Vaisala HMP80 User Manual

Page 1
M212903EN-A
User Guide
Vaisala Handheld Humidity
and Temperature Probes
HMP80 Series
Page 2
PUBLISHED BY
Vaisala Oyj Vanha Nurmijärventie 21, FI01670 Vantaa, Finland P.O. Box 26, FI00421 Helsinki, Finland +358 9 8949 1
www.vaisala.com docs.vaisala.com
© Vaisala 2023
No part of this document may be reproduced, published or publicly displayed in any form or by any means, electronic or mechanical (including photocopying), nor may its contents be modified, translated, adapted, sold or disclosed to a third party without prior written permission of the copyright holder. Translated documents and translated portions of multilingual documents are based on the original English versions. In ambiguous cases, the English versions are applicable, not the translations.
The contents of this document are subject to change without prior notice.
Local rules and regulations may vary and they shall take precedence over the information contained in this document. Vaisala makes no representations on this document’s compliance with the local rules and regulations applicable at any given time, and hereby disclaims any and all responsibilities related thereto.
This document does not create any legally binding obligations for Vaisala towards customers or end users. All legally binding
obligations and agreements are included exclusively in the applicable supply contract or the General Conditions of Sale and General Conditions of Service of Vaisala.
This product contains software developed by Vaisala or third parties. Use of the software is governed by license terms and conditions included in the applicable supply contract or, in the absence of separate license terms and conditions, by the General License Conditions of Vaisala Group.
This product may contain open source software (OSS) components. In the event this product contains OSS components, then such OSS is governed by the terms and conditions of the applicable OSS licenses, and you are bound by the terms and conditions of such licenses in connection with your use and distribution of the OSS in this product. Applicable OSS licenses are included in the product itself or provided to you on any other applicable media, depending on each individual product and the product items delivered to you.
Page 3

Table of contents

Table of contents
1. About this document.....................................................................................7
1.1 Version information.......................................................................................... 7
1.2 Related manuals................................................................................................7
1.3 Documentation conventions............................................................................7
1.4 Trademarks........................................................................................................ 8
2. Product overview............................................................................................ 9
2.1 Basic features and options.............................................................................. 9
2.2 HMP80N and HMP80L probe models............................................................ 9
2.3 Output parameters......................................................................................... 10
2.4 Probe structure................................................................................................. 11
2.5 Safety................................................................................................................ 12
2.6 ESD protection.................................................................................................12
3. Functional description.................................................................................13
3.1 HUMICAP technology..................................................................................... 13
3.2 Sensor purge....................................................................................................13
3.3 Condensation prevention functions..............................................................14
3.4 Environmental compensation........................................................................14
3.5 Filtering factor................................................................................................. 15
4. Attaching cable to probe........................................................................... 16
5. Using probe with Indigo80 handheld indicator................................18
5.1 Indigo80 handheld indicator......................................................................... 18
5.1.1 Indigo80 keypad...................................................................................... 19
5.1.2 Indigo80 main menu................................................................................19
5.2 Connecting probes to Indigo80.................................................................... 19
5.3 Measurement location.....................................................................................21
5.3.1 Configuring environmental compensations.........................................22
5.4 Configuring probe features with Indigo80 handheld indicator................22
5.4.1 Example of configuring filtering factor with Indigo80
handheld indicator.................................................................................. 23
5.5 Configuring environment settings with Indigo80 handheld indicator.... 24
6. Configuration with Insight software.....................................................26
6.1 Vaisala Insight PC software...........................................................................26
6.2 Connecting to Insight software.................................................................... 26
6.3 Configuration options.................................................................................... 27
6.4 Diagnostics in Insight..................................................................................... 27
7. Calibration and adjustment......................................................................29
7.1 Calibration and adjustment overview..........................................................29
7.2 Adjustment points and requirements..........................................................29
3
Page 4
HMP80 Series User Guide M212903EN-A
7.3 Calibration and adjustment with Indigo80 handheld indicator...............30
7.3.1 Performing guided calibration with Indigo80 handheld indicator.... 31
7.3.2 Performing manual adjustment with Indigo80 handheld
indicator....................................................................................................32
7.3.3 Restoring factory adjustment with Indigo80 handheld indicator.....33
7.4 Adjusting measurement with Insight software.......................................... 34
7.5 HMK15 Humidity Calibrator with HMP80 series probes.............................35
8. Maintenance....................................................................................................36
8.1 Cleaning the probe.........................................................................................36
8.1.1 Chemical tolerance................................................................................. 36
8.2 Changing the probe filter.............................................................................. 37
9. Troubleshooting............................................................................................38
9.1 Problem situations......................................................................................... 38
9.2 Error messages............................................................................................... 38
9.3 Restoring factory default settings with Insight software......................... 40
9.4 Restoring factory default settings with Indigo80 handheld indicator..... 41
9.5 Writing a problem report............................................................................... 41
10. Technical data................................................................................................ 42
10.1 HMP80 specifications.................................................................................... 42
10.2 Dimensions......................................................................................................45
10.3 Spare parts and accessories.........................................................................46
10.4 Accuracy of calculated output parameters.................................................46
Appendix A:
Recycling instructions...........................................................48
Maintenance and calibration services.........................................................51
Warranty.............................................................................................................51
Technical support.............................................................................................51
Recycling............................................................................................................51
4
Page 5
List of figures
Figure 1 HMP80N dimensions, side and bottom view............................................10
Figure 2 HMP80L dimensions, side and bottom view............................................ 10
Figure 3 Probe parts (model HMP80N shown)..........................................................11
Figure 4 Indigo80 handheld indicator.........................................................................18
Figure 5 Indigo80 keypad...............................................................................................19
Figure 6 Indigo80 main menu........................................................................................19
Figure 7 Connecting probes to Indigo80.................................................................. 20
Figure 8 Example installation........................................................................................22
Figure 9 HMP80L in Insight software..........................................................................27
Figure 10 HMP80L diagnostics in Insight (advanced mode)..................................28
Figure 11 Indigo80 calibration methods for probes.................................................30
Figure 12 Calibrating a humidity probe with HMK15 using 211302 adapter........35
Figure 13 HMP80N humidity measurement accuracy as a function
of temperature..................................................................................................43
Figure 14 HMP80N temperature measurement accuracy over full range.......... 43
Figure 15 HMP80L humidity measurement accuracy as a function
of temperature..................................................................................................43
Figure 16 HMP80L temperature measurement accuracy over full range........... 43
Figure 17 Dimensions of HMP80L (top) and HMP80N (bottom),
side and bottom view.....................................................................................45
Figure 18 Vaisala Humidity Calculator in Advanced mode..................................... 47
Figure 19 Materials for recycling....................................................................................48
List of figures
5
Page 6
HMP80 Series User Guide M212903EN-A

List of tables

Table 1 Document versions (English)............................................................................7
Table 2 Related manuals................................................................................................... 7
Table 3 HMP80N and HMP80L temperature ranges.................................................9
Table 4 HMP80 series output parameters.................................................................. 10
Table 5 Troubleshooting table.......................................................................................38
Table 6 HMP80 series measurement performance................................................. 42
Table 7 HMP80 series operating environment......................................................... 43
Table 8 HMP80 series inputs and outputs................................................................. 44
Table 9 HMP80 series output parameters................................................................. 44
Table 10 HMP80 series mechanical specifications.................................................... 44
Table 11 HMP80 series compliance............................................................................... 45
Table 12 HMP80 series spare parts and accessories.................................................46
Table 13 Materials for recycling...................................................................................... 48
6
Page 7
Chapter 1 – About this document

1. About this document

1.1 Version information

This document provides instructions for using and maintaining Vaisala HMP80 HUMICAPâ Handheld Humidity and Temperature Probes.
Table 1 Document versions (English)
Document code Date Description
M212903EN-A June 2023 First version.

1.2 Related manuals

For the latest versions of these documents, see docs.vaisala.com.
Table 2 Related manuals
Document code Name
M212722EN Vaisala Indigo80 Handheld Indicator User Guide
M212904EN Vaisala DMP80 Series Handheld Probes User Guide
M212804EN Vaisala DMP80 Series Handheld Probes Quick Guide
M212803EN Vaisala HMP80 Series Handheld Probes Quick Guide
M210185EN Vaisala HMK15 Humidity Calibrator User Guide

1.3 Documentation conventions

WARNING!
follow instructions carefully at this point, there is a risk of injury or even death.
CAUTION!
follow instructions carefully at this point, the product could be damaged or important data could be lost.
Warning alerts you to a serious hazard. If you do not read and
Caution warns you of a potential hazard. If you do not read and
7
Page 8
HMP80 Series User Guide M212903EN-A
Note highlights important information on using the product.

1.4 Trademarks

Vaisalaâ and HUMICAPâ are registered trademarks of Vaisala Oyj.
Microsoftâ and Windowsâ are either registered trademarks or trademarks of Microsoft Corporation in the United States and other countries.
All other product or company names that may be mentioned in this publication are trade names, trademarks, or registered trademarks of their respective owners.
8
Page 9
Chapter 2 – Product overview

2. Product overview

Vaisala HUMICAPâ Handheld Humidity and Temperature Probes HMP80 Series have been designed for portable spot-checking, field calibration, and data logging use with the Indigo80 handheld indicator. For easy access to device analytics and configuration, HMP80 probes can be connected to Vaisala Insight software for Windowsâ.
The HMP80 series probes are available in two lengths oering similar measurement performance. The longer model (HMP80L) is designed for measurements in more extreme temperatures.
The IP66-classified probe handle oers protection against moisture and dust with the probe connection cable attached. Also the cable connection is protected against mechanical stress by the robust design of the handle.
For more information on the Indigo product family, see www.vaisala.com/indigo.
The free Insight PC software is available for download at www.vaisala.com/insight.
2.1

 Basic features and options

• Portable design optimized for industrial spot-checking and field calibration
• Wide temperature measurement range
• Condensation-tolerant
• Sensor purge improves long-term stability and chemical resistance
• Compatible with Indigo80 handheld indicator and Insight PC software
More information
Indigo80 handheld indicator (page 18)
Vaisala Insight PC software (page 26)
2.2

 HMP80N and HMP80L probe models

Both HMP80 series probe models have the same type of sensor, Vaisala HUMICAPâ R2C, with sensor purge and sensor warming functionality.
Table 3 HMP80N and HMP80L temperature ranges
Property HMP80N HMP80L
Relative humidity measurement range
Temperature measurement range
0 … 100 %RH
−20 … +60 °C (−4 … +140 °F) −50 … +120 °C (−58 … +248 °F), short-time measurement range
−50 … +180 °C (−58 … +356 °F)
9
Page 10
Ø12 [0.5]
103 [4.1] 83 [3.3]
23
63 [2.5]
272 [11]
[0.91]
25
[0.98]
47 [1.9]
mm [in]
probe handle detachable
end piece
25
[0.98]
47 [1.9]
mm [in]
103 [4.1]
278 [11]
63 [2.5]
485 [19]
41 [1.6]
Ø 12 [0.5]
probe handle detachable
end piece
HMP80 Series User Guide M212903EN-A
Property HMP80N HMP80L
Operating temperature of
−20 … +60 °C (−4 … +140 °F) −50 … +120 °C (−58 … +248 °F)
probe head
Operating temperature of
−10 … +60 °C (−14 … +140 °F)
probe body
The identifying letter of each probe model is included in the configuration code of the probe, shown on the side of the device. Example codes: N1A1S0 for an HMP80N probe, and L1A1S0 for an HMP80L probe.
Figure 1 HMP80N dimensions, side and bottom view
Figure 2 HMP80L dimensions, side and bottom view
2.3

 Output parameters

Values of all available output parameters are locked (showing the latest valid value) when sensor purge or condensation prevention functions are active.
Table 4 HMP80 series output parameters
Output parameter name and unit
Absolute humidity (g/m3)
Absolute humidity at NTP (g/m3)
Relative humidity (%RH)
Relative humidity (dew/frost) (%RH)
Dew point temperature (°C) Temperature (°C)
10
Page 11
2
5
4
3
6
1
Chapter 2 – Product overview
Output parameter name and unit
Dew/frost point temperature (°C) Water concentration (ppmv)
Dew/frost point temperature at 1 atm (°C) Water concentration (wet basis) (vol%)
Dew point temperature at 1 atm (°C) Water mass fraction (ppmw)
Dew point temperature dierence (°C) Water vapor pressure (hPa)
Enthalpy (kJ/kg) Water vapor saturation pressure (hPa)
Mixing ratio (g/kg) Wet-bulb temperature (°C)
More information
HMP80 specifications (page 42)
2.4

 Probe structure

Figure 3 Probe parts (model HMP80N shown)
1 Protection cap (remove before use) 2 Location of sensors on the probe head,
inside a removable filter
3 Probe shaft 4 Probe handle with type label. Contains
component board and electronics
housing.
5 M12-5M A-coded connector (inside the
probe handle)
6 Detachable end piece of the probe
handle, protecting the cable connector
inside the handle
1)
1) The HUMICAPâ R2C sensor used in
HMP80 series probes is not user­replaceable. To have the sensor replaced, contact a Vaisala Service Center.
11
Page 12
HMP80 Series User Guide M212903EN-A
More information
Recycling instructions (page 48)
Maintenance and calibration services (page 51)

2.5 Safety

WARNING!
has not been exposed to dangerous contamination, and is safe to handle without special precautions.
CAUTION!
are no user serviceable parts inside the probe handle.
CAUTION!
deposit impurities on the probe head.
CAUTION!
condensation prevention warming or sensor purge functions are active.
When returning a product for calibration or repair, make sure it
Do not attempt to open the solid part of the probe handle. There
Do not touch the probe head with your bare hands. Touching will
Be aware that the probe head may become hot to touch when

2.6 ESD protection

Electrostatic discharge (ESD) can damage electronic circuits. Vaisala products are adequately protected against ESD for their intended use. However, it is possible to damage the product by delivering electrostatic discharges when touching, removing, or inserting any objects in the equipment housing.
To avoid delivering high static voltages to the product:
• Handle ESDsensitive components on a properly grounded and protected ESD workbench or by grounding yourself to the equipment chassis with a wrist strap and a resistive connection cord.
• If you are unable to take either precaution, touch a conductive part of the equipment chassis with your other hand before touching ESDsensitive components.
• Hold component boards by the edges and avoid touching component contacts.
12
Page 13
Chapter 3 – Functional description

3. Functional description

3.1 HUMICAP technology

Vaisala HUMICAPâ is a capacitive thin-film polymer sensor consisting of a substrate on which a thin film of polymer is deposited between two conductive electrodes. The sensing surface is coated with a porous metal electrode to protect it from contamination and exposure to condensation. The substrate is typically glass or ceramic.
The thin-film polymer either absorbs or releases water vapor as the relative humidity of the measurement environment rises or falls. The dielectric properties of the polymer film depend on the amount of absorbed water. As the relative humidity around the sensor changes, the dielectric properties of the polymer film change, and so does the capacitance of the sensor. Capacitance is converted into a humidity reading. The HUMICAP sensor is complemented by a temperature sensor, which is bonded together with the capacitive polymer on some of the sensor variants. Additional output parameters such as dew point are calculated from the readings measured by the sensors.
HUMICAP sensors with bonded temperature sensors are also known as composite sensors. Probes with composite sensors can use the sensor purge function to maintain sensor performance. They also implement condensation prevention functions that are a key element in achieving an excellent stability and good condensation resistance in high humidity applications.
3.2

 Sensor purge

Sensor purge heats the sensor to a high temperature, which evaporates excess molecules out of the sensor polymer. This minimizes the drift at the wet end of humidity readings, and removes contaminants from the sensor element. The heating phase is followed by a cooling phase where the sensor cools down to ambient temperature.
During sensor purge the values of the output parameters stay frozen at their last measured values. Sensor purge takes a few minutes to complete, after which measurement continues normally and updated values of output parameters are again available.
HMP80 series probes perform the purge at one day intervals (known as interval purge). If desired, sensor purge can also be performed when the probe is powered on (known as startup purge). Purge settings can be configured with the Insight PC software or the Indigo80 handheld indicator.
To allow the sensor purge to be completed in a reasonable time, it will not be started if the ambient temperature is below 0 °C (+32 °F).
13
Page 14
HMP80 Series User Guide M212903EN-A

3.3 Condensation prevention functions

Condensation prevention prevents water from condensing on the sensor and interfering with humidity measurement. Use of condensation prevention is controlled by a single setting that can be changed with the Insight software or the Indigo80 handheld indicator.
By default, condensation prevention is o.
When condensation prevention is enabled, the probe will automatically use its available condensation prevention functions. When warming is active, values of output parameters that depend on temperature measurement (for example, relative humidity) are unavailable, unless the temperature compensation setting is set to read the temperature from an external source. Output parameters that can be measured or calculated without this external temperature information, such as dew point temperature, are available even without the temperature input.
Sensor warming
The sensor warming feature of the condensation prevention function is intended for improving the condensation tolerance of HMP80 probes. It warms the humidity sensor when necessary to keep its temperature above the dew point of the measurement environment. When sensor warming is activated, all temperature-dependent output parameters become unavailable, unless the temperature compensation setting is set to read the temperature from an external source. The output parameters become available again 4 minutes after the warming is stopped.
Sensor warming is intended for improving the condensation tolerance of the probe in measurement environments that may occasionally cause condensation to form. It is not intended for environments where continuous warming is needed.

3.4 Environmental compensation

Environmental compensations improve the measurement accuracy of the probe. You can configure the settings with the Insight software or the Indigo80 handheld indicator.
Pressure compensation
The pressure of the measured environment aects the accuracy of humidity calculations, especially calculations that are dependent on pressure (for example, mixing ratio). To achieve accurate output of all humidity parameters, you are recommended to update the value of the pressure compensation setpoint to match the environmental pressure when installing the probe.
For example, when you install the probe in a process with a pressure diering from normal atmospheric pressure, update the correct pressure into the pressure compensation setpoint parameter or register of the probe using the Insight software or the Indigo80 handheld indicator.
The entered value is persistent and remains unchanged at device restart.
14
Page 15
Chapter 3 – Functional description
Temperature compensation
You need to set an external source for the temperature compensation setpoint if the sensor warming feature of the condensation prevention function prevents the probe's own temperature measurement, and the output parameters that require temperature information to be known are required.
When warming is active, values of output parameters that depend on temperature measurement (for example, relative humidity) are unavailable, unless temperature information is read from an external source. Output parameters that can be measured or calculated without this external temperature information, such as dew point temperature, are available even without the temperature input.
The external temperature source can be another probe that is connected to the same Indigo80 handheld indicator as your probe.
To get the most accurate measurement values, the external source needs to be able to provide the temperature compensation setpoint value constantly in real time, and the value needs to correspond to the temperature of the probe's measurement environment as accurately as possible. A suitable measurement point is as close to the probe as possible, but yet far enough so that ongoing condensation prevention warming of the probe does not have an eect on the measured value.
At device restart, the value is cleared.
More information
Configuring environment settings with Indigo80 handheld indicator (page 24)
3.5

 Filtering factor

The filtering factor determines how quickly the results of measurement cycles are reflected in values of output parameters. You can use the filtering factor to slow down the response time of the measurement, filtering out rapid changes and measurement noise.
• Maximum value (1) = No filtering, fastest measurement. Output shows 100 % of the most recent measured value.
• Less than maximum value = Filtering is applied and measurement response time is reduced. For example, defining a value that is 90 % of the maximum value means that the filtered output reading = 90 % of the most recent measured value + 10 % of the previous reading.
Filtering factor settings can be configured with the Insight software or the Indigo80 handheld indicator.
More information
Example of configuring filtering factor with Indigo80 handheld indicator (page 23)
15
Page 16
HMP80 Series User Guide M212903EN-A

4. Attaching cable to probe

• Probe connection cable (Vaisala item code 272075SP)
The handle of HMP80 probes has a detachable end piece that protects the probe's cable connector from mechanical stress, such as physical shocks.
To detach the end piece of the handle and expose the cable connector:
1. Hold the probe firmly in your hands.
2. Grip the end piece of the handle and twist it counterclockwise to detach it.
3. Pull out the end piece.
4. You can now attach a probe connection cable in the HMP80 probe's cable connector.
Before attaching the cable, make sure to insert the cable through the end piece first.
5. Position the end piece against the probe and twist it clockwise to secure it into place.
16
Page 17
Chapter 4 – Attaching cable to probe
The probe is delivered with a yellow cap protecting the filter. Leave the protection cap on during storage, but remove it before taking the probe into use.
17
Page 18
HMP80 Series User Guide M212903EN-A

5. Using probe with Indigo80 handheld indicator

5.1 Indigo80 handheld indicator

Figure 4 Indigo80 handheld indicator
Vaisala Indigo80 Handheld Indicator is a portable diagnostics tool that accommodates up to two Vaisala Indigo-compatible probes or transmitters for measuring a wide range of parameters.
With the indicator, you can:
• See real-time measurements and device and status information
• Log measurement data
• Calibrate and adjust the probe
Configure probe features and settings such as condensation prevention, compensation setpoints, sensor purge, and filtering factor. The available features and settings depend on the probe model and firmware version.
Accessing certain configuration options for your probe requires using the free Insight PC software, downloadable at www.vaisala.com/insight.
The help tours in the indicator's user interface guide you through the key features of the indicator. You can access the tours in the Help menu by pressing the
For more information on using the indicator, for example, editing the measurement views and performing data logging, see Indigo80 User Guide (M212722EN).
18
button.
Page 19
1
3
2
1
1
2
3
4
5
Chapter 5 – Using probe with Indigo80 handheld indicator

5.1.1 Indigo80 keypad

Figure 5 Indigo80 keypad
1 Power on/o and main menu button 2 Arrow buttons for navigating menus
and scrolling views
3 Select button for selecting items in the
user interface

5.1.2 Indigo80 main menu

Pressing the button while navigating the Indigo80 menus and views opens the main menu.
Figure 6 Indigo80 main menu
1 Devices menu contains, for example,
options related to sensor purge, calibration, and environment settings (depending on the connected device).
2 Data logging menu for setting logging
interval and duration, and browsing data files.
3 Notifications menu displays
notifications related to Indigo80 and the connected devices.
4 Indigo80 menu for changing the
settings of Indigo80 (for example, date, time, and language) and viewing device information.
5 Help menu contains a Getting started
tour showing the key features of Indigo80, as well as instructions for sending devices to Vaisala for calibration and maintenance.

5.2 Connecting probes to Indigo80

• Probe connection cable
Up to two Vaisala Indigo-compatible probes or transmitters can be connected to the ports located on the bottom of Indigo80. You can connect and disconnect devices both when the indicator is powered on and when it is o.
19
Page 20
1 1
2
3
4
5
HMP80 Series User Guide M212903EN-A
Vaisala recommends using cables provided by Vaisala when connecting devices to the indicator. Cables and other accessories are available to order at store.vaisala.com.
Figure 7 Connecting probes to Indigo80
M12-5F ports on the bottom of Indigo80 for connecting compatible Vaisala devices.
1
Ports are labeled and .
2 M12-5M A-coded cable connector 3 M12-5F A-coded cable connector 4 HMP80 probe, displayed as by Indigo80 5 HMP7 probe, displayed as by Indigo80
1. If the indicator is powered and no devices are connected to it, the text Please connect a probe will be shown on the display.
20
Page 21
Chapter 5 – Using probe with Indigo80 handheld indicator
2. Insert the probe connection cable in one of the ports on the bottom of the indicator.
Note the orientation of the cable connector when inserting it
Hold the connector in place while turning its locking ring clockwise –
never twist the connector body!
3. Connect the probe to the M12-5F end of the probe connection cable.
When the indicator recognizes the connected probe, it shows a notification on the display (for example, HMP80 connected). A probe connected to the leftmost port in the indicator is labeled
4. To change probes, simply detach the cable from the probe and connect a new probe.
on the indicator's display, while the probe in the rightmost port is labeled .
5.3

 Measurement location

When taking measurements with the probe, consider the following:
• Verify the operating environment specification of the probe model. The probe head typically has a much wider operating temperature range than the probe handle.
21
Page 22
1
2
HMP80 Series User Guide M212903EN-A
Figure 8 Example installation
1 Mount the probe head horizontally to prevent any water condensing on the probe head
from running to the sensors
2 Attach Indigo80 to a metallic surface using the optional magnetic hanger accessory
(Vaisala item code ASM214318SP).
5.3.1 Configuring environmental compensations
After installation, configure the pressure compensation setting, if relevant to your probe and its measurement environment.
For example, when you have installed the probe in a process with a pressure diering from normal atmospheric pressure, update the correct pressure into the pressure compensation setpoint parameter or register of the probe. This allows the probe to apply the appropriate pressure compensation into its measurement results.
You can configure the setting with the Insight software or the Indigo80 handheld indicator.
5.4
Configuring probe features with Indigo80
handheld indicator
• Probe connection cable
You can use the Indigo80 handheld indicator to configure probe features such as compensation setpoints, sensor purge settings, and filtering factor. The available features depend on the probe model and firmware version.
22
Page 23
Chapter 5 – Using probe with Indigo80 handheld indicator
Accessing certain configuration options for your probe requires using the free Insight PC software, downloadable at www.vaisala.com/insight.
1. Connect the probe(s) to the indicator.
2. Open the main menu by pressing
3. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
4. Select Settings to access and change the features available for your probe.
5. Exit the menu by pressing or return to the main menu by pressing .
.
5.4.1 Example of configuring filtering factor with Indigo80 handheld indicator
• Probe connection cable
The filtering factor allows averaging the output if the measuring environment produces occasional exceptionally high or low readings.
This example shows how to configure the filtering factor as 0.9 using the Indigo80 handheld indicator.
This means that the new measurement output is a combination of the latest measurement (90 %) and the previous measurement (10 %).
1. Connect the probe(s) to the indicator.
2. Open the main menu by pressing
3. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
4. Select Settings > Filtering.
5. Select Filtering factor.
a. Use the arrow buttons to set the value as 0.9.
b. Select Set.
6. Select Extended filtering on/o > On to enable measurement filtering using the filtering factor defined in the previous step.
7. Exit the menu by pressing
More information
Filtering factor (page 15)
.
or return to the main menu by pressing .
23
Page 24
HMP80 Series User Guide M212903EN-A
5.5 Configuring environment settings with Indigo80 handheld indicator
• Probe connection cable
For optimal measurement accuracy, you can use the Indigo80 handheld indicator to give additional information about the connected probe’s measurement environment.
The environment settings available via Indigo80 correspond to the environmental compensation settings of the probe. Therefore, Environment settings and Settings > Compensation setpoints are interconnected: the configuration set in either menu is applied to both.
1. Connect the probe(s) to the indicator.
2. Open the main menu by pressing .
3. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
4. Select Environment settings to access the settings for your probe.
The available environment settings depend on the probe and probe firmware version.
Pressure
Value for pressure compensation. The entered value is persistent and remains unchanged at device restart.
Temperature
If any value is entered in this field, the device will use it as its temperature measurement value instead of its own sensor reading. This is needed when sensor warming is used, as it will also heat the temperature sensor of the device. At device restart, the value is cleared.
24
The temperature value needs an external source, for example, another probe that is able to provide the value constantly in real time, and the value needs to correspond to the temperature of the measurement environment as accurately as possible. Thus, set the temperature value only if you have another probe connected to the indicator from which you can read the value. Do not set the temperature value manually.
Page 25
Chapter 5 – Using probe with Indigo80 handheld indicator
5. Select a setting from the list, for example, Pressure.
a. Select the unit, if prompted.
b. Use the arrow buttons to set the desired value, and then select OK or Set.
You can also select to read the value from another probe connected to the indicator, provided that the probe can output the required compensation parameter. Press to highlight the Read value from field, and then press
to select the other probe as the
source of the value.
For Temperature, always set the temperature value to be read from another probe. Do not set the temperature value manually. Make sure that both probes are in the same environment to get the most accurate measurement values.
6. Exit the menu by pressing or return to the main menu by pressing .
More information
Environmental compensation (page 14)
25
Page 26
HMP80 Series User Guide M212903EN-A
6. Configuration with Insight software

6.1 Vaisala Insight PC software

Vaisala Insight PC software is a configuration software for Indigo-compatible probes and other supported devices. Insight is available for Microsoft Windowsâ operating systems (64-bit only).
With the Insight software, you can:
• See probe information and status
• See real-time measurement data
• Record data for up to 48 hours and export in CSV format
Configure probe features such as measurement filtering, sensor purge, and condensation prevention
• Calibrate and adjust the probe
Download Vaisala Insight software at www.vaisala.com/insight.
6.2

 Connecting to Insight software

• Computer with a Microsoft Windowsâ operating system (64-bit version) and Vaisala Insight PC software installed
• Indigo USB adapter (item code USB2) or USB connection cable (item code
242659)
CAUTION!
computer may not be able to supply enough power through its USB ports. Use an externally powered USB hub that can supply >2 W for each port.
1. Open the Insight software on your PC.
2. Verify the operating mode of Insight from the Settings menu and change it if appropriate.
Basic mode is suitable for most use cases
Advanced mode provides access to additional configuration options. Use Advanced mode only when instructed to do so by product documentation or Vaisala technical support.
3. Connect the USB adapter to a free USB port on the PC or USB hub.
4. Connect the probe to the USB adapter.
5. Wait for the Insight software to detect the probe.
26
When connecting several devices at the same time, note that your
Page 27
Chapter 6 – Configuration with Insight software
6.3 Configuration options
The complete probe configuration is accessible via the Insight software. After connecting the probe, select > Configure device to access the configuration options.
If Insight has been set to Advanced mode, additional configuration options may be available. Use Advanced mode only when instructed to do so by product documentation or Vaisala technical support.
Figure 9 HMP80L in Insight software
More information
Restoring factory default settings with Insight software (page 40)
6.4

 Diagnostics in Insight

Messages sent by the connected devices are shown automatically in Insight. Additional diagnostic data is available on the Diagnostics page. The data available depends on the probe model and Insight operating mode (basic or advanced). The data can be very useful when diagnosing issues together with Vaisala support, particularly the diagnostic files.
Diagnostic files
If Insight is in Advanced mode, you can retrieve the following diagnostic files from the connected probe:
SSR/Ts histogram: Table of humidity and temperature conditions measured by the probe
Error log: Cumulative total of the error and status events tracked by the probe
Both files contain data from since the probe was manufactured, or since the files were last cleared. Clearing the files is not recommended unless instructed by Vaisala support. The buttons to clear the files are located at the bottom of the Diagnostics page.
27
Page 28
HMP80 Series User Guide M212903EN-A
The files are in comma separated value (CSV) format. They can be opened for viewing in spreadsheet programs such as Microsoft Excel or in text editors.
Figure 10 HMP80L diagnostics in Insight (advanced mode)
28
Page 29
Chapter 7 – Calibration and adjustment

7. Calibration and adjustment

7.1 Calibration and adjustment overview

Calibration means comparing the measurement output of the device to a known reference, such as a known environment in a calibration chamber or the output of a reference instrument. Correcting the reading of the device so that it measures accurately is referred to as adjustment.
The probe is fully calibrated and adjusted as shipped from the factory. To maintain the accuracy of the measurement, calibrate and adjust the probe as needed. Typical calibration interval is one year, but depending on the application it may be necessary to check the accuracy more frequently.
Vaisala recommends sending the probe regularly to Vaisala for factory calibration. The probe is calibrated and adjusted at multiple points along its measurement range. Factory calibration always includes a traceable calibration certificate and adjustment, if needed.
Factory calibration leads to highest accuracy and lowest uncertainty. After factory calibration the instrument performs as new. For up-to-date information about calibration points and oering, visit store.vaisala.com.
WARNING!
has not been exposed to dangerous contamination, and is safe to handle without special precautions.
It is also possible to calibrate and adjust the probe on the field using the Insight PC software or the Indigo80 handheld indicator.
If you think the device is not measuring correctly, calibration and adjustment is not the first thing to do. Check the following first:
• Make sure nothing is interfering with the measurement: heat sources, temperature dierences, or condensation.
• Check that there is no moisture on the probe. If the sensor has become wet, wait for it to dry.
• Always wait for the measurement to stabilize.
When returning a product for calibration or repair, make sure it

7.2 Adjustment points and requirements

You can adjust the humidity measurement in 1 … 5 points and temperature measurement in 1...2 points, depending on the device that you use for the adjustment. Note the following:
• Humidity adjustment in more than 2 points is only available when you use the Insight software in Advanced Mode.
29
Page 30
HMP80 Series User Guide M212903EN-A
• If you are adjusting in more than one humidity point, make sure the first two points are at least 10 %RH apart.
• The probe will reject adjustments that are too large, that is, greater than 10 %RH for humidity and 0.5 °C for temperature. If the probe appears to need such a large correction, perform a sensor purge and repeat the adjustment procedure. Make sure that the measurement has stabilized and the reference environment is reliable. If the required adjustment is still too large, the probe needs to be serviced by Vaisala.
When adjustment of humidity measurement is necessary, Vaisala recommends adjusting in two points, 11 %RH and 75 %RH. These humidities can be produced using Vaisala HMK15 Humidity Calibrator. Adjustment of temperature measurement is typically not necessary.
More information
Calibration and adjustment with Indigo80 handheld indicator (page 30)
Adjusting measurement with Insight software (page 34)
HMK15 Humidity Calibrator with HMP80 series probes (page 35)
7.3
 Calibration and adjustment with Indigo80
handheld indicator
The Indigo80 handheld indicator provides two ways to calibrate and adjust measurements for probes: Guided calibration and Manual adjustment.
Figure 11 Indigo80 calibration methods for probes
• Choose Guided calibration when you want the indicator to guide you through the calibration process step by step. At the end you can also adjust the measurement, if needed.
• Choose Manual adjustment when you want to freely set the reference and measured values to adjust the measurement.
30
Page 31
Chapter 7 – Calibration and adjustment
Calibrating a measurement with the indicator, including purge and adjustment, typically takes 1 ... 2 hours depending on the number of calibration points.

7.3.1 Performing guided calibration with Indigo80 handheld indicator

• Probe connection cable
• Reference environment(s) for producing the desired reference values
In guided calibration, the indicator guides you through the calibration process step by step. At the end, you can also adjust the measurement, if needed.
You can choose to do either 1-point or 2-point calibration, and choose to use the reference environment or instrument of your choice, for example, another probe connected to your indicator.
For relative humidity, you can choose to use Vaisala HMK15 Humidity Calibrator. For more information on HMK15, see HMK15 User Guide (M210185EN).
1. Connect the probe(s) to the indicator.
2. If your probe supports the sensor purge feature, wait for the start-up purge to finish.
3. Open the main menu by pressing .
4. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
5. Select Calibrate / Adjust.
6. Select the parameter, for example, Relative humidity.
7. Select Guided calibration.
8. In Calibration points, select the number of the calibration points: 1 point or 2 points.
9. In Reference, select the desired reference environment(s) or reference instrument(s).
Selection Other reference allows you to use any reference environment(s) of your choice. For 2-point calibration of relative humidity, you can select two salts from the list, one for each calibration point. When you have selected the reference environments for 2-point calibration, press get back to the previous view.
10. Select Start calibration.
The indicator guides you through the calibration process from this point onwards. Press to proceed from one view to another.
to
31
Page 32
HMP80 Series User Guide M212903EN-A

7.3.2 Performing manual adjustment with Indigo80 handheld indicator

• Probe connection cable
• Reference environment(s) for producing the desired reference values
This procedure tells you how to use the indicator to adjust the probe's measurement in 1 ... 2 points.
1. Connect the probe(s) to the indicator.
2. Open the main menu by pressing .
3. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
4. If your probe supports the sensor purge feature, wait for the start-up purge to finish or start the purge manually by selecting Devices > <your device> > Sensor purge.
After the purge has been completed, continue to the next step.
5. Select Calibrate / Adjust.
6. Select the parameter, for example, Relative humidity.
7. Select Manual adjustment.
8. Place the probe to the first reference point and select Reference value, point 1 to open the stabilizing graph view. Wait for the measurement to stabilize fully.
The stabilization typically takes at least 10 ... 15 minutes.
a. When the reading has stabilized, press
b. Use the arrow buttons to set the value of the reference, and then select OK or Set.
.
If you already have the reference and measured values available for your calibration point(s), you can just skip the stabilizing graph view with the button to get to the view where you can set the reference value. Then replace the automatically filled Measured value, point 1 and Measured temperature, point 1 with your actual values.
9. If you want to adjust the measurement in 2 points, place the probe to the second reference point. Then scroll down the <parameter> adjustment view and select Reference value, point 2 to open the stabilizing graph view. Wait for the measurement to stabilize fully.
a. When the reading has stabilized, press
b. Use the arrow buttons to set the value of the second reference, and then select OK or
Set.
32
.
Page 33
Chapter 7 – Calibration and adjustment
10. In <parameter> adjustment view, check the dierence between each reference and measured value. Very large dierences may be due to insucient stabilization time or unsuitable calibration setup.
• To store the adjustments to the probe, scroll down and select Activate adjustment > OK. Confirm by selecting OK.
• To exit without taking the adjustment in use, press to end calibration.
11. Update the calibration information in Calibrate / Adjust > Calibration information. Navigate to the Calibration information menu by pressing
12. Press
to end calibration.
.

7.3.3 Restoring factory adjustment with Indigo80 handheld indicator

• Probe connection cable
If needed, you can restore the factory adjustment for your probe using the Indigo80 handheld indicator.
Note that restoring the factory adjustment is not necessary before a new adjustment: you can just perform the new adjustment on top of the old adjustment. However, you can select to do the restore procedure if your adjustment fails for some reason, or if you want to return to the original factory adjustment.
1. Connect the probe(s) to the indicator.
2. Open the main menu by pressing
3. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
4. Select Calibrate / Adjust.
5. Select any of the available parameters to access the view where you can restore the factory adjustment.
6. Scroll down and select Restore factory adjustment > OK.
7. Confirm by selecting OK.
8. Press
to end calibration.
.
33
Page 34
HMP80 Series User Guide M212903EN-A

7.4 Adjusting measurement with Insight software

• Computer with a Microsoft Windowsâ operating system (64-bit version) and Vaisala Insight PC software installed
• Indigo USB adapter (item code USB2) or USB connection cable (item code
242659)
• Reference environment(s) for producing the desired humidity and/or temperature
This procedure can be used to adjust the probe's humidity or temperature measurement. If you want to adjust both, repeat the procedure.
Because stabilization of temperature and humidity takes time, you should expect the adjustment procedure to take at least 30 minutes for each adjustment point.
1. Connect the probe to the Insight software.
2. If you intend to adjust humidity measurement, first select > Sensor purge to perform sensor purge to condition the sensor. Wait a few minutes for purge to complete.
3. Select
In calibration mode, the device will not use functions that may interfere with calibration and adjustment.
4. Select the type of adjustment to perform: RH adjustment or T adjustment.
5. Define the needed adjustment for the first measurement point:
6. If you want to adjust in more than one point, repeat the previous step for all desired calibration points.
> Calibrate > Yes to switch the probe to calibration mode.
a. Insert the probe head in the reference environment for the first calibration point.
b. Wait for the measurement to stabilize fully. This may take more than 30 minutes.
Monitor the readings to see when both the RH and temperature measurements have stabilized.
c. Click the Reference value, point 1 text box and enter the reference value of the
calibration point. Press ENTER or click outside the text box when done.
d. The probe automatically enters the measured values for the calibration point.
You can adjust humidity measurement in up to five points when Insight is set to Advanced mode.
7. Check the dierence between each reference and measured value. Very large dierences may be due to insucient stabilization time or unsuitable calibration setup.
• To save the adjustment into the device, select Activate adjustment.
• To exit without taking the adjustment in use, select Close.
34
Page 35
Chapter 7 – Calibration and adjustment
8. Check the message that appears at the top of the screen. If the message indicates that the adjustment is activated successfully, your adjustment is stored in the probe.
9. Select the Calibration information tab and update the Calibration date and Calibration text.
10. Select Close to exit the calibration mode.

7.5 HMK15 Humidity Calibrator with HMP80 series probes

CAUTION!
filter is removed. Handle the probe carefully and avoid touching the sensors.
Vaisala HMK15 Humidity Calibrator is a suitable humidity reference for HMP80 series probes. When preparing the salt solutions and using the calibrator, follow the general instructions in the HMK15 User Guide (M210185EN) to avoid common sources of calibration error.
HMP80 series specific considerations:
HMP80L probe model
• Use calibration adapter 211302 to prevent the sensor from being submerged in the salt solution. Adapter 211302 is delivered with the HMK15 calibrator.
• With the filter removed and the adapter attached, place the probe in one of the Ø13.5 mm holes in the HMK15 universal lid.
• Make sure the probe is well supported when using it with the HMK15 calibrator.
HMP80N probe model
• The sensor in the HMP80N is short, so a calibration adapter is not needed.
• With the filter removed, place the probe in the Ø12 mm hole in the HMK15 universal lid.
Sensors on the probe head are vulnerable to damage when the
Figure 12 Calibrating a humidity probe with HMK15 using 211302 adapter
35
Page 36
HMP80 Series User Guide M212903EN-A

8. Maintenance

8.1 Cleaning the probe

Do not attempt to clean the sensors under the filter.CAUTION!
Do not spray anything directly on the probe head, since that may deposit impurities on the sensors.
You can clean the probe, probe handle, and cable by wiping them with a soft, lint-free cloth moistened with water or a suitable cleaning agent, such as isopropyl alcohol. Do not wipe the filter: wiping the filter may block its pores and/or deposit residue on the filter. If the filter is heavily contaminated, replace it.
When cleaning, follow these precautions:
• Avoid touching the filter. If you need to touch the filter, always wear clean gloves (cotton, rubber or similar material). Keep the filter free of any grease or oil.
• Do not scrape the probe or the probe handle, and avoid abrasive cloths when wiping the surface.
• Do not immerse the probe or the probe handle in liquid to clean them.
• Wipe cleaning agents o the probe, probe handle, and the cable after cleaning.
After cleaning the probe, it is recommended to perform sensor purge.

8.1.1 Chemical tolerance

Avoid exposing the probe to cleaning agents for unnecessarily long periods of time.
The following chemicals can be used to clean the probe:
• Deionized water
• Isopropyl alcohol (70 %).
More information
HMP80 specifications (page 42)
36
Page 37
8.2 Changing the probe filter
• New compatible filter
• Clean lint-free gloves
Chapter 8 – Maintenance
CAUTION!
the probe head carefully.
1. Put on clean gloves before touching the filter.
2. Turn the filter counter-clockwise to loosen it.
3. Remove the filter from the probe head. Be careful not to touch the sensors with the filter.
4. Install a new filter on the probe head. Tighten the filter properly.
Sensors are easily damaged when the filter is not in place. Handle
37
Page 38
HMP80 Series User Guide M212903EN-A

9. Troubleshooting

9.1 Problem situations

Table 5 Troubleshooting table
Problem Possible cause Solution
Measurement output seems incorrect
Values of measurement parameters stop changing for a few minutes
Installation location is not representative of actual conditions you want to measure
Heat conduction along probe shaft is interfering with measurement accuracy
Probe is in need of adjustment Calibrate and adjust the probe.
Probe is performing a sensor heating function such as sensor purge or waiting for the sensor to cool down
Verify the installation location and relocate the probe if necessary.
Follow the installation recommendations for cases when temperature of measured environment diers greatly from ambient. See
Measurement location (page 21).
Wait for the function to complete and measurement parameters to be available again.

9.2 Error messages

Probes can communicate with error messages when they are connected to Indigo80 indicator or Insight software. The messages are categorized according to the severity of the status:
Critical errors are fatal to the operation of the device. It may not be able to respond to communication at all, and will not measure correctly.
Errors prevent normal operation of the device. Depending on the problem, errors may resolve themselves. For example, a completely wet humidity sensor may cause a humidity measurement error.
Warnings do not prevent normal operation but may indicate possible problems.
Status indicates a known state of the device.
38
Page 39
Chapter 9 – Troubleshooting
Error message Description Recommended action
Critical errors
Firmware checksum mismatch [49]
The installed firmware is corrupted.
Contact Vaisala technical support.
Device settings corrupted [50] Parameter memory is
Additional configuration
corrupted.
settings corrupted [51]
Sensor coecients corrupted [52]
Main configuration settings corrupted [53]
Factory default settings corrupted [54]
Non-volatile memory read
Hardware fault.
write failure [57]
Errors
Temperature measurement error [44]
Humidity measurement error [45]
Humidity sensor failure [46]
Readings from sensor(s) are missing or out of range.
Inspect probe head and sensors visually. If the probe is completely wet, allow it to dry out.
If the sensors are damaged or missing and the error message(s) stay active, contact Vaisala to have the probe repaired.
Capacitance reference error [47]
Hardware fault. Contact Vaisala technical
support.
Sensor heater error [212]
General device failure [994] High number of measurement
errors, capacitance reference errors, or sensor heater errors written in the error log.
Ambient temperature out of range [48]
Probe handle is too hot or cold. Relocate the probe handle so
that its ambient temperature is within the specified operating range.
Supply voltage out of range [55]
Probe supply voltage is too high or low.
Check and correct the supply voltage.
39
Page 40
HMP80 Series User Guide M212903EN-A
Warnings
Calibration has expired [353] This warning is displayed by
Software restart [679] The probe has automatically
Unexpected device restart [216]
Status messages
Calibration is about to expire [352]
Purge in progress [287] Sensor purge of the humidity
The readings stay frozen until the sensor cools down [288]
More information
Technical support (page 51)
9.3
 Restoring factory default settings with Insight
the calibration reminder functionality if the calibration interval has been exceeded.
restarted itself.
This message is displayed by the calibration reminder functionality.
sensor is ongoing.
Probe is waiting for its sensors to cool down.
Calibrate the probe and update the calibration date information.
Check that supply voltage is stable and operating environment is within
specification.
Calibrate the probe and update the calibration date information.
Wait for sensor purge to complete.
Wait for measurement readings to become available.
software
If needed, you can restore the probe back to its factory default settings using the Insight software. Doing this will also clear any user adjustment and restore the latest adjustment performed by Vaisala.
1. Connect the probe to the Insight software.
2. Select
> Factory default settings > Restore settings.
3. Wait for the probe to be re-detected.
More information
Configuration options (page 27)
40
Page 41
Chapter 9 – Troubleshooting

9.4 Restoring factory default settings with Indigo80 handheld indicator

• Probe connection cable
If needed, you can restore the probe back to its factory default settings using the Indigo80 handheld indicator. The user adjustment of the probe and the settings or log files of the indicator itself are not aected.
1. Connect the probe(s) to the indicator.
2. Open the main menu by pressing
3. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
4. Select Factory default settings > Restore settings > OK.
5. Confirm by selecting OK.
6. Exit the menu by pressing
9.5

 Writing a problem report

When troubleshooting the product, write a problem report including:
• What failed (what worked / did not work)?
• Where did it fail (location and environment)?
• When did it fail (date, immediately / after a while / periodically / randomly)?
• How many failed (only one defect / other same or similar defects / several failures in one unit)?
• What was done when the failure was noticed?
• What was connected to the product and to which connectors?
• Input power source type, voltage, and list of other items (such as lighting, heaters, and motors) that were connected to the same power output.
• Are all parts connected and grounded properly? Take a photo to help the troubleshooting.
.
or return to the main menu by pressing .
41
Page 42
HMP80 Series User Guide M212903EN-A

10. Technical data

10.1 HMP80 specifications
Table 6 HMP80 series measurement performance
Property Description/Value
Relative humidity
Measurement range 0 … 100 %RH
Accuracy at +23 °C (73.4 °F)
Factory calibration uncertainty
T63 response time 15 s
Sensor HUMICAPâ R2C
Temperature
Measurement range HMP80N: −20 … +60 °C (−4 … +140 °F)
Accuracy at +23 °C (+73.4 °F) 1)
Factory calibration uncertainty
Sensor Pt100 RTD Class F0.1 IEC 60751
1)
2)
3)
2)
±0.8 %RH (0 … 90 %RH)
±0.5 %RH (0 … 40 %RH)
±0.8 %RH (40 … 95 %RH)
HMP80L: −50 … +120 °C (−58 … +248 °F), short­time measurement range
−50 … +180 °C (−58 … +356 °F)
±0.1 °C (±0.18 °F)
±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F)
1) Defined against calibration reference. Including non-linearity, hysteresis, and repeatability.
2) Defined as ±2 standard deviation limits. Small variations possible; see calibration certificate.
3) Exposing temperature sensor to temperatures below −20 °C (−4 °F) may cause permanent additional deviation of ±0.1 °C (0.18 °F).
42
Page 43
0.0
0.5
1.0
1.5
2.0
2.5
3.0
-20
-10
0
10
20
30
40
50
60
Accuracy [± %RH]
Temperature [
°
C]
3.5
4.0
90 ... 100 %RH
0 ... 90 %RH
Figure 13 HMP80N humidity measurement
0
0.1
0.2
0.3
0.4
-20
-10
0
10
20
30
40
50
60
Temperature [
°
C]
Accuracy [±
°C]
0.0
0.5
1.0
1.5
2.0
2.5
3.0
Accuracy [± %RH]
Temperature [
°
C]
3.5
4.0
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
90 ... 100 %RH
0 ... 90 %RH
0
0.1
0.2
0.3
0.4
Accuracy [±
°C]
Temperature [°C]
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
-50
accuracy as a function of temperature
Figure 15 HMP80L humidity measurement accuracy as a function of temperature
Table 7 HMP80 series operating environment
Chapter 10 – Technical data
Figure 14 HMP80N temperature measurement accuracy over full range
Figure 16 HMP80L temperature measurement accuracy over full range
Property Description/Value
Operating temperature of probe body −10 … +60 °C (−14 … +140 °F)
Operating temperature of probe head HMP80N: −20 … +60 °C (−4 … +140 °F)
HMP80L: −50 … +120 °C (−58 … +248 °F)
Storage temperature −20 … +60 °C (−4 … +140 °F)
Measurement environment For air, nitrogen, hydrogen, argon, helium,
oxygen, and vacuum
1)
Chemical tolerance Temporary exposure during cleaning:
• Deionized water
• Isopropyl alcohol (70 %)
IP rating of probe handle:
with probe connection cable connected to the probe
IP66: Dust-tight. Protected from powerful water jets from any direction.
43
Page 44
HMP80 Series User Guide M212903EN-A
Property Description/Value
without cable IP55: Limited protection against dust. Protected
from water jets from any direction.
1) Consult Vaisala if other chemicals are present. Consider safety regulations with flammable gases.
Table 8 HMP80 series inputs and outputs
Property Description/Value
Operating voltage 15 … 30 V DC
Current consumption 10 mA typical, 500 mA max.
Digital output RS-485, non-isolated
Table 9 HMP80 series output parameters
Output parameter name and unit
Absolute humidity (g/m3)
Absolute humidity at NTP (g/m3)
Relative humidity (%RH)
Relative humidity (dew/frost) (%RH)
Dew point temperature (°C) Temperature (°C)
Dew/frost point temperature (°C) Water concentration (ppmv)
Dew/frost point temperature at 1 atm (°C) Water concentration (wet basis) (vol%)
Dew point temperature at 1 atm (°C) Water mass fraction (ppmw)
Dew point temperature dierence (°C) Water vapor pressure (hPa)
Enthalpy (kJ/kg) Water vapor saturation pressure (hPa)
Mixing ratio (g/kg) Wet-bulb temperature (°C)
Table 10 HMP80 series mechanical specifications
Property Description/Value
Connector type M12 5-pin A-coded male
Weight HMP80N: 200 g (7 oz)
HMP80L: 300 g (10 oz)
Materials
Probe handle Polyamide (PA) and thermoplastic elastomer
(TPE)
Probe shaft Stainless steel (AISI 316L)
44
Page 45
103 [4.1] 83 [3.3]
23
63 [2.5]
272 [11]
[0.91]
25
47 [1.9]
[0.98]
12 [0.5]
mm [in]
103 [4.1] 278 [11]
41
63 [2.5]
485 [19]
[1.6]
12 [0.5]
Chapter 10 – Technical data
Property Description/Value
Filters
HMP80N: Stainless steel (AISI 316L)
HMP80L: Porous stainless steel (AISI 316L)
1) With holes and without an additional filter membrane. Vaisala item code of filter: DRW255306SP.
2) Vaisala item code: HM47280SP
Table 11 HMP80 series compliance
Property Description/Value
EU directives and regulations EMC Directive (2014/30/EU)
RoHS Directive (2011/65/EU) amended by 2015/863
Electromagnetic compatibility (EMC) EN 613261, industrial environment
Rough handling
IEC 60068231
(excluding sensor inside probe head)
Compliance marks CE, China RoHS, RCM, UKCA

10.2 Dimensions

1)
2)
Figure 17 Dimensions of HMP80L (top) and HMP80N (bottom), side and bottom view
45
Page 46
HMP80 Series User Guide M212903EN-A

10.3 Spare parts and accessories

Information on spare parts, accessories, and calibration products is available online at www.vaisala.com and store.vaisala.com.
Table 12 HMP80 series spare parts and accessories
Property Description/value
Probe connection cable (M12-M12), 1.5 m (4.11 ft) 272075SP
Flat cable for probes (M12-M12), 1.0 m (3.4 ft) CBL210493SP
Indigo USB adapter USB2
Sintered stainless steel filter (HMP80N, HMP80L) HM47280SP
Plastic PPS grid filter (HMP80N, HMP80L) DRW010276SP
PPS grid with SS netting (HMP80N, HMP80L) DRW010281SP
Slotted MIM filter (HMP80N) DRW255306SP
Slotted MIM filter with membrane (HMP80N) ASM214606SP

10.4 Accuracy of calculated output parameters

HMP80 series probes measure relative humidity and temperature, and calculate the other output parameters from those two parameters. Accuracy of the calculated parameters is derived from the accuracy of the original RH and T measurements.
Use the Vaisala Humidity Calculator to convert the accuracy specification of RH measurement for other humidity parameters. You can download it from www.vaisala.com.
46
Page 47
Chapter 10 – Technical data
Figure 18 Vaisala Humidity Calculator in Advanced mode
47
Page 48
Metal waste
Plastic
Electrical and electronic waste
1
4
6
5
7
2
3
HMP80 Series User Guide M212903EN-A

Appendix A. Recycling instructions

These recycling instructions guide you on the end-of-life treatment of this Vaisala product. As waste regulations and infrastructure vary in each country, these instructions only indicate the dierent components to be separated and common ways to handle them. Always follow local requirements when disposing of the product. Vaisala encourages to use the best available recycling practices to minimize related environmental impacts.
Vaisala is committed to meeting the requirements of the EU Waste Electrical and Electronic Equipment (WEEE) Directive. This directive aims to minimize the impact of electrical and electronic goods on the environment, by increasing reuse and recycling, and reducing the amount of WEEE going to landfill. This symbol indicates that the product should be collected separately from other waste streams and treated appropriately.
Figure 19 Materials for recycling
Table 13 Materials for recycling
Part Material Recycling
1 Filter Stainless steel (AISI 316L) Metal waste
2 Sensor Various materials Electrical and electronic
waste
48
Page 49
Appendix A – Recycling instructions
Part Material Recycling
3 Probe shaft Stainless steel (AISI 316L) Metal waste
4 Probe handle Polyamide (PA), glass reinforced
5 Electronics
6 Screw Stainless steel (AISI 316L) Metal waste
7 Component board
More information
housing
and wiring
Probe structure (page 11) HMP80 specifications (page 42)
and thermoplastic elastomer (TPE)
Polyamide (PA), glass reinforced Dispose of or recycle as
Various materials Electrical and electronic
Dispose of or recycle as plastic according to local waste management practices and regulations
plastic according to local waste management practices and regulations
waste
49
Page 50
Page 51

Maintenance and calibration services

Vaisala oers comprehensive customer care throughout the life cycle of our measurement instruments and systems. Our factory services are provided worldwide with fast deliveries. For more information, see www.vaisala.com/
calibration.
• Vaisala Online Store at store.vaisala.com is available for most countries. You can browse the oering by product model and order the right accessories, spare parts, or maintenance and calibration services.
• To contact your local maintenance and calibration expert, see
www.vaisala.com/contactus.

Warranty

For standard warranty terms and conditions, see www.vaisala.com/warranty. Please observe that any such warranty may not be valid in case of damage due to normal wear and tear, exceptional operating conditions, negligent handling or installation, or unauthorized modifications. Please see the applicable supply contract or Conditions of Sale for details of the warranty for each product.

Technical support

Contact Vaisala technical support at helpdesk@vaisala.com. Provide at least the following supporting information as applicable:
• Product name, model, and serial number
• Software/Firmware version
• Name and location of the installation site
• Name and contact information of a technical person who can provide further information on the problem
For more information, see www.vaisala.com/support.

Recycling

Recycle all applicable material according to local regulations.
51
Page 52
Page 53
Page 54
www.vaisala.com
Loading...