HMT360 Series Intrinsically Safe Humidity and Temperature Transmitters for hazardous areas
Technical data
Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 % RH Accuracy (including non-linearity, hysteresis, and repeatability)
with Vaisala HUMICAP® 180 or 180R for typical applications
at +15 ... +25 °C (59 ... +77 °F) ± 1.0 % RH (0 ... 90 %RH)
±1.7 %RH (90 ... 100 %RH) at -20 ... +40 °C (-4 ... +104 °F) ±(1.0 + 0.008 x reading) %RH at -40 ... +180 °C (-40 ... +356 °F) ± (1.5 +0.015 x reading) %RH
with Vaisala HUMICAP® 180L2 for application with demanding
chemical environment at -10... +40 °C (14 ... +104 °F) ± (1.0 + 0.01 x reading) %RH at -40 ... +180 °C (-40 ... +356 °F) ± (1.5 + 0.02 x reading) %RH
Factory calibration uncertainty (+20 °C) ± 0.6 % RH (0 ... 40 %RH)
± 1.0 % RH (40 ... 97 %RH)
(Defined as ±2 standard deviation limits. Small variations
possible, see also calibration certificate.)
Response time (90 %) at +20 °C (+68 °F) in still air
with grid filter 8 s / 17 s* with grid + steel netting filter 20 s / 50 s* with sintered filter 40 s / 60 s*
* with HUMICAP® 180R sensor
TEMPERATURE
Measurement range -40 ... +180 °C (-40 ... +356 °F)
(depends on selected probe) Typical accuracy of electronics at +20 °C (+68 °F) ±0.2 °C (0.36 °F) Typical temparature dependence
of electronics 0.005 °C/°C (0.005 °F/°F) Sensor Pt 1000 RTD 1/3 Class B IEC 751 Accuracy over temperature range
OTHER VARIABLES
Optionally available dewpoint temperature, mixing ratio,
absolute humidity, wet bulb temperature.
Operating environment
Temperature range
operating temp. range for electronics with display -20 ... +60 °C (-4 ... +140 °F) storage -40 ... +70 °C (-40 ... +158 °F)
Pressure range see probe specifications
Complies with EMC standard EN61326-1, Electrical equipment for measurement, control and laboratory use - EMC requirements; Industrial Environment. NOTE! IEC 1000-4-5 complies only when using external EXi approved surge arrester on safe area.
-40 ... +60 °C (-40 ... +140 °F)
Technical data
Inputs and outputs
Operating voltage 12 ... 28 V
with serial port (service mode) 15 ... 28 V Analog outputs two-wire 4 ... 20 mA,
one standard, one optional Typical accuracy of analog outputs at +20 °C ±0.05% full scale Typical temperature dependence
of analog outputs 0.005% / °C (0.005% / °F) full scale Analog outputs connection via safety barriers RS232C serial output for service use connector type RJ45 Display two-line LCD
Mechanics
Connections screw terminals, 0.33...2.0 mm
2 wires (AWG 14-22)
Cable bushings For 7.5...12mm or 10...15mm cable
diameters (M20) Conduit fitting NPT 1/2" (M20) Housing material G-AlSi10Mg (DIN 1725) Housing classification IP66 (NEMA 4X) Housing weight 950 g
Options and accessories
Duct installation kit (for HMP363/367) 210697 Mounting flange (for HMP365) 210696 Ball valve ISO 1/2 with welding joint
(for HMP368) pressure range at +20 °C (+68 °F): 0 ... 20 bar (0 ... 290 psia)
(during installation max. 10 bar (145 psia) Calibration adapter for HMK15 211302 Serial interface cable for PC
connectors RJ45 - D9 female 25905ZZ Galvanic isolator 212483 Zener barrier (USA & Canada) 210664 Protection cover (for use in the presence of combustible dust, ATEX)
BALLVALVE-1
214101
II 1 D IP65 T = 70 °C
Classification with current outputs
EUROPE / VTT
EU (94/9/EC, ATEX100a) II 1 G Ex ia IIC T4 Ga
VTT 09 ATEX 028 X issue No: 1
Safety factors Ui = 28 V, Ii = 100 mA, Pi = 700 mW
Ci = 1 nF, Li negligibly low
Environmental specifications
T
amb
P
amb
Dust classification (with protection cover) II 1 D (IP65 T=70 °C)
USA (FM)
Classes I, II, III, Division 1, Groups A-G and
Safety factors: Vmax = 28 VDC, I
Ci = 1 nF, Li = 0, Pi = 0.7 W, T
JAPAN (TIIS)
Safety factors: Ui = 28 VDC, Ii = 100 mA, Ci = 1 nF,
Pi = 0.7 W, Li = 0, T
CANADA (CSA)
Class I Division 1 and Division 2, Groups A, B, C, D; Class II Division 1 and Division 2, Groups G and
Class III CSA File No: 213862 0 000, CSA Report: 1300863 Safety factors: T
Intrinsically safe when connected as per
CHINA (PCEC)
Standard GB3686.1-2000 and GB3836.4-2000
IECEx (VTT)
Safety factors
Environmental specification
T
amb
P
amb
-40 ... +60 °C (-40 ... +140 °F)
0.8 ... 1.1 bar
VTT 04 ATEX 023X
Division 2, Groups A-D, F and G
FM Project ID: 3010615
= 100 mA,
max
= 60 °C(140 °F), T5
amb
Ex ia IIC T4
Code number: TC17897
= 60 °C (140 °F)
amb
Coal Dust;
= 60 °C, T4,
amb
Installation Drawing DRW213478.
Ex ia II CT5
Certificate No. CE042052
Ex ia IIC T4 Ga
IECEx VTT 09.0002x issue No: 1
Ui = 28 V, Ii = 100 mA, Pi = 700 mW
= 1 nF, Li negligibly low
C
i
-40 ... +60 °C (-40 ... +140 °F)
0.8 ... 1.1 bar
For more information, visit www.vaisala.com or contact us at sales@vaisala.com
Ref. B210890EN-A ©Vaisala 2009
This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. The reproduction, transfer, distribution or storage of information contained in this brochure in any form without the prior written consent of Vaisala is strictly prohibited. All specifications — technical included — are subject to change without notice.