Universal Enterprise C127 User Manual

Page 1
C125 & C127 SENSOR CALIBRATION & REPLACEMENT
QUICK START GUIDE
Calibration Instructions;
Note: Disclaimer of Liability
These instructions are prepared for use by trade professionals familiar with gas analysis and detection, and are familiar or have experience in calibrating gas measurement systems. They are intended to be as complete and simple as possible to eliminate the chance for error. Certain factors related to analyzer field recalibration are beyond the control of UEi, and the user should be aware of these factors such as proper delivery of gas to the analyzer during calibration, the age of the calibration gas, and the accura­cy of the calibration gas.
By using these instructions to recalibrate your instrument, you assume all risks and liability associated with the measurement of carbon monoxide (CO) gas concentrations.
Indemnification
esentatives,
epr
zer, you
ate your analy
xtent permitted b
o the e
T
ee to defend, indemnify, and hold harmless, U
r
ag
ents from and against all claims and expenses, including attorneys’ fees, arising out of the use of
and ag this material.
y applicable law, b
y using these instructions to r
Ei, its ag
encies, officers, emplo
ecalibr
yees, r
SENSOR CALIBRATION & REPLACEMENT QUICK START GUIDE .
Page 2
SENSOR REPLACEMENT
1. Tools required to replace the sensor include a standard Phillips screw driver and a source of calibration gas.
5. Remove the two screws securing the case halves together.
2. Remove the protective boot from the instrument.
6. Separate the case by gently pressing up on the front housing with pulling the top apart.
3. Remove the water trap by sliding down and then away from the housing.
7. Lift the main circuit board up
and away from the instrument.
4. Remove the batteries.
a. Be careful of the wiring
harness that powers the pump and connects to the battery compartment.
8. Place the instrument on a flat surface with the circuit board positioned as shown.
9. Remove the four screws holding the sensor cover in place.
10. Remove the sensor cover.
11. Lift the sensor to be replaced out of the manifold.
ensor located at the top is
S
a.
O2, and the bottom position is CO.
Page 3
Note the position of the
b.
alignment pin on the O2 and pin orientation on the CO sensors. These will only fit into the sensor cover in one position.
12. Install the sensor to be replaced.
13. Reinstall sensor cover. a. Verify that the sensor cover
is seated on the manifold
e tightening screws.
befor
14. Set circuit board back on manifold.
a. Align sensor pins with sockets
on circuit board.
14. Join case halves together starting at the bottom edge as shown
15. Reinstall case screws
Page 4
GAS CALIBRATION
It is recommended to allow the sensors to stabilize a minimum of four hours before calibration.
1. Turn instrument on in fresh air and allow to zero.
5. Select “Cal_CO”.
2. Select “Menu”.
6. Enter gas concentration.
3. Scroll to “Service”.
7. Connect Gas.
4. Enter user code “1
234
”.
8. Adjust regulator to match flow of instrument.
a. Floating ball should be in
the middle of the second
om the bottom.
space fr
b. Note the pitch of the pump.
It should match the pitch when drawing in ambient air.
9. Allow to sample gas three minutes.
a. Reading does not indicate
ppm value.
b. Slight fluctuation of display
is normal.
CANADA: 1-877-475-0648 • Fax: 604.278.8299
GLE Combustion
EA
YRIGHT © 20
COP
07 UEi.
10. Press “Enter” at the end of 3 minutes.
eading is not ppm,
Note r and may still fluxuate slig
11. Scroll to “ press “
htly.
Exit” and then
Enter”.
USA: 1.800.547.5740 • Fax: 503.643.6322
W W W.UEiTEST.COM
Analyzers™ is a trademark of UEi. CALKIT-MAN 11/07
Loading...