UVP MultiDoc-It Imaging System User Manual

MultiDoc-It™ Imaging System
Installation and User Instructions
UVP, LLC Ultra-Violet Products Ltd.
2066 W. 11th Street Unit 1, Trinity Hall Farm Estate Upland, CA 91786 Nuffield Road, Cambridge CB4 1TG UK Phone: (800) 452-6788 / (909) 946-3197 Phone: +44(0)1223-420022 Fax: (909) 946-3597 Fax: +44(0)1223-420561

Web Site: www.uvp.com

81-0214-01 Rev K
MultiDoc-It Imaging System 2
Overhead white light
Benchtop UV

Introduction

The MultiDoc-It Imaging System is a high resolution imager capable of capturing and documenting fluorescent gel images. The system combines:
1. MultiDoc-It Darkroom
2. DigiCam digital color camera
3. Ethidium Bromide (EtBr) Filter
4. Doc-It
5. UV transilluminator (transilluminator instructions are provided in a separate manual)
NOTE: MultiDoc-It TLC Systems (for Thin Layer Chromatography) do not include a transilluminator.
®
LS Acquisition Software
Small footprint

darkroom

UV-blocking Gel
Viewer Window
Wide access door
DigiCam camera
with diopter
Three position filter tray;
EtBr filter included
Side access doors
(optional)
Transilluminator
Darkroom
The MultiDoc-It darkroom provides a dark environment for gel UV illumination and protects the DigiCam camera. Some MultiDoc-It models include side access doors for moving or cutting gels without opening the front door.

UV Filter

The included emission filter, an orange-colored UV blocking bandpass filter, is used to absorb UV and IR radiation from the transilluminator and to enhance the orange/pink bands generated by Ethidium Bromide stained gels. The filter can be removed when imaging non-fluorescent media (protein gels, colony plates, etc.) in order to produce brighter images.

Camera and Diopter

The DigiCam digital color camera includes a built-in optical zoom lens. A close-up diopter is connected to the lens at the UVP factory.
MultiDoc-It Imaging System 3
Doc-ItLS Acquisition Software

Doc-ItLS Software

The Doc-ItLS software loads onto the user’s existing computer. The software controls the camera acquisition functions as well as enhancement and reporting features. For information on operation of the software, refer to the Doc-ItLS software help files.

UV Transilluminator

A UV transilluminator may be shipped with the system. Refer to the Transilluminator manual for operating instructions.

Operating System Requirements

Windows 7, Vista or XP SP2 or later
Internet Explorer 6.0 or higher [To locate the version of Internet Explorer, open Internet
Explorer and click on He lp > About Internet Explorer]
Intel Pentium Processor or equivalent, 1.6 GHz or higher
1 GB of RAM or greater
200 MB of available hard disk space for the program, more for data
CD-ROM drive
One Universal Serial Bus (USB) for connecting the camera
Color monitor, supporting at least 1024 x 768 resolution and 16-bit or better colors; 24-bit or
32-bit color is strongly recommended
NOTE: For 21 CFR Part 11 support functionality, the computer partition must be formatted with NTFS.
Refer to the Doc-ItLS software manual for instructions.