Vaisala PTU200 Technical data

P.O. Box 26, FIN-00421 Helsinki, FINLAND
Email: industrialsales@vaisala.com
PTU200 Combined Pressure, Humidity and Temperature Transmitter
Features/Benefits
Barometric pressure, humidity and
temperature measurement in one transmitter
Available with two barometric
pressure sensors for added reliability
RS232C serial interface with NMEA
protocol for GPS use
Optional local display and RS485
serial interface
NIST traceable (certificate included)
Optional IP65 (NEMA 4) outdoor
installation kit and tripod available
Tel: +358 9 894 91
Fax: +358 9 8949 2485
www.vaisala.com
Three-in-one measurement
The Vaisala Combined Pressure, Humidity and Temperature Transmitter PTU200 is a versatile multipurpose instrument for various applications. The transmitter can be used e.g. in GPS* meteorology and other meteorological applications, as well as in environmental monitoring in calibration laboratories.
The user can select between two pressure accuracy classes. The PTU200 can also be equipped with a second pressure sensor for added reliability. Three different probes can be used with the PTU200 transmitter. Two of the probes are relative humidity and temperature probes and one is a temperature probe.
Serial communication
An RS232C serial interface is standard with PTU200 transmitters. The transmitter software is also compatible
PTU200 combines barometric pressure, humidity and temperature measurement in one transmitter.
with major GPS receivers and NMEA (National Marine Electronics Association) coded messages. An optional RS485 interface is available.
Vaisala's proven sensor technology
The PTU200 transmitter uses the Vaisala BAROCAP® Sensor for pressure measurement. The humidity and temperature probes are fitted with the Vaisala HUMICAP® Sensor. The sensors are known for their high accuracy and excellent long-term stability. The temperature sensor is a platinum RTD sensor.
Outdoor installation kit
For outdoor use, an optimized Vaisala Meteorological Installation Kit PTU200MIK1 is available. The PTU200MIK1 includes a IP65 (NEMA
4) rated enclosure for the barometer, a static pressure head and a radiation shield for the humidity and
Applications
GPS meteorology: estimating
precipitable water vapor in the atmosphere
GPS accuracy improvement
Meteorological applications
Environmental monitoring in
calibration laboratories
temperature probe. In addition, an optional Vaisala Tripod PTU200TRIPOD is available to enable quick field setup.
* GPS = Global Positioning System
Vaisala Meteorological Installation Kit PTU200MIK1 equipped with Vaisala Tripod PTU200TRIPOD.
106 Vaisala Instruments Catalog 2005
Ref. B210435en rev. A
PTU200
BAROMETRIC PRESSURE
Technical Data
Barometric pressure
Pressure range 500...1100 hPa, 50...1100 hPa
Accuracy 500…1100 hPa 50....1300 hPa
Linearity ±0.05 hPa ±0.10 hPa ±0.20 hPa Hysteresis* ±0.03 hPa ±0.03 hPa ±0.08 hPa Repeatability* ±0.03 hPa ±0.03 hPa ±0.08 hPa Calibration uncertainty** ± 0.07 hPa ±0.15 hPa ±0.20 hPa Accuracy at +20 °C*** ± 0.10 hPa ±0.20 hPa ±0.30 hPa
Temperature dependence**** ±0.1 hPa ±0.1 hPa ±0.3 hPa Total accuracy (-40...+60 °C / -40...+140 °F)
±0.15 hPa ±0.25 hPa ±0.45 hPa Long-term stability/year ±0.1 hPa ±0.1 hPa ±0.2 hPa Response time (100% response)
one sensor 2 s * Defined as ±2 standard deviation limits of endpoint nonlinearity, hysteresis error, or repeatability error. ** Defined as ±2 standard deviation limits of inaccuracy of the
working standard including traceability to NIST. ***Defined as the root sum of the squares (RSS) of endpoint
nonlinearity, hysteresis error, repeatability error and calibration uncertainty at room temperature. ****Defined at ±2 standard deviation limits of temperature
dependence over the operating temperature range.
Relative humidity
Measurement range 0.8...100 %RH Output scale 0...100 %RH equals 0...1 VDC Accuracy at +20 °C (including nonlinearity and hysteresis)
against factory references ±1 %RH
field calibration against references ±2 %RH (0...90% RH)
Typical long-term stability better than 1 %RH per year Temperature dependence ±0.05 %RH/°C Response time (90%) at +20 °C 15 sec. with membrane filter Sensor HUMICAP® 180
Temperature
HMP45A-P & HMP45D ProbesHMP45A-P & HMP45D Probes
HMP45A-P & HMP45D Probes
HMP45A-P & HMP45D ProbesHMP45A-P & HMP45D Probes Measurement range -36...+60 °C (-33…+140 °F) (HMP45A-P)
Accuracy
at -40 °C (-40 °F) ±0.5 °C (±0.9 °F)
at +20 °C (68 °F) ±0.2 °C (±0.4 °F)
at +60 °C (140 °F) ±0.4 °C (±0.7 °F) Temperature sensor
HMP 45 A-P Pt 1000 IEC 751 1/3 Class B
HMP 45D Pt 100 IEC 751 1/3 Class B PT100 Sensor HeadPT100 Sensor Head
PT100 Sensor Head
PT100 Sensor HeadPT100 Sensor Head Measurement range -40...+60 °C (-40…+140 °F) Accuracy ±0.2°C (±0.4 °F) Temperature sensor Pt 100 IEC 751 1/4 Class B
Class A Class B
1 s 1 s
±3 %RH (90...100% RH)
-40...+60 °C (-40…+140 °F) (HMP45D)
with local display less than 30 mA (without backlight)
less than 50 mA (with backlight)
hardware shutdown mode less than 0.1 mA
Serial I/O full duplex RS 232C
or
bidirectional TTL level or
half duplex RS 485/422 two-wire code ASCII parity even•, odd, none data bits 7 stop bits 1
or 8
or 2
Baud rates 300, 600, 1200, 2400, 4800, 9600 Pressure units hPa•, mbar, kPa, Pa, inHg,
mmH20, mmHg, torr, psia Humidity units %RH Temperature units °F, °C Resolution 0.1 hPa•, 0.1 %RH, 0.01°C Settling time at power-up (one sensor)
class A 6 s class B 5 s
Acceleration sensitivity negligible Weight
depending on selected probe 2.3/1.3/1.1 kg
Transmitter bodyTransmitter body
Transmitter body
Transmitter bodyTransmitter body Pressure connector M5 (10-32) internal thread Pressure fitting barbed fitting for 1/8" I.D. tubing
quick connector with shutoff valve for 1/8" hose Maximum pressure limit 5000 hPa abs. Electrical connector female 9-pin subD Housing epoxy painted aluminum HMP45A-P and HMP45DHMP45A-P and HMP45D
HMP45A-P and HMP45D
HMP45A-P and HMP45DHMP45A-P and HMP45D Housing ABS plastic Probe cable length 20 m (HMP45A-P)
3.5 m (HMP45D) Housing classification IP65 (NEMA 4) Sensor protection (standard) membrane filter part no. 2787HM Pt100 ProbePt100 Probe
Pt100 Probe
Pt100 ProbePt100 Probe Probe cable length 2 m Complies with EMC standard EN61326-1:1997+ Am1:1998 + Am2:2001; Generic Environment.
Accessories
PTU200MIK1 outdoor mounting kit PTU200TRIPOD tripod stand 213119 DC adapter for PTB220/PTU200 series 213026 DC adapter for PTB220/PTU200 series & RS232 cable for PC AC adapter, universal 100-240 VAC with US, EURO, UK or AUS plugs, to be used together with 213119 or 213026 Factory settings are marked with a
.
General
Temperature range
operating -40...+60 °C (-40…+140 °F) with local display 0...+60 °C (+32…+140 °F) storage -40...+60 °C (-40…+140 °F)
with local display -20...+60 °C (-4…+140 °F) Humidity range non-condensing Supply voltage 10...30 VDC polarity protected Supply voltage sensitivity negligible Current consumption less than 30 mA
Ref. B210435en rev. A
240 (9.45)
24
A
A
(0.94)
118 (4.65)
Removable probe head
¯18.5
0.72)
BAROCAP® and HUMICAP® are registered trademarks of Vaisala. Specifications subject to change without prior notice. ©Vaisala Oyj
A-A
20.5
(0.81)
(0.94)
24
107Vaisala Instruments Catalog 2005