MS-100 Multi-Sense
Optical Radiometer
USER INSTRUCTIONS
Instructions. The Multi-SenseTM MS-100 instrument and plug-in sensors are specifically
designed to be simple to use and maintain. The instrument has no external push buttons or
switches, so taking measurements is very straightforward:
1. Select the sensor for the measurement you want to make.
2. Insert the plug into the receptacle at the top of the MS-100 instrument.
3. The instrument will now be switched on. Take your reading. The units of
measurement will display on the back of the sensor.
4. When finished, unplug the sensor by gripping the plug body and pulling. Do not pull
on the cable, as the plug is a latching type.
5. For the utmost precision, it is possible to set the display accurately to zero, when
required. The adjustment point is accessible through a 4 mm diameter hole on the
left hand side of the instrument. Use a flat blade miniature screwdriver or proprietary
trimmer tool. Ensure that the sensor is properly blanked (e.g. held face down on flat
surface of optical sensors).
6. When the battery requires renewing, a small (bat) indicator will appear in the top-left
of the display window. Open the battery compartment by sliding the cover
downwards. Replace the battery with type PP3, MN1604, 6F22 or equivalent 9-volt
type.
7. If you require help with your measurement techniques or are having difficulty
interpreting your results, please call our offices.
Calibration. UVP recommends this unit be returned for recalibration every six months. Contact
UVP for instructions on sending a meter for calibration.
UVP, Inc. 2066 W. 11th Street, Upland, CA 91786 | (800) 452-6788 | Fax: (909) 946-3197 | E-mail: info@uvp.com
Ultra-Violet Products Ltd, Unit 1, Trinity Hall Farm Estate, Nuffield Road, Cambridge CB4 1TG UK
+44(0)1223-420022 | Fax: +44(0)1223-420561 | E-Mail: uvp@uvp.co.uk
Internet: www.uvp.com
81-0219-01 Rev B