Vaisala Barometric Pressure Transfer
Standard PTB330TS
Technical data
These specifi cations apply when MI70, PTB330 and HMP155 are
used together in the PTB330TS product. For individual specifi cations,
please refer to the product documentation and brochures of the
PTB330 and HMP155.
General
Operating temperature range -10 ... +40 °C (+14 ... +104 )°F
Operating humidity range non-condensing
Maximum pressure limit 5000 hPa abs.
Power supply Rechargeable NiMH battery pack
with AC-adapter or 4xAA-size
alkalines, type IEC LR6
Operation time (using rechargeable battery pack)
Continuous use with PTB330 11 h typical at +20 °C (+68 °F)
Datalogging use up to 30 days
Menu languages English, Chinese, French, Spanish,
German, Russian, Japanese,
Swedish, Finnish
Display LCD with backlight, graphic
trend display of any parameter,
character height up to 16 mm
Data logging capacity 2700 points
Alarm audible alarm function
PTB330TS is in conformity with the following EU directives:
- EMC Directive (2004/108/EC) Complies with the EMC product family standard EN61326-1,
Electrical equipment for measurement control and laboratory use - Basic immunity test
requirements.
- Low Voltage Directive (2006/95/EC)
- ROHS Directive (2002/95/EC)
Performance
Barometric pressure (PTB330)
Measurement range 500 ... 1100 hPa
Linearity* ±0.05 hPa
Hysteresis* ±0.03 hPa
Repeatability* ±0.03 hPa
Calibration uncertainty** ±0.07 hPa
Accuracy at +20 °C (+68 °F) *** ±0.10 hPa
Temperature dependence**** ±0.1 hPa
Total accuracy -40... +60 °C (-40...+140 °F) ±0.15 hPa
Long-term stability ±0.1 hPa/year
Settling time at power-up (one sensor) 4 s
Response time (one sensor) 2 s
Acceleration sensitivity negligible
* Defined as ±2 standard deviation limits of endpoint nonlinearity, hysteresis or repeatability error.
** Defined as ±2 standard deviation limits of inaccuracy of the working standard including
traceability to NIST.
*** Defined as the root sum of the squares (RSS) of endpoint non-linearity, hysteresis error,
repeatability error and calibration uncertainty at room temperature.
**** Defined as ±2 standard deviation limits of temperature dependence over the operating
temperature range.
Relative humidity (HMP155)
Measurement range 0 ... 100 %RH
Accuracy (incl. non-linearity, hysteresis and repeatability) at
+15 ... +25 °C (+59 ... +77 °F) ±1 %RH (0 ... 90 %RH)
-10 ... +40 °C (-4 ... 104 °F) ±(1.0 + 0.008 x reading) %RH
Factory calibration uncertainty (+20 °C /+68 °F)
Humidity sensor HUMICAP180R
Response time at +20 °C in still air with a sintered PTFE fi lter
63 % 20 s
90 % 60 s
* Defined as ±2 standard deviation limits. Small variations possible, see also calibration
certificate.
Temperature (HMP155)
Measurement range -10 ... +40 °C (+14 ... +104 °F)
Accuracy
-10 ... +20 °C ±(0.176 - 0.0028 x
+20 ... +40 °C ±(0.07 + 0.0025 x
Accuracy over temperature range (see graph overleaf)
Temperature sensor Pt100 RTD 1/3 Class B IEC 751
Response time with additional temperature probe in 3 m/s air fl ow
63 % <20 s
90 % <35 s
www.vaisala.com
±1.7 %RH (90 ... 100 %RH)
±0.6 %RH (0 ... 40 %RH)*
±1.0 %RH (40 ... 97 %RH)*
HUMICAP180RC
temperature) °C
temperature) °C
Technical data
∆ °C
0,3
0,2
0,1
0
-0,1
-0,2
-0,3
-20 0 20 40
°C
Accuracy of HMP155 temperature measurement over
temperature range
Available Parameters
Pressure parameters P, P3h, HCP, QFE, QNH
Humidity and temperature RH, T, Tdf, Td, x, Tw
parameters
Inputs and outputs
MI70 probe ports 2
MI70 data interface RS-232 (accessible only with MI70
Link software)
PTB330 supply voltage 10 ... 35 VDC (if not powered by
MI70)
PTB330 data interface RS-232C
PTB330 serial I/O connectors RJ45 (service port)
Male 8-pin M12 (user port)
HMP155 data interface RS-485
HMP155 serial I/O connector Male 8-pin M12
Mechanics
PTB330
Housing material G-AlSi 10 Mg (DIN 1725)
Housing classifi cation IP65
Pressure connector M5 (10-32) internal thread
Pressure fi tting barbed fi tting for 1/8” I.D.
tubing or quick connector with
shutoff valve for 1/8” hose
HMP155
Housing material PC
Housing classifi cation IP66
Additional T-probe cable length 2 m
Cable material PUR
Sensor protection Sintered PTFE
MI70 MEASUREMENT INDICATOR
Housing classifi cation IP54
Housing material ABS/PC blend
TRANSPORT CASE
Housing classifi cation
(when closed) IP67
Plastic parts TTX01®, PP+SEBS, POM
Metal parts stainless steel AISI303
Interior foam material PE and polyether
Weight with all instruments and
typical accessories 5.9 kg
Exterior dimensions (LxWxH) 405×330×165 mm
(15.94×12.99×6.50) inch
Accessories
PTB330
MI70 – PTB330 Spiral Cable 223235SP
USB-RJ45 serial connection cable 219685
Serial connection cable 19446ZZ
Barbed fi tting 1/8” 19498SP
Quick Connector 1/8” 220186
Transport case with interior foams
and tabletop casing for PTB330 224068SP
Tabletop casing for PTB330 224064SP
MI70
USB cable for MI70, 219687
includes MI70 Link software
MI70 Link software MI70LINK
MI70 connection cable to HMT330,
MMT330, DMT340, HMT100, PTB330 211339
MI70 battery pack 26755
variety of AC adapters available
HMP155
HMP155 – MI70 connection cable 221801
Protection set for HMP155 calibration
buttons: protective cover,
2 O-rings and protective plug 221318
USB cable for HMP155 221040
Sintered tefl on fi lter + O-ring 219452SP
Humidity sensor HUMICAP180R
Humidity Calibrator HMK15
Vaisala Oyj
Helsinki, Finland
Tel. (+358 9) 894 91
Fax: (+358 9) 8949 2876
Email: sales@vaisala.com
For other Vaisala
locations visit us at:
www.vaisala.com
Ref. B210786EN-A ©Vaisala 2009
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifi cations — technical included — are subject
to change without notice.