Topcon MC-4S User Manual

USER MANUAL
MIRROR CHART
MC-4S

INTRODUCTION

Thank you for purchasing the TOPCON Mirror Chart MC-4S.

INTENDED USE / INDICATIONS FOR USE

This instrument is the visual acuity chart which is used in conjunction with the CV-5000. (Unless this instrument is connected to the CV-5000, it cannot operate as a visual acuity chart.)

FEATURES

This instrument has the following features:
• The active chart in which the targets are changed and moved is installed. You can observe the patient's eye condition with the target's move.
• The animation playback function is installed. You can check a variety of visual conditions by watching the displayed image.
• You can set the installing distance to 60cm, 80cm, 95cm, 1m and 1.5m*. * Only when selecting the H1 type.

PURPOSE OF THIS MANUAL

To ensure the best use of the instrument (safely and efficiently), please carefully read "DIS­PLAYS AND SYMBOLS FOR SAFE USE" and "GENERAL SAFETY INFORMATION" and familiarize yourself with the correct safety procedures prior to operating the MC-4S. Keep the user manual on hand for future reference.
This instrument should be used in conjunction with a CV system.
There are three methods to operate the MC-4S through the CV-5000.
• Operate the MC-4S with the 1Dial Controller KB-50S.
• Connect the mouse to the monitor and operate the MC-4S with the mouse.
• Connect a personal computer to the CV-5000 and operate the MC-4S with it.
This manual is made according to operation with the KB-50S. In the case of "Mouse opera­tion" and "Personal computer operation", you may use this manual by changing the words to the specified ones as shown in the following table.
Controller unit/CV software screen of 1Dial Controller
Monitor unit/CV software screen of 1Dial Controller
KB-50S operation
Notation Switch Button Button
Action Press Click Click
Display of setting menu
Display of test list
Action Touch Click Click
[Shift]+ [Menu]
[Shift]+[Prog] [Examination
Mouse operation
[Settings] button
list]
Personal computer operation
[Settings] button
[Examination list]
1
2
1. No part of this manual may be copied or reprinted, in whole or in part, without prior written permission.
2. The contents of this manual are subject to change without prior notice and without legal obligation.
3. The contents of this manual are correct to the best of our knowledge. Please inform us of any ambiguous or erroneous descriptions, missing information, etc.
4. Original Instructions This instruction manual was originally written in English.
©2011 TOPCON CORPORATION
ALL RIGHTS RESERVED
CONTENTS
INTRODUCTION .............................................................................................................................. 1
GENERAL SAFETY INFORMATION ............................................................................................... 5
GENERAL MAINTENANCE INFORMATION ................................................................................... 7
DISCLAIMERS ................................................................................................................................. 7
DISPLAYS AND SYMBOLS FOR SAFE USE ................................................................................. 8
POSITIONS OF WARNING AND CAUTION INDICATIONS ........................................................... 9
BEFORE USE ...............................................................................................................................10
STANDARD COMPONENTS ......................................................................................................... 10
NOMENCLATURE .................................................................................................................... 11
NOMENCLATURE OF ALIGNMENT EMITTER ............................................................................ 12
PREPARATION BEFORE USE ....................................................................................... 13
INSTALLING THE INSTRUMENT .................................................................................................. 13
CONNECTING TO COMPU-VISION CV-5000 .............................................................................. 14
CONNECTING THE POWER CORD ............................................................................................. 15
PILOT LAMP STATUS ................................................................................................................... 16
REGISTERING THE VISUAL ACUITY CHART ............................................................................. 16
SETTING THE INSTALLING DISTANCE ...................................................................................... 18
SETTING THE ALIGNMENT EMITTER ......................................................................................... 19
HOW TO ADJUST ALIGNMENT .................................................................................................... 20
INDICATION OF CHART ...................................................................................................... 21
TEST CHART LIST ........................................................................................................................ 21
OPERATION AFTER USE ............................................................................................................. 26
CHART FUNCTIONS UNIQUE TO THE MC-4S .................................................. 27
BLACK-AND-WHITE REVERSING FUNCTION ............................................................................ 27
CHARACTER CONTRAST CHANGING FUNCTION .................................................................... 29
ACTIVE CHART ............................................................................................................................. 34
DISPLAY OF STILL IMAGE/MOVIE .............................................................................................. 39
REGISTRATION (USER CUSTOMIZE) AND DISPLAY/PLAYBACK OF
STILL IMAGES/MOVIES ................................................................................................................ 47
SETTING THE VISUAL ACUITY CHART ...................................................................................... 52
MAINTENANCE .......................................................................................................................... 57
DAILY CHECKUPS ........................................................................................................................ 57
USER MAINTENANCE ITEMS ...................................................................................................... 58
REPLACING THE BATTERY ......................................................................................................... 58
BEFORE REQUESTING SERVICE .............................................................................. 59
TROUBLE SHOOTING GUIDE ...................................................................................................... 59
3
SPECIFICATIONS & PERFORMANCE ..................................................................... 62
SPECIFICATIONS BY ITEMS ........................................................................................................62
OTHER SPECIFICATIONS ............................................................................................................ 62
GENERAL INFORMATION ON USAGE AND MAINTENANCE .............. 63
PURPOSE OF USE/EFFICACY OR EFFECT ............................................................................... 63
ENVIRONMENTAL CONDITIONS FOR USE ................................................................................ 63
STORING METHOD, USAGE PERIOD AND OTHERS ................................................................ 63
ELECTRIC RATING ....................................................................................................................... 64
DIMENSIONS AND WEIGHT ......................................................................................................... 64
SYSTEM CLASSIFICATION ..........................................................................................................64
OPERATION PRINCIPLE .............................................................................................................. 65
CHECKPOINTS FOR MAINTENANCE .......................................................................................... 65
DISPOSAL ..................................................................................................................................... 65
PATIENT’S ENVIRONMENT ......................................................................................................... 66
REQUIREMENTS FOR THE EXTERNAL DEVICE ....................................................................... 66
ELECTROMAGNETIC COMPATIBILITY ....................................................................................... 67
REFERENCE MATERIAL .................................................................................................... 71
SHAPE OF PLUG .......................................................................................................................... 71
4

GENERAL SAFETY INFORMATION

WARNINGS
Preventing Electric Shock and Fire.
To avoid injury, do not touch the terminal while the AC adapter is in use.
To avoid fire and electric shock, install the instrument in a place free of water and other liquids.
To avoid fire and electric shock, do not insert metal objects into any clearance, etc.
To avoid fire in the event of an instrument malfunction, immediately turn off the power switch and remove the power plug from the outlet if you see smoke coming from the instrument or if you detect other problems. Don't install the instrument where it is difficult to disconnect the power plug from the outlet. Ask your dealer for repairs.
5
CAUTIONS
Preventing Electric Shock and Fire.
Set the test distance of the instrument correctly. [The reliability of the measurement result may be lowered.]
Be careful not to damage the chart window or stain it with fingerprints, dust or other contaminants. [The reliability of the measurement result may be lowered.]
Electromagnetic Compatibility (EMC)
This instrument has been tested (with 100V/230V) and found to comply with IEC60601-1-2 Ed.3.0:
2007. This instrument radiates radio frequency energy within standard and may affect other devices in the vicinity. If you have discovered that turning on/off the instrument affects other devices, we recommend you change its position, keep a proper distance from other devices, or plug it into a different outlet. Please consult the dealer from whom you purchased the instrument if you have any additional questions.
Connect devices complying with UL60950/UL60950-1, UL60601-1 or IEC60950/ IEC60950-1 to the MC-4S. Ask the authorized Topcon representative service engineer to connect the device.
6

GENERAL MAINTENANCE INFORMATION

USER MAINTENANCE

To ensure the safety and performance of this instrument, all maintenance work, unless specified in this manual, shall be conducted by trained service engineers. The following maintenance tasks may be done by the user. For details, see "MAINTENANCE" in this manual.

CLEANING THE CHART WINDOW

The chart window of the instrument can be cleaned by the user. For details, refer to “Cleaning the chart window” on page 57 in this manual.
CHECKING THE SCREW FOR THE SUPPORT METAL FIXTURE FOR LOOSE­NESS
The user can check if the screws for the support metal fixture are loose. For details, refer to “USER MAINTENANCE ITEMS” on page 58 of this manual.

DISCLAIMERS

• TOPCON shall not take any responsibility for damage due to fire, earthquakes, actions by third per­sons and other accidents, or damage due to negligence and misuse by the user and any use under unusual conditions.
• TOPCON shall not take any responsibility for damage derived from inability to properly use this instrument, such as loss of business profit and suspension of business.
• TOPCON shall not take any responsibility for damage caused from using this instrument in a man­ner other than that described in this user manual.
• Diagnoses made shall be the responsibility of pertaining doctors and TOPCON shall not take any responsibility for the results of such diagnoses.
7

DISPLAYS AND SYMBOLS FOR SAFE USE

Symbol

IEC/ISO
Publication
Description
IEC 60417-5032 Alternating Current
IEC 60417-5031 Direct current
IEC 60417-5957 For indoor use only
IEC 417-5265
“Off” (only for a part of EQUIPMENT)
IEC 417-5264
“On” (only for a part of EQUIPMENT)
ISO 7010-W001 General warning sign
ISO 7010-M002
Refer to the instruction man­ual.
ISO 7000-2497 Date of manufacture
ISO 7000-2498 Serial number
Not applicable
WEEE Directive 2002/96/ EEC Separate collection is neces­sary
*
*
Symbol
IEC/ISO
Publication
Description
Not applicable
Polarity Centre positive symbol
Not applicable The polarity of the batteries
ISO 7000-3082 Manufacturer
ISO 15223-1
Authorised Representative in the European Community
VGA *
Used to enter the VGA sig­nal. (This is not used when using the product.)
DVI *
Used to enter the DVI signal. (The image cable, which is an accessory, is connected to this.)
POWER *
Used to connect the AC adapter. (The AC adapter, which is an accessory, is connected to this.)
To encourage safe and proper use and to prevent danger to the operator and others or potential damage to property, important cautionary messages are placed on the instrument body and inserted in the instruc­tion manual. We suggest that everyone using the instrument understands the meaning of the following displays, icons and text before reading the "GENERAL SAFETY INFORMATION" and observes all listed instructions.

DISPLAYS

Display Meaning
Indicates a potentially hazardous situation which, if not avoided,
WARNINGS
CAUTIONS
NOTES
could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
Useful functions to know. Paying attention to these will prevent the problems noted.
SYMBOL
*: for USA and Canada
8

POSITIONS OF WARNING AND CAUTION INDICATIONS

1
3
2
4
5
To ensure safety, the machine provides warning displays. Use the instrument correctly by observing the display instructions. If any of the following display labels are missing, contact your TOPCON dealer or your local Topcon office listed on the back cover of this manual.
No. Label Meaning
WARNING
1
2
3
4
5
* Do not connect any equipment to these connections, other than what is listed in the standard compo-
nents table on page 10.
To avoid injury caused by electric shock, do not open the cover. Ask your dealer for service.
WARNING
Do not the touch the terminal while using the AC adapter. Doing this may cause injury.
CAUTION
Only install the maintenance cap when cleaning the instrument. If you do not follow this, the chart may not be displayed correctly and cause an incorrect test result.
CAUTION
Install the support metal fixture. If you fail to install this, the instru­ment may fall and injure someone.
Use an AC adapter of the following rating for this instrument. 100V: 12V 2.5A ( ) 220-240V: 12V 5A ( ) If you use an adapter which is outside of the above rating, the instru­ment may be damaged.
CAUTION
9

BEFORE USE

STANDARD COMPONENTS

The table below shows the standard components. Make sure that you have all of the components. The numerical value in ( ) shows the quantity. If one of the standard components is not included in the package, contact your dealer or your local TOPCON office (listed on the back cover).
User manual /
Power cord (1) AC adapter (1) *
Instruction manual (1)
USER MANUAL
MIRROR CHART
MC-4S
Dust cover (1) Alignment emitter (1) Dry battery (2)
Image cable (1)
* BPM030S12F04 (for USA and Canada)
JMW160KA1200F09 (for outside the USA and Canada)
10
BEFORE USE

NOMENCLATURE

2
3
1
4
123
4
CAUTIONS
Front surface
To avoid electric shock, do not touch the external connection terminal and the patient at the same time.
Chart window: The patient watches the chart through this window. Power switch/Pilot lamp: The pilot lamp indicates the power status of the instrument. The pilot
lamp is lit in orange when the instrument is plugged into the outlet. When the power switch is turned ON, this lamp is lit in green and the
instrument can receive the image signal. Maintenance cap: Use this in maintenance. Sensor detector unit: This unit receives the signal from the alignment emitter to adjust the
eye position.
11
NOMENCLATURE
Lower rear surface
8
5
9
6
10
7
56789
10
Magnet
Battery cover
Light emitting unit
Switch
Cover
NOTES
Analog VGA connector: Usually this is not used. DVI connector: Used to connect with the external devices (CV-5000 only). Ask the
AC adapter jack: Used to connect the AC adapter. Support metal fixture: This reinforces the instrument's stability. Screw for support metal fixture:These affix the support metal fixture onto the instrument. Adjuster: Used to adjust the inclination when installing the instrument.
Cover is attached to the analog VGA connector. Don't use this connector.
service personnel to connect the instrument with the external devices.

NOMENCLATURE OF ALIGNMENT EMITTER

12
NOMENCLATURE

PREPARATION BEFORE USE

INSTALLING THE INSTRUMENT

• Be careful not to damage the chart window or stain it with fingerprints, dust or other contaminants. The reliability of the measurement result may be lowered.
CAUTIONS
On the chart window, there is sometimes noticeably unevenness on the back-
NOTES
ground due to interfering light. To prevent sunlight or interfering light from com­ing in contact with the chart window directly, change the installation location or the direction of the instrument.
• Tighten the screws for the support metal fixture securely so they can­not be loosened. If the screw becomes loosened, the instrument may fall and cause damage and/or injury.
• To remove the support metal fixture, ask your service personnel. If the instrument is not installed along a wall, it may fall and cause injury.
13
PREPARATION BEFORE USE

CONNECTING TO COMPU-VISION CV-5000

CV system
MC-4S
This instrument is connected to the CV system and is used as a visual acuity chart. As shown below, connect the image cable to the DVI connector of this instrument.
This instrument can only be connected to the following devices: CV POWER SUPPLY UNIT WITH BUILT-IN PERSONAL COMPUTER or the personal com-
NOTES
puter connected to the CV POWER SUPPLY UNIT WITH EXTERNAL PER­SONAL COMPUTER. This instrument cannot be connected to the CV power supply unit and the personal computer connected with the CV power supply unit.
CV-5000
Refer to the CV-5000 user manual for the connection method for the CV system.
14
PREPARATION BEFORE USE

CONNECTING THE POWER CORD

• To avoid fire and electric shock in case of leakage, be sure to use a
WARNINGS
grounded outlet. Do not connect to outlets that are not grounded.
• To avoid electric shock, use only the attached AC adapter.
CAUTIONS
To avoid electric shock, do not handle the plugs with wet fingers.
1 Connect the AC adapter to the AC adapter jack on the lower rear surface.
2 Connect the power cord to the AC adapter.
3 Insert the power cord plug into the 3-plug AC outlet with proper grounding.
PREPARATION BEFORE USE
15

PILOT LAMP STATUS

The pilot lamp color in each instrument status is shown below. (1) OFF: AC adapter; Not connected.
Power switch; ON/OFF
(2) Lit in orange: AC adapter; Connected.
Power switch; OFF
(3) Lit in green: AC adapter; Connected.
Power switch; ON

REGISTERING THE VISUAL ACUITY CHART

Operate this instrument through the CV system. This chapter will explain the setting procedure of the visual acuity chart that is necessary to use this instrument.
1 Turn on this instrument and the power supply unit.
* When operating through a personal computer:
When using this instrument by connecting to a personal computer, it is necessary to display the desktop on the screen before starting the CV software. Set “Resolution: 640 x 480, 32­bit color (Screen refresh rate: 70Hz or more)” for the MC-4S monitor and connect it as the sub monitor in advance. Turn on this instrument, the power supply unit and the personal computer to start the CV software.
2 While pressing the [Shift] switch on KB-50S, press the [Menu] switch.
The "Settings" menu is displayed.
16
PREPARATION BEFORE USE
3 Touch the [Chart settings] button.
The menu to select the chart that should be connected is displayed.
4 Set "MC-4S" for "Device" of "For far vision" and then select a type from "Type". Then, select
"MC-4S" for "Chart monitor selection".
5 Touch the [Chart software config.] button and then the [Chart settings] button. The "CHART
SETTINGS" screen appears. Set [Examination distance] according to the real test distance. Touch the [OK] button and then the [Exit] button. The "Chart settings" screen appears again.
6 Touch the [OK] button and then the [Exit] button to save the setting and finish the procedure.
17
PREPARATION BEFORE USE

SETTING THE INSTALLING DISTANCE

• To avoid electric shock, when setting the installing distance do not make any contact with the patient and the sensor detector unit at the
CAUTIONS
same time.
• Before setting the installing distance, make sure that the power switch of the instrument is OFF. If it is ON, you may receive an electric shock.
NOTES
You can change the installing distance to 60cm, 80cm, 95cm, 1m and 1.5m*. * Only when selecting the H1 type. According to the installing distance, it is necessary to change the DIP switch of the instrument.
To change the installing distance setting, ask your service personnel.
1 To perform setting for the instrument, open the cover at the rear of the sensor detector unit.
NOTES
Installing distance
60cm OFF OFF OFF
80cm OFF ON OFF
95cm, 1m, 1.5m OFF OFF ON
To change the setting of the instrument, ask your service personnel.
Instrument
SW01-1 SW01-2 SW01-3
*The installing distance is set to 80cm as the default when shipped.
18
PREPARATION BEFORE USE

SETTING THE ALIGNMENT EMITTER

Battery cover
Battery box
Slide switch S1 Left: H Right: L
Slide switch S3
Slide switch S2 Left: H Right: L
When installing two or more instruments in neighboring areas, it is necessary to set the channels of the instruments and controllers in order to prevent interference. (It is possible to install up to four instruments.)
To avoid electric shock, when changing the setting of the alignment
CAUTIONS
emitter do not make any contact with the patient and alignment emitter at the same time.
NOTES
To change the setting of the alignment emitter, ask your service personnel.
1 Remove the battery cover from the alignment emitter.
2 Change the slide switches inside the battery box according to the channel table.
3 Attach the battery cover. 4 To perform the setting for the instrument, open the cover at the rear of the sensor detector unit.
NOTES
Channel table Setting of alignment emitter and instrument
NOTES
To change the setting of the instrument, ask your service personnel.
Channel
1HHOFFOFF 2LHONOFF 3HLOFFON 4 L L ON ON
Alignment emitter Instrument
S1 S2 SW02-1 SW02-2
"Channel 1" is set as the default when shipped.
19
PREPARATION BEFORE USE

HOW TO ADJUST ALIGNMENT

Place the alignment emitter beside the eye and adjust the height.
Place the alignment emitter beside the examination window and adjust the height.
1 Set 80cm (refer to Note 1) as the distance from the
instrument’s front surface to the patient’s eye. Then, ask the patient to sit down on the chair.
Note 1: The installing distance can be selected from
60cm, 80cm, 95cm, 1m and 1.5m*. * Only when selecting the H1 type.
2 Place the alignment emitter beside the right eye of the
patient and press its switch.
80cm
3 Adjust the chart indication position to the optimum within the range of 150mm at the right and left
positions and 200mm in the upper and lower positions against the front of the chart window.
• When the patient watches the chart through the TOPCON CV-5000 or other instruments, press the alignment emitter switch beside the examination win­dow.
NOTES
• If the alignment emitter is placed too far away in the back-and-forth direction, sometimes the height is not adjusted correctly though it is beside the eye or examination window.
• When using the alignment emitter, let it face the sensor detector unit of the MC-4S.
• If the chart window is in the sunlight or if there is a window behind the patient, the instrument may operate incorrectly or its buzzer may sound continuously to stop its operation. Change the installation location or close the curtain or blind.
20
PREPARATION BEFORE USE

INDICATION OF CHART

The MC-4S has the functions particular to the image indication type. This user manual will explain the chart functions that are unique to the MC-4S. For a general indication of charts, refer to the Compu-Vision CV-5000 user manual.
• The Compu-Vision CV-5000 user manual describes the tests when the MC-4S is used.
• When the instrument is set to indicate the visual acuity, it is indicated at the
NOTES

TEST CHART LIST

The MC-4S can indicate three types of charts (D type, F type and H1 type).
right side of the chart. When a large chart is selected, the visual acuity is indicated at the upper or lower right. Sometimes, even if the system is set to indicate the visual acuity in large size, it is forcedly indicated in small size.
• D type
Hiragana chart: Visual acuity 0.05
Hiragana chart: Visual acuity 0.8, 0.9 and 1.0
Hiragana chart: Visual acuity 0.1 and
0.15
Hiragana chart: Visual acuity 0.9, 1.0 and 1.2
Hiragana chart: Visual acuity 0.2, 0.3 and 0.4
Hiragana chart: Visual acuity 1.2, 1.5 and 2.0
Hiragana chart: Visual acuity 0.5, 0.6 and 0.7
Landolt ring chart: Visual acuity 0.03
Landolt ring chart: Visual acuity 0.05
Landolt ring chart: Visual acuity 0.1
Landolt ring chart: Visual acuity 0.15
Landolt ring chart: Visual acuity 0.2, 0.3 and 0.4
21
INDICATION OF CHART
Landolt ring chart: Visual acuity 0.5, 0.6 and 0.7
Landolt screening chart ETDRS chart (1) ETDRS chart (2) Illustration chart
Landolt ring chart: Visual acuity 0.8, 0.9 and 1.0
Landolt ring chart: Visual acuity 1.2, 1.5 and 2.0
Hiragana screening chart
RG chart RG binocular balance
chart
Reticle heterophoria chart (black back­ground)
Fixation target chart Cyclophoria chart
Cross chart with fixation target (white back­ground)
(white background)
Binocular balance (polarization) chart
Cross chart with fixation target (black back­ground)
Cyclophoria chart (black background)
Reticle heterophoria chart (white back­ground)
Cross ring chart
One vertical line char­acter chart
One horizontal line character chart
22
INDICATION OF CHART
Aniseikonia chart Precise stereoscopic
vision chart
Depth perception test chart
Worth 4 Dots chart Cross cylinder chart (1) Cross cylinder chart (2) Cross cylinder chart (3)
Astigmatism chart (1) Astigmatism chart (2) Astigmatism chart (3)
• F type
Letter E chart: Visual acuity 0.03
Letter E chart: Visual acuity 0.5, 0.6 and 0.7
Landolt ring chart: Visual acuity 0.1 and 0.16
Letter E chart: Visual acuity 0.05
Letter E chart: Visual acuity 0.8, 0.9 and 1.0
Landolt ring chart: Visual acuity 0.2, 0.3 and 0.4
Letter E chart: Visual acuity 0.1 and 0.16
Letter E chart: Visual acuity 1.2, 1.5 and 2.0
Landolt ring chart: Visual acuity 0.5, 0.6 and 0.7
Letter E chart: Visual acuity 0.2, 0.3 and 0.4
Landolt ring chart: Visual acuity 0.05
Landolt ring chart: Visual acuity 0.8, 0.9 and 1.0
Landolt ring chart: Visual acuity 1.2, 1.5 and 2.0
Numeral chart: Visual acuity 0.1 and 0.2
Numeral chart: Visual acuity 0.3, 0.4 and 0.5
Numeral chart: Visual acuity 0.6, 0.7 and 0.8
23
INDICATION OF CHART
Numeral chart: Visual acuity 1.0, 1.2 and 1.5
ETDRS chart (1) ETDRS chart (2) RG chart Binocular balance
Chart for children: Visual acuity 0.2 and
0.3
Chart for children: Visual acuity 0.4, 0.5 and 0.6
Chart for children: Visual acuity 0.7, 0.8 and 1.0
(polarization) chart
Reticle heterophoria chart (white back­ground)
Cross chart with fixation target (white back­ground)
Stereoscopic vision chart
Reticle heterophoria chart (black back­ground)
Cross chart with fixation target (black back­ground)
Worth 4 Dots chart Cross cylinder chart (1) Cross cylinder chart (2)
Cross ring chart Fixation target chart
Aniseikonia chart (verti­cal)
Aniseikonia chart (hori­zontal)
Cross cylinder chart (3) Astigmatism chart (1) Astigmatism chart (2) Astigmatism chart (3)
24
INDICATION OF CHART
• H1 type
Alphabet chart: Visual acuity 400
Alphabet chart: Visual acuity 40, 30 and 25
Alphabet and numeral chart: Visual acuity 25, 20 and 15
Alphabet chart: Visual acuity 200 and 150
Alphabet chart: Visual acuity 30,25 and 20
Letter E chart: Visual acuity 600
Alphabet chart: Visual acuity 100 and 80
Alphabet chart: Visual acuity 20, 15 and 10
Letter E chart: Visual acuity 400
Alphabet chart: Visual acuity 70, 60 and 50
Alphabet chart: Visual acuity 20
Letter E chart: Visual acuity 200
Letter E chart: Visual acuity 150
Letter E chart: Visual acuity 30,25 and 20
Numeral chart: Visual acuity 70, 60 and 50
Letter E chart: Visual acuity 100 and 80
Letter E chart: Visual acuity 20, 15 and 10
Numeral chart: Visual acuity 40, 30 and 25
Letter E chart: Visual acuity 70, 60 and 50
Numeral chart: Visual acuity 200 and 150
Numeral chart: Visual acuity 30,25 and 20
Letter E chart: Visual acuity 40, 30 and 25
Numeral chart: Visual acuity 100 and 80
Numeral chart: Visual acuity 20, 15 and 10
25
INDICATION OF CHART
Chart for children: Visual acuity 100
Chart for children: Visual acuity 80 and 60
Chart for children: acuity 40, 30 and 20
Visual
ETDRS chart (1)
ETDRS chart (2) RG chart Binocular balance
(polarization) chart
Fixation target chart Aniseikonia chart Stereoscopic vision
chart
Cross cylinder chart (1) Cross cylinder chart (2) Cross cylinder chart (3) Astigmatism chart (1)
Cross chart with fixation target (white back­ground)
Worth 4 Dots chart
Astigmatism chart (2) Astigmatism chart (3) Cross chart with fixation
target (black back­ground)

OPERATION AFTER USE

1 Turn off the power switch. 2 Unplug the power cable from a 3-pin AC inlet with grounding.
26
INDICATION OF CHART

CHART FUNCTIONS UNIQUE TO THE MC-4S

BLACK-AND-WHITE REVERSING FUNCTION

The MC-4S has the ability to change the chart indicating black characters with white background to the one indicating white characters with black background in the visual acuity measurement.
Relevant chart
This function is applied to the visual acuity charts except the screening chart.
An example of the test chart icon displayed on the chart page of the KB-50S:
Usable chart
Reversing the black-and-white display condition
1 Select the visual acuity chart which can be reversed from the chart page. 2 Press the [Shift] key on the KB-50S and touch the [Reverse] button among the function buttons
at the bottom of the screen.
3 On the MC-4S, the chart is displayed in the reversed black-and-white condition.
At the same time, the chart is also displayed in the reversed condition on the chart check display area of the KB-50S.
27
CHART FUNCTIONS UNIQUE TO THE MC-4S
• To reverse the black-and-white condition through the mouse/personal com­puter, click the button displayed on the test screen of Step 2.
• The black-and-white reverse mode is kept while the following operations are
NOTES
Canceling the reversed black-and-white display condition
being executed.
• Changing to any other visual acuity chart applicable to this reverse mode
• Setting and moving the vertical mask, horizontal mask and character mask
• Executing the random mode
1 Press the [Shift] key on the KB-50S and touch the [Reverse] button among the function buttons
at the bottom of the screen.
2 The reversed black-and-white condition of the chart indicated on the MC-4S is canceled.
At the same time, the reversed condition of the chart is also canceled on the chart check display area of the KB-50S.
• To cancel the reversed black-and-white condition through the mouse/personal computer, click the button displayed on the test screen of Step 2.
• The black-and-white reverse mode is also canceled by the following opera-
NOTES
tions.
• Changing to the chart which is not applicable to this reverse mode
• Setting the RG mask
• Setting the contrast mode
28
CHART FUNCTIONS UNIQUE TO THE MC-4S

CHARACTER CONTRAST CHANGING FUNCTION

In the MC-4S, the character contrast of the chart can be changed and displayed for visual acuity measure­ment. There are two display methods. "Single rate character contrast mode" changes the character contrast of the whole chart uniformly. "Plural rate character contrast mode" changes the contrast of each column on the chart.
Single rate character contrast mode
In "Single rate character contrast" mode, the character contrast on the chart can be changed in seven steps, 2.5%, 5%, 10%, 12.5%, 25%, 50%, and 100%.
Relevant chart
This function is applied to all the visual acuity charts except the screening chart.
An example of the test chart icon displayed on the chart page of the KB-50S:
Usable chart
Setting "Single rate character contrast"
1 Select the visual acuity chart applicable to the "Single rate character contrast" function from the
chart page.
2 Touch the [Contrast] button among the function buttons at the bottom of the screen on the KB-
50S to display the contrast setting menu.
29
CHART FUNCTIONS UNIQUE TO THE MC-4S
3 Touch the button of the desired contrast from the contrast setting menu. The chart indicated on
the MC-4S is displayed with the selected contrast as shown below.
• When operating the CV system through a mouse/personal computer, you can change the contrast with the same procedure.
NOTES
• When changing the chart to another that is applicable to this mode, the "Sin­gle rate character contrast" mode is kept.
• While the "Single rate character contrast" mode is being executed, you cannot set the vertical, horizontal, character and RG masks.
Input of contrast visual acuity in the single rate character contrast mode
1 Input the visual acuity value. For inputting the visual acuity, refer to the COMPU-VISION CV-
5000 USER MANUAL.
2 The visual acuity value with the contrast value, which is being displayed, is recorded.
30
CHART FUNCTIONS UNIQUE TO THE MC-4S
Canceling "Single rate character contrast"
1 Touch the [Contrast] button among the function buttons at the bottom of the screen on the KB-
50S to display the contrast setting menu.
2 Touch the [Exit] button on the contrast selection menu, and the "Single rate character contrast"
mode is canceled.
• When operating the CV system through a mouse/personal computer, you can
NOTES
cancel the contrast with the same procedure.
• When changing to a chart that is not applicable to this mode, the "Single rate character contrast" mode is also canceled.
Plural rate character contrast mode
In the "Plural rate character contrast" mode, the contrast conditions of the five stages, 25%, 12.5%, 10%, 5% and 2.5%, can be displayed on one chart at a time.
Relevant chart
This function is applied to a visual acuity chart having 3 lines and 5 columns.
An example of the test chart icon displayed on the chart page of the KB-50S:
Usable chart
Unusable chart
31
CHART FUNCTIONS UNIQUE TO THE MC-4S
Setting "Plural rate character contrast"
1 Select the visual acuity chart applicable to the "Plural rate character contrast" function. 2 Touch the [Contrast] button among the function buttons at the bottom of the screen on the KB-
50S to display the contrast setting menu.
3 Touch the [Plural rate] button in the contrast setting menu. As shown below, on the MC-4S, the
chart is displayed under the condition that the columns are different in contrast by stages, 25%,
12.5%, 10%, 5% and 2.5% from the leftmost.
• When operating the CV system through a mouse/personal computer, you can change the contrast with the same procedure.
NOTES
• When changing the chart to another that is applicable to this mode, the "Plural rate character contrast" mode is kept.
• While the "Plural rate character contrast" mode is being executed, you cannot set the vertical, horizontal, character and RG masks.
32
CHART FUNCTIONS UNIQUE TO THE MC-4S
Input of contrast visual acuity in the plural rate character contrast mode
1 Input the visual acuity value. For inputting the visual acuity, refer to the COMPU-VISION CV-
5000 USER MANUAL.
2 The dialog box to select a contrast value is displayed. Touch the desired contrast value.
3 The visual acuity with the touched contrast value is recorded.
Canceling "Plural rate character contrast"
1 Touch the [Contrast] button among the function buttons at the bottom of the screen on the KB-
50S to display the contrast setting menu.
2 Touch the [Exit] button on the contrast selection menu, and the "plural rate character contrast"
mode is canceled.
• When operating the CV system through a mouse/personal computer, you can
NOTES
cancel the contrast with the same procedure.
• When changing to a chart that is not applicable to this mode, the "Plural rate character contrast" mode is also canceled.
33
CHART FUNCTIONS UNIQUE TO THE MC-4S

ACTIVE CHART

In the MC-4S, it is possible to use "Active chart" for testing/checking by making use of the characteristics of the built-in liquid crystal visual acuity chart.
Prism simulation function
With this function, you can move the target indicated to the right/left eyes in the phoria test. This func­tion is useful to explain the eye condition to the patient or check the patient's eye after the test.
Relevant chart
The following charts are applicable to this function:
Executing "Prism simulation"
1 Select the chart applicable to "Prism simulation" from the chart page. 2 Perform the phoria measurement according to normal procedure. 3 Touch the [Simulate] button among the function buttons at the bottom of the screen on the KB-50S.
4 The prism lens, which is set on the measuring head of the CV-5000, is temporarily canceled. As
shown below, the chart is converted so that the targets may be indicated at the right and left eye target positions and then the chart is displayed on the MC-4S. The prism measurement value is rounded to the 0.2 prism step value.
Example: Cross chart
Before simulation After simulation
34
CHART FUNCTIONS UNIQUE TO THE MC-4S
5 You can change the prism amount by using the dial of the KB-50S.
In the "Simulation" mode, the prism movable range is 6.0
4.4
BD to 4.4BU for vertical prism.
Turn the dial, and the target moves according to "Dial Navigation".
Horizontal phoria measurement
Vertical phoria measurement
Example of "Dial Navigation" for cross chart
Turn the dial clockwise, and the vertical line moves left and the horizontal line moves right. Turn the dial counterclockwise, and the vertical line moves right and the horizontal line moves left.
Turn the dial clockwise, and the vertical line moves down and the horizontal line moves up. Turn the dial counterclockwise, and the vertical line moves up and the horizontal line moves down.
6 To finish "Simulation", touch the [Simulate] button again.
The chart of the MC-4S is returned to the display status before executing "Simulation". The prism lens on the measuring head of the CV-5000 is set according to the current prism value.
BI to 6.0BO for horizontal prism and,
• When operating through a mouse/personal computer, do not turn the dial but click the button on the control window. Operate according to "Mouse Naviga­tion".
• The "Simulation" mode cannot be executed in the following case: The preset prism value before accessing the "Simulation" mode is beyond the prism mov­able range of the "Simulation" mode.
NOTES
• When executing the "Simulation" mode with the cross ring chart, the prism movable range is 6.0
BU for vertical prism. The range where the targets are within the chart
4.4 screen is 1.4 tical prism. If you set the prism value exceeding this range, the right eye tar­get moves to the outside of the chart because of the chart shape characteristics.
• When changing to other charts, the "Simulation" mode is also exited.
BI to 1.4BO for horizontal prism and 1.0BD to 1.0BU for ver-
BI to 6.0BO for horizontal prism and 4.4BD to
Aniseikonia simulation function
By using the aniseikona simulation function, the target size at the left side of the vertical aniseikonia chart can be changed in stages and, under this changed status, the chart can be displayed.
Relevant chart
The following chart is applicable to this function:
35
CHART FUNCTIONS UNIQUE TO THE MC-4S
Executing "Aniseikonia simulation"
1 Select the chart applicable to "Aniseikonia simulation" from the chart page. 2 Touch the [Simulate] button among the function buttons at the bottom of the screen on the KB-50S.
3 The "Simulation" mode starts.
Turn the dial of the KB-50S during the "Simulation" mode, and the square size at the left side is changed as shown below. Adjust the size so that the patient can see the right and left targets in the same size.
4 Operate the dial according to "Dial Navigation".
Turn the dial clockwise, and the left target size is decreased by 0.5% step. Turn the dial counterclockwise, and the left target size is increased by
0.5% step.
The changeable size range is -10% to 10% with the size rate of the left square against the right one. The size during changing is displayed on the function button at the bottom of the screen on the KB-50S.
36
CHART FUNCTIONS UNIQUE TO THE MC-4S
5 To record the result, touch the [Record] button among the function buttons at the bottom of the
screen on the KB-50S.
6 To finish "Simulation", touch the [Simulate] button again.
The chart of the MC-4S is returned to the display status before executing "Simulation".
• When operating through a mouse/personal computer, do not turn the dial but
NOTES
click the button on the control window. Operate according to "Mouse Naviga­tion".
• When changing to other charts, the "Simulation" mode is also exited.
Astigmatism chart rotation function
By using this function, on the astigmatism chart where only one radial line color is deeper than others, the deeper line can be changed with the dial. To measure the cylinder axis by using the astigmatism chart, let the patient align the deeper line on the chart with the line whose color is deeper in his/her vision.
Relevant chart
The following charts are applicable to this function:
37
CHART FUNCTIONS UNIQUE TO THE MC-4S
Executing "Astigmatism chart rotation function"
1 Select the chart applicable to the "Astigmatism chart rotation" function from the chart page. 2 Touch the [Rotate] button among the function buttons at the bottom of the screen on the KB-50S.
When the cylinder power and axis are already measured and the lens is already set on the mea­suring head of the CV-5000, the cylinder axis is rounded to the 5-degree step value and the cyl­inder power is temporarily set to 0.00D.
3 The chart on the screen of the MC-4S is changed to the following one.
Turn the dial of the KB-50S, and the deeper color line on the astigmatism chart is changed in the dial turning direction. Let the patient align the deeper color line on the chart with the line whose color is deeper in his/ her vision.
4 To exit the "Astigmatism chart rotation" mode, touch the [Rotate] button again.
The chart of the MC-4S is returned to the status before executing the "Astigmatism chart rotation" function. The cylinder power before executing the "Astigmatism chart rotation" mode is set with the current cylinder axis on the measuring head of the CV-5000.
• When operating through a mouse/personal computer, do not turn the dial but click the button on the control window. Operate according to "Mouse Naviga­tion".
NOTES
• While the "Astigmatism chart rotation" mode is being executed in the KB-50S, the display of "Dial Navigation" is opposite to the real axis value change.
• When changing to other charts, the "Astigmatism chart rotation" mode is also exited.
38
CHART FUNCTIONS UNIQUE TO THE MC-4S

DISPLAY OF STILL IMAGE/MOVIE

It is possible to display still images/movies on the MC-4S. In addition, the 3D still images and movies can be displayed by making use of the polarization chart characteristics. Still images and movies are preinstalled. It is also possible to display the still images and movies prepared by the customer.
Preinstalled contents
When shipped, 20 files of still images, 2 files of movies and 9 files of still images for slideshows for each of D type, F type and H1 type are provided in the MC-4S.
Still image file
• D type
Chart page icon
MC-4S Test screen Remarks
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
39
CHART FUNCTIONS UNIQUE TO THE MC-4S
Chart page icon
MC-4S Test screen Remarks
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
40
CHART FUNCTIONS UNIQUE TO THE MC-4S
Chart page icon
MC-4S Test screen Remarks
The still image is displayed on the MC­4S only.
The still image is displayed on the MC­4S only.
This still image is the 3D still image. Watch this image through the polarized lens set on the measuring head of the CV-5000. It will be seen as a stereo­scopic vision.
This still image is the 3D still image. Watch this image through the polarized lens set on the measuring head of the CV-5000. It will be seen as a stereo­scopic vision.
This still image is the 3D still image. Watch this image through the polarized lens set on the measuring head of the CV-5000. It will be seen as a stereo­scopic vision.
This still image is the 3D still image. Watch this image through the polarized lens set on the measuring head of the CV-5000. It will be seen as a stereo­scopic vision.
This still image is the 3D still image. Watch this image through the polarized lens set on the measuring head of the CV-5000. It will be seen as a stereo­scopic vision.
41
CHART FUNCTIONS UNIQUE TO THE MC-4S
Chart page icon
• F type, H1 type
MC-4S Test screen Remarks
This still image is the 3D still image. Watch this image through the polarized lens set on the measuring head of the CV-5000. It will be seen as a stereo­scopic vision.
Chart page icon
MC-4S Test screen Remarks
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
42
CHART FUNCTIONS UNIQUE TO THE MC-4S
Still image is displayed on MC-4S and the test screen.
Still image is displayed on MC-4S and the test screen.
Chart page icon
MC-4S Test screen Remarks
Still image is displayed on MC-4S and the test screen.
This still image is the 3D still image. Watch this image through the polarized lens set on the measuring head of the CV-5000. It will be seen as a stereo­scopic vision.
This still image is the 3D still image. Watch this image through the polarized lens set on the measuring head of the CV-5000. It will be seen as a stereo­scopic vision.
This still image is the 3D still image. Watch this image through the polarized lens set on the measuring head of the CV-5000. It will be seen as a stereo­scopic vision.
This still image is the 3D still image. Watch this image through the polarized lens set on the measuring head of the CV-5000. It will be seen as a stereo­scopic vision.
This still image is the 3D still image. Watch this image through the polarized lens set on the measuring head of the CV-5000. It will be seen as a stereo­scopic vision.
This still image is the 3D still image. Watch this image through the polarized lens set on the measuring head of the CV-5000. It will be seen as a stereo­scopic vision.
The still image is displayed on the MC­4S only.
43
CHART FUNCTIONS UNIQUE TO THE MC-4S
Chart page
Chart page icon (example)
icon
MC-4S Test screen Remarks
The still image is displayed on the MC­4S only.
The still image is displayed on the MC­4S only.
The still image is displayed on the MC­4S only.
The still image is displayed on the MC­4S only.
Playback of still images
1 Select an optional still image icon from the chart page.
The still image is played.
The still image is displayed on the MC­4S only.
The still image is displayed on the MC­4S only.
NOTES
Select any other chart. The playback of the still image is finished and the selected chart is indicated.
44
CHART FUNCTIONS UNIQUE TO THE MC-4S
Movie file
Chart page icon (example)
• D type
Chart page icon
• F type, H1 type
Chart page icon
MC-4S Remarks
This movie is the 3D movie. Watch this movie through the polarized lens set on the measur­ing head of the CV-5000. It will be seen as a stereoscopic vision.
This movie shows the landscape photographed from the car running on a superhighway.
MC-4S Remarks
Playback of movie
1 Select "MOVIE" from the chart page.
The movie is played.
This movie is the 3D movie. Watch this movie through the polarized lens set on the measur­ing head of the CV-5000. It will be seen as a stereoscopic vision.
This movie is the animation showing the scene "Traveling on the road at night".
NOTES
Select any other chart. The playback of the movie is finished and the selected chart is indicated.
45
CHART FUNCTIONS UNIQUE TO THE MC-4S
Still image file for slideshow
Chart page icon
• D type, F type, H1 type
Chart page icon
MC-4S Remarks
The images are dis­played on the MC­4S in the order shown. The still image dis­play time can be changed in the set­ting screen.
Playback of slideshow
1 Select "Slideshow" from the chart page.
The image slideshow is played.
NOTES
Select any other chart. "Slideshow" is canceled and the selected chart is indi­cated on the instrument.
46
CHART FUNCTIONS UNIQUE TO THE MC-4S

REGISTRATION (USER CUSTOMIZE) AND DISPLAY/PLAYBACK OF STILL IMAGES/MOVIES

User customizable contents
In the MC-4S, it is possible to display/play back the still images/movies prepared by the customer by saving them in the USB memory and connecting it with the CV-5000 system.
NOTES
It is necessary to connect the USB memory to the USB port of the CV-5000 sys­tem before starting this system.
Contents that can be added:
The contents that can be added by the customer are shown below.
2D still image
It is possible to display the photos taken by a digital camera and an illustration made by the cus­tomer. Access this still image from the KB-50S and display it optionally.
3D still image
Register one image for each of the right and left eyes. It is possible to compound these images so that they may be displayed as a 3D image on the MC-4S. To display the stereoscopic image, prepare the still images for the right and left eyes with a little parallax between them. Access this still image from the KB-50S and display it optionally.
Still image to be displayed on the test screen
It is possible to relate the photos taken by a digital camera and an illustration made by the cus­tomer with the 2D or 3D still images and display them on the test screen in a link operation. For instance, it is possible to display a far image on the MC-4S and, on the test screen, display a nearby image.
Movie
It is possible to display the image taken by the video camera.
Still image for slideshow
The photos taken by a digital camera and an illustration made by the customer can be displayed as a slideshow.
CHART FUNCTIONS UNIQUE TO THE MC-4S
47
File format
USB memory
This folder contains all of the user customizable contents. This folder contains the movie files. This folder contains the still image files. Three folders are included in it. This folder contains the still images which are displayed on the test screen. This folder contains the 2D/3D still images which are displayed on the MC-4S. This folder contains the still images for the slideshow that are displayed on the
MC-4S.
The following table shows the conditions of the files that can be displayed and played as user customi­zable contents.
Type Format Size File quantity Filename Remarks
2D still image
JPEG Optional 20 files when
combined with 3D still images
01.jpg - 20.jpg When using a number for the filename, do not use the same number as the 3D still image.
3D still image
Still image to be displayed on the test screen
Movie WMV file
Still image for slideshow
JPEG Optional 20 files when
combined with 2D still images
JPEG 800 × 600 20 files 01.jpg - 20.jpg This still image is related
Max. vertical encoded by CODEC Windows Media Video 7 - 8
JPEG Optional 20 files 01.jpg - 20.jpg
and horizontal
size: 1280×1024
Max. bit rate:
4Mbps or less
Max. frame rate:
30fps
2 files 01.wmv/02.wmv The playback of the movie
How to add the user customizable contents
For right eye: 01R.jpg - 20R.jpg For left eye: 01L.jpg - 20L.jpg
The image size for the right eye must be the same as the left eye. When using a number for the filename, do not use the same number as the 2D still image.
to the 2D/3D still image having the same number as the number in the file­name of this image.
is not smooth in some movie files.
To use the user customizable contents files, save these files in the USB memory and connect it with the CV-5000 system.
Folder configuration of the USB memory
48
CHART FUNCTIONS UNIQUE TO THE MC-4S
Arrange the folders under USB memory in the following configuration and add the file.
Registration to the chart page of the KB-50S
To display the user customizable contents on the MC-4S, it is necessary to add the icons concerned to the chart page of the KB-50S.
1 Press the [Shift] key and [Menu] key of the KB-50S at the same time to display the "Settings"
screen.
2 On the "Settings" screen, touch the [Chart settings] button. 3 The "Chart settings" screen is displayed.
Make sure that "MC4S" is set for "Device" and then touch the [Chart page registration] button.
4 The "Chart page registration" screen is displayed.
49
CHART FUNCTIONS UNIQUE TO THE MC-4S
Drag the user customizable contents icons from the chart icon list at the top of the screen and drop them into the chart page to register them.
These are the icons for 2D/3D still images. The number in the icon corresponds to the filename.
This is the icon for slideshow.
These are the icons for movies. The number in the icon corresponds to the filename.
5 Perform the registration to set the polarized lens onto the measuring head of the CV-5000 auto-
matically when a still image is displayed in order to register the 3D still image. Select the icon to be registered as the 3D still image from the chart page and touch the [Modify] button.
6 The "Change Parameters" screen is displayed.
50
CHART FUNCTIONS UNIQUE TO THE MC-4S
7 Touch the [Aux. lens] button to display the auxiliary lens list. Select the polarized lenses whose
shape is like " ".
8 Touch the [OK] button to close the "Change Parameters" screen. 9 After registering all of the user customizable contents icons, touch the [OK] button to exit from the
"Settings" screen and return to the test screen.
10 After returning to the test screen, make sure that the registered user customizable contents icons
are added in the chart page. Then, make sure that the still images, movies and slideshows are displayed correctly.
Display/playback of the user customizable contents
It is necessary to connect the USB memory where the user customizable con-
NOTES
tents are saved to the USB port of the CV-5000 system before starting this sys­tem.
1 Select the icon of the user customizable contents to be displayed and played from the chart
page.
The selected image of the user customizable contents is displayed and played.
NOTES
Select any other chart. The playback of the user customizable contents is fin­ished and the selected chart is indicated.
51
CHART FUNCTIONS UNIQUE TO THE MC-4S

SETTING THE VISUAL ACUITY CHART

You can set a variety of operations such as the installing distance of the MC-4S, the display status of mask, the display interval of a slideshow, etc. Set the visual acuity chart through the setting screen of the CV system.
Displaying the setting screen
Display the setting screen of the MC-4S by the following procedure:
1 Press the [Shift] key and [Menu] key of the KB-50S at the same time. 2 The "Settings" menu is displayed.
Touch the [Chart settings] button on the "Settings" menu.
3 The "Chart settings" screen appears.
Make sure that "MC-4S" is set for "Device" and touch the [Chart software config.] button.
52
CHART FUNCTIONS UNIQUE TO THE MC-4S
4 The "SETTING" menu of the MC-4S appears.
Chart settings
Touch the [Chart settings] button on the "SETTING" menu of the MC-4S, and the "CHART SETTINGS" screen appears. You can set the following items on the "CHART SETTINGS" screen:
1 Examination distance
Set the installing distance of the MC-4S. Select one of 60cm, 80cm, 95cm, 1m and 1.5m* according to the distance from the MC-4S instrument's front surface to the patient's eye. * You can select this only when selecting the H1 type.
2 Slideshow interval
Set the interval for changing the still image in the preinstalled slideshow and the "User customiz­able contents" slideshow. You can set the interval within the range of 3 to 10 seconds at intervals of 1 second.
CHART FUNCTIONS UNIQUE TO THE MC-4S
53
3 Size of VA indicator
Set the size of the visual acuity value that is displayed at the rightmost side of the chart. Select from "Small", "Large" and "Disable". When you select "Disable", the visual acuity value is not displayed on the MC-4S screen.
4 Masked chart layout
Set the chart display position when setting "Vertical mask", "Horizontal mask" and "Character mask" that are valid for the visual acuity chart. Select "Fixed position", and part of the chart is displayed without changing the chart's display position when setting the mask. Select "Center position", and part of the chart, which is displayed when setting the mask, is dis­played at the center of the screen.
5 Language
The language, which is set currently on the MC-4S, is displayed. Language setting is linked with the setting of the CV-5000 system. You cannot change the lan­guage setting here.
NOTES
Color adjustment
Touch the [Color adjustment] button on the "SETTING" menu of the MC-4S, and the "COLOR ADJUSTMENT" screen appears. You can adjust the chart color on the "COLOR ADJUSTMENT" screen.
Background, R&G chart
Touch the [Page 1] button among the function buttons, and this adjustment screen appears. You can adjust the background color of the chart having the normal white background and the red and green of the Red/Green chart. When this screen appears, the display to check the adjusted color appears on the MC-4S screen.
Touch the [Default] button among the function buttons. The set data for the chart is returned to the setting when shipped.
54
CHART FUNCTIONS UNIQUE TO THE MC-4S
The adjusting procedure is shown below.
1 Select the object of adjustment from [Color to adjust]. You can select from "Background", "Red"
and "Green".
2 Adjust the color selected on [Color to adjust] by changing [R], [G] and [B] within the range of 0 to
255. Set "0", and the color ingredient is zero. Select "255", and the color ingredient is maximum.
3 When changing the value of [R], [G] and [B], the change is effectuated in the color of the MC-4S
screen at the same time. Adjust the object color while watching the screen.
4 After adjusting the color, select the unadjusted item from [Color to adjust] and carry out Step 2
and 3.
NOTES
Binocular test chart (Chart on black)
Touch the [Page 2] button among the function buttons, and this adjustment screen appears. You can adjust the "Red", "Green", "Background" and "Crossed area" (the place where the red crosses the green) of the chart that separates fusion by using the red/green lens (e.g. the cross ring chart or Worth 4 Dots chart). When this screen appears, the display to check the adjusted color appears on the MC-4S screen.
Touch the [Default] button among the function buttons. The set data for the chart is returned to the setting when shipped.
The adjusting procedure is shown below.
1 Set your red/green eyeglasses or the red/green lens on the tentative frame. Adjust the color
while watching the MC-4S screen.
2 Select the object of adjustment from [Color to adjust]. You can select from "Red", "Green",
"Background" and "Crossed area".
CHART FUNCTIONS UNIQUE TO THE MC-4S
55
3 Adjust the color selected on [Color to adjust] by changing [R], [G] and [B] within the range of 0 to
255. Set "0", and the color ingredient is zero. Select "255", and the color ingredient is maximum.
4 When changing the value of [R], [G] and [B], the change is effectuated in the color of the MC-4S
screen at the same time. Adjust the object color while watching the screen.
The following table shows the points for the adjustment.
Adjusted color Right eye (red lens) Left eye (green lens)
Red "Red" can be recognized as red.
The "Crossed area" can be assimi­lated with "Red" and recognized as red.
Green "Green" is assimilated with the back-
ground color on the screen.
Background The background color is assimilated
with "Green" on the screen.
Crossed area The "Crossed area" can be assimi-
lated with "Red" and recognized as red.
"Red" is assimilated with the back­ground color on the screen.
"Green" can be recognized as green. The "Crossed area" can be assimi­lated with "Green" and recognized as green.
The background color is assimilated with "Red" on the screen.
The "Crossed area" can be assimi­lated with "Green" and recognized as green.
5 After adjusting the color, select the unadjusted item from [Color to adjust] and carry out Step 2
and 3.
NOTES
Touch the [Default] button among the function buttons. The set data for the chart is returned to the setting when shipped.
56
CHART FUNCTIONS UNIQUE TO THE MC-4S

MAINTENANCE

To avoid fire or electric shock, turn off the power of the instrument and
WARNINGS
Dust is a formidable foe to the instrument. Cover the machine when not in use. When not in use, turn off the power switch. When the AC adapter is not in use for a long period of time, disconnect the power plug.

DAILY CHECKUPS

Inspection at start
remove the power plug from the outlet before inspecting or cleaning the instrument.
CAUTIONS
You can change the test distance with a setting in the instrument. Ensure that the correct distance is set prior to performing the V. A. test so you do not perform measurement with a different test distance. For the setting of the instrument, refer to "SETTING THE INSTALLING DISTANCE" in this manual.
Cleaning the chart window
• Do not use organic solvent such as thinner. Do not leave the instrument with
NOTES
• Do not apply force when removing the dust on the chart window. Cleaning the
When the chart window is stained, wipe it lightly with an absorbent cotton or a clean cloth. When the chart window is badly stained, prepare a solution of diluted neutral detergent and moisten a soft cloth with it. Wring the cloth thoroughly and wipe the chart window with the cloth. After the chart window has dried, wipe it with a dry soft cloth.
To prevent the reliability of the measurement result from being lowered, set the test distance of the instrument correctly.
water drops on its surface as it is. Keep the chart window surface dry from moisture. The chart window may be deteriorated or damaged. If any liquid enters the instrument, it may malfunction.
chart window with tissue paper may cause the chart window to be damaged.
57
MAINTENANCE
Cleaning the monitor
When there is dust on the monitor, open the maintenance cap and remove dust with a commercial blower or the like.

USER MAINTENANCE ITEMS

Item Inspection time Contents
Inspection Before using • The instrument must operate normally.
• The chart window must be free of stain or damage.
• The screw for the support metal fixture must not be loos­ened.
Cleaning When the part is stained • Chart window
• Instrument body, switches, etc.

REPLACING THE BATTERY

Replacing the battery of alignment emitter
• When replacing the battery, use the AAA size dry battery.
NOTES
• When replacing the battery, fit a new battery to the battery direction indicated on the battery case bottom and then load it correctly.
• When replacing the battery, do not use the old and new batteries together.
1 Remove the battery cover on the rear surface of the instrument. For removing the battery cover,
refer to P.19.
2 Remove the old battery. 3 Load a new battery. 4 Attach the battery cover to the original position.
58
MAINTENANCE

BEFORE REQUESTING SERVICE

TROUBLE SHOOTING GUIDE

To avoid electric shock, do not remove the cover. Contact your local
WARNINGS
Refer to the following check list when you encounter a problem. If, after following the instructions below, you still cannot restore the instrument to a normal condition or if the problem does not fall into any of the categories below, contact your dealer or your local TOPCON office (listed on the the back cover).
Problem Condition Remedial measure Page
authorized dealer or your local Topcon office (listed on the back cover) to repair the instrument.
Check list
The chart window sud­denly turns off.
The pilot lamp is not lit in green.
Error message Remedial measure
"Fail to execute chart software. (Error Code
601)" or "Fail to change the resolution of the selected chart monitor. (Error Code 1)" or "Fail to change the resolution of the selected chart monitor. (Error Code 2)"
The image signal is not input­ted from the CV-5000.
If the pilot lamp is lit in orange, the power switch is not turned on.
If the pilot lamp is OFF, the power cord or AC adapter is not connected to the instru­ment.
MC-4S software message and remedial measure
Check the connection with the CV-5000.
Check the power switch and the power cord or AC adapter connection.
This message is displayed when you cannot change the resolution of the MC-4S. When using the CV system with a KB-50S or the CV system with a mouse, restart the CV system. When using the MC-4S with an external personal com­puter, set the resolution to "640×480" and the color to "32 bits" for the MC-4S on the personal computer con­nected to the MC-4S in advance. While the screen is ON, start the CV software. If the instrument does not reset, ask the TOPCON office to repair it.
14
16
“Fail to change the resolution of the selected chart monitor. (Error Code 3)”
This message is displayed when using the MC-4S with an external personal computer and restarting is neces­sary because the resolution of the MC-4S is changed. Restart the personal computer to which the MC-4S is connected. Then, start the CV software. If the instrument does not reset, ask the TOPCON office to repair it.
59
BEFORE REQUESTING SERVICE
Error message Remedial measure
"Fail to execute chart software. (Error Code
602)”
"Could NOT find the selected chart monitor. Please confirm the chart monitor setting and the connection." or "Fail to execute chart software. (Error Code
603)"
"Fail to execute chart software. (Error Code
604)"
"Fail to execute chart software. (Error Code
605)"
This message is displayed when the MC-4S starts and the necessary files cannot be accessed. Try the follow­ing items. Restart the CV system. Reinstall the software. If the instrument does not reset, ask the TOPCON office to repair it.
It is probable that the monitor cable is not connected or is connected improperly. Connect the cable and restart the CV system. It is probable that the MC-4S is connected through a VGA cable. Use the DVI cable to connect with the MC­4S. After connecting, restart the CV system. It is probable that the connected monitor is not the MC­4S. Make sure that the MC-4S is connected as the monitor. Make sure that "MC-4S" is set in "Chart moni­tor selection". After connecting, restart the CV system.
It is probable that any other chart except the MC-4S is set in "Chart monitor selection" of "Chart settings". Set "MC-4S" in "Chart monitor selection" of "Chart set­tings".
This message is displayed in the following case: When the CV system starts the MC-4S system, it specifies the chart type. The MC-4S cannot recognize this setting. Restart the CV system. If the instrument does not reset, select a chart type again or reinstall the software.
"Fail to execute chart software. (Error Code
606)"
"Fail to open chart software configuration screen because the other task is in progress. Please try again."
"Fail to change chart type because the other task is in progress. Please try again."
"Fail to change chart language setting because the other task is in progress."
"Fail to open chart software configuration screen. (Error Code 300)"
"Fail to open chart software configuration screen. (Error Code 301)"
This message is displayed when the setting file of the MC-4S cannot be accessed. Restart the CV system. If the instrument does not reset, ask the TOPCON office to repair it.
This message is displayed when the command to open the MC-4S setting screen cannot be executed because the MC-4S is executing another task. Access the setting screen again.
This message is displayed when the command to change the type cannot be executed because the MC­4S is executing another task. Change the type again.
This message is displayed when the command to change the language cannot be executed because the MC-4S is executing another task. Change the language again.
Restart the CV system. If the instrument does not reset, ask the TOPCON office to repair it.
This message is displayed when the setting file of the MC-4S cannot be accessed. Restart the CV system. If the instrument does not reset, ask the TOPCON office to repair it.
60
BEFORE REQUESTING SERVICE
Error message Remedial measure
"Fail to execute chart software with selected chart type. Please select different chart type. (Error Code 401)"
"IDS_CHART_LANG_TYPE_ERR (Error Code 501)"
"Fail to save chart software configurations. The changed parameters are canceled. (Error Code 701)"
* When the errors occur frequently, contact your dealer or your local TOPCON office (listed on the back
cover).
This message is displayed when the MC-4S cannot rec­ognize the chart type changed by the setting. Restart the CV system. If the instrument does not reset, select a chart type again or reinstall the software.
This message is displayed when the language that is not usable in the MC-4S software is set in the CV soft­ware. The MC-4S software starts in the English mode. Use the software as it is.
This message is displayed when the setting file of the MC-4S cannot be accessed after exiting the setting. Restart the CV system and change the setting again. When the instrument is not reset, ask your local TOP­CON office to repair it.
61
BEFORE REQUESTING SERVICE

SPECIFICATIONS & PERFORMANCE

SPECIFICATIONS BY ITEMS

Test distance : 5.0m Test chart : Visual acuity test chart for visual acuity 0.03 - 2.0 and visual function test chart Chart display : Charts of one horizontal line, one vertical line, one character, R&G, reversed
B&W are displayed. Contrast is changed. Polarization display is done.
Filter : Polarized filter Luminance : 200 ±20cd/m Contrast : 85% or more Installing distance : 60cm, 80cm, 95cm, 1m or 1.5m (from the vertex of the patient to the chart win-
dow of the instrument)

OTHER SPECIFICATIONS

Display resolution : VGA (640×480 pixels) External input/output port : DVI connector
2
* Specifications and appearance are subject to change without advance notice for product improve-
ment.
* Sometimes there is a missing dot or lighting dot on the screen of the chart window. Its cause is not a
problem with the instrument but the characteristics of the liquid crystal display. So, it has nothing to do with repair or replacement.
Essential performance
NOTES
• The screen display is not to be disturbed.
• The instrument must not reboot.
62
SPECIFICATIONS & PERFORMANCE
GENERAL INFORMATION ON USAGE AND MAINTE­NANCE

PURPOSE OF USE/EFFICACY OR EFFECT

This instrument displays the visual acuity chart data outputted by the following two units and is used as the visual acuity chart in the visual function test.
• CV POWER SUPPLY UNIT WITH BUILT-IN PERSONAL COMPUTER, which is a component of the Compu-Vision CV-5000
• The external personal computer connected to CV POWER SUPPLY UNIT

ENVIRONMENTAL CONDITIONS FOR USE

Temperature : 10°C - 40°C Humidity : 30% - 90% (non-condensing) Air pressure : 700hPa - 1060hPa

STORING METHOD, USAGE PERIOD AND OTHERS

1. Storage (without wrapping (without package))
* Temperature: 10°C - 40°C Humidity: 10% - 95% (without dew condensation) Air pressure: 700hPa - 1060hPa * THIS INSTRUMENT DOES NOT MEET THE TEMPERATURE REQUIREMENTS OF ISO 15004-
1 FOR STORAGE. DO NOT STORE THIS INSTRUMENT IN CONDITIONS WHERE THE TEM­PERATURE MAY RISE ABOVE 40°C OR FALL BELOW 10°C.
2. Storage (with wrapping (with package))
Temperature: -20°C - 50°C Humidity: 10% - 95%
3. Transportation (with wrapping (with package))
Temperature: -40°C - 70°C Humidity: 10% - 95%
4. When storing the instrument, ensure that the following conditions are met:
(1) The instrument should not be splashed with water. (2) Store the instrument away from an environment where air pressure, temperature, humidity, ven-
tilation, sunlight, dust, salty/sulfurous air, etc. could cause damage.
(3) Do not store or transport the instrument on a slope or uneven surface or in an area where it is
subject to vibrations or instability.
(4) Do not store the instrument where chemicals are stored or gas is generated.
5. Usage period
8 years from delivery providing regular maintenance is performed (according to the self-certification [Topcon data])
63

GENERAL INFORMATION ON USAGE AND MAINTENANCE

ELECTRIC RATING

AC adapter rating : BPM030S12F04 (for USA and Canada)
Input : 100-240V, 50-60Hz, 1.0A Output : 12V, 2.5A
: JMW160KA1200F09 (outside of USA and Canada)
Input : 100-240V, 50-60Hz, 1.5A
Output : 12V, 5.0A Power supply voltage : AC 100-240V Frequency : 50-60Hz

DIMENSIONS AND WEIGHT

Instrument body
Dimensions: 440mm (W) × 339mm (D) × 1330mm (H) Weight: 19.5kg

SYSTEM CLASSIFICATION

Type of protection against electric shocks: This instrument is classified as Class I equipment. (MC-4S Mirror Chart)/Internal power supply device (Alignment emitter)
Class I equipment does not depend only on basic insulation for protection against electric shocks, but also provides a means of connection to a protective earth system of facilities so that metal parts that come into contact do not become conductive while the basic insulation is in failure. "Internal power supply device" is a device that can be operated by the internal power supply.
Degree of protection against harmful ingress of water: IPx0 (MC-4S Mirror Chart and alignment emitter)
The MC-4S has no protection against ingress of water. (The degree of protection against harmful ingress of water defined in IEC 60529 is IPx0.)
Classification according to the method(s) of sterilization or disinfection recommended by the manufacturer: not applicable. (MC-4S Mirror Chart and alignment emitter)
The MC-4S has no part to be sterilized or be disinfected.
Classification according to the degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide: Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. (MC-4S Mirror Chart and alignment emitter)
The MC-4S should be used in environments where no flammable anesthetics and/or flammable gases are present.
Classification according to the mode of operation: Continuous operation. (MC-4S Mirror Chart and alignment emitter)
Continuous operation is the operation under normal load for an unlimited period, without the specified lim­its of temperature being exceeded.
64
GENERAL INFORMATION ON USAGE AND MAINTENANCE

OPERATION PRINCIPLE

This symbol is applicable for EU member countries only. To avoid potential damage to the environment and possibly human health, this instrument should be disposed of (i) for EU member countries - in accordance with WEEE (Directive on Waste Electrical and Electronic Equipment), or (ii) for all other countries, in accordance with local disposal and recycling laws.
Receives the test chart data and indicates it on the liquid crystal display. By using the mirror, the patient can check the test chart data visually with the size of the regular distance in front of him/her.

CHECKPOINTS FOR MAINTENANCE

Maintenance by user
1. Regularly maintain and check the instrument and its parts.
2. Make sure that the test distance of the instrument is set correctly.
3. Make sure that the instrument is installed in a stable place.
4. When using the instrument after a prolonged period of inactivity, confirm normal and safe operation
beforehand.
5. If the chart window is stained, clean it.
6. When this instrument is not in use, apply the dust cover to it. For details, refer to "PREPARATION BEFORE USE" and "MAINTENANCE."

DISPOSAL

Dispose of the instrument according to local disposal and recycling laws.
NOTES
GENERAL INFORMATION ON USAGE AND MAINTENANCE
65

PATIENT’S ENVIRONMENT

Radius 1.5m
1.5m1.5m
2.5m
Device applicable to the use in
• USB memory
Do not use a power strip in the patient's environment. Connect the power supply of the device to the commercial power supply.
patient’s environment
When the patient or inspector may touch the devices (including the connecting devices) or when the patient or inspector may touch the person in contact with the devices (including the connecting devices), the patient's environment is shown below. In the patient's environment, use a device conforming to IEC60601-1. If you are compelled to use any device not conforming to IEC60601-1, use an insulation transformer or the common protective earth system.

REQUIREMENTS FOR THE EXTERNAL DEVICE

The external device connected to the analog and digital interfaces must comply with the respective IEC or ISO standards (e.g. IEC 60950 for data processing equipment and IEC 60601-1 for medical equip­ment). Anybody connecting additional equipment to medical electrical equipment configures a medical system and is therefore responsible that the system complies with the requirements for medical electrical sys­tems. Attention is drawn to the fact that local laws take priority over the above mentioned requirements. If in doubt, contact your dealer or TOPCON(see the back cover).
66
GENERAL INFORMATION ON USAGE AND MAINTENANCE

ELECTROMAGNETIC COMPATIBILITY

This product conforms to the EMC standard (IEC 60601-1-2 Ed.3.0: 2007).
a) MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in the ACCOMPANYING
DOCUMENTS. b) Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT. c) The use of ACCESSORIES, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replacement
parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the
EQUIPMENT or SYSTEM. d) The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the EQUIPMENT or SYSTEM should be observed to verify nor-
mal operation in the configuration in which it will be used. e) The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other than those
specified may result in increased EMISSION or decreased IMMUNITY of the EQUIPMENT or SYS-
TEM.
Item Article code Cable Shielded Ferrite Core Length (m)
Image cable 42496 9012 YES YES 3.0
AC adapter 42428 9001 NO YES 1.5
AC adapter 42428 9011 NO YES 1.5
Power cord 42412 2090 NO NO 3.0
Guidance and manufacturer's declaration - electromagnetic emissions
The MC-4S is intended for use in the electromagnetic environment specified below. The customer or the user of the MC-4S should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC61000-3-2
Voltage fluctuations/ flicker emissions IEC61000-3-3
Group 1
Class B
Class A
Complies
The MC-4S uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The MC-4S is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
67
GENERAL INFORMATION ON USAGE AND MAINTENANCE
Guidance and manufacturer's declaration - electromagnetic immunity
The MC-4S is intended for use in the electromagnetic environment specified below. The customer or the user of the MC-4S should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient/burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and Voltage variations on power supply input lines IEC 61000-4-11
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line (s) to line (s)
± 2 kV line (s) to earth
<5% U
t
(>95% dip in Ut) for 0.5 cycle 40% U
t
(60% dip in Ut) for 5 cycles 70% U
t
(30% dip in Ut) for 25 cycles <5% U
t
(>95% dip in Ut) for 5 sec
Compliance
level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line (s) to line (s)
± 2 kV line (s) to earth
<5% U
t
(>95% dip in Ut) for 0.5 cycle 40% U
t
(60% dip in Ut) for 5 cycles 70% U
t
(30% dip in Ut) for 25 cycles <5% U
t
(>95% dip in Ut) for 5 sec
Electromagnetic environment -
guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user or the MC­4S requires continued operation during power mains interruptions, it is recommended that the MC-4S be powered from an uninterruptible power supply or battery.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
t
3 A/m 3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical com­mercial or hospital environment.
68
GENERAL INFORMATION ON USAGE AND MAINTENANCE
Guidance and manufacturer's declaration - electromagnetic immunity
P
P
P
The MC-4S is intended for use in the electromagnetic environment specified below. The customer or the user of the MC-4S should assure that it is used in such an environment.
Immunity test
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
IEC 60601
test level
3 Vrms 150kHz to 80MHz
3 V/m 80MHz to 2.5GHz
Compliance
level
3 V
3 V/m
Electromagnetic environment -
guidance
Portable and mobile RF communications equipment should be used no closer to any part of the MC-4S, including cables, than the recommended separation distance cal­culated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2
d = 1.2 80MHz to 800MHz d = 2.3 800MHz to 2.5GHz
where P is the maximum output power rat­ing of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, ance level in each frequency range.
a
should be less than the compli-
b
Interference may occur in the vicinity of equipment marked with the following sym­bol:
NOTE 1 NOTE 2
At 80 MHz and 800 MHz, the higher frequency range applies. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
abField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be pre­dicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF trans­mitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the MC-4S is used exceeds the applicable RF compliance level above, the MC-4S should be observed to verify normal operation. If abnormal performance is observed, additional mea­sures may be necessary, such as reorienting or relocating the MC-4S.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
69
GENERAL INFORMATION ON USAGE AND MAINTENANCE
Recommended separation distance between
PPP
portable and mobile RF communications equipment and the MC-4S
The MC-4S is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the MC-4S can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the MC-4S as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum output
power of transmitter
W
0.01 0.12 0.12 0.23
0.1 0.380.380.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation dis­tance d in metres (m) can be estimated using the equation applicable to the frequency of the transmit­ter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
150kHz to 80MHz
d = 1.2
80MHz to 800MHz
m
d = 1.2
800MHz to 2,5GHz
d = 2.3
NOTE 1
NOTE 2
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
70
GENERAL INFORMATION ON USAGE AND MAINTENANCE

REFERENCE MATERIAL

SHAPE OF PLUG

Country Voltage/frequency Shape of plug
Mexico 110V/50Hz Type C&E
Argentina 220V/60Hz Type A
Peru 220V/60Hz Type A
Venezuela 110V/50Hz Type C&E
Bolivia & Paraguay 220V/60Hz Type A (Most common)
Type H (Infrequently)
Chile 220V/60Hz Type A
Colombia 110V/50Hz Type C
Brazil 220V/60Hz
127V/60Hz
Ecuador 110V/50Hz Type C&E
USA 120V/60Hz Type A (Hospital Grade)
Canada 120V/60Hz Type A (Hospital Grade)
Type A Type C
71
REFERENCE MATERIAL
Please provide the following information when contacting us regarding questions about this Prod-
5
uct:
Model name: MC-4S
Serial No.: This is described on the rating nameplate on the top surface of the
instrument.
Period of use: Please inform us of the date of purchase.
State of instrument: Please provide us with as much detail as possible on the problem.
Mirror Chart MC-4S
USER MANUAL Version of 2014 (1403-100TH ) Date of issue: March 31, 2014
Published by TOPCON CORPORATION 75-1 Hasunuma-cho, Itabashi-ku, Tokyo, 174-8580 Japan
©2014 TOPCON CORPORATION
ALL RIGHTS RESERVED
Mirror Chart
5
MC-4S
42496 90215
Printed in Japan 1403-100TH
Loading...