 211339
Wall mounting plate (plastic) 214829
Pole installation kit 215108
Rain shield 215109
DIN rail installation set 211477
Duct installation kit, PTU303/307 210697
Cable gland and AGRO, PTU303/307 HMP247CG
Solar radiation shield, PTU303/307/30T DTR502B
Meteorological installation kit HMT330MIK
Duct installation kit (T probe) 215003
TYPE A PPROVED PRODUCT
CERTI FICATE NO.: A-11440
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Ref. B210892EN-A ©Vaisala 2009
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.