mechanical (including photocopying), nor may its contents be communicated to a third
party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for
Vaisala towards the customer or end user. All legally binding commitments and
agreements are included exclusively in the applicable supply contract or Conditions of
Sale.
HMI41 AND PROBES
M210316EN-AOperating Manual
Contents
1. THE HMI41 INDICATOR AND PROBES...............................................................................1
2. TO BE NOTED IN THE M EASUREMENT OF HUM IDIT Y.....................................................3
The HMI41 is an easy-to-use portable humidity and temperature indicator for
a variety of applications, including such as industrial monitoring and
inspections, occupational health and safety, laboratory and research use, spot
checking etc. When equipped with optional calibrati on cables, the HMI41 can
also be used as a field calibrator for most Vaisala transmitters.
There are six different probe types that can be used with the HMI41 indicator.
The indicator recognizes the probe type automatically, so there is no need to
change settings each time the probe is changed. However, please, note that this
feature is active only in indicators and probes with letters ID in the instrument
label. For previous versions, the probe type has to be set manually. The
HMP44/L is an exception, as indicators marked with ID recognize automatically all versions of this probe. All probe types are optimized for different
applications:
•HMP41 probe can be used for measuring humidity and temperature in
various applications, for example in spot checks.
•HMP42 probe head has a diameter of only 4 mm and is 23.5 cm long.
This probe structure is specially indicated for measurements in very
tight places, e.g. in joint spaces between tiles or in air conditioning
channels as well as for measuring the equilibrium humidity of e.g.
timber (separate manual).
•HMP44 and HMP44L are used for measuring humidity in concrete
and other structures (separate manual).
•HMP45 probe is indicated for measurements in channels and other
places that are difficult to reach and therefore require a probe head
with cable.
•HMP46 consists of a 32 cm long probe head of stainless steel; the
HMP46 is optimized for measurements in relatively high temperatures
(up to +100°C, temporarily even +180 °C), in dirty processes and in
general in applications that require a robust probe structure.
1
HMI41 AND PROBES
Operating Manual M210316EN-A
The HMI41 indicator displays relative humidity, temperature and dewpoint
temperature readings. In addition to these, one of the following quantities can
also be chosen: absolute humidity, wet bulb temperature or mixing ratio.
The indicator also features an automatic power-off function which can be
disabled, and a continuously updated display. The display can also be frozen
to show the current readings, and it can be used for checking the minim um and
maximum readings measured during data collecting. The automatic power-off
function is not active during data collecting even if it was previously selected.
The versatile HMI41 indicator also includes a data collecting feature. Data
collecting can be either automatic or manual, and it can be optimized for each
application: both the measurement interval and duration can be set by the user.
During automatic data collecting, the probe takes measurements only just
before storing each measurement. In order to minimize the consumption and to
maximize the battery life, the power is automatically turned off for the
measurement interval and the display is dim except when the readi ngs on the
display are updated (once a minute). If required, the collected data can be
transferred to a PC; an optional serial interface cable is available for this
purpose (order code 19446ZZ).
The humidity measurement range is 0...100 %RH. The temperature
measurement range depends on the probe used. Relative humidity is measured
with the accurate and stable HUMICAP180 humidity sensor which uses an
operating principle based on the changes in the capacitance of the sensor as its
thin polymer film absorbs water molecules.
2
HMI41 AND PROBES
M210316EN-AOperating Manual
2.TO BE NOTED IN THE MEASUREMENT OF HUMIDITY
In the measurement of humidity and especially in calibration it is essential that
temperature equilibrium is reached. Even a small difference in temperature
between the measured object and the sensor causes an error. If the t emperature
is +20 °C (+68 °F) and the relative humidity is 50 %RH, a difference of +1 °C
between the measured object and the sensor causes an error of +3 %RH. When
the humidity is 90 %RH, the corresponding error is + 5.4 %RH.
The error is at its greatest when the sensor is colder or warmer than the
surroundings and the humidity is high. Although the humidity sensor reacts
rapidly to changes in the amount of water vapour in the air, the probe
temperature changes more slowly. To avoid errors caused by temperature
differences the probe must always be left to stabilize to ambient temperature
before starting measurements: the bigger the temperature difference, the
longer the stabilization time.
Indoors relative humidity should be measured in a place where the temperature
is as close to the average temperature of the room as possible. Measurements
taken close to heat sources will not give a true picture of the relative humidity
in the whole room.
10
9
8
7
6
5
4
dRH (%RH)
3
2
1
0
-40-200 20406080100
Temperature (°C)
Figure 2.1Measurement error at 100 %RH when the dif ference
between the ambient and the sen sor temperature is 1 °C
3
HMI41 AND PROBES
Operating Manual M210316EN-A
3.GETTING STARTED
3.1Inserting the batteries
When taking the HMI41 indicator into use, first insert the batteries (4 pcs of
size AA [LR6] batteries). Open the lid on back of the device and insert the
batteries as indicated in the housing. Then close the lid carefully.
3.2Connecting the probes
After having inserted the batteries, connect the probe you wish to use. The
HMP41 is connected to the connector on top of the HMI41: remove the plastic
plug and mount the probe (see Figure 3.2). The HMP45 and the HMP46 both
have a cable to be plugged into the connector at the bottom of the HMI41,
marked PROBE (see Figure 3.2). The other connector is for calibration cables.
HMP45 and HMP46
cable
Figure 3.2Connecting the probes to the HMI41
HMP41
4
HMI41 AND PROBES
M210316EN-AOperating Manual
4.TAKING MEASUREMENTS
4.2Starting the measurements
Before starting the measurements with the HMI41 and appropriate probe,
remember to allow enough time for stabilization.
Turn the power on with the ON/OFF button and the following appears:
RH
Td
T
P
set min maxhyst hold Lo bat Hi
%
°C °F
°C
°F
Pahg/m
Within a couple of seconds, the display changes to show the software version
and then the probe type indication (41.45 or 42.46):
NOTE: if the following appears, check that the probe is correctly connected:
If the software version does not appear, it is smaller than 1.02 and the
indicator cannot be used with the HMP46 probe. If the probe type does not
appear, the indicator or the probe is of a previous version and the automatic
recognition is not possible; give the setting manually (see Chapter 8).
After a couple of seconds, the battery voltage appears on the display with an
indication of the battery charge (high or low):
Hi
bat
If the battery voltage is higher than 4.75 V, the text on the lower righthand
corner is “bat HI” and after a few seconds, the HMI41 automatically displays
the RH and T readings. If the voltage is 4.65...4.75 V, the text is “Lo bat” and
the batteries should be replaced (see Chapter 3.1). If the voltage is lower than
5
HMI41 AND PROBES
Operating Manual M210316EN-A
4.65 V, the indicator turns itself off to prevent erroneous measurements and
readings. Should this happen, replace the batteries.
4.2Measurement readings on display
Relative humidity and temperature readings appear automatically after the
battery charge indication:
RH
T
%
°C
If MODE is pressed, dewpoint temperature (Td) reading is displayed:
Td
T
°C
°C
When MODE is pressed again, one of the following is displayed or the HMI41
returns to showing the RH and T readings according to what has been chosen
for the measurement quantities (see Chapter 8.4):
T
°C
abs
T
g/m
Temperature and absolute humidity
°C
°C
Temperature and wet bulb temperatu re; the
arrow in the lower righthand corner indicates that
wet bulb temperature has been chosen
T
°C
Temperature and mixing ratio; the arrow in the
lower righthand corner indicates that mixing ratio
has been chosen (the unit is g/kg or gr/lb).
6
HMI41 AND PROBES
M210316EN-AOperating Manual
4.3HOLD-mode
In any of the previously mentioned measurement reading displays, the button
HOLD freezes the display to show the current readings, e.g. RH & T readings:
RH
T
%
°C
hold
With MODE or ENTER, the display returns to the normal display mode.
If the indicator turns itself off with the automatic power-off function during
the HOLD-mode, it wakes up in the same mode when turned on again. The
'hold'-text is blinking and the indicator can be returned to the normal display
mode with any button except ON/OFF.
4.3.1MIN-mode
If you wish to know the minimum readings that have been measured after
power up, press HOLD when the indicator is in the HOLD-mode. The
minimum readings of the currently displayed quantities are shown (if the
‘hold’-text is blinking, the HMI41 must first be returned to the normal display
mode in order to activate the MIN-mode; see Chapter 4.3):
RH
T
min
%
°C
With MODE or ENTER, the indicator returns to the normal display mode.
4.3.2MAX-mode
If you wish to know the maximum readings that have been measured after
power up, press HOLD when the indicator is in the MIN-mode. The maximum
readings of the currently displayed quantities appear:
RH
T
max
%
°C
The indicator returns to the normal display mode with any button except
ON/OFF.
7
HMI41 AND PROBES
Operating Manual M210316EN-A
5.CALIBRATION
The HMI41 indicator and probes are fully calibrated at the factory, so there
should be no immediate need for recalibration. Calibration is recommended
once in a year and always when there is a reason to believe that the
adjustments have changed.
5.1Calibration with trimmer potentiometers
5.1.1Humidity calibration
Humidity calibration of the probes is preferably done by adjusting the trimmer
potentiometers. The potentiometers are located under a protective plug. The
potentiometer marked T (temperature) is for factory use only; DO NOT make
any adjustments. To make sure that this potentiometer is not accidentally
touched when adjusting the other two, turn the plastic plug aside without
removing it completely (see Figure 5.1.1).
For adjusting the potentiometers, use a trimming wrench provided with the
probe, or some other suitable tool, e.g. a ceramic 1.5 mm slot screwdriver.
HMP41 &
HMP46
wet
dry
W
D
W
D
HMP45
TWD
wetdry
Figure 5.1.1 Location of trimmer potentiometers
A two point calibration is performed with the HMK15 calibrator or the probe
can be sent to Vaisala. The probes must always be recalibrated when the
sensor is changed.
8
HMI41 AND PROBES
M210316EN-AOperating Manual
The calibration procedure is as follows (see also the calibrator manual):
• Leave the calibrator and the probe at the calibration site for at least 30
minutes before starting the calibration in order to let the probe
temperature stabilize to the room temperature. Unscrew the plastic
grid (HMP41 and HMP45) or the sintered filter (HMP46) of the
probe.
• Insert the probe into a measurement hole of the LiCl salt chamber in
the humidity calibrator.
• Wait until the humidity reading stabilizes (approx. 30 minutes).
Check the temperature and read the closest humidity value in the
calibration table. Adjust the dry end reading with the potentiometer D
(dry) in the probe body to correspond to the value given in the
calibration table; use a suitable trimming screw wrench (see Figure
5.1.1).
• Insert the thermometer into the 13.5 mm hole of the NaCl salt
chamber and the probe into another hole of the NaCl chamber.
NOTE
When calibrating probes that are being used for long
term measurements (over 1 hour) in high humidities
(90 - 100 %RH), use the K2SO4 salt as the high end
reference.
• Wait until the humidity reading stabilizes. Read the salt chamber
temperature from the thermometer and then the closest humidity value
from the calibration table. Adjust the wet end reading with the
potentiometer W (wet) to correspond to the value given in the
calibration table.
*If the LiCl solution is used or stored in temperatures below +18°C, its equilibrium
humidity changes permanently
LiClNaClK2SO
4
9
HMI41 AND PROBES
Operating Manual M210316EN-A
As the D (dry) and W (wet) adjustments may affect each other, check again
the humidity reading in the LiCl salt chamber: insert the probe into the
calibration hole and wait until the reading stabilizes. If necessary, repeat the
adjustments in both the LiCl and the NaCl (K2SO4) salt chambers until the
reading is correct.
5.2Calibration with HMI41 software commands
Calibration can also be done with the HMI41 software commands. When
calibration is done with software commands, the corrections are entered in the
indicator memory with pushbuttons. If only one probe is being used, the
HMI41 software calibration is useful. However, when several probes are in
use, calibration should be done by adjusting the probe potentiometers. Note
that when the sensor is changed, calibration must always be done by adjusting
the potentiometers and it is recommended that the HMI41 is reverted to
factory settings by selecting default calibration.
NOTE
If the factory calibration of the HMI41 is changed, the
correction data refers to the calibrated probe only.
Therefore, if you change the probe, always revert this
data to factory settings by selecting default calibration,
or perform a new calibration with the new probe.
The calibration procedure is included in the HMI41 setup mode. Enter the
setup mode by pressing the ON/OFF button. The following appears:
RH
Td
T
P
set min maxhyst hold Lo bat Hi
%
°C °F
°C
°F
Pahg/m
Then release the ON/OFF button and press within 1 - 2 seconds both ENTER
and MODE buttons until the following text appears on the display:
10
After a few seconds, the text changes automatically to show the following:
°C
set
HMI41 AND PROBES
M210316EN-AOperating Manual
Press ENTER eight times until the following text appears:
set
This indicates that no calibration has been selected. In addition to this, there
are five calibration types available. The desired type is selected with buttons
▲ or ▼. All selections are acknowledged with ENTER. In the following, you
will find a list of these calibration options.
RH
T
Default calibration restores the factory settings of
the humidity and temperature calibrations.
set
RH
One point humidity calibration; select this for
performing humidity calibration at one point. See
Chapter 5.2.1 for further details.
set
RH
Two point humidity calibration; select this for
performing humidity calibration at two points.
See Chapter 5.2.2 for further details.
set
One point temperature calibration; select this for
T
performing temperature calibration at one point.
See Chapter 5.2.3 for further details.
set
Two point temperature calibration; select this for
T
performing temperature calibration at two points.
See Chapter 5.2.4 for further details.
set
5.2.1One point humidity calibration
In one point humidity calibration, one accurate humidity reference is
sufficient. However, note that after one point calibration, the humidity reading
is most accurate near the reference value. For a better accuracy over the whole
range, perform a two point calibration if possible.
Leave the reference instrument (HMK15) and the probe at the calibration site
at least 30 minutes so that the probe temperature stabilizes to the room
temperature. Start the calibration by inserting the probe to the reference
humidity.
11
HMI41 AND PROBES
Operating Manual M210316EN-A
In setup mode, press ENTER repeatedly until the following is displayed:
set
Then press MODE twice, and the following appears:
RH
set
Press ENTER to activate the one point calibration mode. A message similar to
the following appears with the first line blinking:
RH
set
%
The blinking number indicates the humidity reference value in the HMI41
memory. Check the salt chamber temperature, read the closest humidity value
in the calibration table and adjust the display with buttons ▲ and ▼ to
correspond to the value given in the table. For example, if the temperature in
the calibrator's NaCl salt chamber is 20.5 °C, adjust the value to 75.5 %RH:
RH
set
%
Each time the button is pressed, the value changes with 0.1 %. If you keep the
button pressed, the value changes in a faster mode. Press ENTER and the
HMI41 indicator shows the value that the probe is currently measuring,
displaying a message similar to the following:
12
RH%
Wait at least ten minutes for the reading to stabilize and press ENTER to
acknowledge the value. Press ENTER again to conclude the one point
HMI41 AND PROBES
M210316EN-AOperating Manual
calibration. If the calibration has been successful, the following message
appears:
The correction data has now been calculated and stored in the HMI41 memory.
The HMI41 returns automatically to the selection of display units, and can be
turned off. When the indicator is used as a standard indicator and the
correction data differs from the factory settings, an arrow in the upper
righthand corner is displayed:
RH
T
%
°C
If the message “cal pass” does not appear (instead, some other text may
appear, e.g. “too close”, “err offst” or “err gain”), the correction has not been
stored in the memory. The error may be due to an incorrect reference value or
to measured values that are out of the range.
5.2.2Two point humidity calibration
In two point humidity calibration, two accurate references (e.g. the HMK15
Calibrator) are needed. Leave the reference instrument and the probe at the
calibration site for at least 30 minutes so that the probe temperature stabilizes
to the room temperature.
Start the calibration by inserting the probe to the lower reference humidity. In
setup mode press ENTER repeatedly until the following is displayed:
set
Then press MODE three times, and the following appears:
RH
set
13
HMI41 AND PROBES
Operating Manual M210316EN-A
Press ENTER to activate the two point humidity calibration mode. A message
similar to the following appears with the first line blinking:
RH
set
%
Lo
The blinking number indicates the value of the lower humidity reference
stored in the HMI41 memory. Check the salt chamber temperature, read the
closest humidity value in the calibration table and adjust the display with
buttons ▲ and ▼ to correspond to the value given in the table. For example, if
the temperature of the LiCl salt chamber is 22 °C, adjust the value to
11.3 %RH:
RH
set
%
Lo
Press ENTER, and the HMI41 shows the value that the probe is currently
measuring, displaying a message similar to the following:
RH%
Lo
Wait at least ten minutes for the reading to stabilize and press ENTER to
acknowledge the value. Press ENTER again to conclude the lower point
calibration. A message similar to the following appears with the first line
blinking:
RH
set
%
Hi
The blinking number indicates the value of the higher humidity reference
stored in the HMI41 memory. Insert the probe to the higher reference
humidity. Check the salt chamber temperature, read the closest humidity value
in the calibration table and adjust the display with buttons ▲ and ▼ to
correspond to the value given in the table. For example, if the temperature in
the calibrator's NaCl salt chamber is 20.5 °C, adjust the value to 75.5 %RH:
14
HMI41 AND PROBES
M210316EN-AOperating Manual
RH
set
%
Press ENTER and the HMI41 indicator shows the value that the probe is
currently measuring, displaying a message similar to the following:
RH%
Hi
Wait at least ten minutes for the reading to stabilize and press ENTER to
acknowledge the value. Press ENTER again to conclude the calibration. If the
calibration has been successful, the following message appears:
The correction data has now been calculated and stored in the HMI41 memory.
The HMI41 returns automatically to the selection of the display units, and can
be turned off. When the indicator is used as a standard indicator and the
correction data differs from the factory settings, an arrow in the upper
righthand corner is displayed:
RH
T
%
°C
If the message “cal pass” does not appear (instead, some other text may
appear, e.g. “too close”, “err offst” or “err gain”), the correction has not been
stored in the memory. The error may be due to an incorrect reference value or
to measured values that are out of the range.
5.2.3One point temperature calibration
In one point temperature calibration, one accurate temperature reference is
sufficient.
Start the calibration by inserting the probe to the reference temperature.
In setup mode, press ENTER repeatedly until the following is displayed:
15
HMI41 AND PROBES
Operating Manual M210316EN-A
set
Then press MODE four times, and the following appears:
T
set
Press ENTER to activate the one point temperature calibration mode.
A message similar to the following appears with the first line blinking:
T°C
set
The blinking number indicates the temperature reference value stored in the
HMI41 memory. Check the actual temperature reference and change the
display with buttons ▲ and ▼ to correspond to the accurate value, for
example:
T°C
set
Press ENTER and the HMI41 indicator shows the value that the probe is
currently measuring, displaying a message similar to the following:
T°C
16
Wait at least ten minutes for the reading to stabilize and press ENTER to
acknowledge the value. Press ENTER again to conclude the calibration. If the
calibration has been succesful, the following message appears:
HMI41 AND PROBES
M210316EN-AOperating Manual
The correction data has now been calculated and stored in the HMI41 memory.
The HMI41 returns automatically to the selection of the display units, and can
be turned off. When the indicator is used as a standard indicator and the
correction data differs from the factory settings, an arrow in the upper
righthand corner is displayed:
RH
T
%
°C
If the message “cal pass” does not appear (instead, some other text may
appear, e.g. “too close”, “err offst” or “err gain”), the correction has not been
stored in the memory. The error may be due to an incorrect reference value or
to measured values that are out of the range.
5.2.4Two point temperature calibration
In two point temperature calibration, two accurate temperature references are
needed. Note that you must allow enough time for all instruments to stabilize
to the temperature equilibrium.
Start the calibration by inserting the probe to the lower reference temperature.
In setup mode, press ENTER repeatedly until the following is displayed:
set
Then press MODE five times, and the following appears:
T
set
Press ENTER to activate the two point temperature calibration mode. A
message similar to the following appears with the first line blinking:
17
HMI41 AND PROBES
Operating Manual M210316EN-A
T
set
°C
Lo
The blinking number indicates the value of the lower temperature reference
stored in the HMI41 memory. Insert the probe to the lower reference
temperature. Check the reference temperature and change the display with
buttons ▲ and ▼ to correspond to the accurate value, for example:
T
set
°C
Lo
Press ENTER and the HMI41 shows the value that the probe is currently
measuring, displaying a message similar to the following:
T
°C
Lo
Wait at least ten minutes for the reading to stabilize and press ENTER to
acknowledge the value. Press ENTER again to conclude the lower point
calibration. A message similar to the following appears with the first line
blinking:
T
set
°C
Hi
The blinking number indicates the value of the higher temperature reference
stored in the HMI41 memory. Insert the probe to the higher reference
temperature. Check the reference temperature and change the display with
buttons ▲ and ▼ to correspond to the accurate value, for example:
T
°C
18
set
Hi
Press ENTER and the HMI41 shows the value that the probe is currently
measuring, displaying a message similar to the following:
Loading...
+ 50 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.