Vaisala HMT360N Technical data

P.O. Box 26, FI-00421 Helsinki, Finland
115 (4.53)
164 (6.46)
125 (4.92)
39 (1.5)
73 (2.9)
11 (0.43) 62 (2.4)
Email: industrialsales@vaisala.com
www.vaisala.com/HMT360N
HMT360 Series Dewpoint and Temperature Transmitters for Natural Gas Moisture
Features/Benefits
high accuracy, excellent long-term stability, and negligible hysteresis
Excellent performance in harsh
conditions; tolerates glycol and water spikes
Measures dewpoint of water, also
outputs lb/MMscf, ppmv and T
Safe operation with the entire
transmitter in hazardous areas: Division 1 and 2 (USA, Canada), Category 1 / Zone 0 (EU)
One-year maintenance interval
saves lifetime costs
Interchangeable probes
Direct pipeline installation possible
Sampling cell option
NIST traceable (certicate included)
Tel. +358 9 894 91
Fax: +358 9 8949 2485
e HMT362, HMT364 and HMT368 are innovative and reliable instruments for measuring moisture in natural gas.
e Vaisala HUMICAP® Dewpoint and Temperature Transmitter Series HMT360 is a reliable instrument for measuring moisture in natural gas. An accurate moisture measurement is critical for preventing pipeline failures and ensuring the quality of the gas.
Reliable technology now for natural gas
e Vaisala HUMICAP® polymer sensor technology has been successfully used for decades in industrial moisture measurements. e technology is now optimized for measuring moisture in natural gas.
Low maintenance need
e HMT360 is long-lasting and typically needs calibration only once a year, saving lifetime costs and measurement downtime.
When recalibration is needed, the transmitter or just the probe can
20
be sent to Vaisala. All calibration coecients are included in the interchangeable probe unit itself, which means that a calibrated probe can be switched in the eld in a couple of minutes.
Intrinsically safe
e whole HMT360 transmitter can be installed into areas classied as having a constant risk of explosion.
Rugged and reliable
e durable design and trouble-free operation of the HMT360 ensure a long-term solution for monitoring moisture in natural gas.
Three models with interchangeable probes
e three probe options allow dierent measurement setups, for example using a sampling cell, or temporary or permanent installation directly into the pipeline.
Customized calibration and maintenance contracts for the HMT360 series transmitters for natural gas are available on request.
Dimensions
Dimensions in mm (inches)
HMT360 transmitter body
Vaisala Instruments Catalog 2009
Ref. B210768EN rev. A
HMT360 SERIES FOR NATURAL GAS DEWPOINT
Dewpoint temperature (°C)
0
20 40 60
0
40
-50
Process temperature (°C)
10
20
30
50
60
-10
-20
-40
-30
-60
3010
50
-10
-20
Not recommended for long-term use
Accuracy not specified
Accuracy ± 2 °C Td
Technical Data
Performance
Water dewpoint measurement
Measurement range -50 ... +50 °C (-58 ... +122 °F) T Accuracy ±2 °C (±3.6 °F) (see graph below)
Dewpoint accuracy vs. measurement conditions
Response time 63 % [90 %]
-20 -> +10 °C (-4 -> +50 °F) 6 s [32 s] +10 -> -20 °C (+50 -> -4 °F) 120 s [370 s]
Sensor Vaisala HUMICAP® 180M
Temperature
Measurement range -40 ... +100 °C (-40 ... +212 °F) Typical accuracy at +20 °C (+68 °F) ±0.2 °C (±0.36 °F) Sensor Pt 1000 RTD 1/3 Class B IEC 751
Calculated variables (typical ranges)
water content 0 ... 20 lb/MMscf, 0 ... 320 mg/Nm parts per million by volume 0 ... 500 ppm
Operating environment
Temperature range
operating -20 ... +60 °C (-4 ... 140 °F) storage -40 ... +70 °C (-40 ... 158 °F) Pressure range see probe specications
Complies with EMC standard EN61326-1, Electrical equipment for measurement, control and laboratory use ­EMC requirements; Industrial environment. NOTE! IEC 1000-4-5 complies only when using external EXi approved surge arrester on safe area.
Inputs and outputs
Operating voltage 12 ... 28 VDC
with serial port (service mode) 15 ... 28 VDC Analog outputs two-wire 4 ... 20 mA, one standard, one optional Typical accuracy of analog outputs at +20 °C ±0.05% full scale Typical temperature dependence
of analog outputs 0.005% / °C (0.005% / °F) full scale Analog outputs connection via safety barriers RS232C serial output for service use connector type RJ45 Display two-line LCD character size (1st line/2nd line) 12 mm/10 mm
USA (FM) Classes I, II, III, Division 1, Groups A-G and
Division 2, Groups A-D, F and G
d
Safety factors: V
= 1 nF, Li = 0, Pi = 0.7 W, T
C
i
FM Project ID: 3010615
= 28 VDC, I
max
= 60 °C(140 °F), T5
amb
Japan (TIIS) Ex ia IIC T4
Code number: TC17897
Safety factors: U
= 28 VDC, Ii = 100 mA, Ci = 1 nF,
i
= 0.7 W, Li = 0, T
P
i
amb
China (PCEC) Ex ia II CT5
Certicate No. CE042052
Standard GB3686.1-2000 and GB3836.4-2000
Canada (CSA) Class I, Division 1 and Division 2, Groups A, B, C, D;
Class II, Division 1 and Division 2, Groups G and Coal Dust;
CSA File No: 213862 0 000, CSA Report: 1300863
Safety factors: T
connected as per Installation Drawing DRW213478.
= 60 °C, T4, Intrinsically safe when
amb
Russia (STV) Ex ia IIC T4
Certicate No. ROSS FI.GB04.V00634
Safety factors Ui = 28 V, Ii = 100 mA, Pi = 700 mW
Ci = 1 nF, Li = 0 H, T
Mechanics
3
Connections screw terminals, 0.33 ... 2.0 mm
v
= -20 °C ... +60 °C
amb
wires (AWG 14-22) Cable bushing for 7.5..12mm or 10..15mm cable diameters (M20x1,5) Conduit tting NPT 1/2" (M20) Housing material G-AlSi10Mg (DIN 1725) Housing classication IP66 (NEMA 4X) Housing weight 950 g
Options and accessories
Additional analog output 4 ... 20 mA Ball valve ISO 1/2 with welding joint ( for HMP368) BALLVALVE-1
pressure range at +20 °C (+68 °F): 0 ... 40 bar (0 ... 580 psia)
Serial interface cable for PC
connectors RJ45 - D9 female 25905ZZ Sampling cell for HMP362 HMP302SC Rain shield 215109 Calibration adapter for HMK15 211302 Sintered stainless steel lter HM47280 Galvanic isolator (EU) 212483 Zener Barrier (USA & Canada) 210664
(during installation max. 10 bar (145 psia)
= 100 mA,
max
= 60 °C(140 °F)
Class III
2
Classification with current outputs
Europe / CENELEC (PTB)
EU (94/9/EC, ATEX100a) II 1 G EEx ia IIC T4
Safety factors U
Environmental specications
T
amb
P
amb
Ref. B210768EN rev. A
= 28 V, Ii = 100 mA, Pi = 0.7 W
i
= 1 nF, Li = 0 H
C
i
-20 ... +60 °C (-4 ... 140 °F)
PTB 00 ATEX 2112 X
0.8 ... 1.1 bar
Cable bushing.
Conduit tting for wire conduits.
HUMICAP® is a registered trademark of Vaisala. Specications subject to change without prior notice. © Vaisala Oyj