UVP 81-0357-01 User Manual

PCR Workstations and Cabinets
Installation and User Instructions
Web Site: www.uvp.com

81-0357-01 Rev STD

PCR Workstations and Cabinets Page 2
UV1 PCR Cabinet
UV2 PCR Cabinet
UV1 PCR Workstation
UV2 PCR Workstation
95-0437-01 115V
95-0437-03 100V
95-0436-01 115V
95-0436-03 100V
95-0367-01 115V
95-0367-03 100V
95-0439-01 115V
95-0439-03 100V
Light Sources:
Light Sources:
Light Sources:
Light Sources:
Chamber
Features include:
Features include:
UV/Air Circulator
Features include:
Optional: UV/Air Circulator
Features include:
UV/Air Circulator
Dimensions:
in. (500 x 544 mm)
Dimensions:
in. (500 x 544 mm)
Dimensions:
in. (706 x 544 mm)
Dimensions:
in. (706 x 544 mm)

Introduction

The UV1 and UV2 PCR Hoods create an ideal environment for preparing PCR and other samples by reducing the chance of contamination. The built-in high intensity shortwave (254nm) UV provides a source for inactivation of DNA between experiments. Additional contamination control is provided by the specially-coated stainless steel and aluminum design that maintains antimicrobial efficacy. The surface resists growth of destructive bacteria, molds and fungi. The Makrolon assembly blocks UV below 400nm.
In addition to the UV1/UV2 PCR Hood features, HEPA/UV3 PCR Hoods provide additional protection by blowing filtered air, treated by the HEPA/UV system, into the PCR chamber.
The UV/Air Circulator, included as standard on UV2 and HEPA/UV3 PCR Hoods and optional on the UV1 PCR W orkstation (95-0367-xx only), eliminates amplicons of DNA or genomic DNA from a previously-dispensed experiment. For ordering information, refer to the “Maintenance and Replacement Parts/Accessories” section.
®
panel

Safety Information

The PCR Hoods are designed with functionality, reliability and safety in mind. The Hoods provide shortwave (254nm) UV which is a powerful source of UV radiation that will cause damage to unprotected eyes and skin. Before operating, ensure that all personnel are properly protected and that the instructi ons for the use of this equipment are followed. A safety shut-off switch automatically turns off the UV light when the door is open, protecting users from UV exposure. In addition, the Makrolon panels are specially formulated to block UV wavelengths below 400nm.
Disconnect the power supply before assembling or servicing the PCR Hoods.
®
95-0437-02 230V UK 95-0437-04 230V Euro
254nm 8-watt UV:
Chamber
White 8-watt
Chamber
Two power outlets Two small shelves UV Timer Antimicrobial coated
stainless steel (work
surface) & aluminum Makrolon block below 400nm
Exterior: 28.7H x 21.4W x 24D in. (729 x 544 x 610 mm) Interior: 19.7W x 21.4D

Specifications

STANDARD UV (UV1 and UV2) PCR MODELS
95-0436-02 230V UK 95-0436-04 230V Euro
254nm 8-watt UV:
White 8-watt:
Two power outlets Two small shelves UV Timer Antimicrobial coated
®
panels
Makrolon
Exterior: 28.7H x 21.4W x 24D in. (729 x 544 x 610 mm) Interior: 19.7W x 21.4D
Chamber UV/air circulator
Chamber
stainless steel (work surface) & aluminum
®
panels block
below 400nm
95-0367-02 230V UK 95-0367-04 230V Euro
254nm 25-watt UV:
Chamber
White 15-watt:
Chamber
Four power outlets Two shelves UV Timer Antimicrobial coated
stainless steel (work surface) & aluminum
®
Makrolon
below 400nm
Exterior: 28.7H x 29W x 24D in. (729 x 737 x 610 mm) Interior: 27.8W x 21.4D
panels block
95-0439-02 230V UK 95-0439-04 230V Euro
254nm 25-watt UV:
Chamber
254nm 8-watt UV:
UV/air circulator
White 15-watt:
Four power outlets Two shelves UV Timer Antimicrobial coated
stainless steel (work surface) & aluminum
®
Makrolon
below 400nm
Exterior: 28.7H x 29W x 24D in. (729 x 737 x 610 mm) Interior: 27.8W x 21.4D
panels block
PCR Workstations and Cabinets Page 3
Light Sources:
Chamber
Light Sources:
Chamber
Features include:
UV/Air Circulator
Features include:
UV/Air Circulator
Antimicrobial coated stainless steel (horizontal work surface) and aluminum
Exterior
Aluminum powder coated
HEPA/UV3 PCR MODELS
HEPA UV3 PCR Cabinet HEPA UV3 PCR Workstation
95-0434-01 115V 95-0434-02 230V UK 95-0434-04 230V Euro 95-0434-03 100V
254nm 8-watt UV:
Filter area Chamber UV/air circulator
White – 8-watt:
Three stage filters:
Pre-filter Carbon filter HEPA filter
Two power outlets Two small shelves UV Timer Antimicrobial coated stainless steel (work
surface) and aluminum
®
Makrolon below 400nm
Dimensions:
Exterior:
32.5H x 21.4W x 24D in. (826 x 544 x 610 mm) Interior: 19.7W x 21.4D in. (500 x 544 mm)
panels block
95-0438-01 115V 95-0438-02 230V UK 95-0438-04 230V Euro 95-0438-03 100V
254nm UV:
Filter area 8-watt Chamber 25-watt UV/air circulator 8-watt
White – 8-watt:
Three stage filters:
Pre-filter Carbon filter HEPA filter
Four power outlets Two shelves UV Timer Antimicrobial coated stainless steel (work
surface) and aluminum
®
Makrolon below 400nm
Dimensions:
Exterior:
32.5H x 29W x 24D in. (826 x 737 x 610 mm) Interior: 27.8W x 21.4D in. (706 x 544 mm)
panels block
Shelves
Construction Dimensions (Workstation) Dimensions (Cabinet)
Construction
Interior
Door and Side Panels
Electrical
115V - 60 Hz, 230V - 50 Hz, 100V - 50/60Hz
Two - Formed antimicrobial-coated aluminum 13”W x 4.2”D (330 x 107mm)
8.75” x 4”D (222 x 102mm)
Makrolon® panel assembly Blocks UV under 400nm
PCR Workstations and Cabinets Page 4
Max
95-0437-01
95-0438-05
115V
230V
9.1
8.7

Operational Ratings

PCR Hoods are intended to be used indoors only
Altitude must not exceed 2,000m
Ambient temperature must not exceed 5°C to 40°C (41°F to 104°F)
Relative humidity must not exceed 80% for temperatures up to 31°C (88°F) decreasing linearly to
50% relative humidity at 40°C (104°F)
Mains supply voltage fluctuations must not ex ceed ±10% of the nominal voltage
Pollution rating 2 or better for laboratory areas
Part Number Voltage
95-0437-02 95-0437-04 95-0437-03 95-0437-05 95-0436-01 95-0436-02 95-0436-04 95-0436-03 95-0436-05 95-0367-01 95-0367-02 95-0367-04 95-0367-03 95-0367-05 95-0439-01 95-0439-02 95-0439-04 95-0439-03 95-0439-05 95-0434-01 95-0434-02 95-0434-04 95-0434-03 95-0434-05 95-0438-01 95-0438-02 95-0438-04 95-0438-03
230V 230V 100V 230V 115V 230V 230V 100V 230V 115V 230V 230V 100V 230V 115V 230V 230V 100V 230V 115V 230V 230V 100V 230V 115V 230V 230V 100V
Amps
8.8
8.8
9.1
8.8
9.1
8.7
8.7
9.1
8.7
8.8
8.6
8.6
8.8
8.6
9.1
8.7
8.7
9.1
8.7
9.1
8.7
8.7
9.1
8.7
9.1
8.7
8.7
9.1
PCR Workstations and Cabinets Page 5

Assembling the PCR System

Carefully remove the PCR Hood components from the packaging. CAUTION: Use care when unboxing the PCR Hood, as staples with sharp ends may have been used in packaging.
Follow the assembly instructions provided in this manual. NOTE: Place the PCR Hood Base on a level surface large enough to accommodate the entire Base. The surface must be level to ensure proper alignment of the Makrolon
Required Tools
Philips head screwdriver (motorized, if available)
Flathead screwdriver (motorized, if available)
Gloves, to reduce transfer of fingerprints to PCR Hood components
Two people for assembly and lifting of the equipment
NOTE: Assembly drawings show the 95-0436-01 UV2 PCR Cabinet. However, these instr uctions apply for all PCR models.
Outer Back Panel
1. Separate the Inner and Outer Back Panels by removing the 4 screws, as shown: NOTE: The Outer and Inner Back Panels and Shelf Bracket are assembled at the factory for
shipping purposes.
®
panels.
2. Loosen the two screws on the Base Assembly, as shown:
PCR Workstations and Cabinets Page 6
Screw
Screw
Base
Outer Back Panel. Note the notch in
the top of the panel allows room for
NOTE: It is advised to use the remaining bubble wrap to protect the antibacterial-coated surfaces during assembly.
3. Attach the Outer Back Panel to the Base Assembly with two screws, as shown below. Leave the screws loose (do not fully tighten at this time).
the wires.
PCR Workstations and Cabinets Page 7
Inner Back Panel
Shelf Bracket
Position the Inner Back
the frame rests against the
Cutout for wires
Inner Back Panel
Inner Back Panel
Screw
Shelf Bracket
(8 Shown)
Inner Back Panel
1. Position the Inner Back Panel so that the screw holes align with the Outer Back Panel. Attach the Inner Back Panel to the Outer Back Panel with two (2) screws, as shown. Make sure the wires come out of the cutout in the back panel, as shown by the red circle below, for later conne ction to the Top Assembly.
2. Secure the Shelf Bracket to the front of the Inner Back Panel using:
a. Eight (8) screws in the Workstation or b. Six (6) screws in the Cabinet
NOTE: Two people are required to hold the panels in place and insert the screws. Leave the screws loose (do not fully tighten at this time).
Screw
Makrolon
Screws
Panel so that the edge of
stainless steel b ase
®
Panel Assembly
1. Slide the Makrolon Panel so that the Frames are under the Shelf Bracket. Note that the Panel Frames are oriented so that the left Frame cannot be used on the right side and vice versa. See illustrations below.
2. Remove the protective film from the Makrolon
3. Slide the Panels into position between the Frames and the lips on the sides of the Inner Back Panel. Ensure that the Makrolon
4. Press the Frames tightly against the Makrolon
®
Panel Frames for the Makrolon® Panels under the screws on the Inner Back
®
Panels.
®
Panels are pushed all the way against the Inner Back Panel.
®
Panels.
PCR Workstations and Cabinets Page 8
Makrolon® Panel
Makrolon® Panels
Shelf Bracket
5. Tighten all Frame screws to secure the Makrolon® Panels in place.
Frames
Top Assembly and Shelf Assembly
1. Unwrap the Top Assembly.
2. Rest the Assembly on top of the Back Panel Assembly and side Makrolon
CAUTION: Top is heavy! WARNING: Ensure that the unit is not plugged in.
3. Pull all of the colored wires out of the Top Assembly.
4. Connect the bottom and Top Assembly wires together according to their labels, matching wire “A” with wire “A,” wire “B” with wire “B” and so forth, until all labeled wires are connected. Connect the wires by joining the male and female red connectors located on the ends of each wire. Then, connect the two unlabeled green/yellow ground wires together using the red connectors.
After all connections are made, slightly pull each connection to ensure that all connectors are seated together securely.
®
Panels.
PCR Workstations and Cabinets Page 9
Flange
Screw locations
Shelf Bracket
Top Assembly
5. Tuck all of the connected wires between the Inner and Outer Back Panels.
6. Slide the Top Assembly down so that the side Makrolon channels. Make sure that the back flange of the Top Assembly is positioned between the Inner Back Panel and Outer Back Panel.
7. Secure and tighten the Top Assembly to the Back Panel Assembly with screws, pushing from the back to ensure that the screws are secure.
8. Slide the Shelves into the Shelf Bracket.
®
Panels are inside the Top Assembly
PCR Workstations and Cabinets Page 10
Thumb Knobs
Door Assembly
Side Brackets
Magnet
Top Assembly
Door Hinges
Door Brackets
Door Assembly
1. Remove the protective film from the Makrolon® panels.
2. Slide the upper panel of the Door Assembly into slotted area in the Top Assembly. Holding the door in position, use the black thumb screws (qty. 3) to attach the upper door panel to the Top Assembly. NOTE: To help the screw holes align, slide the Door Assembly fully up into the Top Assembly, then slightly lower the Door Assembly.
3. Ensure that the Side Panels are pushed as far back into the Back Panel as possible. Then, secure the upper panel of the Door Assembly to the left and right side panels with the two side brackets. If the brackets and holes don’t align, it may help to push up on the bottom of the Door Assembly until the holes align. NOTE: The brackets are side-specific, meaning th at each bracket will only fit on one side of the panels.
4. A magnet is attached to the bottom right bracket on the door. The purpose of the magnet is to shut off the overhead UV light when the door is open.
Note: Some systems may have a single large bracket running along the width of the entire lower door, rather than the two small brackets shown below.

When the PCR Hood is assembled, plug the power cord into the back of the unit and then into a wall outlet in the laboratory.

WARNING: The PCR Hood is for laboratory use and cannot be connected to a public power supply.
PCR Workstations and Cabinets Page 11
115V
230V UK
230V Euro

Operating the PCR Hood

The PCR Hoods feature:
Three-stage HEPA filtration system with pre filter, carbon filter and HEPA filter (UV3 models only)
Overhead UV for decontamination of the chamber or PCR equipment
UV safety switch and keylock for preventing exposure to UV
Antimicrobial coated stainless steel and aluminum for reducing bacterial growth
Easy-access door
Overhead white light
Shelves for placement of tools
Power outlets built into the chamber interior
Built-in UV/Air Circulator (standard on UV2 and UV3 models; optional for UV1 PCR Workstation)

Use of PCR Equipment in the Chamber

Shelves are provided for placement of small tools for decontamination. The Shelves slide into the Bracket on back of the unit.
Power outlets built into the chamber allow for use of shakers, rockers and other equipment for PCR experiments. Equipment can be decontaminated between uses. To operate equipment within the chamber, plug the equipment into the power plug. Specific power plugs are installed as shown in the illustrations below.

NOTE: W hen operating equipment plugged into the outlets, the combined amperage total is not to exceed 8 amps.

Operating the Overhead UV and Fluorescent White Lights

UV Keylock
The UV PCR Hoods provide an automated process for eliminating contamination by utilizing the germicidal properties of shortwave 254nm ultraviolet light. The UV light is typically operated when no samples are inside the chamber. The chamber is decontaminated, eliminating viable fungi, bacteria, and yeast.
An optional UV/Air Circulator can be mounted into the UV1 PCR Workstation chamber for elimination of amplicons of DNA or genomic DNA from a previously dispen sed experiment.
PCR Workstations and Cabinets Page 12
Power Switch for
Overhead
UV Keylock
30-Minute
UV Timer
Timer
Power Switch
Circulator
Overhead
30-Minute
UV Timer
Timer
Button Hole
HEPA/UV3 Control Panel:
Calibration
Button Hole
UV/Air Circulator
UV1 and UV2 Control Panel:
for UV/Air
White Light
White Light
To use the UV decontamination function, turn the UV Keylock ON, then press the UV Timer to operate the germicidal lamp housed within the chamber. The Timer may take a few seconds to initialize once the Timer has power. The Timer will only operate when the door is shut and the magnetic safety switch on the door is engaged.
The default timer setting is 30 minutes, adjustable at 5 minute increments from 5 to 60 minutes, then at 15 minute increments up to 12 hours. To adjust the timer settings:
1. Turn the system on and wait until the timer reads “OFF”.
2. Locate the Timer Calibration Button in the small hole to the left of the timer. Use a small paper clip to press the button for one second.
3. Press the main timer button to adjust the timer settings. Each press will increase the amount of time by 5 minutes (eg 35:00, 40:00, etc.) until one hour is reached. After one hour is reached, the time will be adjuste d in 15-minute increments and will read without a colon. For ex ampl e , one hour will read “1 00,” one hour and 15 minutes will read “1 15” and so forth.
4. Once the desired time setting is reached, do not press any additional buttons. The timer will then cycle through the remainder of its settings. When the timer reads “OFF” for 5 seconds without blinking, the timer is ready to use with the new time setting.
Turn the key to the Off position after UV irradiation to prevent exposure of samples to the UV
Calibration
NOTE: A safety switch sensor is built into the bottom of the door. When the door is opened, the ultraviolet lights will automatically shut off.
WARNING: Exposure to UV light is harmful.
PCR Workstations and Cabinets Page 13
Decontaminated UV airflow into and out
The Makrolon® door is UV blocking and will not allow UV radiation to pass through the door. The germicidal tubes will shut off if the door is opened. The PCR Workstation contains a powerful source of UV radiation that will cause damage to unprotected eyes and skin. Before operating any unit, be sure all personnel in the area are properly protected. Even though the unit shuts the UV off when the door is open, UV Blocking Eyewear should be worn as well. UVP has a complete line of UV Blocking Eyewear, including Spectacles, Goggles and Faceshield, designed for this purpose.
Extended periods of UV exposure can be used to disinfect the interior chamber by the destruction of organisms. Refer to the Maintenance section for information on measuring the intensity of the UV tubes with a UV meter.
NOTE: Do not attempt to perform PCR procedures with the germicidal lamp on. The germicidal lamp is used to decontaminate the internal chamber in between experiments.
Overhead White Light
The overhead white light can remain lit at any time throughout experiment procedures or the decontamination process. The fluorescent white light provides a bright light within the PCR Hood. To use the white light, press the ON button located on the top panel.

Operating the UV/Air Circulator

The UV/Air Circulator is built into the UV2 and HEPA/UV3 PCR Hoods.
To operate the Circulator, press the UV/Air Circulator switch to the ON position. The Circulator will pull air from inside the chamber through the Circulator, decontaminating the air with shortwave UV and blowing the air back into the chamber.
NOTE: Do not place equipment in front of the air vents, as this will block air flow.
An optional UV/Air Circulator module can be mounted into the top of the UV1 PCR Workstation (95-00367-XX only) for elimination of amplicons of DNA or genomic DNA from a previously dispensed
experiment. The power switch, outlet, and mounting brackets for the Circulator are pre-installed at the factory if users purchase the Circulator at a later time.

Operating the HEPA/UV System (HEPA/UV3 Model s Only)

Push the HEPA/UV power switch ON to operate the HEPA/UV filter system. The airflow fan setting can be adjusted from Low to High, circulating anywhere from 22 to 50 feet per minute (FPM) of air. The filte r system consists of:
Pre-Filter to preserve the life of other filters by capturing large dust particles
Carbon Filter to remove ozone, gases, odor
HEPA Filter to provide a barrier (99.99%) against dust, bacteria and mold down to 0.3 micron
particles (NOTE: The HEPA filter does not prevent particulates from entering the workstation when the Makrolon
®
door is open).
of the chamber
PCR Workstations and Cabinets Page 14

Care and Cleaning

The PCR Hoods are built to provide trouble-free operation.
NOTE: The door and side panels are made of Makrolon® and are subject to scuffing and scratches if improperly cleaned.
NOTE: Crazing is a normal process for Makrolon® panels exposed to UV light. Crazing will occur over a period of time. Reduce crazing by keeping UV exposure to a minimum. Crazing may occur within the warranty period and is regarded as normal wear which is not covered by the warranty. If crazing of the Makrolon part numbers.
The stainless steel and aluminum surfaces are manufactured with an antimicrobial coating to reduce bacterial growth. Care in cleaning and use of the equipment is recommended to reduce wear of the coating. To clean the unit:
Wipe excess water from inside the unit and outside the unit with an absorbent soft cloth or sponge.
Use mild soap and a damp, soft cloth or damp sponge to clean the exterior of the unit.
Use a mild detergent on the interior surfaces.
Warning: Unplug the unit before cleaning around the UV tube contacts and internal plug
outlets.
Clean the Makrolon® panels with a mild detergent; never use organic-based compounds or cleaners containing alcohol or ammonia.
Surface decontaminants such as RNase AWAY laboratory surfaces) and DNA AWAY™ (used to remove DNase and DNA contamination) are safe to use on PCR Hood panels.
For general cleaning of the Makrolon is available from local plastic supply distributors.
Do not use abrasive pads or cleansers.
®
panel occurs, the panel can be replaced. Refer to Replacement Parts for
®
(used to eliminate RNase and DNA from
®
panels, a plastic cleaner solution is recommended and
PCR Workstations and Cabinets Page 15
System/Part Number
Replacement Tube
Tube Description
Qty.
UV PCR Cabinet
34-0007-01
8W 254nm UV
4
95-0437-xx
34-0056-01
8W White
2
UV PCR Workstation
34-0073-01
25W 254nm UV
2
95-0367-xx
34-0087-01
15w White
1
UV2 PCR Cabinet
34-0007-01
8W 254nm UV
5
95-0436-xx
34-0056-01
8W White
2
UV2 PCR Workstation
34-0007-01
8W 254nm UV
1
95-0439-xx
34-0073-01
25W 254nm UV
2
34-0087-01
15w White
1
UV3 HEPA PCR Cabinet
34-0007-01
8 Watt 254nm UV
6
95-0434-xx
34-0056-01
8 Watt White
2
UV3 HEPA PCR Workstation
34-0007-01
8 Watt 254nm UV
2
95-0438-XX
34-0056-01
8 Watt White
2 34-0073-01
25 Watt 254nm UV
2

Maintenance and Replacement Parts/Access ori es

Replacement Parts and Accessories

Replacement parts are listed below. Only authorized UVP service personnel should perform repairs or replacements other than specified in the procedures within this user manual.
Replacement Parts Part Number
Fuse (2 required) 56-0049-01 Shelf, small 8.8”W (for cabinets) 20-1209-01 Shelf, large 13”W (for workstations) 20-0970-02 Key, replacement 53-0178-04 Pre-filter (HEPA/UV3 Models Only) 38-0314-01 Carbon-filter (HEPA/UV3 Models Only) 76-0352-01 HEPA filter (HEPA/UV3 Models Only) 17-0113-01
Makrolon
Side Panels (2 Required) 10-0618-01 Upper Door Panel (Workstation) 10-0619-01
Lower Door Panel (Workstation) 10-0620-01 Upper Door Panel (Cabinet) 10-0621-01
Lower Door Panel (Cabinet) 10-0622-01
Replacement Tubes
®
Panels
Accessories Part Number
Table, stainless steel top 98-0077-01 Shelf, long 27” (for Workstations) 20-0970-01 UV/Air Circulator Module (for 95-0367-xx) 97-0157-xx UVX Radiometer 97-0015-02 UVX Sensor (UVX-25) 97-0016-01 J-225 Meter 97-0004-01 Spectacles, UV blocking (UVC-303) 98-0002-01 Goggles, UV blocking (UVC-503) 98-0002-02 Face Shield, UV blocking (UVC-803) 98-0002-04
PCR Table
PCR Workstations and Cabinets Page 16
Turn the system power off and unplug the unit from the outlet. Open the Door and locate the tubes.
Note that UV tubes are clear and white light tubes are white.
Pre-Filter
UV/Air

Replacing the Fuses

Fuses are located on lower corner of the system. To remove the fuse, use a flathead screwdriver to turn the fuseholder counter-clockwise. Pull the fuse out.
Insert the new fuse and, using the screwdriver, turn the fuseholder clockwise to lock.

Replacing Tubes in the Chamber

Tube has two prongs on each end that first fit into a socket th en rota te i nto pla ce. Car efully hold the tube on the end and insert the tube into the socket. Twist the tube 90° until the tube locks into place.

Replacing the UV/Air Circulator Tube

1. Locate and remove the two thumb screws on the Circulator cap on the outer back panel of the system.
2. Pull the Circulator cap away from the system.
3. The tube, attached to the cap, has two prongs on each end that fit into a socket. Carefully hold the tube on the ends, rotate the tube 90° and remove from the cap. Insert a new tube into the sockets. Twist the tube 90° until the tube locks into place.
4. Place cap back into position and reattach using the two thumb screws.

Replacing the Filters (HEPA/UV3 Models Only)

Recommended filter replacement schedule to maintain optimum filter performance:
HEPA filter: Every 6 months or when significant drop in airflow is observed Carbon filter: Once a year Pre-filter: Once a year
1. To replace any of the filters or the UV tube in the HEPA/UV system, unscrew knobs (2) located on the ceiling of the workstation to release compression in the HEPA/UV system.
2. Open the HEPA/UV door located on the left of the system to access the filters.
3. Pre-Filter: The Pre-Filter (38-0314-01) is located on the inside of the HEPA/UV door. To remove, open the door and slide the filter off of the door.
Circulator
Cap
PCR Workstations and Cabinets Page 17
Knobs
UV Tube
Carbon
Filter
HEPA
Filter
Thumb
Screws
4. Carbon Filter: Slide the filter assembly out of the chamber. The Carbon Filter (76-0352-01) then slides to the right out of the filter assembly. If the filter does not slide out easily, loosen the thumb screws in front of the filter.
5. HEPA Filter: To replace the HEPA filter, slide the filter assembly out of the chamber and remove the carbon filter if not already removed. Turn the knobs counterclockwise to loosen. Open the HEPA door.
Slide the HEPA filter out and replace filter. Close the door and tighten the knobs (turn clockwise). Insert the carbon filter. Tighten the carbon filter thumbscrews if loose. Close filter door. Retighten the knobs (2) on the workstation ceiling.

Replacing UV Tube in the Filter System (HEPA/UV3 Models Only)

Follow steps 1 and 2 under “Replacing the Filters” and remove the carbon filter as described in step 4 above. Step 5 shows location of the UV tube.
The tube has two prongs on each end. Twist the tube 90° then remove the tube from the sockets. To install a new tube, carefully hold the tube on each end and insert the tube into the socket. Twist the tube 90° until the tube locks into place.

Replace the carbon filter and retighten the knobs (2) on the ceiling of the PCR Hood.

PCR Workstations and Cabinets Page 18
Microwatt
seconds/cm2
Microwatt
seconds/cm2
Bacillus anthracis
8700
YEAST
S. enteritidis
7600
Saccharomyces ellipsoideus
13200
B. Megatherium sp. (veg.)
2500
Saccharomyces sp.
17600
B. Megatherium sp. (spores)
5200
Saccharomyces cerevisiae
13200
B. parathyphosus
6100
Brewer’s yeast
6600
B. subtilis
11000
Baker’s yeast
8800
B. subtilis spores
22000
Common yeast cake
13200
Clostridium tetani
22000
Corynebacterium diptheriae
6500
MOLD SPORES
Eberthella typosa
4100
Penicillium roqueforti
26400
Escherichlia coli
6600
Penicillium expansum
22000
Micrococcus cadidus
12300
Penicillium digitatus
88000
Micrococcus sphaeroides
15400
Aspergillus glaucus
88000
Mycobacterium tuberculosis
1000
Aspergillus flavus
99000
Neisseria catarrhalis
8500
Aspergillus niger
330000
Phytomonas tumefaciens
8500
Rhisopus nigricans
220000
Proteus vulgaris
6600
Mucor racemosus A
35200
Pseudomonas aeruginosa
10500
Mucor racemosus B
35200
Pseudomonas fluorescens
6600
Oospora lactis
11000
S. typhimusium
15200
Salmonella
10000
VIRUS
Sarcina lutea
26400
Bacteriophasge (E. coli)
6600
Sarratia marcescens
6160
Tobacco mosaic
44000
Dysentery bacilli
4200
Influenza
6600
Shigella paradyseneriae
3200
Spirillum rubrum
6160
PROTOZOA
Staphlococcus albus
5720
Paramecium
200000
Staphylococcus aereus
6600
Nematode eggs
9200
Streptococcus hemolyticus
5500
Chlorella vulgaris (algae)
22000
Streptococcus lactis
8800
Streptococcus viridans
3800

Measuring the UV Tube Intensity

As the ultraviolet tubes age, the intensity and germicidal destruction rate decreases. It is important to monitor the efficiency of the tubes to ensure that the germicidal requirements are met. The germicidal destruction rate is a function of the UV intensity at a wavelength of 254nm and the exposure time. The lower the tube intensity, the longer the lamps must be on to accomplish the same objective.
The UVX Radiometer with a 254nm sensor will allow the user to measure the 254nm emission from the tubes. When it is time to measure the UV intensity inside the PCR Hood (recommended weekly or as needed), the sensor can be placed on the floor of the Hood. Close the Hood and turn on the UV tubes. Wait 3 minutes to record the measurement on the UV meter. It is recommended that the user record an initial value upon receipt and assembly of the PCR unit for later comparison purposes.
The germicidal destruction rate calculation:
Microwatt seconds/cm2 = microwatts/cm2 x seconds of exposure
For ordering information on the UVX Radiometer, J-225 meter and sensors, refer to the
Replacement Parts and Accessories section in this manu al. Bacterial Destruction Chart: The bacterial destruction chart below indicates the amount of shortwave
(254nm) UV energy required for the complete destr uct ion of v arious orga nis ms .
Bacterial Organisms
Additional Organisms
PCR Workstations and Cabinets Page 19

Technical Assistance

Note: A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before returning any product.
UVP offers free lifetime technical support on all of its products and software. Should you have any questions regarding the product’s use, operation or repair, contact UVP’s offices at the locations below,
www.uvp.com.
or visit
If you are in North America, South
America, East Asia or Australia:
Call (800) 452-6788 or (909) 946-
3197, and ask for Customer Service
during regular business days, between
7:00 am and 5:00 pm, PST.
E-mail your message to:
info@uvp.com
Fax Customer Service, and send it to
(909) 946-3597
Write to: UVP, LLC 2066 W. 11th
Street, Upland, CA 91786 USA
If you are in Europe, Africa, the
Middle East of Western Asia:
Call +44(0) 1223-420022, and ask for
Customer Service during regular
business days between 9:00 am and
E-mail your message to: uvp@uvp.co.uk
Fax Customer Service, and send it to:
+44(0) 1223-420561
Write to: Ultra-Violet Products Ltd
Unit 1, Trinity Hall Farm Estate, Nuffield
Road, Cambridge CB4 1TG UK
Warranty
UVP, LLC warrants its products to be free of detects in materials and workmanship for a period of two (2) years from the date of purchase. The foregoing warranty of UVP shall be of no force and effect if buyer has modified or damaged the product. Tubes and filters are warranted for 90 days.
NOTE: “Crazing” (cracking and/or yellowing) of the Makrolon period. This is regarded as normal wear and tear and is not covered by the warranty.
All warranties or merchantability and fitness for any purpose and all other warranties, expressed or implied, except those expressly set forth herein, are deemed waived and excluded.
UVP’s duty under the warranty is limited to replacement and/or repair of the defective part at the option of UVP, LLC. UVP shall not be liable for any expenses or damages incurred by the purchaser except as expressly set forth herein, and in no event shall UVP be liable for any special, incidental or consequential damages of any kind. This warranty does not supersede any statutory rights that may be available in certain countries.
Makrolon is a registered trademark of Bayer AG. RNase AWAY is a registered trademark of Molecular Bio-Products, Inc. DNA AWAY is a trademark of Molecular BioProducts, Inc.
5:30 pm.
®
panels may occur within the warranty
Loading...