Copyright 2013 Vaisala Oyj. All rights reserved.
Trademarks
Vaisala Oyj, and Vaisala HUMICAP® Humidity and Temperature Transmitter HMT140 are trademarks of Vaisala Oyj.
Windows is a registered trademark of Microsoft Corporation.
All other trademarks referred to are the property of their respec-
tive owners.
Disclaimer and Limitation of Liability
Vaisala Oyj and its subsidiaries assume no responsibility for any
damage or loss resulting from the use of this manual.
Vaisala Oyj and its subsidiaries assume no responsibility for any
loss or claims by third parties which may arise through the use of
this software. Vaisala Oyj and its subsidiaries assume no responsibility for any damage or loss caused by deletion of data as a
result of hardware malfunction. Be sure to make backup copies of
all important data to protect against data loss.
Vaisala Oyj makes no warranties, either expressed or implied,
regarding Vaisala data loggers or software, its merchantability, or
its fitness for any particular purpose. The exclusion of implied
warranties is not permitted by some states. As such, the exclusion
may not apply.
The Vaisala HUMICAP® Humidity and Temperature Transmitter HMT140 has been tested using a variety of network systems.
However, the large number of possible hardware and network
configurations makes testing under every circumstance impossible. If you have trouble using any Vaisala software or hardware,
contact Vaisala Oyj.
Technical Support
For technical support in North America, please call 1-866-8613388, or for customers outside North America, see “Getting Help”
on page 2.
Email: helpdesk@vaisala.com
Document part number: M211488EN-C
Release date: August 2013
No copying, distribution, publication, modification, or incorporation of this document, in whole or part, is permitted for commercial purposes without the express written permission of Vaisala
Oyj.
Thank you for choosing the Vaisala HUMICAP® Humidity and
Temperature Transmitter HMT140.
About this Manual
This manual provides information for installing, operating, and
maintaining the Vaisala HUMICAP® Humidity and Temperature
Transmitter HMT140. It also details the operation of the HMT140
Utility software.
Contents of this Manual
• Chapter 1: General Information, provides safety and
regulatory information for the HMT140 series of transmitters.
• Chapter 2: Product Overview, provides inform ation about the
HMT140 features, models and accessories.
• Chapter 3: Installation, describes how to mount the HMT140
to a wall or duct, and illustrates different wiring options.
• Chapter 4: Setup and Operation, contains the information you
need to install and connect an HMT140 to your network using
the HMT140 Utility software and an HMT140 Configuration
Cable, and instructions on how to operate the HMT140
transmitter.
• Chapter 5: Maintenance, provides information required for
basic maintenance of the product.
• Appendix: Specifications and Troubleshooting, provides
additional technical data.
M211488EN-C1
Getting Help
Related Documentation and Software
For help setting up the HMT140 with basic configuration
requirements, consult the Vaisala HUMICAP® Humidity and
Temperature Transmitter HMT140 Quick Guide.
For help configuring the HMT140 with Vaisala Veriteq viewLinc,
consult the Vaisala Veriteq viewLinc Administrator’s Guide.
Getting Help
North America
Contact Vaisala Canada Inc, 8am - 4pm PST monday to Friday, 1866-861-3388 (or 605-273-6850) or email helpdesk@vaisala.com.
See also http://www.vaisala.com/en/lifescience.
For sales, pricing, quotations, or general information, please call 1800-273-8374 (or 604-273-6850).
Vaisala Headquarters (Finland)
Vanha Nurmijärventie 21
01670 Va ntaa
FINLAND
Industrial Instruments
Phone: +358 9 8949 2658
Fax: +358 9 8949 2295
Japan Service Center
42 Kagurazaka 6-Chome
Shinjuku-ku
Tokyo 162-0825
JAPAN
Phone: +81 3 3266 9611
Fax: +81 3 3266 9610
2VAISALA
Safety
Safety
China Service Center
Floor 2, EAS Building
No. 21, Xiao Yun Road, Dongsanhuan Beilu
Chaoyang District
Beijing 100027
CHINA
Phone: +86 10 5827 4100
Fax: +86 10 8526 1155
The Vaisala HUMICAP® Humidity and Temperature Transmitter
HMT140 delivered to you has been tested for safety and approved as
shipped from the factory.
Do not modify the unit. Improper modification can damage the
product or lead to malfunction.
Batteries
Remove batteries before shipping.
ESD Protection
Electrostatic Discharge (ESD) can cause immediate or latent damage
to electronic circuits. Vaisala products are adequately protected
against ESD for their intended use. However, it is possible to damage
the product by delivering electrostatic discharges when touching,
removing or inserting any objects inside the equipment housing.
M211488EN-C 3
License Agreement
To make sure you are not delivering high static voltages:
• Handle ESD sensitive components on a properly grounded and
protected ESD workbench.
• When this is not possible, ground yourself to the equipment
chassis before touching the boards, or ground yourself with a
wrist strap and a resistive connection cord.
• When neither of the above is possible, touch a conductive part of
the equipment chassis with your other hand before touching the
boards.
• Always hold the boards by the edges and avoid touching the
component contacts.
Recycling
Recycle all applicable material. Dispose of batteries and the unit
according to statutory regulations. Do not dispose with regular
household refuse.
License Agreement
All rights to any software are held by Vaisala or third parties. the
customer is allowed to use the software only to the extent that is
provided by the applicable supply contract or Software License
Agreement.
Warranty
Visit our Internet pages for information about our standard
warranty terms and conditions: http://www.vaisala.com/en/
services/maintenance/Pages/warranty.aspx.
Please observe that any such warranty may not be valid in case of
damage due to normal wear and tear, exceptional operating
conditions, negligent handling or installation, or unauthorized
modifications. Please see the applicable supply contract or
Conditions of Sale for details of the warranty for each product.
4VAISALA
Regulatory Compliances
The Vaisala HUMICAP® Humidity and Temperature Transmitter
HMT140 is in conformity with the provisions of the following
directives/standards:
• FCC: U30-G2M5477 (FCC Part 15 SubpartB: 2008 Class B,
FCC CRF Title 47 Part 15 Subpart C)
• IC: 8169A-G2M5477 (IC RSS-210)
• CE: EU ID # 0681 (ETSI EN301 489-1 V1.8.1, 489-17 V2.1.1,
ETSI EN300 328 V1.7.1, EN55022 Class B, EN61000 4-2, 4-3,
4-6, 4-8, EN60950-1:2001+A11:2004)
• CAN ICES-3(B)/NMB-3(B)
• Wi-Fi Certified: WFA7649
• FCC 47 CRF Part 15-2011 Subpart B
• IC ICES-003:2004
• EN61326-1:2006
• EN61326-2-3:2006
• EN61000-3-2:2006+A1:2009+A2:2009
• EN61000-3-3:2008
• EN61326-1:2006
• MIC R 201-125765
• CMIIT ID: 2013DJ7129
Regulatory Compliances
HMT140 Configuration Cable
The HMT140 Configuration Cable complies with part 15 of the
FCC Rules. Operation is subject to the following two conditions:
• This device may not cause harmful interference.
• This device must accept any interference received, including
interference that may cause undesired operation.
M211488EN-C 5
Regulatory Compliances
6VAISALA
CHAPTER 2
PRODUCT OVERVIEW
The Vaisala range of relative humidity measurement
instruments covers all the applications from ventilation to
process control in demanding conditions. For more
information about other Vaisala relative humidity
instruments, please contact your Vaisala representative or
visit www.vaisala.com.
This chapter provides general information about the
HMT140, including:
•Features
•Models
• Options and accessories
• Transmitter components
M211488EN-C7
Introduction to the HMT140
Introduction to the HMT140
The Vaisala HUMICAP® Humidity and Temperature
Transmitter HMT140 measures relative humidity and
temperature using the connected probe and analog signals –
RTD, Voltage, Current Loops and Boolean Contacts. It is
powered with three 3.6 volt DC batteries (or an optional 930VDC power supply), outputs data via wireless transmitter
and comes with or without an optional LCD display.
The HMT140 transmitter supports several different signal
measurements (channels), which are monitored in viewLinc:
• Relative Humidity (RH)
• Temperature (T)
•Switch Contact
• Voltage and Current.
Channel types are set at the factory during time of order.
Other wireless parameters can be changed at a later date
using an HMT140 Configuration Cable and HMT140 Utility
software.
Models
The following HMT140 models and options are available:
• HMT141: Vaisala Humicap Humidity and Temp Probe
(RH/T)
• HMT143: Dual RTD Temperature Inputs w/5M Probes
(-90C to 90C)
• HMT144: Voltage Inputs (0-5 VDC)
• HMT145: Voltage inputs (0-10 VDC)
• HMT146: Current Inputs (0-20 mA)
• HMT147: Boolean Inputs
• HMT148: RTD Temperature Input w/5M Probe &
Boolean Input
Options and Accessories
Humidity & temperature probe HMP110*
Humidity & temperature replacement probe HMP110R*
8VAISALA
Introduction to the HMT140
Humidity sensor HUMICAP180R
HMP110 Probe mounting flange 226061
HMP110 Probe mounting clamps, 10 pcs 226067
HMP110 Probe cable 3 m HMT140Z300
HMP110 Probe cable 5 mHMT140Z500
HMP110 Probe cable 10 m HMT140Z1000
Duct installation kit215619
Probe mounting flange226061
Boolean Switch Kit with Door Contact236319
Thermal Dampener Blocks236310SP
Replacement Batteries (Pack of 3)236318SP
External Power Supply 236081SP
HMP110 sensor protectioM211488EN-CM211488EN-Cn
Plastic grid filter DRW010522SP
Plastic grid with membrane filter DRW010525
Stainless steel sintered filter HM46670SP
*Separate order form required. Contact Vaisala sales at:
The HMT140 is available either with a fixed probe directly
attached to the transmitter housing or a remote probe with
different (3/5/10 m) cable lengths. Extension cables can be
easily cascaded in order to obtain longer reach, to a
maximum length of 10 m.
Interchangeable Probe
The HMP110 HUMICAP® relative humidity probe used in
the HMT140 transmitter is fully interchangeable. You can
easily remove the probe and replace it with a new one
without having to adjust the transmitter. You have the
following options when purchasing a new probe from
Vaisala:
•Order a new probe and keep your current one.
•Order a new probe and return the old one to Vaisala
(replacement probe).
M211488EN-C9
Introduction to the HMT140
Note:
Only probes that have a compatible digital output (VDIGI mode) can be used with the HMT140
transmitter. Compatible probes have the letter
"V" as the first letter in their order code. The
order code is written on the probe.
Optional Display
The HMT140 is also available with an optional graphical
128x64 pixel resolution LCD display. The display shows the
measurement results of selected parameters in selected units
(defined at the time of ordering). The parameters are
displayed simultaneously as two separate rows on the
display.
Transmitter Components
Figure 1 illustrates the main features of the HMT140. On the
left is a remote probe model without display, and on the right
is a fixed probe model with the optional display. The
numbers and arrows indicate the main components of the
transmitters.
The device is typically installed mounted on the wall with up
to four screws (not included). See Figure 2 on page 15.
For duct mounting, Duct Installation Kits are available as
accessories. See Options and Accessories, on page 8.
To help you position the screws correctly, use the Drilling
Template, available inside the back cover of the Quick Guide.
HMT140 Utility and Configuration Cable
To connect an HMT140 device to your network, you will need
to install the HMT140 Utility software and connect the device
to your PC with the configuration cable (see, Connecting to
your Network, on page 24).
M211488EN-C11
Introduction to the HMT140
12VAISALA
CHAPTER 3
INSTALLATION
This chapter provides the information you need to:
• Mount the HMT140 on a wall or duct
•Connect the HMT140 to your network
• Review optional mounting accessories
M21488EN-C 13
Mounting
Mounting
Depending on your installation requirements, you may wish
to complete the network configuration and setup prior to
mounting (see, Chapter 4 Setup and Operation). Note that
you can complete the configuration and setup procedures
after mounting with a portable PC.
Note:To connect the HMT140 to your network
Wall Mounting
To mount the HMT140 on a wall:
Note:If mounting against a metal surface, be sure to
Open the HMT140 cover. Use your thumb to press and
1
hold the release tab located between the two glands (see
Figure 9 on page 19), then pull the cover out and up with
your other hand.
2Make sure that the HMT140 is correctly aligned and
attach it directly to the wall with up to four screws (not
included in the package).
and view HMT140 data using viewLinc, you
require HMT140 Utility software and an
HMT140 Configuration Cable (shipped with
viewLinc).
keep a minimum of 2 mm between the metal and
the back of the device to ensure an uninterrupted
radio signal.
•Select the size and type of the fastening screws
according to the wall material (for example, wood or
stone). The diameter of the fastening screws is
typically between 3.5 and 4 mm.
14VAISALA
Mounting
Figure 2: Mounting Diagram
Note:It is strongly recommended that you use all four
screws; however, the HMT140 enclosure fastening holes are initially covered with a thin plastic
membrane, so less than four screws could be also
be used without sacrificing the ingress protection
(IP) class of the enclosure.
Duct Installation
The duct installation kit includes a plastic pipe with a flange
(Vaisala part#: 215619).
To install the HMP110 probe with the duct installation kit,
drill a hole into the duct wall, assemble the probe to the duct
installation kit, slide the probe head through the hole, and
attach the flange to the duct wall with four screws.
M21488EN-C 15
Loading...
+ 53 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.