Vaisala HUMICAP HMT330 Series User Manual

Page 1
USER'S GUIDE
Vaisala HUMICAP®Humidity and
Temperature Transmitter Series
M210566EN-D
Page 2
PUBLISHED BY
Vaisala Oyj Phone (int.): +358 9 8949 1 P.O. Box 26 Fax: +358 9 8949 2227 FIN-00421 Helsinki Finland
Visit our Internet pages at http://www.vaisala.com/
© Vaisala 2006
No part of this manual may be reproduced in any form or by any means, electronic or mechanical (including photocopying), nor may its contents be communicated to a third party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for Vaisala towards the customer or end user. All legally binding commitments and agreements are included exclusively in the applicable supply contract or Conditions of Sale.
Page 3
_________________________________________________________________________________

Table of Contents

CHAPTER 1
GENERAL INFORMATION ............................................................................ 9
About This Manual ................................................................... 9
Contents of This Manual ....................................................... 9
General Safety Considerations ........................................... 10
Feedback............................................................................. 10
Product Related Safety Precautions .................................... 11
ESD Protection .......................................................................11
Recycling ................................................................................ 12
Trademarks ............................................................................. 12
License Agreement ................................................................ 12
Warranty .................................................................................. 13
CHAPTER 2
PRODUCT OVERVIEW ................................................................................ 15
Introduction to HMT330 ......................................................... 15
Basic Features and Options................................................ 16
Probe Options ..................................................................... 19
Warmed Probe HMT337 ..................................................... 20
CHAPTER 3
INSTALLATION............................................................................................ 21
Mounting the Housing ........................................................... 21
Standard Mounting without Mounting Plate ........................21
Wall Mounting with Wall Mounting Kit ................................. 22
Mounting with DIN Rail Installation Kit ................................ 23
Pole Installation with Installation Kit for Pole or Pipeline .... 24
Mounting Rain Shield with Installation Kit ........................... 26
Panel Mounting Frame ........................................................ 27
Wiring ......................................................................................28
Cable Bushings ................................................................... 28
Grounding the Cables ......................................................... 29
Grounding the Transmitter Housing .................................... 30
Signal and Power Supply Wiring......................................... 31
Connections to 24 VAC Power Supply ............................... 32
Probe Mounting ...................................................................... 34
General Instructions for Probes with Cable ........................ 35
HMT333 for Ducts and Tight Spaces .................................. 37
HMT334 for High Pressure and Vacuum Applications........ 37
HMT335 for High Temperatures ......................................... 40
HMT337 for High Humidity Applications ............................. 40
Temperature Probe (Optional) ............................................ 40
HMT338 for Pressurized Pipelines...................................... 41
Tightening the Clasp Nut..................................................... 42
VAISALA _____________________________________________________________________ 1
Page 4
User's Guide _______________________________________________________________________
Optional Modules....................................................................
44
Power Supply Module..........................................................44
Installation ......................................................................45
Warnings ........................................................................46
Galvanic Isolation for Output ...............................................48
Third Analog Output ............................................................48
Installation and Wiring .................................................... 49
Relays..................................................................................50
Installation and Wiring .................................................... 50
Selecting the Activation State of the Relay .................... 50
RS-422/485 Interface ..........................................................52
Installation and Wiring .................................................... 52
8-Pin Connector...................................................................55
CHAPTER 4
OPERATION .................................................................................................57
Getting Started........................................................................57
Display/Keypad (Optional) ..................................................... 57
Basic Display .......................................................................57
Graphic History .................................................................... 58
Menus and Navigation......................................................... 59
Language Setting ...........................................................60
Rounding Setting ............................................................61
Display Backlight Setting ................................................ 61
Display Contrast Setting ................................................. 61
Keypad Lock (Key guard) ............................................... 62
Menu PIN Lock ...............................................................62
Factory Settings .............................................................63
MI70 Link Program for Data Handling ..................................63
Serial Line Communication ...................................................63
User Port Connection ..........................................................64
Service Port Connection......................................................65
Terminal Program Settings.................................................. 66
List of Serial Commands .....................................................68
Getting Measurement Message from Serial Line................ 70
Stopping Continuous Outputting ....................................70
S ................................................................................70
Outputting Reading Once ............................................... 70
SEND ........................................................................70
SEND D.....................................................................71
Formatting Serial Line Message..........................................72
TIME and DATE .............................................................72
FTIME and FDATE .........................................................72
FST.................................................................................73
General Settings .....................................................................74
Changing Quantities and Units............................................ 74
Using Display/Keypad ....................................................74
Using Serial Line ............................................................75
FORM........................................................................75
UNIT ..........................................................................76
2 _________________________________________________________________ M210566EN-D
Page 5
_________________________________________________________________________________
Pressure Compensation Setting .........................................
Using Display/Keypad .................................................... 77
Using Serial Line ............................................................ 77
PRES and XPRES .................................................... 77
User Port Serial Settings..................................................... 78
Using Display/Keypad .................................................... 78
Using Serial Line ............................................................ 79
SERI ............................................................................... 79
SMODE .......................................................................... 80
INTV ............................................................................... 80
ECHO............................................................................. 81
Data Filtering ....................................................................... 81
FILT................................................................................ 82
Device Information .............................................................. 82
? .....................................................................................83
HELP.............................................................................. 83
ERRS .............................................................................84
VERS ............................................................................. 84
Resetting Transmitter by Using Serial Line......................... 85
RESET ...........................................................................85
Locking Menu/Keypad by Using Serial Line ....................... 85
LOCK ............................................................................. 85
Data Recording....................................................................... 87
Selecting Data Recording Quantities .................................. 87
DSEL.............................................................................. 87
View Recorded Data ........................................................... 88
DIR ................................................................................. 88
PLAY .............................................................................. 89
Deleting the Recorded Files................................................ 90
DELETE/UNDELETE ..................................................... 90
Analog Output Settings ......................................................... 91
Changing Output Mode and Range .................................... 91
Analog Output Quantities .................................................... 93
AMODE/ASEL................................................................ 93
Analog Output Tests ........................................................... 95
ITEST ............................................................................. 95
Analog Output Fault Indication Setting ............................... 96
AERR .............................................................................96
Operation of Relays ............................................................... 97
Quantity for Relay Output.................................................... 97
Measurement-Based Relay Output Modes ......................... 97
Relay Setpoints .............................................................. 97
Hysteresis ...................................................................... 98
Relay Indicating Transmitter Error Status ........................... 98
Enabling/Disabling Relays ................................................ 100
Setting Relay Outputs ....................................................... 100
RSEL............................................................................ 101
Testing Operation of Relays.............................................. 103
RTEST ......................................................................... 103
77
VAISALA _____________________________________________________________________ 3
Page 6
User's Guide _______________________________________________________________________
Operation of RS-485 Module ...............................................
104
Networking Commands .....................................................104
SERI .............................................................................104
ECHO ...........................................................................105
SMODE ........................................................................105
INTV .............................................................................106
ADDR ...........................................................................106
SEND............................................................................107
OPEN ...........................................................................107
CLOSE .........................................................................107
Sensor Functions .................................................................108
Chemical Purge (Optional) ................................................108
Automatic Chemical Purge (Interval Purge) ................. 109
Manual Chemical Purge ...............................................109
Chemical Purge in Power Up ....................................... 109
Starting and Configuring Chemical Purge .........................110
Using Buttons on Motherboard.....................................110
Using Display/Keypad (Optional) .................................110
Using Serial Line ..........................................................111
PURGE....................................................................111
PUR.........................................................................112
Sensor Heating .................................................................. 113
Setting Humidity Sensor Heating .................................113
XHEAT ....................................................................113
CHAPTER 5
MAINTENANCE..........................................................................................115
Periodic Maintenance...........................................................115
Cleaning ............................................................................115
Changing the Probe Filter .................................................115
Changing the Sensor ...........................................................116
Error States .......................................................................117
CHAPTER 6
CALIBRATION AND ADJUSTMENT ......................................................... 119
Opening and Closing the Adjustment Mode......................119
Relative Humidity Adjustment............................................. 121
Using Push-Buttons........................................................... 121
Using Display/Keypad .......................................................122
Using Serial Line ...............................................................123
CRH..............................................................................123
Relative Humidity Adjustment after Sensor Change ........124
Using Display/Keypad .......................................................124
Using Serial Line ...............................................................124
FCRH............................................................................124
Temperature Adjustment .....................................................125
Using Display/Keypad .......................................................125
Using Serial Line ...............................................................125
CT.................................................................................126
CTA ..............................................................................126
4 _________________________________________________________________ M210566EN-D
Page 7
_________________________________________________________________________________
Analog Output Adjustment..................................................
127
Using Display/Keypad ....................................................... 127
Using Serial Line ............................................................... 127
ACAL............................................................................ 127
Feeding Adjustment Information........................................ 128
Using Display/Keypad ....................................................... 128
Using Serial Line ............................................................... 128
CTEXT ......................................................................... 128
CDATE ......................................................................... 128
CHAPTER 7
TECHNICAL DATA ....................................................................................129
Specifications ....................................................................... 129
Performance...................................................................... 129
Relative Humidity .........................................................129
Temperature (+ Operating Pressure Ranges) ............. 130
Optional Temperature Probe ....................................... 130
Calculated Variables .................................................... 131
Accuracies of Calculated Variables .................................. 131
Accuracy of Dewpoint Temperature °C ....................... 131
Accuracy of Mixing Ratio g/kg (Ambient Pressure 1013
mbar)............................................................................
132
Accuracy of Wet Bulb Temperature °C ........................ 132
Accuracy of Absolute Humidity g/m³ ............................ 132
Dewpoint Temperature (HMT337 Warmed Probe
Option)...............................................................................
133
Operating Environment ................................................ 133
Inputs and Outputs....................................................... 134
Mechanics .................................................................... 134
Transmitter Weights ..................................................... 135
Technical Specifications of Optional Modules .................. 135
Power Supply Module .................................................. 135
Analog Output Module ................................................. 135
Relay Module ............................................................... 136
RS-485 Module ............................................................136
Options and Accessories ............................................. 136
Dimensions (in mm) ............................................................. 139
HMT331............................................................................. 140
HMT333............................................................................. 140
HMT334............................................................................. 141
HMT335............................................................................. 141
HMT337............................................................................. 142
HMT338............................................................................. 142
Temperature Probe ........................................................... 142
Technical Support ................................................................ 143
Return Instructions ..............................................................143
Vaisala Service Centers....................................................... 144
VAISALA _____________________________________________________________________ 5
Page 8
User's Guide _______________________________________________________________________
APPENDIX A
PROBE INSTALLATION KITS AND INSTALLATION EXAMPLES .........145
Duct Installation Kits (for HMT333/337/335)....................... 145
Duct Installation Kit for Temperature Probe (for HMT337)146 Pressure Tight Swagelok Installation Kits (for HMT337)..147
RH Probe Installation.........................................................147
Temperature Probe Installation .........................................147
Examples of Vapor Tight Installations with Cable Gland.148
RH-Probe Installations (for HMT333/337) .........................148
T- Probe Installations (HMT337) .......................................149
Example of Climate Chamber Installation .........................150
Example of Installation Through Roof ...............................151
Ball Valve Installation Kit for HMT338 ................................152
Meteorological Installation Kit (for HMT337) .....................154
APPENDIX B
CALCULATION FORMULAS.....................................................................155
List of Figures
Figure 1 Transmitter Body ......................................................................17
Figure 2 Inside of Open Transmitter .......................................................18
Figure 3 Probe Options...........................................................................19
Figure 4 Standard Mounting ...................................................................21
Figure 5 Mounting with Wall Mounting Kit ..............................................22
Figure 6 Dimensions of Plastic Mounting Plate ......................................22
Figure 7 Mounting with DIN Rail Installation Kit ..................................... 23
Figure 8 Vertical Pole.............................................................................. 24
Figure 9 Horizontal Pole .........................................................................24
Figure 10 Mounting with Metal Wall Mounting Plate ................................25
Figure 11 Dimensions of Metal Mounting Plate (mm) ..............................25
Figure 12 Mounting Rain Shield with Installation Kit ................................26
Figure 13 Panel Mounting with Frame ...................................................... 27
Figure 14 Panel Mounting Dimensions.....................................................27
Figure 15 Cable Bushings......................................................................... 28
Figure 16 Grounding the Screen of Electrical Cable ................................29
Figure 17 Screw Terminal Block on Motherboard ....................................31
Figure 18 Connections to 24 VAC Power Supply .....................................33
Figure 19 Measurement Error at 100 %RH ..............................................34
Figure 20 Horizontal Mounting of Sensor Head........................................ 35
Figure 21 Vertical Mounting of Sensor Head............................................36
Figure 22 HMT344 Probe .........................................................................38
Figure 23 Tightening the Nut ....................................................................38
Figure 24 Cleaning of Tightening Cone ....................................................39
Figure 25 HMT338 Probe .........................................................................41
Figure 26 Sealing of Fitting Body into Process......................................... 42
Figure 27 Tightening the Clasp Nut ..........................................................42
Figure 28 Power Supply Module...............................................................44
Figure 29 Galvanic Output Isolation Module............................................. 48
Figure 30 Third Analog Output ................................................................. 48
Figure 31 Relay Module............................................................................51
Figure 32 RS-485 Module.........................................................................52
6 _________________________________________________________________ M210566EN-D
Page 9
_________________________________________________________________________________
Figure 33
Figure 34 Wiring of Optional 8-Pin Connector.......................................... 55
Figure 35 Basic Display ............................................................................ 58
Figure 36 Graphical Display ..................................................................... 58
Figure 37 Main Menus .............................................................................. 60
Figure 38 Service Port Connector and User Port Terminal on Mother
Figure 39 Connection Example Between PC Serial Port and User Port.. 64
Figure 40 Starting Hyper Terminal Connection ........................................ 66
Figure 41 Connecting to Hyper Terminal ................................................. 67
Figure 42 Hyper Terminal Serial Port Settings ......................................... 67
Figure 43 Device Information on Display ................................................ 82
Figure 44 Current/Voltage Switches of Output Modules ..........................91
Figure 45 Measurement-Based Relay Output Modes .............................. 97
Figure 46 FAULT/ONLINE STATUS Relay Output Modes ...................... 99
Figure 47 Relay Indicators on Display.................................................... 100
Figure 48 Decrease of Sensor Gain ....................................................... 108
Figure 49 Purge Buttons on Motherboard .............................................. 110
Figure 50 Chemical Purge Settings........................................................ 110
Figure 51 Performing Chemical Purge ................................................... 111
Figure 52 Changing the Sensor.............................................................. 116
Figure 53 Error Indicator and Error Message......................................... 117
Figure 54 Adjustment and Purge Buttons .............................................. 120
Figure 55 Adjustment Menu ................................................................... 120
Figure 56 Selecting Point 1 Reference Type.......................................... 122
Figure 57 Accuracy over Temperature Range ....................................... 130
Figure 58 Accuracy in Dewpoint Measurement...................................... 133
Figure 59 Transmitter Body Dimensions ................................................ 139
Figure 60 HMT331 Probe Dimensions ................................................... 140
Figure 61 HMT333 Probe Dimensions ................................................... 140
Figure 62 HMT334 Probe Dimensions ................................................... 141
Figure 63 HMT335 Probe Dimensions ................................................... 141
Figure 64 HMT337 Probe Dimensions ................................................... 142
Figure 65 HMT338 Probe Dimensions ................................................... 142
Figure 66 Optional Temperature Probe Dimensions.............................. 142
Figure 67 Duct Mounting Installation Kit................................................. 145
Figure 68 Duct Mounting Installation Kit for T-Probe ............................. 146
Figure 69 Swagelok Installation Kit for RH-probe .................................. 147
Figure 70 Swagelok Installation Kit for T-Probe ..................................... 147
Figure 71 Cable Installation with Cable Gland ....................................... 148
Figure 72 Probe Head Installation with Cable Gland ............................. 148
Figure 73 Vapor Tight Installation........................................................... 149
Figure 74 Wall Mounting Installation ...................................................... 149
Figure 75 Climate Chamber Installation (not Available from Vaisala) .... 150
Figure 76 Example of Installation through Roof ..................................... 151
Figure 77 Installing Sensor Head through Ball Valve Assembly ............ 152
Figure 78 Meteorological Installation Kit for Outdoor Installation........... 154
4-Wire RS-485 Bus .................................................................. 53
Board ........................................................................................
64
VAISALA _____________________________________________________________________ 7
Page 10
User's Guide _______________________________________________________________________
List of Tables
Table 1 Quantities Measured by HMT330.............................................15
Table 2 Optional Quantities Measured by HMT330 ..............................16
Table 3 HMT338 Probe Dimensions .....................................................41
Table 4 Connecting the Twisted Pair Wires to the Screw Terminals ....52
Table 5 4-Wire (Switch 3: On) ...............................................................53
Table 6 2-Wire (Switch 3: Off) ...............................................................54
Table 7 Wiring of 8-Pin Connector ........................................................ 55
Table 8 Periods for Trend and Max/Min Calculations ...........................59
Table 9 Graph Information Messages in Cursor Mode .........................59
Table 10 Default Serial Communication Settings for the User Port ........64
Table 11 Fixed Communication Settings for Service Port .......................65
Table 12 FORM Command Modifiers ......................................................75
Table 13 Multiplication Factors ................................................................ 78
Table 14 Selection of Output Modes .......................................................80
Table 15 Filtering Levels..........................................................................81
Table 16 Error Messages ......................................................................118
Table 17 Indicator Led Functions ..........................................................120
Table 18 Calculated Variables (Typical Ranges) ..................................131
Table 19 Transmitter Weights (in kg/lb).................................................135
8 _________________________________________________________________ M210566EN-D
Page 11
Chapter 1 _________________________________________________________ General Information
CHAPTER 1

GENERAL INFORMATION

This chapter provides general notes for the manual and the product.

About This Manual

This manual provides information for installing, operating, and maintaining Vaisala HUMICAP® Humidity and Temperature Transmitter Series HMT330.

Contents of This Manual

This manual consists of the following chapters:
- Chapter 1, General Information, provides general notes for the manual and the product.
Chapter 2, Product Overview, introduces the features, advantages,
­and the product nomenclature of HMT330.
-
Chapter 3, Installation, provides you with information that is intended to help you install the product.
- Chapter 4, Operation, contains information that is needed to operate this product.
- Chapter 5, Maintenance, contains information that is needed in basic maintenance of the product.
- Chapter 6, Calibration and Adjustment, provides information and instructions concerning calibration and adjustment of HMT330.
- Chapter 7, Technical Data, provides the technical data of the product.
VAISALA _____________________________________________________________________ 9
Page 12
User's Guide _______________________________________________________________________
- Appendix A, Probe Installation Kits and Installation Examples, presents the installation kits available for HMT330 and provides some installation examples.
- Appendix B, Calculation Formulas, presents the equations used in HMT330 to calculate values of dewpoint, mixing ratio, absolute humidity and enthalpy in normal pressure.

General Safety Considerations

Throughout the manual, important safety considerations are highlighted as follows:
WARNING
CAUTION
NOTE
Warning alerts you to a serious hazard. If you do not read and follow instructions very carefully at this point, there is a risk of injury or even death.
Caution warns you of a potential hazard. If you do not read and follow instructions carefully at this point, the product could be damaged or important data could be lost.
Note highlights important information on using the product.

Feedback

Vaisala Customer Documentation Team welcomes your comments and suggestions on the quality and usefulness of this publication. If you find errors or have other suggestions for improvement, please indicate the chapter, section, and page number. You can send comments to us by e-mail:
manuals@vaisala.com
10 ________________________________________________________________ M210566EN-D
Page 13
Chapter 1 _________________________________________________________ General Information

Product Related Safety Precautions

The Vaisala HUMICAP® Humidity and Temperature Transmitter Series HMT330 delivered to you has been tested for safety and approved as shipped from the factory. Note the following precautions:
WARNING
CAUTION
Ground the product, and verify outdoor installation grounding periodically to minimize shock hazard.
Do not modify the unit. Improper modification can damage the product or lead to malfunction.

ESD Protection

Electrostatic Discharge (ESD) can cause immediate or latent damage to electronic circuits. Vaisala products are adequately protected against ESD for their intended use. However, it is possible to damage the product by delivering electrostatic discharges when touching, removing, or inserting any objects inside the equipment housing.
To make sure you are not delivering high static voltages yourself:
- Handle ESD sensitive components on a properly grounded and protected ESD workbench. When this is not possible, ground yourself to the equipment chassis before touching the boards. Ground yourself with a wrist strap and a resistive connection cord. When neither of the above is possible, touch a conductive part of the equipment chassis with your other hand before touching the boards.
- Always hold the boards by the edges and avoid touching the component contacts.
VAISALA ____________________________________________________________________ 11
Page 14
User's Guide _______________________________________________________________________

Recycling

Recycle all applicable material.
Dispose of batteries and the unit according to statutory regulations. Do not dispose of with regular household refuse.

Trademarks

HUMICAP® is a registered trademark of Vaisala. Microsoft®, Windows®, Windows NT®, Windows® 2000 and Windows® XP are registered trademarks of Microsoft Corporation in the United States and/or other countries.

License Agreement

All rights to any software are held by Vaisala or third parties. The customer is allowed to use the software only to the extent that is provided by the applicable supply contract or Software License Agreement.
12 ________________________________________________________________ M210566EN-D
Page 15
Chapter 1 _________________________________________________________ General Information

Warranty

Vaisala hereby represents and warrants all Products manufactured by Vaisala and sold hereunder to be free from defects in workmanship or material during a period of twelve (12) months from the date of delivery save for products for which a special warranty is given. If any Product proves however to be defective in workmanship or material within the period herein provided Vaisala undertakes to the exclusion of any other remedy to repair or at its own option replace the defective Product or part thereof free of charge and otherwise on the same conditions as for the original Product or part without extension to original warranty time. Defective parts replaced in accordance with this clause shall be placed at the disposal of Vaisala.
Vaisala also warrants the quality of all repair and service works performed by its employees to products sold by it. In case the repair or service works should appear inadequate or faulty and should this cause malfunction or nonfunction of the product to which the service was performed Vaisala shall at its free option either repair or have repaired or replace the product in question. The working hours used by employees of Vaisala for such repair or replacement shall be free of charge to the client. This service warranty shall be valid for a period of six (6) months from the date the service measures were completed.
This warranty does not however apply when the defect has been caused through
a) normal wear and tear or accident;
b) misuse or other unsuitable or unauthorized use
of the Product or negligence or error in storing, maintaining or in handling the Product or any equipment thereof;
c) wrong installation or assembly or failure to
service the Product or otherwise follow Vaisala's service instructions including any repairs or installation or assembly or service made by unauthorized personnel not approved by Vaisala or replacements with parts not manufactured or supplied by Vaisala;
d) modifications or changes of the Product as well
as any adding to it without Vaisala's prior authorization;
e) other factors depending on the Customer or a
third party.
Notwithstanding the aforesaid Vaisala's liability under this clause shall not apply to any defects arising out of materials, designs or instructions provided by the Customer.
This warranty is however subject to following conditions:
a) A substantiated written claim as to any alleged
defects shall have been received by Vaisala within thirty (30) days after the defect or fault became known or occurred, and
b) The allegedly defective Product or part shall,
should Vaisala so require, be sent to the works of Vaisala or to such other place as Vaisala may indicate in writing, freight and insurance prepaid and properly packed and labelled, unless Vaisala agrees to inspect and repair the Product or replace it on site.
This warranty is expressly in lieu of and excludes all other conditions, warranties and liabilities, express or implied, whether under law, statute or otherwise, including without limitation any implied warranties of merchantability or fitness for a particular purpose and all other obligations and liabilities of Vaisala or its representatives with respect to any defect or deficiency applicable to or resulting directly or indirectly from the Products supplied hereunder, which obligations and liabilities are hereby expressly cancelled and waived. Vaisala's liability shall under no circumstances exceed the invoice price of any Product for which a warranty claim is made, nor shall Vaisala in any circumstances be liable for lost profits or other consequential loss whether direct or indirect or for special damages.
VAISALA ____________________________________________________________________ 13
Page 16
User's Guide _______________________________________________________________________
This page intentionally left blank.
14 ___________________________________________________________________ M210566EN-D
Page 17
Chapter 2 ___________________________________________________________ Product Overview
CHAPTER 2

PRODUCT OVERVIEW

This chapter introduces the features, advantages, and the product nomenclature of HMT330.

Introduction to HMT330

Vaisala HUMICAP® Humidity and Temperature Transmitter Series HMT330 provides reliable humidity measurement in wide range of applications. Analog outputs can be chosen between current and voltage signals. Alternatively, digital outputs RS-232 (standard) or RS-422/485 (optional) can be selected. The quantities measured and calculated by HMT330 are presented in Table 1 below. The quantities available as an option are presented in Table 2 below.
Table 1 Quantities Measured by HMT330
Quantity Abbreviati
Relative humidity (RH) RH %RH %RH Temperature (T) T ºC ºF
on
Metric Unit Non Metric
Unit
VAISALA________________________________________________________________________ 15
Page 18
User's Guide _______________________________________________________________________
Table 2 Optional Quantities Measured by HMT330
Quantity Abbreviati
on
Dewpoint / Frostpoint Temperature (T Dewpoint Temperature (Td) TD ºC ºF Absolute humidity (a) A g/m3 gr/ft3 Mixing ratio (x) X g/kg gr/lb Wetbulb temperature (Tw) TW ºC ºF Humid air volume/ dry air volume (by volume or by weight) (H Water vapor pressure (Pw) PW hPa lb/in2 Water vapor saturation pressure (P Enthalpy (h) H kJ/kg Btu/lb Difference of T and T
)
d/f
)
ws
(T) DT ºC ºF
d/f
TDF ºC ºF
H2O ppmv/ppm
O)
2
PWS hPa lb/in
Metric Unit Non Metric
Unit
ppmv/ppmw
w
2

Basic Features and Options

- Several probes for various applications
- User friendly display
- Calculated output quantities available
- Different probe mounting kits, sensor protection options and probe cable lengths
- Transmitter mounting kits for multiple installation purposes
- Chemical purge for applications where interfering chemicals in the measuring environment pose a risk
- Warmed probe and sensor heating for high humidity conditions (HMT337)
- Optional modules: isolated power supply, power supply module, RS-422/485-module, additional analog output module and relay module
- Additional temperature sensor (HMT337)
16 ___________________________________________________________________ M210566EN-D
Page 19
Chapter 2 ___________________________________________________________ Product Overview
0503-001
Figure 1 Transmitter Body
The following The following numbers refer to Figure 1 above: 1 = Signal + powering cable gland
2 = Cable gland for optional module 3 = Cable gland for optional module 4 = Cover screw (4 pcs) 5 = Display with keypad (optional) 6 = Cover LED
VAISALA________________________________________________________________________ 17
Page 20
User's Guide _______________________________________________________________________
0508-010
Figure 2 Inside of Open Transmitter
The following numbers refer to Figure 2 above: 1 = Service port (RS-232)
2 = DIP switches for analog output settings 3 = Power supply and signal wiring screw terminals 4 = Relay/RS-422/485 module (optional) 5 = Grounding connector 6 = Power supply module (optional) 7 = Analog output relay module (optional) 8 = Humidity probe cable 9 = Temperature probe cable (optional) 10 = Output isolation module (optional) 11 = Adjustment buttons (chemical purge buttons) with indicator
led
18 ___________________________________________________________________ M210566EN-D
Page 21
Chapter 2 ___________________________________________________________ Product Overview

Probe Options

0507-033
Figure 3 Probe Options
The following numbers refer to Figure 3 above: 1 = HMT331 for demanding wall-mounted applications
2 = HMT333 for ducts and tight spaces 3 = HMT334 for high pressure and vacuum applications
(up to 100 bars)
4 = HMT335 for high temperatures (up to 180 ºC, vapor tight)
*) Flange available as an option
5 = HMT337 for high humidity applications (optional warmed
and vapor tight probe head)
6 = HMT338 for pressurized pipelines (up to 40 bar)
Probe cable lengths are 2 m, 5 m and 10 m.
VAISALA________________________________________________________________________ 19
Page 22
User's Guide _______________________________________________________________________

Warmed Probe HMT337

Temperature difference between the probe head and external environment can cause a risk of condensation on the sensor. A wet probe cannot observe the actual humidity in the ambient air. If the condensed water is contaminated, the life span of the probe may shorten and calibration may change.
HMT337 probe shall be used in applications where condensation can occur due to high humidity and rapid humidity changes. The warmed probe head is heated continuously so that its temperature is always higher than in environment. This prevents condensation on the probe. The power consumption of the warmed probe is slightly higher than other probes.
20 ___________________________________________________________________ M210566EN-D
Page 23
Chapter 3 ________________________________________________________________ Installation
CHAPTER 3

INSTALLATION

This chapter provides you with information that is intended to help you install the product.

Mounting the Housing

The housing can be mounted either without the mounting plate or with optional mounting plates.

Standard Mounting without Mounting Plate

Mount the housing by fastening the transmitter to the wall with 4 screws, for example M6 (not provided).
96
169
7 Ø
0508-019
Figure 4 Standard Mounting
VAISALA________________________________________________________________________ 21
Page 24
User's Guide _______________________________________________________________________

Wall Mounting with Wall Mounting Kit

When mounting with wall mounting kit the mounting plate (Vaisala order code 214829) can be installed directly on wall or onto a standard wall box (also US junction box). When wiring through back wall, remove the plastic plug from the wiring hole in the transmitter before mounting.
0503-004
Figure 5 Mounting with Wall Mounting Kit
The following numbers refer to Figure 5 above: 1 = Plastic mounting plate
2 = Mount the plate to wall with 4 screws M6 (not provided) 3 = The arched side up 4 = Fasten HMT330 to the mounting plate with 4 fixing screws
M3 (provided)
5 = Holes for wall/junction box mounting
0503-040
Figure 6 Dimensions of Plastic Mounting Plate
22 ___________________________________________________________________ M210566EN-D
Page 25
Chapter 3 ________________________________________________________________ Installation

Mounting with DIN Rail Installation Kit

DIN rail installation kit includes a wall mounting kit, 2 clip-fasteners and 2 screws M4 × 10 DIN 7985 (Vaisala order code 215094).
1. Attach two spring holders to the plastic mounting plate by using the screws provided in the installation kit.
2. Fasten HMT330 to the plastic mounting plate with 4 screws provided for that purpose.
3. Press the transmitter onto the DIN rail so that the clip-fasteners snap into the rail.
0503-002
Figure 7 Mounting with DIN Rail Installation Kit
VAISALA________________________________________________________________________ 23
Page 26
User's Guide _______________________________________________________________________

Pole Installation with Installation Kit for Pole or Pipeline

Installation kit for pole or pipeline (Vaisala order code: 215108) includes the metal mounting plate and 4 mounting nuts for pole mounting. When mounting, the arrow in the metal mounting plate must point upwards; see Figure 10 on page 25 below.
Figure 8 Vertical Pole
The following numbers refer to Figure 8 above: 1 = Fixing brackets (2 pcs) M8 (provided) for 30 ... 102 mm
poles.
2 = Mounting nuts M8 (4 pcs)
0503-007
Figure 9 Horizontal Pole
The following number refers to Figure 9 above:
0503-006
1 = Mounting nuts M8 (4 pcs)
24 ___________________________________________________________________ M210566EN-D
Page 27
Chapter 3 ________________________________________________________________ Installation
Metal mounting plate is included in rain shield with installation kit and installation kit for pole or pipeline.
0503-041
Figure 10 Mounting with Metal Wall Mounting Plate
The following numbers refer to Figure 10 above: 1 = Mount the plate to wall with 4 screws M8 (not provided)
2 = Fasten HMT330 to the mounting plate with 4 fixing screws
M6 (provided)
3 = Note the position of the arrow when mounting. This side
must be up when mounting.
0508-023
Figure 11 Dimensions of Metal Mounting Plate (mm)
VAISALA________________________________________________________________________ 25
Page 28
User's Guide _______________________________________________________________________

Mounting Rain Shield with Installation Kit

0503-008
Figure 12 Mounting Rain Shield with Installation Kit
The following numbers refer to Figure 12 above: 1 = Fasten the rain shield with installation kit (Vaisala order
code: 215109) to the metal mounting plate with 2 (M6) mounting screws (provided).
2 = Fasten the mounting plate with rain shield with installation
kit to the wall or to the pole (see pole installation).
3 = Fasten HMT330 to the mounting plate with 4 fixing screws
(provided).
26 ___________________________________________________________________ M210566EN-D
Page 29
Chapter 3 ________________________________________________________________ Installation

Panel Mounting Frame

A panel mounting frame with adhesive tape is available as an option (Vaisala order code: 216038). After the HMT330 has been installed through the panel, a frame can be used to finish off the sawed surface of the panel.
1. Mount the HMT330 to the panel.
2. The frame is attached to the panel with an adhesive tape attached to the frame. Remove the paper protecting the tape before attaching.
3. Attach the frame to the panel around the HMT330 to finish off the panel mounting as illustrated, see Figure 13 below.
Figure 13 Panel Mounting with Frame
0503-025
Figure 14 Panel Mounting Dimensions
0503-024
VAISALA________________________________________________________________________ 27
Page 30
User's Guide _______________________________________________________________________

Wiring

Cable Bushings

A single electrical cable with screen and three to ten wires is recommended for power and analog/serial connections. The cable diameter should be 8 ... 11 mm. The number of cable bushings depends on the transmitter options. See the following recommendations for the cable bushings:
NOTE
0503-010
Figure 15 Cable Bushings
The following numbers refer to Figure 15 above: 1 = Cable for signal/powering Ø8 ... 11 mm
2 = Cable for optional module Ø8 ... 11 mm 3 = Cable for optional power supply module Ø8 ... 11 mm
When there is high electric noise level (for example, near powerful electric motor) in the operating environment it is recommended to use shielded cable or take care that the signal cables are separated from other cables.
28 ___________________________________________________________________ M210566EN-D
Page 31
Chapter 3 ________________________________________________________________ Installation

Grounding the Cables

Ground the screen of the electrical cable properly to achieve the best possible EMC performance.
0504-049
Figure 16 Grounding the Screen of Electrical Cable
VAISALA________________________________________________________________________ 29
Page 32
User's Guide _______________________________________________________________________
1. Cut back outer sheath to desired length.
2. Cut back screen braiding or screen foil to dimension X (see figure 3).
3. Push the domed cap nut (item 1) and the seal insert with contact socket of the gland (item 2+3) onto the cable as shown in the diagram.
4. Bend over the screen braiding or screen foil by about 90º (item
4).
5. Push the seal insert with the contact socket of the gland (item 2+3) up to the screen braiding or screen foil.
6. Mount lower part (item 5) on the housing.
7. Push the seal with the contact socket of the gland and (item 2+3) flush into the lower part (item 5).
8. Screw the domed cap nut (item 1) onto the lower part (item 5).

Grounding the Transmitter Housing

In case you need to ground the transmitter housing, the grounding connector is found inside the housing, see Figure 2 on page 18. Note anyhow that the probe head is connected to the same potential as the housing. Make sure that different groundings are made to the same potential. Otherwise harmful ground currents may be generated.
If it is needed to have galvanic isolation of the power supply line from the output signals, HMT330 can be ordered with optional output isolation module. This module prevents harmful grounding loops.
30 ___________________________________________________________________ M210566EN-D
Page 33
Chapter 3 ________________________________________________________________ Installation

Signal and Power Supply Wiring

When connecting transmitter with 8-pin connector, see section 8-Pin Connector on page 55. When wiring the power supply module, see section Power Supply Module on page 44.
WARNING
0506-028
Figure 17 Screw Terminal Block on Motherboard
The following numbers refer to Figure 17 above: 1 = Power supply terminals 10 ... 35 VDC, 24 VAC
2 = User port (RS-232 terminals) 3 = Analog signal terminals
Make sure that you connect only de-energized wires.
1. Open the transmitter cover by taking out the four cover screws.
2. Insert the power supply wires and signal wires through the cable
bushing in the bottom of the transmitter; see the grounding instructions in the previous sections.
3. Connect the analog output cables to terminals: Ch1 +, Ch1-,
Ch2+, Ch2-. Connect the RS-232 user port cables to terminals RxD, GND and TxD. For more information about the RS-232 connection refer to section Serial Line Communication on page
63.
VAISALA________________________________________________________________________ 31
Page 34
User's Guide _______________________________________________________________________
4. When wiring RS-485 module, relay module or additional analog output module, see section RS-422/485 Interface on page 52, section Relays on page 50, and section Third Analog Output on page 48.
5. Connect the power supply wires to the connectors: POWER 10 ... 35V+ 24V~ (+) and (-) terminals. If you are using 24 VAC power supply, see the note below before connecting the supply wires.
6. Turn on the power. The indicator led on the cover lit continuously during normal operation.
7. Close the cover and replace the cover screws. The transmitter is ready for use.

Connections to 24 VAC Power Supply

CAUTION
Separate floating supply for each transmitter is recommended (see upper part of Figure 18 on page 33). If you have to connect several transmitters or other instruments to one AC supply, the phase (~) must always be connected to (+) connector of each transmitter (see lower part of Figure 18).
24 VAC POWER SUPPLY USE
To prevent fire and/or damage, if either 24 VAC wire is grounded or connected to a "-", "0", or "GND" terminal of any other device,
you must connect the same wire on the "-" terminal also on this instrument.
32 ___________________________________________________________________ M210566EN-D
Page 35
Chapter 3 ________________________________________________________________ Installation
0507-020
Figure 18 Connections to 24 VAC Power Supply
VAISALA________________________________________________________________________ 33
Page 36
User's Guide _______________________________________________________________________

Probe Mounting

In humidity measurement and especially in calibration it is essential that temperature of the probe and measuring environment is the same. Even a small difference in temperature between the environment and the probe causes an error. As the curve below shows, if the temperature is +20 °C and the relative humidity 100 %RH, a difference of ±1 °C between the environment and the probe causes an error of ±6 %RH.
The graph below illustrates the measurement error at 100 %RH when the difference between the ambient and sensor temperature is 1 °C.
Figure 19 Measurement Error at 100 %RH
0507-023
34 ___________________________________________________________________ M210566EN-D
Page 37
Chapter 3 ________________________________________________________________ Installation

General Instructions for Probes with Cable

Mount the probes with a cable with the sensor head horizontally; this way, any water condensing on the tube cannot flow onto the sensor.
0507-024
Figure 20 Horizontal Mounting of Sensor Head
The following numbers refer to Figure 20 above: 1 = To be sealed.
2 = To be insulated. 3 = Insulate the cable. 4 = Let the cable hang loosely. This prevents condensed water
running to the sensor along the cable.
VAISALA________________________________________________________________________ 35
Page 38
User's Guide _______________________________________________________________________
When there is no alternative but to install the sensor head in the process vertically, the point of entry must be carefully insulated. The cable must also be allowed to hang loosely as this prevents any condensed water from running onto the sensor head along the cable.
0507-022
NOTE
Figure 21 Vertical Mounting of Sensor Head
The following numbers refer to Figure 21 above: 1 = To be sealed.
2 = Insulate the cable. 3 = To be insulated. 4 = Let the cable hang loosely. This prevents condensed water
running to the sensor along the cable.
Please do not attach a heated probe (HMT337) to metal structures to avoid condensation problems caused by heat conduction along the metal.
If the process temperature is much higher than that of the environment, the whole sensor head and preferably plenty of cable must be inside the process. This prevents measuring inaccuracy caused by heat conduction along the cable.
When mounted on the side of a duct or channel, the sensor head must be inserted from the side of the duct. If this is not possible and the sensor head must be inserted from the top, the point of entry must be carefully insulated.
For Vaisala probe installation kits and some installation examples see Appendix A on page 145.
36 ___________________________________________________________________ M210566EN-D
Page 39
Chapter 3 ________________________________________________________________ Installation

HMT333 for Ducts and Tight Spaces

The HMT333 is a small size (ø = 12mm) general-purpose probe suitable for ducts and channels with the installation kit available from Vaisala.
The HMT333 provides for two measuring range options. The first probe version is equipped with a flexible cable and can be used when measuring in environments up to 80 ºC. The second version is suitable for measuring in environments up to 120 ºC.
See Appendix A on page 145 for the following probe installation kits for HMT333 and installation examples.
- Duct mounting kit
- Cable gland.

HMT334 for High Pressure and Vacuum Applications

The HMT334 probe is for the dewpoint measurements in pressurized rooms and industrial processes. The probe is provided with a nut, a fitting screw and a sealing washer. Keep the fitting screw and the nut in place on the body of the sensor head during handling to prevent damage to the highly polished surface of the probe. Follow the instructions below to achieve a leak-tight assembly:
1. Unscrew the fitting screw from the nut and the sensor head.
2. Fasten the fitting screw to the chamber wall with a sealing
washer. Tighten the fitting screw into the threaded sleeve with a torque spanner. The tightening torque is 150 ± 10 Nm (110 ± 7 ft-lbs).
3. Insert the body of the sensor head into the fitting screw and
screw the nut manually to the fitting screw until the connection feels tight.
4. Mark both the fitting screw and the nut hex.
VAISALA________________________________________________________________________ 37
Page 40
User's Guide _______________________________________________________________________
0506-029
Figure 22 HMT344 Probe
The following numbers refer to Figure 22 above: 1 = Tightening cone
2 = Nut 3 = Fitting screw, M22x1.5 or NPT 1/2" 4 = Sealing washer 5 = Sensor head; Ø12 mm.
5. Tighten the nut a further 30º (1/12) turn or if you have a torque spanner tighten it with a torque of 80 ± 10 Nm (60 ± 7 ft-lbs).
0503-034
Figure 23 Tightening the Nut
NOTE
When re-tightening the nut after detachment the nut must be tightened without increased effort.
6. Clean and grease the tightening cone of the fitting screw after every tenth detachment. Change the sealing washer every time the fitting screw is detached. Use high-vacuum grease (for example Dow Corning) or similar grease.
38 ___________________________________________________________________ M210566EN-D
Page 41
Chapter 3 ________________________________________________________________ Installation
0503-033
Figure 24 Cleaning of Tightening Cone
The following numbers refer to Figure 24 above: 1 = Fitting screw
2 = Sealing washer 3 = Tightening cone 4 = Clean cotton stick
CAUTION
NOTE
In pressurized processes it is essential to tighten the supporting nuts and screws very carefully to prevent loosening of the probe by the action of pressure.
When the HMT330 is installed in a process with a pressure differing from normal atmospheric pressure, please enter the pressure value of the process (in hPa or mbar) into the transmitter memory, see command PRES and XPRES on page 77 or use display/keypad.
VAISALA________________________________________________________________________ 39
Page 42
User's Guide _______________________________________________________________________

HMT335 for High Temperatures

HMT335 is installed similarly than the HMT333 probe but without the supporting bar. Refer to Appendix A on page 145 for more information on the duct installation kit for HMT335.
To avoid incorrect humidity readings, the temperature differences between inside and outside of the duct must not be remarkable.

HMT337 for High Humidity Applications

The HMT337 is for environment where relative humidity is very high, near saturation. The warmed probe head prevents the saturation of the sensor. An additional temperature probe is also available.
See Appendix A on page 145 for a presentation of the following probe installation kits for HMT337 with installation examples:
- Duct mounting kit
- Cable gland
- Pressure tight Swagelok connector
- Vaisala's Meteorological Installation kit
The installation kits are available for both humidity and temperature probe.

Temperature Probe (Optional)

An external temperature probe for HMT337 is needed for the measurement of the ambient temperature in case you have a transmitter with a warmed probe. This allows you to measure other humidity quantities apart from dewpoint and mixing ratio.
The temperature probe is connected always in factory.
40 ___________________________________________________________________ M210566EN-D
Page 43
Chapter 3 ________________________________________________________________ Installation

HMT338 for Pressurized Pipelines

Due to the sliding fit the HMT338 is easy to install into and remove from the pressurized process. The probe is especially suitable for the measurements in pipelines. See section Ball Valve Installation Kit for HMT338 on page 152.
0508-077
Figure 25 HMT338 Probe
The following numbers refer to Figure 25 above: 1 = Clasp nut, 24 mm hex nut
2 = Fitting body, 27 mm hex head
The following two fitting body options are available:
- Fitting Body ISO1/2 solid structure
- Fitting Body NPT1/2 solid structure
Table 3 HMT338 Probe Dimensions Probe type Probe Dimension Adjustment Range
Standard 178 mm 120 mm Optional 400 mm 340 mm
VAISALA________________________________________________________________________ 41
Page 44
User's Guide _______________________________________________________________________
0507-025
Figure 26 Sealing of Fitting Body into Process

Tightening the Clasp Nut

1. Adjust the probe to a suitable depth according to the type of installation.
2. Tighten the clasp nut first manually.
3. Mark the fitting screw and the clasp nut.
4. Tighten the nut a further 50 -60º (ca. 1/6 turn) with a wrench. If you have suitable torque spanner, tighten the nut to max 45 ± 5 Nm (33 ± 4 ft-lbs).
0503-032
Figure 27 Tightening the Clasp Nut
The following numbers refer to Figure 27 above: 1 = Probe
2 = Clasp nut 3 = Pen 4 = Fitting screw
42 ___________________________________________________________________ M210566EN-D
Page 45
Chapter 3 ________________________________________________________________ Installation
NOTE
CAUTION
CAUTION
NOTE
Take care not to over tighten the clasp nut to avoid difficulties when opening it.
Take care not to damage the probe body. A damaged body makes the probe head less tight and may prevent it from going through the clasp nut.
In pressurized processes it is essential to tighten the supporting nuts and screws very carefully to prevent loosening of the probe by the action of pressure.
When installed in a process with a pressure differing from normal atmospheric pressure, please enter the pressure value of the process (in hPa or mbar) into the transmitter memory via the serial line (see command PRES and XPRES on page 77) or by using display/keypad.
VAISALA________________________________________________________________________ 43
Page 46
User's Guide _______________________________________________________________________

Optional Modules

Power Supply Module

The AC (mains) power connection may be connected to the power supply module only by an authorized electrician. A readily accessible disconnect device shall be incorporated in the fixed wiring.
0506-027
Figure 28 Power Supply Module
The following numbers refer to Figure 28 above: 1 = Connect AC (mains) voltage wires to these terminals
2 = Grounding terminal 3 = In case the module is not installed in the factory: Connect
wires from these terminals to the POWER 10 ... 36V 24V
terminals of the mother board. 4 = + 5 = -
44 ___________________________________________________________________ M210566EN-D
Page 47
Chapter 3 ________________________________________________________________ Installation
Installation
1. Disconnect the power and open the transmitter cover.
2. Remove the protective plug from the cable gland and thread the wires. In case the power supply module is installed in the factory, continue with the step 5.
3. To attach the module fasten the power module to the bottom of the housing with four screws. See the position Figure 2 on page
18.
4. Connect the wires from the terminals of the power supply module marked with + and - to the terminals POWER 10 ... 35 V 24V on the motherboard of the transmitter.
5. Connect the AC mains voltage wires to the power supply module terminals marked with N and L.
6. Attach the grounding wire to the grounding terminal on the right-hand side of the transmitter.
WARNING
WARNING
WARNING
7. Connect the power. The LED on the cover of the transmitter is lit continuously during normal operation.
Do not detach the power supply module from the transmitter when the power is on.
Do not connect the mains power to power supply module when it is not installed in the transmitter.
Always connect protective ground terminal.
VAISALA________________________________________________________________________ 45
Page 48
User's Guide _______________________________________________________________________
Warnings
Questo prodotto é conforme alla Direttiva sul basso voltaggio (73/23 CEE).
- La conduttura elettrica puó essere collegata al modulo di alimentazione elettrica soltanto da un elettricista autorizzato.
- Non staccare l´alimentazione elettrica dal trasmettitore quando é acceso.
- Non collegare la corrente elettrica al modulo di alimentazione elettrica se non é installato nel trasmettitore HMT330.
- Collegare sempre il morsetto protettivo a terra!
Dette produkt er i overensstemmelse med direktivet om lavspænding (73/23 EØS).
- Netstrømskoblingen til må kun tilsluttes strømforsyningsmodulet af en autoriseret elinstallatør
- Strømforsyningsmodulet må ikke løsgøres fra senderen, mens spændingen er sluttet til.
- Slut ikke netspændingen til strømforsyningsmodulet, når det ikke er installeret i HMT330­senderen
- Forbind altid den beskyttende jordklemme!
Dit product voldoet aan de eisen van de richtlijn 73/23 EEG (Laagspanningsrichtlijn).
- De stroom kan aan de stroomtoevoer module aangesloten worden alleen door een bevoegde monteur.
- Het is niet toegestaan de stroomtoevoer module van de transmitter los te koppelen wanneer de stroom aan is.
- Het is niet toegestaan de stroom aan de stroomtoevoer module aan te sluiten als deze niet in een HMT330-transmitter is gemonteerd.
- Altijd beschermend aardcontact aansluiten!
Este producto cumple con la directiva de bajo voltaje (72/23 EEC).
- La conexión de la alimentación principal al módulo de alimentación sólo puede realizarla un electricista autorizado.
- No desenchufe el módulo de alimentación del transmisor cuando esté encendido.
- No conecte la alimentación principal al módulo de alimentación cuando no esté instalado en el transmisor HMT330.
- Conecte siempre el terminal de protección de conexión a tierra.
See toode vastab madalpinge direktiivile (73/23 EEC).
- Voolukaabli võib vooluallika mooduli külge ühendada ainult volitatud elektrik.
- Ärge ühendage vooluallika moodulit saatja küljest lahti, kui vool on sisse lülitatud.
- Ärge ühendage voolukaablit vooluallika mooduli külge, kui seda pole HMT330-tüüpi saatjasse paigaldatud.
- Ühendage alati kaitsev maandusklemm!
46 ___________________________________________________________________ M210566EN-D
Page 49
Chapter 3 ________________________________________________________________ Installation
Ez a termék megfelel a Kisfeszültségű villamos termékek irányelvnek (73/23/EGK).
- A hálózati feszültséget csak feljogosított elektrotechnikus csatlakoztathatja a tápegységmodulra.
- A bekapcsolt távadóról ne csatolja le a tápegységmodult.
- Ne csatlakoztassa a hálózati feszültséget a tápegységmodulhoz, ha az nincs beépítve a HMT330 távadóba.
- Feltétlenül csatlakoztasson földelő védőkapcsot!
Šis produktas atitinka direktyvą dėl žemos įtampos prietaisų (73/23/EB).
- Elektros tinklą su energijos tiekimo moduliu sujungti gali tik įgaliotas elektrikas.
- Niekada neišimkite energijos tiekimo modulio iš siųstuvo, kai maitinimas yra įjungtas.
- Jei energijos tiekimo modulis nėra įmontuotas HMT330 siųstuve, nejunkite jo į elektros tinklą.
- Visada prijunkite prie apsauginės įžeminimo jungties!
Šis produkts atbilst Zemsprieguma direktīvai (73/23 EEC).
- Strāvas pieslēgumu var pieslēgt pie barošanas avota moduļa tikai autorizēts elektriķis.
- Neatvienot barošanas avota moduli no raidītāja, kad pieslēgta strāva.
- Nepievienot strāvu barošanas avota modulim, ja tas nav uzstādēts HMT330 raidītājā
- Vienmēr pievienot aizsargājošu iezemētu terminālu !
Ten produkt spełnia wymogi Dyrektywy niskonapięciowej (73/23 EEC).
- Napięcie zasilające powinno zostać podłączone do modułu zasilacza tylko przez wykwalifikowanego elektryka.
- Nie wolno odłączać modułu zasilacza od nadajnika, kiedy zasilanie jest włączone.
- Nie wolno podłączać napięcia zasilającego do modułu zasilacza, kiedy nie jest on zamontowany w nadajniku HMT330.
- Zawsze należy podłączać zabezpieczający zacisk uziemiający!
Tento výrobek vyhovuje Směrnici pro nízké napětí (73/23 EEC).
- Připojení síťového napájení k napájecímu modulu smí provádět pouze oprávněný elektrikář.
- Neodpojujte napájecí modul od snímače při zapnutém napájení.
- Nepřipojujte síťové napájení k napájecímu modulu, pokud není instalován ve snímači HMT330.
- Vždy zapojte ochrannou zemnící svorku!
VAISALA________________________________________________________________________ 47
Page 50
User's Guide _______________________________________________________________________

Galvanic Isolation for Output

If galvanic isolation of the power supply line from the output signals is needed, HMT330 can be ordered with optional output isolation module. This module prevents harmful grounding loops.
NOTE
Output isolation module is not needed when using the power supply module.
0506-030
Figure 29 Galvanic Output Isolation Module
The following number refers to Figure 29 above: 1 = Output isolation module

Third Analog Output

0503-030
Figure 30 Third Analog Output
The following numbers refer to Figure 29 above: 1 = Flat cable pins
2 = Screw terminals for signal line 3 = DIP switches to select the output mode and range
48 ___________________________________________________________________ M210566EN-D
Page 51
Chapter 3 ________________________________________________________________ Installation
Installation and Wiring
1. Disconnect the power. In case the analog output module is
installed in the factory, continue with the step 4.
2. To attach the module, open the transmitter cover and fasten the
analog output module to the bottom of the housing with four screws. See the position from the picture on page 11.
3. Connect the flat cable between the analog output module and the
motherboard's pins MODULE 2.
4. Take out the protective plug from the cable gland and thread the
wires.
5. Connect the wires to the screw terminals marked with Ch+ and
Ch- .
6. Select the current/voltage output by setting ON either of the
switches 1 or 2.
7. Select the range by setting ON one of the switches 3 ... 7.
NOTE
Only one of the switches 1 and 2 can be ON at a time.
Only one of the switches 3 ... 7 can be ON at a time.
8. Connect the power.
9. Select the quantity and scale the channel via the serial line or
display/keypad, see section Analog Output Quantities on page
93. For testing the analog output, see Section Analog Output Tests on page 95. For fault indication setting, see section Analog Output Fault Indication Setting on page 96.
VAISALA________________________________________________________________________ 49
Page 52
User's Guide _______________________________________________________________________

Relays

HMT330 can be equipped with one or two configurable relay modules. Each module contains two configurable relays. See the contact ratings in section Technical Specifications of Optional Modules on page 135.
Installation and Wiring
1. Disconnect the power and open the transmitter cover. In case the relay-module is installed in the factory, continue with step 5.
2. To attach the module fasten the relay module to the bottom of the housing with four screws. See the position in page 18.
3. When the mains power is in use attach the grounding wire to the grounding terminal.
Figure 2 on
4. Connect the flat cable between the relay module and the MODULE 1 or MODULE 2 pins of the motherboard.
5. Take out the protective plug from the cable gland and thread the relay wires.
6. Connect the wires to the screw terminals: NO, C, NC.
Selecting the Activation State of the Relay
The middlemost C terminal and either one of the terminals NO/NC must be connected. The polarity can be freely selected.
NO Normally open C Common relay NC Normally closed
Relay NOT activated: C and NC outputs are closed, NO is open Relay IS activated: C and NO outputs are closed, NC is open. Connect the power and close the cover.
NOTE
50 ___________________________________________________________________ M210566EN-D
For instructions on how to operate the relay (for example, select quantity for the relay output and set the relay setpoints) see section Operation of Relays on page 97.
Page 53
Chapter 3 ________________________________________________________________ Installation
WARNING
WARNING
0503-037
Figure 31 Relay Module
The following numbers refer to
Figure 31 above:
1 = Indication led for the relay 1 or 3 2 = Relay test buttons 3 = Flat cable pins 4 = Indication led for relay 2 or 4
The relay module may contain dangerous voltages even if the transmitter power has been disconnected. Before opening the transmitter you must switch off both the transmitter and the voltage connected to the relay terminals.
Do not connect the mains power to relay unit without grounding the transmitter.
VAISALA________________________________________________________________________ 51
Page 54
User's Guide _______________________________________________________________________

RS-422/485 Interface

0503-029
Figure 32 RS-485 Module
The following numbers refer to Figure 32 above: 1 = Flat cable pins
2 = Selection switches 3 = Screw terminals for wiring
Installation and Wiring
1. Disconnect the power. In case the RS-485-module is installed in the factory, continue with the item 4.
2. To attach the module, open the transmitter cover and fasten the RS-485 module to the bottom of the housing with four screws.
3. Connect the flat cable between the RS-485 module and the motherboard's pins MODULE1 (Communications).
4. Pull the network wirings through the cable gland.
5. Connect the twisted pair wires (1 or 2 pairs) to the screw terminals as presented in Table 4 below:
Table 4 Connecting the Twisted Pair Wires to the Screw
Terminals
Screw terminal Data line
(2-wire RS-485)
1 (not connected) RxB 2 (not connected) RxA 3 Data pair shield Data pair shield 4 B TxB 5 A TxA
Data line (4-wire RS-485/422)
52 ___________________________________________________________________ M210566EN-D
Page 55
Chapter 3 ________________________________________________________________ Installation
6. If you use RS-485 (or RS-422) to connect just one HMT330 to a
master computer, enable the internal termination of HMT330 by switching switches 1 and 2 ON. Make sure that the master's end of the line is also terminated (by using master's internal termination or with a separate terminator).
If you are connecting many transmitters to the same RS-485 bus, make sure that switches 1 and 2 are OFF and terminate the bus with separate terminators at both ends. This allows removing any transmitter without blocking the bus operation.
NOTE
If you use the internal termination of the transmitter at the end of the RS-485 bus (instead of using separate terminators) removing that transmitter will block the bus operation.
7. Use the bus type (4-wire/2-wire) to select the selection switch 3.
In 4-wire mode RS-485 master sends data to the HMT330 through terminals RxA and RxB and receives data from HMT330 through terminals TxA and TxB.
Figure 33 4-Wire RS-485 Bus
Table 5 4-Wire (Switch 3: On)
RS-485 master Data HMT330
TxA RxA TxB RxB RxA TxA RxB TxB
VAISALA________________________________________________________________________ 53
Page 56
User's Guide _______________________________________________________________________
Table 6 2-Wire (Switch 3: Off)
RS-485 master Data HMT330
A A B B
8. When operating in communication mode RS-422, set both switches 3 and 4 to ON position (4-wire wiring is required for RS-422 mode).
9. Connect the power and close the cover.
54 ___________________________________________________________________ M210566EN-D
Page 57
Chapter 3 ________________________________________________________________ Installation

8-Pin Connector

0503-026
Figure 34 Wiring of Optional 8-Pin Connector
Table 7 Wiring of 8-Pin Connector
PIN/Terminal Wire Serial Signal Analog Signal
1 White Data out TX A - ­2 Brown (serial GND) (serial GND) Signal GND (for
3 Green - - Ch 2+ 4 Yellow - - Ch 1 + 5 Grey Supply - Supply - Supply ­6 Pink Supply + Supply + Supply + 7 Blue Data in RX B - - 8 Shield/Red Cable shield Cable shield Cable shield
RS-232 (EIA-232) RS-485 (EIA-485)
both channels)
VAISALA________________________________________________________________________ 55
Page 58
User's Guide _______________________________________________________________________
This page intentionally left blank.
56 ___________________________________________________________________ M210566EN-D
Page 59
Chapter 4 _________________________________________________________________ Operation
CHAPTER 4

OPERATION

This chapter contains information that is needed to operate this product.

Getting Started

Within a few seconds after power-up the led on the cover of the transmitter is lit continuously indicating normal operation. When using the optional display and turning the transmitter on the first time, the language selection menu window opens. Select the language with
arrow buttons and press the SELECT button.
The pressure has an effect on humidity calculations and accuracy. Therefore, accurate calculations can be achieved only when the ambient pressure is taken into consideration. For instructions on how to set the pressure, see section Pressure Compensation Setting on page
77.

Display/Keypad (Optional)

Basic Display

Display shows you the measurement values of the selected quantities in the selected units. You can select 1 ... 3 quantities for the numerical basic display (see section Changing Quantities and Units on page 74.)
VAISALA________________________________________________________________________ 57
Page 60
User's Guide _______________________________________________________________________
NOTE
0507-002
Figure 35 Basic Display
The following numbers refer to Figure 35 above: 1 = The Info shortcut button, see section Device Information on
page 82 below
2 = The Graph shortcut button, see section Graphic History on
page 58 below
3 = Quantities selected for display
From any view, a four-second press on the right-hand function button takes you directly to the basic display.

Graphic History

The graphical display shows the data trend of the selected quantities, one at a time. The graph is updated automatically while measuring. Use the following functions in the graphical display:
- Press the NEXT button to have the trend graph and max/min graph
in turns and browse through the quantities selected for display.
- Press the EXIT button to get back the basic display.
0507-001
Figure 36 Graphical Display
Trend graph: Shows you a curve of average values. Each value is a
calculated average over a period. See Table 8 on page 59 below.
58 ___________________________________________________________________ M210566EN-D
Page 61
Chapter 4 _________________________________________________________________ Operation
Max/min graph: Shows you the minimum and maximum values in a form of curve. Each value is max/min over a time period. See Table 8 below.
Table 8 Periods for Trend and Max/Min Calculations
Observation Period Period for Trend/Max/Min
Calculations (Resolution)
20 minutes 10 seconds 3 hours 90 seconds 1 day 12 minutes 10 days 2 hours 2 months 12 hours 1 year 3 days
- Press the arrow buttons to zoom in and out the time in the graph
window.
- Press the arrow buttons to have the cursor mode where you can
observe individual measuring points. Press an arrow button to move the cursor (vertical bar) along the time axis. The numerical value at the cursor position is shown at the left upper corner. Time from the present to the chosen moment is shown at the right upper corner.
Table 9 Graph Information Messages in Cursor Mode
Message Interpretation
Power outage Power failure (marked also with dashed
vertical line) No data Quantity has not been selected for the display Device failure General device failure T meas. failure Temperature measurement/sensor failure RH meas. failure Humidity measurement/sensor failure Adj. mode active Adjustment mode active (data recorded in the
adjustment mode is not displayed)
A question mark after time tells you that at least one power failure (dashed vertical line) has occurred after the chosen moment. In this case, the exact time difference between the present and the cursor position is not exactly known.

Menus and Navigation

You can change settings and select functions in the menus.
1. Open the MAIN MENU by pressing any of the arrow
buttons in the basic (numeric) display mode.
2. Move in the menus by using the arrow buttons.
VAISALA________________________________________________________________________ 59
Page 62
User's Guide _______________________________________________________________________
3. Open a submenu with button.
4. Press to return to the previous level.
5. Function button EXIT returns you back to the basic display.
0507-003
Figure 37 Main Menus
Language Setting
1. Go back to the basic display by keeping the right-hand button pressed for four seconds.
2. Open the MAIN MENU by pressing any of the buttons.
3. Select System (the lowest row), press button.
4. Select Language: ... (the third row marked with a flag icon), press SELECT button (left-hand button).
5. Select the menu language with buttons and press SELECT button (left-hand button).
6. Press EXIT to return to the basic display.
60 ___________________________________________________________________ M210566EN-D
Page 63
Chapter 4 _________________________________________________________________ Operation
Rounding Setting
Round off one decimal by using the Rounding function. The default setting is rounding on. Rounding has no effect on quantities without decimals.
1. Open the MAIN MENU by pressing any of the arrow
buttons.
2. Select Display and confirm by pressing the arrow button.
3. Select Rounding and press ON/OFF button.
4. Press EXIT to return to the basic display.
Display Backlight Setting
As a default the display backlight is always on. In the automatic mode the backlight stays on for 30 seconds from the last press of any button. When pressing any button, the light turns on again.
1. Open the MAIN MENU by pressing any of the arrow
buttons.
2. Select Display, press the arrow button.
3. Select Backlight, press the CHANGE button.
4. Select On/Off/Automatic, press the SELECT button.
5. Press EXIT to return to the basic display.
Display Contrast Setting
1. Open the MAIN MENU by pressing any of the arrow
buttons.
2. Select Display, press the arrow button.
3. Select Contrast, press the ADJUST button.
4. Adjust the contrast by pressing the arrow buttons.
5. Press OK and EXIT to return to the basic display.
VAISALA________________________________________________________________________ 61
Page 64
User's Guide _______________________________________________________________________
Keypad Lock (Key guard)
This function locks the keypad and prevents unintentional key presses.
1. Keep pressing the left-hand function button for 4 seconds to lock the keypad (at any display).
2. To unlock the keypad, press the OPEN button for 4 seconds.
Menu PIN Lock
You can prevent unauthorized changes of the device settings by activating the menu PIN lock. When this function is activated, the basic display and graphical view are available but access to the menus is locked. The key symbol indicates the activation of this feature.
1. Open the MAIN MENU by pressing any of the arrow buttons.
2. Select System, press the arrow button.
NOTE
3. Select Menu PIN, press the ON button.
4. Enter a PIN code by using the arrow buttons. Press OK to confirm the setting. Now the PIN lock is on and a key symbol is shown in a display.
5. Press EXIT to return to the basic display. Returning to the menu is possible only by entering the correct PIN code.
When you want to turn off the PIN lock, go to the menu by giving the PIN code and select System, Menu PIN, press OFF button.
In case you have forgotten the PIN code, open the transmitter cover and press the ADJ button once. Wait for a few seconds and the
adjustment menu opens. Select Clear menu PIN, press CLEAR.
You can also disable the keypad completely with serial command LOCK.
62 ___________________________________________________________________ M210566EN-D
Page 65
Chapter 4 _________________________________________________________________ Operation
Factory Settings
Use the display/keypad to restore the factory settings. This operation does not affect the adjustments. Only settings available in the menus are restored.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select System by pressing the ► arrow button.
3. Select Factory settings and press the REVERT button to
confirm your selection. Press the YES button to reset all settings to the factory defaults.
See section General Settings on page 74 for a description of the other menu options.

MI70 Link Program for Data Handling

The recorded data can be transferred to a PC by using MI70 Link program. You can examine the recorded data easily in Windows environment and transfer it further to a spreadsheet program (such as Microsoft Excel) or virtually to any Windows program in numeric or graphical format. MI70 Link program allows you also to monitor transmitter readings directly with a PC (real-time window function).
MI70 Link program is available from Vaisala, see list of accessories in section Options and Accessories on page 136.
1. Connect the connection cable between the serial port of your PC
and the Service Port of HMT330; see Figure 38 on page 64 below.
2. Check that the HMT330 is powered and start using the MI70
Link program.
Use a MI70 Link version 1.07, or a newer one, to be able to utilize all the functions of HMT330.

Serial Line Communication

Connect the serial interface by using either the user port or the service port.
For permanent interfacing to host system, use the user port. You can change the serial settings and operate in RUN, STOP and POLL modes.
VAISALA________________________________________________________________________ 63
Page 66
User's Guide _______________________________________________________________________
For temporary RS-232 connections use the service port. Service port is always available with fixed serial settings.
0507-026
Figure 38 Service Port Connector and User Port Terminal on
Mother Board
The following numbers refer to Figure 38 above: 1 = Service port connector
2 = User port terminals

User Port Connection

Use suitable serial cable between the user port RxD, GND and TxD screw terminals and the PC serial port, see Figure 39 on page 64 below.
Table 10 Default Serial Communication Settings for the User
Port
Parameter Value
Bauds 4800 Parity Even Data bits 7 Stop bits 1 Flow control None
+
24 V~
RxD
GND
RS-232 10…36 V=
TxD
Ch1+
Ch1–
Ch2+
Ch2–
ANA LOG OUTPUTS USERPORT POWER
1 2 3 4 5
6 7 8 9
IOIOI
0506-033
Figure 39 Connection Example Between PC Serial Port and
User Port
64 ___________________________________________________________________ M210566EN-D
Page 67
Chapter 4 _________________________________________________________________ Operation
Connections to pins 4, 6, 7 and 8 on PC serial port are required only if you are using software requiring hardware handshaking.
After power-up the transmitter (in STOP-mode) outputs the software version and the command prompt.
HMT330/2.02 >
In RUN mode a measurement output starts immediately after power­up.
In POLL mode, the transmitter does not output anything after power­up (see section SMODE on page 105).
NOTE
User port cannot be used when RS-485 module is connected.

Service Port Connection

1. Connect the serial interface cable (optional accessory, order
code: 19446ZZ) between the serial port of your PC and the service port connector on the motherboard, see Figure 38 on page 64.
Table 11 Fixed Communication Settings for Service
Port
Parameter Value
Bauds 19200 Parity No Data bits 8 Stop bits 1 Flow control None
2. Open a terminal program and set the communication settings
(see the following section for more detailed instructions).
3. Power-up the HMT330.
VAISALA________________________________________________________________________ 65
Page 68
User's Guide _______________________________________________________________________

Terminal Program Settings

The following instructions show a connection example with HyperTerminal program (included in the Microsoft Windows).
Follow the instructions below to open a HyperTerminal program:
1. Start HyperTerminal. To get help for starting HyperTerminal, click "Start", select "Help" to open Windows help, and search for "HyperTerminal".
0507-027
Figure 40 Starting Hyper Terminal Connection
2. In the "New Connection" window of the HyperTerminal, define a name for HMT330 serial connection, for example "HMT330". Click OK.
3. In the "Connect using" pull down menu, select the PC communications port where the serial cable is connected. (If your computer has only one COM port, it is called "COM1"). Click OK.
66 ___________________________________________________________________ M210566EN-D
Page 69
Chapter 4 _________________________________________________________________ Operation
0507-028
Figure 41 Connecting to Hyper Terminal
4. Set the port settings in the "Properties" window to match the
settings of your HMT330 user port/service port. For HMT330, "Flow control" must always be set to "None". Finally click OK to start using the serial connection.
0503-015
Figure 42 Hyper Terminal Serial Port Settings
5. Select "File" Æ "Save" in the HyperTerminal main window to
save the serial port settings. To use the saved settings later, start HyperTerminal, click cancel in the "New Connection" window, and select "File" Æ "Open".
VAISALA________________________________________________________________________ 67
Page 70
User's Guide _______________________________________________________________________

List of Serial Commands

The bold text in the brackets is the default setting. Commands are entered by typing them on your computer and pressing the Enter key.
Measurement commands R Start the continuous outputting S Stop the continuous outputting INTV [0…255 S/MIN/H] Set the continuous output interval
(for RUN mode)
SEND [0…99] Output the reading once SEND D Outputting the reading with the
raw data
SMODE [STOP/RUN/POLL] Set the serial interface SERI [baud p d s] User port settings (default: 4800 E
7 1) baud: 300…115200
ADDR [0…99] Set the transmitter address (for
POLL mode)
OPEN [0…99] Open temporarily connection to
the POLL-mode device
CLOSE Close the connection (back to
POLL mode)
Formatting commands FORM Set the output format of SEND and
R commands
TIME Set the time DATE Set the date FTIME [ON/OFF] Add time to output to SEND and R
outputs
FDATE [ON/OFF] Add date to R and SEND outputs FST [ON/OFF] Add the state of probe heating and
chemical purge in connection with SEND and R commands
UNIT Select the metric or non-metric
output units
Data recording commands DSEL Select data recording and display
quantities.
PLAY [0 ... 17] Output recorded data file DIR Display recorded files DELETE Delete the graph files UNDELETE Recovers the deleted files
Chemical purge commands PUR Set the automatic chemical purge PURGE Start the manual chemical purge
68 ___________________________________________________________________ M210566EN-D
Page 71
Chapter 4 _________________________________________________________________ Operation
Calibration and adjustment commands CRH Relative humidity calibration CT Temperature calibration CTA Additional temperature probe
calibration
FCRH Relative humidity calibration after
sensor change
CTEXT Give the text to calibration
information field
CDATE Set the calibration date ACAL Analog output calibration
Setting and testing the analog outputs AMODE View the analog output modes ASEL Select the parameters for the
analog outputs
ITEST Test the analog outputs AERR Change the analog error output
values
Setting and testing the relays RSEL Set and view the relays RTEST Test the relays
Other commands ? Output the information about the
device
?? Output the information about the
device in POLL-state
ECHO [ON/OFF] Turn the serial interface echo
ON/OFF
FIND All devices in POLL mode are sent
their addresses
HELP List the most common commands PRES [hPa] Set the value for pressure
compensations
XPRES [hPa] Set the value for pressure
compensations, temporarily
ERRS List present transmitter errors FILT Set the result filtering VERS Display the software version
information
XHEAT Sensor heating LOCK Lock the menu/disable keypad
VAISALA________________________________________________________________________ 69
Page 72
User's Guide _______________________________________________________________________

Getting Measurement Message from Serial Line

Enter the R command to start output of measurements. Enter the S command, press the Esc button or reset the transmitter to stop outputting. See command SMODE to change the default (power-up) operation mode.
Format the output by using the following commands:
- Outputting interval can be changed with the INTV command.
- Output message format can be changed with a command FORM.
- Status of chemical purge and probe heating can be added with a
command FST.
- Date and time information can be added with commands FDATE
and FTIME
Examples:
>r RH= 11.3 %RH T= 28.8 'C >
>r RH= 11.3 %RH T= 28.8 'C Tdf= -3.7 'C Td= -4.2 'C a= 3.2 g/m3 x= 2.8 g/kg T w= 13.0 'C ppm= 4440 pw= 4.48 hPa pws= 39.53 hPa h=
36.1 kJ/kg dT= 32.5 'C
>
Stopping Continuous Outputting
S
Use the S command to end the RUN mode. After this command all other commands can be used.
Outputting Reading Once
SEND
Use the SEND command to output the reading once in STOP mode:
The output format depends on which parameters the transmitter can output.
70 ___________________________________________________________________ M210566EN-D
Page 73
Chapter 4 _________________________________________________________________ Operation
Examples:
RH= 98.4 %RH T= 31.1 'C
RH= 98.4 %RH T= 31.1 'C Td= 36.0 'C Tdf= 36.0 'C a= 42.4 g/m3 x= 38.8 g/kg Tw= 30.8 'C ppm= 62414 pw= 59.53 hPa pws= 60.52 hPa h= 130.7 kJ/kg
If a value is too long to fit to the allocated space in the output, or if there is an error in outputting the quantity, the value is displayed with stars ‘*’.
Example:
RH=***.* %RH T= 31.0 'C
The output mode can be changed with the commands: FORM, FST, FDATE, and TIME.
Outputting Reading with Raw Data
SEND D
Example:
>send d
24.1720 15.0399 -3.5743 189.2324 15.0709 15.0399
23.9765
Where the readings (from the left) are:
24.1720 = Temperature of the humidity probe ( ºC)
15.0399 = RH (%RH)
-3.5743 = Tdf (C)
189.2324 = Capacitance (pF)
15.0709 = RH raw: calculated from scaled capacitance (%RH)
15.0399 = Enhancement factor corrected RH (%RH)
23.9765 = Temperature of the additional temperature probe
(optional) (ºC)
VAISALA________________________________________________________________________ 71
Page 74
User's Guide _______________________________________________________________________

Formatting Serial Line Message

TIME and DATE
To set time enter the TIME command. To set date enter the DATE command.
TIME DATE These time and date settings are shown on the timestamps of PLAY
command. When you want to include time and date in the R and
SEND commands, use the FTIME and FDATE commands.
Example:
>TIME Current time is 04:12:39 Enter new time (hh:mm:ss) ? 12:24:00 >DATE Current date is 2000-01-01 Enter new date (yyyy-mm-dd) ? 2004-07-05 >
NOTE
Time and date are cleared to 2000-01-01 00:00:00 at reset or at power failure.
FTIME and FDATE
FTIME and FDATE commands will enable/disable output of time and date to the serial line. To add time to R and SEND outputs enter:
FTIME [x] To add date to R and SEND outputs enter:
FDATE [x] where x = ON or OFF
72 ___________________________________________________________________ M210566EN-D
Page 75
Chapter 4 _________________________________________________________________ Operation
N
Example:
>send RH= 98.4 %RH T= 31.0 'C >ftime on Form. time : ON >send 03:47:59 RH= 98.4 %RH T= 31.0 'C >fdate on Form. date : ON >send 2004-07-05 03:48:03 RH= 98.4 %RH T= 31.0 'C >
FST
To output the state of optional probe heating and chemical purge in connection with SEND and R commands enter:
FST [x]
Where x = ON or OFF (default)
Example:
>fst on Form. status : ON >send N 0 RH= 40.1 %RH T= 24.0 'C Td= 9.7 'C Tdf= 9.7 'C a= 8.7 g/m3 x= 7.5 g/kg Tw= 15.6 'C ppm= 11980 pw= 12.00 hPa pws= 29.91 hPa h= 43.2 kJ/kg >purge Purge started, press any key to abort. >send S 134 RH= 40.2 %RH T= 24.1 'C Td= 9.8 'C Tdf= 9.8 'C a= 8.8 g/m3 x= 7.5 g/kg Tw= 15.7 'C ppm= 12084 pw= 12.10 hPa pws= 30.11 hPa h= 43.5 kJ/kg >
For more information on chemical purge, see section Chemical Purge (Optional) on page 108.
Where the state of the probe is indicated by the following letters and values:
... xxx = Normal operation where h xxx = Probe heat power X ... xxx = Sensor heating where xxx = Sensor temperature (ºC) H ... xxx = Chemical purge where xxx = Sensor temperature (ºC) S ... xxx = Sensor cooling
where xxx = Sensor temperature (ºC)
after purge
VAISALA________________________________________________________________________ 73
Page 76
User's Guide _______________________________________________________________________

General Settings

Changing Quantities and Units

To change quantities and units use serial commands or the optional display/keypad. See Table 1 on page 15 for available quantities and Table 2 on page 16 for optional quantities.
NOTE
Only the quantities selected when ordering the device can be selected as a display output quantity.
Using Display/Keypad
Use display/keypad to select the display output quantities.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Press the ► arrow button to select Display.
3. Press ► arrow button to select Quantities.
4. Select the quantity by using the ▲▼ arrow buttons. Confirm the
selection by pressing SELECT. You can select 1 ... 3 display quantities at a time.
5. Press EXIT to return to the basic display.
To select display units:
1. Press any of the arrow buttons to open the MAIN MENU.
2. Press the ► arrow button to select Display.
3. Use the ▲▼arrow buttons to select Units. Confirm the selection by pressing the right-hand arrow button.
4. Use the ▲▼ arrow buttons to select display units. Confirm the selection by pressing CHANGE. The unit changes from metric to non-metric or the other way round.
5. Press EXIT to return to the basic display.
NOTE
74 ___________________________________________________________________ M210566EN-D
Changing the display quantities/units (by using the display/keypad) has no effect on the serial output data.
Page 77
Chapter 4 _________________________________________________________________ Operation
Using Serial Line
Use the serial line command FORM to change the format or select a certain quantities for the output commands SEND and R. Use the serial line command UNIT to select metric or non-metric output units.
FORM
Use the serial line command FORM to change the format or select a certain quantities for the output commands SEND and R.
FORM [x]
where x = Formatter string
Formatter string consists of quantities and modifiers.
When entering the command, use the abbreviations of the quantities. For more information on quantities, see Table 1 and Table 2 on page
15.
The modifiers are presented in Table 12 below.
Table 12 FORM Command Modifiers
Modifier Description
x.y Length modifier (number of digits and decimal places) #t Tabulator #r Carriage-return #n Line feed “" String constant #xxx Special character, code "xxx" (decimal), for example
#027 for ESC U5 Unit field and length ADDR Transmitter address with two characters [00...99] ERR Error flags for P, T, Ta, RH [0000 ... 1111], 0 = no error STAT Transmitter status in 7 character field, for example:
N 0 no heating
h 115 probe heating active, power 115/255
H 159.0 purge heating active, temperature 159ºC
S 115.0 purge cooling active, temperature 115ºC
X 95.0 sensor heating active, temperature 95ºC SN Transmitter serial number TIME Time [hh:mm:ss] DATE Date [yyyy-mm-dd] OK Pressure stability indicator, two characters [OK or " "] CS2 Modulus-256 checksum of message sent so far, ascii
encoded hexadecimal notation
VAISALA________________________________________________________________________ 75
Page 78
User's Guide _______________________________________________________________________
Modifier Description
CS4 Modulus-65536 checksum of message sent so far, ascii
encoded hexadecimal notation
CSX NMEA xor-checksum of message sent so far, ascii
encoded hexadecimal notation
A3H Pressure tendency [* or 0...8]
Example:
>form "RH=" 4.2 rh U5 #t "T=" t U3 #r #n RH= 14.98%RH T= 74.68'F
>send RH= 16.03%RH T= 74.66'F
>form "Tfrost=" tdf U3 #t "Temp=" t U3 #r#n Tfrost= 36.0'C Temp= 31.0'C >
Command ‘FORM /’ will return the default output format. The default output format depends on the device configuration.
NOTE
>form / >send RH= 98.4 %RH T= 31.1 'C >
UNIT
Use the UNIT command to select metric or non-metric output units:
UNIT [x]
where x = M or N
where M = Metric units
N = Non-metric units
This command changes both the serial output and display units to either metric or non-metric units. When you want to output both metric and non-metric units simultaneously on the display, select the display units later by using the display/keypad.
76 ___________________________________________________________________ M210566EN-D
Page 79
Chapter 4 _________________________________________________________________ Operation

Pressure Compensation Setting

The pressure has an effect on humidity calculations and accuracy. Therefore, accurate calculations can be achieved only when the process pressure is taken into consideration.
Note that conversions from mmHg and inHg are defined at 0°C and for mmH2O and inH2O at 4°C.
NOTE
Pressure compensation is intended to be used in normal air only. When measuring in other gases, please contact Vaisala for further information.
Using Display/Keypad
Use display/keypad to set the pressure compensation. To select the pressure unit using display/keypad, see section Changing Quantities and Units on page 74.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Measuring and press the ►arrow button to confirm your
selection.
3. Select Pressure compensation and press the arrow button to confirm you selection.
4. Press SET and enter the pressure value in the chosen unit by using the arrow buttons.
5. Press OK and EXIT to return to the basic display.
Using Serial Line
PRES and XPRES
Command XPRES should be used if the value is changed frequently. Its value is not retained at reset, and when set to 0; last value set with
PRES is used instead. Use the serial line and do the following: PRES [aaaa.a]
XPRES [aaaa.a]
where aaaa.a = Absolute process pressure (hPa)
VAISALA________________________________________________________________________ 77
Page 80
User's Guide _______________________________________________________________________
Example:
>pres Pressure : 1013.00 hPa ? >pres 2000 Pressure : 2000.00 hPa >
Table 13 Multiplication Factors
From To: hPa
mbar 1 Pa N/m2 0.01 mmHg torr 1.333224 inHg 33.86388 mmH2O 0.09806650 inH2O 2.490889 atm 1013.25 at 980.665 bar 1000 psia 1) 68.94757
1) psia = psi absolute.
Example:
29.9213 inHg = 29.9213 × 33.86388 = 1013.25 hPa

User Port Serial Settings

Using Display/Keypad
The communication settings for the user port can be changed via the serial line or by using the optional display/keypad. The communication settings for the service port are fixed and not changeable.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Interfaces and press the ►arrow button to confirm your
selection.
3. Select Serial interface and press the arrow button to confirm
your selection.
4. Select Bit rate/Serial format/Comm. mode by pressing the CHANGE button. Use the ▲▼ arrow buttons to select and press SELECT to confirm your selection.
5. If you selected RUN communication mode, select RUN interval for and press SET to confirm your selection.
78 ___________________________________________________________________ M210566EN-D
Page 81
Chapter 4 _________________________________________________________________ Operation
6. Use the arrow buttons to set the measuring interval and the unit.
Press OK to confirm your settings.
7. If you selected POLL communication mode, select POLL
address and press SET to confirm your selection.
8. Use the arrow buttons to set the transmitter address. Press OK to
confirm the setting.
9. Use the arrow buttons to select ECHO. Press ON to turn to it
on. Press OFF to turn it off.
10. Press EXIT to return to the basic display. The new user port settings set using the display/keypad are effective
immediately.
Using Serial Line
SERI
Use the serial line command SERI [b p d s] to set communication settings for the user port.
SERI [b p d s] where b = Bit rate (110, 150, 300, 600, 1200, 2400, 4800, 9600,19200,
38400, 57600, 115200) p = Parity (n = none, e = even, o = odd) d = Data bits (7 or 8) s = Stop bits (1 or 2)
Example:
>SERI 600 N 8 1 600 N 8 1 >
You need to reset the transmitter to activate the new communication settings set with command SERI.
The settings can be changed one parameter at a time or all parameters at once:
>SERI O changing parity only 4800 O 7 1
>SERI 600 N 8 1 600 N 8 1 >
VAISALA________________________________________________________________________ 79
changing all parameters
Page 82
User's Guide _______________________________________________________________________
SMODE
Use the command SMODE to set the user port start-up operating mode.
SMODE [xxxx]
where xxx = STOP, RUN or POLL
Table 14 Selection of Output Modes
Mode Output Available Commands
STOP RUN Automatic output Only command S POLL
Selected output mode will be activated after power outages.
Only with the SEND command
Only with the SEND [addr] command
All (default mode)
Use with RS-485 buses, see Operation of RS-485 Module on
104.
page
INTV
Use the command INTV to set the outputting interval for the RUN mode.
INTV [xxx yyy]
where xxx = Output interval (0 ... 255). 0: the fastest possible output rate.
yyy = Unit (s, min or h)
Example:
>INTV 10 min Output intrv. : 10 min >
80 ___________________________________________________________________ M210566EN-D
Page 83
Chapter 4 _________________________________________________________________ Operation
ECHO
Use the command ECHO to set the user port echo. The command either enables or disables echo of characters received.
ECHO [x]
where x = ON (default) or
= OFF
NOTE
You can use the SERI, SMODE, INTV and ECHO commands to change/view the user port settings even if you are currently connected to the service port.

Data Filtering

The averaging data filter calculates an average over a certain period of time. The lowest measurement noise is achieved with the extended filtering. There are three filtering levels available.
Table 15 Filtering Levels
Setting Filtering level
OFF No filtering ON (default) Standard = short filtering (approximately 15 s moving
average)
EXTENDED Extended filtering (default: approximately 1 min
average)
Use display/keypad to set the filtering level.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Measuring by pressing the arrow button.
3. Select Filtering and press CHANGE to confirm your selection.
4. Select Off/Standard/Extended and press SELECT to confirm
your selection.
5. Press EXIT to return to the basic display.
VAISALA________________________________________________________________________ 81
Page 84
User's Guide _______________________________________________________________________
FILT
Use the serial line command FILT [xxx] to set the filtering level.
FILT [xxx]
where xxx = OFF, ON or EXT (default = ON)

Device Information

Use the display/keypad or the serial line to display the device information.
Press the INFO button in the basic display to see the following information:
- Current sensor operation (for example, chemical purge), if any, in progress
- Present or past unacknowledged errors, if any
- Device information
- Adjustment information fed by the user
- Measuring settings
- Information on chemical purge settings (when applicable)
- Serial interface information
- Analog output information
- Relay output information (when applicable)
0507-030
Figure 43 Device Information on Display
Proceed in the information views by pressing the MORE button as many times as you get the desired information. You can browse through the information displays also with arrow buttons. Press OK to return to the basic display.
82 ___________________________________________________________________ M210566EN-D
Page 85
Chapter 4 _________________________________________________________________ Operation
?
Use the serial line command ? to check the current transmitter configuration. Command ?? is similar but can also be used if the transmitter is in POLL mode.
Example:
>? HMT330 / 2.02 Serial number : Z4140007 Batch number : Z4020017 Adjust. date : 2005-06-20 Adjust. info : VAISALA/HEL Date : 2000-02-12 Time : 22:58:25 Serial mode : STOP Baud P D S : 4800 E 7 1 Output interval: 0 s Address : 0 Echo : ON Pressure : 1013.25 hPa Filter : OFF Ch1 output : 4...20mA Ch2 output : 4...20mA Ch1 RH low : 0.00 %RH Ch1 RH high : 100.00 %RH Ch2 T low : -40.00 'C Ch2 T high : 60.00 'C Module 1 : not installed Module 2 : not installed >
HELP
Use the command HELP to list the commands.
Example:
>help ? ACAL ADDR AERR ASCL ASEL CDATE CLOSE CODE CRH CT CTA CTEXT DATE DELETE DIR DSEL DSEND ECHO ERRS FCRH FDATE FILT FORM FST FTIME HELP INTV ITEST MODS OPEN PLAY PRES R RESET SEND SERI SMODE TEST TIME UNDELETE UNIT VERS XPRES >
VAISALA________________________________________________________________________ 83
Page 86
User's Guide _______________________________________________________________________
ERRS
Use the command ERRS to display transmitter error messages, see Table 16 on page 118.
Example:
>ERRS NO ERRORS >
Example:
>ERRS FAIL Error: Temperature measurement malfunction Error: Humidity sensor open circuit >
VERS
Use the command VERS to display software version information. Example:
>vers HMT330 / 2.02 >
84 ___________________________________________________________________ M210566EN-D
Page 87
Chapter 4 _________________________________________________________________ Operation

Resetting Transmitter by Using Serial Line

RESET
This command resets the device. The user port switches to start-up output mode selected with command SMODE.

Locking Menu/Keypad by Using Serial Line

LOCK
Use the LOCK command to turn on the menu lock with 4-digit PIN code, for example 4444.
LOCK [x yyyy]
where x = 1 (Menu locked)
yyyy = 4-digit PIN code
Example:
>lock 1 4444 Keyboard lock : 1 [4444] >
VAISALA________________________________________________________________________ 85
Page 88
User's Guide _______________________________________________________________________
Use the LOCK command to turn on the menu lock without PIN code access possibility.
LOCK [x]
where x = 1 (Menu locked)
Example:
>lock 1 Keyboard lock : 1 >
Use the LOCK command to disable the keypad completely.
LOCK [x]
where
NOTE
x = 2 (Keypad disabled)
Example:
>lock 2 Keyboard lock : 2 >
Open the locks with the serial command LOCK 0. You can open the menu lock also by using the keypad, if a PIN code has been set.
86 ___________________________________________________________________ M210566EN-D
Page 89
Chapter 4 _________________________________________________________________ Operation

Data Recording

Data recording function is always on and collects data automatically into the memory of the device. Recorded data do not disappear from the memory when the power is switched off. Collected data can be observed in a form of a graph in the graphical view of the display or it can be listed out by using the serial line or MI70 Link program.

Selecting Data Recording Quantities

If the device is provided with the optional display, the recorded quantities are always those selected for the display. Up to three quantities can be recorded at a time. For instructions on how to select the display quantities with the keypad, see section Changing Quantities and Units on page 74.
DSEL
Use the serial line command DSEL to select the quantities to be recorded if the transmitter is not equipped with display/keypad.
DSEL [xxx]
where xxx = Data recording quantity. See Table 1 on page 15 and Table 2
on page 16 for the quantities.
Example:
>dsel rh t tdf RH T Tdf >
Enter the command without parameters and press ENTER to display current recording parameters.
VAISALA________________________________________________________________________ 87
Page 90
User's Guide _______________________________________________________________________

View Recorded Data

If the device is provided with the optional display, the graphical display shows the data of the selected quantities, one at a time. See section Graphic History on page 58 for details about graphical display.
You may also dump the logged data to the serial line in numeric form with the following commands.
DIR
Use the serial line and enter the DIR command to check the available files.
The device records six files (six observation periods) for each selected quantity. Thus, total amount of the files depends on the amount of the selected quantities being at minimum 6 and at maximum 18. See Table 8 on page 59.
Select, for example, three quantities (RH, T, and Tdf). The last two columns illustrate software information that is not essential for the user.
Example:
>dir 0 RH latest 20 minutes 00-02-12 22:54:27 135 020A 1 RH latest 3 hours 00-02-12 19:54:27 135 025A 2 RH latest 1 day 00-02-11 20:16:57 135 040C 3 RH latest 10 days 00-02-01 17:16:57 135 0802 4 RH latest 2 months 99-12-07 11:16:57 135 080C 5 RH latest 1 year 99-01-03 23:16:57 135 1003 6 T latest 20 minutes 00-02-12 22:54:27 135 020A 7 T latest 3 hours 00-02-12 19:54:27 135 025A 8 T latest 1 day 00-02-11 20:16:57 135 040C 9 T latest 10 days 00-02-01 17:16:57 135 0802 10 T latest 2 months 99-12-07 11:16:57 135 080C 11 T latest 1 year 99-01-03 23:16:57 135 1003 12 Tdf latest 20 minutes 00-02-12 22:54:27 135 020A 13 Tdf latest 3 hours 00-02-12 19:54:27 135 025A 14 Tdf latest 1 day 00-02-11 20:16:57 135 040C 15 Tdf latest 10 days 00-02-01 17:16:57 135 0802 16 Tdf latest 2 months 99-12-07 11:16:57 135 080C 17 Tdf latest 1 year 99-01-03 23:16:57 135 1003 >
88 ___________________________________________________________________ M210566EN-D
Page 91
Chapter 4 _________________________________________________________________ Operation
PLAY
Use the PLAY command to output the selected file to the serial line. Data in the output is <TAB> limited. This is compatible with most spreadsheet programs. Before giving the command, set the correct date and time with TIME and DATE commands, if needed.
PLAY [x]
where x = 0 ... 17
Example:
>play 2 RH latest 1 day 99-12-30 22:33:13 Date Time trend min max yy-mm-dd hh:mm:ss %RH %RH %RH 99-12-30 22:33:13 19.16 18.99 19.33 99-12-30 22:45:13 19.30 19.09 19.55 99-12-30 22:57:13 20.01 19.28 21.17 99-12-30 23:09:13 21.21 20.98 21.44 99-12-30 23:21:13 19.57 17.72 21.11 99-12-30 23:33:13 19.09 18.62 19.84
NOTE
The <ESC> key can be used to interrupt the output listing.
The PLAY -1 command can be used to output all files.
Output of large amounts of recorded data can take a long time. If you are using the user port, select the highest serial baud supported to reduce the time required for output.
VAISALA________________________________________________________________________ 89
Page 92
User's Guide _______________________________________________________________________

Deleting the Recorded Files

Use the keypad/display to delete the recorded files. Note that the transmitter automatically overwrites the old data when the memory gets full, so manual deletion of the recorded files is not necessary.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select System by pressing the ► arrow button.
3. Select Clear graph memories by pressing the CLEAR button.
Press the YES button to confirm the selection.
CAUTION
NOTE
This function clears all the data history from the memory, all graphs included.
DELETE/UNDELETE
Use the serial line to delete or undelete data files.
Use the DELETE command to delete all data files. Use the UNDELETE command to recover the deleted files.
The UNDELETE command will only recover the part of the deleted data that has not been recorded over yet.
90 ___________________________________________________________________ M210566EN-D
Page 93
Chapter 4 _________________________________________________________________ Operation

Analog Output Settings

The analog outputs are set in the factory according to the order form. In case you want to change the settings, follow these instructions. See section Third Analog Output on page 48.

Changing Output Mode and Range

Both output channels have their own DIP switch module with 8 switches; see the position in Figure 2 on page 18 (DIP switches for analog output settings).
1. Select the current/voltage output; switch ON either of the switches, 1 or 2.
2. Select the range; switch ON one of the switches from 3 to 7.
0503-045
Figure 44 Current/Voltage Switches of Output Modules
The following numbers refer to Figure 44 above: 1 = Current/voltage selection output switches (from 1 to 2)
2 = Current/voltage range selection switches (from 3 to 7) in
analog output 1 and 2.
3 = Switches for service use only. Keep in OFF position always.
VAISALA________________________________________________________________________ 91
Page 94
User's Guide _______________________________________________________________________
d
d
d
d
NOTE
Only one of the switches 1 or 2 must be ON at a time.
Only one of the switches 3 to 7 must be ON at a time.
Example: 0 ... 5 V voltage output selected for channel 1 and 4 ... 20 mA selected for channel 2.
OFF ON Selection
1 2
Voltage output selecte
3 4 5 6
0 ... 5 V selecte
7 8
1
Current output selecte
2 3 4
4 ... 20 mA selecte
5 6 7 8
NOTE
92 ___________________________________________________________________ M210566EN-D
If you have customized the error output setting (AERR), check that the set error values are still valid after changing the output mode/range, see section Analog Output Fault Indication Setting on page 96.
Page 95
Chapter 4 _________________________________________________________________ Operation

Analog Output Quantities

Use the display/keypad to change and scale the analog output quantities.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Interfaces by pressing the ► arrow button.
3. Select Analog outputs by pressing the arrow button.
4. Select Output 1/2/3 by pressing the arrow button.
5. Select Quantity by pressing the ▲▼ arrow buttons. Confirm your selection by pressing CHANGE.
6. Select the quantity by using the arrow buttons. Press SELECT to confirm your selection.
7. Select Scale, lower limit, by pressing the ▲▼arrow buttons. Press SET to confirm your selection. Press OK to confirm your setting.
8. Select the upper limit by pressing the ▲▼arrow buttons. Use the arrow buttons to set the upper limit value. Press SET to confirm your selection. Press OK to confirm your setting.
9. Press EXIT to return to the basic display.
AMODE/ASEL
Use the serial line to select and scale the analog output quantities. Connect the transmitter to the PC. Open the terminal connection between your PC and the transmitter.
1. Check the analog output modes with the AMODE command.
Example:
>amode Ch1 output : 0...1V Ch2 output : 0...1V >
2. Select and scale the quantities for the analog outputs with the command ASEL. Note that the optional quantities can be selected only if they have been selected when ordering the device.
VAISALA________________________________________________________________________ 93
Page 96
User's Guide _______________________________________________________________________
ASEL [xxx yyy zzz]
where xxx = Quantity of channel 1
yyy = Quantity of channel 2 zzz = Quantity of the optional analog output channel 3
Enter always all the quantities for all outputs. For quantities and their abbreviations see Table 1 on page 15 and Table 2 on page 16.
Use the command ASEL [xxx yyy] as shown in the example below when using a device with two analog outputs.
Example:
>asel rh t Ch1 (RH ) low : 0.00 %RH ? 0 Ch1 (RH ) high : 100.00 %RH ? 100 Ch2 (T ) low : -40.00 'C ? -50 Ch2 (T ) high : 60.00 'C ? 80 >
94 ___________________________________________________________________ M210566EN-D
Page 97
Chapter 4 _________________________________________________________________ Operation

Analog Output Tests

Use the display/keypad for testing to test the operation of the analog by forcing the outputs to known values. Measure then the outputs with a current/voltage meter.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select System by pressing the ► arrow button.
3. Select Diagnostics by pressing the arrow button.
4. Select Analog output tests by pressing the arrow button.
5. Select one of the testing options Force 0%/50%/100% of scale. Press TEST to confirm your selection. All outputs are tested simultaneously. The actual output value depends on the selected range.
6. Press OK to stop testing. Press EXIT to return to the basic display.
ITEST
Use the serial line to test the operation of the analog outputs. Use the command ITEST to force the analog outputs to entered values. The set values remain valid until you enter the command ITEST without parameters or RESET the transmitter.
ITEST [aa.aaa bb.bbb]
where
aa.aaa = Current or voltage value to be set for channel 1 (mA or V) bb.bbb = Current or voltage value to be set for channel 2 (mA or V)
Example:
>itest 20 5 Ch1 (Td ) : * 20.000 mA H'672A Ch2 (T ) : * 5.000 mA H'34F9 >itest Ch1 (Td ) : -23.204 'C 16.238 mA H'FFFE Ch2 (T ) : 22.889 'C 8.573 mA H'5950 >
VAISALA________________________________________________________________________ 95
Page 98
User's Guide _______________________________________________________________________

Analog Output Fault Indication Setting

Factory default state for analog outputs during error condition is 0 V/ 0 mA. Please be careful when selecting the new error value. The error state of the transmitter should not cause unexpected problems in process monitoring.
Use the display/keypad to set the analog output fault indication.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Interfaces by pressing the ► arrow button.
3. Select Analog Outputs by pressing the arrow button.
4. Select Output 1/2/3 by pressing the arrow button.
5. Select Fault indication. Press SET to confirm your selection.
Enter the fault indication value by using the arrow buttons. Press OK to confirm your setting. This value is outputted if a transmitter error occurs.
NOTE
NOTE
6. Press EXIT to return to the basic display.
AERR
Use the serial line AERR command to change the error output.
AERR
Example:
>aerr Ch1 error out : 0.000V ? 5.0 Ch2 error out : 0.000V ? 5.0 >
The error output value must be within a valid range of the output mode.
The error output value is displayed only when there are minor electrical faults such as humidity sensor damage. When there is a severe device malfunction, the error output value is not necessarily shown.
96 ___________________________________________________________________ M210566EN-D
Page 99
Chapter 4 _________________________________________________________________ Operation

Operation of Relays

Quantity for Relay Output

A relay monitors the quantity chosen for the relay output. Any of the quantities available can be chosen.

Measurement-Based Relay Output Modes

Relay Setpoints
When the measured value is in between the "above" and "below" values, the relay is passive. When choosing lower value as "above" value and higher value as "below" value, the relay is passive when the measured value is not between the setpoints. You can also set only one setpoint. See Figure 45 below for illustrative examples of the different measurement-based relay output modes.
0610-076
Figure 45 Measurement-Based Relay Output Modes
VAISALA________________________________________________________________________ 97
Page 100
User's Guide _______________________________________________________________________
Mode 4 is usually used if an alarm needs to be triggered when the measured value exceeds a safe range. The relay is active when measurement is in range, and is released if the value goes out of range or the measurement fails.
NOTE
NOTE
If the measurement of the selected quantity fails or the transmitter loses its power, the relay is released.
Hysteresis
Hysteresis function is to prevent the relay switching back and forth when the measured value is near to the setpoint values.
Relay is activated when the measured value passes the exact value of the setpoint. When returning and passing the setpoint again relay is not released before the value reaches the setpoint increased/decreased by the hysteresis value.
Hysteresis should be smaller than difference of the setpoints.
Example: When the 'active above' value is 60 %RH and the hysteresis value is 5 %RH, relay activates when the relative humidity reaches 60 %RH. As the humidity decreases, relay releases at 55 %RH.
If both setpoints are specified and "above" setpoint is lower than "below" setpoint, the hysteresis works in the opposite direction, that is, relay is released when the measured value passes the exact value of the setpoint.

Relay Indicating Transmitter Error Status

You can set a relay to follow the operation status of the device. By selecting FAULT/ONLINE STATUS for output quantity a relay changes state on the basis of the operation status as follows:
FAULT STATUS
Normal operation: relay active (C and NO outputs are closed)
Not measuring state (error state or power off): relay released (C and NC outputs are closed)
98 ___________________________________________________________________ M210566EN-D
Loading...