The EpiChemi II (EC2) Darkroom is designed specifically for low light and fluorescence imaging and because of its light-tightness is particularly suitable for chemiluminescent applications. The special design accommodates a Benchtop Transilluminator on the convenient sliding
tray for easy access. For chemiluminescent applications, the chemi sample tray can be lowered for sample positioning. Additionally, the darkroom provides users added flexibility due
to the filter wheel and overhead UV.
UVP offers a large selection of cameras with specially designed mounts for securely connecting the camera to the darkroom. Camera specifications are not discussed in this manual.
Contact UVP or see other documentation included with your system for camera operating
and set-up instructions. Transilluminators are ordered separately and also include their own
documentation.
The EpiChemi II Darkroom features:
Overhead Light
Selector:
White Light
365nm UV
254nm UV
CCD Camera and Mounting Bracket
(Contact UVP or see other documentation
for camera set-up instructions. The
EpiChemi Darkroom may be sold with
various camera options with appropriate
mounting brackets.)
Filter Wheel
Selector
Transilluminator
Control
Master On/Off Switch
Gel Viewer
Compression
Door Lock
Benchtop UV
Transilluminator
Fold-Down
Sample Platform
for Chemi blots
Sliding Transilluminator Tray
1
EpiChemi II Darkroom
Components
The EpiChemi II Darkroom is comprised of:
Darkroom Cabinet
The EpiChemi II Darkroom is manufactured of stainless
steel construction and fabricated to provide a light-tight
chamber. The darkroom is ready to go with connections to
a wall outlet, the camera, mount and transilluminator.
Filter Wheel
The five-position filter wheel is designed for viewing many
fluorescent and visible color stains. Rotate the dial to desired filter as required for you application. The filter wheel is
securely mounted into the darkroom for a clean, protected
environment.
Filter Wheel
EC2 Darkroom Filter Performance
Gel Viewer
The viewer opens for safe viewing of samples. The safety
glass allows viewing without exposure to ultraviolet radiation.
Sample Platform
The sample platform folds down for placement of chemi
blots.
2
Sample Platform and
Transilluminator Tray
EpiChemi II Darkroom
Transilluminator Tray
The slides forward and back for easy access to the transilluminator surface. See transilluminator documentation for specific transilluminator operating instructions.
NOTE: When using the transilluminator with the darkroom door open, UV exposure safety
precautions, including protective eyewear and clothing, must be observed to prevent harmful ultraviolet radiation exposure.
Set-Up and Operating Instructions
DO NOT ATTEMPT TO CONNECT ANY WIRING WHILE THE EQUIPMENT IS
CONNECTED TO ANY POWER SUPPLY.
CAUTION!
Do not install the EpiChemi II Darkroom in places with high moisture, dust or high
temperature.
Do not use any oil or petroleum based cleaner for the cabinet. Use only mild soap or
detergent solution for cleaning. Ensure that the system is turned OFF during cleaning.
Set-Up
1. Connect the camera and mount to the darkroom as
specified in the camera documentation manual.
2. Connect the power cord to the main power port which is
located on the back-top of the darkroom cabinet.
Operation
1. Turn on the master power switch on the front of the
darkroom cabinet. Once the power is switched on, power
is applied to all internal components and to the power port
for the jumper power cable that supplies power to the
transilluminator.
2. The Transilluminator Control knob has three options:
Safety Switch - Transilluminator safety control switch
automatically turns the transilluminator off when door opens
which prevents the user from exposure to UV
Always Off - This button setting puts the transilluminator in
Off mode. Use this switch when the transilluminator is not
required such as when capturing chemi blots.
3
EpiChemi II Darkroom
Always On - This setting overrides the safety switch with
the transilluminator always in the on mode. Users should
wear protective clothing and eyewear when the
darkroom is in the Always On mode to prevent exposure
to the UV.
3. The Overhead Light control knob has four options for
providing overhead diffused 254 and 365nm UV and
fluorescent white light for uniform epi-illumination.
Off - Place the overhead light switch in the Off mode when
not using the darkroom for applications requiring epiillumination.
White - Overhead white light illuminates when white is
selected.
365nm and 254nm - When either ultraviolet wavelength is
selected, the red light will illuminate to alert the user that
the light is in operation. The overhead lights are designed
to automatically turn off when the darkroom door is open.
4. The Transilluminator Tray accommodates a Benchtop
UV Transilluminator. The tray moves forward or back for
easy access to the transilluminator surface.
NOTE: When using the transilluminator with the darkroom
door open, UV exposure safety precautions, including protective eyewear and clothing,
must be observed to prevent
harmful ultraviolet radiation exposure.
5. The Sample Platform folds backwards against the back
of the darkroom cabinet. Pull the platform down into a
horizontal position as shown in the picture. The sample
platform is designed for use when transillumination is not
required. The platform provides a surface for placement of
chemiluminescent blots with minimum background noise
interference.
6. The Gel Viewer has a pressure-sensitive clasp. Press firmly
to open the viewer. The glass is UV blocking while providing
a clear view to the transilluminator or platform surface for
visibility of samples without opening the door.
4
EpiChemi II Darkroom
Troubleshooting
No power to darkroom cabinet.
1. Recheck main power cord connections to the both the EC2 Darkroom and the wall
power (surge protector).
2. Check the fuse located on the main power port. If the darkroom continues to blow
fuses, call UVP Technical Support Department.
Transilluminator will not turn on!
1. Be sure the darkroom cabinets door is completely closed. There is a UV exposure
safety cut-off switch that turns the transilluminator of when the darkroom cabinets door
is opened.
2. Be sure the darkroom cabinets main power switch is lit. If not, refer to No power to
darkroom cabinet above.
3. Be sure the transilluminator switch is on. When the switch on the transilluminator is lit
green, the unit is receiving power!
4. Be sure the transilluminator control switch on the darkrooms front panel is selected to
either safety switch or always on.
5. Call UVP Technical Support Department with any technical questions.
Replacement Parts
To order replacement parts for the EpiChemi II Darkroom, contact UVPs offices. Contact
information is listed on the next page.
UVP offers technical support for all of its products. If you have any questions about the
products use, operation or repair, please call or fax UVP Customer Service or Technical
Support at the following:
If you are located in North America, SouthIf you are in Europe, Africa, the Middle East
America, East Asia or Australia:or Western Asia:
Call (800) 452-6788 or (909) 946-3197 andCall +44(0) 1223-420022 and ask for BioImaging
ask for BioImaging Systems Technical Support.Systems Technical Support. Support is available
Tech Support assistance is available duringduring regular business days, between 8:00am
regular business days, between 7:00am andand 5:30 pm UK time.
5:30 pm PST.
E-Mail your message to:E-Mail your message to
techsupport@uvp.comuvp@uvp.co.uk
Fax your questions to:Fax your questions to:
(909) 946-3597+44(0)1223-420561
Write to:Write to:
BioImaging Systems Technical SupportBioImaging Systems Technical Support
UVP, Inc.Ultra-Violet Products Ltd.
2066 W. 11th StreetUnit 1, Trinity Hall Farm Estate, Nuffield Road
Upland, CA 91786 USACambridge CB4 1TG UK
A Returned Goods Authorization (RGA) number must be obtained from UVPs
Customer Service Department prior to returning any product to UVP.
6
EpiChemi II Darkroom
Wiring Diagram
7
Products available from UVP
LabWorks Image Acquisition and Analysis Software
Ultraviolet Transilluminators
UV/White Light Transilluminators
Visi-Blue Transilluminators and Plates
White Light Transilluminators and Plates
Gel Tools
UV Crosslinkers
UV Viewing Cabinets
UV Lamps
Ultraviolet Intensity Meters
Hybridization Ovens
http://www.uvp.com
UVP, Inc., 2066 W. 11th Street, Upland, California 91786