System Components ................................................................................................................................................... 4
Safety Information ..................................................................................................................................................... 5
Using the Light Source .............................................................................................................................................. 11
Service Procedures .................................................................................................................................................... 13
Replacement Parts and Accessories ....................................................................................................................... 14
Care and Cleaning .................................................................................................................................................. 14
Technical Support ................................................................................................................................................... 15
BioLite Xe MultiSpectral Light Source 3
Introduction
The BioLite Xe MultiSpectral Light Source is a 150 watt xenon light source designed for use with select
UVP Advanced Imaging Systems. The unit features a closed optical path to tightly control the output
spectrum, allowing consistent and repeatable measurements with superior signal to noise. The BioLit e Xe
features a six position dimmer permitting variable intensity , and the integrated filter wheel a cco mmodates up
to eight excitation filters. Sy stem control is via either the VisionWorks
located on the front of the unit.
The unit’s fiber optic bundle, designed uniquely for the BioLite, direc ts lighting to the darkroom. Refer to the
components list included with the BioLite and imaging system to determine the appropriate assembly
instructions.
®
LS software interface or switche s
BioLite Xe MultiSpectral Light Source 4
Fiber Optic Cable
Filter Access Door
(Side)
Press Button to
Transillumination
Illumination Ports: Left for
Master Power
Switch
Xenon Light Power Button:
Light Intensity Selector
Filter Selector Switch and
Position Indicator
System Components
Refer to the packing slip and schematic shown below for specific parts and components included with the
BioLite MultiSpectral Light Source.
The graph below displays the spectral emission curve for the xenon arc light source within the BioLite,
without filtering:
Safety Information
WARNING: The device will overheat if the fan outlet on the back of the BioLite is blocked.
WARNING: The xenon bulb becomes very hot with use. Do not open the bottom access panel while the
BioLite is running or without allowing it to fully cool.
CAUTION: Never touch the inner portion of the xenon bulb. If it accidentally comes in contact with skin,
carefully clean the glass with rubbing alcohol and a clean, lint free cloth.
CAUTION: Never look into the output port or fiber optic bundle of the BioLite when it is on, as the intense
light may hurt your eyes.
BioLite Xe MultiSpectral Light Source 6
USB Cable
Power Cord
Ridge
Side of
Setup Instructions
BioLite MultiSpec tral Light Source Setup
1. Plug the power cord into the receptacle on the back of the unit and t he other end to a power outlet.
Connect the USB cable to the BioLite and the other end to an available port on the computer.
2. Position the BioLite to the left of the darkroom.
Excitation Filter Installation
Filters are installed in a metal black casing from the UVP fac tory . To install the filters into the BioLite filter
wheel:
1. Hold the filter assembly so that it is positioned vertically with the filter hol der’ s ridge facing toward
the front of the BioLite as shown below.
BioLite
2. Slide the filter into the filter wheel as shown below. The small metal tab on the filter holder will
magnetically hold the filter in place within the BioLite. Note: Do not force the filter into the BioLite
wheel slot. If the filter is not fitting smoothly, ensure that the orientation of the fi lter assembly is
correct.
3. To access the other filter positions, manually rotate the filter wheel with the BioLite powered off.
BioLite Xe MultiSpectral Light Source 7
Epi Light Guides
Stopper
Fiber Optic Cable
Attach light guide s to these
Plug the Fiber
Epi Light Guide Installation (BioSpectrum®)
1. Remove the epi light covers at the top of the darkroom.
2. Remove one stopper at the left side of the darkroom.
3. Thread the knurled (non-smooth side) of the fiber optic cable from the inside of the darkroom
through the hole created by removing the stopper until the large bulb separating the smooth and
rough side of the fiber optic cable can no longer be pushed through the opening.
4. The light guides should be on the inside of the darkroom. Make sure the fiber optic cables extend
from the light guide toward the front of the darkroom (toward you). Match the drilled holes in the
light guide holders to the protruding screws on the epi light structure. Use the brass thumb nuts to
secure the light guides to the darkroom.
screws at top of darkroom
using the brass thumb nuts
5. Remove the rubber cap from the input side of the fiber optic trunk and plug the cable into the left
port on the BioLite unit.
Optic Cable into
the left port
BioLite Xe MultiSpectral Light Source 8
Light emitting
platform tray
Cable faces
the darkroom
Light guide fits
bracket
Epi Light Guide Installation (GelDoc-It
1. Remove one stopper at the left side of the darkroom.
2. Remove any existing plates or screws from the m ounting bracket so that the epi light guide bracket
appears as follows:
3. Thread the knurled (non-smooth side) of the fiber optic cable from the inside of the darkroom
through the hole created by removing the stopper until the large bulb separating the smooth and
rough side of the fiber optic cable can no longer be pushed through the openin g.
4. Position the light guide in the mounting bracket as shown below. Make sure the cable protruding
from the light guide faces toward the front of the darkroom (toward the darkroom door) and that the
light emitting portion of the light guide is facing toward the platform tray.
®2/TS2
and ChemiDoc-It
®2/TS2
)
the front of
5. Attach the mounting bracket plate to the mounting bracket using two of the screws provided, as
shown below. Make sure the tab on the plate faces up.
6. Repeat the process for the other light guide within the darkroom.
snugly in mounting
portion of the light
guide is facing
BioLite Xe MultiSpectral Light Source 9
Plug the Fiber
Mounting Plate
Bracket
Light Bulb
7. Remove the rubber cap from the input side of the fiber optic trunk and plug the cable into the left
port on the BioLite unit.
1. Remove one stopper at the left side of the darkroom.
2. If the imaging system already contains a transilluminator, unplug and remove the transilluminator.
3. Insert the fiber optic light table transill umi nator in the imaging system.
4. Thread the knurled (non-smooth side) of the fiber optic cable from the inside of the darkroom
through the hole created by removing the stopper until the large bulb s eparating the smooth and
rough side of the fiber optic cable can no longer be pushed through the opening.
5. Remove the rubber cap from the input side of the fiber optic trunk and plug the cable into the left
port on the BioLite unit.
Lamp Assembly Installation
The BioLite Xe is shipped from the factory with the xenon lamp assembly uninstalled. Prior to initial use, the
lamp assembly must be installed in the BioLite Xe as described below.
The BioLite Xe xenon light bulb is a lamp assembly which includes a mounting plate and bracket. The
assembly is installed through the bottom of the unit as shown:
1. Unplug the power cable and USB cable from the back of the BioLIte Xe.
BioLite Xe MultiSpectral Light Source 10
Three
Nuts
Attach the Male
2. The lamp assembly will have a wire coming from each end of the xenon bulb. One wire has a male
connector on the end and the other wire has a female connector. Attach the male connector from
the lamp to the female connector from the BioLite Xe unit, and vice versa.
CAUTION: DO NOT TOUCH THE GLASS OF THE BULB!
and Female
Connectors
3. Insert the lamp assembly into the bottom of the BioLite Xe, aligning the three all-thread scre ws on
the bottom of the BioLite Xe with the three holes in the lamp assembly mounting plate.
4. Reconnect the plate to the bottom of the unit using the three supplied nuts.
Mounting
5. Plug the power cable into the receptacle on the back of the BioLite Xe and the other end into a
power outlet.
6. Plug the USB cable into the back of the BioLite Xe and the other end into the computer.
BioLite Xe MultiSpectral Light Source 11
Using the Light Source
1. Set the power switch on the front of the system to the ON (I) position.
2. Use the VisionWorksLS “BioLite” module (shown below) to adjust the various BioLite settings:
a. Filter: Select the desired excitation filter position from the dropdown menu.
b. Light Engine: Turn the BioLite bulb on or off using the radio buttons (note that the
BioLite’s green power switch must be in the ON position to use the lighting controls).
When the bulb is on, the l ower indicator light to the left of the switch on the front of t he
BioLite will glow red.
c. Intensity: This setting regulates the amount of illumination in six increments. The relative
output levels are specified in the chart below:
Intensity Setting Relative Intensity
6 100%
5 50%
4 40%
3 25%
2 12%
1 0%
Light Table: Select between either Epi (overhead) or Trans (base lighting) depending on
the application. Note that both the epi and transillumination cannot be used
simultaneously.
NOTE: If using the darkroom to perform light-tight experiments such as capturing a
chemiluminescence sample for more than two minutes, use the black rubber cap to cover the end
of the fiber optic cable. Exposing an image for less than two minutes with the rubber cap removed
will not impact the resultant image, so long as the fiber optic cable is installed in the BioLite.
Without capping the fiber optic cable, light may enter the darkroom through the end of the fiber
optic cable, thus affecting the results of an extended resolution image.
3. It is possible to adjust the BioLite’s settings using the switches located on the front of the BioLite.
Note, however, that any settings modified using the switches will NOT be reflected in
VisionWorksLS’ BioLite software interface.
BioLite Xe MultiSpectral Light Source 12
Fiber Optic Lighting
Light Source Switch
Light Intensity Switch
Filter Position Switch
Selector
a. Fiber Optic Lighting: Press this button to select either Epi or Transillumination lighting.
Note that both epi and transillumination cannot be used simultaneously.
b. Light Source: Press this button to turn the xenon light bulb on or off. When the bulb is on,
the lower indicator light to the left of the switch will glow red.
c. Light Intensity: Press the rocker switch toward Max to increase the relative intensity of
the lighting. Inversely, press the rocker switch toward Min to decre ase the relativ e
intensity of the lighting. Lighting is provided in six increments, indicated on the LCD
display to the left of the rocker switch.
d. Filter Position: Press the rocker switch toward + to move to the next exc i tation filter
position. Inversely, press the rocker switch toward – to move to the previous excitation
filter position. There are a total of eight filter positions; the current position will be indicated
on the LCD display to the left of the rocker switch.
BioLite Xe MultiSpectral Light Source 13
Mounting Plate
Bracket
Light Bulb
Three
Nuts
Attach the Male
Service Procedures
Bulb Replacement
The BioLite Xe xenon light bulb is a lamp assembly which includes a mounting plate and bracket. The
assembly is installed through the bottom of the unit as shown:
1. Unplug the power cable and USB cable from the back of the BioLite Xe and allow the bulb at least
½ hour to cool.
2. The lamp assembly will have a wire coming from each end of the xenon bulb. One wire has a male
connector on the end and the other wire has a female connector. Attach the male connector from
the lamp to the female connector from the BioLite Xe unit, and vice versa.
CAUTION: DO NOT TOUCH THE GLASS OF THE BULB!
3. Insert the lamp assembly into the bottom of the BioLite Xe, aligning the three all-thread scre ws on
the bottom of the BioLite Xe with the three holes in the lamp assembly mounting plate.
4. Reconnect the plate to the bottom of the unit using the three supplied nuts.
and Female
Connectors
Mounting
BioLite Xe MultiSpectral Light Source 14
5. Plug the power cable into the receptacle on the back of the BioLite Xe and the other end into a
power outlet.
Return Procedure
A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before
returning any product. See “Technical Support” below for contact information.
Replacement Parts and Accessories
To order accessories or replacement parts for the BioLite MultiSpectral Light Source, contact UVP’s offices.
Blue Filter (to enhance red and yellow objects) 38-0319-03
Red Filter (darkens green and blue objects) 38-0319-04
Green Filter (increases for red or blue objects) 38-0319-05
Yellow Filter (safelight for photo resist materials) 38-0319-06
Other standard and custom filters are available for a variety of imaging applications. Contact UVP
or visit www.uvp.com for further information.
Troubleshooting
Bulb Does Not Illuminate
1. Recheck the main power cord connection to the BioLite unit.
2. Check the fuses located at the back of the unit next to the power port. A small flat-head screwdriver
or similar tool will be required. Turn the fuseholder cap counterclockwise and the fuse holder will
pop out. Inspect the thin wire within the glass fuse to see if there is a break in the wire. If so,
replace the fuse(s). If fuses are blowing repeatedly, contact UVP Technical Support for additional
troubleshooting.
3. The bulb’s wires may have come loose. Recheck the bulb electrica l connections by following the
“Bulb Replacement” instruction earlier in this manual.
4. The bulb may be burned out. Follow the “Bulb Replacement” instruction earlier in this manual.
.
Care and Cleaning
Use only mild soap or detergent solution for cleaning. Do NOT use oil- or petroleum-based cleaners for the
cabinet. Ensure that the system is turned OFF and unplugged during cleaning.
BioLite Xe MultiSpectral Light Source 15
Technical Support
UVP offers free lifetime technical support on all of its products and software. Should you have any questions
regarding the product’s use, operation or repair, contact UVP’s offices at the locations below, or visit
www.uvp.com.
If you are in North America, South
America, East Asia or Australia:
Call (800) 452-6788 or (909) 946-
3197, and ask for Technical Support
during regular business days, between
7:00 am and 5:00 pm, PST.
E-mail your message to:
info@uvp.com
Fax Customer Service, and send it to
(909) 946-3597
Write to: UVP, LLC 2066 W. 11th
Street, Upland, CA 91786 USA
If you are in Europe, Africa, the
Middle East of Western Asia:
Call +44(0) 1223-420022, and ask for
Customer Service during regular
business days between 9:00 am and
5:30 pm.
E-mail your message to: uvp@uvp.co.uk
Fax Customer Service, and send it to:
+44(0) 1223-420561
Write to: Ultra-Violet Products Ltd
Unit 1, Trinity Hall Farm Estate, Nuffield
Road, Cambridge CB4 1TG UK
VisionWorks, BioSpectrum, GelDoc-It and ChemiDoc-It are registered trademarks of UVP, LLC.
BioLite is a trademark of UVP, LLC.