HM70 Handheld Humidity and
Temperature Meter
For Spot-Checking and Field Calibration
Features
• Designed for spot-checking and field calibration
• Multilingual user interface
• Shows measurement trends graphically
• Proven Vaisala HUMICAPâ sensor technology
• 3 probe alternatives, temperature measurement ranges between -70 and +180 °C (-94 and
+356 °F)
• 2 probes: also dew point and CO2 probes can be connected simultaneously
• Displays various humidity parameters
• Sensor preheat and chemical purge options for demanding conditions
• Data can be logged and transferred to a PC via MI70 Link software
• 6-point traceable calibration (certificate included)
Vaisala HUMICAPâ Handheld Humidity and Temperature Meter HM70 is designed for demanding humidity measurements in spot-checking applications. It is also ideal as an on-site calibrator for Vaisala's fixed humidity instruments.
Vaisala HUMICAPâ Technology
HM70 incorporates the world-class HUMICAPâ sensor, one of the most reliable and stable sensors on the market. The HUMICAPâ sensor copes well with chemical interference and provides accuracy that lasts in demanding conditions.
Chemical Purge
The chemical purge option maintains measurement accuracy in environments with high concentrations of chemicals. The sensor preheat option reduces measurement delays as it keeps the sensor dry when the probe is inserted into hot and humid processes.
Three Probes to Choose From
HMP75 is a general purpose probe, whereas HMP76 is a long, stainless steel probe especially suitable for spotchecking in ducts. HMP77 is a small probe at the end of a 5-meter cable. The probe is ideal for diƒcult-to-reach areas and for on-site calibration of Vaisala's process transmitters. In addition, HM70 supports the use of Vaisala's dew point, carbon dioxide, and moisture in oil probes, allowing measurements in several multiparameter applications.
MI70 Link
The optional MI70 Link Windowsâ software and the USB connection cable form a practical tool for transferring logged data from HM70 to a PC.
On-site calibration with HM70 Handheld Meter
Technical Data
Measurement Performance, HMP75, HMP76, and
HMP77 Probes
Relative Humidity |
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|
Measurement range |
|
|
|
|
0 ... 100 %RH |
|
|||
Accuracy including non-linearity, hysteresis and repeatability. Defined as ±2 |
|||||||||
standard deviation limits: |
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|||
At +15 ... +25 °C (+59 ... +77 °F) |
|
|
±1 %RH (0 ... 90 %RH) |
|
|||||
|
|
|
|
|
|
|
±1.7 %RH (90 ... 100 %RH) |
|
|
At -20 ... +40 °C (-4 ... +104 °F) |
|
|
±(1.0 + 0.008 x reading) %RH |
||||||
|
|
|
|
|
|
||||
At -40 ... +180 °C (-40 ... +356 °F) |
|
±(1.5 + 0.015 x reading) %RH |
|
||||||
Factory calibration uncertainty |
|
|
±0.6 %RH (0 ... 40 %RH) |
||||||
(+20 °C / +68 °F) |
|
|
|
|
±1.0 %RH (40 ... 97 %RH) |
||||
|
|
|
|
||||||
Response time (90%) at +20 °C (+68 °F) in still air: |
|
||||||||
HMP75 (with standard plastic grid) |
|
17 s |
|
|
|||||
|
|
|
|
|
|
|
|
||
HMP76 (with standard sintered bronze |
|
60 s |
|
|
|||||
filter) |
|
|
|
|
|
|
|
||
HMP77 (with standard plastic grid and |
|
50 s |
|
|
|||||
stainless steel netting) |
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
||
Typical long-term stability |
|
|
|
|
Better than 1 %RH / year |
|
|||
Temperature |
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|||
HMP75 measurement range |
|
|
-20 ... +60 °C (-4 ... +140 °F) |
|
|||||
HMP76 measurement range |
|
|
-50 ... +120 °C (-58 ... +248 °F) |
||||||
|
|
|
|
|
|
||||
HMP76 short time measurement range |
|
-50 ... +180 °C (-58 ... +356 °F) |
|
||||||
HMP77 measurement range |
|
|
-70 ... +180 °C (-94 ... +356 °F) |
||||||
|
|
|
|
|
|
|
|||
Accuracy at +20 °C (+68 °F) |
|
|
±0.2 °C (±0.36 °F) |
|
|||||
Accuracy over temperature range: |
|
|
|
|
|||||
0.7 |
°C |
|
|
|
|
|
|
|
|
0.6 |
|
|
|
|
|
|
|
|
|
0.5 |
|
|
|
|
|
|
|
|
|
0.4 |
|
|
|
|
|
|
|
|
|
0.3 |
|
|
|
|
|
|
|
|
|
0.2 |
|
|
|
|
|
|
|
|
|
0.1 |
|
|
|
|
|
|
|
°C |
|
0 |
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
-0.1 |
|
|
|
|
|
|
|
|
|
-0.2 |
|
|
|
|
|
|
|
|
|
-0.3 |
|
|
|
|
|
|
|
|
|
-0.4 |
|
|
|
|
|
|
|
|
|
-0.5 |
|
|
|
|
|
|
|
|
|
-0.6 |
|
|
|
|
|
|
|
|
|
-0.7 |
|
|
|
|
|
|
|
80 100 120 140 160 180 |
|
-80 -60 -40 -20 |
0 |
20 |
40 |
60 |
|
Other Parameters Available
dew point, frost point, absolute humidity, mixing ratio, wet bulb temperature, water content, vapor pressure, saturation vapor pressure, enthalpy, water activity
General, HMP75, HMP76, and HMP77 Probes
Humidity sensor |
HUMICAPâ 180R |
|
HUMICAPâ 180RC (chemical purge, |
|
sensor preheat) |
|
|
Temperature sensor |
Pt100 RTD Class F0.1 IEC 60751 |
Operating temperature range for |
-40 ... +60 °C ( -40 ... +140 °F) |
electronics |
|
|
|
Standard Sensor Protection |
|
HMP75 |
Plastic grid |
|
|
HMP76 |
Sintered bronze filter |
HMP77 |
Grid with SS netting |
Mechanical Specifications, HMP75, HMP76, and
HMP77 Probes
IP rating |
IP65 (NEMA 4) |
|
|
Housing material |
ABS/PC blend |
Probe material |
Stainless steel (AIS316L) |
|
|
Probe cable length (between indicator |
1.9 m (6.2 ft) |
and probe handle) |
|
Probe cable length of HMP77 (from |
5.0 m (16 ft) |
handle to the root of probe) |
|
|
|
Probe diameter |
12 mm (0.47 in) |
Weight |
|
|
|
HMP75 |
250 g (8.8 oz) |
HMP76 |
350 g (12 oz) |
|
|
HMP77 |
500 g (18 oz) |
|
|
MI70 Measurement Indicator
Operating Environment |
|
|
|
Operating temperature |
-10 ... +40 °C (+14 … +104 °F) |
Operating humidity |
0 ... 100 %RH, non-condensing |
|
|
Storage temperature |
-40 ... +70 °C (-40 ... +158 °F) |
Inputs and Outputs |
|
|
|
Max. no of probes |
2 |
Power supply |
Rechargeable NiMH battery pack with |
|
AC adapter or 4xAA size alkalines, |
|
type IEC LR6 |
|
|
PC interface |
MI70 Link software with USB or serial |
|
port cable |
Analog Output |
|
|
|
Scale |
0 ... 1 VDC |
Output resolution |
0.6 mV |
|
|
Accuracy |
0.2 % full scale |
Temperature dependence |
0.002 %/˚C (0.01 %/˚F) full scale |
|
|
Minimum load resistor |
10 kΩ to ground |
Mechanical Specifications |
|
|
|
Housing classification |
IP54 |
Housing materials |
ABS/PC blend |
|
|
Weight |
400 g (14 oz) |
Compatibility |
|
|
|
EMC compliance |
EN61326-1, Portable Equipment |
Other |
|
|
|
Menu languages |
English, Chinese, Spanish, Russian, |
|
French, Japanese, German, Swedish, |
|
Finnish |
Display |
• LCD with backlight |
|
• Graphic trend display of any |
|
parameter |
|
• Character height up to 16 mm |
|
(0.63 in) |
|
|
Alarm |
Audible alarm function |
Data logging capacity |
2700 real time data points |
|
|
Logging interval |
1 s to 12 h |
Logging duration |
1 min ... memory full |
|
|
Resolution |
0.01 %RH, 0.01 °C/°F, 0.01 hPa, |
|
0.01 aw, 10 ppm / 0.01 %CO2 |