Vaisala HMP45D User Manual

USER'S GUIDE
Vaisala HUMICAP® Humidity and
Temperature Probes
HMP45A/D
U274EN-1.2
PUBLISHED BY Vaisala Oyj Phone (int.): +358 9 8949 1
P.O. Box 26 Fax: +358 9 8949 2227 FIN-00421 Helsinki Finland
Visit our Internet pages at http://www.vaisala.com/
© Vaisala 2006 No part of this manual may be reproduced in any form or by any means, electronic or
mechanical (including photocopying), nor may its contents be communicated to a third party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for Vaisala towards the customer or end user. All legally binding commitments and agreements are included exclusively in the applicable supply contract or Conditions of Sale.
HMP45A and HMP45D
U274en-1.2 Operating Manual

1. GENERAL

The HMP45A and HMP45D probes are designed for the measurement of relative humidity and temperature. Humidity measurement is based on the capacitive thin film polymer sensor HUMICAP180. Temperature measurement is based on resistive platinum sensors (Pt 100 and Pt 1000). Both the humidity and temperature sensors are located at the tip of the probe and in standard version protected by a membrane filter. The HMP45A and HMP45D have a similar humidity output, but the temperature output is active in HMP45A (voltage output 0-1V) and passive in HMP45D (resistive output Pt 100).

2. CONNECTIONS

The cable wires are connected as shown in Figure 1.
HMP45D
HMP45A
YEL OUTPUT
0...1V=-40...+60 ˚C
BLU V+ 7...35VDC
BRN OUTPUT
0...1V=0... 100%RH
VIO GND
RED GND (SENSE)
GREY (SHIELD)
YEL
WHT
GRN
BLK
GREY (SHIELD)
Pt100 Pt100
Pt100 Pt100
BLU 7...35VDC
BRN OUTPUT 0...1V=0...100%RH
VIO GND RED GND (SENSE)
Figure 1 Wire colours
Pt100
SIGNAL GROUND is used for output signal in a differential measurement. With
SIGNAL GROUND, the cable can be extended up to 100 metres without
disturbing the measurement accuracy. When outputs are not measured against
SIGNAL GROUND, connect GROUND and SIGNAL GROUND to the same point.
1
HMP45A and HMP45D Operating Manual U274en-1.2

3. CALIBRATION AND MAINTENANCE

Calibration and maintenance of the probes should be performed at regular intervals, depending on the conditions of use and desired accuracy. The recommended calibration interval is one year.
The HMP45A/D probes are easy to maintain and calibrate. The probe consists of a probe head and a handle with cable. All calibration electronics are in the probe head which can be disconnected from the handle without disconnecting the wires (see Figure 2). The handles of all HMP45A and HMP45 probes are fully interchangeable. If you wish to continue the measurement immediately, you can insert a calibrated probe head in place of the disconnected one; this way, the measurement is interrupted for less than a minute.

3.1 Reading the outputs during calibration

adjustment trimmers: W = wet D = dry (T = temperature; for
D
W
T
factory use only!)
pull the probe head off the handle
NOTE: remove the O-ring if you connect the probe directly to the HMI41 indicator
ENTERMODE HOLD ON/OFF
HMI41HMH45
Figure 2 Adjustment trimmers and probe head connection/disconnection
In calibration, the HMP45A/D output is normally read from the output cables. The HMP45A probe head can also be checked with Vaisala’s HMI41 indicator (see Figure 2); the probe head can be connected either to the HMH45 handle, or directly to the HMI41 indicator. However, note that if you connect the probe head directly to the HMI41, you must first remove the O-ring. Measured humidity and temperature appear on the HMI41 display.
A simple field check can also be performed when there seems to be something wrong with the measurement system. It is sufficient to compare the HMP45A reading with the reading measured by a calibrated reference probe head. Check
2
HMP45A and HMP45D
U274en-1.2 Operating Manual
the ambient relative humidity and temperature level with the reference probe head, and then connect the HMP45A probe head on the HMH45 handle and check the output readings.
Prepare the HMI41 indicator for measurements with the HMP45A probe head as follows: turn the HMI41 indicator on and within 1-2 seconds press simultaneously buttons ENTER and MODE until the text SEtUP appears on the display.
Wait a few seconds, and the text Unit appears; press ENTER repeatedly until a text similar to the following appears on the display:
set
Select PROBE TYPE 3 with buttons (number up) or (number down) and turn the indicator off. Turn it on again by pressing the ON/OFF button. After a few seconds, the relative humidity and temperature readings appear automatically on the display. The HMP45D cannot be used w ith the HMI41
indicator as the passive temperature signal results in error messages on the HMI41.

3.2 Humidity calibration

For a high-accuracy two-point calibration use a Vaisala HMK15 or HMK13B calibrator and saturated salt solutions as described in the respective manuals.
Leave the calibrator, the HMI41 and the probe head in the same space for at least four hours so that their temperatures have time to equalize. Unscrew the plastic grid of the probe.
The calibration is done first for the dry end and then for the wet end by adjusting trimmer potentiometers marked “D” (dry, <50 %RH) and “W” (wet, >50 %RH). The potentiometers are located under a protective plug; see Figure
2. Use a ceramic screw driver with 2.5 mm blade for adjusting the potentiometers. Note: if zero point is calibrated in Nitrogen (N2), the minimum output signal of 0.008 V corresponds to a relative humidity of 0.8 %RH.
Greenspan’s calibration table
Temperature °C 15 20 25 30 35 LiCl %RH *) 11.3 11.3 11.3 11.3 NaCl %RH 75.6 75.5 75.3 75.1 74.9
K2SO
4
*) Do not use or store the LiCl solution in temperatures below +18°C as its humidity
equilibrium may change permanently
%RH 97.9 97.6 97.3 97.0 96.7
3
HMP45A and HMP45D Operating Manual U274en-1.2
As the D (dry) and W (wet) adjustments may affect each other, check again the humidity reading at the low end. If necessary, repeat the adjustments in both the low and the high humidity points until the reading is correct.

3.3 Changing the HUMICAP®180 humidity sensor

Unscrew the filter. Remove the damaged sensor and mount a new HUMICAP®180 humidity sensor in its place. Handle the sensor with care. Calibrate the probe using a two-point calibration procedure. Note that if the probe is not calibrated, the accuracy is still better than ±7 %RH.

4. SPARE PARTS AND ACCESSORIES

Order code Description
HUMICAP180 18921 Temperature sensor Pt 1000 IEC 751 1/3 Class B (HMP45A) 19159 Temperature sensor Pt 100 IEC 751 1/3 Class B (HMP45D) 2787HM Membrane filter (standard) 6685 Sintered filter 37 µm 6686 Sintered filter 216 µm 6597 Plastic grid HMP45ASP HMP45A probe head HMP45DSP HMP45D probe head HMH45ASP Probe handle for HMP45A and HMP45D HMI41 Humidity and temperature indicator HMH45 Probe handle for HMP45A with a connector to HMI41 HMK11 Humidity Calibrator HMK13B Humidity Calibrator
Humidity sensor

5. TECHNICAL DATA

5.1 Humidity (HMP45A & HMP45D)

Measurement range 0.8...100 %RH Output scale 0...100 %RH equals 0-1 VDC Accuracy at +20 °C (including non-linearity and hysteresis):
against factory references ±1 %RH field calibration against references ±2 %RH (0...90 %RH)
Typical long-term stability better than 1 %RH per year Temperature dependence ±0.05 %RH/°C Response time (90%) at +20 °C 15 s with membrane filter Humidity sensor HUMICAP
4
±3 %RH (90...100 %RH)
®
180
HMP45A and HMP45D
C
U274en-1.2 Operating Manual

5.2 Temperature

HMP45A

Measurement range -39.2...+60 °C Output scale -40...+60 °C equals 0...1 VDC Accuracy at 20°C ±0.2 °C Accuracy over the whole measurement range:
0.5
0.4
0.3
0.2
0.1 0
-0.1
-0.2
-0.3
-0.4
-0.5
-30-40
-20
-10 0
10
20
30
40
50
60
°
Temperature sensor Pt 1000 IEC 751 1/3 Class B

HMP45D

Measurement range -40...+60 °C Output signal resistive four wire connection Temperature sensor Pt 100 IEC 751 1/3 Class B

5.3 General

Operating temperature range -40...+60 °C Storage temperature range -40...+80 °C Supply voltage 7...35 VDC Settling time 500 ms
Power consumption < 4 mA Output load >10 kohm (to ground) Weight 350 g (including package) Cable length 3.5 m Housing material ABS plastic Housing classification (electronics) IP 65 (NEMA 4) Sensor protection (standard) membrane filter, part no. 2787HM Dimensions in mm (inches):
5
HMP45A and HMP45D Operating Manual U274en-1.2
240 (9.45)
118 (4.65)
A
3500 (137.8)
24
A
A-A
20.5
(0.81)
(0.94)

5.4 Electromagnetic compatibility

5.4.1 Emissions

Radiated interference, test setup according to EN55022

5.4.2 Immunity

Test: Test setup Performance:
according to:
24
(0.94)
18.5
( 0.72)
REMOVABLE PROBE HEAD
(132 mm)
Radiated interference IEC 1000-4-3 HMP45A level 1 (3V/m)
HMP45D level 3 (10 V/m)
Electrostatic discharge IEC 801-4 level 4 (HMP45A&D)

GUARANTEE

Vaisala issues a guarantee for the material and workmanship of this product for one (1) year from the date of delivery. Damage due to exceptional operating conditions, careless handling or misapplication will void the guarantee. Detailed warranty information is given in the Warranty and the Standard Conditions of Sale of Vaisala Oy.
6
*U274
EN
*
www.vaisala.com
Loading...