UVP BioDoc-It Installation and User Instructions

BioDoc-It
® Imaging System
Installation and User Instructions
UVP, LLC Ultra-Violet Products Ltd.
2066 W. 11th Street Unit 1, Trinity Hall Farm Estate Upland, CA 91786 Nuffield Road, Cambridge CB4 1TG UK Phone: (800) 452-6788 / (909) 946-3197 Phone: +44(0)1223-420022 Fax: (909) 946-3597 Fax: +44(0)1223-420561

Web Site: www.uvp.com

81-0301-01 Rev E
BioDoc-It Imaging Systems 2

System Introduction

BioDoc-It® Imaging System

The BioDoc-It Imaging System enables simple documentation of fluorescent and non-fluorescent gels, membranes, blots, film plates and assays with the ability to save images to a USB storage device for later quantitative analysis or enhancement for publication.
Images are saved in 8-bit TIFF or JPEG formats, allowing saved im ages to be read by most PC or MAC programs. The BioDoc-It Imaging System is a cost effective solution for capturing quality images in a compact standalone pack age, as no external computer is required.
The BioDoc-It comes in two configurations:
BioDoc-It 220 Imaging System includes a high resolution, 1.3 megapixel camera, 8-48mm zoom
lens and side access doors.
BioDoc-It 210 Imaging System includes a 0.3 megapixel camera and an 8-48mm zoom lens.
BioDoc-It Imaging Systems 3
8” LCD Touch Screen
White Light and Darkroom Power Switches
Wide Access Door
USB Drive
Filter Tray (located under the Main System Power Button
UV Gel Viewer Window
Transilluminator
FluorCam 210 or 220
UV to White Light Converter

System Components

BioDoc-It Components

Refer to the packing slip and pictured components for parts included with the system.
with Tilt Capability
Camera and Zoom Lens
camera) with EtBr Filter
Plate (order separately)
Side Access Doors (on BioDoc-It 220 models)
BioDoc-It Imaging Systems 4

Camera

BioDoc-It 220 Models:
settings.

Lens

The BioDoc-It is equipped with an 8-48mm zoom
length of the BioDoc-It Imaging System’s darkroom.

Built-In Computer

Ethidium Bromide Filter (EtBr)

The Ethidium Bromide (EtBr) filter (50mm sq) UV
System Specifications
The BioDoc-It 220 is equipped with a FluorCam 220 Camera (resolution of 1.3MP).
BioDoc-It 210 Models:
The BioDoc-It 210 is equipped with a Fluor C am 210 Camera (resolution of 0.3MP).
Both the FluorCam 220 and FluorCam 210 use a USB 2.0 PC connection. All camera settings are factory pre-set for optimum performance when viewing gels, films, or membranes under low light level conditions. Contact UV P T echnic al Suppor t before making any adjustments to the camera
lens fitted with a close-up diopter and step-up ring, allowing the camera to be held in a fixed position. The diopter is for focusing on objects at the focus
Internal Storage: 60GB or more
USB 2.0 Ports: 1 front, 4 in the back USB Storage Device Capacity: 1GB (minimum) Software: TS software
blocking band pass interference filter blocks UV and IR radiation emitted from the transilluminator. The filter is placed in the filter tray below the camera/lens assembly.
The filter allows visualization of fluorophores from 580-630nm, targeting the EtBr emission peak of 605nm.
The EtBr filter can be substituted for other specific fluorophore filters or removed when imaging non­fluorescent media (protein gels, colony plates, etc.) in order to produce brighter images.
BioDoc-It Imaging Systems 5

Darkroom

Darkroom features include:
UV gel viewer window Overhead white light Wide access door with UV safety interlock switch

Transilluminator

A transilluminator is included w ith the BioDoc-It system. UVP offers a variety of transilluminators, from benchtop models with 8-watt, single wav eleng th and sin gle inte nsi ty to models with multiple wavelengths and high/low intensities. UVP also offers the high-end FirstLight®, a highly uniform, 302nm UV transilluminator.

LCD Touch Screen

The display is an 8-inch VGA color touch screen with tilt adjustment connected to the darkroom cabinet. The touch screen allows the user to preview, snap, save and print images, as well as select certain preference options, without the need to use an external mouse or keyboard. However, should the user desire, a mouse and/or keyboard (not included) can be used with the BioDoc-It.
A stylus pen is included for increased pointer control for selecting software options on the touch screen.

UVP Focus Target

The UVP Focus Target fluoresces when placed on a transilluminator or when exposed to overhead UV. The target provides sharp, fluorescent images to aid in adjusting the focus of the camera.

Removable USB Stick

The removable USB stick, included with the system, has 1GB memory (minimum) and connects to the system to allow for saving and transferring of images.
UVP Focus Target
BioDoc-It Imaging Systems 6
UV/White Converter Plate

Optional Equipment

Refer to the Replacement Parts and Accessories section of this manual for optional equipment part numbers.

Thermal Printer

The Thermal Printer provides archive quality, 256 grayscale prints and five optional cost-effective print sizes.

Converter Plates

The UV/White Converter Plate allows imaging of non-fluorescent stained media with an ultraviolet transilluminator. The converter plate is specially coated to convert 302nm UV output to white light.
The Visi-Blue™ Converter Plate (not shown) converts UV to 460­470nm and is designed for use with blue excitation samples such as SYBR Green, SYPRO Orange and GFP stains.
Thermal Printer
BioDoc-It Imaging Systems 7
2. The filter tray is located under the camera
3. Place the filter inside the filter tray. Ensure
The camera and zoom lens are assembled at the
2. If not already assembled, attach the step-up
Step-up ring
Diopter

Setup Instructions

Components

BioDoc-It Darkroom CCD Camera and Zoom Lens Camera/Lens Mounting Bracket and Hardware Power, Jumper and USB Cables Transilluminator USB Storage Device (1GB minimum)
WARNING: DO NOT ATTEMPT TO CONNECT ANY WIRING WHI LE THE EQUIPMENT IS CONNECTED TO THE POWER SUPPLY!
CAUTION: Do not install the system in places with high moisture, dust, or high temperatures. Do not use any oil or petroleum based cleaners on the cabinet. Use only a mild soap or detergent solution for cleaning. Ensure that the system is turned OFF and is unplugged from its power source during cleaning.
Keep the equipment away from motors or other large magnetic equipment apparatus.

Installing the Filter

Systems include a 50mm sq Ethidium Bromide (EtBr) filter. To install the filter in the BioDoc-It darkroom:
1. Carefully remove the filter from the protective plastic case, holding the filter at the edges to prevent fingerprints.
bracket on the top right side of the darkroom. Slide the tray open.
that the text on the edge of the filter is positioned so it is right side up when facing the installer.
Additional filters are available. Refer to Replacement
Parts and Accessories for ordering information.

Camera Setup

UVP factory.
Note: The zoom lens shipped may appear different than the one pictured at right.
1. Remove the cap from the lens.
ring and diopter to the lens. The step up ring and diopter will only fit one way.
BioDoc-It Imaging Systems 8
3. Using the four brass thumb screws provided,
4. Slide the camera and lens assembly into the
5. Insert and tighten the black thumb screw and
6. Plug the camera cable into the top of the
Camera and lens assembly
Camera USB cable
Brass thumb screw s
Connection for
Connection for the USB port
Camera bracket
screw
Base
secure the bracket to the base.
camera bracket. Slide the lens through the center hole in the gasket on the base and use the gasket to form a light-tight seal around the lens.
washer in the central hole in the camera to secure the camera to the bracket.
Black thumb
secure the bracket to the base
camera and the other end into a USB port on back of the darkroom.
7. Connect the jumper cord from the darkroom to the transilluminator.
8. Connect the main power cord from the back of the darkroom to a surge-protected power outlet.

Stylus and Holder

The BioDoc-It screen is touch sensitive. Use the stylus included with the system to operate the software controls. For easy access to the stylus when operating the system, remove the tape from the back of the stylus holder and adhere it to the BioDoc-It. Slide the stylus into the holder.
main darkroom power
jumper cord to the transilluminator
BioDoc-It Imaging Systems 9
1. Ensure that the main power cord is connected from
Main Power
Black

Using the System

Powering Up the System

the back of the darkroom to a surge-protected power outlet.
2. Power up the system by pressing once on the Main Power button located on the right side of the unit. All internal components will now have power and the TS software will automatically load.
3. To power down the system, push the Main Power button.

Adjusting the Screen Angle

1. To adjust the angle of the screen, loosen the black rubberized knobs on each side of the monitor. Twist each knob counter-clockwise (note that the knobs will be turning opposite directions).
2. Tilt the screen to the desired angle.
3. Turn each knob clockwise to tighten.
Rubberized Knob
Button
BioDoc-It Imaging Systems 10
Metal tab

Operating the TS Software

Once the Main Power button is pressed, the system will go through the boot-up process. Upon com plet ion, the LCD monitor will display the TS software screen similar to the one below:
To shut down the TS software, touch the X on the bott om left corner of the screen. To turn the system off, press the Main Power button or power down from the Windows icon on the bottom right
corner of the screen. Refer to the TS Software User Manual for further instructions on using the software.

Using the Transilluminator

Once the BioDoc-It darkroom power is turned on, power is supplied to all components. This includes the jumper cable that supplies power to the transilluminator. To use the transilluminator, turn ON the two transilluminator power switches. One switch is on the front of the BioDoc-It unit and is labeled UV Trans. The other switch is located on the front of the transilluminator itself, directly below the darkroom door.
Refer to the Transilluminator Manual for additional instructions on using the transill uminator.

Using the UV Gel Viewer Window

The UV Gel Viewer Window has a metal cover to block out ambie nt lighting and to protect the UV-safe glass. The cover is held in the closed position by two small magnets built into the UV-safe glass. To open the window cover, pull down on the metal tab. The cover will pop open. To close the cover, pivot it up until it reengages with the two small magnets.
The glass is UV blocking while providing a clear, UV-safe view to the transilluminator surface for visibility of samples.
BioDoc-It Imaging Systems 11
Aperture
Zoom

Min / Sec / mSec

Decrease
Increase

Image Focusing

Prior to capturing any images, prepare the image foc us cap a bilit ie s of the system:
1. Turn on the transilluminator and place the UVP Focus Targe t on the transilluminator surface.
NOTE: The BioDoc-It darkroom has a UV safety switch that turns
the transilluminator off when the door is open. After closing the door, be sure the overhead white light switch is turned OFF.
2. Press the LIVE button on the TS software screen. While watching the LCD screen, rotate the Aperture adjustment ring on the lens so that t he image is as bright as possible.
3. Using the integration time settings at the bottom of the screen, press the mSec time interval selector square. Then, use the + button to increase integration time until the desired brightness is achieved. Readjust the Aperture adjustment ring if needed.
4. Rotate the Focus adjustment ring on the lens. Adjust this so that the image appears in clear focus on the LCD monitor. Note: Once the proper zoom range is set, the lens will have to be refocused when zoomed in completely.
5. Rotate the Zoom adjustment ring on the lens so that the image is as large as possible. Readjust the Focus ring on the lens, making the image clear. Adjust the zoom so that the object of interest is within the picture on the LCD monitor.
Focus
Integration
Time
Time Interval Selectors
Integration
Time

Printer Setup

The thermal printer is an optional accessory. If this order did not include a thermal printer, move on to the next section. Otherwise, to set up the printer:
1. Plug the power cable from the printer into the power source.
2. Press the Open button to open the lid of the printer and load the paper by allowing the loose end of the roll to come off the top of the roll.
3. After shutting the printer lid, tear off the excess paper that is visible.
4. Plug in the USB cable from the printer into the BioDoc-It USB port.
5. With the BioDoc-It turned on, press the “Windows” button on the lower-right portion of the TS software screen.
6. To set the printer settings click “Start” at the lower left of the computer. Next, click “Printers and Faxes”, then right-click on the appropriate printer icon and select “Printing Preferences”.
7. Choose the following settings: High Density paper, 1280X1280, Portrait mode.
8. Click on the “Option” tab and click the “Enlarge Image to Fit Paper” box. Then, click “Apply” and “OK”.
9. Restart the BioDoc-it system.
If the thermal printer does not work correctly, contact UVP Technical Support for assistance. See the Technical Support section of this manual for contact infor ma tion .
BioDoc-It Imaging Systems 12

Service Procedures

Return Procedure

Replacement Parts and Accessories

Troubleshooting

A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before returning any product.
To order accessories or replacement parts for the BioDoc-It Imaging System, contact UVP’s offices.
Part Description Part Number
Filter, Ethidium Bromide, 50mm Square 38-0220-01 Filter, SYBR Green, 50mm Square 38-0219-01 Filter, SYBR Gold, 50mm Square 38-0221-01
White Light Converter Plate, 21x26cm 38-0191-01 White Light Converter Plate, 25x26cm 38-0191-04 Visi-Blue Converter Plate, 21x26cm 38-0200-01 Visi-Blue Converter Plate, 25x26cm 38-0200-04
Gel-Cutter 85-0002-01 Gel-Ruler 85-0003-01 Gel-Scooper 85-0006-01 Gel-Tray, small 85-0007-01 Gel-Sentry DNA Preparation Plate 97-0076-01 Fluorescent Standard Step Tablet 33-0014-02
®
LS Analysis Software 97-0185-02
Doc-It Spectacles, UV Blocking (UVC-303) 98-0002-01
Goggles, UV Blocking (UVC-503) 98-0002-02 Faceshield, UV Blocking (UVC-803) 98-0002-04
No Power to the Darkroom or Transilluminator
1. Recheck the main power cord connection to the BioDoc-It darkroom as well as the power cables
between the darkroom and transilluminator.
2. Check the fuses located at the back of the unit next to the power port. A flat-head screwdriver will
be required. Turn the fuseholder cap counterclockwise and the fuse holder will pop out. Inspect the thin wire within the glass fuse to see if there is a break in the wire. If so, replace the fuse(s). If fuses are blowing repeatedly, contact UVP Technical Support for additional troubleshooting.
Transilluminator Will Not Turn On
1. Make sure to turn ON the two transilluminator power switches and that both switches are glowing
green. One switch is on the front of the BioDoc-It unit and is labeled UV Trans. The other switch is located on the front of the transilluminator itself, directly below the darkroom door.
If the switches do not glow green, refer to “No Power to the Darkroom or Transilluminator” above.
2. Be sure the darkroom cabinet’s door is completely closed. There is a UV exposure safety cut-off
switch that turns the transilluminator off when the darkroom cabinet’s door is opened.
3. Be sure the transilluminator’s power jumper cord is securely connected at both the ends of the
cable.
BioDoc-It Imaging Systems 13
Touch Screen Not Accurate
1. If touching or clicking on the buttons becomes inaccurate, the touch screen may need to be
recalibrated. Please follow all instructions in the touch screen calibration instructions included in the TS software CD.
Error Messages Appear on the Screen
1. An error message that is related to the TS software interface or Microsoft Windows may appear on
the screen. If the message is related to Microsoft Windows, such as a reminder to activate or update the copy of Windows, please contact your system administrator for assistance.
2. If an error message appears repeatedly and your system administrator does not recognize it as a
Microsoft Windows error, contact UVP Technical Support for further assistance.
Technical Support
UVP offers technical support on all of its products. If there are questions about the product’s use, operation or repair, please contact UVP’s offices at the locations below. Or go to UVP’s web site and click the Tech Support > BioImaging Systems.
If in North America, South America,
East Asia or Australia:
Call (800) 452-6788 or (909) 946-
3197, and ask for Customer Service
during regular business days, between
7:00 am and 5:00 pm, PST.
E-mail your message to:
techsupport@uvp.com
Fax Customer Service, and send it to
(909) 946-3597
Write to: UVP, LLC 2066 W. 11th
Street, Upland, CA 91786 USA
If in Europe, Africa, the Middle East of
Western Asia:
Call +44(0) 1223-42002, and ask for
Customer Service during regular
business days between 9:00 am and
5:30 pm.
E-mail your message to: uvp@uvp.co.uk
Fax Customer Service, and send it to:
+44(0) 1223-420561
Write to: Ultra-Violet Products Ltd
Unit 1, Trinity Hall Farm Estate, Nuffield
Road, Cambridge CB4 1TG UK
Warranty
UVP's products are guaranteed to be free of defect s in materi al s, workmanship and manufacture for one (1) year from date of purchase. Consumable and disposable parts inc luding, but not limited to tubes and filters, are guaranteed to be free from defects in manufacture a nd materi als for ninet y (90) day s fr om da te of purchase. Transilluminators carry a two (2) year war ranty fr om date of purchase. If equipment failure or malfunction occurs during the warranty period, UVP shall ex amine the inoperativ e equipment and have the option of repairing or replacing any part(s) which, in the judgment of UVP, were originally defective or became so under conditions of normal usage and service.
No warranty shall apply to this instrument, or part thereof, that has been subject to accident, n egligence, a lt eration, abuse or misuse by the end-user. M or eov er, UVP makes no war ranties whatsoever with respect to parts not supplied by UVP or that have been installed, used and/or serviced other than in strict complia nce with instructions appearing in this manual.
In no event shall UVP be responsible to the end-user for any incidental or consequential damages, whether foreseeable or not, including but not limited to property damage, inability to use equipment, lost business, lost profits, or inconvenience arising out of or connected with the use of instruments produced by UVP. Nor is UVP liable or responsible for any personal injuries occurring as a result of the use, installation and/or servicing of equipment.
This warranty does not supersede any statutory r ights that may be av ailable in certain countries.
BioDoc-It, Doc-It and FirstLight are registered trademarks of UVP, LLC. Visi-Blue is a trademark of UVP, LLC.
Loading...