No part of this manual may be reproduced in any form or by any means, electronic or
mechanical (including photocopying), nor may its contents be communicated to a third
party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for
Vaisala towards the customer or end user. All legally binding commitments and
agreements are included exclusively in the applicable supply contract or Conditions of
Sale.
- Appendix A, Probe Installation Kits and Installation Examples,
presents the installation kits available for HMT330 and provides
some installation examples.
- Appendix B, Calculation Formulas, presents the equations used in
HMT330 to calculate values of dewpoint, mixing ratio, absolute
humidity and enthalpy in normal pressure.
General Safety Considerations
Throughout the manual, important safety considerations are
highlighted as follows:
WARNING
CAUTION
NOTE
Warning alerts you to a serious hazard. If you do not read and follow
instructions very carefully at this point, there is a risk of injury or
even death.
Caution warns you of a potential hazard. If you do not read and
follow instructions carefully at this point, the product could be
damaged or important data could be lost.
Note highlights important information on using the product.
Feedback
Vaisala Customer Documentation Team welcomes your comments
and suggestions on the quality and usefulness of this publication. If
you find errors or have other suggestions for improvement, please
indicate the chapter, section, and page number. You can send
comments to us by e-mail:
Chapter 1 _________________________________________________________ General Information
Product Related Safety Precautions
The Vaisala HUMICAP® Humidity and Temperature Transmitter
Series HMT330 delivered to you has been tested for safety and
approved as shipped from the factory. Note the following precautions:
WARNING
CAUTION
Ground the product, and verify outdoor installation grounding
periodically to minimize shock hazard.
Do not modify the unit. Improper modification can damage the
product or lead to malfunction.
ESD Protection
Electrostatic Discharge (ESD) can cause immediate or latent damage
to electronic circuits. Vaisala products are adequately protected
against ESD for their intended use. However, it is possible to damage
the product by delivering electrostatic discharges when touching,
removing, or inserting any objects inside the equipment housing.
To make sure you are not delivering high static voltages yourself:
- Handle ESD sensitive components on a properly grounded and
protected ESD workbench. When this is not possible, ground
yourself to the equipment chassis before touching the boards.
Ground yourself with a wrist strap and a resistive connection cord.
When neither of the above is possible, touch a conductive part of
the equipment chassis with your other hand before touching the
boards.
- Always hold the boards by the edges and avoid touching the
component contacts.
Dispose of batteries and the unit according to statutory regulations.
Do not dispose of with regular household refuse.
Trademarks
HUMICAP® is a registered trademark of Vaisala. Microsoft®,
Windows®, Windows NT®, Windows® 2000 and Windows® XP are
registered trademarks of Microsoft Corporation in the United States
and/or other countries.
License Agreement
All rights to any software are held by Vaisala or third parties. The
customer is allowed to use the software only to the extent that is
provided by the applicable supply contract or Software License
Agreement.
Chapter 1 _________________________________________________________ General Information
Warranty
Vaisala hereby represents and warrants all Products
manufactured by Vaisala and sold hereunder to be
free from defects in workmanship or material
during a period of twelve (12) months from the date
of delivery save for products for which a special
warranty is given. If any Product proves however to
be defective in workmanship or material within the
period herein provided Vaisala undertakes to the
exclusion of any other remedy to repair or at its
own option replace the defective Product or part
thereof free of charge and otherwise on the same
conditions as for the original Product or part
without extension to original warranty time.
Defective parts replaced in accordance with this
clause shall be placed at the disposal of Vaisala.
Vaisala also warrants the quality of all repair and
service works performed by its employees to
products sold by it. In case the repair or service
works should appear inadequate or faulty and
should this cause malfunction or nonfunction of the
product to which the service was performed Vaisala
shall at its free option either repair or have repaired
or replace the product in question. The working
hours used by employees of Vaisala for such repair
or replacement shall be free of charge to the client.
This service warranty shall be valid for a period of
six (6) months from the date the service measures
were completed.
This warranty does not however apply when the
defect has been caused through
a) normal wear and tear or accident;
b) misuse or other unsuitable or unauthorized use
of the Product or negligence or error in storing,
maintaining or in handling the Product or any
equipment thereof;
c) wrong installation or assembly or failure to
service the Product or otherwise follow
Vaisala's service instructions including any
repairs or installation or assembly or service
made by unauthorized personnel not approved
by Vaisala or replacements with parts not
manufactured or supplied by Vaisala;
d) modifications or changes of the Product as well
as any adding to it without Vaisala's prior
authorization;
e) other factors depending on the Customer or a
third party.
Notwithstanding the aforesaid Vaisala's liability
under this clause shall not apply to any defects
arising out of materials, designs or instructions
provided by the Customer.
This warranty is however subject to following
conditions:
a) A substantiated written claim as to any alleged
defects shall have been received by Vaisala
within thirty (30) days after the defect or fault
became known or occurred, and
b) The allegedly defective Product or part shall,
should Vaisala so require, be sent to the works
of Vaisala or to such other place as Vaisala may
indicate in writing, freight and insurance
prepaid and properly packed and labelled,
unless Vaisala agrees to inspect and repair the
Product or replace it on site.
This warranty is expressly in lieu of and excludes
all other conditions, warranties and liabilities,
express or implied, whether under law, statute or
otherwise, including without limitation any implied
warranties of merchantability or fitness for a
particular purpose and all other obligations and
liabilities of Vaisala or its representatives with
respect to any defect or deficiency applicable to or
resulting directly or indirectly from the Products
supplied hereunder, which obligations and
liabilities are hereby expressly cancelled and
waived. Vaisala's liability shall under no
circumstances exceed the invoice price of any
Product for which a warranty claim is made, nor
shall Vaisala in any circumstances be liable for lost
profits or other consequential loss whether direct or
indirect or for special damages.
This chapter introduces the features, advantages, and the product
nomenclature of HMT330.
Introduction to HMT330
Vaisala HUMICAP® Humidity and Temperature Transmitter Series
HMT330 provides reliable humidity measurement in wide range of
applications. Analog outputs can be chosen between current and
voltage signals. Alternatively, digital outputs RS-232 (standard) or
RS-422/485 (optional) can be selected. The quantities measured and
calculated by HMT330 are presented in Table 1 below. The quantities
available as an option are presented in Table 2 below.
Table 1 Quantities Measured by HMT330
Quantity Abbreviati
Relative humidity (RH) RH %RH %RH
Temperature (T) T ºC ºF
Dewpoint / Frostpoint
Temperature (T
Dewpoint Temperature (Td) TD ºC ºF
Absolute humidity (a) A g/m3 gr/ft3
Mixing ratio (x) X g/kg gr/lb
Wetbulb temperature (Tw) TW ºC ºF
Humid air volume/ dry air
volume
(by volume or by weight) (H
Water vapor pressure (Pw) PW hPa lb/in2
Water vapor saturation
pressure (P
Enthalpy (h) H kJ/kg Btu/lb
Difference of T and T
)
d/f
)
ws
(∆T) DT ºC ºF
d/f
TDF ºC ºF
H2O ppmv/ppm
O)
2
PWS hPa lb/in
Metric Unit Non Metric
Unit
ppmv/ppmw
w
2
Basic Features and Options
- Several probes for various applications
- User friendly display
- Calculated output quantities available
- Different probe mounting kits, sensor protection options and probe
cable lengths
- Transmitter mounting kits for multiple installation purposes
- Chemical purge for applications where interfering chemicals in the
measuring environment pose a risk
- Warmed probe and sensor heating for high humidity conditions
(HMT337)
- Optional modules: isolated power supply, power supply module,
RS-422/485-module, additional analog output module and relay
module
Temperature difference between the probe head and external
environment can cause a risk of condensation on the sensor. A wet
probe cannot observe the actual humidity in the ambient air. If the
condensed water is contaminated, the life span of the probe may
shorten and calibration may change.
HMT337 probe shall be used in applications where condensation can
occur due to high humidity and rapid humidity changes. The warmed
probe head is heated continuously so that its temperature is always
higher than in environment. This prevents condensation on the probe.
The power consumption of the warmed probe is slightly higher than
other probes.
When mounting with wall mounting kit the mounting plate (Vaisala
order code 214829) can be installed directly on wall or onto a standard
wall box (also US junction box). When wiring through back wall,
remove the plastic plug from the wiring hole in the transmitter before
mounting.
0503-004
Figure 5 Mounting with Wall Mounting Kit
The following numbers refer to Figure 5 above:
1 = Plastic mounting plate
2 = Mount the plate to wall with 4 screws M6 (not provided)
3 = The arched side up
4 = Fasten HMT330 to the mounting plate with 4 fixing screws
Pole Installation with Installation
Kit for Pole or Pipeline
Installation kit for pole or pipeline (Vaisala order code: 215108)
includes the metal mounting plate and 4 mounting nuts for pole
mounting. When mounting, the arrow in the metal mounting plate
must point upwards; see Figure 10 on page 25 below.
Figure 8 Vertical Pole
The following numbers refer to Figure 8 above:
1 = Fixing brackets (2 pcs) M8 (provided) for 30 ... 102 mm
A panel mounting frame with adhesive tape is available as an option
(Vaisala order code: 216038). After the HMT330 has been installed
through the panel, a frame can be used to finish off the sawed surface
of the panel.
1. Mount the HMT330 to the panel.
2. The frame is attached to the panel with an adhesive tape
attached to the frame. Remove the paper protecting the tape
before attaching.
3. Attach the frame to the panel around the HMT330 to finish off
the panel mounting as illustrated, see Figure 13 below.
A single electrical cable with screen and three to ten wires is
recommended for power and analog/serial connections. The cable
diameter should be 8 ... 11 mm. The number of cable bushings
depends on the transmitter options. See the following
recommendations for the cable bushings:
NOTE
0503-010
Figure 15 Cable Bushings
The following numbers refer to Figure 15 above:
1 = Cable for signal/powering Ø8 ... 11 mm
2 = Cable for optional module Ø8 ... 11 mm
3 = Cable for optional power supply module Ø8 ... 11 mm
When there is high electric noise level (for example, near powerful
electric motor) in the operating environment it is recommended to use
shielded cable or take care that the signal cables are separated from
other cables.
2. Cut back screen braiding or screen foil to dimension X (see
figure 3).
3. Push the domed cap nut (item 1) and the seal insert with contact
socket of the gland (item 2+3) onto the cable as shown in the
diagram.
4. Bend over the screen braiding or screen foil by about 90º (item
4).
5. Push the seal insert with the contact socket of the gland (item
2+3) up to the screen braiding or screen foil.
6. Mount lower part (item 5) on the housing.
7. Push the seal with the contact socket of the gland and (item 2+3)
flush into the lower part (item 5).
8. Screw the domed cap nut (item 1) onto the lower part (item 5).
Grounding the Transmitter Housing
In case you need to ground the transmitter housing, the grounding
connector is found inside the housing, see Figure 2 on page 18. Note
anyhow that the probe head is connected to the same potential as the
housing. Make sure that different groundings are made to the same
potential. Otherwise harmful ground currents may be generated.
If it is needed to have galvanic isolation of the power supply line from
the output signals, HMT330 can be ordered with optional output
isolation module. This module prevents harmful grounding loops.
When connecting transmitter with 8-pin connector, see section 8-Pin
Connector on page 55. When wiring the power supply module, see
section Power Supply Module on page 44.
WARNING
0506-028
Figure 17 Screw Terminal Block on Motherboard
The following numbers refer to Figure 17 above:
1 = Power supply terminals 10 ... 35 VDC, 24 VAC
2 = User port (RS-232 terminals)
3 = Analog signal terminals
Make sure that you connect only de-energized wires.
1. Open the transmitter cover by taking out the four cover screws.
2. Insert the power supply wires and signal wires through the cable
bushing in the bottom of the transmitter; see the grounding
instructions in the previous sections.
3. Connect the analog output cables to terminals: Ch1 +, Ch1-,
Ch2+, Ch2-. Connect the RS-232 user port cables to terminals
RxD, GND and TxD. For more information about the RS-232
connection refer to section Serial Line Communication on page
4. When wiring RS-485 module, relay module or additional analog
output module, see section RS-422/485 Interface on page 52,
section Relays on page 50, and section Third Analog Output on
page 48.
5. Connect the power supply wires to the connectors: POWER 10 ... 35V+ 24V~ (+) and (-) terminals. If you are using 24 VAC
power supply, see the note below before connecting the supply
wires.
6. Turn on the power. The indicator led on the cover lit
continuously during normal operation.
7. Close the cover and replace the cover screws. The transmitter is
ready for use.
Connections to 24 VAC Power
Supply
CAUTION
Separate floating supply for each transmitter is recommended (see
upper part of Figure 18 on page 33). If you have to connect several
transmitters or other instruments to one AC supply, the phase (~) must
always be connected to (+) connector of each transmitter (see lower
part of Figure 18).
24 VAC POWER SUPPLY USE
To prevent fire and/or damage, if either 24 VAC wire is grounded or
connected to a "-", "0", or "GND" terminal of any other device,
you must connect the same wire on the "-" terminal also on this
instrument.
In humidity measurement and especially in calibration it is essential
that temperature of the probe and measuring environment is the same.
Even a small difference in temperature between the environment and
the probe causes an error. As the curve below shows, if the
temperature is +20 °C and the relative humidity 100 %RH, a
difference of ±1 °C between the environment and the probe causes an
error of ±6 %RH.
The graph below illustrates the measurement error at 100 %RH when
the difference between the ambient and sensor temperature is 1 °C.
When there is no alternative but to install the sensor head in the
process vertically, the point of entry must be carefully insulated. The
cable must also be allowed to hang loosely as this prevents any
condensed water from running onto the sensor head along the cable.
0507-022
NOTE
Figure 21 Vertical Mounting of Sensor Head
The following numbers refer to Figure 21 above:
1 = To be sealed.
2 = Insulate the cable.
3 = To be insulated.
4 = Let the cable hang loosely. This prevents condensed water
running to the sensor along the cable.
Please do not attach a heated probe (HMT337) to metal structures to
avoid condensation problems caused by heat conduction along the
metal.
If the process temperature is much higher than that of the
environment, the whole sensor head and preferably plenty of cable
must be inside the process. This prevents measuring inaccuracy
caused by heat conduction along the cable.
When mounted on the side of a duct or channel, the sensor head must
be inserted from the side of the duct. If this is not possible and the
sensor head must be inserted from the top, the point of entry must be
carefully insulated.
For Vaisala probe installation kits and some installation examples see
Appendix A on page 145.
The HMT333 is a small size (ø = 12mm) general-purpose probe
suitable for ducts and channels with the installation kit available from
Vaisala.
The HMT333 provides for two measuring range options. The first
probe version is equipped with a flexible cable and can be used when
measuring in environments up to 80 ºC. The second version is suitable
for measuring in environments up to 120 ºC.
See Appendix A on page 145 for the following probe installation kits
for HMT333 and installation examples.
- Duct mounting kit
- Cable gland.
HMT334 for High Pressure and
Vacuum Applications
The HMT334 probe is for the dewpoint measurements in pressurized
rooms and industrial processes. The probe is provided with a nut, a
fitting screw and a sealing washer. Keep the fitting screw and the nut
in place on the body of the sensor head during handling to prevent
damage to the highly polished surface of the probe. Follow the
instructions below to achieve a leak-tight assembly:
1. Unscrew the fitting screw from the nut and the sensor head.
2. Fasten the fitting screw to the chamber wall with a sealing
washer. Tighten the fitting screw into the threaded sleeve with a
torque spanner. The tightening torque is 150 ± 10 Nm (110 ± 7
ft-lbs).
3. Insert the body of the sensor head into the fitting screw and
screw the nut manually to the fitting screw until the connection
feels tight.
5. Tighten the nut a further 30º (1/12) turn or if you have a torque
spanner tighten it with a torque of 80 ± 10 Nm (60 ± 7 ft-lbs).
0503-034
Figure 23 Tightening the Nut
NOTE
When re-tightening the nut after detachment the nut must be
tightened without increased effort.
6. Clean and grease the tightening cone of the fitting screw after
every tenth detachment. Change the sealing washer every time
the fitting screw is detached. Use high-vacuum grease (for
example Dow Corning) or similar grease.
In pressurized processes it is essential to tighten the supporting nuts
and screws very carefully to prevent loosening of the probe by the
action of pressure.
When the HMT330 is installed in a process with a pressure differing
from normal atmospheric pressure, please enter the pressure value of
the process (in hPa or mbar) into the transmitter memory, see
command PRES and XPRES on page 77 or use display/keypad.
HMT335 is installed similarly than the HMT333 probe but without the
supporting bar. Refer to Appendix A on page 145 for more
information on the duct installation kit for HMT335.
To avoid incorrect humidity readings, the temperature differences
between inside and outside of the duct must not be remarkable.
HMT337 for High Humidity
Applications
The HMT337 is for environment where relative humidity is very high,
near saturation. The warmed probe head prevents the saturation of the
sensor. An additional temperature probe is also available.
See Appendix A on page 145 for a presentation of the following probe
installation kits for HMT337 with installation examples:
- Duct mounting kit
- Cable gland
- Pressure tight Swagelok connector
- Vaisala's Meteorological Installation kit
The installation kits are available for both humidity and temperature
probe.
Temperature Probe (Optional)
An external temperature probe for HMT337 is needed for the
measurement of the ambient temperature in case you have a
transmitter with a warmed probe. This allows you to measure other
humidity quantities apart from dewpoint and mixing ratio.
The temperature probe is connected always in factory.
Due to the sliding fit the HMT338 is easy to install into and remove
from the pressurized process. The probe is especially suitable for the
measurements in pipelines. See section Ball Valve Installation Kit for
HMT338 on page 152.
0508-077
Figure 25 HMT338 Probe
The following numbers refer to Figure 25 above:
1 = Clasp nut, 24 mm hex nut
2 = Fitting body, 27 mm hex head
The following two fitting body options are available:
- Fitting Body ISO1/2 solid structure
- Fitting Body NPT1/2 solid structure
Table 3 HMT338 Probe Dimensions
Probe type Probe Dimension Adjustment Range
1. Adjust the probe to a suitable depth according to the type of
installation.
2. Tighten the clasp nut first manually.
3. Mark the fitting screw and the clasp nut.
4. Tighten the nut a further 50 -60º (ca. 1/6 turn) with a wrench. If
you have suitable torque spanner, tighten the nut to max 45 ± 5
Nm (33 ± 4 ft-lbs).
0503-032
Figure 27 Tightening the Clasp Nut
The following numbers refer to Figure 27 above:
1 = Probe
Take care not to over tighten the clasp nut to avoid difficulties when
opening it.
Take care not to damage the probe body. A damaged body makes the
probe head less tight and may prevent it from going through the clasp
nut.
In pressurized processes it is essential to tighten the supporting nuts
and screws very carefully to prevent loosening of the probe by the
action of pressure.
When installed in a process with a pressure differing from normal
atmospheric pressure, please enter the pressure value of the process
(in hPa or mbar) into the transmitter memory via the serial line (see
command PRES and XPRES on page 77) or by using
display/keypad.
The AC (mains) power connection may be connected to the power
supply module only by an authorized electrician. A readily accessible
disconnect device shall be incorporated in the fixed wiring.
0506-027
Figure 28 Power Supply Module
The following numbers refer to Figure 28 above:
1 = Connect AC (mains) voltage wires to these terminals
2 = Grounding terminal
3 = In case the module is not installed in the factory: Connect
wires from these terminals to the POWER 10 ... 36V 24V
1. Disconnect the power and open the transmitter cover.
2. Remove the protective plug from the cable gland and thread the
wires. In case the power supply module is installed in the
factory, continue with the step 5.
3. To attach the module fasten the power module to the bottom of
the housing with four screws. See the position Figure 2 on page
18.
4. Connect the wires from the terminals of the power supply
module marked with + and - to the terminals POWER 10 ... 35 V 24V on the motherboard of the transmitter.
5. Connect the AC mains voltage wires to the power supply
module terminals marked with N and L.
6. Attach the grounding wire to the grounding terminal on the
right-hand side of the transmitter.
WARNING
WARNING
WARNING
7. Connect the power. The LED on the cover of the transmitter is
lit continuously during normal operation.
Do not detach the power supply module from the transmitter when
the power is on.
Do not connect the mains power to power supply module when it is
not installed in the transmitter.
If galvanic isolation of the power supply line from the output signals
is needed, HMT330 can be ordered with optional output isolation
module. This module prevents harmful grounding loops.
NOTE
Output isolation module is not needed when using the power supply
module.
0506-030
Figure 29 Galvanic Output Isolation Module
The following number refers to Figure 29 above:
1 = Output isolation module
Third Analog Output
0503-030
Figure 30 Third Analog Output
The following numbers refer to Figure 29 above:
1 = Flat cable pins
2 = Screw terminals for signal line
3 = DIP switches to select the output mode and range
1. Disconnect the power. In case the analog output module is
installed in the factory, continue with the step 4.
2. To attach the module, open the transmitter cover and fasten the
analog output module to the bottom of the housing with four
screws. See the position from the picture on page 11.
3. Connect the flat cable between the analog output module and the
motherboard's pins MODULE 2.
4. Take out the protective plug from the cable gland and thread the
wires.
5. Connect the wires to the screw terminals marked with Ch+ and
Ch- .
6. Select the current/voltage output by setting ON either of the
switches 1 or 2.
7. Select the range by setting ON one of the switches 3 ... 7.
NOTE
Only one of the switches 1 and 2 can be ON at a time.
Only one of the switches 3 ... 7 can be ON at a time.
8. Connect the power.
9. Select the quantity and scale the channel via the serial line or
display/keypad, see section Analog Output Quantities on page
93. For testing the analog output, see Section Analog Output
Tests on page 95. For fault indication setting, see section Analog
Output Fault Indication Setting on page 96.
HMT330 can be equipped with one or two configurable relay
modules. Each module contains two configurable relays. See the
contact ratings in section Technical Specifications of Optional
Modules on page 135.
Installation and Wiring
1. Disconnect the power and open the transmitter cover. In case the
relay-module is installed in the factory, continue with step 5.
2. To attach the module fasten the relay module to the bottom of
the housing with four screws. See the position in
page 18.
3. When the mains power is in use attach the grounding wire to the
grounding terminal.
Figure 2 on
4. Connect the flat cable between the relay module and the
MODULE 1 or MODULE 2 pins of the motherboard.
5. Take out the protective plug from the cable gland and thread the
relay wires.
6. Connect the wires to the screw terminals: NO, C, NC.
Selecting the Activation State of the Relay
The middlemost C terminal and either one of the terminals NO/NC
must be connected. The polarity can be freely selected.
NO Normally open
C Common relay
NC Normally closed
Relay NOT activated: C and NC outputs are closed, NO is open
Relay IS activated: C and NO outputs are closed, NC is open.
Connect the power and close the cover.
For instructions on how to operate the relay (for example, select
quantity for the relay output and set the relay setpoints) see section
Operation of Relays on page 97.
1 = Indication led for the relay 1 or 3
2 = Relay test buttons
3 = Flat cable pins
4 = Indication led for relay 2 or 4
The relay module may contain dangerous voltages even if the
transmitter power has been disconnected. Before opening the
transmitter you must switch off both the transmitter and the voltage
connected to the relay terminals.
Do not connect the mains power to relay unit without grounding the
transmitter.
6. If you use RS-485 (or RS-422) to connect just one HMT330 to a
master computer, enable the internal termination of HMT330 by
switching switches 1 and 2 ON. Make sure that the master's end
of the line is also terminated (by using master's internal
termination or with a separate terminator).
If you are connecting many transmitters to the same RS-485 bus,
make sure that switches 1 and 2 are OFF and terminate the bus
with separate terminators at both ends. This allows removing
any transmitter without blocking the bus operation.
NOTE
If you use the internal termination of the transmitter at the end of the
RS-485 bus (instead of using separate terminators) removing that
transmitter will block the bus operation.
7. Use the bus type (4-wire/2-wire) to select the selection switch 3.
In 4-wire mode RS-485 master sends data to the HMT330
through terminals RxA and RxB and receives data from
HMT330 through terminals TxA and TxB.
This chapter contains information that is needed to operate this
product.
Getting Started
Within a few seconds after power-up the led on the cover of the
transmitter is lit continuously indicating normal operation. When
using the optional display and turning the transmitter on the first time,
the language selection menu window opens. Select the language with
arrow buttons and press the SELECT button.
The pressure has an effect on humidity calculations and accuracy.
Therefore, accurate calculations can be achieved only when the
ambient pressure is taken into consideration. For instructions on how
to set the pressure, see section Pressure Compensation Setting on page
77.
Display/Keypad (Optional)
Basic Display
Display shows you the measurement values of the selected quantities
in the selected units. You can select 1 ... 3 quantities for the numerical
basic display (see section Changing Quantities and Units on page 74.)
The following numbers refer to Figure 35 above:
1 = The Info shortcut button, see section Device Information on
page 82 below
2 = The Graph shortcut button, see section Graphic History on
page 58 below
3 = Quantities selected for display
From any view, a four-second press on the right-hand function button
takes you directly to the basic display.
Graphic History
The graphical display shows the data trend of the selected quantities,
one at a time. The graph is updated automatically while measuring.
Use the following functions in the graphical display:
- Press the NEXT button to have the trend graph and max/min graph
in turns and browse through the quantities selected for display.
- Press the EXIT button to get back the basic display.
0507-001
Figure 36 Graphical Display
Trend graph: Shows you a curve of average values. Each value is a
calculated average over a period. See Table 8 on page 59 below.
Max/min graph: Shows you the minimum and maximum values in a
form of curve. Each value is max/min over a time period. See Table 8
below.
Table 8 Periods for Trend and Max/Min Calculations
Observation Period Period for Trend/Max/Min
Calculations (Resolution)
20 minutes 10 seconds
3 hours 90 seconds
1 day 12 minutes
10 days 2 hours
2 months 12 hours
1 year 3 days
- Press the arrow buttons to zoom in and out the time in the graph
window.
- Press the arrow buttons to have the cursor mode where you can
observe individual measuring points. Press an arrow button to
move the cursor (vertical bar) along the time axis. The numerical
value at the cursor position is shown at the left upper corner. Time
from the present to the chosen moment is shown at the right upper
corner.
Table 9 Graph Information Messages in Cursor Mode
Message Interpretation
Power outage Power failure (marked also with dashed
vertical line)
No data Quantity has not been selected for the display
Device failure General device failure
T meas. failure Temperature measurement/sensor failure
RH meas. failure Humidity measurement/sensor failure
Adj. mode active Adjustment mode active (data recorded in the
adjustment mode is not displayed)
A question mark after time tells you that at least one power failure
(dashed vertical line) has occurred after the chosen moment. In this
case, the exact time difference between the present and the cursor
position is not exactly known.
Menus and Navigation
You can change settings and select functions in the menus.
1. Open the MAIN MENU by pressing any of the arrow
Round off one decimal by using the Rounding function. The default
setting is rounding on. Rounding has no effect on quantities without
decimals.
1. Open the MAIN MENU by pressing any of the arrow
buttons.
2. Select Display and confirm by pressing the arrow button.
3. Select Rounding and press ON/OFF button.
4. Press EXIT to return to the basic display.
Display Backlight Setting
As a default the display backlight is always on. In the automatic mode
the backlight stays on for 30 seconds from the last press of any button.
When pressing any button, the light turns on again.
1. Open the MAIN MENU by pressing any of the arrow
buttons.
2. Select Display, press the arrow button.
3. Select Backlight, press the CHANGE button.
4. Select On/Off/Automatic, press the SELECT button.
5. Press EXIT to return to the basic display.
Display Contrast Setting
1. Open the MAIN MENU by pressing any of the arrow
buttons.
2. Select Display, press the arrow button.
3. Select Contrast, press the ADJUST button.
4. Adjust the contrast by pressing the arrow buttons.
5. Press OK and EXIT to return to the basic display.
This function locks the keypad and prevents unintentional key presses.
1. Keep pressing the left-hand function button for 4 seconds to
lock the keypad (at any display).
2. To unlock the keypad, press the OPEN button for 4 seconds.
Menu PIN Lock
You can prevent unauthorized changes of the device settings by
activating the menu PIN lock. When this function is activated, the
basic display and graphical view are available but access to the menus
is locked. The key symbolindicates the activation of this feature.
1. Open the MAIN MENU by pressing any of the arrow
buttons.
2. Select System, press the arrow button.
NOTE
3. Select Menu PIN, press the ON button.
4. Enter a PIN code by using the arrow buttons. Press OK to
confirm the setting. Now the PIN lock is on and a key symbol is
shown in a display.
5. Press EXIT to return to the basic display. Returning to the menu
is possible only by entering the correct PIN code.
When you want to turn off the PIN lock, go to the menu by giving the
PIN code and select System, Menu PIN, press OFF button.
In case you have forgotten the PIN code, open the transmitter cover
and press the ADJ button once. Wait for a few seconds and the
adjustment menu opens. Select Clear menu PIN, press CLEAR.
You can also disable the keypad completely with serial command
LOCK.
Use the display/keypad to restore the factory settings. This operation
does not affect the adjustments. Only settings available in the menus
are restored.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select System by pressing the ► arrow button.
3. Select Factory settings and press the REVERT button to
confirm your selection. Press the YES button to reset all settings
to the factory defaults.
See section General Settings on page 74 for a description of the other
menu options.
MI70 Link Program for Data Handling
The recorded data can be transferred to a PC by using MI70 Link
program. You can examine the recorded data easily in Windows
environment and transfer it further to a spreadsheet program (such as
Microsoft Excel) or virtually to any Windows program in numeric or
graphical format. MI70 Link program allows you also to monitor
transmitter readings directly with a PC (real-time window function).
MI70 Link program is available from Vaisala, see list of accessories in
section Options and Accessories on page 136.
1. Connect the connection cable between the serial port of your PC
and the Service Port of HMT330; see Figure 38 on page 64
below.
2. Check that the HMT330 is powered and start using the MI70
Link program.
Use a MI70 Link version 1.07, or a newer one, to be able to utilize all
the functions of HMT330.
Serial Line Communication
Connect the serial interface by using either the user port or the service
port.
For permanent interfacing to host system, use the user port. You can
change the serial settings and operate in RUN, STOP and POLL
modes.
The following instructions show a connection example with
HyperTerminal program (included in the Microsoft Windows).
Follow the instructions below to open a HyperTerminal program:
1. Start HyperTerminal. To get help for starting HyperTerminal,
click "Start", select "Help" to open Windows help, and search
for "HyperTerminal".
0507-027
Figure 40 Starting Hyper Terminal Connection
2. In the "New Connection" window of the HyperTerminal, define
a name for HMT330 serial connection, for example "HMT330".
Click OK.
3. In the "Connect using" pull down menu, select the PC
communications port where the serial cable is connected. (If
your computer has only one COM port, it is called "COM1").
Click OK.
4. Set the port settings in the "Properties" window to match the
settings of your HMT330 user port/service port. For HMT330,
"Flow control" must always be set to "None". Finally click OK
to start using the serial connection.
0503-015
Figure 42 Hyper Terminal Serial Port Settings
5. Select "File" Æ "Save" in the HyperTerminal main window to
save the serial port settings. To use the saved settings later, start
HyperTerminal, click cancel in the "New Connection" window,
and select "File" Æ "Open".
Enter the R command to start output of measurements. Enter the S
command, press the Esc button or reset the transmitter to stop
outputting. See command SMODE to change the default (power-up)
operation mode.
Format the output by using the following commands:
- Outputting interval can be changed with the INTV command.
- Output message format can be changed with a command FORM.
- Status of chemical purge and probe heating can be added with a
command FST.
- Date and time information can be added with commands FDATE
If a value is too long to fit to the allocated space in the output, or if
there is an error in outputting the quantity, the value is displayed with
stars ‘*’.
Example:
RH=***.* %RH T= 31.0 'C
The output mode can be changed with the commands: FORM, FST,
FDATE, and TIME.
Outputting Reading with Raw Data
SEND D
Example:
>send d
24.1720 15.0399 -3.5743 189.2324 15.0709 15.0399
23.9765
Where the readings (from the left) are:
24.1720 = Temperature of the humidity probe ( ºC)
15.0399 = RH (%RH)
-3.5743 = Tdf (C)
189.2324 = Capacitance (pF)
15.0709 = RH raw: calculated from scaled capacitance (%RH)
15.0399 = Enhancement factor corrected RH (%RH)
23.9765 = Temperature of the additional temperature probe
For more information on chemical purge, see section Chemical Purge
(Optional) on page 108.
Where the state of the probe is indicated by the following letters and values:
... xxx = Normal operation where h xxx = Probe heat power
X ... xxx = Sensor heating where xxx = Sensor temperature (ºC)
H ... xxx = Chemical purge where xxx = Sensor temperature (ºC)
S ... xxx = Sensor cooling
To change quantities and units use serial commands or the optional
display/keypad. See Table 1 on page 15 for available quantities and
Table 2 on page 16 for optional quantities.
NOTE
Only the quantities selected when ordering the device can be selected
as a display output quantity.
Using Display/Keypad
Use display/keypad to select the display output quantities.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Press the ► arrow button to select Display.
3. Press ► arrow button to select Quantities.
4. Select the quantity by using the ▲▼ arrow buttons. Confirm the
selection by pressing SELECT. You can select 1 ... 3 display
quantities at a time.
5. Press EXIT to return to the basic display.
To select display units:
1. Press any of the arrow buttons to open the MAIN MENU.
2. Press the ► arrow button to select Display.
3. Use the ▲▼arrow buttons to select Units. Confirm the selection
by pressing the right-hand arrow button.
4. Use the ▲▼ arrow buttons to select display units. Confirm the
selection by pressing CHANGE. The unit changes from metric
to non-metric or the other way round.
Use the serial line command FORM to change the format or select a
certain quantities for the output commands SEND and R. Use the
serial line command UNIT to select metric or non-metric output units.
FORM
Use the serial line command FORM to change the format or select a
certain quantities for the output commands SEND and R.
FORM [x]
where
x = Formatter string
Formatter string consists of quantities and modifiers.
When entering the command, use the abbreviations of the quantities.
For more information on quantities, see Table 1 and Table 2 on page
15.
The modifiers are presented in Table 12 below.
Table 12 FORM Command Modifiers
Modifier Description
x.y Length modifier (number of digits and decimal places)
#t Tabulator
#r Carriage-return
#n Line feed
“" String constant
#xxx Special character, code "xxx" (decimal), for example
#027 for ESC
U5 Unit field and length
ADDR Transmitter address with two characters [00...99]
ERR Error flags for P, T, Ta, RH [0000 ... 1111], 0 = no error
STAT Transmitter status in 7 character field, for example:
N 0 no heating
h 115 probe heating active, power 115/255
H 159.0 purge heating active, temperature 159ºC
S 115.0 purge cooling active, temperature 115ºC
X 95.0 sensor heating active, temperature 95ºC
SN Transmitter serial number
TIME Time [hh:mm:ss]
DATE Date [yyyy-mm-dd]
OK Pressure stability indicator, two characters [OK or " "]
CS2 Modulus-256 checksum of message sent so far, ascii
Command ‘FORM /’ will return the default output format. The
default output format depends on the device configuration.
NOTE
>form /
>send
RH= 98.4 %RH T= 31.1 'C
>
UNIT
Use the UNIT command to select metric or non-metric output units:
UNIT [x]
where
x = M or N
where
M = Metric units
N = Non-metric units
This command changes both the serial output and display units to
either metric or non-metric units. When you want to output both
metric and non-metric units simultaneously on the display, select the
display units later by using the display/keypad.
The pressure has an effect on humidity calculations and accuracy.
Therefore, accurate calculations can be achieved only when the
process pressure is taken into consideration.
Note that conversions from mmHg and inHg are defined at 0°C and
for mmH2O and inH2O at 4°C.
NOTE
Pressure compensation is intended to be used in normal air only.
When measuring in other gases, please contact Vaisala for further
information.
Using Display/Keypad
Use display/keypad to set the pressure compensation. To select the
pressure unit using display/keypad, see section Changing Quantities
and Units on page 74.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Measuring and press the ►arrow button to confirm your
selection.
3. Select Pressure compensation and press the ►arrow button to
confirm you selection.
4. Press SET and enter the pressure value in the chosen unit by
using the arrow buttons.
5. Press OK and EXIT to return to the basic display.
Using Serial Line
PRES and XPRES
Command XPRES should be used if the value is changed frequently.
Its value is not retained at reset, and when set to 0; last value set with
PRES is used instead. Use the serial line and do the following:
PRES [aaaa.a]
mbar 1
Pa N/m2 0.01
mmHg torr 1.333224
inHg 33.86388
mmH2O 0.09806650
inH2O 2.490889
atm 1013.25
at 980.665
bar 1000
psia 1) 68.94757
1) psia = psi absolute.
Example:
29.9213 inHg = 29.9213 × 33.86388 = 1013.25 hPa
User Port Serial Settings
Using Display/Keypad
The communication settings for the user port can be changed via the
serial line or by using the optional display/keypad. The
communication settings for the service port are fixed and not
changeable.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Interfaces and press the ►arrow button to confirm your
selection.
3. Select Serial interface and press the ►arrow button to confirm
your selection.
4. Select Bit rate/Serial format/Comm. mode by pressing the CHANGE button. Use the ▲▼ arrow buttons to select and
press SELECT to confirm your selection.
5. If you selected RUN communication mode, select RUN interval
for and press SET to confirm your selection.
Use the command ECHO to set the user port echo. The command
either enables or disables echo of characters received.
ECHO [x]
where
x = ON (default) or
= OFF
NOTE
You can use the SERI, SMODE, INTV and ECHO commands to
change/view the user port settings even if you are currently connected
to the service port.
Data Filtering
The averaging data filter calculates an average over a certain period of
time. The lowest measurement noise is achieved with the extended
filtering. There are three filtering levels available.
Table 15 Filtering Levels
Setting Filtering level
OFF No filtering
ON (default) Standard = short filtering (approximately 15 s moving
average)
EXTENDED Extended filtering (default: approximately 1 min
average)
Use display/keypad to set the filtering level.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Measuring by pressing the ► arrow button.
3. Select Filtering and press CHANGE to confirm your selection.
4. Select Off/Standard/Extended and press SELECT to confirm
Use the serial line command FILT [xxx] to set the filtering level.
FILT [xxx]
where
xxx = OFF, ON or EXT (default = ON)
Device Information
Use the display/keypad or the serial line to display the device
information.
Press the INFO button in the basic display to see the following
information:
- Current sensor operation (for example, chemical purge), if any, in
progress
- Present or past unacknowledged errors, if any
- Device information
- Adjustment information fed by the user
- Measuring settings
- Information on chemical purge settings (when applicable)
- Serial interface information
- Analog output information
- Relay output information (when applicable)
0507-030
Figure 43 Device Information on Display
Proceed in the information views by pressing the MORE button as
many times as you get the desired information. You can browse
through the information displays also with arrow buttons. Press OK to
return to the basic display.
Use the serial line command ? to check the current transmitter
configuration. Command ?? is similar but can also be used if the
transmitter is in POLL mode.
Example:
>?
HMT330 / 2.02
Serial number : Z4140007
Batch number : Z4020017
Adjust. date : 2005-06-20
Adjust. info : VAISALA/HEL
Date : 2000-02-12
Time : 22:58:25
Serial mode : STOP
Baud P D S : 4800 E 7 1
Output interval: 0 s
Address : 0
Echo : ON
Pressure : 1013.25 hPa
Filter : OFF
Ch1 output : 4...20mA
Ch2 output : 4...20mA
Ch1 RH low : 0.00 %RH
Ch1 RH high : 100.00 %RH
Ch2 T low : -40.00 'C
Ch2 T high : 60.00 'C
Module 1 : not installed
Module 2 : not installed
>
HELP
Use the command HELP to list the commands.
Example:
>help
? ACAL ADDR AERR ASCL
ASEL CDATE CLOSE CODE CRH
CT CTA CTEXT DATE DELETE
DIR DSEL DSEND ECHO ERRS
FCRH FDATE FILT FORM FST
FTIME HELP INTV ITEST MODS
OPEN PLAY PRES R RESET
SEND SERI SMODE TEST TIME
UNDELETE UNIT VERS XPRES
>
Data recording function is always on and collects data automatically
into the memory of the device. Recorded data do not disappear from
the memory when the power is switched off. Collected data can be
observed in a form of a graph in the graphical view of the display or it
can be listed out by using the serial line or MI70 Link program.
Selecting Data Recording
Quantities
If the device is provided with the optional display, the recorded
quantities are always those selected for the display. Up to three
quantities can be recorded at a time. For instructions on how to select
the display quantities with the keypad, see section Changing
Quantities and Units on page 74.
DSEL
Use the serial line command DSEL to select the quantities to be
recorded if the transmitter is not equipped with display/keypad.
DSEL [xxx]
where
xxx = Data recording quantity. See Table 1 on page 15 and Table 2
on page 16 for the quantities.
Example:
>dsel rh t tdf
RH T Tdf
>
Enter the command without parameters and press ENTER to display
current recording parameters.
If the device is provided with the optional display, the graphical
display shows the data of the selected quantities, one at a time. See
section Graphic History on page 58 for details about graphical display.
You may also dump the logged data to the serial line in numeric form
with the following commands.
DIR
Use the serial line and enter the DIR command to check the available
files.
The device records six files (six observation periods) for each selected
quantity. Thus, total amount of the files depends on the amount of the
selected quantities being at minimum 6 and at maximum 18. See
Table 8 on page 59.
Select, for example, three quantities (RH, T, and Tdf). The last two
columns illustrate software information that is not essential for the
user.
Use the PLAY command to output the selected file to the serial line.
Data in the output is <TAB> limited. This is compatible with most
spreadsheet programs. Before giving the command, set the correct
date and time with TIME and DATE commands, if needed.
PLAY [x]
where
x = 0 ... 17
Example:
>play 2
RH latest 1 day 99-12-30 22:33:13
Date Time trend min max
yy-mm-dd hh:mm:ss %RH %RH %RH
99-12-30 22:33:13 19.16 18.99 19.33
99-12-30 22:45:13 19.30 19.09 19.55
99-12-30 22:57:13 20.01 19.28 21.17
99-12-30 23:09:13 21.21 20.98 21.44
99-12-30 23:21:13 19.57 17.72 21.11
99-12-30 23:33:13 19.09 18.62 19.84
NOTE
The <ESC> key can be used to interrupt the output listing.
The PLAY -1 command can be used to output all files.
Output of large amounts of recorded data can take a long time. If you
are using the user port, select the highest serial baud supported to
reduce the time required for output.
Use the keypad/display to delete the recorded files. Note that the
transmitter automatically overwrites the old data when the memory
gets full, so manual deletion of the recorded files is not necessary.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select System by pressing the ► arrow button.
3. Select Clear graph memories by pressing the CLEAR button.
Press the YES button to confirm the selection.
CAUTION
NOTE
This function clears all the data history from the memory, all graphs
included.
DELETE/UNDELETE
Use the serial line to delete or undelete data files.
Use the DELETE command to delete all data files. Use the
UNDELETE command to recover the deleted files.
The UNDELETE command will only recover the part of the deleted
data that has not been recorded over yet.
The analog outputs are set in the factory according to the order form.
In case you want to change the settings, follow these instructions. See
section Third Analog Output on page 48.
Changing Output Mode and Range
Both output channels have their own DIP switch module with 8
switches; see the position in Figure 2 on page 18 (DIP switches for
analog output settings).
1. Select the current/voltage output; switch ON either of the
switches, 1 or 2.
2. Select the range; switch ON one of the switches from 3 to 7.
0503-045
Figure 44 Current/Voltage Switches of Output Modules
The following numbers refer to Figure 44 above:
1 = Current/voltage selection output switches (from 1 to 2)
2 = Current/voltage range selection switches (from 3 to 7) in
analog output 1 and 2.
3 = Switches for service use only. Keep in OFF position always.
If you have customized the error output setting (AERR), check that
the set error values are still valid after changing the output
mode/range, see section Analog Output Fault Indication Setting on
page 96.
Use the display/keypad to change and scale the analog output
quantities.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Interfaces by pressing the ► arrow button.
3. Select Analog outputs by pressing the ► arrow button.
4. Select Output 1/2/3 by pressing the ► arrow button.
5. Select Quantity by pressing the ▲▼ arrow buttons. Confirm
your selection by pressing CHANGE.
6. Select the quantity by using the arrow buttons. Press SELECT
to confirm your selection.
7. Select Scale, lower limit, by pressing the ▲▼arrow buttons.
Press SET to confirm your selection. Press OK to confirm your
setting.
8. Select the upper limit by pressing the ▲▼arrow buttons. Use
the arrow buttons to set the upper limit value. Press SET to
confirm your selection. Press OK to confirm your setting.
9. Press EXIT to return to the basic display.
AMODE/ASEL
Use the serial line to select and scale the analog output quantities.
Connect the transmitter to the PC. Open the terminal connection
between your PC and the transmitter.
1. Check the analog output modes with the AMODE command.
Example:
>amode
Ch1 output : 0...1V
Ch2 output : 0...1V
>
2. Select and scale the quantities for the analog outputs with the
command ASEL. Note that the optional quantities can be
selected only if they have been selected when ordering the
device.
Use the display/keypad for testing to test the operation of the analog
by forcing the outputs to known values. Measure then the outputs with
a current/voltage meter.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select System by pressing the ► arrow button.
3. Select Diagnostics by pressing the ► arrow button.
4. Select Analog output tests by pressing the ► arrow button.
5. Select one of the testing options Force 0%/50%/100% of scale.
Press TEST to confirm your selection. All outputs are tested
simultaneously. The actual output value depends on the selected
range.
6. Press OK to stop testing. Press EXIT to return to the basic
display.
ITEST
Use the serial line to test the operation of the analog outputs. Use the
command ITEST to force the analog outputs to entered values. The
set values remain valid until you enter the command ITEST without
parameters or RESET the transmitter.
ITEST [aa.aaa bb.bbb]
where
aa.aaa = Current or voltage value to be set for channel 1 (mA or V)
bb.bbb = Current or voltage value to be set for channel 2 (mA or V)
Example:
>itest 20 5
Ch1 (Td ) : * 20.000 mA H'672A
Ch2 (T ) : * 5.000 mA H'34F9
>itest
Ch1 (Td ) : -23.204 'C 16.238 mA H'FFFE
Ch2 (T ) : 22.889 'C 8.573 mA H'5950
>
Factory default state for analog outputs during error condition is 0 V/
0 mA. Please be careful when selecting the new error value. The error
state of the transmitter should not cause unexpected problems in
process monitoring.
Use the display/keypad to set the analog output fault indication.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Interfaces by pressing the ► arrow button.
3. Select Analog Outputs by pressing the ► arrow button.
4. Select Output 1/2/3 by pressing the ► arrow button.
5. Select Fault indication. Press SET to confirm your selection.
Enter the fault indication value by using the arrow buttons. Press
OK to confirm your setting. This value is outputted if a
transmitter error occurs.
NOTE
NOTE
6. Press EXIT to return to the basic display.
AERR
Use the serial line AERR command to change the error output.
AERR
Example:
>aerr
Ch1 error out : 0.000V ? 5.0
Ch2 error out : 0.000V ? 5.0
>
The error output value must be within a valid range of the output
mode.
The error output value is displayed only when there are minor
electrical faults such as humidity sensor damage. When there is a
severe device malfunction, the error output value is not necessarily
shown.
A relay monitors the quantity chosen for the relay output. Any of the
quantities available can be chosen.
Measurement-Based Relay Output
Modes
Relay Setpoints
When the measured value is in between the "above" and "below"
values, the relay is passive. When choosing lower value as "above"
value and higher value as "below" value, the relay is passive when the
measured value is not between the setpoints. You can also set only
one setpoint. See Figure 45 below for illustrative examples of the
different measurement-based relay output modes.
Mode 4 is usually used if an alarm needs to be triggered when the
measured value exceeds a safe range. The relay is active when
measurement is in range, and is released if the value goes out of range
or the measurement fails.
NOTE
NOTE
If the measurement of the selected quantity fails or the transmitter
loses its power, the relay is released.
Hysteresis
Hysteresis function is to prevent the relay switching back and forth
when the measured value is near to the setpoint values.
Relay is activated when the measured value passes the exact value of
the setpoint. When returning and passing the setpoint again relay is
not released before the value reaches the setpoint increased/decreased
by the hysteresis value.
Hysteresis should be smaller than difference of the setpoints.
Example: When the 'active above' value is 60 %RH and the hysteresis
value is 5 %RH, relay activates when the relative humidity reaches 60
%RH. As the humidity decreases, relay releases at 55 %RH.
If both setpoints are specified and "above" setpoint is lower than
"below" setpoint, the hysteresis works in the opposite direction, that
is, relay is released when the measured value passes the exact value
of the setpoint.
Relay Indicating Transmitter Error
Status
You can set a relay to follow the operation status of the device. By
selecting FAULT/ONLINE STATUS for output quantity a relay
changes state on the basis of the operation status as follows:
FAULT STATUS
Normal operation: relay active (C and NO outputs are closed)
Not measuring state (error state or power off): relay released (C and
NC outputs are closed)